d104096 rev m surgical training manual pdf

Upload: marihesa138

Post on 14-Apr-2018

237 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    1/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 1 of 69Otologics Confidential

    Otologics LLCIn United States currently in Clinical Trials

    CAUTION: Investigational Device. Limited by Federal Law (USA) to Investigational Use.

    In Europe CE Marked

    SURGICAL TRAINING MANUAL

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    2/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 2 of 69Otologics Confidential

    Contact Information

    Worldwide Headquarters

    Otologics LLC

    5445 Airport BlvdBoulder, CO 80301US Tel: +1 (303) 448-9933

    US Fax: +1 (303) 448-9955

    EU Tel: +33 4 92 92 52 00

    Website: www.otologics.com

    E-Mail: [email protected]

    E-Mail: [email protected]

    European Representative

    AR-MED Ltd

    Runnymede Malthouse

    Egham TW20 9BD

    United Kingdom

    Product: METTM (Middle Ear Transducer) Fully Implantable Ossicular Stimulator

    Document: Surgical Training ManualDocument No: D104096

    Copyright 2007 Otologics LLCNo part of this manual may be reproduced without permission.

    All trademarks used in text are property of their respective owners.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    3/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 3 of 69Otologics Confidential

    Contents

    INTRODUCTION......................................................................................................................................... 4

    OTOLOGISTS TRAINING FOR THE FULLY-IMPLANTABLE MET OSSICULAR STIMULATOR..... 4

    MET OSSICULAR STIMULATOR DESCRIPTION.................................................................................... 5

    CANDIDACY CRITERIA............................................................................................................................. 8Technical Specifications.........................................................................................................................................9Implant....................................................................................................................................................................9Charger System.......................................................................................................................................................9Transducer Loading Assistant...............................................................................................................................10Remote..................................................................................................................................................................10Programming System............................................................................................................................................10

    NEW IMPLANTATION OF THE FULLY-IMPLANTABLE MET OSSICULAR STIMULATOR - FIMOS. 11Step 1: Pre-Surgical Implant Preparation.............................................................................................................12Step 2: Making the Incision .................................................................................................................................13Step 3: Drilling the Atticotomy............................................................................................................................14Step 4: Positioning the Mounting Bracket ...........................................................................................................15Step 5: Fastening the Mounting Bracket ..............................................................................................................19Step 6: Creating the Laser Hole ...........................................................................................................................20Step 7: Preparing the Bone Beds and Lead Channels ..........................................................................................22Step 8: Placing the MET Ossicular Stimulator ....................................................................................................23Step 9 Loading the MET Ossicular Stimulator and Diagnostics...........................................................................25Step 10: Attaching the Electronics Capsule to the Transducer .............................................................................27Step 11. Securing the Electronics Capsule and Closing the Incision...................................................................30Step 12. Post Surgical Care and Activation .........................................................................................................32

    UPGRADE IMPLANTATION FROM THE SEMI-IMPLANTABLE OSSICULAR STIMULATOR (SIMOS)TO THE FULLY-IMPLANTABLE MET OSSICULAR STIMULATOR (FIMOS)....................................... 33

    Step 1: Pre-Surgical Preparation for the SIMOS to FIMOS Implantation............................................................34Step 2: Making the Incision and Removal of the Semi-Implantable Electronics.................................................35Step 3: Connector Disassembly ...........................................................................................................................36Step 4: Attaching the Electronics Capsule to the Transducer ...............................................................................37Step 5: Enlarging the Bone Bed for the Electronics Capsule, Pendant Microphone, and Closing.......................40Step 6: Post Surgical Care and Activation ...........................................................................................................41

    SURGICAL STEPS FOR REPLACING THE FULLY-IMPLANTABLE MET OSSICULAR STIMULATOR(FIMOS) WITH THE FULLY-IMPLANTABLE MET OSSICULAR STIMULATOR (FIMOS).................... 42

    Step 1: Pre-Surgical Preparation for FIMOS to FIMOS Surgery .........................................................................43Step 2: Making the Incision and Removal of the FIMOS Electronics Capsule ...................................................44Step 3: Connector Disassembly ...........................................................................................................................45Step 4: Attaching the New FIMOS Electronics Capsule to the Old Transducer...................................................46

    Step 5: Securing the Electronics Capsule, Pendant Microphone, and Closing ....................................................49Step 6: Post Surgical Care and Activation ...........................................................................................................50

    APPENDICES........................................................................................................................................... 52Appendix 1: Otologics Surgical Equipment ........................................................................................................52Appendix 2: Troubleshooting an Implanted Device During Surgery ..................................................................57Appendix 3: Surgical Training Checklist.............................................................................................................58Appendix 4: Surgical Revision Training Checklist .............................................................................................64Appendix 5: Surgical Procedure Step By Step ....................................................................................................66

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    4/69

    ____________________________________________________________________________________MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 4 of 69Otologics Confidential

    INTRODUCTION

    This manual describes the surgical procedure for the

    implantation of the MET Fully-Implantable

    Ossicular Stimulator, both for a new implantationand for an upgrade implantation from the MET

    Semi-Implantable Ossicular Stimulator. Before

    implanting the device for the first time, the

    implanting surgeon and clinical staff responsible forthe implantation must participate in a surgical course

    provided by Otologics, LLC and obtain certification.

    Certification training consists of a temporal bone

    course review of the Otologics device, surgical tools,and surgical procedure. An Otologics

    representatives presence is mandatory for at least

    the first two implantations. Surgical/Hospital stafftraining is also provided by Otologics. In addition to

    this manual, there is a surgical video available. It is

    strongly recommended that surgeons review this

    video before performing surgery.

    The surgical steps section of this manual gives keypoints and detailed steps to follow when implanting

    the device. Pictures and descriptions of the required

    surgical tools are included in each step. Safetyinstructions are also listed with each step. The

    Appendices contain detailed drawings of the surgical

    kits and instructions in case corrective action isrequired during the surgery.

    At the fitting, the patient also receives instructions

    on the use of the MET Fully-Implantable OssicularStimulator and its accessories, the Charger system

    and Remote Control.

    OTOLOGISTS TRAINING

    FOR THE FULLY-

    IMPLANTABLE MET

    OSSICULAR STIMULATOR

    The objective of the Otologists Training is toprovide an effective and consistent method for

    training otologists so they can perform the

    implantation of the MET Ossicular Stimulatorsuccessfully. Only otologists should be performing

    the implantation procedure and receiving the

    Otologists Training for the MET Ossicular

    Stimulator.

    In order to accomplish this objective, the following

    protocol has been created:

    1. Each implanting Otologic surgeon will gothrough an information protocol and hands on

    training including information on:

    a. the Otologics device

    b. the Otologics surgical instrumentsc. the Otologics implantation procedure

    2. An Otologics Representative will review theabove information through a temporal bone

    session or a half skull model demonstration,after which a surgical training checklist is

    completed and signed. This review session must

    be successfully completed before the firstimplantation surgery and returned to Otologics.

    3. Two tools have been created to assist in thetraining session:

    a. The Otologics Surgical TrainingManual (D104096 This manual)

    b. The Surgical Procedure TrainingChecklist and Signature Page(Appendix 3)

    NOTE: If more than six months passes without

    participation in a MET Ossicular Stimulator

    implantation surgery, the otologist will be

    required to go through another review session

    with an Otologics Representative to review the

    tools and procedure and to be updated on any

    new information. The review core will consist of,

    at a minimum:

    1. A review of the tools and procedure using ahalf skull model and the actual surgicaltools. An additional temporal bone session

    can be scheduled, if requested.

    2. The Surgical Procedure Checklist will befilled out in full and signed again.

    3. Any new information will be reviewed indetail

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    5/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 MSurgical Training Manual Page 5 of 69

    Otologics Confidential

    MET OSSICULAR STIMULATOR DESCRIPTION

    The MET Fully-Implantable Ossicular Stimulator is

    a fully implantable device intended for patients withmoderate to severe sensorineural hearing loss. The

    device is designed to eliminate the receiver found in

    conventional hearing aids and drive the ossicles

    directly. The potential benefits include eliminatingouter ear canal occlusion, improving sound quality,

    and providing greater gain without feedback for

    optimal speech recognition.

    DESCRIPTION:

    The Otologics MET Fully-Implantable Ossicular

    Stimulator consists of four primary components: theimplant, the programming system, the charger, and

    the remote control. The implant component of the

    MET Fully-Implantable Ossicular Stimulator

    consists of two primary parts: the electronics capsule

    and the Middle Ear Transducer (Figure 1). Theelectronics capsule contains the microphone, battery,

    magnet, digital signal processor and connector.

    Figure 1

    Sounds are picked up by a sensitive microphone,

    amplified according to the wearers needs, andconverted into an electrical signal. The signal is sent

    down the lead and into the transducer. The tip of the

    MET Fully-Implantable Ossicular Stimulator

    transducer is mounted in a laser-drilled hole in the

    body of the incus. Thetransducer translates theelectrical signals into a mechanical motion that

    directly stimulates the ossicles and enables thewearer to perceive sound.

    For upgrading from the MET Semi-ImplantableOssicular Stimulator to the MET Fully-Implantable

    Ossicular Stimulator (not approved in the US), the

    size of the bone bed is increased to accommodate thefully implantable electronics capsule, which is then

    attached to the transducer. The fully implantable

    electronics capsule is secured, and the wound isclosed. Two to six weeks after surgery and medical

    clearance the device can be programmed and

    activated.

    The Otologics Programming System (Figure 2)

    consists of fitting and diagnostic software, a radiofrequency coil that, when placed over the implant

    site, magnetically adheres to the side of the wearers

    head, and the NOAHlink programming interface,

    which is worn around the neck. Using OtoFitFitting Software, the NOAHlink interface receives

    signals from the computer through the wireless

    connection and sends the signals to the implant viathe radio frequency coil.

    Programming the implant is done in the same

    manner as programming traditional digital hearingaids. In addition, the Otologics Programming System

    provides the ability for extensive testing and

    diagnostics of the MET Fully-Implantable Ossicular

    Stimulator.

    Figure 2The charger system consists of the base station,

    charging coil, and charger body. To charge the

    implant, the wearer removes the charger body from

    the base station and places the coil on the skin, over

    the implant site (Figure 3). The charger bodycontains a clip that allows the charger to be attached

    to the belt or waistband of the wearer duringcharging (Figure 4). Typically, charging time will be

    about one hour if performed daily. While rechargingthe implant, the wearer can perform normal daily

    activities, turn the implant on and off, and adjust the

    volume.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    6/69

    ____________________________________________________________________________________MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 6 of 69Otologics Confidential

    Figure 3

    Figure 4

    A remote is used for controlling the MET Fully-

    Implantable Ossicular Stimulator when the device is

    not being charged. The remote allows the wearer toturn the implant on and off, and to adjust the volume.

    To use the remote control, the wearer simply holds

    the remote against the skin over the implant (Figure

    5).

    Figure 5

    SURGERY OVERVIEW:

    The transducer is implanted through an extended

    atticotomy approach that exposes the incus body andmalleus head and is intended to minimize thesurgical risk. This technique circumvents any need

    to dissect close to the facial and chorda tympani

    nerves, associated with the facial recess approach.

    The implant surgery is a relatively simple procedure

    and not difficult in practiced hands. It does require a

    set of proprietary surgical instruments. The

    instruments are detailed in Appendix 1 of thismanual. Also, each surgical step lists the

    instruments required for that step and has pictures

    where applicable.

    The basic steps of the surgery are as follows: non-

    sterile preparation, patient preparation and incision,

    atticotomy drilling, mounting bracket placement,laser hole creation, device placement and closing.

    The surgery, from opening incision to closing,

    typically can take up to 2-2.5 hours for surgeons

    familiar with the procedure.

    Upgrading a wearer who already has the semi-

    implantable device to a fully-implantable device is

    approved in the EU, This procedure is not approvedfor the US Clinical Trial. This is a simple surgical

    procedure which may be performed under local

    anesthesia. To upgrade, the semi-implantableelectronics capsule is disconnected from the

    transducer via the IS-1 connector, leaving the

    transducer and mounting bracket in place.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    7/69

    ____________________________________________________________________________________MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 7 of 69Otologics Confidential

    PROCEDURAL RISKS:

    The implantation of the MET Fully- Implantable

    Ossicular Stimulator requires a surgical procedure

    involving removal of tissue and bone. As with anysurgical procedure involving a general anesthetic,

    certain risks exist.

    Risks that may occur during surgery:The risks involved in surgery and complicationsrelated to surgery include the normal risks of the

    operative procedure and general anesthesia. Theseinclude pain, blood clots, and/or circulatory problems,infection, cardiac arrest, cholesteatoma, and even

    death. Specific risks related to operating on themastoid include infection, inflammation, numbness orstiffness about the ear, disturbance of taste or balanceand noticeable change in head noise.

    The potential risks related to inserting the implantinclude disarticulation of the ossicles and increased

    pressure on the ossicular chain. Increased pressure

    during insertion may cause damage to the inner earand increased hearing loss.

    Risks that may occur after surgery:Potential risks after surgery include inflammation,facial paralysis and perilymph fistula with the

    possible development of meningitis, tinnitus andcomplications associated with general anesthesia.Other postoperative problems to consider arelabyrinthitis, vertigo, puncture or displacement of the

    eardrum, and otitis media.

    Potential risks related to the implant followingsurgery include infection, reparative granuloma and

    inflammatory reaction.

    It is possible that dislocation of the implant, a reactionto the implant materials, and/or failure of theimplanted device can occur and may necessitate anoperation to remove and/or replace the implant.

    SAFETY INSTRUCTIONS HIGHLIGHTED INTHIS MANUAL

    Important Information!

    Special information that supports correct handling of

    the implants components and helps avoid

    operational errors is denoted as importantinformation and printed in italics.

    Important Information!

    This is an example of information that helps avoid

    intraoperative errors.

    CAUTION!Important information that helps prevent damage

    occurring to the implants components is signaled by

    the word CAUTION! and printed in bold type.

    CAUTION!

    This is an example that helps avoid damaging the

    implants components.

    DANGER:

    If the instructions have to be followed strictly to

    exclude the possibility of damage or injury to the

    MET Ossicular Stimulator, patient or to a third party,

    this is signaled by the word DANGER and framed

    for emphasis.

    DANGER:

    This is an example of instructions that help avoid

    damage to the implant or injury to the patient.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    8/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 MSurgical Training Manual Page 8 of 69

    Otologics Confidential

    CANDIDACY CRITERIA

    Selection Criteria for the MET Fully-

    Implantable Ossicular Stimulator1

    18 years of age or older

    Bilateral moderate to severe sensorineuralhearing loss within the Candidacy Audiogram

    range shown below2 (Figure 6)

    Unaided NU-6 or W-22 score 40% at 80 dBHL or 40 dB SL in the ear to be implanted

    Post linguistic onset of hearing loss

    Non-fluctuating, stable hearing loss

    Normal middle ear anatomy

    Realistic expectations for the middle ear implant

    3 months experience over the past year withappropriate hearing aids (either binaural

    amplification or a monaural fitting in the ear tobe implanted).3 (Note: This criteria applies to

    the US Clinical Trial only.)

    1 See Audiologists Manual for further detailed informationon candidacy.2 Moderate to severe loss is defined in terms of the puretone average (PTA) at three frequencies; 500, 1000, and2000 Hz or a high frequency pure tone average (HFPTA =average of 1000, 2000, 3000, and 4000 Hz) of 40-80 dB HL.3 The hearing aid(s) meet the insertion gain prescriptivetarget for an appropriate prescriptive formula for the hearingaid circuitry within 10 dB at 500, 1000, and 2000 Hz andwithin 15 dB at 4000 Hz for a 70-dB SPL input signal.

    0

    20

    40

    60

    80

    100

    0.25 0.5 1 1.5 2 3 4 6

    Frequency [kHz]

    Hearing

    loss[dBH

    L]

    Figure 6

    Contraindications

    Vestibular disorder, including MenieresDisease

    Osteodegenerative disorders, including PagetsDisease

    Middle ear pathology, including a history ofrecurrent otitis media

    Conductive or mixed hearing loss4(Note: This

    applies to the US Clinical Trial only. Patientswith a MEI in the contra lateral ear or patients

    that are to be upgraded from the Otologics

    semi-implantable device to the fully-implantable

    device will often exhibit conductive components)

    Non-organic hearing loss

    Retrocochlear hearing loss / central auditorynervous system disorder

    Pre-linguistic onset of hearing loss

    Medical contraindications to surgery or use ofthe device

    4 Air-bone gap no greater than 10 dB, 500-4000 Hz

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    9/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 MSurgical Training Manual Page 9 of 69

    Otologics Confidential

    Technical Specifications

    Implant

    Figure 7

    Nominal Operating Parameters

    Battery life: > 5 years

    Battery recharge time: 60 minutes (typical for dailycharge)

    Battery recharge frequency: Daily (up to 3 days

    possible, but longer recharging time needed)

    Peak Output: 121 dB MET Frequency Range: < 200 to 6250 Hz

    Processor Type: 12 Band, 2 Channel Digital

    Processor Materials: Biocompatible titanium, gold, ceramic,

    epoxy, silicone elastomer, platinum-iridium alloy,

    ruby

    Connector Technology: IS-1

    Charger System

    Figure 8

    Charger Body recharges implant battery

    Charger Base recharges Charger Bodybattery

    Turns implant on and off

    Adjusts implant volume control

    Implant functions normally during batterycharging

    Charger battery life: > 5 years

    Charging Coil

    Charger Body

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    10/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 MSurgical Training Manual Page 10 of 69

    Otologics Confidential

    Transducer Loading Assistant

    Assists the surgeon in loading the

    transducer by providing an indication of tip

    contact Monitors current to the transducer to

    determine tip contact

    Powered via USB from the computer thatshows graphical loading information

    Remote

    Figure 9

    Turns implant on and off

    Adjusts implant volume

    Range: Skin contact

    Keychain sized

    Programming System

    Performs diagnostics to assess implant andbattery status

    Programs implant according to hearing

    needs of the wearer with flexibility to adjustgain, compression, and crossover frequency

    Uses OtoFit Fitting Software inconjunction with NOAHlink.

    Figure 10

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    11/69

    ____________________________________________________________________________________MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 11 of 69Otologics Confidential

    NEW IMPLANTATION OF THE FULLY-IMPLANTABLE MET

    OSSICULAR STIMULATOR - FIMOS

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    12/69

    ____________________________________________________________________________________MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 12 of 69Otologics Confidential

    Step 1: Pre-Surgical Implant Preparation

    Required Equipment Purpose

    Implant To charge prior to implantation

    Charger System For charging implant prior to implantation

    Important Information!The MET Fully-Implantable Ossicular Stimulator must be charged prior to surgery. This should be done the day

    before surgery.

    Charging the Implant:

    Plug in the Charger Base Station, and make sure that the Base Station indicator light is green, verifying that the

    charger system is connected to a power source.

    Confirm that the charging system is fully charged by checking to see if the charger power indicator is green.

    (The charger indicator is the light on the far left side of the charger). A full recharge time for the charger body

    is 6 hours.

    To charge the implant, place the charging coil over the implant while it is still in its sterile packaging. The

    charging will take place through the sterile packaging. Fully charging an implant can take up to 3 hours

    (Figure 11). Once the implant is fully charged the Implant charger indicator will turn green. (The Implant

    charger indicator is the light on the far right side of the charger.) Consult the Charger User Guide foradditional information (D105809).

    Figures 11. Charging through the external packaging or through the internal packaging

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    13/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 MSurgical Training Manual Page 13 of 69

    Otologics Confidential

    Step 2: Making the Incision

    Required Equipment Purpose

    Capsule Template (included in MET-1050) Determine size of incision required to allow placement

    Key points:

    1. Ensure that the incision is large enough to permit placement of the electronics capsule.2. Make the incision so that the Charging Coil will not rest directly on it.

    Important Information!Review of a CT Scan taken prior to surgery is helpful to indicate the width of the attic space. The transducer is

    6 mm in width and at least this much width is required for implantation of the device.

    Prior to Surgery:

    Prior to surgery an implant model or capsuletemplate is used to identify the optimal implant

    placement and location of the incision. See dottedline in Figure 12. The optimal position of the

    implant capsule typically lies between lines of 30

    and 90 degree angle relative to the horizontal with

    the silicone charging coil superior to the titanium

    capsule body (see Figure 12) and where the capsuleand coil will sit flat on the patients skull. Although

    the implant capsule and coil have been designed

    for long term implantation, care must be takenwhen inserting the coil into a tissue pocket.

    Excessive manipulation or bending of the coil can

    damage it. In addition, the coil should be placed

    such that it is not bent more than 10 relative to theimplant capsule.

    Prior to starting the surgery, an implant template

    can be placed over the skin and a sterile penmarker used to trace the shape of the implanttemplate. Then, the incision line can be drawn with

    the sterile pen to assist in making sure the incision

    will avoid the electronics and the MET Fully-Implantable Ossicular Stimulator.

    Preoperative Preparation of the Patient:

    Place the patient on the operating room table in the

    standard manner for otologic surgery. Induce

    anesthesia. Prophylactic antibiotics can beadministered at this time. The patients head may be

    shaved over the incision site.

    Any standard cochlear implant incision may be

    applied (Figure 13).

    Retract the auricular flap inferiorly and incise and

    retract the periosteum. Define the lower border of thetemporalis muscle and retract it to expose the base of

    the zygoma anteriorly and to identify the superior edgeof the canal wall and the spine of Henle. It is

    important to extend the incision anterior to the pinna

    for adequate visualization of the ossicles whiledrilling.

    Use the Capsule Template (Figure 14) to determine the

    extent of the incision required to properly place theelectronics capsule. This template represents the size

    of the entire implant, including the microphone and

    integral fixation strap.

    Figure 13 Incision Patterns

    Figure 14 Implant Template

    Figure 12 Implant Placement

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    14/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 MSurgical Training Manual Page 14 of 69

    Otologics Confidential

    Step 3: Drilling the Atticotomy

    Required Equipment Purpose

    Atticotomy Template (SUR-1110) Determine if size of atticotomy is sufficient

    Key Points:1. Ensure that the incus body and malleus head is exposed.2. Create as wide a space as possible between the ear canal and the dura by thinning the bone down over both

    structures.3. The atticotomy must extend 20-25 mm beyond the incus body posteriorly.4. Remove bony ledges deep in the atticotomy since they will impede placement of the transducer.

    The atticotomy is started with a 6 mm burr at 12

    oclock superior to the bony external ear canal at the

    root of the zygoma (Figure 15). This 12 oclock

    position is defined by extending a vertical tangent upfrom the anterior canal wall. The atticotomy should be

    extended anteriorly to expose the head of the malleus

    and posteriorly along the temporal (dural) line,

    approximately 20-25 mm beyond the incus body, to awidth of 10-12 mm. It is not necessary to expose the

    dura, only to define the plane and maintain the

    atticotomy inferior to that plane. If the anterior edge

    of the atticotomy does not adequately expose the

    malleus head, visualization of the laser hole andplacement of the probe tip will be difficult. Keep the

    atticotomy borders vertical in order to preserve the

    surrounding cortex for the mounting bracket screws.

    When drilling the atticotomy, skeletonize the bony

    external canal and tegmen down to Krners septumjust superior to the incus. Krners septum forms the

    roof of the middle ear, is approximately 10 to 12 mmdeep from the cortex. Once Krners septum isidentified, it is carefully removed using a diamond

    burr to expose the epitympanum, proximal antrum,

    the incus and head of the malleus. Malleus head

    exposure is important in order to obtain a clear view

    of the incus body during laser use and probe tip

    placement (Figure 16).

    If required, use the Oval Fitment Gauge (Figure 17)

    as a guide to approximate the size and shape of the

    atticotomy. The oval fitment gauge represents theminimum atticotomy size. Typically the atticotomy

    will extend further anterior for good visualization of

    the incus body and malleus head. It is possible to

    test the width of the bottom of the atticotomy by

    ensuring that a 5mm burr will fit into the opening.

    Important Information!

    It is important to create as wide a space aspossible between the dura and the ear canal

    for device insertion.

    Remove bony ledges near the incus. Theymay hamper advancement of the transducer

    and the probe tip into the laser hole.

    Figure 16 - Atticotomy

    Figure 15 - Atticotomy location (Left Ear)

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    15/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 MSurgical Training Manual Page 15 of 69

    Otologics Confidential

    DANGER: While it is important to create as

    wide a space as possible between the dura and the

    ear canal, it is also important to avoid penetrating

    the dura and to mitigate any penetration of the

    external ear canal.

    Step 4: Positioning the Mounting Bracket

    Required Equipment Purpose

    Mounting bracket with Bending Template attached.

    (included in MET-1050)

    Affix MET Ossicular Stimulator to skull

    Laser Guide Assembly (SUR-1140) Assess size and location of atticotomyDetermine optimal placement for laser hole drilling

    Bending Pliers (SUR-1120) Used to bend legs of mounting bracket (if necessary)

    Cutting Pliers (SUR-1130) Used to remove mounting bracket legs (if necessary)

    Key Points:

    1. The mounting bracket must be placed so that the tip of the Laser Guide assembly points directly at thelocation on the incus where the laser hole will be made (Figures 19 and 20).

    2. The Laser Guide tip angle may be altered by loosening the locking ring and moving the handle.3. The mounting bracket legs can be bent or removed (with the cutting pliers) to allow flush placement

    against the patients cortex.

    The Laser Guide Assembly, which reproduces the geometry and dimensions of the transducer assembly, is

    used to position the optical fiber on the body of the incus. The hospital surgical staff will place the Laser

    Guide Assembly (Figure 18a) in the mounting bracket assembly (Figure 18b). Test fit the mounted LaserGuide Assembly in the atticotomy (Figure 19).

    If necessary, enlarge the atticotomy until the mounting bracket fits in the hole and the Laser Guide Assembly

    tip is just above the center of the body of the incus. The axis of the tip must be as perpendicular to the incusbody as possible. It is important to avoid the edge of the incus, or the connection may not be secure (Figure

    20). The angle of the Laser Guide Assembly can be adjusted by slightly unscrewing the locking ring tool and

    moving the Laser Guide tip both angularly and axially. The tip should point directly at the spot intended for

    the laser hole and should be able to touch the incus. Once the tip has been properly aligned, retighten the

    locking ring tool.

    Figure 17 - AtticotomyTemplate

    10 mm wide x 20 mm lon

    Locking RingTool

    Handle

    Laser Guide Tip

    Figure 18a Laser Guide Assembly.

    Bending Template,

    comes pre-assembled

    on Bone Bracket.

    Bone Bracket

    Figure 18b Bone Bracket with Bending Template.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    16/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 MSurgical Training Manual Page 16 of 69

    Otologics Confidential

    While holding the Laser Guide assembly in position, bend the Surgical Bending Template legs (shown blue in

    Figure 19) to conform to the surface of the skull. Once satisfied, remove the Mounting Bracket/Laser Guide

    assembly and bend the legs on the mounting bracket to match the bends previously created on the template.(See Figure 21a)

    Figure 20 Correct Position of Laser Hole

    DANGER: When introducing the Laser Guide

    Assembly, ensure that the tip does not make

    contact with the incus to avoid the possibility of

    disarticulating the ossicles.

    Figure 19 - Placing Mounting Bracket (Right Ear)

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    17/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 MSurgical Training Manual Page 17 of 69

    Otologics Confidential

    It will be necessary to bend the legs of the mounting bracket with the Bending Pliers to achieve a flush fit at

    the mounting location on the cortex. (Figure 21a & 21c)

    Once the Bone Bracket is fit to the surface of the cortex, the bending template (blue) can be removed from themounting bracket using the bending pliers. The center tab is bent upwards and the blue template is pulled off

    the mounting bracket. (Figure 21b)

    DANGER:

    The Surgical Bending Template must be removed prior to securing the Mounting Bracket to the

    cortex. It is not intended to be left inside the body, due to sharp edges, though it is of the same

    titanium material as the bone bracket.

    Figure 21a Bending Bone Bracket Legs to Match Template.

    Figure 21b Removing Surgical Bending Template from Bone Bracket.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    18/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 MSurgical Training Manual Page 18 of 69

    Otologics Confidential

    The bone screws should not be used to pull a mounting bracket leg down to achieve a flush fit of the mountingbracket to the skull. The legs should be bent (using bending pliers) to achieve this purpose. If necessary, one

    mounting bracket leg can be removed using the Cutting Pliers (Figure 22a & 22b).

    Figure 22b

    Cutting Pliers

    Figure 21c Correct Fit of Bracket to Skull

    Figure 22a Bending Pliers

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    19/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 MSurgical Training Manual Page 19 of 69

    Otologics Confidential

    Step 5: Fastening the Mounting Bracket

    Required Equipment Purpose

    Drill bit (included in MET- 1050) Drill proper size and depth holes for mounting bracket

    Mounting screwdriver (Blade, SUR-1160; Handle, SUR-

    1170)

    Screw mounting bracket to skull

    4 mm Bone screws (included in MET-1050) Affix mounting bracket to skull

    Oversize bone screws (included in MET-1050) Substitute for 4 mm bone screws (if necessary)

    Key Points:

    1. Ensure that the Laser Guides tip is aligned to the exact desired location for the incus laser hole beforedrilling bone screw holes.

    2. Screw down each leg with at least one screw per leg.3. Use the oversize bone screws if necessary.

    Once the Laser Guide tip points at the location forthe laser hole and the mounting bracket legs have

    been adjusted to fit flush against the cortex, use

    the drill bit provided in the surgical kit to drill a

    hole through one of the mounting holes located atthe end of one of the mounting bracket legs. Use

    the Otologics screwdriver (Figure 23) and a 4 mm

    screw (Figure 24) to secure the mounting bracket

    leg to the skull. Repeat this process for all of theremaining legs of the mounting bracket utilizing

    one bone screw per leg.

    It is not necessary to place more than one bonescrew per mounting bracket leg, but in order to

    ensure a secure mount there should be at least one

    screw in all four legs. If it is not possible to put ascrew in a particular leg, that leg should be

    removed using the cutting pliers.

    Important Information!

    A minimum of 3 legs must be attached tothe skull. If 3 legs cannot be attached,

    remove and reposition the mounting

    bracket!

    All legs must have at least one screwsecurely fastening the mounting bracket

    to the skull. If a leg is not securely

    fastened to the skull pressure on the leg

    may cause probe tip movement and

    decrease patient performance or

    transducer damage (Figure 25)

    If for any reason, the laser guide assembly tipcannot be properly aligned, the screws may be

    removed and the mounting bracket repositioned.

    In order to use the same hole again, it may be

    necessary to use the oversize bone screw.

    Alternatively, a new hole can be drilled into theskull through one of the other holes on the

    mounting bracket leg.

    Important Information!

    As the screws are fastened, ensure thatthe mounting bracket does not move and

    change the intended approach angle for

    the tip of the Laser Guide Assembly.

    The mounting bracket should be fastenedwith 4mm bone screws. The oversize

    bone screws should be used if the bonescrewholes become stripped.

    Manual Drilling: If for any reason thedrill bit provided does not fit in the

    otologic drill, the screwdriver blade can

    be removed from the screwdriver handle

    and the drill bit inserted in its place. The

    proper sized hole can then be drilled by

    twisting the drill bit/handle by hand.

    Figure 23 - Screwdriver

    Figure 24 - Bone Screw

    Figure 25 Bend Leg to Skull

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    20/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 20 of 69Otologics Confidential

    Step 6: Creating the Laser Hole

    Required Equipment Purpose

    Laser guide tip Used to determine laser alignment to incus.

    Laser fiber Used to deliver laser energy to precise point on theincus

    Hypodermic Tube (included in MET-1050) Used to handle and align laser fiber

    Hand auger (included in MET-1050) Used to remove debris from laser hole

    Laser Used to produce energy to cut hole in incus

    Laser fiber cutter (SUR-2110) May be required to prepare laser fiber tip

    Laser fiber stripper (SUR-2120) May be required to prepare laser fiber tip

    Key Points:1. Line up the Laser Guide Assembly with the exact position of the laser hole2. Verify the sheath on the end of the laser fiber has been stripped back a minimum of 2mm from then end.

    If not, strip the sheath using the Laser Fiber Stripper Tool.

    3. Generate a laser hole at least 0.75mm deep (measured to the shoulder of the hand auger)4. Wear appropriate eye protection.

    Ensure the Laser Guide Assembly is locked into

    position by tightening the locking ring tool. TheLaser Guide Assembly should be mounted so that

    the tip is just touching the incus.

    The Laser Guide tip is then removed from the LaserGuide Assembly (Figure 26). The Hypodermic Tube

    (Figure 27) is placed over the end of the optical

    fiber. The laser fiber should extend out of the

    hypodermic tube no more than 1 mm (Figure 28).

    The Hypodermic Tube and optical fiber are theninserted through the laser guide and the optical fiber

    is used to deliver laser energy to the body of incus.

    The laser fiber should be in contact with the incus forproper laser cutting. The laser fiber may be prepared

    for cutting by test firing several pulses on a sterilewooden spatula.

    Typical settings with a KTP 532 nm laser energy are

    0.1 second pulse bursts at 0.5 second intervals with a

    power setting of 2.4W. This laser should be run in

    50 pulse sets and can take between 300 and 450

    Figure 27Hypodermic Tube

    Figure 26

    Laser Guide (Separated)

    Figure 28

    Laser Fiber Protruding 1 mm out of

    Hypodermic Tube

    Laser Guide Tip

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    21/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 21 of 69Otologics Confidential

    pulses on average to complete the hole. Typicalsettings with a DIODE 810nm laser energy are 0.2

    second pulse at 0.7 pulse rate with a power setting of

    3W. This laser should be run in 20 pulse sets and

    can take 60 to 80 pulses on average to

    complete the hole. Consult your Laser representative

    and your Otologics representative to determine the

    correct settings for your laser model and type. It isrecommended that irrigation be used throughout the

    creation of the laser hole. In order to remove the

    char after sufficient laser pulses, the tube and fiberare removed from the laser guide and replaced by the

    hand auger. Debris is removed using the manually

    operated hand auger (Figures 29 and 30) and suction-

    irrigation.

    The hand auger is best utilized by rolling the handle

    back and forth between the thumb and index finger.The laser ablation process is repeated until the hole

    is approximately 0.75 mm deep at the 3:00 and 9:00

    positions of the hole, as determined by the shoulder

    at the tip of the hand auger (Figure 31). The holeshould be surrounded by bone at least 0.25 mm deep

    on the shallowest side. The laser guide is thenremoved from the mounting bracket in preparation

    for transducer placement.

    DANGER: Ensure all personnel are aware of the

    use of the laser and are taking proper

    precautions.

    DANGER: Ensure the correct diameter fiber is

    in use to avoid creating an oversized laser hole.

    Should the initial hole form a trench reposition the

    laser guide and cut a second hole (Figure 32)

    Figure 30Rotating the Hand Auger Between the

    Thumb and Forefinger (Right Ear)

    Figure 31

    Correct Depth of Laser Hole

    Figure 32

    Second Hole Drilled After an Incorrect

    Trench

    Auger Tip

    Figure 29

    Hand Auger

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    22/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 22 of 69Otologics Confidential

    Step 7: Preparing the Bone Beds and Lead Channels

    Required Equipment Purpose

    Capsule Template (included in MET-1050) Determine size and location of bone bed

    Key Points:

    1. Cover the atticotomy during bone bed drilling2. Use the capsule template to assess proper location for the bone bed.

    After removing the laser guide from the mounting

    bracket, fill and/or cover the atticotomy with

    protective material to keep the area free of debrisduring bone bed drilling.

    Using the capsule template (Figure 33)for sizing,

    bone-beds are created in the skull at the positionsdefined prior to surgery, keeping in mind the

    recommended location for the microphone, shown in

    the area without muscle, directly posterior to theexternal auditory canal (Figure 12, page 13).

    The Capsule bone bed (Figure 34) should be drilledto the depth represented by the thickest portion of the

    template, for the entire size of the implant, not

    including the Antenna Coil area. The intent of thebone bed is to reduce the curvature of the skull,

    providing the minimum attainable deflection of the

    coil. The capsule placement shown in Figure 12

    Implant Placement represents the portion of the

    skull with the minimum curvature as a starting point.

    The capsule must not be placed so that it touches anyof the other components or their mounting straps.

    The Capsule Antenna Coil should be centered underthe position representing optimal placement of the

    Charging Coil. If the total thickness of the tissue

    flap, including skin and muscle, is greater than 6

    mm, it must be thinned down to 6 mm to ensureproper magnetic fixation of the Charging Coil.

    The bone bed for the microphone may be done one

    of two (2) ways (Figure 34). (1) At a minimum, acircular depression may be made to accommodate

    the raised portion on the under or back-side of the

    microphone, or (2) A larger area to fit the shape for

    the entire device may be drilled. A lead channel

    should be drilled for the microphone lead in

    either case. Also, a channel should be drilled to

    accommodate the transducer lead. Forperformance reasons the microphone, and its straps,

    must not touch any of the other implant components.

    It is important to consider the placement of the

    microphone, capsule, and leads relative to the temple

    pieces of eyeglasses or other head gear. Themicrophone should not be placed under or adjacent

    to muscle tissue, such as the temporalis muscle, as

    the localized tension created by the muscle will

    affect the performance of the microphone.

    Adequate space is required to allow for the

    installation of the silicone straps on the implant to

    fit securely into the bone bed for the implant.

    DANGER: Cover atticotomy during bone bed

    drilling to prevent debris from entering the

    atticotomy.

    Figure 33- Capsule Template

    Figure 34

    MET Fully-Implantable Ossicular

    Stimulator Bone Beds and Lead Channels

    (Right Ear)

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    23/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 23 of 69Otologics Confidential

    Step 8: Placing the MET Ossicular Stimulator

    Required Equipment Purpose

    MET Ossicular Stimulator Implanted into atticotomy

    Transducer Insertion Instrument (SUR-1190) To roughly position the MET Ossicular Stimulator tip

    Locking ring tool (SUR-1150) To tighten locking ring in place when tip is alignedLocking Ring (included in MET-1050) Tightened by locking ring tool to hold MET Ossicular

    Stimulator in place

    Key Points:

    1. When threading the MET Ossicular Stimulator into the mounting bracket, take great care with the probetip.

    2. A flatter approach to the mounting bracket may make it easier to thread the device into the bracket.3. Position the probe tip within 2mm of the laser hole in the incus.

    The Transducer Insertion Instrument, Locking ringtool and Locking Ring are assembled onto the

    transducer (Figure 35 a & b).

    The transducer is then threaded into the mounting

    bracket (Figure 35 c & d) with the transducer in a

    fully retracted position. Slightly loosen thelocking ring tool to allow the Transducer Insertion

    Instrument to move. Using the Transducer

    Insertion Instrument, make angular and grossdepth adjustments to align the transducer probe tip

    with the laser-ablated hole in the body of the incus

    (Figure 36). Position the probe tip of the

    transducer 1-2 mm from the body of the incus.Using the locking ring tool, tighten the mountingsystem locking ring and lock the transducer in

    place. Remove locking ring tool and Transducer

    Insertion Instrument.

    Figure 35 d

    Installing MET Fully Implantable Ossicular

    Stimulator in Mounting Bracket (Right Ear)

    Figure 35 a

    Assembling the Transducer Insertion Instrument,Locking Ring tool and locking ring onto the rear of

    Figure 35 b

    Figure 35 c

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    24/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 24 of 69Otologics Confidential

    Important Information!

    Position the tip within 2mm of thelaser hole before attempting to

    advance the tip into the hole with

    the advancement screwdriver.Advancing the probe tip beyond

    5mm (20 turns) may decouple the

    transducer body from the

    mounting shaft.

    The transducer motor willdecouple from the mounting shaft

    if the transducer is extended more

    than 6mm (24 turns).

    CAUTION!

    The probe tip is extremely sensitive

    and must not make contact with anyobject.

    DANGER: To ensure probe tip protection, fully

    retract the probe tip before threading the

    transducer into the mounting bracket.

    DANGER: While threading the transducer

    into the mounting bracket, hold the Transducer

    Insertion Instrument to prevent it from falling

    into the atticotomy and contacting the incus.

    Figure 36

    Aligning Probe Tip to Hole in Incus (Right Ear)

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    25/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 MSurgical Training Manual Page 25 of 69

    Otologics Confidential

    Step 9 Loading the MET Ossicular Stimulator and Diagnostics

    Required Equipment Purpose

    Adjustment Screwdriver (SUR-1180) Used to advance MET Ossicular Stimulator tip into hole in

    incus

    Transducer Loading Assistant (TLA) Used to promote appropriate loading

    Diagnostic System Use to confirm operational state of the implant

    Key Points:

    1. Align the probe tip directly with the hole in the incus.2. Test mobility of the incus prior to and after inserting the probe tip into the incus hole.3. Advance the transducer probe tip until the incus moves and then advance the tip turn of the adjustment

    screwdriver.4. If retraction is necessary, be sure to retract the probe tip AT LEAST TURN MORE than the stopping-

    point desired, or completely retract probe tip out of the hole and begin process again.

    5. Final transducer placement should always be an ADVANCING motion.

    CAUTION!

    When using the TLA, the coil must be

    wrapped in a sterile bag to be placed over theimplant coil.

    Prior to advancing the probe tip, the mobility of the incus should be tested by moving it gently with a surgical pick

    (Figure 39). The degree of movement should be noted.

    The Adjustment Screwdriver is fitted to the hex nut on top of the mounting shaft to make the final probe tipinsertion adjustment.

    If using the Transducer Loading Assistant (TLA), refer to the TLA Instructions for Use (D107454-US or

    D106317-EU) at this point for specific instructions for loading the transducer.

    If not using the TLA, proceed as follows. To achieve proper loading of the ossicular chain with the transducer,the adjustment screwdriver (Figure 37) is turned clockwise to advance the transducer probe tip until it makes

    contact with the bottom of the hole and the incus moves (Figure 38). The tip should then be advanced an additional turn of the adjustment screwdriver. A surgical pick should be used again to test mobility of the ossicles. Upon

    final loading, the ossicles should demonstrate similar pre-placement mobility (Figure 39).

    During normal use the transducer tip vibrates in and out 4 um, thus it is essential that the tip be in contact with the

    bottom of the hole. If the transducer has been advanced too far, it must be retracted to avoid creating a conductive

    hearing loss.

    Figure 37

    Adjustment Screwdriver

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    26/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 26 of 69

    Otologics Confidential

    When retracting the transducer, be sure to retract AT LEAST TURN MORE than the stopping-point desired.

    That is, if the transducer has been over-advanced by turn, then the transducer must be retracted a minimum of

    turn and then advanced turn to result in a net retraction.

    CAUTION! Final transducer placement should always be anADVANCING motion.

    Once the transducer is in its final position, a pick should again be used to test the mobility of the ossicles. If thereappears to be stiffening in the ossicular chain, the entire procedure should be repeated until there is minimal

    stiffening of the ossicular chain (Figure 39).

    Important Information!

    Do not overextend the probe tip by trying to advance it more than 20 turns (5 mm).

    One complete turn (360) of the screwdriver is equivalent to 0.25 mm of probe tip insertion or retraction.

    The transducer mounting shaft has been designed for subsequent readjustment if necessary to obtainoptimum performance.

    DANGER: Overloading may cause a conductive hearing loss, damage to the ossicles and/or damage to the

    cochlea.

    Figure 38

    Advancing probe tip (right ear)

    Figure 39

    Incus Mobility Test

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    27/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 27 of 69

    Otologics Confidential

    Step 10: Attaching the Electronics Capsule to the Transducer

    Required Equipment Purpose

    Disposable Torque Wrench Used to tighten connector block set screws securing transducer lead to

    electronics capsule

    Transducer with lead To be attached to electronics capsule to form complete unit

    Electronics Capsule To be attached to transducer to form complete unit

    The transducer can now be attached to the

    electronics capsule by inserting the male connectorof the transducer lead into the female connector of

    the electronics capsule (Figure 40). Once the

    connector has been fully inserted into the

    electronics capsule, the two screws are tightened(clockwise) with the torque wrench provided using

    the following procedure:

    Figure 40

    1. The electronics capsule is constructed so that

    the male transducer connector may be insertedinto the female electronics connector.

    Lubricate the male connecting pin with sterilesaline.

    2. While firmly holding the electronics capsuleby the pin connector block between the thumband forefinger (taking care to not touch the

    microphone), insert the male connecting pin

    into the female connecting block until the pin

    hits the positive stop on the inside of theconnector and can be seen protruding past the

    pin connector block (Figure 41 & 42).

    3. Two locking screws will be tightened with thetorque wrench to hold the connector in place.(Figure 44)

    4. While firmly supporting the pin connectingblock between the thumb and forefingergently push on the torque wrench while

    turning clockwise to engage the wrench headinto the screw.

    5. Once the wrench has been felt to engage thescrew, turn the wrench clockwise until the

    torque wrench clicks once. This will take

    between and turn.6. While still firmly supporting the pin

    connecting block between the thumb and

    forefinger remove the torque wrench, and

    gently pull on the lead to confirm contact.

    7. Repeat steps 5-7 above for the ring lockingscrew while continuing to firmly support the

    electronics capsule by holding the ring

    connecting block between the thumb and

    forefinger.8. Install the septum Plug into the connector

    blocks by pressing each end of the molded

    septum plug into the appropriate connectorblock.

    Figure 41

    Set Screws

    torqued with

    Torque Wrench

    Tip protrusion

    Figure 42

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    28/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 28 of 69Otologics Confidential

    Caution!

    Over tightening may cause the screwhead to be stripped and may cause the

    connector blocks to be pulled free from

    the capsule.

    Always support the connector blocks

    while tightening or loosening thescrews. Failure to do so may damage

    the electronics capsule.

    Never attempt to turn the connectorscrews with anything other than the

    supplied torque wrench. Failure to do

    so may result in implant damage.

    Failing to insert the transducerconnector fully into the connector

    blocks may cause damage to the

    transducer when the set screws are

    secured.

    1. The recess above the screws will besealed with Septum Plug ) The SeptumPlug should be installed using the

    following procedure:

    a. Wipe the area with sterile wipeto assure it is dry of blood or

    saline solution.

    b. Align the septum plug with thetwo connector blocks..

    c. Apply pressure to seat theseptum plug into the screw

    holes. (Figure 43a and b)d. Verify septum plug is seated into

    the connector block surface and

    is covering the entire connector

    block and screw surfaces.

    Important Information!

    Never tighten connector block screws without the

    male connecting pin inserted into the female

    connecting blocks.

    Always lubricate the male connecting pin with

    saline prior to insertion into the connector block.

    Failure to do so will result in an improperconnection. Never fill the female connecting

    block with fluid. To do so will result in an

    improper connection.

    Figure 43a Top View

    Figure 43b Side View

    After the Septum Plugs are firmly seated, a pair of Silicone Bands must be slid over the implant to secure them and

    ensure a proper seal. As the Capsule is already connected to the Transducer, the Silicone Bands will need to be slid

    over the Coil end. Care should be exercised not to bend the Coil portion of the Capsule more than a few degrees.Sterile saline may be used as a lubricant to allow the Bands to slide completely over the Septum Plugs (Figure 45a).

    The bands should be placed so they do not lie on the curved portion of the Capsule. This is to prevent them from

    slipping off (Figure 45b).

    Figure 44 Torque Wrench

    Figure 45a Silicone Bands Figure 45b Silicone Bands Installed

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    29/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 29 of 69Otologics Confidential

    Surgical Diagnostics:

    Some tests should be performed during

    implantation. These tests help to make sure that

    the transducer and microphone are fully functionaland will help detect if the device has been

    damaged, e.g. during handling or in the OR during

    implantation. The data collected can be used laterin the clinic for diagnostics and troubleshooting.

    These tests are:

    Battery test (reads charge state)

    SAFI test:Microphone functionality

    Functional Test

    Transducer Impedance

    CAUTION!

    For the surgical tests, the programming coil

    must be wrapped in a sterile bag to be placed

    over the implant coil.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    30/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 30 of 69Otologics Confidential

    Step 11. Securing the Electronics Capsule and Closing the Incision

    Required Equipment Purpose

    Flap Thickness Gauge (SUR-1200) Used to measure skin flap thickness over the device to

    ensure proper device function.

    Key Points:

    1. If the tissue flap is thicker than 6mm, it must be thinned down to ensure that the Charging Coil will remainattached to the patient during charging of the implant and the Programming Coil remains fixated during

    the fitting process.

    2. Only bi-polar cauterization is permitted once the implant has been placed.

    Important Information!3. Integral straps are provided for securing the electronics capsule and the microphone pendant to the bone

    bed. Additional bending and fitting of the straps may be required. The straps will constrain the implant

    and the microphone in a fixed location following the installation of the bone screws. For this reason

    each strap will require the use of two bone screws. The devices should be tested for movementfollowing the installation of the bone screws. If movement is present additional adjustment to the straps

    must be performed. Additional bending and shaping of the strap may be performed using the surgical

    pliers used to form the bone-mounting bracket. Precautions for bending are described in detail in the

    Instructions for Use included with the Bone Bracket.

    4. CAUTION! The leads of the Transducer and Microphone should not be placed in contact with the legs ofthe mounting bracket to avoid damage to the leads.

    5. CAUTION! Place the Microphone in the recommended location as in figure below, directly posterior tothe external auditory canal and route the lead from the microphone to the Capsule so that there is no

    tension on either end of the lead. DO NOT pull on the leads or create tight bends on them, especially inthe areas where they enter the Pendant and Capsule bodies.

    Implant Location

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    31/69

    ____________________________________________________________________________________MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 31 of 69Otologics Confidential

    The thickness of the tissue flap is measured using

    the flap measurement tool (Figure 44) and noted.

    If it is thicker than 6mm, the tissue must be

    thinned down. If it is too thick, the magnet in theCharging Coil may not be strong enough to remain

    attached to the magnet in the electronics capsule

    and it may result in poor conduction. The tissueflap is sutured down to cover the electronics

    capsule and the wound is closed in layers (Figure45).

    DANGER: Once the electronics capsule

    and pendant microphone have been

    implanted, only bi-polar cauterization is

    allowed. The use of mono-polar cauterizing

    equipment may damage the implanted

    electronics.

    Important Information! Make sure that the

    implant is positioned with THIS SIDE OUT

    facing away from the skull (lateral).

    At the surgeon's discretion, the implant

    electronics may be placed under or over themuscle to achieve a total tissue thickness over

    the coil and magnet of 6mm (Figure 46).

    At this point, tests should be conducted using

    the Surgical Assistant FIMOS Implant software

    (SAFI) to make sure that the transducer and

    microphone are fully functional and whether thedevice has been damaged. Refer to the SAFI

    Instructions for Use D107455.

    Figure 47, Measuring Flap Thickness

    Figure 48

    Cross Section Showing Capsule

    Placement

    Figure 46, Flap Thickness Gauge

    DANGER: Excessive manipulation or

    bending of the coil can damage it. In addition,

    the coil should be placed such that it is not

    bent more than 10relative to the implant

    ca sule.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    32/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 32 of 69Otologics Confidential

    Step 12. Post Surgical Care and Activation

    Post Surgery Care:

    Follow up patient care is performed in the usual manner for otologic surgery. The patient is given the samepostoperative instructions provided after standard otologic surgery. If the patient was wearing amplification

    prior to surgery, he or she may wear amplification in the non-implant ear, but should not wear amplification inthe implant ear until after the eight week healing period is complete and medical clearance is given.

    Activation:

    Following a 8-week healing period and medical clearance, the MET Fully-Implantable Ossicular Stimulator

    will be programmed according to the hearing needs of the individual patient. Detailed procedures for fitting

    the MET Fully-Implantable Ossicular Stimulator are provided in the MET Fully-Implantable Ossicular

    Stimulator Fitting Guide.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    33/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 33 of 69Otologics Confidential

    UPGRADE IMPLANTATION FROM THE SEMI-

    IMPLANTABLE OSSICULAR STIMULATOR (SIMOS) TOTHE FULLY-IMPLANTABLE MET OSSICULAR

    STIMULATOR (FIMOS)

    (E.U. ONLY)

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    34/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 34 of 69Otologics Confidential

    Step 1: Pre-Surgical Preparation for the SIMOS to FIMOS

    Implantation

    Important Information!

    The MET Fully-Implantable Ossicular Stimulator must be charged prior to surgery using the Charger and

    instructions for use for the Charger.

    Important Information!

    A CAT scan or a lateral x-ray must be performed to determine location of the implanted lead and mounting

    bracket.

    Important Information!Integral straps are provided for securing the electronics capsule and the microphone to the bone bed.

    Additional bending and fitting of the straps may be required. The straps should constrain the implant or the

    microphone in a fixed location following the installation of the bone screws. For this reason, two bone

    screws must be used in each strap. The devices should be tested for movement following the installation ofthe bone screws. If movement is present additional adjustment to the straps must be performed. Additional

    bending and shaping of the Bone Bracket may be performed using the surgical pliers used to form the bone-

    mounting bracket. Precautions for bending are described in detail in the Instructions for Use included with the

    straps. The recommended locations for the straps are also indicated in the Instructions for Use provided withthe brackets.

    CAUTION! The leads of the Transducer and Microphone should not be placed in contact with the legs of themounting bracket to avoid damage to the leads.

    CAUTION! Place the Microphone in the recommended area (Figure 12 page 13), directly posterior to the

    external auditory canal, and route the lead from the microphone to the Capsule so that there is no tension on

    either end of the lead. DO NOT pull on the leads or create tight bends on them, especially in the areas wherethey enter the Pendant and Capsule bodies.

    CAUTION: For an upgrade implantation, only bi-polar cauterization is permissible. The use

    of mono-polar cauterizing equipment may damage the implanted transducer.

    Charging the Implant:

    Plug in the Charger Base Station, and make sure

    that the Base Station indicator light is green,verifying that the charger system is connected to a

    power source.

    Confirm that the charging system is fully charged

    by checking to see if the charger power indicator isgreen. (The charger indicator is the light on the far

    left side of the charger). A full recharge time for

    the charger body is 6 hours.

    To charge the implant, place the charging coil over

    the implant while it is still in its sterile packaging.The charging will take place through the sterile

    packaging. Fully charging an implant can take up

    to 3 hours. Once the implant is fully charged the

    Implant charger indicator will turn green. (The

    Implant charger indicator is the light on the far right

    side of the charger.) Consult the Charger User

    Guide for additional information.

    Determine location of implanted electronics:

    The surgeon should determine the location of the

    implanted electronics using a CAT scan or x-ray

    and manual palpation. Once the location of theelectronics has been identified, the patient may be

    prepared for surgery.

    Preoperative Preparation of the Patient:

    The patient is placed on the operating room tablein the standard manner for otologic surgery and

    anesthesia is induced. Prophylactic antibiotics can

    be administered at this time as well. The patients

    head may be shaved over the incision site.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    35/69

    ____________________________________________________________________________________MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 35 of 69Otologics Confidential

    Step 2: Making the Incision and Removal of the Semi-

    Implantable Electronics

    The original incision site should be utilized (Figure 48), and should allow for adequate access to the implant

    electronics capsule.

    DANGER: Care must be taken to not cut the electronics capsule during the incision or

    dissection.

    Figure 49

    Once the implant electronics has been located, incise the sheath covering the implant electronics, and cut any

    sutures that may be holding the capsule in place.

    Remove the implant from its tissue bed taking care not to stress the transducer lead.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    36/69

    ____________________________________________________________________________________MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 36 of 69Otologics Confidential

    Step 3: Connector Disassembly

    Required Equipment Purpose

    Disposable Torque Wrench Used to tighten or loosen septum screw securing transducer lead to electronics capsule

    Transducer with lead To be attached to electronics capsule to form complete unit

    Electronics Capsule To be attached to transducer to form complete unit

    Using a scalpel, cut around the front and rear

    septums to expose the allen-head screws of the

    front and rear connector blocks (Figure 49).

    Place the head of the torque wrench (Figure 53page 35) into the allen slot in the front connector

    block and turn counter clockwise to loosen the

    screw.

    Grasp the lead strain relief and gently pull the

    male connecting pin (Figure 49) from theelectronics capsule.

    Take care not to damage the male connecting pin

    in any way. Clean tissue and blood from theconnecting pin and inspect the front and rear seals

    for cracks and tears.

    Measure, record and report to Otologics the tissue

    thickness. If the tissue that over-lies the

    electronics capsule is thicker than 6 mm, thin the

    tissue to 6mm prior to connecting the new

    electronics capsule.

    Figure 49 (left ear) Figure 50

    Pin connector

    block

    Ring Connector block

    Male connector pin

    Female connector

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    37/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 37 of 69Otologics Confidential

    Step 4: Attaching the Electronics Capsule to the Transducer

    Required Equipment Purpose

    Disposable Torque Wrench Used to tighten or loosen septum screw securing transducer lead to electronics capsule

    Transducer with lead To be attached to electronics capsule to form complete unit

    Electronics Capsule To be attached to transducer to form complete unitBone Wax Minimally

    Resorbable (E.U. Only)

    If used in place of provided Septum plug and bands, to create a moisture barrier on

    outside of Implant connector block set screws

    The transducer can now be attached to theelectronics capsule by inserting the male connector

    of the transducer lead into the female connector of

    the electronics capsule (Figure 51). Once the

    connector has been fully inserted into the

    electronics capsule, the two screws are tightened(clockwise) with the torque wrench provided using

    the following procedure:

    Figure 51

    1. The electronics capsule is constructed so thatthe male transducer connector may be inserted

    into the female electronics connector.

    Lubricate the male connecting pin with sterilesaline.

    2. While firmly holding the electronics capsuleby the pin connector block between the thumband forefinger (taking care to not touch the

    microphone), insert the male connecting pin

    into the female connecting block until the pin

    hits the positive stop on the inside of theconnector and can be seen protruding past the

    pin connector block (Figure 52 & 53).

    3. Two locking screws will be tightened with thetorque wrench to hold the connector in place.(Figure 55)

    Figure 52

    4. While firmly supporting the pin

    connecting block between the thumb andforefinger gently push on the torque

    wrench while turning clockwise to

    engage the wrench head into the screw.

    5. Once the wrench has been felt to engagethe screw, turn the wrench clockwiseuntil the torque wrench clicks once. This

    will take between and turn.

    6. While still firmly supporting the pinconnecting block between the thumb andforefinger remove the torque wrench, and

    gently pull on the lead to confirm contact.

    7. Repeat steps 5-7 above for the ringlocking screw while continuing to firmly

    support the electronics capsule by holding

    the ring connecting block between the

    thumb and forefinger.8. Install the septum Plug into the connector

    blocks by pressing each end of the

    molded septum plug into the appropriate

    connector block.

    Tip protrusion

    Figure 53

    Set Screws

    torqued with

    Torque Wrench

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    38/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 38 of 69Otologics Confidential

    Caution!

    Over tightening may cause the screwhead to be stripped and may cause the

    connector blocks to be pulled free from

    the capsule.

    Always support the connector blockswhile tightening or loosening the

    screws. Failure to do so may damagethe electronics capsule.

    Never attempt to turn the connectorscrews with anything other than the

    supplied torque wrench. Failure to do

    so may result in implant damage.

    Failing to insert the transducerconnector fully into the connector

    blocks may cause damage to the

    transducer when the set screws are

    secured.

    2. The recess above the screws will be

    sealed with Septum Plug ) The SeptumPlug should be installed using the

    following procedure:

    a. Wipe the area with sterile wipeto assure it is dry of blood orsaline solution.

    b. Align the septum plug with thetwo connector blocks..

    c. Apply pressure to seat theseptum plug into the screwholes. (Figure 54a and b)

    d. Verify septum plug is seated intothe connector block surface and

    is covering the entire connector

    block and screw surfaces.

    Important Information!

    Never tighten connector block screws without the

    male connecting pin inserted into the female

    connecting blocks.

    Always lubricate the male connecting pin with

    saline prior to insertion into the connector block.

    Failure to do so will result in an improperconnection. Never fill the female connecting

    block with fluid. To do so will result in an

    improper connection.

    Figure 54a Top View

    Figure 54b Side View

    After the Septum Plugs are firmly seated, a pair of Silicone Bands must be slid over the implant to secure them and

    ensure a proper seal. As the Capsule is already connected to the Transducer, the Silicone Bands will need to be slid

    over the Coil end. Care should be exercised not to bend the Coil portion of the Capsule more than a few degrees.Sterile saline may be used as a lubricant to allow the Bands to slide completely over the Septum Plugs (Figure 56a).

    The bands should be placed so they do not lie on the curved portion of the Capsule. This is to prevent them from

    slipping off (Figure 56b).

    Figure 55 Torque Wrench

    Figure 56a Silicone Bands Figure 56b Silicone Bands Installed

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    39/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 39 of 69Otologics Confidential

    Important information:

    Never tighten connector block screws without the male connecting pin inserted into the female connecting blocks.

    Always lubricate the male connecting pin with saline prior to insertion into the connector block. Failure to do so

    will result in an improper connection.

    Never fill the female connecting block with fluid. To do so will result in an improper connection.

    Surgical Diagnostics:

    Some tests should be performed during implantation.

    These tests help to make sure that the transducer and

    microphone are fully functional and will help detectif the device has been damaged, e.g. during handling

    or in the OR during implantation. The data collected

    can be used later in the clinic for diagnostics and

    troubleshooting.

    These tests are:

    Battery test (reads charge state)

    SAFI tests:Microphone functionality

    Functional Test

    Transducer Impedance

    CAUTION!

    For the surgical tests, the TLA programming coil

    must be wrapped in a sterile bag to be placed

    over the implant coil.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    40/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 40 of 69Otologics Confidential

    Step 5: Enlarging the Bone Bed for the Electronics Capsule,

    Pendant Microphone, and Closing

    The size of the bone bed is increased and a new bed

    is created to accommodate the fully-implantable

    electronics capsule and microphone. The capsuleshould be placed in a relatively flat location, and the

    bone bed should be drilled under the entire capsule.

    The bone bed should be enlarged to accommodatethe implant with the silicone bands installed. (Figure

    57, 58, 59). The microphone should not be placed

    under or adjacent to muscle tissue, such as thetemporalis muscle, as the localized tension created

    by the muscle will affect the performance of the

    microphone.

    When inserting the capsule in to the bone bed, ensure

    that the coil is not severely bent. Secure the fully-implantable capsule, microphone, and close the

    wound in the normal manner.

    Figure 57 MET Semi-Implantable Ossicular

    Stimulator bone bed

    Figure 58 Bone bed enlarged and microphone

    bed created for MET Fully-Implantable OssicularStimulator

    Figure 59 MET Fully-Implantable Ossicular

    Stimulator in situ

    Important Information!

    Integrated straps are provided for securing theelectronics capsule and the microphone pendant to

    the bone bed. Additional bending and fitting may be

    required. The straps should constrain the implant or

    the microphone pendant in a fixed location followingthe installation of the bone screws. For this reason,

    two screws must be used in each strap. The

    devices should be tested for movement following the

    installation of the bone screws. If movement is

    present additional adjustment to the straps must beperformed. Additional bending and shaping of the

    straps may be performed using the surgical pliers

    used to form the bone-mounting bracket.Precautions for bending are described in detail in the

    Instructions for Use included with the straps.

    Important Information!

    The bone bed for the microphone may be done one

    of two (2) ways (Figure 34). (1) At a minimum, a

    circular depression may be made to accommodatethe raised portion on the under or back-side of the

    microphone, or (2) A larger area to fit the shape forthe entire device may be drilled. A lead channel

    should be drilled for the microphone lead in

    either case. Also, a lead channel should be drilled

    to accommodate the transducer lead.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    41/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 M

    Surgical Training Manual Page 41 of 69Otologics Confidential

    Step 6: Post Surgical Care and Activation

    Post surgery care:

    Follow up patient care is performed in the usual manner for otologic surgery. The patient is given the same

    postoperative instructions provided after standard otologic surgery. If the patient was wearing amplification priorto surgery, he or she may wear amplification in the non-implant ear, but should not wear amplification in theimplant ear until after the eight week healing period is complete and medical clearance is given.

    Activation:

    Following a 8-week healing period and medical clearance, the MET Fully-Implantable Ossicular Stimulator will

    be programmed according to the hearing needs of the individual patient. Detailed procedures for fitting the MET

    Fully-Implantable Ossicular Stimulator are provided in the MET Fully-Implantable Ossicular Stimulator Fitting

    Guide.

  • 7/27/2019 D104096 Rev M Surgical Training Manual PDF

    42/69

    ____________________________________________________________________________________

    MET Fully-Implantable Ossicular Stimulator D104096 MSurgical Training Manual Page 42 of 69

    Otologics Confidential

    SURGICAL STEPS FOR REPLACING THE FULLY-

    IMPLANTABLE MET OSSICULAR STIMULATOR (FIMOS)

    WITH THE FULLY-IMPLANTABLE MET OSSICULAR

    STIMULATOR (FIMOS)

    Step 1: