Gas Analysis NDIR gas analyser in single or dual channel version BA 6000 - IR DE550008 04/2019 page 1 / 4 Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen Tel. +49 (0) 21 02 / 49 89-0, Fax: +49 (0) 21 02 / 49 89-20 E-Mail: [email protected] Internet: www.buehler-technologies.com The BA 6000 - IR gas analysers us the NDIR dual stream push- pull principle and highly selectively measure gases with absorption bands in the infrared wave range of 2 to 9 µm, e.g. CO, CO 2 , NO, SO 2 , NH 3 , H 2 O, CH 4 and other hydrocarbons. Single channel units measure one gas component. Dual channel units measure two separate gas components simultaneously. The units are designed for use in non-explosive areas. Four measuring ranges per channel, programmable, all measuring ranges linear (including with suppressed zero point) Electrically isolated measurement output 0/2/4 to 20 mA per channel Automatic or manual measuring range switchover option Menu-guided control for ease of use Fast response time Low long-term drift Internal pressure sensor for correcting fluctuations in barometric pressure Programmable automatic measuring range adjustment Operation based on NAMIR Sample gas monitor (flow rate and pressure) RS 485 serial port per channel

Upload: others

Post on 06-May-2022

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: DE550008 BA 6000 IR - buehler-technologies.com

Gas Analysis

NDIR gas analyser in single or dual channel version BA 6000 - IR

DE550008

04/2019

page 1 / 4

Bühler Technologies GmbH, Harkortstr. 29, D-40880 RatingenTel. +49 (0) 21 02 / 49 89-0, Fax: +49 (0) 21 02 / 49 89-20

E-Mail: [email protected]: www.buehler-technologies.com

The BA 6000 - IR gas analysers us the NDIR dual stream push-pull principle and highly selectively measure gases withabsorption bands in the infrared wave range of 2 to 9 µm, e.g.CO, CO2, NO, SO2, NH3, H2O, CH4 and other hydrocarbons.

Single channel units measure one gas component. Dualchannel units measure two separate gas componentssimultaneously.

The units are designed for use in non-explosive areas.

Four measuring ranges per channel, programmable, allmeasuring ranges linear (including with suppressed zeropoint)

Electrically isolated measurement output 0/2/4 to 20 mA perchannel

Automatic or manual measuring range switchover option

Menu-guided control for ease of use

Fast response time

Low long-term drift

Internal pressure sensor for correcting fluctuations inbarometric pressure

Programmable automatic measuring range adjustment

Operation based on NAMIR

Sample gas monitor (flow rate and pressure)

RS 485 serial port per channel

Page 2: DE550008 BA 6000 IR - buehler-technologies.com

BA 6000 - IR

Design and Features

Housing

– 19" rack, 4 HE for installation in swing frames

– 19" rack, 4 HE for installation in cabinets, with or without telescopic rails

– Front panel can be swivelled down (laptop connection)

– Internal pressure sensor for correcting fluctuations in barometric pressure

– Internal gas paths: FPM (Viton) tube or titanium pipe

– Measuring chamber (with or without compensation path under flow) made from stainless steel or tantalum for highly cor-rosive sample gases. (such as HCL, CL2, SO2, SO3, etc.)

– Gas connections for sample gas inlet and outlet as well as (possible reference gas): Pipe diameter 6 mm or 1/4”

Display and control panel

– large LCD panel simultaneously displays:

– Measurement (digital and analog display)

– Status line

– Measuring ranges

– LCD panel contract adjustable in menu

– permanent LED backlight

– wipe-clean membrane keyboard with five soft keys

– menu-guided operation for parametrisation, test function and adjustment

– Plain text help

– graphic display of concentration gradient; programmable time intervals

– Control software in two languages: German/English, English/Spanish, French/English, Spanish/English, Italian/English

Inputs/outputs

– One analog output per measurement value

– two analog outputs configurable to e.g. cross-correction

– six programmable binary inputs for e.g. measuring range changeover

– six programmable relay outputs for e.g. failure, maintenance requests, maintenance switch, limit alarm, external solenoidvalves

– Optional expansion, eight each additional binary inputs and relay outputs for automatic adjustment with max. four testgases

Port

– RS 485 serial per channel

2 Bühler Technologies GmbH We reserve the right to amend specification. DE550008 ◦ 04/2019

Page 3: DE550008 BA 6000 IR - buehler-technologies.com

BA 6000 - IR

Function

The BA 6000 - IR gas analyser uses the infrared push-pull alternating light principlewith double layer detector and optical coupler.

The measuring principle is based on infrared light absorption. The absorbedwavelengths are characteristic for individual gases. However, with a gas mixture,some spectra may overlap. On the BA 6000 - IR these cross-sensitivities are minim-ised by:

• the gas-filled beam splitter (6)

• the double-layer detector with optical coupler (11-13)

• optical filters based on analysis purpose (5)

A transducer (4) heated to about 700 °C which can then be shifted to balance thesystem generates two equal beams (measurement and reference beam) inside thebeam splitter (6). The beam splitter also serves as filter cuvette. Whilst the referencebeam impinges upon the right side of the detector chamber (11) through a referencecuvette (10) filled with N2 (not infrared-active gas) practically unattenuated, themeasurement beam flows through the measurement cuvette (9) with sample gasflowing through it and, weakened more or less depending on the concentration ofthe sample gas, impinges on the left side of the detector chamber (13). The detectorchamber is filled with a defined concentration of the gas component to be meas-ured.

The detector has a double-layer design. The top detector layer preferably absorbsthe middle of the absorption bands, whilst the sides of the bands is absorbed aboutequally in the bottom and top layer. The top and bottom detector layer are pneu-matically connected via the microflow sensor (3). This degeneration results a verynarrow-band spectral sensitivity.

The optical coupler (12) optically extends the bottom detector chamber layer. Chan-ging the slider position (14) generates a pulsing stream between the detector cham-bers, which the microflow sensor (3) converts into an electric signal.

The microflow sensor consists of two nickel grids which are heated to 120 °C, whichtogether with two additional legs forms a Wheatstone bridge. The pulsing flowcombined with a very dense grid arrangement changes the resistance of the Nigrids. This causes bridge detuning which varies by the concentration of the samplegas.

1 234567891011121314

Sample gas inlet Sample gas outletMicroflow sensorTransducer, adjustableOptical filterBeam splitter (gas filter)Diaphragm wheelEddy current driveMeasuring cuvetteReference cuvetteDetector chamber, rightOptical couplerDetector chamber, leftSlider, adjustable

Technical Data

BA 6000 IR Technical DataMeasuring ranges 4 per channel, auto changeover measuring ranges also possibleSmallest possible measuring range varies by application

e.g.: CO: 0 to 10 vpm, CO2: 0 to 5 vpmLargest possible varies by applicationCharacteristic linearisedLinearity deviation < 0.5% of spanControl panel LCD with LED backlight and adjustable contrast, soft keys, number keypad and function

keysMeasurement display 5 character, resolution varies by the measuring range selected, decimal places selectionEMCImmunity as required by standard NAMUR NE21 (05/93); CE mark EN 50081-1, EN 50082-2 and EN

61010 (electromagnetic compatibility)Position of use Front panel verticalDimensions (H x W x D) 19", 4 HE = 177 x 483 x 476 mmWeight approx. 15 kg for single channel unit

approx. 21 kg for dual channel unitAuxiliary energyElectric supply AC 100 to 120 V, 48 to 63 Hz (rated range 90 V to 132 V)

AC 200 to 240 V, 48 to 63 Hz (rated range 180 V to 264 V)

3Bühler Technologies GmbHWe reserve the right to amend specification.DE550008 ◦ 04/2019

Page 4: DE550008 BA 6000 IR - buehler-technologies.com

BA 6000 - IR

BA 6000 IR Technical DataPower input approx. 40 VA for single channel unit

approx. 75 VA for dual channel unitGas inlet conditionspressure 0.5 to 1.5 bar absolutePressure flow 20 to 90 L/h (0.3 to 1 L/min)Temperature 0 to 50 °CHumidity < 90% RH1) or varies by measuring taskTime behaviourHeat-up time at room temperature: < 30 minlag (T90 time) varies by dead time and programmable dampingDamping 0 to 100 s, programmable (electric time constant)Time lag approx. 0.5 to 5 s depending on version (washout time of the gas path inside the device

at 1 L/min.)Time for internal signal processing < 1 sPressure correction rangePressure sensor (internal or external) 700 to 1200 mbarMeasuring performance2)

Static + 1 % of the smallest possible measuring range per type plateZero drift < ± 1 % from measuring range / weekMeasurement value drift < ± 1 % from measuring range / weekReproducibility between 0.1 % and 1 % from the respective measuring range depending on versionFactors3)

Ambient temperature < 1% from measuring range / 10KSample gas pressure with pressure compensation off: < 0.15 % from target value 1 % air pressure changeSample gas flow rate negligibleAuxiliary energy < 0.1% from output signal range at a rated voltage of ± 10%Electric inputs and outputsAnalog output 0 / 2 / 4 to 20 mA, potential-free, load ≤ 750 ꭥRelay outputs 6 with change-over contacts, programmable e.g. to measuring range identification; rat-

ing: AC/DC 24 V / 1 A, potential-free, non-sparkingAnalog inputs 2, configured for 0 / 2 / 4 to 20 mA for external pressure sensor and carrier gas influence

(cross-gas correction)Binary inputs 6, configured for 24 V, potential-free, freely parametrisable e.g. to measuring range

changeoverPort RS 485Options Add-on electronics with 8 each additional binary inputs and relay outputs, e.g. to trigger

automatic adjustmentPermissible climatic conditionsAmbient temperature - 30 to + 70 °C during storage and transport, + 5 to + 45 °C during operationHumidity < 90% RH1) annual average during storage and transport4)

IP rating IP 20 (EN 60529)1) RH: relative humidity2) highest accuracy is reached after 2 hours3) based on sample gas pressure 1 bar absolute, 0.5 L/min sample gas flow and 25 °C ambient temperature

4 Bühler Technologies GmbH We reserve the right to amend specification. DE550008 ◦ 04/2019