Design and construction of a
mid-IR SPIDER apparatus
09/10/2012Malte Christian BrahmsImperial College London
09/10/2012
![Page 2: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/2.jpg)
2
Contents
• Purpose of SPIDER• Working principle• Design of mid-IR SPIDER• User’s manual• Outlook
09/10/2012
![Page 3: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/3.jpg)
3
Purpose
• Pulse characterised completely by – Field as fct. of time or – Spectral amplitude and phase
• Oscillations in on the order of fs– Temporal resolution required too high
• Instead: Measure spectral amplitude and phase:
Spectral Phase Interferometry for Direct Electric Field Reconstruction
(SPIDER)09/10/2012
![Page 4: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/4.jpg)
4
Purpose
• SPIDERs available commercially • Why build one?– Unusual wavelength in mid-IR– Specific variable wavelength SPIDER
needed for TOPAS
09/10/2012
![Page 5: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/5.jpg)
5
Contents
• Purpose of SPIDER• Working principle• Design of mid-IR SPIDER• User’s manual• Outlook
09/10/2012
![Page 6: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/6.jpg)
6
Working principle
• Based on two replicas of the test pulse
• Displaced relative to each other – In time: Delay τ– In frequency: Shear Ω
• In frequency domain:
09/10/2012
Frequency Shear
Time delay
![Page 7: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/7.jpg)
7
Working principle
Interferogram:
09/10/2012
![Page 8: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/8.jpg)
8
Working principle
09/10/2012
![Page 9: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/9.jpg)
9
Working principle
09/10/2012
![Page 10: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/10.jpg)
10
Contents
• Purpose of SPIDER• Working principle• Design of mid-IR SPIDER• User’s manual• Outlook
09/10/2012
![Page 11: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/11.jpg)
11
Design
• Two problems to solve:– Time delay– Frequency shear
• Solutions:– Delay: Split mirror• Allows variable delay
– Shear: SFG in BBO with chirped pulse
09/10/2012
![Page 12: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/12.jpg)
12
Design – Time delay
09/10/2012
Split mirrors
![Page 13: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/13.jpg)
13
Design – Frequency shearing
• Use SFG with chirped ancillary pulse replica– Chirp: Carrier frequency depends linearly on
time:
– SFG with same original frequency at two different times different output frequencies:
• Obvious solution: Piece of glass– But: Most glass almost dispersion-free at ca.
1500nm09/10/2012
Stretched ancilla
Test pulse replicas
![Page 14: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/14.jpg)
14
Design – Frequency shearing
Chirping the ancillary pulseGrating Compressor:
09/10/2012Image: 2010 J. Phys. B: At. Mol. Opt. Phys. 43 103001 , p.5
Walmsley et al.: The role of Dispersion in ultrafast optics, Rev. Sci. Instrum., Vol. 27,
No.1, Jan 2001, p. 7
![Page 15: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/15.jpg)
15
Design – Recombination
09/10/2012
Sum-frequency generation• Focus delayed pulse-pair and chirped
ancillary pulse into BBO crystal
Image: 2010 J. Phys. B: At. Mol. Opt. Phys. 43 103001 p. 25
SHG + Original
SHG + Original
SFG
![Page 16: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/16.jpg)
16
Design – Constraints
• On delay τ: Spectrometer resolution and shear– Fringe spacing 2π/τ must be sufficiently large
τ cannot be too large– τ must be sufficiently large resolve peaks in FT
• On shear Ω:– Small enough to satisfy the sampling theorem:
– Satisfied in practice, usual: Ω=10% of bandwidth
09/10/2012
T: Reconstruction window
![Page 17: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/17.jpg)
17
Design – Constraints
• On chirp : – Only SFG with monochromatic field
conserves shape of spectrum:• In time domain: • In frequency domain (if Eanc is
monochromatic):
– Need slow-varying, quasi-monochromatic frequency lower limit on
09/10/2012
![Page 18: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/18.jpg)
18
Design - Constraints
• Resolution:
• Shear size:
• Lower limit:
09/10/2012
![Page 19: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/19.jpg)
19
Design – Constraints
• With numbers:
At 1300-2000nm for a ca. 40fs pulse
• Both the constraints and the value are wavelength-dependent!
• Chirp determined by compressor angle and separation– Can vary angle easily, use mostly– If necessary, enough space to move one
grating
09/10/2012
![Page 20: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/20.jpg)
20
Design – Setup
09/10/2012
Spectrometer
A1A2
A3 f=100mm
f=100mm
D-mirror
D-mirror
BS
(400-1100nm)
![Page 21: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/21.jpg)
21
Contents
• Purpose of SPIDER• Working principle• Design of mid-IR SPIDER• User’s manual• Outlook
09/10/2012
![Page 22: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/22.jpg)
22
User’s Manual – Practical issues
• Calibration by simply adding ignores frequency-dependent behaviour (noise) in apparatus– Need to calibrate for this– Use signal without shear instead:
• Depending on wavelength and spectrometer:– Measure calibration either in SHG (blue) or original
(red)– SPIDER in SFG
• For this SPIDER at 1300-2000nm:– Measure calibration and SPIDER signal in blue, first
order
09/10/2012
![Page 23: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/23.jpg)
23
User’s Manual – Practical issues
• Spectral amplitude:– Determined from calibration signal• Take magnitude instead of phase angle:
• Determining shear Ω:–Measure compressor parameters or
block one of sheared pair (spatially)– In practice: Use compressor parameters
• τ determined by software
09/10/2012
![Page 24: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/24.jpg)
24
User’s manual – Procedure
• Good signal:– High fringe contrast– Well separated fringes– Near saturation
• Needed for this:– Phase-matching and crystal position– Spatial overlap (focus on crystal)– Temporal overlap (compressor translation)– Right amount of delay– SHG beams blocked (use aperture A3)
09/10/2012
![Page 25: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/25.jpg)
25
User’s manual – Procedure
1. Take reading without ancillary pulse (calibration)– Block after beam splitter
2. Take SPIDER signal– Unblock ancillary, take trace
3. Analyse• OR: Use Tobi’s TwinSPIDER (live
operation)– Ask Tobi for more advice
09/10/2012
![Page 26: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/26.jpg)
26
User’s Manual – Software
09/10/2012
![Page 27: Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121](https://reader035.vdocuments.site/reader035/viewer/2022062620/551ac194550346b2288b53e5/html5/thumbnails/27.jpg)
27
Outlook
• This week or next: Use on TOPAS• Later: Test on few-cycle pulses• Intended for long-term use on TOPAS