CT-SCAN
Dr.H.Sigit Widodo,Sp.RBagian RadiologiFakultas Kedokteran Universitas Trisakti
![Page 2: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/2.jpg)
Computed Tomography (CT)Scan
Basis Principle
The Interval Structure Of Any 3 Dimensional Subject Can be Reconstructed from many different Projections / Views of that subject
![Page 3: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/3.jpg)
Minimal diperlukan 180 Projections yang berbeda untuk memperoleh Radiograph yang bernilai Diagnostik (X-Ray Beam Were Directed Through The Tissue at 1° Interval Until 180 Readings were made)
![Page 4: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/4.jpg)
Fungsi Scan Unit
Collecting Large Amount of Transmission Data for Processing By the Computer
![Page 5: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/5.jpg)
![Page 6: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/6.jpg)
![Page 7: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/7.jpg)
CT-Scan Generasi -1
EMI Mark-I 80x80 Pixel Resolution (th 1972)
1 Slice 4.5 Min Scan time 1.5 Min Reconstruction time
Karena Long Acquisition Time & Cardiac & Respiration Motion
Semula CT-Scan hanya untuk Head Scan Rotate / translate, pencil Beam hanya 2
(dua) detector, Keuntungan : Scatter ↓ (Pencil Beam )
![Page 8: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/8.jpg)
CT-Scan ,4 bagian Utama :
1.Scan Unit
a.Patient Table / Couch
b.Gantry Central Opening Gantry Aperture
*Ada X-Ray Tube Radiation Detector Array
![Page 9: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/9.jpg)
2.Processing Unit :
Computer takes raw data and convert it into picture form
Computer yang membedakan CT dari Conventional Radiography
![Page 10: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/10.jpg)
3.Display unit / Direct Display Console Gray-Scale Television Screen Enlargement Contrast Scale Manipulatorry Substraction Operator Console
![Page 11: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/11.jpg)
4.Storage System Main Memory , Floppy Disc. Magnetic
Tape Radiograph / Polaroid Film
![Page 12: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/12.jpg)
CT-SCAN GENERASI 2
Rotate / translate / Narrow Fan Beam Detector 30 X-Ray : 30x D.P.I Scan Time / 1 Slice : 18 Sec
(15 x > cepat D.P 1 )
![Page 13: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/13.jpg)
CT-SCAN GENERASI 3
Rotate / translate / Wide Fan Beam Detector > 800
Joined X-Ray tube dan detector rotate together around the patient without translation Scan time 0.5 – 5 Sec
Problem : Ring artefacts
![Page 14: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/14.jpg)
The Scope of Diagnostic Imaging
![Page 15: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/15.jpg)
![Page 16: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/16.jpg)
Scanner Principles Sejak CT generasi 1(th 1972), Data Gathering
has Progressed through 4 (four) generations. Generasi 1 perlu 4 ½ Minute per slice Generasi berikutnya mempersingkat Scan
Time CT generasi 4 : Scan Time per slice 2- 10
second Ada ≥ 600 Detector melingkari pasien dalam
Gantry 1 (Satu) X-Ray tube berotasi 360º Data
Collection
![Page 17: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/17.jpg)
CT-SCAN GENERASI 4 (AKHIR 70)
Rotate/ Stationary Meniadakan Problem Ring Artefact
4800 Detector Removed From Rotating Gantry Placed in Stationary 360º Ring Around the patient
![Page 18: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/18.jpg)
CT-SCAN GENERASI 5 Stationary Khusus untuk jantung Tidak digunakan Conventional X-Ray
Tube ART Of Tungsten Encircles The Patient,opposite to detector Ring
High Energy Electron Beam Strike Tungsten X-Ray Scan Time 50 MSec Dapat CT-Movies of Beating Heart
disebut juga Electron Beam Scanner
![Page 19: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/19.jpg)
CT-SCAN GENERASI 6 (Helical Spiral)
The X-Ray Tube Rotates around the patient in a helical pattern, while the patient and the table are translated through the gantry
Entire scan time can be performed within a single breth hold (30 second for entire abdomen)
![Page 20: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/20.jpg)
CT-SCAN GENERASI 7
Multiple Detector Array
![Page 21: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/21.jpg)
![Page 22: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/22.jpg)
DETECTOR
1.Xenon (Gas) High Pressere 25 ATM Position Fixed to X-Ray Source
Tidak bisa pada generasi 4 Ionization Detector Electronic
Signal Amplified & Digitized
![Page 23: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/23.jpg)
2.Solid Detector Crystalline Dielectric CdWO4 /
Yurium / Gadolinium Ceramic X-Ray Scintillator Light
Photo Detector Electrical Signal Dipakai pada generasi 4
![Page 24: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/24.jpg)
CT-Scan Cranium
Irisan Sejajr OML (Orbito Meatal Line)
Dari Foramen Magnum s/d Vertex Ketebalan Irisan 5-10 mm Polos – Kontras Window : 1.Otak 2.Tulang
![Page 25: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/25.jpg)
OML = Reids
![Page 26: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/26.jpg)
Deskripsi = Evaluasi
I.Differensiasi antara White Matter (Substansia Alba) dengan Grey Matter (Substansia Grisea)
II.1.Brain Stem (Med.Oblongata, Pons) dan Cerebellum2.Cerebral Hemisphere (Lobus Frontalis, Temporalis, Parietalis Occipitalis)3.Basal ganglia, Thalamus
Rongga Cairan Cerebrospinal Sistem ventrikel Cisterna, Fissura, SulciIII.Tulang - tulang
![Page 27: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/27.jpg)
CT-Scan Abdomen
Potongan Axial = Transversal Whole Abdomen : Proc.Xyphoideus
s/d Symphisis Pubis Abdomen Atas : Proc.Xyphoideus s/d
Spina Iliaca Superior (Ginjal) Abdomen Bawah (Pelvis) : L-5 s/d
Buli-buli (Prostat)
![Page 28: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/28.jpg)
Puasa : 8 jam Media Kontras :1. BaSO4 1 jam . ½ jam 5- 10 menit sebelumnya @300-400
cc (Gaster-Usus Halus) 2- 3 jam
sebelumnya 400-600 cc(Colon Proximal)
Enema 200 – 300 cc (Colon Distal)2.Intra Venous
![Page 29: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/29.jpg)
Bone +1000
+400
Hematoma +100Grey MatterWhite MatterOedemaCSFWater 0Fat Tissue
- 100
AIR - 1000
![Page 30: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/30.jpg)
CT-Scan Thorax
Potongan Axial dari Apex PAru s/d Diapgragma langsung dengan kontras I.V
Tebal 5-10 mm Dicetak dalam 2 window level
paru dan mediastinum
![Page 31: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/31.jpg)
Evaluasi :1.Pulmo Parenchyme : Nodule / Mass Bronchus : Terbuka A.V Pulmonalis : Tapering Hilus :KGB Pleura
![Page 32: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/32.jpg)
2.Mediastinum :KGB Pengelompokkan/ Pembesaran
3.Cor, Aorta Thoracalis, V.Cava Superior
4.Tulang dinding thorax (Vert.Thoracalis, Costae, Sternum)
![Page 33: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/33.jpg)
Kelebihan CT-Scan
1. Cross-Section Imaging,Over Laping(-)2. Contrast Resulotiun baik.Dapat bedakan
Soft Tissue(Grey Matter dari White Matter,Brain)
3. Geometric Accuracy,Distortion Free dapat mengukur tepat : Jarak,diameter, tebal, displacement.Digunakan untuk rencana operasi / Radioterapi
4. Tissue characterization Menggunakan CT Number Soft Tissue Mengukur Bone Mineral Content
![Page 34: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/34.jpg)
Kekurangan CT-Scan
1.Geometric Miss GM Contrast Miss CM Nodule yang luput antara 2 (dua) slice :
GM
Window widh yang lebar (contrast Kurang) Window Widh yang sempit pada window
level yang salah : CM
![Page 35: CT-SCAN DAN USG](https://reader036.vdocuments.site/reader036/viewer/2022062308/55cf8ede550346703b9674ab/html5/thumbnails/35.jpg)
2.Artefact Partial volume avera ging Geometric oversight High Density oversight High Density Artefact Motion Artefact