demystifying cardiac assist · pdf filedemystifying cardiac assist devices 1 2014 cynthia...
TRANSCRIPT
DEMYSTIFYING CARDIAC ASSIST DEVICES
1
2014 Cynthia Webner DNP, RN, CCNS, CCRN-CMC
“I ATTRIBUTE MY
SUCCESS TO
THIS – I NEVER
GAVE OR TOOK
ANY EXCUSES.”
~ FLORENCE
NIGHTINGALE
2
Objectives
Discuss pacing modes utilized to minimize right
ventricular pacing and preserve normal cardiac
function.
Review cardiac pacing modalities used in the
treatment of heart failure and the prevention of
sudden cardiac death.
Identify criteria for consideration of mechanical
assist devices in acute decompensated heart failure
or shock states.
3
PACING BASICS
4
Indications for Pacing
Symptomatic
bradycardia
Sinus node
dysfunction
AV conduction system
dysfunction - #1
Drug induced
bradycardia
5
Indications for Pacing 6
Symptomatic bradycardia
Symptomatic heart blocks
Chronic bifasicular and trifasicular blocks
Sick Sinus Syndrome
Neurocardiogenic Syncope
Hypertrophic Cardiomyopathy
Cardiac support for treatment of arrhythmias requiring ablation and / or medications resulting in bradycardia
Pacing for termination of tachyarrhythmias (part of ICD therapy)
CHF (biventricular pacing)
Types of Cardiac Pacing 7
Temporary Transvenous Pacing
Transcutaneous Pacing
Epicardial Pacing
Permanent Pacing
Single chamber
Dual chamber
Biventricular
Rate adaptive pacing
Temporary Pacemakers 8
Transcutaneous Pacer 9
Epicardial Pacer 10
Permanent
Pacemakers 11
Pacer Parts
Battery / Brains Leads
12
Implantation of Permanent Pacer
13
14
What kind of Bundle Branch Block does RV
pacing mimic?
14
15
Left Bundle Branch Block
V1 = QS
V6 = wide R
QRS = .12 sec or more
V1 = rS
Pacemaker Function 16
Bipolar vs. Unipolar 17
18 Pace
• Ability of the pacemaker to send a stimulus to the myocardium
• Identified by a pacemaker spike on the ECG
Capture
• Ability of the pacing stimulus to depolarize chamber being paced
• Identified by a pacemaker spike that is immediately followed by a P wave or a QRS complex on the ECG
Sense
• Ability of the pacemaker to recognize and respond to intrinsic cardiac depolarization
• Identified by pacing when no intrinsic beats and not pacing when intrinsic beats are present
Pacing 19
Identify automatic pacing interval (pacing rate)
Two consecutive pacer spikes
Spikes should appear regularly unless pacer is
inhibited by sensed intrinsic activity
Capture 20
Pacing stimulus results in depolarization of chamber
being paced
Each spike should be followed by a QRS unless it
falls in heart’s refractory period
Sensing 21
Pacemaker sees and responds to intrinsic activity
Must be given opportunity to sense
Must be in demand mode
There must be intrinsic activity to be sensed
22
Position I Position II Position
III Position
IV Position V
Chamber(s) Paced
Chamber(s) Sensed
Response to Sensing
Rate Modulation
Multisite
O=None O=None O=None O=None O=None
A=Atrium A=Atrium T=Triggered R=Rate
modulation P=Paced
V=Ventricle V=Ventricle I=Inhibited S=Shocks
D=Dual (A+V)
D=Dual (A+V)
D=Dual (T+I)
D=Dual (P+SV)
(Bernstein et al., 2002)
Revised NASPE/BPEG Generic
Code for Antitachycardia Pacing
UNDERSTANDING PACEMAKER
FUNCTION
23
24
25
AAI Pacing – Atrial Inhibited
AAI
Paces the Atrium
AAI
Senses the Atrium
AAI
Atrial sensing inhibits atrial pacing
Pacing Modes AAI 26
VVI Pacing – Ventricular Inhibited 27
VVI
Paces the Ventricle
VVI
Senses the Ventricle
VVI
Ventricular sensing inhibits ventricular pacing
VVI Pacing 28
Dual Chamber Pacers 29
Provide AV synchrony Maintains atrial kick
Improves hemodynamics in those with heart blocks
Tracks atrial activity Ventricular pacing occurs in response to atrial activity
Improved hemodynamics
Decreased incidence of pacemaker syndrome
Basic Pacemaker Timing 30
AV Interval
Period of time between an atrial event (sensed “P” wave or atrial pace) and a paced ventricular event
VV Interval
Period of time from ventricular complex to ventricular complex
VA Interval
Ventricular complex to atrial activity
Also called AEI or atrial escape interval
Basic Pacemaker Timing
Refractory Period
Brief period of time when pacer is not allowed to
look for intrinsic events
Absolute Refractory Period
Nothing can be sensed
Relative Refractory Period
Allows sensing but pacer will not respond
31
Basic Pacemaker Timing 32
Low Rate
Lowest rate allowed by the pacer before a paced
beat is initiated
High Rate
Upper rate limit
Highest rate that can be achieved and still maintain
AV synchrony
DDDR Pacing 33
DDDR
Paces both Atrium and Ventricle
DDDR
Senses both Atrium and Ventricle
DDDR
1. Atrial sensing inhibits atrial pacing and triggers ventricular pacing
2. Ventricular sensing inhibits ventricular and atrial pacing
34
DDD Pacing:
AV Sequential Pacing State 35
DDD Pacing:
Atrial Pacing State 36
DDD Pacing:
Atrial Tracking State 37
DDD Pacing:
Atrial Sensing and Ventricular Sensing State 38
Minimizing Right Ventricular Pacing
1/9/2014
39
1/9/2014 40
41
Left Bundle Branch Block
V1 = QS
V6 = wide R
QRS = .12 sec or more
V1 = rS
V1
V6
Understanding Dyssynchrony 42
Normal function:
Septum moves towards left ventricle during ventricular
contraction
Mitral valve papillary muscles contract slightly before
LV free wall
Understanding Dyssynchrony 43
Electrical Abnormalities with LBBB
RV depolarizes normally
Septum depolarizes after activated by the right bundle
branch and before the left ventricle
Left ventricle depolarizes late
Understanding Dyssynchrony 44
Mechanical abnormalities with LBBB LV activation delayed Septum completes contraction before LV contracts Septum bulges into RV when LV contracts Mitral valve papillary muscle contract late allowing
leaflets to open into LA resulting in mitral regurgitation
Minimizing RV Pacing with Dual
Chamber Pacers 45
Increased hospitalizations for HF (DAVID Trial)
Increased mortality (DAVID Trial)
No improvement in mortality, HF hospitalizations or stroke free survival when compared to VVI (MOST Trial, CTOPP Trial)
AAI pacing demonstrates improved outcomes
Reducing RV pacing to less than 10% in patients with dual chamber pacemakers reduced the relative risk of developing persistent atrial fibrillation by 40% compared to conventional dual chamber pacing (SAVE PACe Trial)
Minimizing Right Ventricular Pacing 46
RV pacing results in mechanical dysynchrony (mechanical LBBB)
Similar changes occur as do with LBBB normally: LV remodeling
Systolic dysfunction
Decreased perfusion
Wall motion abnormalities
Mitral valve regurgitation
Increased risk of AF and HF
Minimizing Right Ventricular Pacing 47
Pacer Lead Placement Options
His Bundle
RV outflow tract
RV septal sites
Dual pacers in RV
LV pacing
Biventricular pacing
Minimizing Right Ventricular Pacing 48
Programming Options
DDIR mode
AAIR mode with mode switching
VVI mode with low rate for those being paced as
defibrillation back up only
Long AV delays
49
DDI Pacing
DDI
Paces both Atrium and Ventricle
DDI
Senses both Atrium and Ventricle
DDI
1. Atrial sensing inhibits atrial pacing and DOES NOT trigger ventricular pacing
2. Ventricular sensing inhibits ventricular and atrial pacing
DDI Pacing 50
LRI
Managed Ventricular Pacing 51
Promotes Intrinsic Conduction
Reduces unnecessary RV pacing
Risk of atrial fibrillation increases as the percentage of
ventricular pacing increases
AAI(R) pacing
Back up dual chamber system available
Medtronic Program
VIP (Ventricular Intrinsic Preference) – St. Jude
Managed Ventricular Pacing 52
Managed Ventricular Pacing 53
AV Search Hysteresis 54
Guidant Program
Functions in DDD/R mode and automatically searches for
intrinsic AV conduction by extending the AV delay by 10% -
100% (programmable value) to look for intrinsic conduction
If intrinsic conduction is present, AV delay remains long until
conduction fails, then pacer returns to DDD/R mode
Search AV Operation 55
• Search AV periodically measures AV intervals
• Determine the effect of Search AV delay in reducing unnecessary ventricular pacing, especially in patients with 1:1 conduction
• Encourage intrinsic conduction
Search AV extension
Cardiac Resynchronization Therapy 56
Cardiac Resynchronization Therapy 57
Treatment modality for heart failure not just pacing
Treatment modality in conjunction with drug therapy
Goals:
Improve hemodynamics by restoring synchrony of
ventricular contraction
Improve quality of life
Decrease mortality and morbidity
Cardiac Resynchronization Therapy 58
Improves ventricular systolic function with reduced metabolic
costs
Decreases functional mitral regurgitation
Induces favorable remodeling with reduction of cardiac
chamber dimensions
Mortality reduction – 24-36%
Reduction in hospitalizations - 30%
Improved 6 minute walk tests
Improvement by at least one NYHA class
Clinical improvement in quality of life
Improved ejection fraction
Improved peak oxygen consumption
Cardiac Resynchronization Therapy 59
•Septum contracts with LV
•Increased LA filling time – improved “atrial kick”
•Improves papillary muscle contraction - ↓ MR
•Reverse remodeling
Understanding Dyssynchrony 60
Normal function:
Septum moves towards left ventricle during ventricular
contraction
Mitral valve papillary muscles contract slightly before
LV free wall
Understanding Dyssynchrony 61
Electrical Abnormalities with LBBB
RV depolarizes normally
Septum depolarizes after activated by the right bundle
branch and before the left ventricle
Left ventricle depolarizes late
Understanding Dyssynchrony 62
Mechanical abnormalities with LBBB LV activation delayed Septum completes contraction before LV contracts Septum bulges into RV when LV contracts Mitral valve papillary muscle contract late allowing
leaflets to open into LA resulting in mitral regurgitation
Indications for CRT 63
For patients who have LVEF less than or equal
to 35%
QRS duration greater than or equal to 0.12
seconds
Sinus rhythm
NYHA functional Class III or ambulatory Class IV
heart failure symptoms with optimal recommend
medical therapy
64
Lead Location
HOW DO YOU KNOW YOUR PATIENT
IS RECEIVING RESYNCHRONIZATION
THERAPY BY LOOKING AT THE
RHYTHM STRIP?
65
66
67
68
Internal Monitoring with CRT 69
Heart Rate Variability
Patient Activity
Night Heart Rate
Impedance
CRT 70
Goal: Force biventricular pacing
Goal: Ventricular Pacing 90% of time or greater
Causes of Loss of Bi V pacing:
Long AV Delays
Prolonged PVARP
ST with 1 degree AV Block
Lead dislodgement
Dealing with Non Responders 71
30% are “nonresponders”
Hemodynamics
Clinical Assessment
Echocardiogram
Location of lead
Programming
AUTOMATIC IMPLANTABLE
CARDIOVERTER DEFIBRILLATORS
72
Sudden Cardiac Death 73
Cardiac arrest with cessation of cardiac function
Ventricular fibrillation or pulseless VT
Pulseless electrical activity
Asystole
Occurs most often in patients with:
CAD
Cardiomyopathy
Indications for ICD
Survivors of cardiac arrest due to VF or
hemodynamically unstable sustained VT after
evaluation to define the cause of the event and to
exclude any completely reversible causes.
Structural heart disease and spontaneous sustained
VT, whether hemodynamically stable or unstable.
Syncope of undetermined origin with clinically
relevant, hemodynamically significant sustained VT
or VF induced at electrophysiological study.
74
Indications for ICD
LVEF <35% due to prior MI > 40 days old and NYHA functional Class II or III.
Nonischemic DCM, LVEF <35% and NYHA functional Class II or III.
LV dysfunction due to prior MI > 40 days old, LVEF <30%, and are in NYHA functional Class I.
Nonsustained VT due to prior MI, LVEF <40%, and inducible VF or sustained VT at electrophysiological
Other Indications 76
Patients with long QT syndrome and prior cardiac arrest, syncope, and strong family history of sudden death, or intolerance to beta blockers.
Patients with HCM with family history of sudden death, syncope, nonsustained VT, abnormal blood pressure response to exercise, left ventricular wall thickness > 30 mm.
Secondary prevention in patients with arrhythmogenic right ventricular dysplasia, but primary prevention in patients with syncope or inducible VT.
ICD Device 77
Pulse Generator
Single chamber, dual chamber, or biventricular pacing
Back up pacing
Antitachycardia pacing
Implanted subcutaneously – same as pacemaker
Defibrillator lead
Detects arrhythmias
Delivers therapy
Defibrillator lead capable of pacing and defibrillating
Placed in right ventricle
ICD Function – Rhythm Detection 78
Heart Rate
Monitors ventricular rate and delivers therapy when rate
exceeds programmed tachycardia detection rate
Defined rate boundaries
Tachycardia zones
Sudden Onset
Detects sudden shortening of cycle length
ICD Function – Rhythm Detection 79
Interval stability
Looks for variability in cycle lengths
Differentiates regular from irregular rhythms
Morphology
Measures width of electogram
Only treats if width is greater than programmed value
ICD Termination Therapies 80
ATP-Antitachycardia Pacing
Painless
“Slow” VT’s
Burst
Ramp
Decremental Scanning
Cardioversion Shock
Defibrillating Shock
ICD Function
ATP-Anti tachycardia Pacing
Tiered Antiarrhythmia Therapies
81
ICD Functions
Cardioversion Shock Delivers shocks from 0.1 to 30 joules synchronized on the R wave
82
ICD Function 83
Defibrillating Shock Delivers high energy (20-34 joules) unsynchronized shock for
VF
Other Device Features 84
Brady Pacing
Atrial Diagnostics
Differentiates between SVT and VT
Stored Electrograms
Store arrhythmia event
Noninvasive EPS
EP study through implantable leads
Care of Patient With ICD 85
Implantation
Cath Lab, EP Lab, OR
Similar to permanent pacer
General anesthesia or conscious sedation
Note Status of Device
Is device on or off
Know what therapies will be delivered for detection of ventricular arrhythmias: pacing, cardioversion, defibrillation
Care of Patient With ICD 86
Emergency Care for VT/VF
Device will deliver therapy within 10-15 seconds and will
continue to deliver therapy as programmed
DO NOT WAIT for device to deliver all its therapies if
patient is hemodynamically unstable or in VF
Defibrillate if necessary – avoid placing paddles directly
over device
Assure patient
Document rhythm
Care of the Patient with ICD 87
Inappropriate firing of ICD Device may fire with SVT’s
Notify physician to have device deactivated
All ICDs can be turned off using a programmer
A round magnet over the generator will deactivate arrhythmia detection
Removal of magnet will reactivate arrhythmia detection
Considerations when the patient requires surgery
NOTE: Magnet works differently for ICD’s than pacemakers!
Care of the Patient with ICD
Patient / Family Education
Reason for ICD, how it works, what to expect
Carry ID card always
Continue to take antiarrhythmic medications if on them
Importance of follow up visits
Every 4-6 moths
Family should learn CPR
Activities
Contact sports restricted
Driving may be restricted
Swimming and boating OK but not alone
Support groups
88
What To Do When ICD Fires 89
If patient is aware of rhythm sit or lie down
If receive only one shock – notify MD
If receive multiple shocks or feels terrible after one
shock– call 911
If device fires and patient does not wake up
immediately call 911.
Device Options
Acute Decompensated HF or Shock
States 90
0
1
4
3
2
5
20 18 16 14 12 10 8 6 4 2 32 30 28 26 24 22 34 36
Preload: PWP, lung sounds (dry or wet)
Fo
rwa
rds F
low
:
CI,
Sk
in t
em
p (
wa
rm o
r co
ld)
Normal Hemodynamics (I) No pulmonary congestion:
• PWP < 18; Dry lungs No hypoperfusion:
• CI > 2.2; Warm skin
Backwards Failure (II) Pulmonary congestion
• PWP > 18; Wet lungs No hypoperfusion
• CI > 2.2; Warm skin
Forwards Failure (III) No pulmonary congestion
• PWP < 18; Dry lungs Hypoperfusion
• CI < 2.2; Cold skin
The Shock Box (IV) Pulmonary congestion
• PWP > 18; Wet lungs Hypoperfusion
• CI < 2.2; Cold skin
91
Hemodynamic and Clinical Subsets
Treatment for Acute Decompensated
Heart Failure
Congestion with
Adequate Perfusion
Subset II
Reduce Preload
Hypoperfusion with
No Congestion Subset III
Increase contractility
Assure adequate preload
Hypoperfusion with
Congestion
Subset IV
Reduce Afterload
92
Bridge to transplant (BBT) for those who are
transplant eligible
Destination therapy (DT) for those who are not
transplant eligible.
Careful consideration for all therapies
Some patients may be too ill with multisystem issues to
benefit from MCS
Some decisions are best made in the hands of the most
experienced centers
Mechanical Circulatory Support in ADHF 93
Intra Aortic Balloon Pump
Most often 1st step in cardiogenic shock treatment
Minimally invasive
The IAB is a volume displacement device
Decreases LV afterload
Increases coronary perfusion
95
Impella
Mechanical Cirulatory
Support Device
“Percutaneous VAD”
Minimally invasive
Unloads ventricle reducing
myocardial workload
Produces 2.5 liters of
cardiac output
Recommended for up to 7
days
Bridge to Recovery 96
ECMO
Extracorporeal Membrane Oxygenation
Used to treat medically refractory cardiogenic
shock with poor oxygenation
Provides biventricular support
Not good for long term durability
Used in a short term situation
Requires perfusion support
Bridge to Recovery 97
Left Ventricular Assist Devices
Profound failure Mean blood pressure < 60 mmHg,
Systolic blood pressure < 90 mmHg, Cardiac index < 2.0 L/min/m2
Temporary replacement of pumping function of the left ventricle
Blood diverted from LA and LV to the LVAD
Blood returned to the aorta
Continuous flow vs. pulsatile flow
Portability
98
Courtesy of Thoratec Heart Mate II
Left Ventricular Assist Devices
LVADs
Bridge to transplant
“Destination Therapy”
Can improve quality of life
Prolonged LVAD support may result in enough
recovery of myocardial function to allow for
life after explantation
Occurs more frequently in patients with acute CHF
and no CAD
99
100
Final Quote:
Our grand business in life
is not to see what lies
dimly at a distance,
but to do what lies clearly at hand.
Thomas Carlyle (1795-1881)
BE THE BEST THAT YOU CAN BE
EVERY DAY. YOUR PATIENTS ARE
COUNTING ON IT!
101