Obtaining peripheral blood mononuclear cells (PBMCs) from blood samples with traditional methods, such as density gradient centrifugation, can be time consuming and difficult to automate. In addition, isolating cells from older blood samples can result in red blood cell (RBC) and granulocyte contamination, requiring additional RBC lysis and centrifugations.
Why Use EasySep™ Direct Human PBMC Isolation Kit?
FAST. Isolate human PBMCs in as little as 20 minutes without the need for lysis or centrifugation.
EASY. Avoid laborious and time consuming methods such as density gradient centrifugation and RBC lysis.
HIGH PURITY. Obtain highly purified mononuclear cells, even on older blood samples.
CONVENIENT. Automate blood sample processing with RoboSep™ instruments to increase laboratory throughput.
FLEXIBLE. Obtain cells from whole blood, cord blood, bone marrow, buffy coats, leukapheresis products, and leukocyte reduction system (LRS) cones.
How Does It Work?
Pour off sample into a new tube, add RapidSpheres™ and mix.
Add EasySep™ Direct Human PBMC Isolation Cocktail and mix well.
Add recommended medium to top up sample. Add RapidSpheres™ to sample and mix.
Place the tube into the magnet and incubate for 5 minutes*.
Incubate for 5 minutes
*Times will vary depending on the EasySep™ magnet platform that is used.
Place the tube into the magnet and incubate for 5 minutes*.
Pour off sample into a new tube.
Repeat steps 5 and 6. Cells are now ready for downstream applications.
1 2 3 4 5 6
EasySep TM
EasySep TM
EasySep TM
EasySep TM
EasySep TM
EasySep TM
EasySep TM
EasySep TM
Sample Protocol Video: www.stemcell.com/easysepdirectvideo
Using EasySep™ Direct Human PBMC Isolation Kit (Catalog #19654), you can immunomagnetically obtain PBMCs without lysis or centrifugation in as little as 20 minutes. The resulting cells are highly purified with no RBC contamination, even on older samples (Figure 1 and 2).
To meet all of your laboratory needs, the EasySep™ Direct Human PBMC Isolation Kit can be used on:
• Whole blood• Cord blood• Bone marrow• Buffy coats• Leukapheresis products• Leukocyte reduction system (LRS) cones
Increase sample processing throughput by automating your PBMC isolation with RoboSep™ instruments. Automation can minimize sample handling and free up valuable hands-on time (Table 1, page 3).
2 STEMCELL Technologies Inc.
ISOLATE MONONUCLEAR CELLS DIRECTLY FROM WHOLE BLOOD
PBMC Isolation with EasySep™ DirectSkip centrifugation and isolate PBMC immunomagnetically directly from whole blood with EasySep™ Direct Human PBMC Isolation Kit. With immunomagnetic cell isolation PBMCs are separated based on cell surface markers rather than cell density, resulting in highly purified PBMCs and fewer contaminating cells, even when processing older blood samples (Figure 2C).
Figure 1. Typical EasySep™ Direct Human PBMC Isolation Profile
(A) Starting with human whole blood from normal healthy donors, the typical mononuclear cell content of the non-lysed final isolated fraction is 98.3 ± 2.8% (gated on CD45). In the above example, the mononuclear cell content of the whole blood start sample (lysed by ammonium chloride) and non-lysed final isolated fraction is 27.0% and 98.6% (not gated on CD45), respectively. (B and C) Mononuclear cells were isolated from whole blood samples using either density gradient centrifugation or EasySep™ Direct Human PBMC Isolation Kit. (B) Representative FSC vs. SSC flow cytometry plots. (C) Representative t-SNE plots of PBMCs stained with 19 markers and analyzed with mass cytometry (CyTOF). Cells are clustered and colored based on the combination of markers they express. Both density gradient centrifugation and EasySep™ Direct Human PBMC Isolation Kit resulted in comparable cell populations with the exception of the contaminating granulocyte population (CD16+ CD15+ CD11b+ CD66b+ CD45low).
Figure 2. EasySep™ Direct Human PBMC Isolation Kit Results in Fewer Contaminating Cells Compared to Density Gradient Centrifugation
Mononuclear cells were isolated from whole blood samples using either density gradient centrifugation or EasySep™ Direct Human PBMC Isolation Kit. Cells were counted and analyzed by flow cytometry. (A) Both density gradient centrifugation and EasySep™ Direct Human PBMC Isolation Kit resulted in an equivalent total number of nucleated cells recovered from 24-hour-old blood samples (mean ± SD; n=14). (B) Using EasySep™ Direct Human PBMC Isolation Kit to obtain mononuclear cells from 24-hour-blood samples resulted in lower numbers of residual platelets (CD41+), red blood cells (Glycophorin A+/CD45−), and granulocytes (CD66b+) compared to density gradient centrifugation (mean ± SD; n=15). (C) Cell isolation from 24-, 48-, 72-, and 96-hour-old blood samples using EasySep™ Direct Human PBMC Isolation Kit resulted in lower numbers of residual granulocytes compared to cell isolation using density gradient centrifugation (mean ± SD; n=3).
Figure 3. PBMCs Isolated with EasySep™ Direct Human PBMC Isolation Kit Proliferate and Maintain High RNA Integrity
Mononuclear cells were isolated from whole blood samples using either density gradient centrifugation or EasySep™ Direct Human PBMC Isolation Kit. (A) Isolated mononuclear cells were used for downstream RNA isolation. There was no significant difference in RNA integrity as measured with the Agilent RNA Bioanalyzer (mean ± SEM, n=3). (B) Isolated mononuclear cells were labeled with Proliferation Dye eFluor 450 and stimulated with ImmunoCult™ Human CD3/CD28 T Cell Activator and 0.5 ng/ml IL-2. After 4 days in culture, cells were analyzed for proliferation by flow cytometry. Representative histogram showing dividing cells (eFluor 450low).
105
104
103
102
-102
-102 103 104 105
Gly
A P
E
CD41 FITC
Start: No Gate
105
104
103
102
-102
-102 103 104 105
Gly
A P
E
CD41 FITC
Isolated: No Gate
SSC
262144
196608
131072
65536
00 6553665536
262144
196608
131072
65536
0131072131072 196608196608 262144262144
0
FSCFSC
Isolated: EasySep™ Direct Human PBMC Isolation Kit (No Gate)
SSC
Isolated: Density Gradient Centrifugation (No Gate)
A A
A C
Total Number of Cells B B
B
Residual Cells C Residual Granulocytes
RNA Integrity
RN
A In
teg
rity
Nu
mb
er
Density Gradient Centrifugation
EasySep™ Direct Human PBMC Isolation Kit
10.0
9.5
9.0
8.5
EasySep™ Direct Human PBMC Isolation Kit
Co
un
t
eFluor 450
1000
750
500
250
-102 103 104 105 106 107
0
97.69%
Functional Data – Whole BloodT Cell Proliferation Data
Frann AntignanoMelina Messing
NK cells - CD45+ CD16+ CD56+ CD8+ CD11blow
B cells - CD45+ CD19+ HLA-DR+
CD8 T cells - CD45+ CD3+ CD8+ CD25low
Antigen presenting cells - CD45+ CD19+
HLA-DR+ CD16+ CD123low CD4low
T cell - CD45+ CD3+
CD4 T cells - CD45+ CD3+ CD4+
Hematopoietic progenitor subtype - CD45+ CD38+ CD123+ CD294+
Density Gradient Centrifugation
tSNE1
tSN
E2
Functional Data – Whole BloodT Cell Proliferation Data
Frann AntignanoMelina Messing
Hematopoietic progenitor subtype - CD34+
CD117+ CD294+
Granulocytes - CD16+ CD15+ CD11b+
CD66b+ CD45low
Monocytes - CD45+ CD11c+ CD14+ CD11b+
HLA-DR+ CD38 CD4low
CD45-
CD8 T cell subset - CD45+ CD3+ CD8+
CD294+ CD25low
tSNE1
tSN
E2
EasySep™ Direct Human PBMC Isolation Kit
Tota
l Nu
mb
er o
f C
ells
R
eco
vere
d p
er m
L B
loo
d
% R
esid
ual
Cel
ls
% R
esid
ual
Gra
nu
locy
tes
Timepoint
4x106 100
Density Gradient Centrifugation
EasySep™ Direct Human PBMC Isolation Kit
80
80
60
60
4040
20 20
0 0
3x106
2x106
1x106
0
PlateletsRBCs
Granulocytes 24h 48h 72h 96h
3WWW.STEMCELL.COM
Automating PBMC Isolation with RoboSep™
Minimize sample handling and free up valuable hands-on time by automating PBMC isolation with the EasySep™ Direct Human PBMC Isolation Kit and the RoboSep™-S instrument. By automating all sample labeling and magnetic separation steps you can perform simultaneous cell isolations from up to 4 samples and increase sample throughput.
Select desired protocol and load samples, EasySep™ reagents, buffer, and tips.
1
Return in 30 minutes to collect your separated cells.3
Press “Run”.2
How Does It Work?
Step Duration (minutes)
Manual Density Gradient
Centrifugation
ManualEasySep™ Direct
Automated EasySep™ Directon RoboSep™-S
Preparation of tubes with density gradient medium and layering sample
7 - -
Loading samples and reagentsonto the RoboSep™-S instrument
- - 5
Cell isolation 45 20 30
Cell collection 10 1 1
Optional: Wash step toremove platelets
15 - -
Optional: Red blood cell lysis 15 - -
Final centrifugation step 15 Optional Optional
Total time for cell separation ≥ 77 Minutes ~ 21 Minutes ~ 36 Minutes
Total hands-on time ≥ 32 Minutes ~ 21 Minutes ~ 6 Minutes
Table 1. Automated PBMC Isolation Using RoboSep™-S Reduces the Overall Cell Separation Time Compared to Density Gradient Centrifugation
Why Automate PBMC Isolation with RoboSep™?
EFFICIENT. Minimize hands-on time and increase sample throughput.
HIGH PURITY. Obtain highly purified mononuclear cells with improved RBC, granulocyte, and platelet clearance compared to density gradient centrifugation.
CONVENIENT. Load your samples and reagents, press “Run”, and come back to your isolated cells.
FLEXIBLE. Obtain cells from cord blood, bone marrow, buffy coats, leukapheresis products, and leukocyte reduction system (LRS) cones.
Steps and approximate times to isolate monononuclear cells from whole blood samples when following using density gradient centrifugation or EasySep™ Direct Human PBMC Isolation Kit (manual separation or automated on the RoboSep™-S). For density gradient centrifugation, additional washing steps and red blood cell lysis might be required.
ISOLATE MONONUCLEAR CELLS DIRECTLY FROM WHOLE BLOOD
STEMCELL TECHNOLOGIES INC.’S QUALITY MANAGEMENT SYSTEM IS CERTIFIED TO ISO 13485. PRODUCTS ARE FOR RESEARCH USE ONLY AND NOT INTENDED FOR HUMAN OR ANIMAL DIAGNOSTIC OR THERAPEUTIC USES UNLESS OTHERWISE STATED.
TOLL FREE PHONE 1 800 667 0322 • PHONE +1 604 877 0713 • [email protected] • [email protected]
FOR GLOBAL CONTACT DETAILS VISIT WWW.STEMCELL.COM DOCUMENT #27156 VERSION 2.0.0 MARCH 2020
To see RoboSep™ instruments in action, visit www.RoboSep.com.
Copyright © 2020 by STEMCELL Technologies Inc. All rights reserved including graphics and images. STEMCELL Technologies & Design, STEMCELL Shield Design, Scientists Helping Scientists, EasySep, and RoboSep
are trademarks of STEMCELL Technologies Canada Inc. Lymphoprep is a registered trademark of Alere Technologies. All other trademarks are the property of their respective holders. While STEMCELL has made all
reasonable efforts to ensure that the information provided by STEMCELL and its suppliers is correct, it makes no warranties or representations as to the accuracy or completeness of such information.
A B C
Tota
l Nu
mb
er o
f M
on
on
ucl
ear
Cel
ls R
eco
vere
d p
er m
L
Tota
l Nu
mb
er o
f M
on
ocy
tes
and
Lym
ph
ocy
tes
Rec
ove
red
per
mL
% R
esid
ual
Cel
ls
4x108 4x108 80 Density Gradient Centrifugation
EasySep™ Direct on RoboSep™-S
3x108 3x108 60
2x108 2x108 40
1x108 1x108 20
0 0 0
Monocytes
Platelets
LymphocytesRBCs
Granulocytes
Figure 4. PBMC Isolation from a Full-Size Leukapheresis Pack Using the EasySep™ Direct Human PBMC Isolation Kit Automated with RoboSep™-S Results in Fewer Contaminating Cells Compared to Density Gradient Centrifugation
Mononuclear cells were isolated from single concentrated leukapheresis packs using either density gradient centrifugation with Lymphoprep™ density gradient medium (Density Gradient Centrifugation) or EasySep™ Direct Human PBMC Isolation Kit automated on the RoboSep™-S instrument (EasySep™ Direct on RoboSep™-S). Cells were counted and analyzed by flow cytometry. Compared to density gradient centrifugation, EasySep™ Direct Human PBMC Isolation Kit automated on the RoboSep™-S instrument resulted in (A) equivalent numbers of total mononuclear cells, (B) equivalent numbers of total monocytes and lymphocytes and (C) lower numbers of residual platelets (CD41+), red blood cells (Glycophorin A+ /CD45− ), and granulocytes (CD66b+ ) (mean ± SD n=6).
ProductBlood Volume
ProcessedCatalog #
EasySep™ Direct Human PBMC Isolation Kit
100 mL 19654
EasySep™ Direct Human PBMC Isolation Kit for RoboSep™ 100 mL 19654RF
RoboSep™ Instruments for Automated Cell Isolation
RoboSep™-S and Robosep™-16 fully automate all cell labeling and separation steps of the EasySep™ procedure, minimizing sample handling and freeing up technician time. Setup is simple: just load your samples and reagents and walk away. Return later to separated cells, which are immediately ready for downstream applications.
RoboSep™-S
Catalog #21000
RoboSep™-16
Catalog #23000
NEED FRESH WHOLE BLOOD PRODUCTS?Get fresh whole peripheral blood and leukapheresis packs when you need it and specify your donor requirements.*
Learn more at www.stemcell.com/PrimaryCells
Special Application
Automate PBMC Isolation from a Full-Size Leukapheresis Pack
Obtain PBMCs from a full-size leukapheresis pack in a single run with EasySep™ Direct Human PBMC Isolation Kit and RoboSep™-S. This automated cell isolation system results in fewer contaminating cells compared to density gradient centrifugation (Figure 4). First, concentrate the leukapheresis pack (see below). Then, load the tubes with the resuspended cells, as well as the EasySep™ reagents, buffers, and tips onto the instrument. Start the instrument’s isolation protocol and come back in 30 minutes to collect your isolated cells.
Below is an example of how to concentrate an entire leukapheresis pack for downstream PBMC isolation using RoboSep™-S. For more information please consult the product information sheet for your specific product or contact our Product & Scientific Support team at [email protected].
1. Transfer the entire contents of the leukapheresis bag into tubes
and centrifuge at 300 x g for 10 minutes with the brake on.
2. Carefully remove the supernatant and resuspend the cells in EasySep™ Buffer (Catalog #20144) to obtain a final volume of 20 mL and approximate concentration of 4x108 cells/mL.
3. Add EDTA at 6 mM final concentration and aliquot the 20 mL into four 14 mL polystyrene round-bottom tubes (e.g. Catalog #38008). Each tube will now contain 5 mL of resuspended cells.
4. Select the appropriate protocol on the screen.
5. Follow the onscreen prompts and load the 4 x 14 mL tubes, reagents, tips and buffer onto the RoboSep™-S carousel.
6. Press “Run”.
7. Come back in 30 minutes after the run is complete and collect the isolated cells.
*Certain products only available in select territories.