Determination of Trace Elements in Rice Products by Flame and Graphite Furnace Atomic Absorption Spectrometry Rebecca Price, Thermo Fisher Scientific, Cambridge, UK. Application Notes 43019 Key Words Atomic Absorption, Essential Elements, Flame, Graphite Furnace, Rice,Toxic Elements. Key Benefits • The Thermo Scientific iCE 3000 Series Atomic Absorption Spectrometers offers both flame and furnace capabilities for precise and accurate analyses at parts per million and parts per billion levels • Analysis of nutritional and toxic elements in rice products is accurately determined with optimized methods • The wizard-driven Thermo Scientific SOLAAR software allows quick and easy optimization for both flame and furnace analyses. Summary The Thermo Scientific iCE 3000 Series Atomic Absorption Spectrometers are the ideal tool for the accurate and rapid determination of multiple trace elements in rice products. Flame atomic absorption can be used as a fast screening tool for the analysis of nutritional elements such as copper, manganese and zinc, while graphite furnace atomic absorption can be used for the accurate determination of toxic elements such as cadmium and lead. Simple sample preparation and easy method optimization provide a rapid and effective solution for accurate and reliable analysis at minor and trace levels. Introduction Rice is the second most prevalent cereal crop in the world with an annual global production of approximately 600 million tons. It is the staple food of most Asian countries, with a daily consumption per person of between 200 and 400 g. Trace elemental analysis of this crop and its products is therefore important on an essential, nutritional and toxicological level. The analysis of heavy metals is of particular relevance to human health following numerous incidents such as the mass cadmium poisoning of hundreds of people in the Toyama Prefecture, Japan. During the early 20th Century, cadmium was released into the Jinzu River by mining companies in the mountains. The river water was used to irrigate rice fields and cadmium was subsequently absorbed by the growing rice. The effect on local people was softening of bones, anemia, and kidney failure. It became known as “Itai-Itai Byo”, a phrase adopted by the locals to represent the pain caused by the poisoning. Global legislation now exists to regulate the permissible levels of cadmium in foodstuffs, with both China and the EU setting an upper limit of 0.2 mg/kg of cadmium in rice 1,2 . Such legislation has produced a requirement to monitor rice and other foodstuffs for trace metal content and this application note discusses the analysis of copper, zinc, manganese, cadmium and lead in rice products by flame and furnace atomic absorption for this purpose.

Upload: others

Post on 14-Jul-2020

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Determination of Trace Elements in Rice Products by Flame ...static.thermoscientific.com/images/D03249~.pdfResults for the CRM, rice flour and whole rice are shown in Table 1, along

Determination of Trace Elements in Rice Products by Flame and Graphite Furnace Atomic Absorption SpectrometryRebecca Price, Thermo Fisher Scientific, Cambridge, UK.

Ap

plica

tion

No

tes 4

30

19

Key WordsAtomic Absorption, Essential Elements, Flame, Graphite Furnace, Rice,Toxic Elements.

Key Benefits• TheThermoScientificiCE3000SeriesAtomicAbsorptionSpectrometersoffersbothflameandfurnacecapabilitiesforpreciseandaccurateanalysesatpartspermillionandpartsperbillionlevels

• Analysisofnutritionalandtoxicelementsinriceproductsisaccuratelydeterminedwithoptimizedmethods

• Thewizard-drivenThermoScientificSOLAARsoftwareallowsquickandeasyoptimizationforbothflameandfurnaceanalyses.

SummaryTheThermoScientificiCE3000SeriesAtomicAbsorptionSpectrometersaretheidealtoolfortheaccurateandrapiddeterminationofmultipletraceelementsinriceproducts.Flameatomicabsorptioncanbeusedasafastscreeningtoolfortheanalysisofnutritionalelementssuchascopper,manganeseandzinc,whilegraphitefurnaceatomicabsorptioncanbeusedfortheaccuratedeterminationoftoxicelementssuchascadmiumandlead.Simplesamplepreparationandeasymethodoptimizationprovidearapidandeffectivesolutionforaccurateandreliableanalysisatminorandtracelevels.

IntroductionRiceisthesecondmostprevalentcerealcropintheworldwithanannualglobalproductionofapproximately600milliontons.ItisthestaplefoodofmostAsiancountries,withadailyconsumptionperpersonofbetween200and400g.Traceelementalanalysisofthiscropanditsproductsisthereforeimportantonanessential,nutritionalandtoxicologicallevel.TheanalysisofheavymetalsisofparticularrelevancetohumanhealthfollowingnumerousincidentssuchasthemasscadmiumpoisoningofhundredsofpeopleintheToyamaPrefecture,Japan.

Duringtheearly20thCentury,cadmiumwasreleasedintotheJinzuRiverbyminingcompaniesinthemountains.Theriverwaterwasusedtoirrigatericefieldsandcadmiumwassubsequentlyabsorbedbythegrowingrice.Theeffectonlocalpeoplewassofteningofbones,anemia,andkidneyfailure.Itbecameknownas“Itai-ItaiByo”,aphraseadoptedbythelocalstorepresentthepaincausedbythepoisoning.Globallegislationnowexiststoregulatethepermissiblelevelsofcadmiuminfoodstuffs,withbothChinaandtheEUsettinganupperlimitof0.2mg/kgofcadmiuminrice1,2.Suchlegislationhasproducedarequirementtomonitorriceandotherfoodstuffsfortracemetalcontentandthisapplicationnotediscussestheanalysisofcopper,zinc,manganese,cadmiumandleadinriceproductsbyflameandfurnaceatomicabsorptionforthispurpose.

Page 2: Determination of Trace Elements in Rice Products by Flame ...static.thermoscientific.com/images/D03249~.pdfResults for the CRM, rice flour and whole rice are shown in Table 1, along

2 MethodReagentsFor Flame Analysis:• Nitricacid,Traceanalysisgrade.

• Copper,manganeseandzincmasterstandards,1000mg/l.

• Multi-elementstandardspreparedat0,0.5,1,2and5ppmcopper,manganeseandzincusingmasterstandards.Dilutedtovolumewith1%nitricacid.

For Graphite Furnace Analysis:• Nitricacid,Traceanalysisgrade.

• Cadmiumandleadmasterstandards,1000mg/l.

• Cadmiumstandardpreparedat5ppbusingmasterstandard.Dilutedtovolumewith1%nitricacid.

• Leadstandardpreparedat10ppbusingmasterstandard.Dilutedtovolumewith1%nitricacid.

• Furnacematrixmodifier:ammoniumnitrate,20μgin10μlinjectionforcadmium,50μgin10μlinjectionforlead.

All standards and reagents from Fisher Scientific, Loughborough, UK.

Sample Preparation Threesampleswereanalyzed:riceflourCRM(IRMM-804,LGC,Teddington,UK)andtwosamplesofretailproductsfromalocalsupermarket,riceflourandwholewhiterice.Samplesweredriedfor12hoursat85˚C.Thesampleswerecooledandportionsofapproximately0.25gwereaccuratelyweighedandtransferredtomicrowavedigestionvessels.4mlnitricacidwasaddedtoeachvesselandleftuncoveredfor1hour.Afurther4mlnitricacidwasaddedtoeachvessel,thevesselssealedandsamplesdigestedinahighpressureclosedmicrowavedigestionsystem,byrampingover20minutesto170˚C.Sampleswerelefttocoolbeforebeingmadeupto250mlwithdeionizedwaterforcadmiumanalysis.Duplicatesampleswerepreparedfortheanalysisofcopper,lead,manganeseandzinc(fromthesamesample)withthedigestsmadeupto50mlwithdeionizedwater.

Aspikedsampleofthesupermarketriceflourwaspreparedtoassessthesamplepreparationandsubsequentanalysisofcadmiumandleadbygraphitefurnace.Spikeswereaddedtoobtainafinalconcentrationapproximately75%oftheCRMriceflourstandard.

Results and DiscussionFlame: Copper, Manganese and Zinc Copper,manganeseandzincwereanalyzedbyflameatomicabsorptionastheseelementscanbefoundasnaturalconstituentsinsoilsandwater,thereforedetectionwithinthelowmg/kg(ppm)rangewasrequired.Thetransverseandlateralburnerpositionandtheimpactbeadweremanuallyoptimizedusinga5ppmcopperstandardpriortoanalysis.BurnerheightandfuelgasflowwereoptimizedforeachelementusingtheOptimize Gas Flow and Burner HeightWizardintheSOLAARsoftware,asshowninFigure1.Thisallowedquickandeasymethoddevelopmentwithoptimizedparametersenteredautomaticallyintotheanalyticalmethod.

Figure 1. Gas Flow and Burner Height Optimization Wizard.

Expected Concentration

(mg/kg)

Measured Concentration

(mg/kg)

Percentage Recovery

(%)

Method Detection Limit

(mg/kg)

Characteristic Concentration

(mg/kg)

Copper

CRM 0.013 0.012 108.3

Rice Flour 0.007 0.0042 0.0646

Whole Rice 0.014

Manganese

CRM 0.163 0.162 100.6

Rice Flour 0.038 0.0379 0.0102

Whole Rice 0.026

Zinc

CRM 0.110 0.111 99.1

Rice Flour 0.091 0.0019 0.0187

Whole Rice 0.063

Table 1. Table of results and percentage recoveries of copper, manganese and zinc in rice products. Method Detection Limits and Characteristic Concentrations shown are based on an initial mass of 0.25 g sample.

Page 3: Determination of Trace Elements in Rice Products by Flame ...static.thermoscientific.com/images/D03249~.pdfResults for the CRM, rice flour and whole rice are shown in Table 1, along

3Thesamplespreparedtototalvolumeof50mlwereanalyzedforcopper,manganeseandzinc.Eachstandardandsamplewasanalyzedintriplicateusingthefastre-samplingoption.Acontinuoussignalwasmeasuredfor4secondsforeachresample.Thetotalanalysisduration,includingoptimization,calibrationandanalysisofsampleswas25minutes.Thisincludeda5pointcalibrationandanalysisofthethreesamplesforeachelement(Cu,MnandZn).Calibrationcurvesforcopper,manganeseandzincareshowninFigure2.

Figure 2. Calibration curves for the determination of copper, manganese and zinc in rice products

ComparisonstotheCRMexpectedconcentrationweremadetoassessthesuitabilityofthemethod.ResultsfortheCRM,riceflourandwholericeareshownin

Table1,alongwiththemethoddetectionlimits(MDLs)andcharacteristicconcentrations(CCs)determinedusingtheautomatedInstrument PerformanceWizardintheSOLAARsoftware.

Furnace: Cadmium and LeadAstoxicelementswithnonutritionalbenefit,detectionofcadmiumandleadwasrequired.Asthiswasattracelevel,graphitefurnaceatomicabsorptionwasselectedastheappropriateanalysistechnique.Whilealongeranalysistimeisrequiredpersample,theuseoftheintegratedautosamplerallowedsamplestoberununattended.Ammoniumnitratematrixmodifierwasusedforboththeanalysisofcadmiumandlead.Aworkingvolumeof20μlwasusedforstandardandsampleinjection,withanadditional10μlaliquotofmatrixmodifieraddedinacombinedwetinjection.20μgofammoniumnitratewasaddedforcadmiumand50μgforlead.GraphiteFurnaceTeleVision(GFTV)wasusedtoobserveinjectiondeposi-tionandthedryingphase.ThisuniquefeatureoftheiCE3000SeriesAtomicAbsorptionSpectrometersallowsdirectvisualizationoftheinsideofthecuvette,displayingahighresolutionimageonthePCmonitorforenhancedmethoddevelopment.TheOptimize Furnace ParametersWizardintheSOLAARsoftwarewasusedtooptimizetheashandatomizephases.Zeemanbackgroundcorrectionwasselectedforallanalysestoeliminatetheeffectofpotentialstructuredbackgroundinterferences.FurnaceparametersforcadmiumandleadareshowninTable2:

Table 2. Furnace parameters for the analysis of cadmium and lead in rice products.

Calibrationswithintherange0-5ppbforcadmiumand0-10ppbforleadwereobtainedusingthemasterstandards.StandardswereautomaticallydilutedusingthefixedvolumestandardpreparationoptionintheSOLAARsoftwareasshowninFigure3:

Figure 3. Sampling conditions for the analysis of cadmium and lead in rice products.

TheCRMwasanalyzedtoassessthesuitabilityofthemethod,followedbythericeflourandwholerice.Spikedriceflourwasalsoanalyzedtoassessrecoveryperfor-manceofthemethod.Threerepeatsofeachstandardandsamplewereanalyzed.ThecalibrationlinegeneratedbytheSOLAARsoftwareforleadisshownin

Figure4,withacorrelationcoefficientof0.9957.ResultsfortheCRM,riceflourandwholericeareshownbelowinTable3alongwiththemethoddetectionlimits(MDLs)andcharacteristicconcentrations(CCs)determinedusingtheautomatedInstrument PerformanceWizardintheSOLAARsoftware.Cadmiumandleadwerenotdetectedineitheroftheretailproductsdisplayingcompliancewithcurrentinternationalregulations.

Figure 4. Calibration line for the analysis of lead in rice products.

Phase Temp. ˚C Time / s Ramp ˚C/s

1: Drying 100 Cd, 115 Pb 30 10

2: Ashing 700 Cd, 1000 Pb 20 150

3: Atomizing 1300 Cd, 1500 Pb 5 Cd, 3 Pb n/a

4: Cleaning 2500 3 n/a

Page 4: Determination of Trace Elements in Rice Products by Flame ...static.thermoscientific.com/images/D03249~.pdfResults for the CRM, rice flour and whole rice are shown in Table 1, along

Thermo Electron Manufacturing Ltd (Cambridge) is ISO Certified.

thermoscientific.com © 2012 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

Africa-Other +27 11 570 1840Australia +61 2 8844 9500Austria +43 1 333 50 34 0Belgium +32 53 73 42 41Canada +1 800 530 8447China +86 10 8419 3588Denmark +45 70 23 62 60Europe-Other +43 1 333 50 34 0

Finland /Norway/Sweden +46 8 556 468 00France +33 1 60 92 48 00Germany +49 6103 408 1014India +91 22 6742 9434Italy +39 02 950 591Japan +81 45 453 9100Latin America +1 608 276 5659

Middle East +43 1 333 50 34 0Netherlands +31 76 579 55 55South Africa +27 11 570 1840Spain +34 914 845 965Switzerland +41 61 716 77 00UK +44 1442 233555USA +1 800 532 4752

AN43019_E 05/12C

Ap

plica

tion

No

tes 4

30

19

Conclusion TheresultsdemonstratetheeasewithwhichtheThermoScientificiCE3000SeriesAtomicAbsorptionSpectrometerscanbeusedforthemultielementanalysisofriceproducts.Flameatomicabsorptionwasusedasafastanalysistoolforthedeterminationofcopper,manganeseandzinc.Detectionlimitswerebelow5ug/kgforcopperandzincandbelow40ug/kgformanganese.Graphitefurnaceatomicabsorptionwasusedasanaccurateanalysistoolforthedeterminationofcadmiumandlead,withexcellentdetectionlimitsof0.03and0.2ug/kgobtainedrespectively,easilyensuringregulatorycompliance.

ACertifiedReferenceMaterialwasusedtoverifyboththeflameandfurnacemethodswithexcellentrecoveriesobtainedTheeaseofuseofthesystemisexemplifiedwiththemanyautomatedwizardswhichenablefastaccurateresultsregardlessofanalystexperience.TheThermoScientificiCE3000SeriesAtomicAbsorptionSpectrometersthereforeprovideanidealsolutionfortheaccurateandrapidmultielementdeterminationofminorandtraceelementsinriceatpartspermillionandpartsperbillionconcentrations.

References1.PeoplesRepublicofChina,FAIRSProductSpecificMaximumLevelsofContaminantsinFood2006

2.CommissionRegulation(EC)No.1881/2006of19December2006settingMaximumLevelsforCertainContaminantsinFoodstuffs

Measured Concentration

(mg/kg)

Expected Concentration

(mg/kg)

Percentage Recovery

(%)

Method Detection Limit

(µg/kg)

Characteristic Concentration

(µg/kg)

Cadmium

CRM 1.5 1.43 +/- 0.07 105%

Rice Flour <DL 0.0303 0.0600

Spiked Rice Flour 1.06 1.00 106%

Whole Rice <DL

Lead

CRM 2.03 2.00 +/- 0.07 102%

Rice Flour <DL 0.1960 0.7119

Spiked Rice Flour 1.24 1.20 103%

Whole Rice <DL

Table 3. Table of results and recoveries of cadmium and lead in rice. MDLs and CCs shown are based on an initial mass of 0.25g sample.