This report is for the exclusive use of Intertek's Client and is provided pursuant to the agreement between Intertek and its Client. Intertek's responsibility and liability are limited to the terms and conditions of the agreement. Intertek assumes no liability to any party, other than to the Client in accordance with the agreement, for any loss, expense or damage occasioned by the use of this report. Only the Client is authorized to copy or distribute this report and then only in its entirety. Any use of the Intertek name or one of its marks for the sale or advertisement of the tested material, product or service must first be approved in writing by Intertek. The observations and test results in this report are relevant only to the sample tested. This report by itself does not imply that the material, product, or service is or has ever been under an Intertek certification program. FOR THE SCOPE OF ACCREDITATION UNDER NVLAP LAB CODE 100402-0. DLC REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Project No. G101977042 Date: January 26, 2015 REPORT NO. 101977042CRT-001 TEST OF ONE 2x2 RETROFIT KIT FAMILY RETROFIT KIT MODEL NO.: DL-R22-A-XXXX-4K0-XXX-XX FIXTURE MODEL NO. METALUX 2GR8U6T8R DRIVER MODEL NO. XXXX=2035, 3035, OR 2535 LED MODEL NO. SAMSUNG 5630 RENDERED TO DIOLUCE, LLC 185 MAIN ST NEW BRITAIN, CT 06051 TEST: Electrical and Photometric tests as required to the IESNA test standard. STATEMENT OF LIMITATION: This report must not be used by the client to claim product certification, approval, or endorsement by NVLAP, NIST, or any agency of the federal government. AUTHORIZATION: The testing performed was authorized by signed quote number Q500574369. STANDARDS USED: The following American National Standards or Illuminating Engineering Society of North America Test Guides were used in part or totally to test each specimen: IESNA LM-79 (2008): Approved Method for Electrical and Photometric Measurements of Solid-State Lighting Products ANSI ANSLG C38.377 (2012): Specifications of the Chromaticity of Solid State Lighting Products Energy Star Manufacturer’s Guide Version 2.1 (2010): Guide for Qualifying Solid State Lighting Luminaires DESCRIPTION OF SAMPLE: The client submitted three prototype retrofit sample of model number DL- R22-XXXX-4K0-XXX and six drivers (two per model.) The samples were received by Intertek on January 15, 2015, in undamaged condition, and five combinations were tested as received. DATES OF TESTS: January 15, 2015 through January 21, 2015

Upload: others

Post on 27-Jun-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

This report is for the exclusive use of Intertek's Client and is provided pursuant to the agreement be tween Intertek and its Client. Intertek's responsib ility and liability are limited to the terms and conditions o f the agreement. Intertek assumes no liability to a ny party, other than to the Client in accordance wi th the agreement, for any loss, expense or damage occasion ed by the use of this report. Only the Client is au thorized to copy or distribute this report and then only in its entirety. Any use of the Intertek name or one of it s marks for the sale or advertisement of the tested material, product or service must first be approve d in writing by Intertek. The observations and test results in this report are relevant only to the sample tested. Thi s report by itself does not imply that the material , product, or service is or has ever been under an Intertek certi fication program.

FOR THE SCOPE OF ACCREDITATION UNDER NVLAP

LAB CODE 100402-0.

DLC REPORT

3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Project No. G101977042 Date: January 26, 2015

REPORT NO. 101977042CRT-001

TEST OF ONE 2x2 RETROFIT KIT FAMILY

RETROFIT KIT MODEL NO.: DL-R22-A-XXXX-4K0-XXX-XX FIXTURE MODEL NO. METALUX 2GR8U6T8R

DRIVER MODEL NO. XXXX=2035, 3035, OR 2535 LED MODEL NO. SAMSUNG 5630

RENDERED TO

DIOLUCE, LLC 185 MAIN ST

NEW BRITAIN, CT 06051 TEST: Electrical and Photometric tests as required to the IESNA test standard. STATEMENT OF LIMITATION: This report must not be used by the client to claim product certification,

approval, or endorsement by NVLAP, NIST, or any agency of the federal government.

AUTHORIZATION: The testing performed was authorized by signed quote number Q500574369. STANDARDS USED: The following American National Standards or Illuminating Engineering Society of

North America Test Guides were used in part or totally to test each specimen: IESNA LM-79 (2008): Approved Method for Electrical and Photometric Measurements of Solid-State Lighting Products ANSI ANSLG C38.377 (2012): Specifications of the Chromaticity of Solid State Lighting Products Energy Star Manufacturer’s Guide Version 2.1 (2010): Guide for Qualifying Solid State Lighting Luminaires DESCRIPTION OF SAMPLE: The client submitted three prototype retrofit sample of model number DL-

R22-XXXX-4K0-XXX and six drivers (two per model.) The samples were received by Intertek on January 15, 2015, in undamaged condition, and five combinations were tested as received.

DATES OF TESTS: January 15, 2015 through January 21, 2015

Page 2: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 2 of 22 Date: January 26, 2015

SUMMARY

Model No.: DL-R22-A-XXXX-4K0-XXX-XX DLC

Category: #31 – Retrofit Kits for 2x2 Luminaires for Ambient Lighting of Interior Commercial Spaces

Description:

Dioluce 2x2 Retrofit kit, with 3 LED bars, 4000K, where: XXXX = driver: (2035=120Vdriver, 3035=277V driver, or 2535 = 120V IR driver) XXX = lens: 001 (Clear) or 002 (Translucent) XX = control setting: 100 (High) or IR (Wireless dimming)

Results – Goniometer Assessment / (Requirement)

Model No: DL-R22-A-2035-

4K0-001-100 DL-R22-A-2035-

4K0-002-100 DL-R22-A-3035-

4K0-001-100 DL-R22-A-3035-

4K0-002-100 DL-R22-A-2535-

4K0-001-IR DL-R22-A-2535-

4K0-002-IR --

Driver: 2035=120V 2035=120V 3035=277V 3035=277V 2535=120V 2535=120V -- Lens: 001=Clear 002=Translucent 001=Clear 002=Translucent 001=Clear 002=Translucent --

Control Setting: 100=High 100=High 100=High 100=High IR = Wireless

Dimming on High IR = Wireless

Dimming on High --

Total Lumen Output

(Lumens) 3020.5 2927.2 3015.8 2927.4 3058.6 2968.6

Complies (>2000 lum)

Total Power (W)

28.699 28.685 31.297 31.138 29.34 29.277 --

Power Factor () 0.996 0.995 0.955 0.960 0.996 0.996 Complies

(≥0.9) Luminaire

Efficacy (LPW) 105.25 102.05 96.36 94.01 104.25 101.40 Complies

(>85 LPW) Spacing

Criteria (0-180) 1.28 1.20 1.26 1.20 1.28 1.20

Complies (1.0 – 2.0)

Spacing Criteria (90-

270) 1.26 1.22 1.26 1.22 1.26 1.22

Complies (1.0 – 2.0)

Zonal Lumens 0-60° (%)

82.8 79.6 82.9 79.3 83.30 79.4 Complies (≥75%)

Page 3: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 3 of 22 Date: January 26, 2015

SUMMARY (cont’d)

Criteria Results - Sphere Assessment / (Requirement)

Model No: DL-R22-A-2035-4K0-001-100 DL-R22-A-2035-4K0-002-100 DL-R22-A-3035-4K0-002-100 -- Driver: 2035=120V 2035=120V 3035=277V Lens: 001=Clear 002=Translucent 002=Translucent

Control Setting: 100=High 100=High 100=High

Total Lumen Output (Lumens) 3125.8 3006.3 3096.7 Complies

(>2000 lum) Total Power (W) 29.16 29.13 31.26 --

Luminaire Efficacy (LPW) 107.19 103.20 99.06 Complies

(>85 LPW) Power Factor 0.995 0.995 0.952 Complies (≥0.9)

Current ATHD (%) 5.40 5.35 15.36 Complies (≤20%)

Correlated Color Temperature (CCT - K) 4583 4255 4584 Complies (≤5000K)

Color Rendering Index (CRI) – Ra 85.0 85.0 85.0 Complies (≥80) Color Rendering Index (CRI) - R9 17.8 18.6 17.8 --

Duv 0.000 0.002 0.000 -- Chromaticity Coordinate (x) 0.358 0.369 0.358 -- Chromaticity Coordinate (y) 0.362 0.366 0.362 -- Chromaticity Coordinate (u’) 0.216 0.222 0.216 -- Chromaticity Coordinate (v’) 0.492 0.495 0.492 --

7-step MacAdam ellipse w/ LED quadrangle – 4000K

0.3500.3550.3600.3650.3700.3750.3800.3850.3900.3950.4000.4050.410

0.360 0.365 0.370 0.375 0.380 0.385 0.390 0.395 0.400 0.405

y

x

ANSI 4000/4100K MACADAM ELLIPSE, 7 STEPW/ LED QUADRANGLE

Page 4: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 4 of 22 Date: January 26, 2015

SUMMARY (cont’d)

Result

Assessment / (Requirement)

Model No: DL-R22-A-2035-4K0-002-

100 --

Driver: 2035=120V -- Lens: 002=Translucent --

Control Setting: 100=High -- Maximum In-Situ Source Temperature

Point (°C) 49.3 Complies (<107C)

EQUIPMENT LIST

Equipment Used Model Number Control Number Last

Calibration Date Calibration Due Date

LSI High Speed Mirror Goniometer 6440 --- 12/16/14 01/16/15 Elgar Power Supply CW1251 --- VBU VBU Yokogawa Power Analyzer WT210 E464 04/17/14 04/17/15 ExTech Hygro Thermometer 445703 T1357 12/10/14 12/10/15 Fisher Scientific Stopwatch 14-649-9 N1405 8/25/14 8/25/15 M-D Building Products Digital Level Smart Tool L112 3/14/2014 3/15/15 Yokogawa Power Analyzer WT1600 E474 03/07/14 03/07/15

LABSPHERE 3M W/ CDS

1100 N307 VBU VBU

Fluke Temperature Meter 53 II T1318 03/21/14 03/21/15 Elgar Power Supply CW1251 --- VBU VBU Extech Hygro-Thermometer 445703 T1355 12/10/14 12/10/15 SORENSEN POWER SUPPLY XFR 150-8 --- VBU VBU NIST Spectral Flux Standard Source RF1024 --- 9/18/2010 100 hrs of use

Extech Hygro-Thermometer 445703 T1366 12/10/14 12/10/15 Fluke Multimeter 87 E259 03/20/14 03/20/15

Fluke Temperature Meter 53 II N1324 03/21/14 03/21/15

Page 5: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 5 of 22 Date: January 26, 2015

TEST METHODS Seasoning in Sample Orientation – LED Products No seasoning was performed in accordance with IESNA LM-79. Photometric and Electrical measurements – Distribution Method A LSI Type C High Speed Model 6440 Mirror Goniometer was used to measure the intensity (candelas) at each angle of distribution for each sample. Ambient temperature was measured equal to the height of the sample mounted on the Goniometer equipment. Each sample was operated at input rated voltage in its designated orientation. Each sample was allowed to stabilize for at least thirty minutes before measurements were made. Electrical measurements including voltage, current, and power were measured using the Xitron or Yokogawa Power Analyzer. Some graphics were created with Photometrics Plus software. Photometric and Electrical Measurements – Integrating Sphere Method A Labsphere Model CDS 1100 CCD Array Spectroradiometer and Two Meter or Ten Foot Sphere was used to measure correlated color temperature, chromaticity coordinates, and the color rendering index for each SSL unit. Ambient temperature was measured at a position inside the sphere. Each SSL unit was operated on the client provided driver at the rated input voltage in its designated orientation. Each SSL unit was allowed to stabilize for at least thirty minutes before measurements were made. Electrical measurements including voltage, current, and power were measured using the Xitron or Yokogawa Power Analyzer. The calibration of the sphere photometer-spectroradiometer system is traceable to the National Institute of Standards and Technology. In-Situ Maximum Measured Power Supply Case and LED Source Point Temperature Power supply case and/or LED source operating temperature measurements were taken on one test sample per model with a thermocouple and Fluke 87 temperature meter. The SSL sample was allowed to reach thermal equilibrium for seven and a half hours before measurements were taken. Power supply or source temperature measurements were measured at the TMPPS or TS point as indicated by the included diagram in accordance with manufacturers declared hot spot location, or at a hot spot location found with a thermal camera when no diagram from the manufacturer is given. The maximum temperature was recorded for the sample. A simulated ceiling or other enclosure may be used in accordance to UL 1598 or UL 153 as applicable.

Page 6: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 6 of 22 Date: January 26, 2015

RESULTS OF TESTS Photometric and Electrical Measurements at 25°C – Integrating Sphere Method

Intertek Sample No. Desc.

Base Orientation

Input Voltage (Vac)

Input Current

(mA)

Input Power (Watts)

Input Power Factor

Current ATHD

(%)

Absolute Luminous

Flux (Lumens)

Lumen Efficacy (Lumens Per Watt)

DL-R22-2035-4K0-

001

120V driver,

clear lens Horizontal 120.0 244.20 29.16 0.995 5.40 3125.8 107.19

DL-R22-2035-4K0-

002

120V driver,

translucent lens

Horizontal 120.0 243.95 29.13 0.995 5.35 3006.3 103.20

DL-R22-3035-4K0-

002

277V driver,

translucent lens

Horizontal 277.0 118.49 31.26 0.952 15.36 3096.7 99.06

Intertek Sample No.

Correlated Color

Temperature (K)

CRI -Ra

CRI -R9

DUV

CIE 31’ Chromaticity Coordinate

(x)

CIE 31’ Chromaticity Coordinate

(y)

CIE 76’ Chromaticity Coordinate

(u’)

CIE 76’ Chromaticity Coordinate

(v’) DL-R22-2035-

4K0-001 4583 85.0 17.8 0.000 0.358 0.362 0.216 0.492

DL-R22-2035-4K0-002

4255 85.0 18.6 0.002 0.369 0.366 0.222 0.495

DL-R22-3035-4K0-002

4584 85.0 17.8 0.000 0.358 0.362 0.216 0.492

Page 7: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 7 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) Photometric and Electrical Measurements – Distribution Method

Intertek Sample No. Description

Input Voltage (Vac)

Input Current (mA)

Input Power (Watts)

Input Power Factor

Absolute Luminous

Flux (Lumens)

Lumen Efficacy (Lumens Per Watt)

DL-R22-A-2035-4K0-001-100

120V Driver, 4000K, clear

lens, high control setting

120.04 240.14 28.699 0.9956 3020.5 105.25

Intensity (Candlepower) Summary at 25°C - Candelas

Angle 0 22.5 45 67.5 90 0 1162 1162 1162 1162 1162 5 1165 1159 1157 1160 1157

10 1155 1153 1139 1149 1156 15 1132 1126 1133 1152 1146 20 1087 1082 1110 1116 1116 25 1033 1036 1044 1051 1042 30 986 996 1011 994 990 35 920 935 934 911 899 40 809 827 816 819 802 45 724 731 720 712 705 50 641 624 609 610 608 55 537 518 504 514 514 60 410 407 407 420 431 65 286 296 318 335 343 70 173 201 231 251 260 75 88 124 157 180 187 80 34 65 98 121 129 85 9 25 51 59 63 90 0 0 0 0 0

Page 8: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 8 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) Illumination Plots

Mounting Height: 10 ft. Illuminance - Cone of Light Isoillumination Plot

Zonal Lumen Summary and Percentages at 25°C

Zone Lumens % Luminaire 0-30 912.9 30.2 0-40 1486.1 49.2 0-60 2499.9 82.8 60-90 520.6 17.2 0-90 3020.5 100

90-180 0 0 0-180 3020.5 100

Page 9: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 9 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) Photometric and Electrical Measurements – Distribution Method

Intertek Sample No. Description

Input Voltage (Vac)

Input Current (mA)

Input Power (Watts)

Input Power Factor

Absolute Luminous

Flux (Lumens)

Lumen Efficacy (Lumens Per Watt)

DL-R22-A-2035-4K0-002-100

120V driver, 4000K,

translucent lens, high

control setting

120.01 240.15 28.685 0.9953 2927.2 102.5

Intensity (Candlepower) Summary at 25°C - Candelas

Angle 0 22.5 45 67.5 90 0 1130 1130 1130 1130 1130 5 1126 1123 1121 1126 1125

10 1110 1106 1103 1107 1108 15 1078 1076 1074 1078 1078 20 1024 1024 1026 1037 1039 25 955 953 955 966 969 30 896 898 905 918 922 35 824 827 830 843 846 40 726 728 738 754 759 45 636 641 659 678 684 50 552 556 578 602 609 55 468 473 501 525 532 60 385 392 420 448 458 65 299 309 341 369 379 70 213 230 263 288 299 75 136 157 189 216 228 80 72 92 126 154 168 85 20 40 70 83 88 90 0 0 0 0 0

Page 10: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 10 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) Illumination Plots

Mounting Height: 10 ft. Illuminance - Cone of Light Isoillumination Plot

Zonal Lumen Summary and Percentages at 25°C

Zone Lumens % Luminaire 0-30 853.8 29.2 0-40 1373.0 46.9 0-60 2328.7 79.6 60-90 598.5 20.4 0-90 2927.2 100

90-180 0.0 0 0-180 2927.2 100

Page 11: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 11 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) Photometric and Electrical Measurements – Distribution Method

Intertek Sample No. Description

Input Voltage (Vac)

Input Current (mA)

Input Power (Watts)

Input Power Factor

Absolute Luminous

Flux (Lumens)

Lumen Efficacy (Lumens Per Watt)

DL-R22-A-3035-4K0-001-100

277V driver, 4000K, cleaR

lens, high control setting

277.0 118.39 31.297 0.9549 3015.8 96.36

Intensity (Candlepower) Summary at 25°C - Candelas

Angle 0 22.5 45 67.5 90 0 1168 1168 1168 1168 1168 5 1164 1160 1159 1165 1163

10 1154 1149 1144 1153 1152 15 1133 1133 1133 1140 1147 20 1089 1087 1102 1113 1121 25 1027 1036 1047 1053 1048 30 981 997 1014 995 987 35 911 927 926 907 908 40 803 822 814 809 800 45 723 726 720 708 699 50 646 626 612 608 605 55 543 520 508 514 511 60 418 409 408 423 429 65 283 296 314 335 344 70 167 198 229 249 260 75 84 121 155 174 187 80 34 64 98 118 129 85 8 25 52 57 62 90 0 0 0 0 0

Page 12: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 12 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) Illumination Plots

Mounting Height: 10 ft. Illuminance - Cone of Light Isoillumination Plot

Zonal Lumen Summary and Percentages at 25°C

Zone Lumens % Luminaire 0-30 913.8 30.3 0-40 1484.8 49.2 0-60 2499.3 82.9 60-90 516.6 17.1 0-90 3015.8 100

90-180 0 0 0-180 3015.8 100

Page 13: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 13 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) Photometric and Electrical Measurements – Distribution Method

Intertek Sample No. Description

Input Voltage (Vac)

Input Current (mA)

Input Power (Watts)

Input Power Factor

Absolute Luminous

Flux (Lumens)

Lumen Efficacy (Lumens Per Watt)

DL-R22-A-3035-4K0-002-100

277V driver, 4000K,

translucent lens, high

control setting

277.0 122.01 31.138 0.9604 2927.4 94.01

Intensity (Candlepower) Summary at 25°C - Candelas

Angle 0 22.5 45 67.5 90 0 1123 1123 1123 1123 1123 5 1120 1116 1114 1118 1118

10 1105 1099 1095 1098 1101 15 1073 1070 1068 1070 1071 20 1019 1017 1021 1030 1033 25 950 949 951 960 964 30 894 895 902 915 919 35 823 828 833 844 846 40 726 732 740 756 760 45 637 643 658 679 685 50 551 556 575 601 610 55 467 470 496 522 532 60 385 387 420 448 459 65 300 307 345 372 382 70 217 231 267 293 304 75 138 160 193 220 231 80 73 96 128 156 170 85 22 40 69 86 92 90 0 0 0 0 0

Page 14: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 14 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) Illumination Plots

Mounting Height: 10 ft. Illuminance - Cone of Light Isoillumination Plot

Zonal Lumen Summary and Percentages at 25°C

Zone Lumens % Luminaire 0-30 849.3 29.0 0-40 1368.7 46.8 0-60 2322.7 79.3 60-90 604.8 20.7 0-90 2927.4 100

90-180 0 0 0-180 2927.4 100

Page 15: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 15 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) Photometric and Electrical Measurements – Distribution Method

Intertek Sample No. Description

Input Voltage (Vac)

Input Current (mA)

Input Power (Watts)

Input Power Factor

Absolute Luminous

Flux (Lumens)

Lumen Efficacy (Lumens Per Watt)

DL-R22-A-2535-4K0-

001-IR

120V driver, 4000K, clear lens, wireless control set on

high

120.0 245.35 29.34 0.9959 3058.6 104.25

Intensity (Candlepower) Summary at 25°C - Candelas

Angle 0 22.5 45 67.5 90 0 1191 1191 1191 1191 1191 5 1187 1181 1183 1186 1184

10 1175 1169 1172 1173 1174 15 1152 1148 1154 1158 1161 20 1108 1109 1119 1137 1135 25 1049 1053 1068 1070 1067 30 1008 1015 1028 1024 1020 35 937 951 951 937 930 40 822 839 837 830 820 45 736 747 733 726 714 50 652 640 620 618 612 55 556 530 514 512 516 60 430 414 411 419 427 65 297 302 314 328 338 70 174 202 223 244 253 75 87 123 151 172 179 80 35 64 93 114 122 85 9 24 46 51 54 90 0 0 0 0 0

Page 16: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 16 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) Illumination Plots

Mounting Height: 10 ft. Illuminance - Cone of Light Isoillumination Plot

Zonal Lumen Summary and Percentages at 25°C

Zone Lumens % Luminaire 0-30 930.3 30.4 0-40 1516.4 49.6 0-60 2547.2 83.3 60-90 511.4 16.7 0-90 3058.6 100

90-180 0 0 0-180 3058.6 100

Page 17: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 17 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) Photometric and Electrical Measurements – Distribution Method

Intertek Sample No. Description

Input Voltage (Vac)

Input Current (mA)

Input Power (Watts)

Input Power Factor

Absolute Luminous

Flux (Lumens)

Lumen Efficacy (Lumens Per Watt)

DL-R22-A-2535-4K0-

002-IR

120V driver, 4000K,

translucent lens, wireless control set on

high

120.0 244.84 29.277 0.9958 2968.6 101.4

Intensity (Candlepower) Summary at 25°C - Candelas

Angle 0 22.5 45 67.5 90 0 1139 1139 1139 1139 1139 5 1135 1130 1131 1136 1135

10 1118 1112 1112 1118 1120 15 1086 1083 1084 1089 1089 20 1032 1030 1037 1047 1050 25 964 962 968 977 978 30 908 909 919 929 932 35 836 838 847 855 854 40 740 740 750 763 767 45 648 651 667 686 695 50 560 564 584 611 621 55 471 479 504 532 543 60 384 396 426 456 466 65 301 315 349 376 385 70 219 235 270 294 305 75 140 162 194 220 232 80 70 96 129 157 171 85 21 40 70 86 95 90 0 0 0 0 0

Page 18: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 18 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) Illumination Plots

Mounting Height: 10 ft. Illuminance - Cone of Light Isoillumination Plot

Zonal Lumen Summary and Percentages at 25°C

Zone Lumens % Luminaire 0-30 862.2 2.0% 0-40 1389.0 46.8 0-60 2357.5 79.4 60-90 611.1 20.6 0-90 2968.6 100

90-180 0.0 0 0-180 2968.6 100

Page 19: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 19 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) In-Situ Maximum Measured LED Source Temperature Manufacturer Supplied Documentation: LED identified as: Samsung 5630

Page 20: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 20 of 22 Date: January 26, 2015

RESULTS OF TESTS (cont’d) In-Situ Maximum Measured LED Source Temperature Maximum Junction Temperature from LED specification (Tj) = 110°C Thermal Resistance Formula from LED specification = 16°C/W Maximum Forward Voltage (Vf) from LED specification = 3.2V Measured LED Current = 54.8mA Calculated LED Wattage = Vf x Measured LED Current = 0.1754W Maximum Source Temperature (Ts) = Tj – (LED Wattage x Thermal Resistance) = 107°C Maximum Measured Manufacturer Designated Source Temperature

Intertek Sample No.

Maximum Measured Source Temperature (°C) Location

Test Box Used

Maximum Rated Source Temperature(°C)

DL-R22-2035-4K0-002-100

49.3 Per diagram above Non-IC

Recessed 107

In-Situ Picture – Ts

In-Situ Picture – Ts location

Page 21: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 21 of 22 Date: January 26, 2015

Pictures (not to scale) Clear lens

Translucent lens

Page 22: DLC REPORT ACCREDITATION UNDER NVLAP - ggitc.comggitc.com/abc/Home Page_Renewal/innovation/101977042CRT-001.pdfdlc report 3933 us route 11 cortland, new york 13045 project no. g101977042

Report No. 101977042CRT-001 page 22 of 22 Date: January 26, 2015

CONCLUSION

The results tabulated in this report are representative of the actual test samples submitted for this report only. The data is provided to the client for further evaluation. Compliance to the referenced specification requirements was not determined in this report. In Charge Of Tests: Report Reviewed By:

Jacki Swiernik David Ellis Staff Engineer Senior Project Engineer Lighting Division Lighting Division Attachment: IES FILES: DL-R22-A-2035-4K0-001-100 DL-R22-A-2035-4K0-002-100 DL-R22-A-3035-4K0-001-100 DL-R22-A-3035-4K0-002-100 DL-R22-A-2535-4K0-001-IR DL-R22-A-2535-4K0-002-IR