Digital FLUORO
Summer 2008
![Page 2: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/2.jpg)
DIGITAL FLUORO
![Page 3: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/3.jpg)
• Digital fluoroscopy is currently most commonly configured as a conventional fluoroscopy system in which the analog video signal is converted to a digital format with an analog-to-digital converter (ADC). An in-depth discussion of digital detector technologies (eg, flat-panel "direct" detection of x rays and charge-coupled device technology) is beyond the scope of this article. After a review of several fundamental digital imaging concepts including binary numbers, pixels, and gray levels, emphasis will be placed on discussions of the digital imaging tools specific to digital fluoroscopy and digital subtraction angiography (DSA).
![Page 4: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/4.jpg)
• Digital fluoroscopy is currently most commonly configured as a conventional fluoroscopy system in which the analog video signal is converted to a digital format with an analog-to-digital converter (ADC).
• With discussions of the digital imaging tools specific to digital fluoroscopy and digital subtraction angiography (DSA).
![Page 5: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/5.jpg)
Early Fluoroscopy
![Page 6: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/6.jpg)
CONVENTIONAL FLUOROSCOPYINVENTED BY THOMAS EDISON
![Page 7: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/7.jpg)
Direct Fluoroscopy: obsolete
In older fluoroscopic examinations radiologist stands behind screen and view the pictureRadiologist receives high exposure; despite protective glass, lead shielding in stand, apron and perhaps goggles
Main source staff exposure is NOT the patient but direct beam
![Page 8: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/8.jpg)
![Page 9: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/9.jpg)
![Page 10: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/10.jpg)
pixels• the smallest element of a digital image.• A digital image is normally composed of a two-
dimensional (square) matrix of pixels.• The matrix size of an image is used to describe
the number of pixels in each row and column of the image.
![Page 11: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/11.jpg)
The size of a pixel determines the smallest detail visible in the image,
the number of pixels (of a given size) determines the field of view of the image. pixel value is assigned to a certain color or gray level for visualization. The imaged object is more faithfully reproduced as matrix size increases and pixel size decreases.
![Page 14: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/14.jpg)
Digital fluoroscopy• is most commonly configured as a conventional
fluoroscopy system (tube, table, image intensifier, video system) in which the analog video signal has been digitized with an ADC.
• Alternatively, digitization may be accomplished with a digital video camera (eg, a charge-coupled device) or via direct capture of x rays with a flat-panel detector.
• For digital fluoroscopy systems in which the analog video signal is digitized with an ADC, the resolution is limited by the resolution of the video camera, which is typically 1–2 line pairs per millimeter.
![Page 15: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/15.jpg)
![Page 16: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/16.jpg)
![Page 17: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/17.jpg)
![Page 18: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/18.jpg)
![Page 19: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/19.jpg)
Digital Image recording
• In newer fluoroscopic systems film recording replaced with digital image recording.
• Digital photospots acquired by recording a digitized video signal and storing it in computer memory.
• Operation fast, convenient.• Image quality can be enhanced by application
of various image processing techniques, including window-level, frame averaging, and edge enhancement.
• But, the spatial resolution of digital photospots is less than that of film images.
![Page 20: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/20.jpg)
Digital radiography principle
Clock
MemoryADCI
Iris
t
t
ANALOGUESIGNAL
DIGITALSIGNAL
See more in Lecture L20
![Page 21: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/21.jpg)
Digital spot film images
• and photospot images may be acquired by using the same digital fluoroscopy system.
• Individual frames from a digital fluoroscopy sequence can be stored digitally and can be used instead of conventional spot film and photospot images.
• Digital photospot images will have the same characteristics (eg, resolution) as digital fluoroscopic images.
![Page 22: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/22.jpg)
Digital format – print to film
![Page 23: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/23.jpg)
Remote – over the table tube
![Page 24: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/24.jpg)
Different fluoroscopy systems
• Remote control systems– Not requiring the presence of
medical specialists inside the X Ray room
• Mobile C-arms– Mostly used in surgical
theatres.
![Page 25: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/25.jpg)
![Page 26: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/26.jpg)
![Page 27: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/27.jpg)
![Page 28: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/28.jpg)
![Page 29: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/29.jpg)
![Page 30: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/30.jpg)
• The digital image data from digital fluoroscopy may be processed by using many useful image processing techniques.
• These techniques may serve to decrease radiation exposure to the patient and medical imaging staff or improve visualization of anatomy.
• Processing options include last image hold, gray-scale processing, temporal frame averaging, and edge enhancement.
• Additional processing is available when digital fluoroscopy data are used to perform DSA.
![Page 31: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/31.jpg)
![Page 32: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/32.jpg)
![Page 33: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/33.jpg)
![Page 34: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/34.jpg)
![Page 35: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/35.jpg)
![Page 36: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/36.jpg)
![Page 37: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/37.jpg)
![Page 40: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/40.jpg)
![Page 41: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/41.jpg)
RADIATION PROTECTION
STAT CH 7 & 8
DHS-RHB
RAD PROT/FLUORO SYLLABUS
![Page 42: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/42.jpg)
Radiation Safety and Fluoroscopy
Time Distance Shielding
Also see Intro to Fluoro LectureFrom RT 124 Spring 2007
![Page 43: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/43.jpg)
Patient Protection
• Tabletop exposure rate– Maximum 10 R/min– Typically 1 – 3 R/min
![Page 44: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/44.jpg)
Patient Protection
• Minimum source-to-skin distance– 12” for mobile equipment– 15” for stationary systems
• Audible alarm at 5 mins.
• Same rules for collimation
![Page 45: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/45.jpg)
Patient Protection
• Typical exposure rates– Cinefluorography
• 7.2 R/min
– Cassettes• 30 mR/exposure
– 105 mm film• 10 mR/exposure
![Page 46: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/46.jpg)
Protection of Radiographer and Radiologist
• Lead apron– 0.25 mm Pb/eq
• Highest energy scatter– 90o angle to the incident beam– Same level as radiographer /radiologist’s
gonads
![Page 47: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/47.jpg)
The Patient & Scatter
![Page 48: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/48.jpg)
WHATS WRONG?
![Page 49: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/49.jpg)
![Page 50: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/50.jpg)
Remote – over the table tube
![Page 51: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/51.jpg)
Under table tube vs Over table tube (remote units)
Need moving shield for Remote rooms
![Page 52: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/52.jpg)
IsoExposure CurvesWhere is it SAFE??
![Page 53: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/53.jpg)
ISOEXPOSURE CURVES
![Page 54: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/54.jpg)
Patient entrance skin exposure (ESE) is higher when the
fluoroscopic x-ray tube is too close to the tabletop.
![Page 55: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/55.jpg)
KEEP I.I. CLOSE TO PATIENT
![Page 56: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/56.jpg)
![Page 57: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/57.jpg)
PATIENT EXPOSURE
• REDUCE DISTANCE OF IMAGE INTESIFIER
• INCREASE DISTANCE FROM THE TUBE
![Page 58: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/58.jpg)
![Page 59: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/59.jpg)
PATIENT EXPOSURE
• REDUCE SIZE OF
COLLIMATED BEAM
WHEN POSSIBLE
![Page 60: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/60.jpg)
ESE FOR FLUORO
• TLD PLACED AT SKIN ENTRACE POINT
• 1 – 5 R/MINUTE AVE IS 4 R/MIN
• INTERGRAL DOSE –• 100 ERGS OF TISSUE = 1 RAD EXPOSURE• OR 1 GM RAD = 100 ERGS
![Page 61: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/61.jpg)
1 RAD -100 ERGS – 1 GRAM
1 rad (of ionizing radiation) deposits 100 ergs of energy per gram
![Page 62: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/62.jpg)
DOSE REGULATIONS
• BEFORE 1974 - AT TABLETOP
• 5R/MIN (WITHOUT AEC)
• 5R/MIN (WITHOUT AEC) – BOOST MODE
• After 1974 with AEC
• 10 R/MIN 20R/MIN BOOST
![Page 63: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/63.jpg)
SSD – TUBE TO SKIN DISTANCE
• FIXED UNITS
• 18” PREFERRED
• 15 “ MINIMUM
MOBILE UNITS ( C-ARMS)
• 12’ MINIMUM
![Page 64: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/64.jpg)
Conventional “FIXED” Fluoro
![Page 65: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/65.jpg)
![Page 66: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/66.jpg)
Mobile – C- arm fluoro
![Page 67: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/67.jpg)
![Page 68: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/68.jpg)
![Page 69: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/69.jpg)
![Page 70: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/70.jpg)
![Page 71: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/71.jpg)
![Page 72: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/72.jpg)
![Page 73: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/73.jpg)
![Page 74: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/74.jpg)
FLUOROSCOPY –
ABOVE AND BELOW THE TABLE TUBE EXPOSURE TO TECH
![Page 75: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/75.jpg)
![Page 76: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/76.jpg)
![Page 77: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/77.jpg)
Pulsed Fluoro
• Some fluoroscopic equipment is designed for pulsed-mode operation. With the pulsed mode, it can be set to produce less than the conventional 25 or 30 images per second. This reduces the exposure rate.
• Collimation of the X ray beam to the smallest practical size and keeping the distance between the patient and image receptor as short as possible contribute to good exposure management.
![Page 79: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/79.jpg)
Dose rate to patients
![Page 80: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/80.jpg)
![Page 81: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/81.jpg)
![Page 82: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/82.jpg)
Mini c-arm
![Page 83: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/83.jpg)
![Page 84: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/84.jpg)
![Page 85: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/85.jpg)
![Page 86: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/86.jpg)
![Page 87: Digital fluoro](https://reader033.vdocuments.site/reader033/viewer/2022061301/54d61cb04a795979598b4713/html5/thumbnails/87.jpg)