dacell loadcell controller
TRANSCRIPT
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 1/32
Ver 5.0
MODEL : DN-10W
DIGITAL INDICATOR
USER MANUAL
DACELL CO.,LTD.
Address : 681-1 Cheoksan-Ri, Nami-Myeon, Cheongweon-Gun,
Chung-buk
TEL :82-43-260-2242
FAX : 82-43-260-2245
http://www.dacell.com
E-Mail : [email protected]
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 2/32
Contents
1. Features ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 2
1-1. High-speed conversion method ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 2
1-2. Calibration method ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 2
1-3. HOLD & PEAK HOLD ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 2
1-4. Comparison Output ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 2
1-5. Data Back-up ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 2
1-6. Watch dog ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 3
1-7. Option ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ 3
1-8. Power ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ 3
2. Attention ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 4
2-1. Attention for installation ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 4
2-2. Attention for use ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 4
3. Specification ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 5
4. Front Panel ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 6
5. Rear Panel ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 7
6. Wiring Diagram ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 9
7. Components & Function ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 10
7-1. Flow Chart ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 10
7-2. How to use Hold mode ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 11
7-3. How to use comparison output function ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 12
8. Setting mode ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 14
8-1. Types of Setting mode & Set-up ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 14
8-2. Function mode (Function Set-up) ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 15
8-3. Digital calibration (Calibration by sensor output value) ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 22
8-4 . Rea l-we igh t Ca l i b ra t i on ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ 23
8-5. SPAN constant calibration⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅ ⋅
248-6. Lock Set-up ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 25
9. Product Inspection ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 26
10. OPTION (OP-02, OP-03)
Option 02 (RS232C) ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 27
Option 03 (RS485) ⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅⋅ 29
1
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 3/32
1. Features
We thank you for using our product. Please refer to this manual or contact our office if
you find any problems during using our product.
This product is an indicator that emplifies the micro voltage signals from each sensor
and display in digital. It is mainly used to measure physical quantities of the items using
strain guage such as loadcell, pressure sensor, displacement sensor and torque sensor.
This product has the following special features.
1-1. Medium & High-speed Conversion Method
16bit High-speed A/D converter that can detect the sensor’s input signal at a rate of100 times per second is used.
For analog output, 12bit D/A converter with 100 times per second is used so that there
is almost no deviate between display value and output value.
1-2. Calibration Method
2 types of calibration systems are adopted. Calibration by actual load (standard
weight) and calibration by rated output of sensor (mV/V).
1-3. Hold & Peak Hold
You can choose Analog peak hold (High-speed) or Digital preak hold (Low-speed) as
per your need.
1-4. Comparison Output
You can do set-up the maxium and lower limit value by the keys on the front panel
and print with contact point signal of the rear panel.
1-5. Data Back-up
All the set-up values will be memorized on the flash memory so the inputted data can
be saved and no need to do the re-setting even in case of the interruption of the
electric power or power disconnection.
2
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 4/32
1-6. Watch dog
This function is for automatic reset in case the system is stopped due to the external
factors such as noise.
1-7. Option
You can use RS232C, RS485 as the optional specification.
1-8. Power
AC 85~265V 50/60Hz free voltage
3
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 5/32
2. Attention
For the efficient and safe use, please carefully read and be fully aware of the following
details before using this product.
It is strictly forbidden to use this product for any other purpose of use or to attempt to
make any alteration on this product.
2-1. Attention for installation
• Please keep it out of wet places.
• Do not set it up near vibration & impulse, high temperature and humidity.
Keep it out of the direct rays of the sun. Set it up where there is less dust, and Keep itout of direct air including salt and ion.
• Do not use when there is inflammable gas or heavy machinery, and smog.
• Ground earth-terminal ( ).
• Make wire separately from power system wiring and noise wiring.
• Make sure the use of 4 line sealed cable as a sensor cable.
Too long cable leads to measurement error due to wiring resistance (around
10meters).
2-2. Attention for use
During calibration, Do not input free-load state and real-weight load until it becomes
stable. Pressing Enter Key in unstable condition leads to calibration error.
Do not press any Key in use at one's discretion. Please refer to 7. Setting-up mode
for the function and method of Key.
4
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 6/32
3. Specification
• Available sensor : Strain Gauge sensor (Bridge 120Ω, 350Ω, 700Ω)
• Upper limit Display : -19999 ~ +99999
• A/D Converter : 16bit, 100 times/sec
• D/A Converter : 12bit
• Temperature Characteristic (Amp Characteristic)
Zero : 0.5μV/
Span : 50ppm/
• Front Panel
Measured value display: 7 segment 5 Digit, Character Height 14mm
Status display : Red LED 3 pcsKey Switch : 4 pcs
• Output
Comparison output : Upper limit(HI), Lower limit(LO), Normal(OK) Output
Contact Capacity - AC 250V 0.5A/ DC 50V 1A
Analog Output : DC 0~10V/4~20mA (Optional at the time of order)
• Temperature Range : -10∼60, less than 80% RH (no dewing.)
• External dimension : 96 × 48 × 128 mm (W × H × D)
•
Panel cutting size : 91.5 × 44.5 mm• Weight : About 800g
• Power Supply : AC 85~265V 50/60Hz free voltage
• Option : RS232C (OP-02), RS485 (OP-03)
5
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 7/32
4. Front Panel
① HI, LO Indication LED
: H
: I
20
i or LO LED will be lighted when measured value exceeds the setting value.
② HOLD indication LED
: This LED will be lighted when the measured value is on Hold.
③ Measured value Indication
: It indicates the measured value and each setting value.
④ ZERO/SPAN setting up VR
t is used when ZERO and SPAN calibration of Analog output (DC 0 ~ 10V or 4 ~
mA) is carried out.
⑤
m
SE
re
Measruing Mode : Once this Key is pressed for more than 1 second, the current
easuring value will be Zero (0) and the Analog output will be 0V (4mA) as well.
T-UP Mode : Once this is pressed on the Function Set-up Mode, you will
turn to the measuring mode.
⑥
Measuring Mode : Once this Key is pressed, the lower limit setting value will be
displayed and this value also can be changed.
SET-UP Mode : The location of row for the number flickering can be moved.
⑦
Measuring Mode : Once this Key is pressed, the upper limit setting value will be
displayed and the value can be changed.
SET-UP Mode : The flickering number will be increased by 1 and 1.
⑧
Me
key
asuring Mode : Once this Key is pressed, HOLD will be selected and once this
is pressed again, HOLD will be cancelled.
6
SET-UP Mode : Save each set-up value.
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 8/32
5. Rear Panel
Please check the location of terminal and its use. While the projected button is being
pressed, please insert the cable into the lower hole completely. As soon as you release
the button, the connection will be completed. At the point, please slightly pull the calbe
and check whether the cable is come off or not.
(The most suitable calbe is Φ 0.5~1. Linking cable must be brazing or used with I
terminal)
① AC IN : Main Power Supply Wiring Terminal
② : Ground Terminal (as an independent ground connection.)
③ OUT : Analog (DC 0 ~10V/DC 4 ~ 20mA) Output Terminal
④ EXC+ : Sensor Supply Voltage + Connection Terminal
⑤ EXC- : Sensor Supply Voltage + Connection Terminal
⑥ SIG+ : Sensor Ouput Signal + Connection Terminal
⑦ SIG- : Sensor Output Signal – Connection Terminal
⑧ GND : SHIELD connection terminal of sensor
⑨ GND, TXD, RXD : RS232C SERIAL INTERFACE (RS485 : TXD → TX+, RXD → TX-)
⑩ GND : External Input Common Terminal
⑪ IN1 : External HOLD Input Terminal
⑪ IN1 : External HOLD Input Terminal
⑫ IN2 : External ZERO Input Terminal
⑬ IN3 : External printer signal Input Terminal (when PT-100 use)
⑭ IN4 : No use
7
⑮ COM : RELAY Output Common Terminal
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 9/32
⑯ RY1 : RELAY 1(Lower Limit) Output Terminal
17 RY2 : RELAY2 (Upper Limit) Output Terminal
18 RY3 : RELAY3( Normal) Output Terminal (OK )
19 RY4 : RELAY 4 Output Terminal
20 ANALOG OUT LOW PASS FILTER Selet Switch
SW1 : 10Hz
SW2 : 100Hz
SW3 : 1kHz
Attention
1. When wiring, please pull out the power plug. (Main power supply)
2. Please use the thick calbe for ground terminal ( ) so that you can avoid any troubledue to the impulse voltage or surge. If possible, please keep it as an independent
ground.
(Th
de
is must be used in the area with heavy noise. If the ground is made with other
vice, it will be affected by noise.)
3. Please check the function of the terminal before you do the wiring so that you can
prevent any malfunction in advance.
4. Manufacture will be relieved of its responsibility for any damge or injury due to the
disassembly and alteration made without any agreement. Also, no A/S will be made.
8
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 10/32
6. Wiring Diagram
<Wiring Diagram>
9
<Example - External Input PLC Connection>
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 11/32
7. Components & Function
7-1. Flow Chart
10
Sensor input
A/D conversion
100 times/s
Peak & Sample hold
circuit
Fixed filter
Average : 3 times
Output count : 32 times/s
Display filter
Display update rate : 1~32times/s
Average : 1 ~ 32times/s
Display D/A conversion
1 ~32 times/s
Comparator
OP-02(RS232C)
OP-03(RS485)Display
Analog output
Relay output
Serial output
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 12/32
7-2. How to use Hold mode
Hold mode is largely divided into Peak Hold and Sample Hold. Please select Analog,
Digital or Display Hold as per your purpose of use (Hi or Low Speed)
To input Hold, you can use Hold key on the front panel or external input. For the
operation method, please refer to the drawing below.
1) Peak Hold : To Hold the maxium value among the measured values.
pe
va
Ab
ma
ak hold mode : It is a low-sped Peak hold. It is to hold and display the maximum
lue of (+) direction.
solute peak hold mode : It is a low-speed Peak hold. It is to hold and display the
ximum value of the absolute value (+/-).
2) Sample hold : It is to hold and display the value at the time of Hold signal inputamong the measured values.
11
On
Off
On
Off
off
On
Front panel HOLD LED
External HOLD input termanal
Front panel HOLD Key
<Peak HOLD> <Sample HOLD>
②
①
time
②
①
time
M e a s u r e d v a l u e
① − : sensor input
value
② − : Display value &
analog output value
M e a s u r e d v a l u e
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 13/32
7-3. How to use comparison ouput function
For comparison output function, there are 3 different modes such as Decision, High
limit, Low limit. It displays through the relay of rear panel comparing each setup value.
On High limit and Low limit mode, Hysteresis can be used.
To set up upper limit (High) and lower limit(Low), please use the key on the front panel.
1) Decision mode : Measured Value ≤ Lower limit setup value ⇒ RY1 ON (LO)
Measured Value ≥ Upper limit setup value ⇒ RY2 ON (HI)
Lower limit setup value < Measured value < Upper limit setup value ⇒ RY3 ON(OK)
2) High limit mode : Measured Value ≥ RY1 setup value ⇒ RY1 ON
Measured Value ≥ RY2 setup value ⇒ RY2 ON
Measured Value ≥ RY3 setup value ⇒ RY3 ONMeasured Value ≥ RY4 setup value ⇒ RY4 ON
Measured Value < RY1 setup value – Hysteresis value⇒ RY1 OFF
Measured Value < RY2 setup value - Hysteresis value⇒ RY2 OFF
Measured Value < RY3 setup value – Hysteresis value⇒ RY3 OFF
Measured Value < RY4 setup value - Hysteresis value⇒ RY4 OFF
3) Low limit mode : Measured Value ≤ RY1 setup value ⇒ RY1 ON
Measured Value ≤ RY2 setup value ⇒ RY2 ON
Measured Value ≤ RY3 setup value ⇒ RY3 ONMeasured Value ≤ RY4 setup value ⇒ RY4 ON
Measured Value > RY1 setup value + Hysteresis value⇒ RY1 OFF
Measured Value > RY2 setup value + Hysteresis value⇒ RY2 OFF
Measured Value > RY3 setup value + Hysteresis value⇒ RY3 OFF
Measured Value > RY4 setup value + Hysteresis value⇒ RY4 OFF
4) Low & High limit mode
Measured Value ≤ RY1 setup value ⇒ RY1 ON
Measured Value ≤ RY2 setup value ⇒ RY2 ON
Measured Value ≥ RY3 setup value ⇒ RY3 ON
Measured Value ≥ RY4 setup value ⇒ RY4 ON
Measured Value > RY1 setup value + Hysteresis value ⇒ RY1 OFF
Measured Value > RY2 setup value + Hysteresis value ⇒ RY2 OFF
Measured Value < RY3 setup value - Hysteresis value ⇒ RY3 OFF
12
Measured Value < RY4 setup value - Hysteresis value ⇒ RY4 OFF
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 14/32
13
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 15/32
8. Setting Modes
8-1. Types of Setting mode & Set-up
For Setting mode, there are 4 different types of mode such as Function mode, Digital
calibration mode, Actual load calibration mode and SPAN constant calibration mode.
14
Mesuring mode
While key is being
Press, press key
Function mode
END
Digital calibration
mode
END
END
Actual load
calibration mode
① Function modeAcess to each function setup mode. Please refer 8-2 function mode.
② Digital calibration mode
It is to calibrate into the sensor’s output value. No need to prepare for the actual
load (stand weight). Please refer 8-3 how to calibrate.
③ Actual load calibration mode
It is to calibrate by adding the actual load (standard weight or the load you know).
Please refer 8-4 how to calibrate.
④ SPAN constant calibration mode
It is to calibrate with the S.CAL value written down for load calibration. Please refer
8-5 how to calibrate.
END
SPAN constant
calibration mode
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 16/32
8-2. Function mode
1) How to set function.
15
Measuring mode
While key is being
Press, press key
END
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 17/32
2) How to set RELAY data
① Decision mode (Decision mode: mode 0)
② Limit mode (Low & High limit mode: mode 1 ~ 3)
16
Measuring mode
Press key
END
Input
value
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 18/32
Function mode list
Name Function Setting
The standard
set-up value
F-01 Decimal point 0, 1, 2, 3 1
F-02 Division 1, 2, 5, 10, 20, 50 1
F-03 Display filter 0, 4, 8, 16, 32 16
F-04 Hold mode Sample hold, Peak hold, Absolute peak hold Peak hold
F-05Comparison mode Decision, High limit, Low limit,
Low&High limitDecision
F-06 Hysteresis 0 ~ 99 0
F-07 DAC zero -19999 ~ +99999 0
F-08 DAC capacity -19999 ~ +99999 10000
F-09 ID Number 0 ~ 32 0
F-10 Baud rate & PRINT 2400, 4800, 9600, 19200, print 9600
F-11 Auto zero tracking 0 ~ 99 0
F-12 Auto zero tracking time 0.0 ~ 5.0 sec 0.0
F-13 Force unit Kg(kg/), N, ℓb, bar, MPa Kg
F-14 Key disabling Zero key, Lo key, Hi key, Hold key 0000
17
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 19/32
F-01. Decimal point (Decimal point Set-up)
(Standard setup value: 1)
Display data Setting
0 00000 : No decimal point
1 0000.0 : One decimal place
2 000.00 : Two decimal places
3 00.000 : Three decimal places
F-02. Division (Minimum display unit setup)
(Standard setup value: 1)
Display data Setting
1 Displayed In 1 (0, 1, 2, 3, 4 …….)
2 Displayed in 2 (0, 2, 4, 6, 8 …….)
5 Displayed in 5 (0, 5, 10, 15 …….)
10 Displayed in 10 (0, 10, 20, 30 …….)
20 Displayed in 20 (0, 20, 40, 60 …….)
50 Displayed in 50 (0, 50, 100, 150 …….)
F-03. Display filter (Display speed setup)
(Standard setup value: 16)
Display data Setting
0 No filter
4 Average time 1/8 sec
8 Average time 1/4 sec
16 Average time 1/2 sec
32 Average time 1 sec
F-04. Hold mode
(Standard setup value: 1)Display data Setting
0 Sample Hold: To hold the display value at the time of Hold
signal input.
1 Peak Hold (+): To hold the maximum value of display values
during Hold signal input.
2 Absolute Peak Hold(+/-): To hold the maximum absolute value of display
values during Hold signal input.
18
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 20/32
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 21/32
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 22/32
Note) During calibration, set up the unit you want after calibrating it into Kg.
F-14. Key disabling (Front key locking set up)
(Standard setup value: 0000)
Display data Setting
0 0 0 0 Hold key Lock (1), Release (0)
Hi key Lock (1), Release (0)
Lo key Lock (1), Release (0)
Zero key Lock (1), Release (0)
<Set-up example>
1001 : Zero & Hold key Lock, Hi and Lo key Release
21
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 23/32
8-3. Digital calibration (Calibration by sensor output value)
At the time of purchasing sensor, the rated capacity (R.C) and rated output (R.O)
declared on the calibration sheet can be used for the calibration for easier calibration.
and
key. And then. Press key
22
Measuring mode
While key is being
Press, press key
Display Function mode
Display Sensor output value calibration mode
Display R.O. value input mode
Display R.C. value input mode
Input the rated output value by using and
key. And then. Press key
Calibration end
Input the rated output value by using
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 24/32
8-4. Actual load calibration
23
This is a calibration method by adding actual load on the sensor. Standard weight is needed.
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 25/32
Display Shunt CAL value (please write to S.CAL)
Cal. Date R
24
Measuring mode
While key is being Press, press key
Display Function mode
Display Sensor output value calibration mode
Display Actual load calibration mode
Display actual load mode
Input the rated output value by using and key. And
then. Press key
Calibration end
Display zero mode. press key without any load on the
sensor
About 1 sec later
After adding the actual load (standard weight) on the
sensor, press key.
About 1 sec later
Display R.O. value input mode
Input the rated output value by using and key. And then.
Press key (Input 3.0000 if do not know the R.O. value)
.O. value calibration constant vlaue
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 26/32
8-5. Calibratuib by Shunt CAL Value
It is to calibrate with the S.CAL value written down for load calibration. You can
calibrate without any standard weight.
25
Measuring mode
While key is being
Press, press key
Display Function mode
Display Sensor output value calibration
Display Actual load calibration indicate
Input the rated output value by using and key. And
then. Press key
Calibration end
Display Shunt cal value calibration mode
Display R.O. value input mode
Display Shunt cal value input mode
Input Shunt cal value by using and key. And then.
Press key
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 27/32
26
8-6. Lock Set-up
You can prevent any accidental operation due to the unnecessary key control by Lock
set-up. After finishing calibration, it is recommended to set the Lock.
At the first stage, please start while the power is OFF.
Related Function when Lock is set : Function related to calibration
Power off
While key is being Press, and
then the power is ON
Please keep pressing key while this is being displayed
or
Set-up end
Pllease select lock or release by pessing key
ON : Setting is impossible. Off :Setting is possible
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 28/32
9. Product Inspection
Symptom Cause Action Remark
When Display
trembles.
Load cell is damaged.
Load cell insulation
resistance.
Indirect occurrence
Load cell input, output.
Check resistance
Check load cell’s
insulation resistance.
Insulation
resistance
(Cable & Case >
1000 Mohm)
Loadcell faultyCheck load cell’s
insulation resistance.
Loadcell connection is
insufficient.
Check the wiring
between load cell and
the main device.Check the load cell’s
calbe’s disconnection.
When weight
changes into (-).
Loadcell wiring is
reversed.
Check load cell’s ouput
cable connection.
Output : (+SIG)
(-SIG)
Displayed as
“OVER” or
“UNDER”
Load cell is damaged.
Load cell connection is
bad.
Check the load cell’s
condition and calbe
connection.
27
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 29/32
10. OPTION
Option-02 (RS232C)
Since RS232C Interface is very sensitive of electric noise. So please do the wiring
from AC Power and electric wires separately. Also you must use the shield calbe always.
Indicator Host PC
TX(Transmission Data) RXD(Receive Datea), No.2 Pin
RX(Receive Datea) TXD(Transmission Data), No.3 Pin
GN(Ground) GND(Ground), 5 Pin
1. TYPE : EIA-232C
2. Method : Half-duplex, asynchronous method.
3. Baud-rate : Select one of 2400, 4800, 9600, 19200bps
4. Parity : No Parity
5. Data bit : 8 bit
6. Stop bit : 1bit
7. Stream mode (Ex. Data +1234.5 transmission)
CODE BYTE1 BYTE2 BYTE3 BYTE4 BYTE5 BYTE6 BYTE7 BYTE8
ASCII S T , N T , + 0
HEX 53H 54H 2CH 4EH 54H 2CH 2BH 30H
CODE BYTE9 BYTE10 BYTE11 BYTE12 BYTE13 BYTE14 BYTE15 BYTE16
ASCII 1 2 3 4 . 5 CR LF
HEX 31H 32H 33H 34H 2EH 35H 0DH 0AH
1) BYTE
. DATA S DAT U S
. DATA DAT L
2) BYTE
3) BYTE
1, BYTE2
table : S T . A Unstable :
OVERFLOW: O L . A UNDERFLOW : U
3 ~ BYTE6 : fixed character (, N T ,)
28
7 ~ BYTE14 : DATA 8 BYTE(including +/-)
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 30/32
4) BYTE CARR
5) BYTE LINE
OP
15 : IAGE RETURN
16 : FEED
8. Command mode
29
-03 : Refer to RS485
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 31/32
Option-03 (RS485)
S
A
ince RS485 Interface is very sensitive of electric noise. So please do the wiring from
C Power and electric wires separately. Also you must use the shield calbe always.
1. TYPE : RS485
2. Method : Half-duplex, asynchronous method.
3. Baud-rate : Select one of 2400, 4800, 9600bps
4. Parity : No Parity
5. Data bit : 8 bit
6. Stop bit : 1bit
Please set up the device No. referring to INDICATOR Manual.
(Can setup from 1 to 32 channel.)
7. Command form (PC -> INDICATOR)
CODE BYTE1 BYTE2 BYTE3 BYTE4 BYTE5
ASCII I D 0 1 P
HEX 49H 44H 30H 31H 50H
1) BYTE
2) BYTE umb
3) BYTE
1, BYTE2 : Fixed character (ID)
3, BYTE4 : device n er (1 ~ 32)
5 : command order (P, H, R, Z)
8. Command chart
Command
ASCII HEX
P 50H Transmit the current value of order equipment.
H 48H Hold for order equipment.
R 52H Release hold for order equipment.
Z 5AH Operate the current value of order equipment as ZERO.
30
7/24/2019 Dacell Loadcell Controller
http://slidepdf.com/reader/full/dacell-loadcell-controller 32/32
9. Transmission DATA form (INDICATOR -> PC)
CODE BYTE1 BYTE2 BYTE3 BYTE4 BYTE5 BYTE6 BYTE7 BYTE8
ASCII I D 0 0 1 , + 0
HEX 53H 54H 30H 30H 31H 2CH 2BH 30H
CODE BYTE9 BYTE10 BYTE11 BYTE12 BYTE13 BYTE14 BYTE15 BYTE16
ASCII 1 2 3 4 . 5 CR LF
HEX 31H 32H 33H 34H 2EH 35H 0DH 0AH
1) BYTE
2) BYTE
3) BYTE4) BYTE
5) BYTE CARR
6) BYTE LINE
1, BYTE2 : Fixed character (ID)
3 ~ BYTE5 : Device number (1 ~ 32)
6 : Fixed character (,)7~BYTE14 : DATA 8byte (including +/-)
15 : IAGE RETURN
16 : FEED