Fundamental Imaging Characteristics and
Resolution of Metal Alloys
By
Marc G. Apple, M.D.
Mark Michael
![Page 2: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/2.jpg)
FWM’s Mission Statement
Fort Wayne Metals was founded and continues to grow because of our dedication to the medical device field
“Our mission is to continuously improve the quality of our products, the speed and efficiency with which we respond to our customers requests, and the knowledge of the materials we work with and recommend.”
![Page 3: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/3.jpg)
Typical Wire Application Need
DiameterStrengthFatigueCorrosion ResistanceElectrical ResistanceFormability
But Now…
![Page 4: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/4.jpg)
Imaging
What size wire should I use?Which alloy will work best?Will flat wire be better than round?How can we find answers?
– What modality will you use?– What is a good image?– What technology will be used?
![Page 5: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/5.jpg)
Where did we start
Most implantable wires lose imaging capability under .004”(.1mm)
The best still perform around .001”(.025mm)
![Page 6: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/6.jpg)
The Concept
Choose wire diameters from .001” to .011” (.025mm to .28mm)
Choose common alloys and candidate alloys
Mount in stable fashionCharacterize using common
modalities
![Page 7: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/7.jpg)
Wire Diameters#100 #50 #25 #20 #15 #10 #5 #4 #3 #2 #1 .25 .18 .127 .114 .098 .080 .057 .051 .044 .036 .025.0100 .0071 .0050 .0045 .0040 .0032 .0023 .0020 .0017 .0014 .0010(100x) (50x) (25x) (20x) (15x) (10x) (5x) (4x) (3x) (2x) (1x)
90 Pt 10 Ni
![Page 8: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/8.jpg)
Candidate Alloys and Materials
MP35NCommercially Pure TitaniumTantalumNitinolPlatinum-10 NiDBS(7)MP-DFT® 28AgConi-DFT® 33Pt
![Page 9: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/9.jpg)
Candidate Alloys and Materials
304V316LVMGoldPlatinumL605Custom 455
![Page 10: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/10.jpg)
Selecting Modalities
Dr. Marc Apple at Parkview Oncology Center
Discuss imaging from a clinician point of view
Choose common modalities: Ximatron Hi-Res FluoroscopyPicker PQS CT Scanning SimulatorSiemans Magnetic Resonance Imaging
![Page 11: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/11.jpg)
Ximatron Hi-Res Fluoroscopy100 cm source to surface distance110kv; 25ma; 3mas, 1.5mag latitude for technician
![Page 12: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/12.jpg)
Typical Fluoroscopic Image
![Page 13: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/13.jpg)
Ximatron Image Comparison
Tantalum
90 Pt-10Ni
Coni-DFT®
-33Pt
MP-DFT®
-28Ag
MP35N
DBS(7)
Nitinol
CP Titanium
Bolus Air
![Page 14: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/14.jpg)
Clinician Comments
Amazed at variation due to equipment set up
Observed “zebra stripe effect” in bird’s eye view
![Page 15: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/15.jpg)
Picker “PQS” System “CT Scan”
130kv; 150ma; pitch=1.5
latitude for technician
![Page 16: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/16.jpg)
CT scan Image Reconstruction Options
Name Windows Level Map Type Description K#
Default 300 40 Ramp Simple Default 0
Lung 1 600 -300 Ramp Simple Scanner 5
Lung 2 120 -650 Ramp Simple Scanner 6
Modified head bone
2516 258 Ramp Simple Scanner 1
Normal Bone
0 210 Step Simple DRR 43
Fat 150 -100 Boxcar Simple 3-D 23
Chest 1060 -900 Ramp Simple 2-D/MPR 21
![Page 17: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/17.jpg)
CT Scan of Wire FrameThanks toKatie Shively, R.T.T.Parkview HospitalRegional Cancer CenterFort Wayne, Indiana
![Page 18: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/18.jpg)
![Page 19: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/19.jpg)
![Page 20: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/20.jpg)
![Page 21: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/21.jpg)
Picker Image Comparison
![Page 22: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/22.jpg)
Variables in Imaging
Media- Impact of bolus was significant
Default equipment settings
![Page 23: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/23.jpg)
Siemens “Magnetom Harmony” magnetic resonance imaging
![Page 24: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/24.jpg)
MRI Measurement Parameters
“Magnetom Harmony” Class1.0 Tesla magnetic fieldOperated at a 42 M#2 frequencyUsed head coil with IR=415.015 internal slices with 3-D reconst3.00mm slice thickness, 0 skip192 x 256 matrixContrast setting W=792-798, C=369-376All images obtained using T-1 weighting
![Page 25: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/25.jpg)
Typical MRI Image
![Page 26: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/26.jpg)
MRI Comparison
![Page 27: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/27.jpg)
Variables in Imaging
Head coil was necessary to obtain images
Resolution was better than presented images
![Page 28: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/28.jpg)
Next Step Options
Explore impact of form on imaging (flat wire, strand, cable, shapes)
Explore surface condition impact (plating, oxides, textures)
Move on to other Modalities (Ultrasound)
![Page 29: Fundamental Imaging Characteristics and Resolution of Metal Alloys By Marc G. Apple, M.D. Mark Michael This presentation will probably involve audience](https://reader035.vdocuments.site/reader035/viewer/2022062619/5517c3a45503461b658b487c/html5/thumbnails/29.jpg)
Conclusions
Novel pilot study to preemptively evaluate tissue equivalent characteristics before prototype devices
This protocol may prove useful as reference data for diameter and material selection