Design and Control Challenges for a Surgical Radiographic Imaging System
Tim ElmoreOrthopaedic Biomechanics Laboratory
Department of Mechanical & Aerospace EngineeringUniversity of Florida
![Page 2: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/2.jpg)
The Motivation
What surgeons can see… What surgeons want to see…
http://www.theploughhundon.co.uk/steak%20tuesday.jpghttp://www.flickr.com/photos/stanfordmedicine/3660598599/
![Page 3: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/3.jpg)
The Motivation
The standard in industry …
Over 100 ft3 total volume!
Current O-arm technology remains in sterile field, giving surgeon limited access to patient
![Page 4: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/4.jpg)
+ =
The Motivation
Existing project development… Gator-Ray !
![Page 5: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/5.jpg)
The Motivation
Existing project development… Gator-Ray !
![Page 6: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/6.jpg)
Design Challenges for Surgical GatoRay
• Kinematic– DOF– Obstacles– Reach– Accuracy
• Kinetic– Payload mass– Velocity/Acceleration– Bearing life
OR table, surgeons, equipment
Under OR table to lateral view
+/- 1mm, +/-1 deg
30-60kg (65-130lbs)
0.25m/s, 2.5m/s^2
1,500,000 cycles (100/day, 40yr)
5 degrees of freedom
![Page 7: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/7.jpg)
Mitsubishi PA10-6CE Specifications
Six Degree-of-Freedom Robot Arm
3-phase AC Brushless Servo Motors
Harmonic Drive Transmissions
Manufacturer Specified Positional Repeatability of
±0.1mm ANDJoint Resolver Accuracy
= ±0.44 arc min (0.0073º)
10kg payload limit
![Page 8: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/8.jpg)
Mitsubishi PA10-6CE Specifications
This payload would result in a 700% overload!
0 20 40 60 80 100 120 140 160 1800
5
10
15
20
25
30
35
Torque Limit
Inertia Limit
Payload
Rbar, distance from EE to CoM
kg
lim
it
![Page 9: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/9.jpg)
Alternative Manipulators
Panasonic VR-032 RobotAxes: 6Payload: 32 kgH-Reach: 1866 mmRepeatability: ±0.15 mmRobot Mass: 580 kgMounting: floor, ceiling
Disadvantages:- Size- Mass- EE rotations are translation
dependent- Cost ($20,000 used)
Table Height:32 inches/810mm
![Page 10: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/10.jpg)
Alternative Manipulators
Parallel ManipulatorsDegrees of Freedom: 6Payload: high (>30kg)Repeatability: good (<1mm)Robot Mass: low (<200 kg)
Disadvantage:- Limited rotation angles- Limited translations
![Page 11: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/11.jpg)
Alternative Manipulators
Ballscrew/Rotary tablesDegrees of Freedom: 1 per tablePayload: high (>100kg)Repeatability: good (<1mm)Robot Mass: low (<200 kg)
Advantages:- Each axis is customizable- Mass supported by frame, not
torque
![Page 12: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/12.jpg)
The Proposed System (v0.1)
X-Ray Source
Undertable
Robot
![Page 13: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/13.jpg)
The Proposed System (v1.0)
X-Ray Sensor
PA-10
X-Ray Source
Under table
Robot
![Page 14: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/14.jpg)
The Proposed System (v1.0)
X-Ray Source
Undertable
Robot
X
YZ
![Page 15: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/15.jpg)
The Proposed System (v1.0)
![Page 16: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/16.jpg)
Design Goals for Surgical GatoRay
• Kinematic– 5 DOF (3 translation, 2 rotation)– Reach around table into lateral view– Displacement: 3 translation, 2 rotation– Accuracy 0.2mm/axis, 1mm total
• Kinetic– 30-60kg payload– 0.5m/sec velocity, 5m/sec2 acceleration– 1,500,000 cycles at 30kg load
![Page 17: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/17.jpg)
Kinematics
• By inspection, task space dependencies:– Rx = f(xjs,Θzjs)
– Ry = f(yjs,Θzjs)
– Rz = f(zjs)
– Θx = f(Θxjs)
– Θz = f(Θxjs, xjs, yjs) X
YZ
![Page 18: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/18.jpg)
Kinematics
![Page 19: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/19.jpg)
Kinetics
• Motor choices– Servo
• Pros: torque, speed• Cons: size, price, control
– Stepper• Pros: size, price, control• Cons: torque, speed
![Page 20: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/20.jpg)
Kinetics
Servo Control
Stepper Control
Input Output
Input Output
![Page 21: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/21.jpg)
Stepper Motors
![Page 22: Design and Control Challenges for a Surgical Radiographic Imaging System Tim Elmore Orthopaedic Biomechanics Laboratory Department of Mechanical & Aerospace](https://reader036.vdocuments.site/reader036/viewer/2022062321/56649f145503460f94c2931e/html5/thumbnails/22.jpg)
Stepper Motors
125kg horizontal payload 30kg vertical payload
time time