digital indicator user’s manualdacell.com/.../8087af52d985f2bc153b87f00ee1cba5.pdf · digital...

31
DIGITAL INDICATOR USER’S MANUAL DN-130T (VER 4.0) DN-130L (VER 4.0) DN-230A (VER 1.0) DACELL CO.,LTD. Address : 681-1 Cheoksan-Ri, Nami-Myeon, Cheongweon-Gun, Chung-buk TEL :82-43-260-2242 FAX : 82-43-260-2245 http://www.dacell.com E-Mail : [email protected]

Upload: others

Post on 19-May-2020

16 views

Category:

Documents


0 download

TRANSCRIPT

DIGITAL INDICATOR

USER’S MANUAL DN-130T (VER 4.0) DN-130L (VER 4.0) DN-230A (VER 1.0)

DACELL CO.,LTD.

Address : 681-1 Cheoksan-Ri, Nami-Myeon, Cheongweon-Gun, Chung-buk

TEL :82-43-260-2242

FAX : 82-43-260-2245

http://www.dacell.com

E-Mail : [email protected]

Contents

1. Sepcial Features ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 2

1-1. High-speed conversion method ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 2

1-2. Calibration method ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 2

1-3. HOLD & PEAK HOLD ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 2

1-4. Comparison Output ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 2

1-5. Data Back-up ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 2

1-6. Watch dog ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 2

2. Attention ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 3

2-1. Attention for installation ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 3

2-2. Attention for use ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 3

3. Specification ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 4

4. Front Panel ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 5

5. Rear Panel ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 6

6. Wiring Diagram ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 8

7. Components & Function ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 10

7-1. How to use Hold mode ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 10

7-2. How to use comparison output function ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 11

8. Setting mode ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 12

8-1. Types of Setting mode & Set-up ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 12

8-2. Function mode (Function Set-up) ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 13

1) How to set up function ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 13

2) How to set up Relay output ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 14

8-3. Digital calibration (Calibration by sensor output value) ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 20

1) DN-130T, DN-230A ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 20

2) DN-130L ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ 21

8-4. Actual load calibration (Real-weight Calibration) ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 22

1) DN-130T, DN-230A ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 22

2) DN-130L ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ 23

8-5. SPAN constant calibration (Calibration by Shunt CAL value) ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 24

1) DN-130T, DN-230A ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 24

2) DN-130L ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ 25

8-6. Data Back-up & Restore ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ 26

8-7. Lock Set-up ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 27

9. Product Inspection ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 28

10. OPTION

Option 02 (RS232C) ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 29

Option 03 (RS485) ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 30

1

1. Special Features

We are grateful to you for picking up our product. If you experience problems while

using it, you may refer to the user manual or contact the Technical Support Department

of the company.

This product is an indicator which displays Torque and RPM in digital mode and it

has the following features.

1-1. High-speed Conversion Method

24bit High-speed A/D converter that can detect the sensor’s input signal at a rate of

1000times per second is used. For analog output, 16bit D/A converter with 1000times per

second is used so that there is almost no devuatuib betweeb display value and output value.

1-2. Calibration Method

2 types of calibration systems are adopted. Calibration by actual load (standard weight) and

calibration by rated output of sensor (mV/V).

1-3. Hold and Peak Hold

You can choose Peak hold or Sample hold as per your need.

1-4. Comparison Output

You can do set-up the maxium and lower limit value by the keys on the front panel and output

contact point signal of the rear panel.

1-5. Data Back-up

All the set-up values will be memorized on the flash memory so the inputted data can be

saved and no need to do the re-setting even in case of the interruption of the electric power or

power disconnection.

1-6. Watch dog

This function is for automatic reset in case the system is stopped due to the external factors

such as noise.

2

2. Attention

For the efficient and safe use, please carefully read and be fully aware of the following

details before using this product.

It is strictly forbidden to use this product for any other purpose of use or to attempt to

make any alteration on this product.

2-1. Attention for installation

• Please keep it out of wet places.

• Do not set it up near vibration & impulse, high temperature and humidity.

Keep it out of the direct rays of the sun. Set it up where there is less dust, and Keep

it out of direct air including salt and ion.

• Do not use when there is inflammable gas or heavy machinery, and smog.

• Ground earth-terminal ( ).

• Make wire separately from power system wiring and noise wiring.

• Make sure the use of 4 line sealed cable as a sensor cable. Too long cable leads to

measurement error due to wiring resistance (around 10meters).

2-2. Attention for use

During calibration, Do not input free-load state and real-weight load until it becomes

stable. Pressing Enter Key in unstable condition leads to calibration error.

Do not press any Key in use at one's discretion. Please refer to 7. Setting-up mode for

the function and method of Key.

3

3. Specification

SPEC. Model DN-130T DN-130L DN-230A

SIGNAL CH1 : 0.5~3mV/V

CH2 : PULSE

CH1,CH2 :

0.5~3mV/V

CH1 : 0~2Vrms

CH2 : PULSE

ANALOG OUT DC 0 ~ 10V, 4 ~ 20mA(in accordance with customer request)

EXCITATION CH1 : 5V

CH2 : 12V CH1,CH2 : 5V

CH1 : 2Vrms/5kHz

CH2 : 12V

MAX. Display CH1: -19999~+99999

CH2: 0 ~ 60000

CH1,CH2: -

19999~+99999

CH1 : -19999~+99999

CH2 : 0 ~ 60000

A/D Converter 24bit 1000time/sec

D/A Converter 16bit, 1000time/sec

Temper

ature

Charact

eristic

ZERO 0.5μV/℃

SPAN 50ppm/℃

Measured

value display 7 segment LED 14mm

Status display Red LED 6 PCS

Key Switch 4 PCS

Comparison

output Upper limit(HI), Lower limit(LO)

Contact

Capacity AC 125V 0.5A

Temperature

Range -10℃∼60℃, less than 80% RH (no dewing.)

Weight About 650g

Power Supply AC 90 ~ 240V, 50/60㎐

OPTION OP-02 : RS232C, OP-03 : RS485

External

dimension 97 × 97 × 141 mm (W × H × D)

Panel cutting

size 92.5 × 92.5 mm

4

4. Front Panel

① HI, LO Indication LED

: Hi or LO LED will be lighted when measured value exceeds the setting value.

② HOLD indication LED : This LED will be lighted when the measured value is on Hold.

③ Measured value Indication : It indicates the measured value and each setting value.

Measruing Mode : Once this Key is pressed for more than 1 second, the current

measuring value will be Zero (0) and the Analog output will be 0V (4mA) as well.

SET-UP Mode : Once this is pressed on the Function Set-up Mode, you will

return to the measuring mode.

Measuring Mode : Once this Key is pressed, the lower limit setting value will be

displayed and this value also can be changed.

SET-UP Mode : The location of row for the number flickering can be moved.

Measuring Mode : Once this Key is pressed, the upper limit setting value will be

displayed and the value can be changed.

SET-UP Mode : The flickering number will be increased by 1 and 1.

Measuring Mode : Once this Key is pressed, HOLD will be selected and once this

key is pressed again, HOLD will be cancelled.

SET-UP Mode : Save each set-up value..

5

5. Rear Panel

Please check the location of terminal and its use. While the projected button is being

pressed, please insert the cable into the lower hole completely. As soon as you release

the button, the connection will be completed. At the point, please slightly pull the calbe

and check whether the cable is come off or not.

(The most suitable calbe is Φ 0.5~1. Linking cable must be brazing or used with 1

terminal)

① AC_L, N : Main Power Supply Wiring Terminal

② : Ground Terminal (as an independent ground connection.)

③ E1+ : CH1 Sensor Supply Voltage + Connection Terminal(FED)

④ E1- : CH1 Sensor Supply Voltage - Connection Terminal(White)

⑤ S1+ : CH1 Sensor Ouput Signal + Connection Terminal(Black)

⑥ S1- : CH1 Sensor Output Signal – Connection Terminal(Green)

⑦ GND : SHIELD connection terminal of sensor (Tare)

⑧ E2+ : CH2 Sensor Supply Voltage + Connection Terminal(Red)

⑨ E2- : CH2 Supply Voltage - Connection Terminal(White)

⑩ S2+ : CH2 Sensor Ouput Signal + Connection Terminal(Black)

⑪ S2- : CH2 Sensor Ouput Signal - Connection Terminal(GREEN)

6

⑫ GND : SHIELD connection terminal of sensor(Tare)

⑬ OUT1, 2 : Analog (DC 0 ~10V/DC 4 ~ 20mA) Output Terminal

⑭ TXD, RXD : RS232C SERIAL INTERFACE (RS485 : TXD → TX+, RXD → TX-)GND

: External Input Common Terminal

⑮ IN1 : External HOLD Input Terminal (CH1)

⑯ IN2 : External ZERO Input Terminal (CH1)

○17E

A IN3 : External HOLD Input Terminal (CH2)

A○18E

A IN4 : External ZERO Input Terminal (CH2,DN130L)

A○19E

A COM : RELAY Output Common Terminal

A○20E

A RY1 : CH1 Lower Limit Output Terminal

A○21E

A RY2 : CH1 Upper Limit Output Terminal

A○22E

A RY3 : CH2 Lower Limit Output Terminal

A○23E

A RY4 : CH2 Upper Limit Output Terminal

◀Attention▶

1. When wiring, please pull out the power plug. (Main power supply)

2. Please use the thick calbe for ground terminal ( ) so that you can avoid any trouble

due to the impulse voltage or surge. If possible, please keep it as an independent

ground.

(This must be used in the area with heavy noise. If the ground is made with other

device, it will be affected by noise.)

3. Please check the function of the terminal before you do the wiring so that you can

prevent any malfunction in advance.

4. Manufacture will be relieved of its responsibility for any damge or injury due to the

disassembly and alteration made without any agreement. Also, no A/S will be made.

7

5. Rear Panel

Please check the location of terminal and its use. While the projected button is being

pressed, please insert the cable into the lower hole completely. As soon as you release

the button, the connection will be completed. At the point, please slightly pull the calbe

and check whether the cable is come off or not.

(The most suitable calbe is Φ 0.5~1. Linking cable must be brazing or used with 1

terminal)

1) DN-130T

2) DN-130L

8

3) DN-230A

1) <Example - External Input PLC Connection>

9

7. Components & Function

7-1. How to use Hold mode

Hold mode is largely divided into Peak Hold and Sample Hold. Please select Analog,

Digital or Display Hold as per your purpose of use (Hi or Low Speed)

To input Hold, you can use Hold key on the front panel or external input. For the

operation method, please refer to the drawing below.

1) Peak Hold : To Hold the maxium value among the measured values.

Peak hold mode : It is a high-speed Peak hold. It is to hold and display the

maximum value of (+) direction.

Absolute peak hold mode : It is to hold and display the the maximum value of the

absolute value (+/-).

2) Sample hold : It is to hold and display the value at the time of Hold signal input

among the measured values.

On

Off

On

Off

On

off

① − : sensor input value

② − : Display value &

analog output value

Front panel HOLD LED

External HOLD input termanal

Front panel HOLD Key

Measure

d va

lue

time

Measure

d va

lue

time

<Peak HOLD> <Sample HOLD>

10

7-2. How to use comparison ouput function

For comparison output function, there are 3 different modes such as Decision, High

limit, Low limit. It displays through the relay of rear panel comparing each setup value.

On High limit and Low limit mode, Hysteresis can be used.

To set up upper limit (High) and lower limit(Low), please use the key on the front panel.

1) Decision mode : Measured Value ≤ Lower limit setup value ⇒ CH1 : RY1 ON, CH2 : RY3 ON

Measured Value ≥ Upper limit setup value ⇒ CH1 : RY2 ON, CH2 : RY4 ON

2) High limit mode : Measured Value ≥ RY1 setup value ⇒ CH1 : RY1 ON, CH2 : RY3 ON

Measured Value ≥ RY2 setup value ⇒ CH1 : RY2 ON, CH2 : RY4 ON

Measured Value < RY1 setup value – Hysteresis value⇒ CH1:RY1 OFF, CH2:RY3 OFF

Measured Value < RY2 setup value - Hysteresis value⇒ CH1:RY2 OFF, CH2:RY4 OFF

3) Low limit mode : Measured Value ≤ RY1 setup value ⇒ CH1 : RY1 ON, CH2 : RY3 ON

Measured Value ≤ RY2 setup value ⇒ CH1 : RY2 ON, CH2 : RY4 ON

Measured Value > RY1 setup value + Hysteresis value⇒ CH1:RY1 OFF, CH2:RY3 OFF

Measured Value > RY2 setup value + Hysteresis value⇒ CH1:RY2 OFF, CH2:RY4 OFF

11

8. Setting Modes

8-1. Types of Setting mode & Set-up

For setting mode, there are 4 different types of mode such as Function mode, Digital

calibration mode, Actual load calibration mode and SPAN constant calibration mode.

Mesuring mode

While key is being

Press, press key

Function mode

END

Digital calibration

mode

END

END

Actual load

calibration mode

① Function mode

Access to each function setup mode. Please refer 8-2 function mode.

② Digital calibration mode

It is to calibrate into the sensor’s output value. No need to prepare for the actual

load (stand weight). Please refer 8-3 how to calibrate.

③ Actual load calibration mode

It is to calibrate by adding the actual load (standard weight or the load you know).

Please refer 8-4 how to calibrate.

④ SPAN constant calibration mode

It is to calibrate with the S.CAL value written down for load calibration. Please

refer 8-5 how to calibrate.

END

SPAN constant

calibration mode

12

8-2. Function mode

DN-130T, DN-230A : CH1 (TORQUE), CH2 (RPM)

DN-130L : CH1 (LOAD CELL), CH2 (LOAD CELL)

1)How to set function.

Measuring mode

While key is being, Press, press key

END

13

3) How to set RELAY data

key : Lower limit setup key

key : Upper limit setup key

Measuring mode

Press key or press key

END Input value

14

# Function mode list #

Name Function Setting The standard

set-up value

F-01 Decimal point 0, 1, 2, 3 1

F-02 Division 1, 2, 5, 10, 20, 50 1

F-03 Display filter 0, 4, 8, 16, 32 8

F-04 Hold mode Sample hold, Peak hold, Absolute peak hold Peak hold

F-05 Comparison mode Decision, High limit, Low limit Decision

F-06 Hysteresis 0 ~ 99 0

F-07 Pulse/circle 0 ~ 360 60

F-08 DAC capacity -19999 ~ +99999 10000

F-09 ID Number 0 ~ 32 0

F-10 Baud rate & PRINT 2400, 4800, 9600 9600

F-11 Auto zero tracking 0 ~ 99 0

F-12 Auto zero tracking time 0.0 ~ 5.0 sec 0.0

F-13 Force unit Kg(kg/㎠), N, ℓb, bar, MPa Kg

F-14 Key disabling Zero key, Lo key, Hi key, Hold key 0000

F-01. Decimal point (Decimal point set up)

(Standard setup value: 1)

Display data Setting Remark

0 00000 : No decimal point DN-130T,DN-230A

: Only set up CH1 .

DN-130D

: Set up CH1 and CH2

1 0000.0 : One decimal place

2 000.00 : Two decimal places

3 00.000 : Three decimal places

15

F-02. Division (Minimum display unit set up)

(Standard setup value: 1)

Display data Setting Remark

1 Displayed In 1 (0, 1, 2, 3, 4 …….)

CH1, CH2

Individual setup

2 Displayed in 2 (0, 2, 4, 6, 8 …….)

5 Displayed in 5 (0, 5, 10, 15 …….)

10 Displayed in 10 (0, 10, 20, 30 …….)

20 Displayed in 20 (0, 20, 40, 60 …….)

50 Displayed in 50 (0, 50, 100, 150 …….)

F-03. Display filter (Display speed set up)

(Standard setup value: 8)

Display data Setting Remark

0 No filter

Only CH1

4 Average time 1/8 sec

8 Average time 1/4 sec

16 Average time 1/2 sec

32 Average time 1 sec

F-04. Hold mode

(Standard setup value: 1)

Display

data Setting Remark

0 Sample Hold : To hold the display value at the time of Hold signal

input.

CH1

setup

1 Peak Hold : To hold the maximum value of display values during

Hold signal input.

2 Absolute Peak Hold : To hold the maximum absolute value of display values

during Hold signal input.

F-05. Comparision mode (Comparision output mode set up)

(Standard setup value: 0)

Display data Setting Remark

0 Decision mode : Relay RY1, RY2, RY3, RY4 output CH1, CH2

Individual

setup

1 High limit mode : Relay RY1, RY2, RY3, RY4 output

2 Low limit mode : Relay RY1, RY2, RY3, RY4 output

16

F-06. Hysteresis

(Standard setup value: 00)

Display data Setting Remark

00 00 : Hysteresis – not used CH1,CH2

Individual

setup

~

99 01 ~99 : Hysteresis – used(Decision mode is not applied)

F-07. Pulse/circle (Set up pulse per circle)

(기준설정값 : 60)

Display data Setting 비고

01 ~ 360 Set up pulse per circle DN-130T,DN-230A

: Set up CH2

F-08. DAC capacity (Analog Outupt value set up)

(Standard setup value: 10000)

Display

data

Setting Remark

-19999 Analog output capacity set-up

<Setting value & output>

Setting

value

Voltage output(±10V) Current output(4 ~20mA)

Display value output Display value output

+10000

-10000 -10V -10000 -

0 0V 0 4mA

+10000 +10V +10000 20mA

-10000

-10000 +10V -10000 20mA

0 0V 0 4mA

+10000 -10V +10000 - 1

CH1, CH2

Individual

setup

~

+99999

F-09. ID Number (Communication Device Number set up)

(Standard setup value: 00)

Display data Setting Remark

00 00 : Device number is not set-up(Stream mode:always transmit data )

CH1

setup

~

32 01~32 : Device number is set-up(Command mode:Transmit data by command)

CH2 ID Number : CH1 ID Number + 1

17

F-10. Baud rate & Print (Communication Speed set up)

(Standard setup value: 9.60)

Display data Setting Stream mode Command mode Remark

2.40 2400 bps O O CH1

setup 4.80 4800 bps O O

9.60 9600 bps O O

F-11. Auto zero tracking (zero operation range set up)

(Standard setup value: 00)

Display data Setting Remark

00 00 : Auto zero is not used Only CH1

~

99 01 ~99 : Set up the operation range of auto zero

F-12. Auto zero tracking time (Auto Zero tracking time set up)

(Standard setup value: 0.0)

Display data Setting Remark

00 0.0 : Auto zero is not used

Only CH1

~

5.0 0.1~5.0 : set up the auto zero operation time (0.1~5.0 sec)

F-13. Force unit (Conversion unit set up)

(Standard setup value: 0)

Display data Setting Remark

0 kg.f kg/㎠

Only CH1

1 N kgf × 9.8 - -

2 ℓb kgf × 2.2 - -

3 - - Bar kg/㎠ × 0.98

4 - - MPa kg/㎠ × 0.098

Note) During calibration, set up the unit you want after calibrating it into Kg.

F-14. Key disabling (Front key locking set up)

(Standard setup value: 0000)

Display data Setting Remark

0 0 0 0 Hold key Lock (1), Release (0)

Hi key Lock (1), Release (0)

Lo key Lock (1), Release (0)

Zero key Lock (1), Release (0)

<Set-up example>

1001 : Zero & Hold key Lock, Hi and Lo key Release

CH1 setup

18

8-3. Digital calibration (Calibration by sensor output value)

At the time of purchasing sensor, the rated capacity (R.C) and rated output (R.O)

declared on the calibration sheet can be used for the calibration for easier calibration.

1)DN-130T, DN-230A

Mesuring mode

While key is being press

press key

Display Function mode

Display Sensor output value calibration mode

Display R.O. value input mode

Display R.C. value input mode

Input the rated output value by using and

key. And then. Press key

Calibration end

Input the rated output value by using and

key. And then. Press key

19

2)DN-130L

Mesuring mode

While key is being press,

press key

Display Function mode

Display Sensor output value calibration mode

Select to Channel

Display R.O. value input mode

Display R.C. value input mode

Input the rated output value by using and

key. And then. Press key

Calibration end

Input the rated output value by using and

key. And then. Press key

20

8-4. Actual load calibration

This is a calibration method by adding actual load on the sensor. Standard weight is needed.

1) DN-130T, DN-230A

Measuring mode

Display Function mode

Display Sensor output value calibration mode

Display Actual load calibration mode

Display actual load mode

Input the rated output value by using and key. And

then. Press key

Calibration end

Display zero mode. press key without any load on the

sensor

About 1 sec later

After adding the actual load (standard weight) on the

sensor, press key.

About 1 sec later

Display R.O. value input mode

Input the rated output value by using and key. And then.

Press key (Input 3.0000 if do not know the R.O. value)

Display Shunt CAL value (please write to S.CAL)

While key is being Press, press key

Cal. Date R.O. value calibration constant

21

2) DN-130L

Measuring mode

Display Function mode

Display Sensor output value calibration mode

Display Actual load calibration mode

Display actual load mode

Input the rated output value by using and key.

And then. Press key

Calibration end

Display zero mode. press key without any load on the

sensor

About 1 sec later

After adding the actual load (standard weight) on the

sensor, press key.

About 1 sec later

Display R.O. value input mode

Input the rated output value by using and key. And then.

Press key (Input 3.0000 if do not know the R.O. value)

Display Shunt CAL value (please write to S.CAL)

Channel Date R.O. Value Cal. value

CH1

CH2

While key is being Press, press key

Select to channel 1 or 2

22

8-5. Calibratuib by Shunt CAL Value

It is to calibrate with the S.CAL value written down for load calibration. You can

calibrate without any standard weight.

1) DN-130T, DN-230A

Measuring mode

While key is being Press, press key

Display Function mode

Display Sensor output value calibration

Display Actual load calibration indicate

Input the rated output value by using and key. And

then. Press key

Calibration end

Display Shunt cal value calibration mode

Display R.O. value input mode

Display Shunt cal value input mode

Input Shunt cal value by using and key. And then.

Press key

23

2) DN-130L

Measuring mode

While key is being Press, press key

Display Function mode

Display Sensor output value calibration

Display Actual load calibration indicate

Input the rated output value by using and key. And

then. Press key

Calibration end

Display Shunt cal value calibration mode

Select to Channel 1 or 2.

Display R.O. value input mode

Display Shunt cal value input mode

Input Shunt cal value by using and key. And then.

Press key

24

8-6. DATA BACK-UP & RESTORE

You can save all the set-up values of the device and then restore them to the current

set-up state as per your need.

◦ DATA BACK-UP : Save the current set-up state.

◦ RESTORE : Restore the current set-up state.

Measuring mode

While key is being press,

press key

About 1 sec later

Display

Choose yes or no by using key.

YES : DATA back-up. NO : DATA back-up no.

1) DATA BACK-UP

Power off

While key being press

power On

About 1 sec later

Display

Choose yes or no by using key.

YES : DATA restore. NO : DATA restore no.

2) RESTORE

25

8-7. Lock Set-up

You can prevent any accidental operation due to the unnecessary key control by Lock

set-up. After finishing calibration, it is recommended to set the Lock.

At the first stage, please start while the power is OFF.

Related Function when Lock is set : Function related to calibration

Power off

While key is being press,

Power on

Please keep pressing key while this is being displayed

or

Setup end

Please select lock or release by pessing key

ON : Setting is impossible. Off : Setting is possible

26

9. Product Inspection

Symptom Cause Action Remark

When Display

trembles.

• Load cell is damaged.

• Load cell insulation

resistance.

• Indirect occurrence

• Load cell input, output.

• Check resistance

• Check load cell’s

insulation resistance.

• Insulation

resistance

(Cable & Case >

1000 Mohm)

When weight

goes up at a

regular ratio or

zero returns are

not made.

• Loadcell faulty • Check load cell’s

insulation resistance.

• Loadcell connection is

insufficient.

• Check the wiring

between load cell and the

main device.

• Check the load cell’s

calbe’s disconnection.

When weight

changes into (-).

• Loadcell wiring is

reversed.

• Check load cell’s ouput

cable connection.

• Output : (+SIG)

(-SIG)

Displayed as

“OVER” or

“UNDER”

• Load cell is damaged.

• Load cell connection is

bad.

• Check the load cell’s

condition and calbe

connection.

27

10. OPTION

#Option-02 (RS232C)

Since RS232C Interface is very sensitive of electric noise. So please do the wiring

from AC Power and electric wires separately. Also you must use the shield calbe always

Indicator Host PC

TX(Transmission Data), No.3 Pin RXD(Receive Datea), No.2 Pin

RX(Receive Data), No.2 Pin TXD(Transmission Data), No.3 Pin

GN(Ground), No.1 Pin GND(Ground), 5 Pin

1. TYPE : EIA-232C

2. Method : Half-duplex, asynchronous method.

3. Baud-rate : Select one of 2400, 4800, 9600, 19200bps

4. Parity : No Parity

5. Data bit : 8 bit

6. Stop bit : 1 bit

7. Stream mode (Ex. Data +1234.5 transmission)

0BCODE BYTE1 BYTE2 BYTE3 BYTE4 BYTE5 BYTE6 BYTE7 BYTE8

1BASCII S 1 , N T , + 0

HEX 53H 31H 2CH 4EH 54H 2CH 2BH 30H

CODE BYTE9 BYTE10 BYTE11 BYTE12 BYTE13 BYTE14 BYTE15 BYTE16

ASCII 1 2 3 4 . 5 CR LF

2BHEX 31H 32H 33H 34H 2EH 35H 0DH 0AH

1) BYTE1 : fixed character (S)

2) BYTE2 : Channel character (CH1 : 1, CH2 : 2)

3) BYTE3 ~ BYTE6 : fixed character (, N T ,)

4) BYTE7 ~ BYTE14 : DATA 8 BYTE(including +/)

5) BYTE15 : CARRIAGE RETURN

6) BYTE16 : LINE FEED

8. Command mode

OP-03 : Refer to RS485

28

#Option-03 (RS485)

Since RS485 Interface is very sensitive of electric noise. So please do the wiring from

AC Power and electric wires separately. Also you must use the shield calbe always.

1. TYPE : RS485

2. Method : Half-duplex, asynchronous method.

3. Baud-rate : Select one of 2400, 4800, 9600bps

4. Parity : No Parity

5. Data bit : 8 bit

6. Stop bit : 1 bit

Please set up the device No. referring to INDICATOR Manual.

(Can set up from 1 to 32 channels.)

7. Command form (PC -> INDICATOR)

3BCODE BYTE1 BYTE2 BYTE3 BYTE4 BYTE5

4BASCII I D 0 1 P

HEX 49H 44H 30H 31H 50H

1) BYTE1, BYTE2 : Fixed character (ID)

2) BYTE3, BYTE4 : device number (1 ~ 32)

3) BYTE5 : command order (P, H, R, Z)

8. Command chart

Command

ASCII HEX

P 50H Transmit the current value of order equipment.

H 48H Hold for order equipment.

R 52H Release hold for order equipment.

Z 5AH Operate the current value of order equipment as ZERO.

29

9. Transmission DATA form (INDICATOR -> PC)

CODE BYTE1 BYTE2 BYTE3 BYTE4 BYTE5 BYTE6 BYTE7 BYTE8

ASCII I D 0 0 1 , + 0

5BHEX 53H 54H 30H 30H 31H 2CH 2BH 30H

CODE BYTE9 BYTE10 BYTE11 BYTE12 BYTE13 BYTE14 BYTE15 BYTE16

ASCII 1 2 3 4 . 5 CR LF

6BHEX 31H 32H 33H 34H 2EH 35H 0DH 0AH

1) BYTE1, BYTE2 : Fixed character (ID)

2) BYTE3 ~ BYTE5 : Device number (1 ~ 32)

3) BYTE6 : Fixed character (,)

4) BYTE7~BYTE14 : DATA 8byte (including +/-)

5) BYTE15 : CARRIAGE RETURN

6) BYTE16 : LINE FEED

30