CTM : PROVING
By: Vinay Varghese
![Page 2: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/2.jpg)
INDEX• CTM• FLOW METERS– Coriolis flow meter– Ultrasonic Flow meter– Turbine Flow meter
• Proving– Why proving ?– Basic principle– M-Factor– K-Factor
![Page 3: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/3.jpg)
INDEX• Pulse Interpolation (Double Chronometry method)• Topics that can be discussed.
– Water draw.– Leak test.– Working principle of the Compact prover.– Daily interpolation.
![Page 4: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/4.jpg)
Custody Transfer Metering• We use custody transfer metering to measure how much
product do we supply to the customers and in the end to bill our clients for the same.
• There are three major components of the CTM.1. Primary Devices: Flow measurement devices2. Secondary Devices: Pressure, Temperature, BS&W, Density
measuring devices3. Tertiary Devices: Flow Computers
• The Primary devices in the CTM are Flow meters and it is important to have an understanding of the working of the various type of flow meters normally used in Custody Transfer.
![Page 5: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/5.jpg)
Flow Meters• There are different types of flow meters that are used in the
CTM.• Mainly there are three:
1. Coriolis Flow meters2. Ultrasonic (Transit Time Based) Flow meters3. Turbine meters
• All three type of flow meters work on different principles.
![Page 6: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/6.jpg)
CORIOLIS FLOW METER
Flow is measured by using velocity sensors to detect the twist in the tube and transmit electrical signals having a relative phase shift that is proportional to mass flow.
Coriolis meters also measure density, whereby the resonant frequency of the forced rotation as a function of fluid density.
When the fluid is flowing, it is led through two parallel tubes. An actuator (not shown) induces a vibration of the tubes. The two parallel tubes are counter-vibrating, to make the measuring device less sensitive to outside vibrations. The actual frequency of the vibration depends on the size of the mass flow meter, and ranges from 80 to 1000 vibrations per second.
When no fluid is flowing, the vibration of the two tubes is symmetrical.
![Page 7: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/7.jpg)
ULTRASONIC FLOW METERS• Types of ultrasonic flowmeters
1. Transit time method2. Doppler method3. Pulse-Doppler Method4. Phase difference method
• Transit times are measured from sound pluses travelling diagonally across the pipe, downstream with the flow and upstream against the liquid flow
• The difference in these transit times is related to the average liquid flow velocity along multiple acoustic paths.
• Numerical calculation techniques are then used to compute the average axial liquid flow velocity and the liquid volume flow rate at line conditions through the meter.
![Page 8: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/8.jpg)
MEASURING FLOW USING USMt21,12 = Lw and Lw
c + va cos(ᶲ) c - va cos(ᶲ)• t21 is the time taken for the sound pulse to travel in the upstream
direction.• t12 is the time taken for the sound pulse to travel in the downstream
direction.• C is the velocity of sound.• V is the flow velocity.• ᶲ is the angle at which the transducer and receiver are positioned with
respect to the pipe.• L is the distance between the two transducers.• Since we have the time, we can compute the velocity of the fluid.
![Page 9: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/9.jpg)
• There are 4 sets of transducers and receivers in the LUSM.• Each set is covering a particular region of the pipe and so, we cannot give
equal weightage to the flow measured by each set.
![Page 10: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/10.jpg)
DATA CALCULATION SUMMARY
![Page 11: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/11.jpg)
TURBINE FLOW METERS• When fluid flows past a turbine meter, the meter's vaned
rotor rotates• The angular velocity of the rotor is proportional to the
velocity of the fluid. • A magnetic pickup coil is mounted outside the pipe. The
turbine or paddlewheel blades are made of or contain magnetic material. As the fluid flows and each blade passes by the coil, the magnetic flux through the coil changes, causing a voltage pulse. A sensor measures the pulse rate and from that determines the flowrate.
![Page 12: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/12.jpg)
WHY DO WE NEEED PROVING ?• What is a Meter Proving
– Meter proving is the physical test used to determine the accuracy and performance of a liquid meter
• Why is Proving performed– The need for proving arises because operating conditions differ
significantly from the conditions under which the meter is calibrated• Why Prove
– Establish reference point under operating conditions.– Account for changes in meter accuracy– Verify repeatability / linearity of meter– Contractual requirement for custody transfer
![Page 13: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/13.jpg)
BASIC PROVING PRINCIPLE• Regardless of the prover type, the same• Basic proving principle applies• Meter Factor = Known_Volume Measured Volume
![Page 14: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/14.jpg)
SMALL VOLUME METER FACTOR• API Definition:• Small volume provers are provers that will collect less than
10,000 pulses in a single pass.• We need 10,000 pulses per pass is necessary to
mathematically generate a meter factor accurately up to 4 decimal points.
• The small volume prover method is possible because of the precision of optical switches and the method of double chronometry pulse interpolation that is used in small volume provers.
![Page 15: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/15.jpg)
K-FACTOR AND M-FACTOR• The operational point of view for proving is simple enough:
– We take the flow in a desired stream through the prover take-off valve into the prover.
– See to it that the flow remains steady with minimal deviation from the flow at which we want to check the accuracy of the flow meter.
– Get the repeatability of the flow meter within the desired range and feed the K and M factors computed by the software to the system and enjoy another three months of peaceful operations.
• What really is important while understanding proving is how this K and M factor is derived for the flow meters.
![Page 16: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/16.jpg)
M- FACTOR• M- Factor is pretty straightforward.• In this process the volume registered by the flow meter being
tested is compared to the volume registered with the flow meter which we consider as a reference(Master Meter), giving us the formula of
Multiplied by
To correct for the pressure and temperature variation from the basetemperature and pressure ( 15oC and 1 bar)
![Page 17: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/17.jpg)
• How do we know that the flow measured by the reference meter is accurate ?
• For that we need to prove the master meter with reference to the compact prover which is essentially the most complicated part, both operation wise and in terms of understanding.
• Compact prover is of a known volume established by the water draw test.
• Essentially water draw test is a physical measurement of the volume of water displaced in one pass of the prover in the downstream stroke.
• We get to know that one pass of the prover is complete because of the optical switches installed along the piston movement of the prover.
![Page 18: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/18.jpg)
K-FACTOR
• These optical switches are going to come in use soon, when we establish the K-factor for the reference meter.(Master Meter)
• As we have seen earlier that the turbine type of flow meter works on the principle of cutting the magnetic field with each rotation of the blades, w.r.t the flow through it.
• The master meter is also a turbine meter, when the magnetic field is cut it generates a pulse as a signal of a unit volume that has passed through the flow meter.
![Page 19: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/19.jpg)
Contd.• To set the master meter as a reference for the other flow
meters to be tested we need to know the no. of pulses it generates for a particular volume displaced.
• This we achieve by double chronometry pulse interpolation while proving the master meter w.r.t the compact prover.
• For this pulse interpolation method we need:– Time taken to displace voume in the compact prover.– No. of pulses generated by the flow meter during the pass.– Time taken for the generation of pulses in the flow meter.– Base volume of the compact prover.
![Page 20: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/20.jpg)
![Page 21: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/21.jpg)
![Page 22: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/22.jpg)
![Page 23: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/23.jpg)
Since we need to find out the exact no. of pulses generated in that pass, we use double chronometry pulse interpolation.
![Page 24: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/24.jpg)
![Page 25: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/25.jpg)
![Page 26: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/26.jpg)
TOPICS THAT WE CAN DISCUSS
• Water draw.• Leak test.• Working principle of the Compact prover.• Daily interpolation.
![Page 27: CTM PROVING](https://reader036.vdocuments.site/reader036/viewer/2022062503/587f53a31a28ab0d378b5bf3/html5/thumbnails/27.jpg)
• Proving is a very elaborate topic. What I have covered is something that I found interesting in the process of proving, and thought to be apt for this short session.