Devices senza cateteri transvenosi. Tra realtà consolidata:
ICD sottocutaneo e prospettive future:
PM intracavitario Dr. Sebastiano Lumera
U.O.C . Cardiologia UTIC
Ospedale “R. Guzzardi “ Vittoria
Direttore G.V. Lettica
1957
PACEMAKER
ESTERNO
1958
PACEMAKER
IMPIANTABILE
1980
PACEMAKER
BICAMERALE
1986
PACEMAKER
RATE-RESPONSE
1998
PACEMAKER
BIVENTRICOLARE
ENDOCARDICO ?
PACEMAKER
BIOLOGICO
2008
HOME
MONITORING
1980
DEFIBRILLATORE
1995
DEFIBRILLATORE
PETTORALE
STORIA DEI DISPOSITIVI
1994
PACEMAKER
BIVENTRICOLARE
EPICARDICO
Il punto debole dei Devices transvenosi è rappresentato dall’elettrocatetere. Più ne vengono impiantati e maggiori sono le possibili complicanze
• Dislocazione • Frattura • Infezione • Problemi per l’estrazione • Trombosi venosa • Dissezione del seno coronarico
(CRT) • Perforazione cardiaca • Pneumotorace • Endocardite • Danno valvolare
Il Sistema S-ICD: una realtà…..
Nessun catetere all’interno o sul cuore
Sistema venoso preservato
Impianto semplice chirurgicamente
Puri riferimenti anatomici
In generale senza bisogno di fluoroscopia
Tempo di impianto prevedibile
ALTER
NA
TO
I componenti del Sistema S-ICD
• Generatore di impulsi, SQ-RX™
• Elettrodo Q-TRAK™
• Tunnellizzatore, Q-GUIDE™
• Programmatore, Q-TECH™
• 80J max output
– max 5 shock per episodio
• Forma d’onda bifasica, 50% tilt
• Polarità shock adattiva
– Cambio di polarità in caso di shock inefficace
• ~ 9s tempo carica per 80 J (inizio vita)
• No ATP
• No anti-brady pacing
• Post shock pacing on demand, max 30s, 50 bpm
• Durata stimata batteria: 5 anni.
Defibrillazione efficace • Capacità di erogare adeguata
energia di defibrillazione con un adeguato margine di sicurezza
• Eragare questa energia in un periodo di tempo breve per ridurre il rischio di sincope e permettendo la eventuale risoluzione spontanea dell’aritmia
Alta capacità di sensing VT/TV • Sensing appropriato di eventi ventricolari ad elevata frequenza senza over-sensing su eventi non ventricolari • Adeguata capacità di discriminazione tra eventi VT/VF dalle aritmie SVT.
Sicurezza di un sistema di ICD
Caratteristiche essenziali di un ICD
Clinical Studies Summary: S-ICD S-ICD® System STUDY STUDY TYPE PATIENTS SITES COMMENTS
Feasibility and Proof of Concept Studies
Acute 78 + 49 Study completed: 2005 Published in NEJM
Initial Human Validation Study Chronic Safety and
Performance 6 2
Study completed: 2008 Published in NEJM
CE Clinical Study Chronic Safety and
Performance 55 8
Study completed: 2009 Published in NEJM
START Study Simulation 64 5 Study completed: 2011 Published in J Cardiov. Electrophy.
IDE Clinical Study Chronic IDE 330 33 Study completed: 2011 Presented at HRS 2012 Submitted to Circulation
EFFORTLESS Registry Chronic Post-Market
OUS Registry 436 as of Feb 2013;
Goal: 1,000 28 Still enrolling
The S-ICD system appropriately withheld therapy in 98% of cases
Significant improvement over inappropriate shock rates with conventional ICDs*
START Study
The S-ICD System Algorithm
Specificity Results for the S-ICD® System and Transvenous Systems**
S-ICD System (n = 50)
Transvenous Systems***
Single Chamber (n = 150)
Dual Chamber (n = 147)
Appropriately withheld therapy 49 102 99
Inappropriate Defibrillation 1 48 48
Defibrillation Specificity 98% 68% 67.30%
*Sustained rate timers (i.e. SRD) programmed OFF if nominally active (BSX = 3 min, SJM = 30 secs) **Systems Compared ***Pooled results from 3 manufacturers (Boston Scientific’s Teligen® DR, St. Jude Medical’s Atlas® II + HF, Medtronic’s Secura® VR and Virtuoso® DR) with devices programmed in single chamber or dual chamber mode.
Gold MR, Theuns DA, Knight BP et al. Head-to-head comparison of arrhythmia discrimination performance of subcutaneous and transvenous ICD arrhythmia detection algorithms: The START Study. J Cardiovasc Electrophysiol, 2012;23:359-366
TRA
NSV
ENO
US RA
R
V
SHO
CK
1
2
3
SUB
CU
TAN
EOU
S
Gold et al, JCE 2011
START Study
Sample Shockable Arrhythmia…
TRA
NSV
ENO
US RA
R
V
SHO
CK
1
2
3
SUB
CU
TAN
EOU
S
Gold et al, JCE 2011
START Study
Sample Non-Shockable Arrhythmia…
IDE Clinical Study: Spontaneous Episodes
Evaluation VT/VF Episodes
Total
IDE Study
Episodes 119
Patients 21
All episodes successfully converted with 80J or spontaneously converted
At least 1 year follow up and 304 patients
* Burke M, et al. Safety and Efficacy of a Subcutaneous Implantable-Defibrillator (S-ICD System US IDE Study). Late-Breaking Abstract Session. HRS 2012.
IDE Clinical Study
Primary Safety Endpoint Met
0,0
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
1,0
0 30 60 90 120 150 180
180-Day Device-Related Complication-Free Rate
99.0% (97.9% LCL)
Freedom from Type I
Complication
Post-op Days
180-day Performance Goal (79%)
Freedom from Device-Complications
* Burke M, et al. Safety and Efficacy of a Subcutaneous Implantable-Defibrillator (S-ICD System US IDE Study). Late-Breaking Abstract Session. HRS 2012.
Type I: Device-Related; Type II: Labeling-Related; Type III: Procedure-Related
IDE Clinical Study
Primary Safety Endpoint Met
Type I: Device-Related; Type II: Labeling-Related; Type III: Procedure-Related
0,0
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
1,0
0 30 60 90 120 150 180
180-day Performance Goal (79%)
Freedom from Type I – III
Complication
Post-op Days
Freedom from all Device-, Labeling-, & Procedure-related Complications
180-day Device & Procedure-Related Complication-Free Rate:
92.1% (88.9% LCL)
* Burke M, et al. Safety and Efficacy of a Subcutaneous Implantable-Defibrillator (S-ICD System US IDE Study). Late-Breaking Abstract Session. HRS 2012.
IDE Clinical Study
Infections
• All infections requiring explant occurred early in the study before investigators meeting • No infections requiring explant occurred in
last 214 patients implanted
• No endocarditis or systemic blood stream infections
• 32 out of 33 patients with previous
Transvenous-ICD had no reports of infection with the S-ICD® System
Few Infections Led to Explant
1.3% Explant Rate for Infections
* Burke M, et al. Safety and Efficacy of a Subcutaneous Implantable-Defibrillator (S-ICD System US IDE Study). Late-Breaking Abstract Session. HRS 2012.
IDE Clinical Study
Inappropriate shocks
Reduction of Inappropriate Shocks Associated with Dual Zone Programming
• 38 patients experienced shock due to non-VT/VF event
o 15 patients (4.6%) – SVT> Discrimination zone
o 24 patients (7.4%) – Oversensing
• No patients experienced a shock due to discrimination error in Conditional Shock (dual) zone
• Study demonstrated that dual zone programming is important tool for reducing inappropriate shock rate
Rate of inappropriate therapy is consistent with current transvenous ICDs*
* S-ICD® System Summary of Safety and Effectiveness Data. 2012.PMA P110042.
13,9
10,2
6,1 3,1
Oversensing Rate > Shock Zone
Ina
pp
rop
ria
te
Sh
ocks (
%)
Single Zone Dual Zone
56% Relative
Reduction 70%
Relative Reduction
EFFORTLESS Registry
Objective:
To demonstrate the early, mid-term, and long-term clinical outcome and cost
effectiveness of the S-ICD System
Outcome measures:
• Perioperative (30 days post implant) S-ICD System Complication Free rate
• 360 day S-ICD System Complication Free Rate
• Percentage of Inappropriate Shocks for AF/SVT
• Prospective Quality of Life
Preliminary Results of the EFFORTLESS S-ICD System Registry as of April 27, 2012* • 210 Active Patients • 98.5% effective conversion of induced VT/VF within 1 procedure • No inappropriate shocks have been recorded for AF/SVT within a programmed conditional
shock zone • Annual inappropriate shock rate of 7% with some of these patients receiving shocks due to Rate
> Shock Zone
* Heart Rhythm - May 2012; Vol 9:5(S1-33) AB07-2.
OUS Registry Designed to Capture Clinical Outcome
and Cost Effectiveness Data of S-ICD® System
L’S-ICD E’ ADATTO A TUTTI I PAZIENTI INDICATI PER UN ICD
Tranne coloro che richiedono:
Pacing antibradicardico
Terapia CRT
ATP
Indicazioni a S-ICD
• Giovani e sportivi
• Brugada, sindromi LQT/SQT
• CMP ipertrofica (ostruttiva)
• Precedenti ICD TV espiantati
– Infezione/endocardite
– Malfunzionamento
• Ponte al trapianto
• Sistema venoso non accessibile per ICD TV
– Accesso venoso a sinistra non utilizzabile e accesso venoso a destra non disponibile per dialisi
– Anomala anatomia cardiaca (congenita)
• Elevato rischio di infezione
– Disfunzione renale
– Diabete
Storia
First External
Pacemaker
Synergist
1989
Activitrax
1986
Byrel
1979
5858
1970
5800
1958
Chardack - Greatbatch
1960
MicroMinix
1990
Elite
1991
Kappa®
1998
Thera™
1995
Adapta®
2006
EnPulse
2004
First Implantable
Pacemaker
Pediatric Asynchronous
Pulse Generator
Rate response Radically
smaller size
1st Micro-
processor-based,
Mode switching
Full automaticity
MVP, Full
automaticity
Rate response via activity
& minute ventilation
Dual chamber
rate response
2011
EnRhythmMRI™
1st MRI-
Conditional
Advisa MRI®
2nd MRI-
Conditional
2013
Micra®
Transcatheter
Pacing System
2015
34 MEDTRONIC CONFIDENTIAL – FOR INTERNAL USE ONLY
Leadless Pace Maker: il futuro?.....
The St. Jude Medical Nanostim leadless pacemaker
The Medtronic Micra Transcatheter Pacing System
Caratteristiche dei due devices
The St. Jude Medical Nanostim leadless pacemaker
The Medtronic Micra Transcatheter Pacing System
Conclusioni • Il sistema S-ICD è sicuro e ben tollerato ed efficace nel trattamento delle
aritmie ventricolari. • La miniaturizzazione dei componenti del PM ha permesso lo sviluppo di
PM senza fili. • In atto questi nuovi devices sono limitati ai pazienti che necessitano una
stimolazione monocamerale in ventricolo destro. È probabile che una comunicazione wireless tra più devices impiantati all’interno del cuore può aprire ulteriori prospettive per l’elettrostimolazione.
• In ogni caso sono necessari ulteriori studi ed evidenze riguardo la sicurezza, la performance a lungo termine e la estraibilità di questi devices in caso di infezione.
Questioni aperte • Riescono questi leadless PM a rimanere performanti in un
lungo periodo di osservazione? • Possono essere estratti e come devono essere trattati i
pazienti che vanno incontro ad una infezione? • In caso di scarica della batteria è possibile lasciare il device
ed impiantarne un altro considerate le ridotte dimensioni? • Possono esserci delle complicanze trombo-emboliche nel
tempo? • Quale è il rapporto costo-beneficio?