developing and implementing calibration programs...the calibration range must be, at a minimum,...

77
Developing and Implementing Calibration Programs March 19, 2013 Pennsylvania Convention Center Philadelphia, PA USA Andy Ferrell PCI CEO & President

Upload: others

Post on 27-Jan-2020

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Developing and

Implementing Calibration Programs

March 19, 2013 Pennsylvania Convention Center Philadelphia, PA USA

Andy Ferrell PCI CEO & President

Page 2: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Wm. Andy Ferrell - Bio

• Bachelors and Masters degrees from East Carolina University • Engineer with big pharma companies from 1989 to 1996 • Speaking on topics such as:

– Smart Asset Management – Emerging Trends in Calibration Compliance – Predictive Maintenance and Calibration Programs: Improving Operations

• Founded PCI in 1996 • PCI is 110 person firm. Headquartered in RTP, NC with 5 field offices and

national consulting & technical service coverage • PCI’s Technical Services: delivering calibration, maintenance, and repair

services for both manufacturing and laboratory equipment • PCI’s Consulting Services: building calibration & maintenance programs

including developing business workflows and policies, implementing software systems, FDA remediation plan development & execution. We development world class maintenance & calibration programs

Page 3: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Introductions: Who are you?

Name Your role Company Industry Any Regulatory Concerns What are you seeking to get

from this workshop

Page 4: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

What we will cover

Background Info …. Calibrations Basic Calibration Program Requirements What is a Calibration Program? Calibration Program Requirements Responsibilities of a Calibration Department Critical Elements of a Calibration Program Paperless: Changing Existing Cultures Benefits of going to a Paperless Program

Page 5: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Calibration: Definition

“A comparison of a measurement standard (Calibration Standard) of known accuracy with another instrument of unknown accuracy to detect, correlate, report, or eliminate by adjustment any variation in the accuracy of the item being compared.”

Page 6: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Every day, industries throughout the world perform a huge number of measurements.

The results of these measurements are used to make decisions that could affect people’s lives both personally and in the workplace.

Why Calibrate?

Page 7: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Calibration is warranted based on good engineering and business practices, and must adhere to standards and regulations in the following ways:

Good engineering and business sense. Instrumentation wears and settings drift, either mechanically or because of environmental effects. These conditions can have a negative effect on instrumentation and can ultimately affect the quality of a product.

Why Calibration?

Page 8: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Without Calibrations

All of us would pay more at the Gas station.

Food would be weighed incorrectly.

No more watching your favorite television shows or listening to your favorite songs on the radio, because due to calibration errors, the frequency would be off enough that TV and radios would be useless.

Page 9: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Without Calibrations

Ingredients in your prescription drugs could cost more or even more important cause illness or death.

Because of incorrect calibrations, criminals could be either not convicted or released on Bad evidence.

Page 10: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

In pharmaceutical manufacturing and other life science businesses, companies take measurements throughout the various stages of the product life cycle to ensure product integrity.

The certification and calibration of instrumentation, combined with a robust calibration management system, helps achieve product integrity throughout all production processes.

With Calibrations

Page 11: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

The Calibration Train wreck

In January 2005, during the annual Measurement Science Conference (MSC) held in in Anaheim, California, one of the participants from Germany explained to the group that he had been a guest professor at an University in China. He explain that they had an input of 8,000 students every year studying measurement techniques. He went on to explain that the Chinese government has increased the attendance to almost 12,000 students.

Then the question came up, why doesn’t the United States or

Germany train in the field of metrology and calibration the way the Chinese are doing?

His answer was “ Because we haven’t had the train wreck yet.” We haven’t had the train wreck where someone says, “ Calibration was the Problem.” Where there was a great loss of life or limb, or many businesses were effected.”

Page 12: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Regulations & Guidances

Page 13: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Regulations & Guidances

Regulations are Law.

Example: the FDA has GMPs (Good Manufacturing Practices) or CFR 21.

CFR 21 is Code of Federal Regulations for Food and Drugs.

Regulations are usually vague.

Page 14: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Regulations & Guidances

Guidances are not Law. Usually published by industry organizations. Much more specific than Regulations. Following Guidances does not provide 100%

certainty of zero observations or violations.

Page 15: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Regulation & Guidance

Guidance Documents - Benchmarking

ISPE GAMP Calibration Management (second edition)

ISPE GAMP Validation of Lab Computerized Systems

NCSLI RP-6 Calibration Control Systems

ISO 17025 General Requirements for Calibration Labs

ISO 98 Guide for Measurement Uncertainties

Page 16: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Regulation & Guidance

IF we follow all the Guidelines, we are guaranteed to be in compliance with the Regulations.

True? No, False All Guidelines are fallible.

Page 17: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Industry Guidelines

Page 18: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Industry Guidances

NCSL International Recommended Practice RP–6 ISO/IEC 17025:2005 specifies the general

requirements for the competence to carry out tests and/or calibrations, including sampling (proficiency testing)

ANSI/NCSL Z540.3-2006 Requirements for the Calibration of Measuring and Test Equipment

The ISPE GAMP® Good Practice Guide: Calibration Management

Page 19: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Implementing a Calibration Program

Page 20: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Calibration Program Basics

Calibration Policy and support Procedures (SOP) Calibration Records Qualified Personnel Calibration Training Calibration Labeling Calibration Intervals Calibration Logistics / Scheduling Define Out of Tolerance conditions or Non Conformance Process for Investigation of Non Conformances Adequacy of Calibration Standards Traceability (and Reverse Traceability) Analysis of Calibration Data Continuous Improvements (Audits)

Page 21: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Calibration Software Requirements

Page 22: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Implementing a Calibration Software

1. Need to put together a team

2. Obtain Site/Management “Buy In”

3. Obtain appropriate funding

4. Put together timeline and resource allocation

Page 23: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Calibration Software Considerations Implementation budget Platform supported Networked/ web based Number of assets to be managed Number of transactions annually Number of years required to be maintained

User Requirements (URS)

Page 24: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful
Page 25: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Ability to capture unique instrument identification

Ability to assign classification for each instrument in the program

Ability to capture the calibration test points for each instrument

Ability to schedule calibration services for assigned specified intervals

Ability to handle multiple test ranges Ability to record calibration data

User Requirements (URS)

Page 26: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Ability to run a matrix report showing, for a specific time period and by various selection criteria: • Total # of scheduled calibrations • Total # of active and total # of inactive

instruments at the beginning of time period • Total # of scheduled calibrations [redundant, 1st

above] • Total # of active and total # of inactive units as

they were at the end of time period • Net change in total # of scheduled calibrations

User Requirements (URS)

Page 27: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Both internal corporate compliance and third-party regulatory bodies conduct Audits throughout the pharmaceutical industry to ensure that manufacturers are following their own procedures and to see how they react to issues and deviations to their processes.

A robust Calibration Program is vital to ensure that

when similar audit points are challenged, it can be demonstrated that systems, processes, and procedures are in place to address potential noncompliance issues.

Calibration Program Checks and Balances

Page 28: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Performance Metrics

Page 29: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Make strategic objectives clear, in order to focus and bring together the total organization

Tie the core business processes to the objectives

Focus on critical success factors for each of the processes, recognizing there will be variables

Track performance trends and highlight progress and potential problems

Identify possible solutions to the problems

Performance Metrics: Purpose

Page 30: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

What Gets Measured Gets Done.

What Gets Measured, Improves!!!

Performance Metrics

Page 31: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

80.0%82.0%84.0%86.0%88.0%90.0%92.0%94.0%96.0%98.0%100.0%

050

100150200250300350400450500

Jan-04 Feb-04 Mar-04 Apr-04 May-04 Jun-04 Jul-04 Aug-04 Sep-04 Oct-04

% Calibrations performed: Laboratory Instruments

Calibrations Scheduled % calibrations performed

Performance Metrics: Example

Page 32: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Auditing

Page 33: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Developing & Implementing Calibration Programs:

The 4 Most Critical Elements

Page 34: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

The 4 Most Critical Elements of Calibration Programs

• Classification • Intervals • Ranges • Limits

Page 35: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Classification, Intervals, Ranges, and Limits

An instrument/device is classified based on its potential to affect a product or products.

This classification is defined and documented in a paperless system using such terms as “critical,” “non-critical,” or “reference only.” Characteristically, where process parameters are monitored, controlled, or recorded, and are used to determine product quality, the associated instruments would warrant a critical classification.

Most instruments associated with noncritical and utility classifications are used to provide indications of general equipment conditions and for troubleshooting.

Page 36: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Classification, Intervals, Ranges, and Limits

Interval or service frequencies or schedule.

For regulated environments, it is typical to assign Intervals of weekly, monthly, quarterly, semi annual, and annual.

Bi-annual are used but are uncommon.

Daily are typically not used. Calibration Before Use is typically assigned instead of daily intervals.

Page 37: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

• Does a measured and controlled parameter on product have a direct effect on product safety or quality (e.g., sterility)?

• What is the instrument’s (OEM) stated

reliability and accuracy? • What is the time from product manufacture to use?

If this time is relatively short, adulterated product could be distributed & used if improper intervals are set.

Interval Assignment Considerations

Page 38: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

o Instrument Design Range The design range or maximum range of operability of the

instrument, as specified by the manufacturer. This must be wide enough to ensure that the instrument is capable of reliably operating across the entire qualified range for the equipment.

o Process Operating Range Specified range within which the process must operate. o Instrument Calibration Range The calibration range must be, at a minimum, equal to the

qualification range. A wider calibration range, however, can be useful for non-linear instruments or whenever the qualified range is too small to prove linearity of the instrument.

Classification, Intervals, Ranges, and Limits

Page 39: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Considerations • Given to the criticality (or classification) of

the instrument • Given to the environment in which the

instrument is used • Given to the users • Will the instrument be used for the same

test consistently

Classification, Intervals, Ranges, and Limits

Page 40: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

An Example: Instrument’s Design Range 0 deg C to 200 deg C

Operating Range 20 deg C to 80 deg C

Instrument Calibration Range 10 deg C to 100 deg C

Classification, Intervals, Ranges, and Limits (tolerance)

Page 41: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

The calibration tolerance is the maximum allowable deviation of an instrument from true value before there is an effect on product and is utilized for critical instruments/devices. This predetermined tolerance is used as an out-of-tolerance investigation trigger.

The Manufacturer’s Accuracy Specifications should not

be used as the Calibration Tolerance Limit

Classification, Intervals, Ranges, and Limits (tolerance)

Page 42: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

An Example: Instrument’s (OEM) Design Accuracy +/-1.0 deg C

Your Adjustment Limit +/- 1.5 deg C

Your Calibration Tolerance +/- 2.0 deg C

Classification, Intervals, Ranges, and Limits (tolerance)

Page 43: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Instrument Tolerance Relationship .

Page 44: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Calibration Tolerances

Page 45: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Classification, Intervals, Ranges, and Limits (tolerance)

General Rule

If an investigation of an Out of Tolerances does not result in a product rework or recall, the Process Tolerance should be set to avoid doing the same investigation if the instrument fails calibration by a similar or less value

Page 46: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

The Rule of Thumb for Setting Calibration Tolerances

All instruments drift. No exceptions.

Therefore, Never use the Manufacturer’s Specifications for Accuracy as the Calibration Tolerance!

Rule of Thumb: double the OEM’s stated accuracy for your calibration tolerance. If that doesn’t meet your needs, purchase another instrument.

Page 47: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Break

Page 48: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Your Bonus Jean Piaget was a scientist who in the early 1900's studied the tendency to focus attention on only one characteristic of shape. In particular, he studied the human inability to distinguish identical volumes in different sized containers. Piaget studied the inability of small children to understand that cylinders with different aspect ratios could have the same volume. Small children have difficulty conceiving the shift in volume from one dimension (say the width of a glass) to another dimension (say the height of a glass). They have difficulty visualizing that there is no more liquid in a tall, slender glass than in the shorter, wider glass. He showed that for certain individuals, this tendency could extend into adulthood.

Page 49: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

This is illustrated in the following Figure:

Page 50: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

The Visual Measure The tendency is to believe that even though the two glasses have the same volume, the tall, slender glass has more volume because the height of the liquid is greater. This fact explains why studies show that mixed drinks at bars that use short, wide glasses are typically stronger then the same drink at bars that use tall, slender glasses. The height of a 1.5 oz. shot of alcohol in the short, wide glass is not as high as the same 1.5 oz. shot in the tall, slender glass. When bartenders pour a shot in a short, wide glass they have the tendency to think that since the liquid level is not very high they have mis-poured, and so add a little more alcohol to the drink. The moral of this study is always get your mixed drink from bars that use short, wide glasses.

Page 51: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

The Beer Gauge

This misconception of relative volume is what leads us to think that the relatively small top portion of a pint glass could not contain such a large percentage of the total volume. In fact, if beer is poured into a pint glass to about 1/2 inch from the top, 13% of the beer is GONE.

A pour to about 1 inch from the top of the glass leaves out 25% of the beer. This is illustrated below, where the "Beer Gauge" is used to indicate the amount of beer missing from the pint glass.

Page 52: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful
Page 53: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Your Bonus

The PCI Logo’d BEER GAUGE!

Use Responsibly

Page 54: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Paperless Calibration Programs

Why Paperless Calibration?

Page 55: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Why Paperless Calibration?

Digitizing and automating calibration has decreased downtime and improved efficiency

Through better calibration testing and setting of tolerances, firms will find a great cost reductions in operations and fewer Out of Tolerance investigations.

Understand and accept that in the world of Calibration, everything around us is changing, it doesn’t matter if it’s the calibration procedure, calibration records, our calibration standards, instrument owners, or instrument use.

Page 56: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Investigation Activity

Page 57: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Features of a Paperless Calibration Program

Page 58: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Features of a Paperless Calibration System

Investment in a paperless system is easy to justify; a few of the basic advantages include: • Fewer errors occur because planners, managers, and

technicians are reliably notified of pending calibration events.

• Calibration obligations cannot be forgotten or overlooked. • Technicians can no longer perform a calibration without

properly documenting. • The use of unsuitable or out-of-calibration test standards is

no longer an option. • A paperless system ensures that everyone who needs to

know of a specific event is notified in a timely fashion Example: if a device is found out of tolerance or if a calibration failure occurs, notifications are automated.

Page 59: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Features of a Paperless Calibration System

• Worker training and qualifications • Facility documentation • Work flow diagrams • Sample details • Work requested • Work performed (procedures, calibrations, etc.) • Results, reports, graphs, etc.

Page 60: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Records are auditable Records are accessible (by others, and from

year to year) Provides a framework to implement/prove

work controls Automated report generation Records tend to be more complete than

notebook entries. That is, all supporting materials (spreadsheets, plots, models, figures, references, protocol, etc.) are attached, too

Features of a Paperless Calibration System

Page 61: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Overall Program Compliance Improvement

Can be set up to ensure Technicians do not leave

instrumentation out of compliance or hard document GMP issues via the business rules of the application and real time warnings, alerts, and work order application rules.

All Measure Data Test Points can be loaded and

Technicians must follow decision data of Calibration Tolerance limits,

Adjustment Limits and Process Limits. Each of these can

have a business rule that correspond to an application of the system electronically notifying the designated party via email, and shows pending days, hours, etc of work pending.

Page 62: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Culture Challenges

• Anything new is initially resisted: the historical notebook and notebook pages are gone.

• Scientists may hate paper, but experience separation-anxiety in its absence.

• Your notebook is not your private space anymore: viewed by all, pre-signature.

• E-signature compliance is easier to monitor, but resented.

Page 63: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Culture Challenges

•Data and results may be stored on separate computers

•Handwritten notes on paper

•Generated data files are stored on a network server (Linux based, dedicated to and physically located in my lab, backed up corporately) that has terabytes of storage available, and room to expand

•Experimental results are communicated to customers via memos, printouts, email, etc.

Page 64: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Break Anyone?

Page 65: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Paperless Program Considerations

This is a PROJECT.

Page 66: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Paperless Benefits: Business Areas/End Users Engineering Department QC Laboratories Manufacturing Departments R&D Laboratories Identification and specification of business area

and end user needs Identification and specification of the intended

uses of the system Review and approval of the User Requirements

specification document

Page 67: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Paperless Program Considerations Realistic Expectations

Few Companies Achieve 100% Paperless We see a Continuum from 0 to 100%

“Paperless” It doesn’t have to be all or nothing

Corporate Attitudes to support it vary greatly

Investment rates vary

Page 68: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Facilitating regulatory compliance Completeness of records Ease of searching records Managing who has access to information Review and approvals

- Time savings when generating reports - When recreating the past - When archiving data

Cost savings (by-product of the previous benefits listed)

Benefits of a Paperless Calibration System

Page 69: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Benefits of a Paperless Calibration System

Saves time Reduce human error Full Standards traceability Automatic tolerance calculations Automatic flags and notifications for

calibration failure

Page 70: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Benefits of a Paperless Calibration System

Productivity Improvement Print Calibration certificates with pre-

populated data Measurement Data Template reuse Reduce redundant data entry Reduces head count or maximizes resource

allocation

Page 71: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Benefits of a Paperless Calibration System

Improved Compliance – Automated reviews – Less data errors – Faster inspections and audits – Lowers costs related to Paper, filing,

storage

Page 72: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Benefits of a Paperless Calibration System

Management Of Work Work Scheduling Work Assignment Calibration Planning SOP Measurement Data Template Standards

Page 73: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Benefits of a Paperless Calibration System (Recap)

History Record Permanent record Reviews & Approvals Consistent Service Delivery Auto Notifications Audit Trail

Page 74: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Making It Happen

Assess Infrastructure Wireless Mobile Computing Laptops, tablets, netbooks PDA’s, cell phones, etc.

Network Drops Cellular cards

Page 75: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

The ‘Take Away’

A solid Calibration Program is a must in business and science today

A paperless Calibration Program is a commitment from the top

The benefits of a paperless Calibration Program far outweigh the cost IF this approach is right for your organization

Page 76: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

Developing & Implementing Calibration Programs

Q & A

Page 77: Developing and Implementing Calibration Programs...The calibration range must be, at a minimum, equal to the qualification range. A wider calibration range, however, can be useful

THANK YOU

Wm. Andy Ferrell PCI, LLC

'Providing Instrument & Equipment Compliance Solutions Since 1996'

8100 Brownleigh Ste 100-A

Raleigh, NC 27617 [email protected] (919)781-7787 Office