“dosimetry and optimisation of cone beam ct for dentomaxillofacial applications”
DESCRIPTION
“Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”. S afety and E fficacy of a New and Emerging Dent al X -ray Modality. Cone Beam CT. Ruben Pauwels Oral Imaging Center Faculty of Medicine. Introduction CBCT SEDENTEXCT project CBCT dosimetry - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/1.jpg)
“Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”
Safety and Efficacy of a New and Emerging Dental X-ray Modality
Cone Beam CT
Ruben PauwelsOral Imaging CenterFaculty of Medicine
![Page 2: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/2.jpg)
• Introduction CBCT• SEDENTEXCT project• CBCT dosimetry
Effective dose (phantom)Standardised dose index
• CBCT OptimisationQC phantom development & analysis
• Work in progress
![Page 3: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/3.jpg)
• Introduction CBCT
![Page 4: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/4.jpg)
Cone Beam CT / Digital Volume Tomography
• Dentomaxillofacial • Radiotherapy• Cardiac imaging
![Page 5: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/5.jpg)
Cone Beam CT / Digital Volume Tomography
• Dentomaxillofacial
![Page 6: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/6.jpg)
Cone Beam CT / Digital Volume Tomography
• Dentomaxillofacial
![Page 7: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/7.jpg)
Cone Beam CT / Digital Volume Tomography
Advantages• 3D (isotropic voxels)• Detail (voxel size 0.08-0.4mm)• Cost, size, ease of use, … • Applications• ‘Low dose’ (generally DE=20-100 µSv)• Viewing possibilities
![Page 8: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/8.jpg)
Cone Beam CT / Digital Volume Tomography
• Dentomaxillofacial
![Page 9: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/9.jpg)
Can CBCT replace traditional radiographic methods?
![Page 10: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/10.jpg)
Can CBCT replace traditional radiographic methods?
• Excessive use: collective dose ↑• Used in primary care (by dentists)• Optimisation? ALARA?
Justification Optimisation (dose & image quality) Referral criteria Training
![Page 11: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/11.jpg)
• Introduction CBCT• SEDENTEXCT project
![Page 12: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/12.jpg)
![Page 13: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/13.jpg)
![Page 14: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/14.jpg)
WP 0 Financial and administrative management
WP 1 Justification and Guideline development
WP 2 Dosimetry
WP 3 Optimisation
WP 4 Diagnostic accuracy
WP 5 Cost effectiveness
WP 6 Training and Valorisation
![Page 15: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/15.jpg)
• Introduction CBCT• SEDENTEXCT project• CBCT dosimetry
![Page 16: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/16.jpg)
WP 2: Dosimetry
Patient dosimetry
Protection of personnel
Standardised dose index
Phantom effective dose
In vivo dose
Mathematical models
![Page 17: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/17.jpg)
• Introduction CBCT• SEDENTEXCT project• CBCT dosimetry
Effective dose (phantom)
![Page 18: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/18.jpg)
Which organs are the major contributors to the effective dose for dental CBCT imaging?
Which acquisition parameters influence the effective dose, and in what way?
kV, mAs, filtration, …FOV, ROI
How can we use this info to reduce patient dose?
WP 2: Dosimetry
![Page 19: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/19.jpg)
WP 2: Dosimetry• Thermoluminescent dosimeters (LiF:Mg,Ti)• Harshaw Reader 6600
![Page 20: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/20.jpg)
WP 2: Dosimetry• Calibration: C → gU → Gy
![Page 21: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/21.jpg)
WP 2: Dosimetry• Alderson Radiation Therapy phantom• 147 organ TLDs, 10 background TLDs, 20 calibration TLDs • ICRP 103 tissue weighting factors
![Page 22: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/22.jpg)
WP 2: DosimetryOrgan ICRP 60 ICRP 103
Gonads 0,2 0,08
Red bone marrow 0,12 0,12
Colon 0,12 0,12
Lung 0,12 0,12
Stomach 0,12 0,12
Bladder 0,05 0,04
Breast 0,05 0,12
Liver 0,05 0,04
Oesophagus 0,05 0,04
Thyroid 0,05 0,04
Skin 0,01 0,01
Bone surface 0,01 0,01
Brain 0 0,01
Salivary glands 0 0,01
Remainder 0,05 0,12
![Page 23: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/23.jpg)
WP 2: Dosimetry
Scanora 3D
Galileos
Picasso Trio
ProMax 3D
Kodak 9000 3D
![Page 24: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/24.jpg)
WP 2: DosimetryScanora 3D Upper jaw 100x75mm
Upper and lower jawLower jaw
Galileos Upper and lower jaw 150x150mm
Kodak 9000 3D Lower jaw (part) 50x37mm
Picasso Trio Lower jaw 120x70mm
ProMax 3D Lower jaw 80x80mm
![Page 25: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/25.jpg)
WP 2: Dosimetry
![Page 26: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/26.jpg)
WP 2: Dosimetry
![Page 27: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/27.jpg)
WP 2: Dosimetry
![Page 28: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/28.jpg)
WP 2: Dosimetry
![Page 29: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/29.jpg)
WP 2: Dosimetry
![Page 30: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/30.jpg)
WP 2: Dosimetry
![Page 31: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/31.jpg)
WP 2: Dosimetry
![Page 32: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/32.jpg)
WP 2: Dosimetry
OrganContribution (%)to effective dose
Red bone marrow 15
Thyroid 44
Skin 1,4
Bone surface 1,5
Brain 1,1
Salivary glands 36,9
![Page 33: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/33.jpg)
Large dose range Thyroid & salivary glands are crucial FOV reduction should always be considered mAs should be selected in function of diagnostic
needs Thyroid shielding & avoidance
Ideal CBCT practice?→ FOV selection (single tooth, …, full head)→ Resolution (=mAs) selection→ Training!!!
WP 2: Dosimetry
![Page 34: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/34.jpg)
• Introduction CBCT• SEDENTEXCT project• CBCT dosimetry
Effective dose (phantom)Standardised dose index
![Page 35: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/35.jpg)
Routine measurements → predict effective dose Conversion factors
(input: dose index + positioning) Existing methods (CTDI) Also include:
- patient (skin) entrance dose- simulations
WP 2: Dosimetry
![Page 36: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/36.jpg)
WP 2: Dosimetry
![Page 37: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/37.jpg)
WP 2: Dosimetry
Typically available dosimeters
![Page 38: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/38.jpg)
WP 2: Dosimetry
Central measurement?
Ionization chamber
![Page 39: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/39.jpg)
WP 2: Dosimetry
Accuracy?
TLD
![Page 40: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/40.jpg)
WP 2: Dosimetry
Sensitivity?Calibration?
Film
![Page 41: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/41.jpg)
WP 2: Dosimetry
FOV size and shapeUniform Distribution?(Practical) application for CBCT?
DAP
![Page 42: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/42.jpg)
WP 2: Dosimetry→ Combine measurements→ Assesment of distribution (anthropomorph / homogeneous)
![Page 43: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/43.jpg)
• Introduction CBCT• SEDENTEXCT project• CBCT dosimetry
Effective dose (phantom)Standardised dose index
• CBCT Optimisation
![Page 44: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/44.jpg)
Technical image quality assessment Image quality vs. Radiation dose! Diagnostical performance < technical parameters Optimising beyond available options Optimising on others levels: also in progress
Monte Carlo simulations (G.Zhang, Leuven)Reconstruction (e.g. artefact reduction)
WP 3: Optimisation
![Page 45: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/45.jpg)
What are the most relevant image quality parameters in dentomaxillofacial imaging, and how are they best measured using a test object?
How do CBCT devices relate to each other for these parameters?How do CBCT pixel values correlate to actual densities and CT
numbers (Hounsfield Units)? Are pixel values uniform throughout the reconstructed volume? Which are the most relevant acquisition parameters affecting the stability and linearity of pixel values?
Are CBCT datasets truly isotropic? How can metal artefacts be evaluated and which parameters
influence their appearance?
WP 3: Optimisation
![Page 46: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/46.jpg)
WP 3: Optimisation
Development of QA tools
PMMAcylinder
Inserts
![Page 47: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/47.jpg)
WP 3: Optimisation• Spatial resolution (XY and Z)• Point spread function• Line spread function• Contrast resolution • CT number• Artefact analysis• Geometric accuracy (XY and Z)• Homogeneity • Uniformity• Dose?
![Page 48: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/48.jpg)
Scanora 3D
Galileos
Picasso TrioProMax 3D
Kodak 9000 3D
WP 3: Optimisation
3D Accuitomo
i-CAT SkyView
![Page 49: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/49.jpg)
WP 3: Optimisation• Spatial resolution (XY and Z)• Point spread function• Line spread function• Contrast resolution • CT number• Artefact analysis• Geometric accuracy (XY and Z)• Homogeneity • Uniformity• Dose?
![Page 50: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/50.jpg)
WP 3: Optimisation
![Page 51: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/51.jpg)
WP 3: Optimisation
![Page 52: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/52.jpg)
WP 3: Optimisation
XY plane
![Page 53: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/53.jpg)
Resolution < 3 lp/mm Artefacts Saturation XY plane
→ line pair pattern on PMMA background
WP 3: Optimisation
![Page 54: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/54.jpg)
WP 3: Optimisation• Spatial resolution (XY and Z)• Point spread function• Line spread function• Contrast resolution • CT number• Artefact analysis• Geometric accuracy (XY and Z)• Homogeneity • Uniformity• Dose?
![Page 55: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/55.jpg)
WP 3: Optimisation
0.1 mm 0.25 mm
Steel wire in air
![Page 56: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/56.jpg)
![Page 57: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/57.jpg)
WP 3: Optimisation
![Page 58: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/58.jpg)
WP 3: Optimisation
i-CAT HR
mm5,24,84,443,63,22,82,421,61,20,80,40
pixe
l val
ue
0,24
0,22
0,2
0,18
0,16
0,14
0,12
0,1
0,08
0,06
0,04
0,02
0
Cauchy distribution
![Page 59: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/59.jpg)
WP 3: Optimisation3D Accuitomo
mm4,64,44,243,83,63,43,232,82,62,42,22
pixe
l val
ue
0,13
0,12
0,11
0,1
0,09
0,08
0,07
0,06
0,05
0,04
0,03
0,02
0,01
0
Galileos
mm5,65,24,84,443,63,22,82,421,61,20,80,40
pixe
l val
ue
0,4
0,36
0,32
0,28
0,24
0,2
0,16
0,12
0,08
0,04
0
i-CAT LR
mm6,45,64,843,22,41,60,80
pixe
l val
ue
0,26
0,24
0,22
0,2
0,18
0,16
0,14
0,12
0,1
0,08
0,06
0,04
0,02
0
![Page 60: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/60.jpg)
WP 3: OptimisationKodak 9000 3D
mm65,65,24,84,443,63,22,82,421,61,20,80,40
pixe
l val
ue
0,064
0,056
0,048
0,04
0,032
0,024
0,016
0,008
0
Scanora 3D
mm65,65,24,84,443,63,22,82,421,61,20,80,40
pixe
l val
ue
0,13
0,12
0,11
0,1
0,09
0,08
0,07
0,06
0,05
0,04
0,03
0,02
0,01
0
SkyView LR
mm5,24,84,443,63,22,82,421,61,20,80,40
pixe
l val
ue
0,18
0,16
0,14
0,12
0,1
0,08
0,06
0,04
0,02
0
![Page 61: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/61.jpg)
WP 3: Optimisation
Picasso Trio
mm65,65,24,84,443,63,22,82,421,61,20,80,40
pixe
l val
ue
0,3
0,28
0,26
0,24
0,22
0,2
0,18
0,16
0,14
0,12
0,1
0,08
0,06
0,04
0,02
0
SkyView HR
Histogram Cauchy
mm5,65,24,84,443,63,22,82,421,61,20,80,40
pixe
l val
ue
0,16
0,14
0,12
0,1
0,08
0,06
0,04
0,02
0
i-CAT HR
mm5,24,84,443,63,22,82,421,61,20,80,40
pixe
l val
ue
0,24
0,22
0,2
0,18
0,16
0,14
0,12
0,1
0,08
0,06
0,04
0,02
0
![Page 62: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/62.jpg)
WP 3: Optimisation
![Page 63: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/63.jpg)
WP 3: Optimisation• Spatial resolution (XY and Z)• Point spread function• Line spread function• Contrast resolution • CT number• Artefact analysis• Geometric accuracy (XY and Z)• Homogeneity • Uniformity• Dose?
![Page 64: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/64.jpg)
WP 3: OptimisationPTFE cube
PMMA surrounding
ESF↓
LSF↓
MTF
Gumbel distribution
![Page 65: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/65.jpg)
![Page 66: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/66.jpg)
WP 3: Optimisation
i-CAT HR
mm1211109876543210
pixe
l val
ue
0,03
0,028
0,026
0,024
0,022
0,02
0,018
0,016
0,014
0,012
0,01
0,008
0,006
0,004
0,002
0
Gumbel distribution
![Page 67: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/67.jpg)
WP 3: OptimisationGalileos
mm1211109876543210
pixe
l val
ue
0,072
0,064
0,056
0,048
0,04
0,032
0,024
0,016
0,008
0
i-CAT HR
mm1211109876543210
pixe
l val
ue
0,03
0,028
0,026
0,024
0,022
0,02
0,018
0,016
0,014
0,012
0,01
0,008
0,006
0,004
0,002
0
i-CAT LR
mm1211109876543210
pixe
l val
ue
0,06
0,056
0,052
0,048
0,044
0,04
0,036
0,032
0,028
0,024
0,02
0,016
0,012
0,008
0,004
0
Kodak 9000 3D
mm131211109876543210
pixe
l val
ue
0,028
0,026
0,024
0,022
0,02
0,018
0,016
0,014
0,012
0,01
0,008
0,006
0,004
0,002
0
![Page 68: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/68.jpg)
WP 3: OptimisationScanora 3D
mm1211109876543210
pixe
l val
ue
0,026
0,024
0,022
0,02
0,018
0,016
0,014
0,012
0,01
0,008
0,006
0,004
0,002
0
SkyView HR
mm1211109876543210
pixe
l val
ue
0,1
0,09
0,08
0,07
0,06
0,05
0,04
0,03
0,02
0,01
0
SkyView LR
mm1211109876543210
pixe
l val
ue
0,09
0,08
0,07
0,06
0,05
0,04
0,03
0,02
0,01
0
Picasso Trio
mm1211109876543210
pixe
l val
ue
0,032
0,028
0,024
0,02
0,016
0,012
0,008
0,004
0
![Page 69: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/69.jpg)
WP 3: Optimisation
![Page 70: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/70.jpg)
WP 3: Optimisation• Spatial resolution (XY and Z)• Point spread function• Line spread function• Contrast resolution • CT number• Artefact analysis• Geometric accuracy (XY and Z)• Homogeneity • Uniformity• Dose?
![Page 71: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/71.jpg)
WP 3: Optimisation• Air• Aluminium• Hydroxyapatite (50, 100, 200)
![Page 72: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/72.jpg)
WP 3: Optimisation• Spatial resolution (XY and Z)• Point spread function• Line spread function• Contrast resolution • CT number• Artefact analysis• Geometric accuracy (XY and Z)• Homogeneity • Uniformity• Dose?
![Page 73: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/73.jpg)
WP 3: OptimisationCNR = MPV1-MPV2 (SD1+SD2)/2
![Page 74: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/74.jpg)
WP 3: Optimisation
![Page 75: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/75.jpg)
WP 3: Optimisation
![Page 76: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/76.jpg)
WP 3: Optimisation
• Pixel value: relation HA 50,100,200?• Work in progress
![Page 77: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/77.jpg)
WP 3: Optimisation
![Page 78: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/78.jpg)
WP 3: Optimisation
![Page 79: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/79.jpg)
WP 3: Optimisation• Spatial resolution (XY and Z)• Point spread function• Line spread function• Contrast resolution • CT number• Artefact analysis• Geometric accuracy (XY and Z)• Homogeneity • Uniformity• Dose?
![Page 80: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/80.jpg)
WP 3: OptimisationTitanium rods
![Page 81: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/81.jpg)
WP 3: Optimisation
![Page 82: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/82.jpg)
WP 3: Optimisation
![Page 83: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/83.jpg)
WP 3: OptimisationSmoothing + subtracting background
2 ROIs, ignore black streaks
![Page 84: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/84.jpg)
WP 3: OptimisationMeasure along intervals, keep extreme values
Distribute (e.g. Gumbel) and evaluate parameters
![Page 85: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/85.jpg)
ProMax 3D
84 kV 9 mA
ProMax 3D
84 kV 14 mA
WP 3: Optimisation→ Inter-device variability→ Intra-device variability?
![Page 86: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/86.jpg)
WP 3: Optimisation• Spatial resolution (XY and Z)• Point spread function• Line spread function• Contrast resolution • CT number• Artefact analysis• Geometric accuracy (XY and Z)• Homogeneity • Uniformity• Dose?
![Page 87: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/87.jpg)
WP 3: Optimisation
Cfr. spatial resolution
![Page 88: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/88.jpg)
WP 3: Optimisation• Spatial resolution (XY and Z)• Point spread function• Line spread function• Contrast resolution • CT number• Artefact analysis• Geometric accuracy (XY and Z)• Homogeneity • Uniformity• Dose?
![Page 89: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/89.jpg)
WP 3: Optimisation
![Page 90: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/90.jpg)
WP 3: Optimisation
Small FOV (truncation, …) Non-homogeneous slices
![Page 91: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/91.jpg)
• Introduction CBCT• SEDENTEXCT project• CBCT dosimetry
Effective dose (phantom)Standardised dose index
• CBCT OptimisationQC phantom development & analysis
• Work in progress
![Page 92: “Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications”](https://reader036.vdocuments.site/reader036/viewer/2022062316/56814a54550346895db77673/html5/thumbnails/92.jpg)
Dose index Effective dose / organ dose analysis (+paediatric) QC phantom: second prototype Image analysis: artefacts, LSF, MTF Pixel homogeneity / uniformity Compare devices & evaluate differences