development of a quantification method for new drugs of abuse in serum using uplc_msms_luiza_tworek
TRANSCRIPT
Development of a Quantification Method for New Drugs of Abuse in Serum Using UPLC – MS/MSAuthor: Luiza Tworek
Dept. of Laboratory MedicineProgramme in Biomedical Laboratory ScienceDegree project, C-level, 15 points Biomedical Laboratory Science
Spring term 2012
Tutor:Olof Beck, professor, Department of Clinical Pharmacology Pharmacology laboratory, Karolinska University Hospital, Huddinge
Co – Tutor: Yasir Ahmad Al – Saffar, BMADepartment of Clinical Pharmacology Pharmacology laboratory, Karolinska University Hospital, Huddinge
Dept. of Laboratory Medicine
Programme in Biomedical Laboratory Science
Degree project, C-level
Biomedical Laboratory Science
Spring term 2012
Development of a Quantification Method for New Drugs of Abuse in Serum Using UPLC – MS/MS
Abstract
Abuse of new unregulated synthetic psychoactive substances, also called designer drugs is a growing problem that brings a need for new sensitive and efficient method of analysis. The aim of this study was to develop and validate an ultra – performance liquid chromatography tandem mass spectrometry method (UPLC – MS/MS) for new drugs of abuse. Twelve new designer drugs were determined with this method methylenedioxypyrovalerone, bk-N-methylbenzodioxolylpropylamine, methylone, para-fluoroamphetamine, 4-methylmethcathinone, 5,6-methylenedioxy-2-aminoindane, para-methoxymethcathinone, naphylprovalerone, 4-hydroxyl-N-methyl-Nethyltryptamine, 2-diphenylmethylpiperidine, O-desmethyl- cis -tramadol and metoxetamine. Samples were prepared by protein precipitation using acetonitrile. Chromatographic separation was performed on Acquity UPLC BEH C18 (2.1 mm x 100 mm) particles size 1.7 µm using 0.1% formic acid in water and 100% acetonitrile in a gradient mode at a flow rate of 0.500 mL/min, injection volume 2 µL.The mass spectrometric analysis used electrospray ionization and compounds were detected with MS/MS with multiple reaction monitoring for two transitions.Calibration curves were linear (r2 > 0.99) in the concentration range 0 – 500 ng/ml in serum. Limit of detection and quantification was documented for each analyte. QC samples intra – and inter – assay precision and repeatability were satisfactory (CV < 10%). The developed method was successfully applied on, serum samples from urine - positive patient (n = 57). This study showed that this method is fully reproducible and specific for all analytes and can be introduced to routine analysis.
Abbreviations:..........................................................................................................31. Summary:.............................................................................................................52. Introduction:.........................................................................................................63. Materials and methods.......................................................................................10
3.1 Chemicals and reagents................................................................................103.2 Stock and standard solutions...................................................................103.3 Patient samples.............................................................................................113.4 Instrumentation............................................................................................11
3.4.1 LC conditions........................................................................................123.4.2 MS/MS conditions................................................................................12
3.5 Sample preparation - protein precipitation:.................................................133.6 Method validation........................................................................................13
3.6.1 Recovery and matrix effect...................................................................143.6.2 Accuracy, precision...............................................................................143.6.3 Calibration curve; linearity...................................................................143.6.4 Limit of detection and limit of quantification.......................................153.6.5 Selectivity and specificity.....................................................................15
3.6.5.1 Interferences...................................................................................153.6.5.2 Matrix effect...................................................................................15
3.6.6 Carryover..............................................................................................163.6.7 Sample stability.....................................................................................16
3.7 Application of method.................................................................................163.8 Statistical methods...................................................................................17
4 Results and discussion...................................................................................174.1 Sample preparation......................................................................................184.2 LC optimization...........................................................................................194.3 Internal Standards........................................................................................214.4 MS/MS conditions.......................................................................................23
4.4.1 Product ions determination...................................................................244.4.2 MRM establishment..............................................................................26
4.5 Method validation........................................................................................294.5.1 Recovery (see also section 4.1 Sample preparation).............................29
4.5.2 Matrix effect..............................................................................................29
1
4.5.2 Carryover..............................................................................................304.5.3 Accuracy and precision.........................................................................314.5.4 Limit of detection and quantification....................................................324.5.5 Calibration curve; linearity...................................................................334.5.6 Selectivity and specificity.....................................................................354.5.7 Sample stability.....................................................................................37
4.5.8 Application of method..............................................................................37Conclusion.............................................................................................................385. References..........................................................................................................39Appendix I.............................................................................................................41Appendix II............................................................................................................44Appendix III...........................................................................................................45
2
Abbreviations:
2- DPMP Desoxypipradol (2 – diphenylmethylpiperidine)
4 – FA 4 – fluoroamphetamine
4 – OH – MET 4 – hydroxyl – N – methyl – N –ethyltryptamine
AcCN Acetonitrile
BEH Ethylene – bridged hybrids
CV Coefficient of variation
ESI Electron spray ionization
FA Formic acid
IS Internal Standard
LC Liquid chromatography
LGC Laboratory of The Government Chemist
LOD Limit of detection
LOQ Limit of quantification
O – DM – TRA O – desmethyl – cis – tramadol
MDAI 5,6 – methylenedioxy – 2 – aminoindane
MDPV Methylenedioxypyrovalerone
ME Matrix effect
3
SRM Selected reaction monitoring (in this text replaced by MRM (multiple reaction monitoring))
MS Mass spectrometry
M/Z ratio Mass to charge ratio
QC Quality control
R2 Correlation coefficient
RE Recovery
RT Retention time
SKL Statens krimineltekniska laboratorium
S/N ratio Signal to noise ratio
STRIDA Samverkansprojekt kring toxicitetutredning och riskbedömning av Internetdroger baserat på Laboratorieanalyser
UPLC Ultra Performance Liquid Chromatography
4
1. Summary:
Abuse of new unregulated synthetic psychoactive substances, also called designer drugs, is a growing problem that brings a need for new sensitive and efficient methods of analysis. The aim of this study was to develop and validate an ultra – performance liquid chromatography tandem mass spectrometry method (UPLC – MS/MS) for new drugs of abuse. Twelve new designer drugs were determined with this method methylenedioxypyrovalerone, bk-N-methylbenzodioxolylpropylamine, methylone, para-fluoroamphetamine, 4-methylmethcathinone, 5,6-methylenedioxy-2-aminoindane, para-methoxymethcathinone, naphylprovalerone, 4-hydroxyl-N-methyl-Nethyltryptamine, 2-diphenylmethylpiperidine, O-desmethyl- cis -tramadol and metoxetamine. Samples were prepared by protein precipitation using acetonitrile. Chromatographic separation was performed on Acquity UPLC BEH C18 (2.1 mm x 100 mm) particles size 1.7 µm using 0.1% formic acid in water and 100% acetonitrile in a gradient mode at a flow rate of 0.500 mL/min, injection volume 2 µL. The mass spectrometric analysis used electrospray ionization and compounds were detected with MS/MS with multiple reaction monitoring for two transitions.Calibration curves were linear (r2 > 0.99) in the concentration range 0 – 500 ng/ml in serum. Limit of detection and quantification was documented for each analyte. QC samples intra – and inter – assay precision and repeatability were satisfactory (CV < 10%). The developed method was successfully applied on serum samples from urine - positive patient (n = 57). This study showed that this method is fully reproducible and specific for all analytes and can be introduced to routine analysis.
5
Key words: designer drugs, fragmentation, chromatography, mass spectrometry, validation
2. Introduction:
Abuse of new unregulated synthetic psychoactive substances, also called “designer
drugs” or “legal highs” is a dramatically increasing problem, which brings a need
for a new sensitive and efficient method of analysis.
Designer drugs are analogues or derivatives of illegal substances classified as
narcotics. Their chemical structures are changed with modifications such as
addition, substitution, deletion of one or more chemical groups. Designer drugs
have similar psychoactive effects to the illegal variant. Although these drugs are
considered a safe alternative to real narcotics, there are many cases of lethal
intoxications. These recreational substances can be easily obtained through the
internet and they are much cheaper than classical drugs. The problem however is
their high toxicity and lack of sufficient pharmacological information (1–4).
This need led to the beginning of a new project called STRIDA
(Samverkansprojekt kring toxicitetutredning och riskbedömning av Internetdroger
baserat på Laboratorieanalyser) that involves cooperation between Karolinska
University Laboratory and the Swedish Poison Information Centre. STRIDA’s
main goals are to gather information about new incoming designer drugs, their
toxicity and to develop new methods allowing for easy and efficient detection (3).
Designer drugs can be classified according to chemical structure or based on
clinical effects (2). In this paper classification according to chemical structure is
6
used, based on the review article “Clinical toxicology of newer recreational
drugs” S.L. Hill, S.H.L. Thomas (2).
Hill and Thomas categorize designer drugs in four distinct chemical families,
piperazines, phenetylamines, tryptamines, piperidines and non-classified
compounds (Figure 1).
Figure 1Classification of designer drugs according to chemical structure, A. piperazines, B. phenetylamines, C. tryptamines, D. piperidines.
Substances used in this study can be classified as follows: phenetylamines
(MDPV, methylone, butylone, 4-fluoroamphetamin, MDAI, methedrone,
mephedrone, naphyrone); tryptamines (4-OH-Met); piperidines (2-DPMP) and
non-classified (O-DM-Tra, MEX). Chemical structures are presented in figure 2.
Phenetylamines are a large group that includes illegal drugs of abuse such as
amphetamine, metamphetamine and a great number of their derivatives. Most of
7
the phenetylamines in this study belong to β-cathinones that are plasma membrane
catecholamine uptake transporters inhibitors. Psychoactive effects are similar to
those produced by amphetamine and MDMA (ecstasy). The most common
symptoms are hallucinations, motorical problems, tachycardia and temperature
rise (3).
4-OH-Met (tryptamine) is a potent hallucinogenic drug that is not classified as a
narcotic in Sweden. STRIDA had 10 registered cases of intoxication with this
substance during 2010. Mydriasis, tachycardia, agitation are those symptoms
which were usually observed in patients (3).
Piperidines are a relatively new group of designer drugs. Psychoactive activity is
probably an effect of inhibition of the dopamine reuptake transporters. There is no
efficient data about piperidines toxicity. The few cases that were registered
showed symptoms such as agitation, tachycardia and mydriasis (2).
O-DM-Tramadol is a metabolite of a common analgesic tramadol. Tramadol is a
µ-opioid receptor agonist and inhibits the reuptake of serotonin and
norepinephrine. O-DM-Tramadol has approximately 300 times higher receptor
affinity than the mother substance (5). This substance is found in a mixture, called
Krypton, with the herbal drug Kratom. Krypton is sold on the internet in large
quantities and because of O-DM-Tramadol it can leads to unintentional overdose
(6).
Methoxetamine, MEX, is an analog of ketamine. Ketamine is a highly addictive
psychedelic drug. There is no scientific information about MEX pharmacology or
toxicity. All described MEX effects come from web forums. Because of a great
8
similarity between MEX and ketamine, psychoactive effects of MEX can be
assumed to be related to those of ketamine (7).
Figure 2 Chemical structures of designer drugs used in this study.
The aim with this study was to develop, optimize and validate a quantification
method for new designer drugs in serum based on a presently validated method in
9
urine (8). The measured drug concentration should correlate between drug
toxicology and patient’s symptoms.
3. Materials and methods
3.1 Chemicals and reagents
2 – DPMP (2 – diphenylmethylpiperidine) and O – DM – cis – Tramadol (O –
desmethyl – cis - tramadol) were purchased by LGC Standards (Laboratory of The
Government Chemist). 4 – OH – MET (4 – hydroxyl – N – methyl – N –
ethyltryptamine) was provided by THC – Pharm. Mephedrone, Methedrone,
MDPV (Methylenedioxypyrovalerone), Butylone, Methylone, 4 – FA (4 –
fluoroamphetamine) were provided by SKL (Statens krimineltekniska
laboratorium). Naphyrone was purchased from Internet characterised by SKL.
Methoxetamine was approved by Läkemedelsverket/SKL. MDAI was supplied by
Chiron. Acetonitrile (AcCN) and formic acid (FA) were provided by VWR
International and IT Baker,Tamro. Water purified with the Milli – Q Water
purification system was used throughout the experiment.
Serum used throughout the experiment was supplied by Department of
Transfusion Medicine, Huddinge.
Stock solution of methamphetamine-d5, pethidine-d4 and psilocin-d10 were
purchased by SKL.
3.2 Stock and standard solutions
10
Reference substances, methanolic solutions, methoxetamine (conc. 2.044 mg/mL),
methedrone (conc. 1.2467 mg/mL), mephedrone (conc. 0.536 mg/mL), naphyrone
(conc. 0.81 mg/mL), butylone (conc. 1.057 mg/mL), O – DM – cis – Tramadol
(conc. 1000 µg/mL), MDAI (conc. 1000 µL/mL), 4 – OH – MET (conc. 893.14
µg/mL), 4 – FA (conc. 1.1528 mg/mL), 2 – DPMP (conc. 964.2 µl/mL) were
kindly supplied by Pharmacological Laboratory (Karolinska University Hospital,
Huddinge).
Stock solution at concentration 40 µg/mL was prepared by diluting reference
substances in 0.1% FA.
Internal standard (IS) working solution at concentration 1000 ng/mL was prepared
by diluting from methamphetamine-d5 in methanol (conc. 100 µg/mL),
phetidine-d4 in methanol (conc. 1.0 mg/mL) and psilocin-d10 (conc. 100 µg/mL)
stock solutions in 0.1% FA.
All solutions were stored at – 20 °C.
3.3 Patient samples
57 patient samples used during the experiment were left-over aliquots from the
routine flow. All samples showed positive result in urine for at least one analyte.
Serum samples were stored at - 80ºC until use.
3.4 Instrumentation
11
All analyses were performed using UPLC – MS/MS Waters system. System
contained Waters Acqutiy UPLC system, Waters Acquity Autosampler, Xevo –
TQ MS # VBA 492 and MassLynx 4.1 SCN810 software.
Optimal instrumental parameters used under this study were based on already
known, designer drugs in urine by Department of Pharmacology Laboratory
(Karolinska University Hospital, Huddinge) (8). New MS/MS condition for
methamphetamine – D5 were determined using screening mode, MassLynx
software (Table I).
3.4.1 LC conditions
Reversed phase column, Acquity UPLC BEH C18 (2.1 mm x 100 mm) with
particles size 1.7 µm, was used to separation of analytes. The flow rate 0.600
mL/min and 0.500 mL/min were tested with an injection volume of 2 µL. Final
flow rate was set at 0.500. Two gradient elution were tested using two mobile
phases A (0.1% formic acid in water) and B (100% acetonitrile). Final gradient
was carried out at a 4 % B, increasing to 45 % B by 2.7 min. At 3.7 min was set to
95 % B for 0.4 min and then returned to starting conditions. Target column
temperature was set at 60.0 °C. Run time was 4.5 min.
Strong needle wash and weak wash was used to avoid carry-over between
injections.
3.4.2 MS/MS conditions
12
Optimal MS/MS parameters used under this study were based on already known,
designer drugs in urine by Department of Pharmacology Laboratory (Karolinska
University Hospital, Huddinge) (8).
MS/MS was operated in positive ESI (electron spray ionization) mode with
multiple reaction monitoring (MRM) using three time segment. Ionization source
capillary voltage was set at 500 V. Cone voltage were set at 14.00 V. Extractor
was set at 3.00 V. Source temperature was set at 150.0 °C and desolvation
temperature at 600 °C. Argon was used as a collision gas with flow set at 0.15
mL/Min.
MRM (multiple reaction monitoring) time segments for analytes detection (two
transitions/analyte, one transition/IS) were maintained at (1) 0 – 2.48 min, (2) 2.48
– 3.14 min and (3) 3.14 – 4.50 min.
Methamphetamine – d5 trace and quantification ion was determined in an
injection experiment using Waters Xevo TQ MS Detector software function. The
sample spiked with 100 µL IS was screened to establish presence of a
methamphetamine-d5 molecule (protonated molecule mass: 154.39 g/mole).
Thereafter, the methamphetamine-d5 product ions were isolated and their
molecular masses were determined.
3.5 Sample preparation - protein precipitation:
IS, sample and AcCN proportion used under this study were based on already
known, designer drugs in urine by Department of Pharmacology Laboratory
(Karolinska University Hospital, Huddinge) (8).
13
50 µL of IS and 400 µL of AcCN was added to 100 µL of a serum sample
followed by centrifugation at 3050 x g for 10 minutes. 300 µL of the supernatant
was transferred to a vial and analysed.
3.6 Method validation
A partial validation was performed according to the FDA Guidance for Industry,
Bioanalytical Method Validation (9)
3.6.1 Recovery and matrix effect
For determination of recovery (RE) and matrix effect (ME), three sets of extracts
were prepared, each in triplicates. (1) included analytes at concentration of 350
ng/mL, spiked with serum, prepared according to protein precipitation method,
there after centrifugation 500 µL of aliquot was diluting with 100 µL 0.1%FA.
(2) 0 - samples prepared as before, there 500 µL of supernatant was added to 100
µL of solution at concentration 350 ng/mL in 0.1%FA.
(3), 1 mL working solutions at concentration of 350 ng/mL in 0.1% FA was used.
Samples were mixed with 50 µL IS, 400 µL ACCN and vortexed.
3.6.2 Accuracy, precision
Accuracy and precision experiment was performed by preparing quality control
samples (QC) in number of 5 at 3 different concentrations: 25; 50; 500 ng/mL
which were analysed for the period of 5 days. All samples were prepared
according to protein precipitation method.
14
3.6.3 Calibration curve; linearity
Calibration standards with concentration of 5; 10; 25; 50; 100; 150; 250; 350; 500
ng/mL analytes in 0 – serum were used. All standards included 50 µL IS (1000
ng/mL). Standards were prepared and analysed as an unknown samples using
protein precipitation method. This experiment was performed by preparing six
calibration curves.
3.6.4 Limit of detection and limit of quantification
Limit of detection (LOD) and limit of quantification (LOQ) were determined
theoretically using data for calibration solution 5 from “Calibration curve;
linearity”. Thereafter the empirical value of LOD and LOQ were established by
analysis of serum samples spiked to obtain following concentrations (ng/mL)
between 0.02 – 10.0. Signal to noise ratio (s/n ratio) was calculated for each
analyte using MassLynx software s/n ratio function.
3.6.5 Selectivity and specificity
3.6.5.1 Interferences
The potential for interference was investigated by testing ten serum samples
collected from patients there other xenobiotic than analytes of interest were
present (sertraline, clozapine, ganciclovir, lamictal, valproate, vancomycin,
zuclopenthixol, haloperidol, rifabutin/rifampicin, meropenem), ten blanks
collected from 10 different drug -free sources and ten calibration solutions each
15
spiked with one of the following reference substances: sertraline, clonazepine,
ganciclovir, buprenorphine, aripirazole, azole – antimycotica, all common
benzodiazepines, haloperidol/ zuclopenthixol, rifabutin/rifampicin,
meropenem/ceftazedim.
3.6.5.2 Matrix effect
Ion suppression was determined for MDAI, methylone and methedrone by
simultaneously injection and infusion of 10 µg/min of two samples, one blank and
one spiked with analytes. Two sets were needed to perform the experiment. Set 1
contained 0-serum sample and set 2 consisted of 0.1 % FA as a blank, in both sets
were used the same sample containing 1000 µL 0.1%FA spiked with MDAI,
methylone, methedrone reference substances to obtained concentration of 100
ng/mL.
3.6.6 Carryover
Carryover experiment was performed by analysing two blank samples after the
analysis of each solution at concentration of 5000; 1000; 500 ng/mL.
3.6.7 Sample stability
Short-term stability was tested for one batch consisted of 57 patient samples
extracted as described before, calibration curve solutions 5 – 500 ng/mL and 0 –
serum. After the first run, samples were stored in 72 hours in autosampler at 8 ºC
and run again with a new prepared calibration curve.
16
3.7 Application of method
Serum samples from 57 patients, which showed positive results in urine for at
least one analyte, were tested. Tested batch included a new prepared calibration
curve solutions 5 – 500 ng/mL, one blank and following numbers of positive
samples: 2 of methylone, 10 of O – DM –tramadol, 3 of 4 – fluoroamphetamine, 1
of methedrone, 4 of butylone, 2 of mephedrone, 13 of metoxetamine, 14 of
MDPV, 5 of 2 – DPMP, 2 of naphyrone and 12 of 4- OH – Met. All samples were
prepared and analysed as described before.
3.8 Statistical methods
The regression analysis was used to calculate correlation coefficient for
calibration curve for each analyte. To determine method repeatability (coefficient
of variation, CV %), inter - assay precision (CV %) and intra – assay precision
(CV %) one - way ANOVA with 5% significance level was used.
4 Results and discussion
This report describes an optimization and validation of a screening method for
new designer drug in serum using highly selective and sensitive method UPLC-
MS/MS. This new method is based on an existing one for those drugs in urine.
The experiment included adjustment of chromatographic parameters and
optimization of mass spectrometry.
17
In the last few years, a number of LC – MS methods have been developed and
validated for different designer drugs both for urine, plasma and serum. This
newly developed method for UPLC MS/MS is simple, rapid and uses simple
sample preparation. It allows detection of 12 analytes belonging to different
chemical classes. The method is developed for serum which decreases problems
with samples collection in acute intoxication cases. Use of tandem MS allows for
determination of a great number of compounds and samples, which is required for
routine analysis. Sample preparation is simple that minimizes errors. Analysis is
performed automatically, and once optimized it does not need further adjustments.
Validation procedure confirmed that this method is specific for the analytes and
suitable for routine use. Recovery results were not as expected but on acceptable
levels. Inter -, intra – assay precision and method repeatability values showed
results between 0.2 – 9.0 % variance. Method accuracy values were between 96 –
108%, which is highly satisfactory. Measuring range was established at 0 – 500
ng/mL. With this method a concentration as low as <5 ng/mL can be detected for
most analytes. This study showed that analytes can be identified in samples as old
as 2 years, when stored properly. Short–term stability results did not shown any
significant differences in peak areas. Long–term stability tests should be
performed to established samples durability when stored at -20ºC and at rum
temperature.
4.1 Sample preparation
Samples were prepared using simple protein precipitation with AcCN. All twelve
analytes prepared according to protein precipitation could be separated and
18
identified. Drying with nitrogen gas of supernatant and resuspension with
10%AcCN in 0.1% FA was tested and results showed no differences in recovery
results. Moreover, a difficulty was observed that 4 – OH – MET could not be
detected. Choice of the sample preparation method was mainly based on recovery
results which refer to extraction efficiency (1). Precipitation with AcCN showed
recovery higher than 85%, which is satisfactory. However, the expected value of
recovery is higher than 92% with this method of purification (10). Sample
purification from potential interference sources is a crucial step preceding liquid
chromatography and mass spectrometry. Preparation accomplished high sample
purity by precipitation of proteins with AcCN. Polson et al. showed in theirs
study that efficiency of protein precipitation with AcCN is approximately 92% at
volume ratio > 2:1 (AcCN; sample). AcCN is the most common organic
precipitant which does not interfere with analytes (10). Moreover, pure mobile
phase used as a precipitant causes greater ionization effects according to Polson et
al. This procedure also allowed less time for sample preparation by excluding the
drying and resuspension steps. However, the high AcCN volume ratio which was
used in this experiment did influence chromatography and will be discussed in the
next section.
4.2 LC optimization
Reversed phase column, Acquity UPLC BEH C18 (2.1 mm x 100 mm) particle
size 1.7 µm, was used throughout the experiment to separate analytes.
19
Ethylene – bridged hybrids (BEH), provided by Waters, is a technology used to
create particles that allow performance at difficult conditions such as high
pressure, temperature and wide pH range (11). 0.1% FA (A) and 100% AcCN (B)
was used as mobile phases according to Water recommendation. 0.1% FA is the
first choice of mobile phase A because of low molecular weight, which decrease
interferences and background noise. AcCN, used as mobile phase B, has a high
efficiency of ESI as mentioned before, good elution strength and applies less
pressure to the column. However, AcCN has a tendency to cause peak tailing or
fronting. In this study, a significant fronting was observed for IS substance
psilocin – d10 (RT 1.35) (figure 9). Psilocin – d10 is the first eluting compound
which indicates that high AcCN volume ratio causes rapid elution and a
noticeable peak fronting (12). This problem can be solved by adding 0.1 % FA to
the supernatant at the last stage of sample preparation or gradient adjustment.
Figure 3 Chromatogram of psilocin - d10, RT 1.37. A. Calibration solution 25; B. Calibration solution 250. Upper chromatogram shows samples diluting with 100 µL 0.1%FA, lower shows standard sample.
One attempt, as described before, was performed to analyse 0.1% FA influence on
analyte response and peak shape (figure 9). Results from the experiment showed
more symmetric-shaped peak but it was still not satisfactory. This experiment
should be tested with different 0.1% FA: supernatant ratio.
20
The mobile phase gradient and flow rate was adjusted to optimize analytes
separation. The goal was to avoid elution of analytes or IS with the void volume
and to achieve a good separation in a short run time. The flow rate was set at
0.500 mL/min to achieve rapid analysis, duration time 4.5 min. Final gradient was
optimized according to analysis duration time as following 0 – 0.5 min: 4 % B;
0.5 – 2.7 min: 45 % B; 2.7 – 4.0 min: 95 % B; 4.1 – 4.5 min: 4 % B (figure 3).
Analytes retention times (RT) were measured for each analyte and summarized in
Figure 4.
In this experiment, no analyte was eluted with the void volume. All analytes were
fully separated and no double peaks were observed.
Figure 4 Mobile phases gradient (%B) used in chromatography throughout the experiment.
4.3 Internal Standards
Internal standard is essential to quantify an analyte. In ideal situation, IS should be
a compound that has as similar as possible chemical structure and properties to the
analyte. IS should also give a similar response but different enough to be
21
recognized by instrument (13). In this case, it was not possible to choose a more
suitable IS for each substance. Mostly because, reference substance for new
designer drugs do not exist and a procedure to acquire a certificate for a new
reference product is complicated and expensive (14).
Psilocin – d10 and Pethidine – d4 and Methamphetamine – d5 were chosen as IS
due to their retention times, stability, chemical similarities to the analytes as well
as availability. A retention time for each IS was measured under the same
conditions as the analytes and determined to 1.35 min Psilocin d-10, 1.98 min
Methamphetamine d-5, 2.78 min Pethidine d – 4 (figure 4).
22
Figure 5 Retention times were measured for each analyte (A) and IS (B). To elution 0.1 % FA and 100 % AcCN were used as mobile phases.
23
4.4 MS/MS conditions
Mass spectrometry function is based on three fundamental units: ion source,
analyzer and detector. The knowledge of their function is important to achieve
good optimization of the instrument leading to good results (15).
The ion source used was electrospray ionization (ESI) operating in positive mode
which was the most suitable for the analyte chemical structures. All ESI
parameters were optimized to accomplish the most effective ionization of the
analytes without any fragmentation. The schematic of a ESI with all used
parameters is presented in figure 11 (picture modifide from 16). To maximize the
operating parameters it is important to be sure that analytes are ionized. The main
problem that can occur is when there is too much interfering molecules in the
matrix that compete with the analytes. This mechanism is known as ion
suppression and must be taken into consideration when experiments are designed.
Ion suppression mechanism will be described in later section.
24
Figure 6 (A) Schematic representation of ESI with accurate parameters which were used throughout the experiment. (B) Schematic of the mechanism of ion formation using ESI.
4.4.1 Product ions determination
Two transitions were used as analyte determination for each substance. All used
transitions were based on earlier method for urine (8). Experiment was performed
only for methamphetamine – d5. Trace and quantification ion were established by
screening the sample with MS/MS software. Three fragments were detected:
91.85; 121.02; 98.95. The fragment: 121.02 was chosen as a trace and
quantification ion.
25
To determine a unique trace and quantification ion (daughter ions) is essential to
identify compounds such as designer drugs. The main reason is that the analyzed
substances can have similar structure to the endogenous compounds or medicine.
Cleavage pattern is important to determine right ions and depends on the chemical
structure of the analyte, atoms and groups and theirs position. Fragmentation
pattern is presented in an example in Figure 13 (17).
Figure 7 Fragmentation pattern for MDPV (Mol. Mass 275.343 g/mol) and methylone (Mol. Mass 207.23 g/mol).
In some cases, when compounds have isomers, a specific isomer cannot be easily
differentiated. Fluoroamphetamine is one of those substances.
Fluoroamphetamine has three isomers para (4) -, meta (3) - and orto (2) –
fluoroamphetamine. All three isomers give fragments which have exactly the
same molecular masses (figure 14, modified from (18–20)) (21,22)
26
Figure 8 Isomers of fluoroamphetamine. Two transitions resulting from fragmentation are presented.
In this study 4 – fluoroamphetamine was included as an analyte. However, when
patient sample is tested the results should indicate that the sample can contain any
one of those three isomers. In cases where patients claim to have consumed 4 –
fluoroamphetamine, the same precaution must be taken because no isomer
determination can be made.
4.4.2 MRM establishment
The principles of tandem mass spectrometry are fragmentation of an ionized
analyte and detection of those fragments with reference to mass-to-charge ratio
(m/z ratio). In tandem mass spectrometry the first quadruple (Q1) acts as a filter
with fixed voltage, there the protonated mother molecule is selected.
Fragmentation of the chosen molecule occurs in the collision cell with an inert gas
(the second quadruple, Q2). The third quadruple like Q1 is a fixed filter where
trace and quantification ions are selected (figure 12, modified from 15 ) (15).
When Q1 and Q3 are simultaneously set to obtain different m/z, it allows
27
scanning specific ions of interest. This highly selective system is called selected
(multiple) reaction monitoring (SRM/MRM).
In this study three MRM time segments were used, created according to retention
time of analytes (see figure 2 and table 1 ) (23).
Figure 9 Schematic of tandem mass spectrometer.
Fragments which were detected by each function and theirs conditions were
summarized in Table I.
Table I Summary of MRM function parameters and transition for each analyte and IS used
throughout the experiment.
No Parent (Da) Daughter (Da)QuantifierQualifier
Cone (V) Collision (eV)
Compound (equivalent IS)
Time segment 1
1 153.98 108.70 25.00 21 4- fluoroamphetamine
(pethidin –d4)137.00 9
2
155.08 121.02 20.00 10.00 Metamphetamine-D5
3 177.98 102.80 22.00 25.00 MDAI
28
(psilocin – d10)
160.90 11.00
4 178.01 118.98 20.00 20.00 Mephedrone
(pethidin –d4)144.91 20.00
5 194.07 161.00 22.00 19.00 Methedrone
(metamphetamine – d5)176.00 13.00
6 207.98 131.94 20.00 26.00 Methylone
(metamphetamine – d5) 190.04 12.00
7 215.20 164.00 22.00 19.00 Psilocin–D10
8 219.20 132.00 23.00 27.00 4 – OH – MET
(psilocin – d10)160.00 19.00
9 222.10 174.10 28.00 17.00 Butylone
(pethidin –d4)204.10 13.00
10 250.12 57.80 22.00 15.00 O – DM – Tramadol
(metamphetamine – d5)232.10 11.00
Time segment 2
1 248.19 175.00 30.00 19.00 Methoxetamine
(pethidin –d4)203.00 13.00
2 252.25 224.10 44.00 23.00 Phetidine-D4
3 276.29 126.00 38.00 25.00 MDPV
(pethidin –d4)135.00 25.00
Time segment 3
1 252.14 128.30 36.00 25.00 2-DPMP
29
(pethidin –d4)166.90 25.00
2 282.04 126.11 32.00 28.00 Naphyrone
(pethidin –d4)154.93 28.00
4.5 Method validation
4.5.1 Recovery (see also section 4.1 Sample preparation)
The efficiency of the preparation method was evaluated by performing recovery
test for each analyte and IS. The results obtained for samples spiked in serum
were compared with the results obtained for post – spiked samples. To calculate
the recovery the ratio of the peak areas were used:
RE = [Area (1) /area (2)]*100%, there area was calculated as mean of 3 samples.
Recovery for all analytes and IS were higher than the minimal required ratio
which is 75%. Recovery data is summarized and shown in Table II. Data used to
recovery calculation is displayed in Appendix I.
4.5.2 Matrix effect
Matrix effect (ME) is a common problem when biological samples are analyzed.
Presences of endogenous molecules like plasma proteins, phospholipids or other
xenobiotic are a great interference source which must be eliminated or decreased.
The sample cleanup and the chromatography are crucial to minimize this problem.
Matrix effect was calculated by comparison of peaks are for post - spiked samples
and samples spiked with 0.1% FA:
ME = [Area (2)/Area (3)]*100%, there area was calculated as mean of 3 samples.
30
Results were higher than 100 % for MDAI and 4 – OH –Met, two analytes that
elutes first. For remaining ten analytes, calculated matrix effect was between 88
and 94 % (Table II).
Table II Recoveries (%) were calculated for each analyte and IS. Matrix effect (%) was calculated for analytes.
Analyte Recovery (%) Matrix
effect (%)
Analyte Recovery
(%)
Matrix
effect (%)
MDPV 86 93 4 –FA 87 89
O – DM –cis –
Tramadol
86 94 Mephedrone 86 93
4 – OH – MET 94 147 MDAI 87 119
2 – DPMP 90 89 Methedrone 86 93
Butylone 87 91 Metoxetamine 87 93
Methylone 85 93 Naphyrone 92 88
IS Recovery (%)
Phetidine – d4 97
Psilocin – d10 94
Methamphetamin
e – d5
99
4.5.2 Carryover
31
Carryover is a problem that can have a great effect on the accuracy and precision.
The carryover detected in this experiment was considered to be a result of needle
contamination by a previous sample at a high concentration (24).
Carryover was tested for samples at concentration 5000, 1000 and 500 ng/mL.
Results were calculated by comparison peak area of analyte in blank with peak
area of analyte in high concentrated solution that was analysed before blank
sample. Calculation showed that carry over values were below 0.30 % (between
0.02 – 0.27) and only three analytes, O – DM –tramadol, 4 – OH – met, butylone
showed no or non-significant carryover (ng/mL) when compared with empirical
LOD and LOQ. Other analytes had carryover (ng/mL) higher than empirical LOD
or LOQ. However, after establishing cutoff and method precision (determined
from QC samples), carryover results were considered to be non-significant.
Calculated data (percentage, ng/mL) is presented in Appendix II.
4.5.3 Accuracy and precision
Five replicates of quality control (QC) samples at three different concentrations
were analysed to determine intra – assay, inter – assay precision and method
repeatability. All results were expresses as a coefficient of variance (CV %)
calculated for QC concentrations. Accuracy for each concentration was calculated
as deviation of the mean from the nominal value. All accuracy and precision
results were within the reference intervals of 20 % for lower calibration solutions
and 15% for other concentrations (Table III).
Table III Accuracy (deviation %) and precision calculation (CV %).
Analyte Repeatability Inter – assay Intra – assay Accuracy (deviation %)
32
(CV %) precision (CV %) precision (CV %)
25 50 500 25 50 500 25 50 500 25 50 500
MDAI 4.7 6.3 5.3 5.7 4.4 3.7 7.4 6.3 5.3 108 106 98.8
O – DM –Tramadol 5.7 3.3 2.6 0.2 4.6 2.7 5.7 5.67 3.7 99.2 96.9 100.2
4 – OH – MET 5.7 5.7 4.3 4.0 6.2 3.8 5.7 8.44 5.7 100.3 96.3 103.9
2 – DPMP 6.1 3.5 3.0 3.7 4.5 1.1 6.1 5.66 3.0 103 100.7 99.4
Butylone 7.4 4.9 4.9 3.6 3.1 3.1 7.4 4.89 4.9 103.2 100.5 100.1
Methylone 5.5 3.7 3.3 1.7 4.9 2.5 5.5 6.14 3.3 98.5 98.8 102
4 –FA 5.6 3.8 2.8 6.2 3.7 4.5 8.3 5.31 5.3 107.4 102.3 99.3
Mephedrone 6.3 3.7 2.6 1.1 4.0 1.4 6.3 5.47 2.6 103.4 101.8 98.4
MDPV 5.4 2.7 2.7 5.1 5.9 3.1 7.4 6.51 4.1 102.8 99.7 100.5
Methedrone 6.0 3.1 2.4 5.9 4.2 2.1 8.4 5.25 2.4 102.3 98.6 101.3
Metoxetamine 5.1 3.4 2.3 7.3 4.1 2.4 8.9 5.29 3.3 103.8 99.7 99.9
Naphyrone 5.8 4.0 3.1 5.6 4.0 2.1 8.0 5.64 3.1 104.1 100.6 100.7
4.5.4 Limit of detection and quantification
LOD and LOQ values were established theoretically from calibration solution 100
using following equations 3.3*standard derivation (SD)/slope and 10*SD/slope.
SD were calculated for six samples of calibration solution 100 and slope were
calculated as a medium slope for six calibrations curves (25). Theoretical LOD
and LOQ were corrected empirically and adjust to instrument capacity. The
theoretically calculated values were below 0.1 ng/mL for LOD and below 0.2
ng/mL for LOQ. However, empirical values established using MassLynx software
at s/n ratio ≥ 3 for LOD were 0.02 – 0.75 ng/mL and s/n ratio ≥ 10 for LOQ were
33
0.1 – 5.0 ng/mL (figure 5) for all analytes except O – DM –Tramadol. LOD and
LOQ for O – DM – Tramadol were 10 and 25 respectively (Table IV). The main
reason that O-DM-Tramadol shows high LOD and LOQ values is problem with
fragmentation.
Figure 10 Determination of s/n ratio for establishment of LOQ for naphyrone (both transitions are shown). Sample concentration 0.02 ng/mL.
4.5.5 Calibration curve; linearity
Six calibration curve were analysed and a correlation coefficient (r2) were
calculated to be higher than 0.99 for each analyte. Calibration curve for Butylone
is shown in figure 6. The correlation coefficient for each analyte is shown in
Table IV.
34
0 100 200 300 400 500 6000
20000400006000080000
100000120000140000160000180000200000
f(x) = 380.664540710285 x + 453.501637722722R² = 0.99995433348773
BUTYLON
BUTYLON Linear (BUTYLON )
Figure 11 Calibration curve (f (concentration) = area) for Butylone, correlation coefficient R2 = 1.
Table IV Calibration curves correlation coefficient, empirical limit of detection and limit of quantification (ng/mL).
Analyte Correlation coefficient (r2) LOD (ng/mL) LOQ (ng/mL)
MDAI 0.997 0.02 0.75
O – DM –cis –
Tramadol
0.999 10 25
4 – OH – MET 0.998 0.75 5.0
2 – DPMP 1.0 0.1 0.75
Butylone 1.0 2.0 5.0
Methylone 0.999 0.5 2.0
4 –FA 0.998 0.5 5.0
Mephedrone 1.0 0.75 2.0
MDPV 1.0 0.02 0.1
35
Methedrone 0.999 0.5 5.0
Metoxetamine 0.999 0.02 0.1
Naphyrone 1.0 0.02 0.1
4.5.6 Selectivity and specificity
Selectivity and specificity was tested for the method by analysis of samples
containing different xenobiotic and ten calibration solutions spiked with known
drugs other than analytes of interests. Moreover, blanks collected from ten
different sources were analysed, for the potential interferences. Chromatograms
example for butylone are presented in figure 7. Analysis of above-mentioned
solutions showed no interference with tested analytes. The method is found highly
selective.
36
Figure 12 Chromatograms for interferences experiment presented for butylone (2 transitioner):patinet sample (sertraline), calibration solution (sertraline), blank.
Ion suppression was tested for MDAI, Methedrone and Methylone to establish a
presence of any source of interference from substances that can exist in a new
matrix, serum. Ion suppression or enhancement (loss/gain in signal) is one
mechanism which is a form of matrix effect. There are many possibilities that can
cause ME. The most well-known cause is the competition between the analyte and
matrix molecule in the ion source. Matrix molecules can co-elute with the analyte
and be easier ionized that causes decreased ionization of molecule of interest
(26,27).
Results of the experiment showed that a significant the ion suppression exist
between 0.4 – 0.9 min. No significant ion suppression was detected what is shown
37
for MDAI (transitions 102.8) in Figure 8.Chromatograms can be found in
appendix III.
Figure 13 Ion suppression for MDAI (102.8 transition is shown). Multiple reaction chromatograms for postcolumn infusion for MDAI, methylone and methedrone. 0.1 % FA is a reference sample (red), matrix 0 - serum (green) and peak for MDAI obtained by standard analysis of calibration solution 100 (violet).
4.5.7 Sample stability
Short – term stability test was performed by analyzing 57 patient samples
containing 68 analytes (se section 3.6 “Sample stability”), thereafter stored in
autosampler at 8ºC for 72 hours and analyzed again. Only two samples showed
significant differences, one consisting mephedrone and one O – DM –tramadol.
Mephedrone was detected on day 1 but not after 72 hours. O – DM –tramadol was
not detected on day one but could be detected after 72 hours. Both samples
showed concentrations below the cutoff, this could explain the results. 53 samples
38
showed positive results there analyte could be detected on both occasions and 14
that gave negative results on day 1 and after 72 hours.
4.5.8 Application of method All tested serum samples were collected from urine–positive patients over a 2-
years period (2010 February – 2012 February) and stored at -80ºC. From 68
analytes 53 were detected. Sample concentration cannot be compared because of
different matrices.
In a few cases results from serum samples were different than results in urine.
In two butylone positive urine-samples showed negative result in serum, it was
probably an error in detection in the urine sample. Small peaks at methylone RT
could be seen on chromatograms. However, quantification was not possible.
In one case, the serum sample showed negative for 4-OH-Met and positive for O-
DM-Tramadol when the urine analysis showed the other way about. 4-OH-Met is
an unstable analyte and degrades easily. In low concentration samples detection
can be a problem. O-DM-Tramadol could have been missed in the urine analysis.
The other samples with negative results could not be explained. It was not
possible to attain data about the initially measured concentrations. No further data
analysis could be performed.
Conclusion
By using UPLC with triple quadruple and MRM a highly selective and specific
method could be developed. Two unique transitions could be determined for each
analyte. Analytes that belong to different chemical classes could be detected in
39
matrix with high levels of interference. Quantification of all substances was
possible. The method was validated and successfully applied to patient samples.
5. References
1. Davies S, Wood DM, Smith G, Button J, Ramsey J, Archer R, et al. Purchasing “legal highs” on the Internet--is there consistency in what you get? QJM 2010;103(7):489–93.
2. Hill SL, Thomas SHL. Clinical toxicology of newer recreational drugs. Clin toxicol 2011; 49(8):705–19.
3. Helander A, Beck O, Hägerkvist R, Hultén P. STRIDA i kampen mot ( o ) lagliga Internetdroger. Läkaretidingen 2011; 108(Xx).
4. Maurer HH. Chemistry, pharmacology, and metabolism of emerging drugs of abuse. Ther Drug Monit. 2010; 32(5):544–9.
40
5. de Moraes NV, Lauretti GR, Napolitano MN, Santos NR, Godoy ALPC, Lanchote VL. Enantioselective analysis of unbound tramadol, O-desmethyltramadol and N-desmethyltramadol in plasma by ultrafiltration and LC-MS/MS: application to clinical pharmacokinetics. J Chromatogr B Analyt Technol Biomed Life Sci. 2012; 880(1):140–7.
6. Kronstrand R, Roman M, Thelander G, Eriksson A. Unintentional fatal intoxications with mitragynine and O-desmethyltramadol from the herbal blend Krypton. J Anal Toxicol 2011; 35(4):242–7.
7. Hofer KE, Grager B, Müller DM, Rauber-Lüthy C, Kupferschmidt H, Rentsch KM, et al. Ketamine-like Effects After Recreational Use of Methoxetamine. Ann Emerg Med 2012; 9:1–3.
8. Al-Saffar YA. Muntlig meddelande: utförandet, metod internetdroger i urin 2012, mars.
9. U.S. Department of Health and Human Services, Food and Drug Administration. Guidance for Industry Bioanalytical Method Validation. Veterinary Medicine; 2001.
10. Polson C, Sarkar P, Incledon B, Raguvaran V, Grant R. Optimization of protein precipitation based upon effectiveness of protein removal and ionization effect in liquid chromatography-tandem mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci 2003; 2:263–75.
11. Corporation W. A Review of Waters Hybrid Particle Technology . Part 2 . Ethylene-Bridged [ BEH Technology ] Hybrids and Their Use in Liquid Chromatography. Technology 2000.
12. Wilson K. Chromatographic techniques. In: Wilson K, Walker J, editors. Principles and techniques of biochemistry and molecular biology. Cambrige: CAMBRIGE University Press; 2010. p. 439; 441.
13. Wilson K. Internal Standard. In: Wilson K, Walker J, editors. Principles and techniques of biochemistry and molecular biology. Cambrige: CAMBRIGE University Press, 2010:35.
14. Archer RP, Treble R, Williams K. Reference materials for new psychoactive substances. Drug Test Anal. 2011; 3(7-8):505–14.
15. Aitken A. Mass spectrometry techniques. In: Wilson K, Walker J, editors. Principles and techniques of biochemistry and molecular biology. Cambrige: CAMBRIGE University Press, 2010: 352 - 354.
41
16. Fenn J, Mann M, Meng C, Wong S. Electrospray ionization for mass spectrometry of large biomolecules. Science 1989; 6.
17. Reitzel L a, Dalsgaard PW, Müller IB, Cornett C. Identification of ten new designer drugs by GC-MS, UPLC-QTOF-MS, and NMR as part of a police investigation of a Danish Internet company. Drug Test Anal. 2011.
18. PubChem C. 2-amino-3-fluoro-1-phenylpropane - Compound Summary (CID 121531). 2005.
19. PubChem C. 4-fluoroamphetamine - Compound Summary (CID 9986). 2005.
20. PubChem C. 1-(3-fluorophenyl)propan-2-amine - Compound Summary (CID 121501). 2005.
21. Johansen SS, Hansen TM. Isomers of fluoroamphetamines detected in forensic cases in Denmark. Int J Legal Med 2012
22. Röhrich J, Becker J, Kaufmann T, Zörntlein S, Urban R. Detection of the synthetic drug 4-fluoroamphetamine (4-FA) in serum and urine. Forensic Sci Int. 2012; 215(1-3):3–7.
23. Cox D, Zhong F, Du M, Duchoslav E. Multiple reaction monitoring as a method for identifying protein posttranslational modifications. J Biomol Tech. 2005;16(2):83–90.
24. Hughes NC, Wong EYK, Fan J, Bajaj N. Determination of carryover and contamination for mass spectrometry-based chromatographic assays. AAPS J. 2007 Jan;9(3):E353–60.
25. Agency EM. European Medicines Agency. Reproduction. 2006; 2(November 1994):1–15.
26. Annesley TM. Ion suppression in mass spectrometry. Clinl chem 2003; 49(7):1041–4.
27. Chambers E, Wagrowski-Diehl DM, Lu Z, Mazzeo JR. Systematic and comprehensive strategy for reducing matrix effects in LC/MS/MS analyses. J Chromatogr B Analyt Technol Biomed Life Sci. 2007; 852(1-2):22–34.
42
Appendix I
id17 name20 desc_sample area 1 area 2 area 3 medel SD
1 MDAI sample 0,1%FA 158725,18
8143556,84
4168809,79
7157030,
612711,4
2 Metylon490882,28
1485187,90
6514767,06
3496945,
815694,1
3 O-DM-tramadol571300,37
5 477815571010,68
8 54004253890,3
4 4-Fluoroamfetamin 91836,102 85299,969 92741,516 89959,24060,32
5 Metedron544774,31
3 532404,5 557968545048,
912783,9
6 Butylon147809,67
2142089,70
3145222,71
9145040,
72864,32
7 Mefedron767760,56
3745136,18
8742869,06
3751921,
913763,4
8 Metoxetamin274498,43
8259491,78
1254188,92
2262726,
410534,0
9 MDPV521053,62
5489798,46
9481832,09
4497561,
420731,1
43
10 2-DPMP146085,31
3135333,62
5133252,78
1138223,
96887,21
11 Nafyron478926,46
9443671,28
1438886,12
5 45382821867,2
12 4-OH-MET102817,22
7 93835,844 104494,18100382,
45731,16
desc_sample_a area 1a area 2a area 3a medel_a SD_a CV_a
sample 350 S 160685,93
8 175203,5158505,78
1164798,40
6 9076,77 5,5388091,06
3422295,84
4385334,62
5398573,84
420590,0
3 5,2428452,53
1464528,65
6433050,34
4 442010,5119636,3
2 4,4
69592,523 75832,438 66968,29770797,752
74553,31
6 6,4
431963,75466428,90
6429198,18
8442530,28
120742,9
6 4,7113827,80
5125429,63
3112735,64
1117331,02
67034,82
6 6,0574303,12
5 641595,5 605377,75607092,12
533678,9
3 5,5207921,67
2227938,95
3207990,70
3214617,10
911537,1
1 5,4381207,78
1429071,34
4390848,62
5400375,91
725314,1
7 6,3108505,30
5119819,53
1109092,52
3112472,45
36369,52
7 5,7358739,09
4398389,15
6359000,40
6372042,88
522816,9
1 6,1136159,18
8145583,62
5136889,59
4139544,13
65243,08
6 3,8
44
desc_sample_b area 1b area 2b area 3b medel_b SD_b CV_bsample 350 B 181926,2 197049 184066,14 187680,443 8183,66
462097,7 483669,69 449306,97 465024,782 17367,4489419 548803,38 492708 510310,125 33376,7
78071,82 85369,852 78168,305 80536,659 4185,95492442,2 531489,13 508358,34 510763,208 19634,3130445,8 140685,33 129003,65 133378,26 6369,06
676786 743009 686150,88 701981,958 35837,7237635,4 261779,11 240481,77 246632,094 13194,7445891,4 493627,19 449171 462896,542 26664122046,4 125892,95 123504,16 123814,492 1941,98395053,7 407281,47 405050 402461,719 6511,84
144993 154642,45 144005,53 147880,333 5876,94
utbyte sample350 S/sample 350 B utbyte sample B/sample 0,1FA87,81 119,5285,71 93,5886,62 94,4987,91 89,5386,64 93,7187,97 91,9686,48 93,3687,02 93,8786,49 93,0390,84 89,5892,44 88,6894,36 147,32
45
ANALYTKONCENTRATIO
N carryover % carryover
ng/mL JON I JON II JON I4-OH-MET 500 - - - 1000 - - - 5000 - - -BUTYLON 500 0,045 - 0,225 1000 0,024 - 0,24 5000 0,028 - 1,4O-DM-TRAMADOL 500 - - - 1000 0,021 - 0,21 5000 0,022 - 1,1MEFEDRON 500 - 0,071 - 1000 0,097 0,068 0,97 5000 0,104 0,047 5,2METEDRON 500 0,044 - 0,22 1000 0,032 0,062 0,32 5000 0,041 0,035 2,05METYLON 500 0,046 0,084 0,23 1000 0,039 0,037 0,39 5000 0,019 0,017 0,954-FLUOROAMFETAMIN 500 0,168 0,245 0,84 1000 0,187 0,172 1,87 5000 0,164 0,168 8,2MDAI 500 0,046 0,134 0,23 1000 0,037 0,087 0,37 5000 0,055 0,053 2,75MDPV 500 0,044 0,059 0,22 1000 0,039 0,053 0,39 5000 0,032 0,027 1,6METOXETAMIN 500 0,053 0,048 0,265 1000 0,04 0,027 0,4 5000 0,012 0,016 0,6NAFYRON 500 0,204 0,233 1,02 1000 0,166 0,171 1,66 5000 0,141 0,11 7,052-DPMP 500 0,264 - 1,32 1000 0,163 - 1,63 5000 0,095 - 4,75
46
Appendix II
Appendix III
47
48