Page 1: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

Mark Bailey

Arne Miller

Risø High Dose Reference Laboratory

Dosimetry and process control for using low energy electron beams for sterilization or decontamination of surfaces

Page 2: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Contents

•Introduction

–Applications of low energy electrons

•Pharmaceutical filling lines

– Issues

–Dosimetry requirements

•Dosimetry with low energy electrons: Dµ

–Absorbed dose to water in first micrometer at surface

•Routine process measurements

•Uncertainty budget

•Process limits

•Ongoing developments / improvements

2 25.04.2017

Page 3: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Introduction

•Low energy electron beams are used in different applications, including:

–Curing / crosslinking of polymers

–Disinfection / sterilisation of surfaces

•Pharmaceutical manufacture requires aseptic filling processes

–Disinfection / sterilisation of the outside of tubs containing syringes of vaccines, for example

3

Page 4: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Pharmaceutical tubs

4

Sterilised with empty syringes, and then filled under aseptic conditions

Page 5: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

5

Pharmaceutical filling line

Page 6: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

6

Pharmaceutical tub filling line

Page 7: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Placement of dosimeters on product

• Think like an electron…!

• How do electrons get into those hidden corners?

7

Page 8: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Electron beams…

8 25.04.2017

Low energy electrons scatter more than high energy electrons

They behave more like a cloud as they scatter...

Page 9: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Range of electrons

•10 MeV electrons:

–Water: 5 cm

–Air: 35 m

9 25.04.2017

•80 keV electrons:

–Water: 0.14 mm

–Air: 10 cm

Product

99%

Ti window and air

Product:

24%

Air

35%

Ti window

41%

Page 10: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Depth-dose distributions

10 25.04.2017

0.00

0.20

0.40

0.60

0.80

1.00

1.20

0 50 100 150 200

Do

se n

orm

alis

ed

to

su

rfac

e

Depth, µm, water

Normalised depth - dose distribution in water

80 keV

125 keV

150 keV

200 keV

Page 11: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Depth-dose distributions

11 25.04.2017

0.00

0.20

0.40

0.60

0.80

1.00

1.20

0 50 100 150 200

Do

se n

orm

alis

ed t

o s

urf

ace

Depth, µm, water

Normalised depth - dose distribution in water

125 keV

18 µm B3

50 µm FWT-60

135 µm alanine (density 1.35)

Average response in dosimeter will not reflect dose at surface

Problem gets worse as dosimeter thickness increases

Page 12: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Measuring the depth-dose distribution

12 25.04.2017

Dosimeter film

Thin Mylar films

Beam

Dosimeter films

Dosimeter film under a stack of e.g. Mylar.

Stack of dosimeter film

•Enables us to measure Dµ (absorbed dose in first µm) and relate this to apparent dose measured in reference dosimeters (e.g. alanine, thickness 135 µm)

•Dµ gives traceability to national standards of absorbed dose via calibration of alanine dosimetry system

Page 13: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Measuring the depth-dose distribution

13 25.04.2017

Depth-dose stack (here, B3) Scan and then analyse using RisøScan

Page 14: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Measuring the depth-dose distribution

14 25.04.2017

0

2

4

6

8

10

12

14

16

18

20

0 50 100 150

Do

se, k

Gy

Depth in Water, µm

Measured depth dose curve

1st measurement

2nd measurement

3rd measurement

Here, stack of B3 films

Polynomial fit to measured doses

Extrapolate to surface for Dµ

(Depth 0.5 µm)

Page 15: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Convert to surface dose Dµ (1)

15 25.04.2017

•Calibration verifications carried out with alanine film dosimeters over several doses

–Cover the range of doses expected to be used

–Cover the range of dosimeters to be used

Alanine films

Risø B3

GEX / B3 DoseStix

Page 16: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Convert to surface dose Dµ (2)

16 25.04.2017

•Note: Dose response is nonlinear for alanine!

–At high surface dose, response curve through alanine film is complex

–Alanine response also changes with temperature

–Alanine response measured using ESR spectrometer

•For these reasons, these calibrations are carried out by calibration laboratories

•Enables use of B3 / GEX DoseStix for routine process monitoring of surface treatment

•Uncertainty is currently quite large!

Page 17: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Routine process dosimetry

17 25.04.2017

•Use (e.g.) GEX DoseStix

–B3 in identified card holder

•Use reference dosimetry tub

–Maintains distance from emitter to dosimeter, so that dosimeter does not saturate

•Use measured and otherwise identified uncertainty components to set process limits on current and conveyor speed to achieve required doses

Page 18: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Example Uncertainty Budget

18 25.04.2017

Calibration

Routine process measurement

Dose mapping

Reference dosimeter - Alanine film

Calibration 1.51

Measurement 1.77

Determination of Dµ (at 150 keV) 5.00

Alanine films total uncertainty 5.30

Routine dose - DoseStix

Calibration 5.3

Dosimeter reproducibility 1.0

E-beam facility variability 7.1

Routine dose uncertainty 7.4

Routine dose total uncertainty 9.5

Dose map - Risø B3 / RisøScan

Calibration 6.1

Dosimeter reproducibility 2.00

E-beam facility variability 7.1

Minimum dose uncertainty 10.9

Minimum dose total uncertainty 12.5

Page 19: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Dose limits – minimum dose

19 25.04.2017

•Sterilisation or disinfection

–No measurement of bioburden is possible!

–Device has been sterilised prior to arrival at filling line

– It is then removed from sterile packaging and placed into filling line machinery

•Acceptance criteria for minimum dose: Usual recommendations:

–Measured dose at minimum dose locations on the surface of the irradiated tubs must exceed 15 kGy

–Average minimum dose should exceed 25 kGy

Page 20: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Dose limits

20 25.04.2017

•Acceptance criteria for maximum dose:

–Most of tub is constructed from polystyrene of about 1mm thickness

–Maximum dose of significantly above 100 kGy will not cause problems

–However, dose under

Tyvek lid / liner should be below 2 kGy

–Maximum dose to the top of the Tyvek lid is therefore measured to be below 100 kGy

Page 21: Dosimetry and process control for using low energy ... · 4/25/2017  · DTU Nutech, Technical University of Denmark Add Presentation Title in Footer via ”Insert”; ”Header &

DTU Nutech, Technical University of Denmark

Add Presentation Title in Footer via ”Insert”; ”Header & Footer”

Improvements and developments

21 25.04.2017

•Currently, development is under way with UK National Physical Laboratory (NPL) of a new graphite calorimeter

–Measure directly energy in graphite from low energy electrons (80 – 200 keV)

–Relate response in alanine at low energies to response in this calorimeter

•Significantly reduce the uncertainty in measurement of Dµ

Vacuum window and support mesh

Graphite calorimeter bodies Alanine films

Pipe to vacuum pump