comparative dosimetry of high-detail computed tomography using the siemens somatom … · 2014. 4....

2
354 Comparative Dosimetry of High-Detail Computed Tomography Using the Siemens Somatom-2 and Complex Motion Tomography for Examination of the Sella Turcica Robert D. Zwicker,' Robert G. Zamenhof," 2 and Samuel M. Wolpert 2 Recent studies have questioned the value of hypocycloi- dal poly tomography of the sella turcica for the diagnosis of pituitary microadenomas because of the substantial number of false-positive and false-negative diagnoses [1, 2]. With the improving image quality available with state-of-the-art computed tomographic (CT) scanners, high-resolution CT appears at least as sensitive as poly tomography in diagnos- ing se ll ar adenomas [3, 4]. Thus, it is probable that CT scanning will become the optimum test for establishing the diagnosis. In view of this, we believe it is of interest to c ompare the radiation dosimetry for these two examinations. Therefore, this study seeks to compare CT and polytomo- graphic examination of the sella turcica in terms of radiation expos ures received by the thyroid, skin, lens of the eye, and pituitary gland, and also in terms of the " integral exposure, " which is a measure of the total ionizing energy absorbed by the head and often is a useful single-number index repre- senting the radiation hazard associated with a radiographic proc edure. Materials and Methods Th e dos imetric study was carried out using a wax head phantom and thermoluminescenc e dosimeters (TLDs). The wax head phan- tom, which was modeled on the Rando phantom head, neck, and shoulder sections, had a biparietal diameter of 15 c m and an anteroposterior diameter of 20 cm. The TLDs (Harshaw TLD-1 00 chips, 3 x 3 x 0.9 mm) were affixed on the surface and inside the wax head phantom (fig. 1). Their locations wer e intended to mea- sure exposures to " skin," " lenses," and " pituitary ," and to permit the calculations of " integral exposures. " The wax head phantom loaded with TLDs was subjected first to a simulated polytomo- graphic examination of the sella, and then (with fresh TLDs) to a simulated CT study of the sella. Two sets of calibration fac tor s were derived for the TLDs to allow for th e difference between the CT and poly tome x-ray beam qualities. Using a diagnostic x-ray unit ad- justed to match each of the beam qualities in turn, the TLDs were exposed " in air " to a level of a few roentgens, with the exposure simultaneously monitored by a 6 cm 3 ionization cha mber and elec- Received September 10, 198 1; accept ed November 29, 198 1. trometer (MDH Industries model 1 015C) . After this, the TLDs were stored in darkness for 48 hr and then read out on a Victoreen 2800 TLD reader. After subtraction of background readings, individual calibration factors were calculated, and all subsequently measured exposures were converted to roentgens using the appropriate cal- ibration factor for each beam quality. The independence of the calibration factor on exposure level was checked and verified. The CT examination of the sella is diagrammed in figure 1 A. Ten e-.-. . _.-._._._ . _._._._. A 8 Fig . 1.- TLD pl acement and geometry of CT scans (A) and poly tome fields (8) . 'Department of Therapeutic Radiology, Tufts-New England Medi ca l Center, 171 Harrison Ave., Boston, MA 0 2 111 . Add ress reprint requests to R. D. Zwicker. ' Department of Di ag nostic Radiology, Tuft s- New England Medical Center, Boston, MA 0 2 111 . Addr ess reprint requests to R. D. Zwicker. AJNR 3: 354 - 355 , May / June 19820195-6 108/82/ 0303 -0 35 4 $00 .00 © Ameri ca n Roentgen Ray Soc iety

Upload: others

Post on 27-Feb-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Comparative Dosimetry of High-Detail Computed Tomography Using the Siemens Somatom … · 2014. 4. 4. · Somatom-2 CT 2.0 0.2 3.8' 3.8 1,300 • Worst case (assuming intersection

354

Comparative Dosimetry of High-Detail Computed Tomography Using the Siemens Somatom-2 and Complex Motion Tomography for Examination of the Sella Turcica Robert D. Zwicker,' Robert G. Zamenhof," 2 and Samuel M. Wolpert2

Recent studies have questioned the value of hypocycloi­dal poly tomography of the sella turcica for the diagnosis of pituitary microadenomas because of the substantial number of false-positive and false-negative diagnoses [1, 2]. With the improving image quality available with state-of-the-art computed tomographic (CT) scanners, high-resolution CT appears at least as sensitive as poly tomography in diagnos­ing se llar adenomas [3, 4]. Thus, it is probable that CT scanning will become the optimum test for establishing the diagnosis . In view of this, we believe it is of interest to compare the radiation dosimetry for these two examinations. Therefore, this study seeks to compare CT and polytomo­graphic examination of the sella turcica in terms of radiation exposures received by the thyroid, skin , lens of the eye, and pituitary gland, and also in terms of the " integral exposure, " which is a measure of the total ionizing energy absorbed by the head and often is a useful single-number index repre­senting the radiation hazard associated with a radiographic procedure.

Materials and Methods

The dosimetric study was carried out using a wax head phantom and thermoluminescence dosimeters (TLDs). The wax head phan­tom, which was modeled on the Rando phantom head, neck, and shoulder sections, had a biparietal diameter of 15 cm and an anteroposterior diameter of 20 cm. The TLDs (Harshaw TLD-1 00 chips, 3 x 3 x 0 .9 mm) were affixed on the surface and inside the wax head phantom (fig. 1). Their locations were intended to mea­sure exposures to " skin ," " lenses," and " pituitary ," and to permit the calculations of " integral exposures. " The wax head phantom loaded with TLDs was subjected first to a simulated polytomo­graphic examination of th e sella, and then (with fresh TLDs) to a simulated CT study of the sella . Two sets of calibration factors were derived for the TLDs to allow for the difference between the CT and poly tome x-ray beam qualiti es. Using a diagnostic x-ray unit ad­justed to match each of the beam qualities in turn, the TLDs were exposed " in air" to a level of a few roentgens, with the exposure simultaneously monitored by a 6 cm3 ionization chamber and elec-

Received September 10, 1981; accepted November 29 , 1981.

trometer (MDH Industries model 1 015C). After this , the TLDs were stored in darkness for 48 hr and then read out on a Victoreen 2800 TLD reader. After subtraction of background readings, individual calibration factors were calculated, and all subsequently measured exposures were converted to roentgens using the appropriate cal­ibration factor for each beam quality. The independence of the calibration factor on exposure level was checked and verified.

The CT examination of the sella is diagrammed in figure 1 A. Ten

e-.-. ._.-._._._ ._._._._.

A

8 Fig . 1 .- TLD placement and geometry of CT scans (A) and poly tome fields

(8).

' Department of Therapeutic Radiology, Tufts-New England Medical Center, 17 1 Harrison Ave., Boston, MA 0 2111 . Address reprint requests to R. D. Zwicker.

' Department of Diag nostic Radiology, Tufts-New England Medica l Center , Boston, MA 0 2111 . Address reprint requests to R. D. Zwicker.

AJNR 3:354- 355, May / June 19820195-6108 / 82 / 0303-0354 $00 .00 © American Roentgen Ray Soc iety

Page 2: Comparative Dosimetry of High-Detail Computed Tomography Using the Siemens Somatom … · 2014. 4. 4. · Somatom-2 CT 2.0 0.2 3.8' 3.8 1,300 • Worst case (assuming intersection

AJNR:3, May / June 1982 DOSIMETRY OF CT AND POL YTOMOGRAPHY 355

TABLE 1: Phantom Exposures in Poly tomography and Computed Tomography of the Sella Turcica

Exposure Site

Pituitary Lens Thyroid Entrance (skin) Integral (g-R)

Exposu res (R)

Philips Poly tome

4.3 5.6'

o 12.9

2,000

Somatom-2 CT

2.0 0.2 3.8 ' 3.8

1,300

• Worst case (assuming intersection of thyroid gland by primary x-ray field in CT, and direct, unshielded exposure of lenses in poly tomography).

contiguous 2-mm-thick coronal scans were obtained using a Sie­mens Somatom-2 scanner, start ing at 10 mm posterior to and ending at 10 mm anterior to the sella. Technique factors used were: 10 sec scan, 460 mAs, and 125 kVp. The filtration of the x-ray beam was 0.4 mm of copper plus 2.5 mm of aluminum (effective). The correct alignment of the scan planes with the plane of the TLDs was verified by examination of the CT scan images.

The poly tomographic study of the sella is diagrammed in figure 1 B. Static exposures were made on a Philips Poly tome three-phase unit with field size of 6.3 x 4.1 cm and a focus-to-fulcrum distance of 122 cm. (In actual patient studies, 49° hypocycloidal motion is used ; however, the type of motion will not significantly affect ex­posures measured .) A lateral scout film was first obtained , followed by 23 lateral 1-mm-thick tomograms starting at 11 mm to the left and ending 11 mm to the right of the midline. Technique factors used for the lateral exposures were 54 kVp, 120 mAs. These were then followed by an anteroposterior scout film and five anteropos­terior tomograms of thickness 1 mm, starting at 2 mm posterior and ending 2 mm anterior to the sella. Technique factors used for the anteroposterior exposures were 60 kVp and 180 mAs. Kilovoltage accuracy was verified using a Victoreen Wisconsin Cassette.

For the CT study, the integral exposure was obtained by linear interpolation between measured " skin " exposures and double ex­ponential fitting of the exposures measured inside the head phan­tom. Single exponential fitting was used to determine the poly tome integral exposure.

Results

Table 1 summarizes the results of the study. Pituitary exposure is about twice as high in the Poly tome examination as in the CT study. Without protective eye shields, the lens exposure in the Poly tome study is much greater than with

CT, although, with the use of eye shields , the exposure would be greatly reduced . Entrance exposure is about three

times as high in the Poly tome examination as with CT . Thyroid exposure is far higher in the CT study if the se­quence of scans intersects the thyroid gland . Finally , inte­gral exposure is about 1 .5 times as high in th e Poly tome examination as in the CT study.

Discussion

In general, the examination of the pituitary by CT appears to be less hazardous to the patient in terms of radiation exposure than by poly tomography. In poly tomography, th e lens exposure can be significantly reduced by using protec­tive eye shields. In CT, the examination should be performed so as not to allow the thyroid gland to fall within the primary radiation field. It should be noted that th e technique factors used in the poly tome study reflect the fact that a fast rare­earth screen-film combination was used . If non-rare-earth intensifying screens were used, all the Poly tome exposures would be at least doubled . However, th e Somatom-2 CT

scanner is also a very dose-efficient machine, and, if a less dose-efficient CT scanner were used , the CT exposures would also be increased. In this context, we believe that the comparison we have made in this study is not parti cularly biased toward either imaging method. The fact that the head phantom used was fabricated out of wax rather than a more realistically tissue-equivalent material would tend to bias the pituitary and integral exposures in favor of th e CT exami­nation. However, we do not believe that this factor would

affect our general conclusions that, at this institution, the CT examination of the sella turcica involves lower radiation exposure to the patient than does the poly tomographic examination.

REFERENCES

1. Burrow GN, Wortzman G, Rewcastle NB, Holgate RC, Kovacs K. Microadenomas of the pituitary and abnormal sellar tomo­grams in an unselected autopsy series. N Engl J Med 1981 ;304: 156-158

2. Turksi PA , Newton TH , Horten BH . Sellar contour: anatomic­poly tomographic correlat ion. AJNR 1981 ;2: 331 - 334

3. Naidich TP, Pinto TS, Kushner MJ, et al. Evaluation of sellar and parasel lar masses by computed tomography. Radiology 1976;120:91-99

4. Syvertsen A, Haughton VM , Williams AL , Cusick JF. The com­puted tomographic appearance of the normal pituitary gland and pituitary microadenomas. Radiology 1979; 133: 385- 39 1