conductr: an r program to measure stem, leaf or root ...€¦ · conductr: an r program to measure...
TRANSCRIPT
conductR: an R program to measure stem, leaf or root hydraulic conductance
with an electronic balance or pipette
Duncan D. Smith
Dec 2015 last updated September 2016 for v.12
this document is best viewed in full screen / presentation mode
program available from http://botany.wisc.edu/mcculloh-lab-methods.htm
send feedback to [email protected]
Setting up balance communications………………………. Initiating/loading the program……………………………… Measuring stems or roots with a balance…………………. File management (what the program expects)…………… Saved data format…………………………………………… Plotting vulnerability…………………………………………. Measuring stems or roots with a pipette………………….. More file management (using multiple methods)………… Measuring leaves with a balance or pipette……………… Input errors…………………………………………………… Appendix A: all functions and their parameters………….. Appendix B: known issues………………………………….. Appendix C: anticipated questions…………………………
3 8 15 16 28 47 49 58 66 73 74 78 79
2
Outline - how to use the program and potentially useful details
Setting up balance communications
You need
•Balance (we use a Sartorius with 4-point precision)•USB cable for balance (note: Sartorius cables are wired differently from others) •Driver for cable - may not be necessary for newer operating systems
•e.g. http://www.ftdichip.com/Drivers/VCP.htm for a Sartorius cable•e.g. http://www.prolific.tw for some other cables
•R (https://cran.r-project.org/)•tcltk (a package that comes with the standard R installation)•X11 or XQuartz (for Macs only - this runs tcltk; http://www.xquartz.org/)•R Studio (https://www.rstudio.com/; optional but provides a better interface on *PC* computers than standard R. Standard R is still needed)
3
Setting up balance communications
Communication settings
• Determine port settings. Your balance manual should specify how to find these and change them. Here, I will assume:
• baud=9600• parity=odd• data bits=7• stop bits=1• handshake=software/xonxoff
• Determine port name. This will be COM# on a PC or /dev/cu.usbserial-# on a Mac, where # will vary between ports and computers
• To find #, turn on the balance and connect it to the computer. For a PC, find # in Device Manager —> Ports. For a Mac, open Terminal (in the Utilities folder) and input: ls -l /dev/cu.* For either system, there may be multiple ports so disconnect and reconnect the balance to determine the correct one. On my Mac, #=00001004, which I’ll use in the following examples
4
Setting up balance communications
Try basic communication
• Open R• Load the tcltk package by typing “require(tcltk)” in the R console.• Input the following into the R console, replacing the red text with your own port ID (in
the first line) and port settings (in the second line):.Tcl('set serial [open /dev/cu.usbserial-00001004 r+]' ) .Tcl('fconfigure $serial -mode \"9600,o,7,1\" -blocking 0 -handshake xonxoff')
• Now press 'print' on the balance and then in R input:.Tcl('return [read $serial]') .Tcl('close $serial')
• The weight on the balance should appear in R. If not, check port settings
5
Setting up balance communications
Let the program tell the balance to print
• We need a print command which is Esc-P-CR-LF for Sartorius balances (CR=carriage return and LF=line feed). In the proper format, this command is \x1B\x50\xD\xA (see http://www.asciitable.com/). For Mettler balances, the command is S-I-CR-LF (\x53\x49\xD\xA)
• Now, run the following lines in R - replacing the red text with your port ID (line 1), port settings (line 2), and print code (line 3):.Tcl('set serial [open /dev/cu.usbserial-00001004 r+]' ) .Tcl('fconfigure $serial -mode \"9600,o,7,1\" -blocking 0 -handshake xonxoff') .Tcl('puts $serial \x1B\x50\xD\xA') .Tcl('flush $serial') Sys.sleep(0.1) .Tcl('return [read $serial]') .Tcl('close $serial')
6
Setting up balance communications
Let the program tell the balance to print (continued)
• R should return the weight on the balance. Note that the flush command forces the print command to send and Sys.sleep(0.1) waits 0.1 s, which should be long enough for a response. If it does not work, I would try a simpler command like ‘tare’ (Esc-T-CR-LF = \x1B\x54\xD\xA) because it will be clear whether the balance received the command or not (i.e. the balance should tare). You might also increase the Sys.sleep() time.
Now to use the program
7
>>>>>>>>>>>>>>
1 ########################### conductR #####################################2 ##########################################################################3 #### an R program to measure plant hydraulic conductance with an #########4 #### electronic balance or pipette. #####################################5 ##########################################################################6 #### Written by Duncan D. Smith 2012-2015#################################7 #### free to use and alter for all non-commercial purposes ##############8 #### Conceptually based on John Sperry's conduct.ver1.xls ################ 9 ##########################################################################10 require(tcltk) #load tcltk ##############11 ##########################################################################1213 ##Computer-specific settings - *three* things ############################14 kwd="/users/duncan/desktop" #working directory ###########15 .Tcl('set port /dev/cu.usbserial-00001004') #port number ############16 # source("/users/duncan/documents/conductR/conductR_12.R") ###############17 #run the above line (without #) to 'load' the program18 ##########################################################################1920 ##Communication/Plot Settings#############################################21 .Tcl('set baud 9600') #baud rate ###############22 .Tcl('set prty o') #parity ##################23 .Tcl('set dbit 7') #data bits ###############24 .Tcl('set sbit 1') #stop bits ###############25 .Tcl('set prnt \x1B\x50\xD\xA') #print code (Esc-P-CR-LF) ################26 par(mar=c(5,5,1,1)) #plot margins etc ###########27 ##########################################################################28 29 ##Physical Settings#######################################################30 Tsoln.C=25 #solution temp (C) - will standardize to 20 ##############31 elev=280 #elevation (masl) ##############################32 lat=43 #latitude (deg north) ##############################
Initial setup
opening conductR_X opens the console and script
8
>>>>>>>>>>>>>>
1 ########################### conductR #####################################2 ##########################################################################3 #### an R program to measure plant hydraulic conductance with an #########4 #### electronic balance or pipette. #####################################5 ##########################################################################6 #### Written by Duncan D. Smith 2012-2015#################################7 #### free to use and alter for all non-commercial purposes ##############8 #### Conceptually based on John Sperry's conduct.ver1.xls ################ 9 ##########################################################################10 require(tcltk) #load tcltk ##############11 ##########################################################################1213 ##Computer-specific settings - *three* things ############################14 kwd="/users/duncan/desktop" #working directory ###########15 .Tcl('set port /dev/cu.usbserial-00001004') #port number ############16 # source("/users/duncan/documents/conductR/conductR_12.R") ###############17 #run the above line (without #) to 'load' the program18 ##########################################################################1920 ##Communication/Plot Settings#############################################21 .Tcl('set baud 9600') #baud rate ###############22 .Tcl('set prty o') #parity ##################23 .Tcl('set dbit 7') #data bits ###############24 .Tcl('set sbit 1') #stop bits ###############25 .Tcl('set prnt \x1B\x50\xD\xA') #print code (Esc-P-CR-LF) ################26 par(mar=c(5,5,1,1)) #plot margins etc ###########27 ##########################################################################28 29 ##Physical Settings#######################################################30 Tsoln.C=25 #solution temp (C) - will standardize to 20 ##############31 elev=280 #elevation (masl) ##############################32 lat=43 #latitude (deg north) ##############################
9
Initial setup
set working directory (line 14), port number
(line 15), and the location of the conductR
file (line 16)
Do these whenever setting up a new
computer
>>>>>>>>>>>>>>
1 ########################### conductR #####################################2 ##########################################################################3 #### an R program to measure plant hydraulic conductance with an #########4 #### electronic balance or pipette. #####################################5 ##########################################################################6 #### Written by Duncan D. Smith 2012-2015#################################7 #### free to use and alter for all non-commercial purposes ##############8 #### Conceptually based on John Sperry's conduct.ver1.xls ################ 9 ##########################################################################10 require(tcltk) #load tcltk ##############11 ##########################################################################1213 ##Computer-specific settings - *three* things ############################14 kwd="/users/duncan/desktop" #working directory ###########15 .Tcl('set port /dev/cu.usbserial-00001004') #port number ############16 # source("/users/duncan/documents/conductR/conductR_12.R") ###############17 #run the above line (without #) to 'load' the program18 ##########################################################################1920 ##Communication/Plot Settings#############################################21 .Tcl('set baud 9600') #baud rate ###############22 .Tcl('set prty o') #parity ##################23 .Tcl('set dbit 7') #data bits ###############24 .Tcl('set sbit 1') #stop bits ###############25 .Tcl('set prnt \x1B\x50\xD\xA') #print code (Esc-P-CR-LF) ################26 par(mar=c(5,5,1,1)) #plot margins etc ###########27 ##########################################################################28 29 ##Physical Settings#######################################################30 Tsoln.C=25 #solution temp (C) - will standardize to 20 ##############31 elev=280 #elevation (masl) ##############################32 lat=43 #latitude (deg north) ##############################
10
Initial setup
change balance communication
settings and any plot preferences
Do this when setting up a new balance
>>>>>>>>>>>>>>
1 ########################### conductR #####################################2 ##########################################################################3 #### an R program to measure plant hydraulic conductance with an #########4 #### electronic balance or pipette. #####################################5 ##########################################################################6 #### Written by Duncan D. Smith 2012-2015#################################7 #### free to use and alter for all non-commercial purposes ##############8 #### Conceptually based on John Sperry's conduct.ver1.xls ################ 9 ##########################################################################10 require(tcltk) #load tcltk ##############11 ##########################################################################1213 ##Computer-specific settings - *three* things ############################14 kwd="/users/duncan/desktop" #working directory ###########15 .Tcl('set port /dev/cu.usbserial-00001004') #port number ############16 # source("/users/duncan/documents/conductR/conductR_12.R") ###############17 #run the above line (without #) to 'load' the program18 ##########################################################################1920 ##Communication/Plot Settings#############################################21 .Tcl('set baud 9600') #baud rate ###############22 .Tcl('set prty o') #parity ##################23 .Tcl('set dbit 7') #data bits ###############24 .Tcl('set sbit 1') #stop bits ###############25 .Tcl('set prnt \x1B\x50\xD\xA') #print code (Esc-P-CR-LF) ################26 par(mar=c(5,5,1,1)) #plot margins etc ###########27 ##########################################################################28 29 ##Physical Settings#######################################################30 Tsoln.C=25 #solution temp (C) - will standardize to 20 ##############31 elev=280 #elevation (masl) ##############################32 lat=43 #latitude (deg north) ##############################
11
Initial setup
check physical settings (location and
temperature)
>>>>>>>>>>>>>>
1 ########################### conductR #####################################2 ##########################################################################3 #### an R program to measure plant hydraulic conductance with an #########4 #### electronic balance or pipette. #####################################5 ##########################################################################6 #### Written by Duncan D. Smith 2012-2015#################################7 #### free to use and alter for all non-commercial purposes ##############8 #### Conceptually based on John Sperry's conduct.ver1.xls ################ 9 ##########################################################################10 require(tcltk) #load tcltk ##############11 ##########################################################################1213 ##Computer-specific settings - *three* things ############################14 kwd="/users/duncan/desktop" #working directory ###########15 .Tcl('set port /dev/cu.usbserial-00001004') #port number ############16 # source("/users/duncan/documents/conductR/conductR_12.R") ###############17 #run the above line (without #) to 'load' the program18 ##########################################################################1920 ##Communication/Plot Settings#############################################21 .Tcl('set baud 9600') #baud rate ###############22 .Tcl('set prty o') #parity ##################23 .Tcl('set dbit 7') #data bits ###############24 .Tcl('set sbit 1') #stop bits ###############25 .Tcl('set prnt \x1B\x50\xD\xA') #print code (Esc-P-CR-LF) ################26 par(mar=c(5,5,1,1)) #plot margins etc ###########27 ##########################################################################28 29 ##Physical Settings#######################################################30 Tsoln.C=25 #solution temp (C) - will standardize to 20 ##############31 elev=280 #elevation (masl) ##############################32 lat=43 #latitude (deg north) ##############################
12
Initial setup
save the script if changes were made
1 ########################### conductR #####################################2 ##########################################################################3 #### an R program to measure plant hydraulic conductance with an #########4 #### electronic balance or pipette. #####################################5 ##########################################################################6 #### Written by Duncan D. Smith 2012-2015#################################7 #### free to use and alter for all non-commercial purposes ##############8 #### Conceptually based on John Sperry's conduct.ver1.xls ################ 9 ##########################################################################10 require(tcltk) #load tcltk ##############11 ##########################################################################1213 ##Computer-specific settings - *three* things ############################14 kwd="/users/duncan/desktop" #working directory ###########15 .Tcl('set port /dev/cu.usbserial-00001004') #port number ############16 # source("/users/duncan/documents/conductR/conductR_12.R") ###############17 #run the above line (without #) to 'load' the program18 ##########################################################################1920 ##Communication/Plot Settings#############################################21 .Tcl('set baud 9600') #baud rate ###############22 .Tcl('set prty o') #parity ##################23 .Tcl('set dbit 7') #data bits ###############24 .Tcl('set sbit 1') #stop bits ###############25 .Tcl('set prnt \x1B\x50\xD\xA') #print code (Esc-P-CR-LF) ################26 par(mar=c(5,5,1,1)) #plot margins etc ###########27 ##########################################################################28 29 ##Physical Settings#######################################################30 Tsoln.C=25 #solution temp (C) - will standardize to 20 ##############31 elev=280 #elevation (masl) ##############################32 lat=43 #latitude (deg north) ##############################
13
Initiating the program
>>>>>>>>>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)
“load” the script by copying the source command into the
console
you can close the script
14
>>>>>>>>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)>
An empty plotting window will open if one doesn’t exist already
Initiating the program
XQuartz or X11 or Tk (on a PC) will open as well
15
>>>>>>>>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)
Making measurements with a balance
Measurements will be made with the cond
function
cond() requires you input an ID (‘numbr’ e.g. stem 21), a group
name (‘sp’ as the group is typically a species e.g. “bepa’ for Betula
papyrifera) and a treatment pressure (‘Ptreat.MPa’ e.g. 0 for a
flushed stem*)
*If measuring native conductance and you didn’t measure native xylem pressure, choose some identifying number for Ptreat.MPa (e.g. 999). The program needs a number because it will always try to calculate PLC relative to the most favorable pressure.
16
>>>>>>>>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)
Making measurements with a balance
Note the use of quotes. Quotes tell R this is a word and to not try to interpret
it. Things without quotes are either numbers or defined variables. Quotes
are needed for numbr if letters are involved (e.g. “21a”)
conductR requires numbr, sp and Ptreat.MPa for grouping measurements together.
• All measurements with the same sp will be pooled in the same .csv files. • All measurements with the same sp, numbr and P.MPa (I call this nomID) will be treated as part of the same conductance measurement. • Measurements with the same sp and numbr (I call this stemID) will be grouped for calculating PLC
There are other options for cond but the defaults are fine for now
(see Appendix A)
17
>>>>>>>>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)
Making measurements with a balance
I’ll start with no applied pressure to the stem.
For an overview of the methodology of measuring hydraulic conductance see
Anna Jacobsen’s document https://www.csub.edu/~ajacobsen/Conductivity%20Methods_2014.pdf
18
>>>>>>>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESC
Making measurements with a balance
Individual flow rates (black) will start to plot every 6
seconds (by default) as they accumulate, followed by the
running means (cyan)
-0.00020
-0.00016
-0.00012
-0.00008
bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
-0.00020
-0.00016
-0.00012
-0.00008
bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
19
>>>>>>>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESC
Making measurements with a balance
Each bar represents one unit of resolution: My balance resolution is
0.0002 g and measurements are every 6
s, so the resolution is 0.000033 g/s
-0.00020
-0.00016
-0.00012
-0.00008
bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
20
>>>>>>>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESC
Making measurements with a balance
Balance resolution can be changed in cond,
e.g., cond(…,resl=0.0001)
(if it’s wrong it doesn’t affect calculations it’s mostly used to assess
variability)
-0.00020
-0.00016
-0.00012
-0.00008
bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
21
>>>>>>>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESC
Making measurements with a balance
Wait for stability
This looks pretty good (repeating pattern)
-0.00020
-0.00016
-0.00012
-0.00008
bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
22
>>>>>>>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESC
Making measurements with a balance
Click the “Done” button
(the button is part of XQuartz (or X11 or Tcl), so select that
program if you see no button)
-0.00020
-0.00016
-0.00012
-0.00008
bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
12345678
9
23
>>>>>>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)
Making measurements with a balance
choose which running mean to
keep follow the prompts
-0.00020
-0.00016
-0.00012
-0.00008
bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
12345678
9
24
>>>>>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)ending pressure? (cm)
Making measurements with a balance
what was the pressure driving flow? (0 in this case)
-0.00020
-0.00016
-0.00012
-0.00008
bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
12345678
9
25
>>>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)ending pressure? (cm)no previous diameter. d.mm= 5.2no previous length. l.mm= 242
Making measurements with a balance
What are the stem’s diameter and length?
-0.00020
-0.00016
-0.00012
-0.00008
bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
12345678
9
26
>>>>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)ending pressure? (cm)no previous diameter. d.mm= 5.2no previous length. l.mm= 242previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=
Making measurements with a balance
Check the solution temperature change if needed
-0.00020
-0.00016
-0.00012
-0.00008
bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
12345678
9
27
>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)ending pressure? (cm)no previous diameter. d.mm= 5.2no previous length. l.mm= 242previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data written in NEW file as: 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”
Making measurements with a balance
A “flowdata.csv” file with all measurements and important settings has been created in
your working directory
The program will add to this file every time you measure this
species (bepa)
*this is the old format to add a full timestamp (yyyy-mm-dd hh-mm-ss), but beyond ‘dd’ serves no purpose (it used to have a purpose).
The saved flowdata.csv file
looks like this
Making measurements with a balance
time of each measurement
(seconds since 1970 - R’s choice not
mine)time (s) between
measurements
readings from balance
each flow rate (g/s)
running means of flow rates (g/s)
first reading
last reading
ver 10 also saves time as decimal day (not shown) 28
Making measurements with a balance
pressure driving flow (user input)
pressure driving flow converted
to MPa
stem diameter
stem length
stem number
species
treatment pressure
species and
number combined
species, number
treatment pressure
combinedsolution
temperature
which point was chosen
as finalwhether point was used to
calculate conductance
(will make more sense later)
flowdata continued
29
ver 11+ also includes a
measurement number,
which mostly helps the program
when fixing errors
30
> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)ending pressure? (cm)no previous diameter. d.mm= 5.2no previous length. l.mm= 242previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data written in NEW file as: 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”> cond(21,”bepa”,0)
Making measurements with a balance
Now I’ll measure with applied pressure to the stem.
Run cond exactly as before
(press the up arrow to recall previous commands in the
console)
0.00160
0.00170
0.00180
0.00190 bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
31
>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)ending pressure? (cm)no previous diameter. d.mm= 5.2no previous length. l.mm= 242“data written in NEW file as: 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”> cond(21,”bepa”,0)"CLICK DONE! WHEN FINISHED - DO NOT HIT ESC!"
Making measurements with a balance
When there are many points, only the recent ones are shown in the main plot. All running means are in the inset
Inset size (as a fraction of total plot size) is controlled by ovrl in
cond. See Appendix A
0.00160
0.00170
0.00180
0.00190 bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
12345
678
9
1011
121314
15
1617
1819
1020
32
>> source(“/users/duncan/desktop/conductR/conductR_12.R”)> cond(21,”bepa”,0)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)ending pressure? (cm)no previous diameter. d.mm= 5.2no previous length. l.mm= 242“data written in NEW file as: 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”> cond(21,”bepa”,0)"CLICK DONE! WHEN FINISHED - DO NOT HIT ESC!"
Making measurements with a balance
Reasonably stable. Note that the y range of the plot is the range of all the data, so while it’s noisy,
there is not much variation in terms of measurement
resolution
0.00160
0.00170
0.00180
0.00190 bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
12345
678
9
1011
121314
15
1617
1819
1020
33
“CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)ending pressure? (cm)no previous diameter. d.mm= 5.2no previous length. l.mm= 242“data written in NEW file as: 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”> cond(21,”bepa”,0)"CLICK DONE! WHEN FINISHED - DO NOT HIT ESC!”which point to keep? (blank=last)ending pressure? (cm) 66.6previous d.mm=5.2 blank=use otherwise d.mm=previous l.mm=242 blank=use otherwise l.mm=
Making measurements with a balance
As before, click done, choose a point and input the pressure
This time the program knows the diameter and
length from before
0.00160
0.00170
0.00180
0.00190 bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s)
12345
678
9
1011
121314
15
1617
1819
1020
34
no previous length. l.mm= 242“data written in NEW file as: 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”> cond(21,”bepa”,0)"CLICK DONE! WHEN FINISHED - DO NOT HIT ESC!”which point to keep? (blank=last)ending pressure? (cm) 66.6previous d.mm=5.2 blank=use otherwise d.mm=previous l.mm=242 blank=use otherwise l.mm=previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”
Making measurements with a balance
All these measurements are added to the
existing flowdata.csv file
0 1 2 3 40.0000
0.0005
0.0010
0.0015
elapsed time (min)
mea
n fl
ow ra
te, Q
(g/s
)
35
10-24-27 bepa flowdata.csv to /users/duncan/desktop”> cond(21,”bepa”,0)"CLICK DONE! WHEN FINISHED - DO NOT HIT ESC!”which point to keep? (blank=last)ending pressure? (cm) 66.6previous d.mm=5.2 blank=use otherwise d.mm=previous l.mm=242 blank=use otherwise l.mm=previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”“(go back one plot to see all mean flow rates)”
Making measurements with a balance
When there are two or more sets of
measurements, the program shows all running
means over time. Red points are those we chose
to keep
The program shows this plot for only 1 second before making the next plot. To toggle through plots, select
Quartz window and use left and right arrow keys while holding ⌘
0 10 20 30 40 50 600.0000
0.0005
0.0010
0.0015
bepa 21 0 MPa
pressure (cm)
mea
n fl
ow ra
te, Q
(g/s
)1
2k = 0.2675r2 = 1
36
10-24-27 bepa flowdata.csv to /users/duncan/desktop”> cond(21,”bepa”,0)"CLICK DONE! WHEN FINISHED - DO NOT HIT ESC!”which point to keep? (blank=last)ending pressure? (cm) 66.6previous d.mm=5.2 blank=use otherwise d.mm=previous l.mm=242 blank=use otherwise l.mm=previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”“(go back one plot to see all mean flow rates)”
Making measurements with a balance
And final running means vs pressure
0 10 20 30 40 50 600.0000
0.0005
0.0010
0.0015
bepa 21 0 MPa
pressure (cm)
mea
n fl
ow ra
te, Q
(g/s
)1
2k = 0.2675r2 = 1
37
10-24-27 bepa flowdata.csv to /users/duncan/desktop”> cond(21,”bepa”,0)"CLICK DONE! WHEN FINISHED - DO NOT HIT ESC!”which point to keep? (blank=last)ending pressure? (cm) 66.6previous d.mm=5.2 blank=use otherwise d.mm=previous l.mm=242 blank=use otherwise l.mm=previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”“(go back one plot to see all mean flow rates)”
Making measurements with a balance
Note conductance (k) show here has units of g s-1 MPa-1, corrected to
20 C - even though plot gives pressure in cm
0 10 20 30 40 50 600.0000
0.0005
0.0010
0.0015
bepa 21 0 MPa
pressure (cm)
mea
n fl
ow ra
te, Q
(g/s
)1
2k = 0.2675r2 = 1
38
> cond(21,”bepa”,0)"CLICK DONE! WHEN FINISHED - DO NOT HIT ESC!”which point to keep? (blank=last)ending pressure? (cm) 66.6previous d.mm=5.2 blank=use otherwise d.mm=previous l.mm=242 blank=use otherwise l.mm=previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”“(go back one plot to see all mean flow rates)”exclude points? (input number to exclude, blank=no change)
Making measurements with a balance
You have the option to remove bad points. This can be more useful later
39
> cond(21,”bepa”,0)"CLICK DONE! WHEN FINISHED - DO NOT HIT ESC!”which point to keep? (blank=last)ending pressure? (cm) 66.6previous d.mm=5.2 blank=use otherwise d.mm=previous l.mm=242 blank=use otherwise l.mm=previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”“(go back one plot to see all mean flow rates)”exclude points? (input number to exclude, blank=no change)
Making measurements with a balance
Next I’ll measure flow with no applied
pressure, using cond just as before
-0.00016
-0.00012
-0.00008
bepa 21 0 MPa
elapsed time
flow
rate
, Q (g
/s) 12
34
56
7
8910
111213
14
40
ending pressure? (cm) 66.6previous d.mm=5.2 blank=use otherwise d.mm=previous l.mm=242 blank=use otherwise l.mm=previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”“(go back one plot to see all mean flow rates)”exclude points? (input number to exclude, blank=no change)> cond(21,”bepa”,0)> “CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)
Making measurements with a balance
initiate measurements as before and wait for stability. Then choose point to keep, input pressure, change or keep diameter and length
0 2 4 6 8 100.0000
0.0005
0.0010
0.0015
elapsed time (min)
mea
n fl
ow ra
te, Q
(g/s
)
“(go back one plot to see all mean flow rates)”exclude points? (input number to exclude, blank=no change)> cond(21,”bepa”,0)> “CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)ending pressure? (cm) 0previous d.mm=5.2 blank=use otherwise d.mm=previous l.mm=242 blank=use otherwise l.mm=previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”
41
Making measurements with a balance
Again, all points over time are shown…
0 10 20 30 40 50 600.0000
0.0005
0.0010
0.0015
bepa 21 0 MPa
pressure (cm)
mea
n fl
ow ra
te, Q
(g/s
)1
2
3
k = 0.2613r2 = 0.998
42
Making measurements with a balance
Followed by finals vs pressure
exclude points? (input number to exclude, blank=no change)> cond(21,”bepa”,0)> “CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)ending pressure? (cm) 0previous d.mm=5.2 blank=use otherwise d.mm=previous l.mm=242 blank=use otherwise l.mm=previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”exclude points? (input number to exclude, blank=no change)
This looks good. The background points (1 and 3) are quite
similar
0 10 20 30 40 50 600.0000
0.0005
0.0010
0.0015
bepa 21 0 MPa
pressure (cm)
mea
n fl
ow ra
te, Q
(g/s
)1
2
3
k = 0.2551r2 = 1
Followed by finals vs pressure
43
Making measurements with a balance
This looks good. The background points (1 and 3) are quite
similar but I can delete* #1, just to show what happens
*don’t worry, those data aren’t really gone
> cond(21,”bepa”,0)> “CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)ending pressure? (cm) 0previous d.mm=5.2 blank=use otherwise d.mm=previous l.mm=242 blank=use otherwise l.mm=previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”exclude points? (input number to exclude, blank=no change) 1exclude points? (input number to exclude, blank=no change)
0 10 20 30 40 50 600.0000
0.0005
0.0010
0.0015
bepa 21 0 MPa
pressure (cm)
mea
n fl
ow ra
te, Q
(g/s
)1
2
3
k = 0.2551r2 = 1
44
Making measurements with a balance
> “CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last)ending pressure? (cm) 0previous d.mm=5.2 blank=use otherwise d.mm=previous l.mm=242 blank=use otherwise l.mm=previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”exclude points? (input number to exclude, blank=no change) 1exclude points? (input number to exclude, blank=no change)> cond(21,”bepa”,0,justcalc=T)
It was fine with #1 but there’s no ‘undo’
To get #1 back, run cond just as before, but add justcalc=T
This will skip making measurements and allow you
to calculate and save conductance
0 10 20 30 40 50 600.0000
0.0005
0.0010
0.0015
bepa 21 0 MPa
pressure (cm)
mea
n fl
ow ra
te, Q
(g/s
)1
2
3
k = 0.2613r2 = 0.998
45
Making measurements with a balance
previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to 2015-12-09 10-24-27 bepa flowdata.csv to /users/duncan/desktop”exclude points? (input number to exclude, blank=no change) 1exclude points? (input number to exclude, blank=no change)> cond(21,”bepa”,0,justcalc=T)“(go back one plot to see all mean flow rates)”exclude points? (input number to exclude, blank=no change)save conductance? (y=yes blank=no) y
Now you’re asked to save conductance. When we
excluded a point, it didn’t ask. By default, the
program requires at least 3 measurements,
which is the bare minimum to assess if Q vs P is linear. The minimum requirement
can be changed. See Appendix A
46
Making measurements with a balance
> cond(21,”bepa”,0,justcalc=T)“(go back one plot to see all mean flow rates)”exclude points? (input number to exclude, blank=no change)save conductance? (y=yes blank=no) y“data written in NEW file as: 2015-12-09 11-05-23 bepa condcalc.csv to /users/duncan/desktop” sp numbr P.MPa nomID stemID l.mm1 bepa 21 0 bepa 21 0 MPa bepa 21 242 d.mm k.g.s.MPa r2 K.g.mm.s.MPa 1 5.2 0.261289 0.9982949 63.231945Ks.g.s.MPa.mm plc1 2.97741 0
Conductance calculations are saved in a new
“condcalc.csv” file* in your working directory
and the program shows you all data within that file
saved conductances and conductivities are
standardized to 20 C.
*shown is the old name format with full timestamp (yyyy-mm-dd hh-mm-ss), but beyond ‘dd’ serves no purpose (it used to have a purpose).
>>>>>>>>>>>>>> showvuln(“qudo")
if constructing a vulnerability curve, the showvuln function plots all data in
condcalc.csv for the specified species (note this is qudo not bepa)
Ks is plotted by default (see Appendix A)
47
0 1 2 3 4 50.0
0.1
0.2
0.3
0.4
0.5
P.MPa
Ks.g.s.MPa.mm
qudo 3qudo 5
Viewing vulnerability
0 1 2 3 4 50
2040
6080
P.MPa
PLC
qudo 3qudo 5
>>>>>>>>>>>>> showvuln(“qudo)> showvuln(“qudo”,as.plc=T)
plotting PLC is another option (see Appendix A)
48
Viewing vulnerability
Measurements with a pipette: manual mode
Most details of the program are the same except you control the measurements
Instead of the program measuring change at fixed times, you tell the program when a fixed unit of fluid has
moved into (or out of) the xylem
49
50
Making measurements with a pipette
>>>>>>>>>>>>>> cond(1,”test”,0,resl=0.01,manul=T)
Use the cond function just as with the balance but:
use manul=T; resolution (resl) will be the
volume increment (ml) of each measurement
(with the balance, resl had units of g and setting it was not
crucial)
51
>>>>>>>>>>>>>> cond(1,”test”,0,resl=0.01,manul=T)
Here I’m measuring stem 1 of species “test”
(These data will be made up)
For these “measurements” I will use three different pressure
heads
Making measurements with a pipette
52
>>>>>>>>>>>>>> cond(1,”test”,0,resl=0.01,manul=T)
Note this layout of windows is useful:
The console has to be selected to make measurements so keep it to the side of the plot so you can see
the data
Making measurements with a pipette
53
>>>>>>>>>>>> cond(1,”test”,0,resl=0.01,manul=T)ready to start! increment = 0.01 ml (press enter)
Once initiated, the program is ready to start the timer
Making measurements with a pipette
54
>> cond(1,”test”,0,resl=0.01,manul=T)ready to start! increment = 0.01 ml (press enter)record n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=done
Making measurements with a pipette
Hit enter every time the meniscus crosses a tick mark
on the pipette
Flow rates and running means accumulate
0.000
0.002
0.004
0.006
0.008
test 1 0 MPa
elapsed time
flow
rate
, Q (m
l/s)
55
>> cond(1,”test”,0,resl=0.01,manul=T)ready to start! increment = 0.01 ml (press enter)record n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=done
Making measurements with a pipette
Just hitting enter tells the program that the meniscus has
moved one tick mark.
If you missed one tick, you could input “2” and hit enter on
the next tick
0.000
0.002
0.004
0.006
0.008
test 1 0 MPa
elapsed time
flow
rate
, Q (m
l/s)
0.000
0.002
0.004
0.006
0.008
test 1 0 MPa
elapsed time
flow
rate
, Q (m
l/s)
123456789
56
record n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=done d
Making measurements with a pipette
To finish, there is no “done” button, instead input “d” and
hit enter as prompted
0.000
0.002
0.004
0.006
0.008
test 1 0 MPa
elapsed time
flow
rate
, Q (m
l/s)
123456789
57
record n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=done dwhich point to keep? (blank=last)ending pressure? (cm) 8d.mm blank and no previous. d.mm= 3.2l.mm blank and no previous. l.mm= 152previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=
Making measurements with a pipette
As with the balance, choose a point, input the pressure,
diameter and length Check solution temp
58
record n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=done dwhich point to keep? (blank=last)ending pressure? (cm) 8d.mm blank and no previous. d.mm= 3.2l.mm blank and no previous. l.mm= 152previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data written in NEW file as: 2015-12-09 12-24-27 test flowdata manul.csv to /users/duncan/desktop”
Making measurements with a pipette
A new flowdata .csv file of all the measurements gets
created*
*this .csv will have nearly the same layout as one created when using the balance. The key difference is units will be in ml instead of g. Note that ‘manul’ appears in the file name. This prevents an error if you happen to use the balance and pipette on the same species
0.000
0.002
0.004
0.006
0.008
test 1 0 MPa
elapsed time
flow
rate
, Q (m
l/s)
123456789
As with the balance, choose a point, input the pressure,
diameter and length Check solution temp
0.000
0.001
0.002
0.003
0.004
0.005
test 1 0 MPa
elapsed time
flow
rate
, Q (m
l/s)
59
12-24-27 test flowdata manul.csv to /users/duncan/desktop”> cond(1,”test”,0,resl=0.01,manul=T)ready to start! increment = 0.01 ml (press enter)record n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=done d
Making measurements with a pipette
Repeat for a different pressure head
wait for stability
60
record n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=done dwhich point to keep? (blank=last)ending pressure? (cm) 4previous d.mm=3.2 blank=keep otherwise d.mm =previous l.mm=152 blank=keep otherwise l.mm =previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to: 2015-12-09 12-24-27 test flowdata.csv to /users/duncan/desktop”
Making measurements with a pipette
All running means vs time
0.0 0.5 1.0 1.50.0025
0.0035
0.0045
elapsed time (min)
mea
n fl
ow ra
te, Q
(ml/s
)
61
record n increments; blank=1; d=donerecord n increments; blank=1; d=done dwhich point to keep? (blank=last)ending pressure? (cm) 4previous d.mm=3.2 blank=keep otherwise d.mm =previous l.mm=152 blank=keep otherwise l.mm =previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to: 2015-12-09 12-24-27 test flowdata.csv to /users/duncan/desktop”“(go back one plot to see all mean flow rates)”exclude points? (input number to exclude, blank=no change)
Making measurements with a pipette
Final running means vs pressure
4 5 6 7 80.0025
0.0035
0.0045
test 1 0 MPa
pressure (cm)
mea
n fl
ow ra
te, Q
(ml/s
)
1
2
k = 5.88r2 = 1
62
blank=no change)> cond(1,”test”,0,resl=0.01,manul=T)ready to start! increment = 0.01 ml (press enter)record n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=done d
Making measurements with a pipette
Repeat for third pressure head.
0.0000
0.0005
0.0010
0.0015
0.0020
0.0025 test 1 0 MPa
elapsed time
flow
rate
, Q (m
l/s)
63
record n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=donerecord n increments; blank=1; d=done dwhich point to keep? (blank=last)ending pressure? (cm) 2previous d.mm=3.2 blank=keep otherwise d.mm =previous l.mm=152 blank=keep otherwise l.mm =previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to: 2015-12-09 12-24-27 test flowdata manul.csv to /users/duncan/desktop”
Making measurements with a pipette
All running means vs time
0 1 2 3 4 5 60.002
0.003
0.004
0.005
elapsed time (min)
mea
n fl
ow ra
te, Q
(ml/s
)
64
record n increments; blank=1; d=donerecord n increments; blank=1; d=done dwhich point to keep? (blank=last)ending pressure? (cm) 2previous d.mm=3.2 blank=keep otherwise d.mm =previous l.mm=152 blank=keep otherwise l.mm =previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to: 2015-12-09 12-24-27 test flowdata manul.csv to /users/duncan/desktop”“(go back one plot to see all mean flow rates)”exclude points? (input number to exclude, blank=no change)
Making measurements with a pipette
Final running means vs pressure
2 3 4 5 6 7 80.002
0.003
0.004
0.005
test 1 0 MPa
pressure (cm)
mea
n fl
ow ra
te, Q
(ml/s
)
1
2
3
k = 5.88r2 = 1
65
flowdata manul.csv to /users/duncan/desktop”“(go back one plot to see all mean flow rates)”exclude points? (input number to exclude, blank=no change)save conductance? (y=yes blank=no) y“data written in NEW file as: 2015-12-09 12-55-33 test condcalc manul.csv to /users/duncan/desktop” sp numbr P.MPa nomID stemID l.mm1 test 1 0 test 1 0 MPa test 1 152 d.mm k.g.s.MPa r2 K.g.mm.s.MPa 1 3.2 5.900161 0.9997041 896.8244Ks.g.s.MPa.mm plc1 111.511 0
Making measurements with a pipette
A new condcalc .csv file gets created
Here are the data in that file
2 3 4 5 6 7 80.002
0.003
0.004
0.005
test 1 0 MPa
pressure (cm)
mea
n fl
ow ra
te, Q
(ml/s
)
1
2
3
k = 5.88r2 = 1
Measuring leaves (balance or pipette)
The program may be configured to measure hydraulic conductance of leaves with the evaporative flux method
The protocol of measuring background flow rates doesn’t lend itself to leaf hydraulics very well. The flow through a transpiring
leaf depends on leaf water potential, which cannot be measured without disrupting hydraulic conductance (e.g. pressure bomb or
chamber psychrometry).
Given this limitation, the program calculates leaf hydraulic conductance from a single flow rate and water potential
66
67
>>>>>>>>>>>>>> cond(12,”bepa”,0,dt=6,kleaf=T)
Making leaf measurements (shown with a balance)
Initiate measurements using the cond function as before, but
add kleaf=T
which will let the program know there will only be one measurement before
conductance can be calculated and what information to prompt from you
manul may be T (pipette) or F (balance; default)
0.00188
0.00192
0.00196
0.00200 bepa 12 0 MPa
elapsed time
flow
rate
into
leaf
, Q (g
/s)
68
>>>>>>>>>>>>> cond(12,”bepa”,0,dt=6,kleaf=T)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESC
Making leaf measurements (shown with a balance)
Measurements will accumulate as before
(note that rates are into the leaf - out of
the balance)
0.00180
0.00190
0.00200
bepa 12 0 MPa
elapsed time
flow
rate
into
leaf
, Q (g
/s)
123
456
7
89
10
11
1213
1415
10
69
>>>>>>>>>>> cond(12,”bepa”,0,dt=6,kleaf=T)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last) infeed pressure (cm)
Making leaf measurements (shown with a balance)
After asking for point to keep, program asks for the pressure of water just before it enters the leaf.
This may be measured as hydraulic head. Zero is ideal - and
a good assumption if unknown.
0.00180
0.00190
0.00200
bepa 12 0 MPa
elapsed time
flow
rate
into
leaf
, Q (g
/s)
123
456
7
89
10
11
1213
1415
10
70
>>>>>>>> cond(12,”bepa”,0,dt=6,kleaf=T)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last) infeed pressure (cm) 0leaf pressure (MPa) 1previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=
Making leaf measurements (shown with a balance)
Next, the leaf must be removed from the tubing and pressure-bombed to get the driving force for
flow*
*The assumption is that leaf pressure is a negative number. If your input is positive, the program will make it negative.
0.0 0.2 0.4 0.6 0.8 1.00.0000
0.0005
0.0010
0.0015
bepa 12 0 MPa
pressure difference (MPa)
mea
n fl
ow ra
te in
to le
af, Q
(g/s
)
1k = 0.00166r2=NA
71
>>>>>> cond(12,”bepa”,0,dt=6,kleaf=T)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last) infeed pressure (cm) 0leaf pressure (MPa) 1previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to: 2015-12-09 12-24-27 bepa flowdata kleaf.csv to /users/duncan/desktop”
Making leaf measurements (shown with a balance)
The flow data gets saved* and the flow vs pressure
relationship will be shown. Since we only have one point, the line runs through the origin
The “pressure difference” on the x-axis is infeed - leaf*
*this .csv will have a different layout from one created when measuring stems (or roots). Note that ‘kleaf’ appears in the file name. This prevents an error if you happen to measure leaves and stems or roots on the same species
0.0 0.2 0.4 0.6 0.8 1.00.0000
0.0005
0.0010
0.0015
bepa 12 0 MPa
pressure difference (MPa)
mea
n fl
ow ra
te in
to le
af, Q
(g/s
)
1k = 0.00166r2=NA
72
>>> cond(12,”bepa”,0,dt=6,kleaf=T)“CLICK DONE! WHEN FINISHED - DO NOT HIT ESCwhich point to keep? (blank=last) infeed pressure (cm) 0leaf pressure (MPa) 1previous Tsoln.C= 25 blank=use, otherwise Tsoln.C=“data APPENDED to: 2015-12-09 12-24-27 bepa flowdata kleaf.csv to /users/duncan/desktop”exclude points? (input number to exclude, blank=no change)save conductance? (y=yes blank=no) y
Making leaf measurements (shown with a balance)
Next, you’re prompted to exclude points if needed and save
conductance. Done
A number of features prevent the user from inputting values that cause the program to error. For example, when asked which point to keep, the input can only be an integer between 1 and the total number of mean flow rates, which often helps if you accidentally enter pressure instead. Pressure, diameter, length and temperature must be numbers. The program will ask for new values if these criteria are not met. Some accidents are accepted by the program though. Most common errors are fixed via cond(…,fix=T), where … must contain the number, species and treatment pressure of the measurement you want to fix. Using fix=T will prompt you for which measurement to fix (i.e. the numbers shown in the Q vs P plot), what to fix (options depend on measurement type - leaf or stem/root), and what the corrected value is. Changes are made to the flowdata csv file then the program proceeds as if a measurement was just completed.
Dealing with input errors
73
Appendix A: Functions and their parameters
cond(numbr,sp,Ptreat.MPa=0,runmean=4,dt=6,resl=0.0002,mxplot=10,
justcalc=F,manul=F,testmode=F,
ovrl=0.3,
kleaf=F,
nmin=3,
fix=F)
Parameters of the cond function (and their defaults)# stem number (or letter, use quotes with letters)# species (must be in quotes)# treatment pressure in MPa (must be a number)# number of values in running mean# target time (s) between balance readings# resolution of balance (g) or pipette (ml)# leave space for this many measurements when making a new plot after existing one is full# option to skip measurements and just calculate conductance# manual mode (T for use with pipette)# create random numbers instead of reading the balance (useful for demonstration or when altering code)# size of inset showing overall flow data - as a fraction of plot size (0=no inset)# is measurement on a leaf? If T, will ask for leaf pressure and not diam or length. Also nmin=1# the minimum number of points required to calculate k from Q vs P# option to skip measurements and change an input in a previous measurement
74
showvuln=(sp,yax=c(“k.”,"K.","Ks")[3],
as.plc=F,
horiz=90,
retrn=F)
Parameters of the showvuln function (and their defaults)
# species# what to plot on y axis (conductance, conductivity or specific conductivity (default), respectively)# option to plot percent loss conductivity, if T, yax is ignored# add a horizontal line at this height in PLC plot# option to return entire condcalc.csv file
Note: this function will search the working directory for .csv files that contain “sp condcalc”. If there are multiple matches, it will ask which to use
75
Appendix A: Functions and their parameters
pcent(rpm,rmax.mm,r=0)
Miscellaneous functions and their parameters
# returns pressure induced by centrifuge in MPa # RPM of centrifuge# radius from center of rotation to meniscus in rotor cup# point of interest in stem relative to center of rotationNote: function uses density of water = 998.2 kg m-3 (20 C)
rpmcent(p,rmax.mmr=0)
# returns RPMs needed to induce given pressure # desired pressure in MPa (needs to be negative)# radius from center of rotation to meniscus in rotor cup# point of interest in stem relative to center of rotationNote: function uses density of water = 998.2 kg m-3 (20 C)
phead(p.cm,rho,grav)
# converts pressure head in cm to MPa # head in cm# density of water in kg m-3# gravity in m s-2
76
Appendix A: Functions and their parameters
Internal functions
cond() uses functions corner() and resbar() to respectively add text and resolution bars to plots
Times are converted from “seconds since 1970” to decimal day via s2dec(), which uses nsplit() and subsplit()
77
Appendix A: Functions and their parameters
After measurement, warning(s) stating NAs introduced by coercion are sometimes given. These may be do to bad balance outputs or bad inputs for pressure, diameter or length. This should not affect measurements but the warning may cause concern.
The code was written using Sartorius balances, which makes them very easy to set up. For other balances (e.g. Mettler), how the program processes the output is quite difficult to tailor. One needs to understand the variation in balance output and how to consistently parse it into a number. If a simple solution can be found, it will be added.
Appendix B: Known issues
78
Anticipated questions:
How do I save my data? Data is automatically saved to your working directory after every measurement. You need not save anything in R except the conductR code if you alter it
How do I change the defaults in cond()? In the conductR code where it says cond=function(…) you’ll see all the function arguments and their defaults. Change as desired, save and re-source the code into the R console.
What if I input the wrong ____? Your options are to either 1) start over (good when you just started a measurement and, say, used the wrong ID) 2) use cond(…,fix=T) and follow the prompts or 3) manually edit the saved data file(s). If diameter and/or length are wrong and you need to make another measurement before calculating conductance, just input the correct d or l when prompted and the program will use this value. If you calculated conductance with a wrong input, manually recalculate K and/or Ks.
Why is this not an R package? It seemed simpler this way to keep everything organized in one file. Plus, this way the user has full access to and control over the code, which allows things like default arguments to be changed. However, there is potential for problems if your variable names overlap with mine. Besides the functions (see Appendix A) and .Tcl variables (i.e. port settings), I use: kwd, Tsoln, elev, lat, rho, grav
Appendix C