Features:• Highresolution15inchtouch-screendisplay

• 8-waveformdisplaywithlargewaveformsanddigits

• Simultaneousmulti-leadECGmonitoring

• Autosettingofalarmlimits

• Color-codedalarms

• Graphicalandtabulartrends(72hours)

• Patientdataentry

• Batterybackup(2removablebatteries)

• Networkingcapability(wirelessandnon-wireless)

• Standbybutton

• Colorofthewaveformscanbechanged

• Newfor2015!MasimoSET®PulseOximetry

TheDREWavelineProcanbeusedonsmall,medium,orlargeanimals.Anapplicablemonitorforveterinarysurgicalsuitesandresearchfacilities.

TheDREWavelineProdisplaysasmanyaseightcolorfulwaveformsonacrisp15”highresolutiondisplay.Itislightweight,portableandmeasuresmultipleparametersandcomeswithstandardveterinaryaccessories.

O.R. monitor provides superior flexibility and performance — Now with Masimo SET® Pulse Oximetry!

NIBP, 3/5 Lead ECG, Dual Temperature, Respiration, SpO2, Invasive Blood Pressure, and Capnography

Also available with Invasive Blood Pressures, CO2 (Mainstream or Sidestream) and Drager Built-in Anesthetic Agent Module

1800 Williamson Ct. • Louisville, Kentucky 40223 USA800-477-2006 • 502-244-4444 • FAX: 502-244-0369 www.dreveterinary.com

Page 2: DRE aveline Pro Multi-Function Patient Monitor 3/5 Lead ECG, Dual Temperature, Respiration, SpO2, Invasive Blood Pressure, and Capnography

Multi-Function Patient MonitorDRE Waveline Pro

Physical Dimensions & Weight BaseUnit: 12.1”/309mm(H)x14.4”/365mm(W)x6.3”/159mm(D)Weight: 17.6lbs/8.0kgs(includinganestheticagentsmoduleandbatteries)

Application Neonatal,pediatricandadultpatients

Performance Specifications Display: 15”colorTFT,Resolution1024x768Trace: 8waveformsIndicator: Alarmindicator Powerindicator QRSbeepandalarmsound

ECG Input: 5-leadECGcableandstandardAAMIlineforconnectionLeadChoice: I,II,III,aVR,aVF,aVL,V,V1-V6,TESTGainChoice: x0.5,x1,x2,x4FrequencyCharacteristic: 0.05~35HZ(+3dB)ECGWaveforms: 7channelsPenetrationVoltage: 4000VAC50/60HzSweepSpeed: 12.5,25,50and100mm/sec

(lefttorightorrighttoleft)HRDisplayRange: 30~300bpmAccuracy: ±1bpmor±1%,whicheverisgreaterAlarmLimitRangeSetting: Upperlimit100~200bpm;Lowerlimit30~100

bpm

RESPMeasureMethod: RA-LLimpedanceRange: 0~120rpmAccuracy: ±3rpmAlarmLimitSetting: Upperlimit6~120rpm:Lowerlimit3~120rpmSweepSpeed: 12.5,25,50and100mm/sec

(lefttorightorrighttoleft)SpO2 ASpO

2: Anti-motionSpO

2SpO

2%Range: 0–100%

SpO2Accuracy: ±2%(70–100%,non-motion)±3%(70–100%,motion)

PulseRateRange: 30–250bpmPulseRateAccuracy: ±2bpm(non-motion)±3bpm(motion)AlarmLimitSetting: Upperlimit70–100%,lowerlimit70–100%SpO

2Probe: RedlightLEDwavelength;660nm±5nm

InfraredlightLEDwavelength;940nm±10nm

NIBPMeasuringTechnology: AutomaticoscillatingmeasurementCuffInflating: <30s(0~300mmHg,standardadultcuff )MeasuringPeriod: AVE<40sMode: Manual,AutoMeasuringIntervalin AUTOMode: 2min~4hrsPulseRateRange: 30~250(bpm)MeasuringRange:Adult/PediatricMode SYS:40~250(mmHg) DIA:15~200(mmHg)NeonatalMode SYS:40~135(mmHg) DIA:15~100(mmHg)Resolution: 1mmHgAccuracy: MaximumMeanerror: ±5mmHg MaximumStandardDeviation: 8mmHgOverpressureProtection: AdultMode:300(mmHg)

NeonatalMode:160(mmHg)AlarmLimitSetting: SYS:50~240mmHg DIA:15~180mmHg

TEMPRange: 25°–50°(C)Accuracy: ±0.2°C(25.0°~34.9°C) ±0.1°C(35.0°~39.9°C) ±0.2°C(40.0°~44.9°C) ±0.3°C(45.0°~50.0°C)DisplayResolution: 0.1°CAlarmLimitSetting: Upperlimit0°~50°C,lowerlimit0°~50°CChannel: 2channels

IBPMeasurementRange: -50~300mmHgChannel: 2channelsPressureTransducer: Sensitivity,5μV/V/mmHgImpedanceRange: 300~3000ΩTransducerSites: ART,PA,CVP,RAP,LAP,ICPUnit: mmHg/kPaselectableResolution: 1mmHgAccuracy: ±1mmHgor±2%,whicheverisgreaterAlarmRange: -10~300mmHg

EtCO2 (Dräger) CO

2MeasurementRange: 0~99mmHg

Accuracy: ±2mmHg(0~38mmHg) 39-99mmHg±5%ofreading+0.08%for every1mmHg(above38mmHg)SamplingRate: 50ml/minInitializationTime: 30seconds(typical),reaches±5%steady-state accuracywithin3minutesRespirationRate: 0~150breaths/minMode: Adult,neonate

Anesthetic Agents Method: InfraredabsorptionGasSorts: Halothane,Isoflurane,Enflurane,Sevoflurane, Desflurane,CO2,N2O(optionalAutomaticAgentID)

MeasurementRange: Halothane,Isoflurane: 0~8.5% Enflurane,Sevoflurane: 0~10% Desflurane: 0~20% CO

2: 0~10%

N2O: 0~100%

O2: 0~100%

Bias: Halothane,Isoflurane,Enflurane, Sevoflurane,Desflurane: ±(0.15Vol%+15%rel.) CO

2: ±(0.5Vol%+12%rel.)

N2O: ±(2Vol%+8%rel.)

O2: ±3Vol%

Networking Industrystandard802.11b/gwirelessnetwork

Power Source: ExternalACpowerorinternalbatteryACPower: 100~240VAC,50/60Hz,150VABattery: Built-in&rechargeablelithiumionChargeTime: 3+hours

Environmental Specifications Temperature: Operating: 5°~40°C Storage: -20°~65°CHumidityrange: Operating: ≤80% Storage: ≤80%

Alsoavailable:Automatic Agent I.D. Consult your sales representative for specifications.

Specifications subject to change without notice.

1800 Williamson Ct. • Louisville, Kentucky 40223 USA800-477-2006 • 502-244-4444 • FAX: 502-244-0369 www.dreveterinary.com