Criterion Series TE 3.0 CO Analyzer Copyright © 2014 Tisch International All Rights Reserved Consolidated Analytical Systems 4014 Spring Grove Ave Cincinnati OH 45223 513-542-1200 [email protected] www.tischinternational.com Tisch Environmental (TE) is the benchmark for continuous sampling and monitoring of criteria pollutants including gases, particulates, metals, volatiles, and specialty monitoring equipment. Since the company's inception in 1953 as General Metal Works, the TE product line has expanded from the first high volume air sampler to include continuous gas and particulate analyzers. The Criterion TE 3.0 Carbon Monoxide analyzer delivers precise and reliable performance at excellent value. It uses proven NDIR gas filter correlation technology to measure CO in ambient air (LDL <40 ppb, range 0-200ppm). Bluetooth connectivity allows for easy programming, troubleshooting, and connectivity. Allowing the operator to maintain, download and email data without touching the instrument. Approvals: US EPA approval (RFCA–0509–174). www.cas-en.com Guaranteed performance with warranty. Comprehensive data logging and remote viewing of over 200 operational parameters. Versatile interfacing through RS232, USB, Bluetooth, analogue and TCP/IP. Removable flash memory stores 10 years of data, including up to 12 individual parameters and event logs. Operational parameters can also be transferred to memory for easy retrieval. Latest firmware updates can easily be installed using the USB flash memory drive. Minimizes time spent on site performing maintenance through superior remote instrument control, diagnostic viewing, and calibration. Programming, viewing, downloading and emailing of data is made simpler using bluetooth connectivity. Enhanced operator safety through the use of 12VDC internal voltages. Reduced cost of spare parts, accessories and consumable items through extensive use of standard components across the Criterion series. Advanced rack slide design makes accessing internal components and removing the analyzer from a rack cabinet easy. Low power internal pump. System power usage can be reduced due to the instrument’s low power demand and its ability to operate over a wide temperature range.

Upload: others

Post on 03-Apr-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Criterion Series - Tisch Environmental...US EPA approval (RFCA–0509–174). Guaranteed performance with warranty. Comprehensive data logging and remote viewing of over 200 operational

Criterion Series

TE 3.0 CO Analyzer

Copyright © 2014 Tisch International All Rights Reserved

Consolidated Analytical Systems4014 Spring Grove Ave

Cincinnati OH 45223513-542-1200

[email protected]

www.tischinternational.com

Tisch Environmental (TE) is the benchmark for continuous sampling and monitoring of criteria pollutants including gases, particulates, metals, volatiles, and specialty monitoring equipment. Since the company's inception in 1953 as General Metal Works, the TE product line has expanded from the first high volume air sampler to include continuous gas and particulate analyzers.

The Criterion TE 3.0 Carbon Monoxide analyzer delivers precise and reliable performance at excellent value. It uses proven NDIR gas filter correlation technology to measure CO in ambient air (LDL <40 ppb, range 0-200ppm). Bluetooth connectivity allows for easy programming, troubleshooting, and connectivity. Allowing the operator to maintain, download and email data without touching the instrument.

Approvals:

US EPA approval (RFCA–0509–174).

www.cas-en.com

Guaranteed performance with warranty.

Comprehensive data logging and remote viewing of over 200 operational parameters.

Versatile interfacing through RS232, USB, Bluetooth, analogue and TCP/IP.

Removable flash memory stores 10 years of data, including up to 12 individual parameters and event logs. Operationalparameters can also be transferred to memory for easy retrieval.

Latest firmware updates can easily be installed using the USB flash memory drive.

Minimizes time spent on site performing maintenance through superior remote instrument control, diagnostic viewing, and calibration.

Programming, viewing, downloading and emailing of data is made simpler using bluetooth connectivity.

Enhanced operator safety through the use of 12VDC internal voltages.

Reduced cost of spare parts, accessories and consumable items through extensive use of standard components across the Criterion series.

Advanced rack slide design makes accessing internalcomponents and removing the analyzer from a rack cabinet easy.

Low power internal pump.

System power usage can be reduced due to the instrument’s low power demand and its ability to operate over a widetemperature range.

Page 2: Criterion Series - Tisch Environmental...US EPA approval (RFCA–0509–174). Guaranteed performance with warranty. Comprehensive data logging and remote viewing of over 200 operational

Continuous CO Analyzer

Copyright © 2014 Tisch International All Rights Reserved

Criterion SeriesSpecification Sheet

Criterion TE 3.0 Specifications

Criterion TE 3.0 CO Analyzer:

Concentration Display: 3 3

User selectable mg/m , µg/m , ppm, ppb, pptRanges: Autoranging 0-200ppm (Analog ranges 0-50ppb to 0-200ppm)Noise: <0.02ppm or 0.2% of concentration reading whichever is greaterLower Detectable Limit: <0.04ppm or 0.2% of readin g whichever is greater Linearity: <1% of full scale (0-50ppm)Precision: 0.1ppm or 0.1% of reading whichever is greaterZero Drift: 24 hours: <0.1ppm 7 Days: <0.1ppmSpan Drift: 7 Days: <0.5% of readingResponse Time: 60 seconds to 95%STP Reference: 0ºC, 20ºC, 25ºC, at 101.3 kPaSample Flow Rate: 1000cc/min

Physical Specifications

Operating Temperature Range: 0-40ºCDimensions (H x W x D): 3.5 RU 6.89” (175mm) x 16.89” (429mm) x 25.19” 638mm)Net Weight: 39.2lbs (17.8kg)

Contact Information

Tisch Environmental145 South Miami AvenueVillage of Cleves, OH [email protected]

www.tischinternational.com

www.tisch-env.com

Power Requirements:99-132 VAC, 198-264 VAC 47-63Hz260VA (Max Startup)150VA (Nominal)

Electrical Specifications

USB Port (Digital communication)Bluetooth (Digital communication)TCP/IP Ethernet network connectionRS232 Port 1: Digital communicationRS232 Port 2: Multidrop portProtocols: MODBUS RTU/TCP, Bavarian, Advanced

Communication Specifications

Menu selectable current or voltage output 0-20mA, 4-20mA,or 0-5VDC (optional 0-10VDC)Menu selectable offset of 0%, 5%, or 10%Auto-ranging from 0-50ppb to 0-20ppm (between two user specified full scale values)3 scalable analog inputs, 0-5v, 160µV resolution8 logic level digital status inputs/8 open collector digital outputs

Inputs/Outputs: 25 pin I/O port

Options

External AC powered pump (capable of operating 3 analyzers, excluding NO )X

Internal long life, low power maintenance free pump (Max 6 watts)12VDC Power supply, dual filter (particulate)High pressure zero/span valvesPrinted manual (Electronic manual supplied with analyzer)

Data Logging (Removable USB Memory)