Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan
Children Hospital Srebrnjak Zagreb, Croatia
Ruđer Bošković Institute, Zagreb, Croatia
COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN
DIAGNOSTIC RADIOLOGY
![Page 2: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/2.jpg)
Our wish is that all children are safe and protected in radiology department!
![Page 3: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/3.jpg)
INTRODUCTION
Pulmonary X-rays are essential in the diagnostics of lung diseases of children and youth.
Chest radiography represents the majority of radiological examinations.
The starting basis for radiation protection is the exact determination of doses.
![Page 4: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/4.jpg)
FEATURES:
very low doses at low and variable energies have to be measured
there exists a considerable variation in radiation doses delivered to patients (different X-ray equipment, different staff, etc.).
![Page 5: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/5.jpg)
AIM
to test a new Shimadzu X-ray unit used for thorax examination of children
to compare a thermoluminescence (TL) dosimetry system based on LiF: Mg,Cu,P with the radiophotoluminescent (RPL) glass dosimetry system (FGD-200).
![Page 6: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/6.jpg)
MATERIALS AND METHODSIrradiations:
137Cs gamma rays in air – for calibration
ISOVOLT 420 X Ray Unit (40-300 kV, 1-20 mA) at the SSDL in air and on the water phantom (plastic bottle, Φ=11 cm, V=2.5 l) – energy dependence
Energies: 33, 48, 65 keV
Dnom= 2 mGy (air kerma)
150 kV Shimadzu CH-200M unit in air
Phantoms:
water phantom (plastic bottle)
doll phantom
![Page 7: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/7.jpg)
BABY PHANTOM
![Page 8: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/8.jpg)
Voltage: 70 kVQuantity of charge: 1.6 mAsTime of irradiation: 5 msSize of the focus: 0.6 mmDistance: 150 cm
![Page 9: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/9.jpg)
BABY FIX
![Page 10: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/10.jpg)
![Page 11: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/11.jpg)
Dosimetry systems
![Page 12: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/12.jpg)
RESULTS
Energy dependence in SSDL
Doses in diagnostic X-ray unit
![Page 13: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/13.jpg)
Energy dependence in SSDL
Relative dose in air: the mean values of doses measured (Dmeasured) “in air” relative to delivered doses specified as “air-kerma free-in-air” (Ka).On phantom: the mean values of the doses measured on the phantom relative to delivered doses specified as air kerma free-in-air
![Page 14: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/14.jpg)
The energy dependence of TL and RPL dosimeters in SSDL
0.00.2
0.40.6
0.81.0
1.21.4
0 20 40 60 80
Mean photon energy (keV)
Re
lativ
e d
ose
0
0.5
1
1.5
2
0 20 40 60 80
Mean photon energy (keV)
Re
lativ
e d
ose
In air On phantom
: TLD : RPLD : TLD : RPLD
● : Calculated values of Hp(10)/Ka
![Page 15: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/15.jpg)
Mean value and standard deviation (SD) of doses measured on “phantoms “ in
diagnostic unitPhantom Doll (unknown plastic) Bottle (water)
Dosimeter RPL TL RPL TL
Place of dosimeter
Back Sternum Back Sternum Entr. Exit Entr. Exit
Mean dose (mGy) 0.040 0.019 0.049 0.022 0.041 0.004 0.030 0.002
SD (mGy) 0.002 0.006 0.008 0.007 0.002 0.001 0.006 0.004
SD(%) 4.3 30.6 15.5 30.0 3.9 28.2 19.5 154.9
Dentrance/Dexit 2.5 2.3 11.3 13.0
Entr.: Entrance
![Page 16: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/16.jpg)
Doses in diagnostic X-ray unitOn the doll: agreement of the dose values of RPL and TL dosimeters in entrance and exit beams On the water phantom: difference between the mean values measured in the ingoing beam with the two dosimeters
Reason: ▪ different materials of phantoms ▪ different energy absorption characteristics of the two dosimeters (below 50 keV) on water phantom
![Page 17: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/17.jpg)
CONCLUSION
TLD (LiF:Mg, Cu, P )(termoluminiscent dosimeter)
RPL(radiophotoluminiscent glass dosimeter)
![Page 18: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/18.jpg)
TLD (LiF: Mg, Cu, P ) (termoluminiscent dosimeter)
High sensitivity
In spite of its anomalous energy dependence nearly tissue- equivalence
Agreement with earlier results
![Page 19: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/19.jpg)
RPL (radiophotoluminiscent glass dosimeter)
Higher sensitivityEnergy dependance “in air” is better than for LiF:Mg, Cu, P (33-65 keV mean energies)Energy dependance curve on the water phantom changes in opposite direction than the calculated Hp(10) values
The absolute difference from Hp(10) is not larger than for LiF:Mg, Cu, P
![Page 20: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/20.jpg)
CONCLUSION
The measured dose values in X-ray diagnostic unit are in accordance with the characteristics found in SSDL for both dosimeters.
The RPL system seems to be suitable for dosimetric measurements in X- ray diagnostics.
![Page 21: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/21.jpg)
ZAGREB
![Page 22: Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia](https://reader036.vdocuments.site/reader036/viewer/2022062308/56649f075503460f94c1cc54/html5/thumbnails/22.jpg)
Thank you for your
attention!
The authors are grateful to Chiyoda Technol Corporation, Japan for the support of this work.