column selection - microsoft · column selection general ... general considerations for column...
TRANSCRIPT
�
ListofTables..................................................................................................................................................2 ListofFigures.................................................................................................................................................2
ColumnSelection GeneralConsiderationsforColumnSelection......................................................................................... 3-11 YourZebronCapillaryGCColumn......................................................................................................... 12-29
ColumnInstallation Pre-InstallationCheckList...........................................................................................................................30 InstallationToolsandSupplies.....................................................................................................................30 DetailedColumnInstallationInstructions.....................................................................................................31 CapillaryColumnPreparation CriticalColumnInstallationSteps................................................................................................................32 InjectorInstallation DetectorInstallation InstallingtheColumn...................................................................................................................................33 ConditioningandTestingtheCapillaryColumn TroubleshootingInstallationProblems........................................................................................................34 HintsandTips ProtectingtheColumn.................................................................................................................................34 Liners............................................................................................................................................................35 GuardColumns...................................................................................................................................... 36-37 ColumnBakeOut.........................................................................................................................................36 PhaseRatio(ß)....................................................................................................................................... 37-38 ShorteningRunTimes............................................................................................................................ 39-40 CarrierGasSelectionandFlowOptimization..............................................................................................41 TemperatureProgramming...........................................................................................................................42 CheckingforLeaks......................................................................................................................................42 InjectionTechniques.....................................................................................................................................43 GasFlowSettings........................................................................................................................................44 CalculatingSplitRatioandColumnFlowRate............................................................................................44 ColumnDeadTimesandMarkers................................................................................................................45 SampleCapacity..........................................................................................................................................46 SolventRinsingofZebron™CapillaryColumns...........................................................................................47 ChemicalCompatibility................................................................................................................................48 “BakingOut”theColumn.............................................................................................................................48 ColumnStorage...........................................................................................................................................48
TaBLeOFCOnTenTS
�
Appendices References............................................................................................................................................................49 Glossary ........................................................................................................................................................ 50-51 equations........................................................................................................................................................ 52-54
ListofTablesTable1: UsesandTrendsbyColumnLength.......................................................................................................4Table2: UsesandTrendsbyColumnID..............................................................................................................6Table3: UsesandTrendsbyColumnFilmThickness..........................................................................................8Table4: PhasePolarity.......................................................................................................................................11Table5: PhaseRatio(ß)Values...........................................................................................................................38Table6: InjectionModesandSelectedSpecifications.......................................................................................43Table7: GasesUsedwithCommonDetectors..................................................................................................43Table8: TypicalHeadpressures(forcolumnpurging).........................................................................................44Table9:expansionVolumesofCommonSolvents............................................................................................44Table10:SplitlessHoldTimes.............................................................................................................................45Table11:RecommendedDeadVolumeTimeMarkers........................................................................................45Table12:RecommendedMethaneRetentionTimes...........................................................................................46Table13:approximateSampleCapacity(maxforsinglecomponent)................................................................46Table14:PhaseCompatibilitywithRinsingSolvents..........................................................................................47Table15:RinsingConditions................................................................................................................................47
ListofFiguresFigure1:ProperandImproperCutCapillaryend................................................................................................31Figure2:CuttingFusedSilicaTubing..................................................................................................................31Figure3:ColumnHanging..................................................................................................................................32Figure4:MeasureInjectionPortDistance...........................................................................................................32Figure5:Z-GuardConnection.............................................................................................................................36Figure6:GuardianColumn.................................................................................................................................36Figure7:PhaseRatio..........................................................................................................................................38Figure8&9:ButaneIsomerSeparations............................................................................................................39Figure10,11,&12:PaHSeparations..................................................................................................................40Figure13:CarrierGasSelectionandVelocityOptimizationPlots........................................................................41Figure14:InletPressurevs.Velocity.....................................................................................................................42Figure15:DeadVolumePeakShapeTest............................................................................................................46
�
Todefineacapillarycolumn,fourparametersmustbespecified:�. LENGTH: Lengthisdirectlyrelatedtooverallefficiencyofthecolumnandtooverallanalysistime.
Therelationbetweenlengthandresolutionisasquarerootrelation,whileanalysistimeisdirectlyrelatedtocolumnlength.
�. INTERNAL dIAmETER:Columninternaldiameterhasamajorimpactoncolumnefficiency(andthusonresolution)andonthesamplecapacityofthecolumn.Itcanalsodictatelimitationsontheinjectionanddetectiontechniquesusedandviceversa.Smaller(InternalDiameter)giveshigherk,butlowercapacity.
�. FILm THICKNESS: Filmthicknessdeterminessoluteretentionandthussoluteelutiontempera-
tures.Itwillalsoplayanimportantroleinthesamplecapacityofthecolumn.Thinfilmsarefasterwithhigherresolution,butofferlowercapacity.
4. PHASE: Thestationaryphaseisthemostinfluentialcolumnparameter.Itnotonlydeterminesthefinalresolutionobtained(itdeterminestherelativeretentionofthesolutes),butduetoitsspecificcharacteristics,itwillinfluencevirtuallyeverycolumnselectionparameter.
Othercriterialikeinstrumentavailability,feasibility,budgetconsiderations,etc.canalsoimposelimitationsoncolumnselection.experience,literaturedataandyourPhenomenexTechnicalRepresentativeareinvalu-abletoolstoguideyouthroughthelabyrinthofcolumnselection.
General Considerations for Column Selection
4
I. Length
A. Influence of length on resolution. Increasingthelengthwillincreaseresolutioninasquarerootrelation.Dueto
disadvantagesofverylongcolumns,increasingresolutionbyincreasingthelengthislimited.
B. Influence of length on sample capacity. Longcolumnshaveahighersamplecapacity(squarerootrelation).C. Influence of length on analysis time. Inordertoincreaseresolutionorsamplecapacitybyincreasingthelength,the
totalanalysistimewillalsoincrease.Decreasinganalysistimebyincreasingcarriergasvelocitywillcausethecolumnefficiencytodecrease.
General GuidelinesThelengthofthecolumnshouldbechosenprimarilytoobtainanadequateefficiency. • Shortcolumnsforsamplescreening. • Longercolumnswhenbetterseparationisneeded. • Longcolumnsforcomplexsamplesorcloselyelutingpeaks.
STARTING AdVICE: Length �0 meter
Table1:UsesandTrendsbyColumnLength
�5 m Applications Chromatographic Trends �05 m Applications as Length Increases Rapidanalysis Increasedretentiontimes Lowboilers Highefficiencyseparations Increasedefficiency Morecomplexmixtures Screening Greaterresolution Lessactivesamples Simplemixtures Programmedtemperatureanalysis Highmolecularweightcompounds Morechemicallyactivecomponents
�5 m
L ~ tR
�05 m
General Considerations for Column Selection (continued)
√ L ~ R
√ L ~ SC
5
II. Internal diameter (Id)Thenewdevelopmentsofcapillarycolumntechnology,andspecificallyincolumnID,aretowardinternalsmallerdiameters.Thesedevelopmentsincreasecapillarycolumnapplicability.
A. Influence of internal diameter on resolution. Thesmallertheinternaldiameter,thehighertheefficiency. Therelationbetweenefficiencyandresolutionisasquarerootrelation.Theevolutionofever-
smallerdiametersresultsinextremelyhighefficiencycolumns,ideallysuitedforhighspeedanaly-sisandhighdetectionsensitivity.notethattheinfluenceofinternaldiameteronefficiencyismostimportantwithlargeβvalues(thinfilmcolumns).
B. Influence of internal diameter on sample capacity. narrow-borecolumnshavelowsamplecapacity.
Wide-borecolumnshavehighsamplecapacity. Oneofthedrawbacksofcapillarycolumnscomparedtopackedcolumns
wasreducedsamplecapacity.ThedevelopmentofwideboreFSOT(0.53mmID)gaveanextradimensiontocapillarychromatography.Thewide-borecolumnscombinethegeneraladvantagesofcapillarieswiththehighsamplecapacityofpackedcolumns.Itisnotwithoutreasonthesearereferredtoas“thepackedcolumnalternative”.evenasmallincreaseofinternaldiameter(frome.g.,0.25to0.32mm)resultsinasignificantincreaseinsamplecapacity.
C. lnfluence of internal diameter on analysis time. asageneralrule,thesmallertheinternaldiameter,theshortertheanalysistimewillbeforagiven
resolution.
1. Wide-borecolumnsarelessefficient(lesstheoreticalplates/length).alongercolumnisneededtoobtainthesamecolumnefficiencyandL~tR.
2. Theinternaldiameterdeterminestheoptimalvelocity(uopt)ofyourcolumn.narrow-borecolumnshavehighoptimalvelocitiesandthelossinefficiencyatvelocitieshigherthanuoptisminimal(seeFigure11,p.40).
SC ~ r�
General Considerations for Column Selection (continued)
�
d. Other factors to take into consideration when selecting the internal diameter. Id (mm) Instrument Selection / Constraints 0.10 Highoperatingpressuresareneeded. Mightrequiresystemmodifications. easilyoverloaded,requiresskilledtechnicians. 0.18 Similaroperatingparameterstothe0.25mmID. Increasedefficiencyallowsforshorterruntimes. Goodfirststepwhenlookingtousenarrowborecolumns. Suitedforusewithmassspectrometers. 0.20 Smalllincreaseinefficiencyvs.0.25mmIDcolumns. Suitedformassspectrometers. 0.25 Mostpopulardimension. On-columninjectionpossible. Suitedformassspectrometricdetection. 0.32 easytoworkwith;goodcompromisebetweenresolutionandstability. Routineon-columninjection. Compatiblewithnewermassspectrometers. 0.53 Canbeinstalledinanadaptedpackedgaschromatograph. easyon-columninjection. Goodforusewithdirtyorhighlevelsamples.*Smallercolumnsyieldhigherdetectionsensitivitybutlowersamplecapacity.
STARTING AdVICE: 0.�5 mm Internal diameter Column
Table2:UsesandTrendsbyColumnID 0.�0 mm Applications Chromatographic Trends 0.5� mm Applications as Id Increases
Complexsamples Increasedsamplecapacity LesscomplexsamplesHighresolutionanalysis Decreasedresolvingpower Samplecomponentswithwide Lowerefficiency rangeofconcentrationsLowsampleconcentrations Lowerdetectability HighsampleconcentrationsCanuseonlycapillary Loweroptimalcarriergasvelocities Suitablewitheitherpackedorcolumninlet capillarycolumninlets
General Considerations for Column Selection (continued)
0.�0 mm 0.5� mm
�
III. Film Thickness
A. Influence of film thickness on resolution (see equation) �. Influence on capacity factor (k).
Theretentionofthesolutebythestationaryphaselargelydependsonthetemperature(KD~T)andontheβvalue.Forlowksolutes(volatilecomponents),retentioncanbeincreasedthroughtheβvalue(higherfilmthickness).notethatthegaininresolutionthroughincreasingthecapacityfactorislimited.
�. Influence on efficiency (see equation).Twogeneralrulescanbeapplied:
Thickfilmsarelessefficient.Forthickfilmcolumns(lowβvalue),theresistancetomass transferintheliquidphase(CL)cannolongerbeignored.
non-polarthickfilmcolumnshaveahigherefficiencythanpolarthickfilmcolumns.Thediffu-sivityintheliquidphase(DL)isdeterminedbythenatureofthesolute,thetemperatureandthenatureofthephase.Ingeneral,themorepolarthestationaryphase,themoreDLwilldecrease.Verythickfilmcoatingswithgoodefficiencycanbemadefornon-polarcolumns.Themaximalfilmthicknessofpolarcolumnsisrestrainedduetotheoverwhelminglossinefficiency.
B. Influence on sample capacity.Thickfilmcolumnshavegreatersamplecapacitythanthinfilmcolumns.C. Influence on analysis time.Thethickerthefilm,thelongertheretentiontimeofthesoluteswillbe.Forhighlyvolatilecomponents,thickfilmcolumnsareanecessity.Lessvolatilecomponentsmayrequireverythinfilms(0.1µm)toenablethemtoelutefromthecolumn.
k = = and ~ RKd
β Kd�df
rk
k+�
df ~ SC
df ~ k
General Considerations for Column Selection (continued)
df ~ �/N and √N ~ R
�
d. Other factors to take into consideration. • Foranalyseswherenormallysubambienttemperaturesarerequired,athickfilmcolumnmay allowtheanalysttoworkatambientoventemperatures.However,theboilingpointrangeofthe samplecomponentsshouldnotbetoowide,otherwisethosewiththehighestboilingpointwill notelutefromthecolumn,evenatmaximaloperatingtemperatures. •athickfilmcoatingwillresultinamoreinertcolumn.Thisisespeciallyimportantifthesolute componentshavereactivefunctionalgroups. •Thickfilmshaveahigherbleedandlowermaximaloperatingtemperatures.Thismakesthem lesssuitedtouseincombinationwithmassspectrometricdetectionwhenhighoven temperaturesarerequired.
General Guidelines Thinfilmcolumnsfortheanalysisofhighboilingcompounds. Mediumfilmcolumnsfor: •soluteswithwideboilingpointrange. •mediumboilingcompounds. Thickfilmcolumnsforvolatilecompounds.
STARTING AdVICE: Thin Film: 0.�0 µm medium Film: 0.�5 µm Thick Film: �.0 µm
Table3:UsesandTrendsbyColumnFilmThickness
0.�0 µm Applications Chromatographic Trends 5.0 µm Applications as Film Thickness Increases
Rapidanalysis Increasedretentiontimes Lowboilers,e.g.,gases,solvents, purgeables(BP<25°C)Highefficiencyseparations Increasedresolution UsedmorewithwideIDcolumns (forlowboilingcompounds)HighMWcompounds Uppertemperaturelimitdecreases MorechemicallyactivecomponentsLesschemicallyactive Bleedincreases componentsMSapplications Morenarrowboilingpointrangefor analytes
0.�0 µm
General Considerations for Column Selection (continued)
5.0 µm
�
IV. The Right Phase
A. Influence of the stationary phase on resolution. �. Stationary phase determines selectivity.
Resolutionbetweentwocomponentsismainlydeterminedbytheselectivity(α)ofthestationaryphase.Thisselectivitydependsonthenatureofthecomponents,onthenatureofthestationaryphaseandonthetemperature.
Selectivityisbasedondifferencesininteractionbetweenanalytesandthestationaryphase.Inter-
activeforcesinclude: • Dispersiveforces • Dipole(permanentorinduced)interactions
• acid-baseinteractions• non-polarphasesofferprimarilydispersiveinteractions,andseparationbetweencomponents willbebasedonboilingpoint• Morepolarphases(withdifferentchemicalgroups)willhavemorepossiblewaysofinteraction. Separationisnotonlybasedsolelyonboilingpoint,butalsoduetointeractionsbetween functionalgroups.Retentionwilloccuraccordingtorelativechemicalfunctionality
�. Intrinsic phase characteristics determine the achievable efficiency. Greaterefficiencywillleadtogreaterresolution,providedthatselectivityisavailable.non-polar stationaryphaseswillprovidethehighestefficienciesduetoabetterdiffusivityandahigher coatingefficiency.
B. Influence of stationary phase on sample capacity.Thesamplecapacityforagivensolutealsodependsonthesortofinteractionswhichoccur,andthusonthenatureofthestationaryphaseandofthesolute(i.e.,onanon-polarphase,verypolarcompo-nentswillshoweasyoverloading).
General Considerations for Column Selection (continued)
R ~ α - �α
�0
C. Influence of stationary phase on analysis time. Toobtainacertainresolution,theanalysistimecanbedecreasedwithahigherselectivityandeffi-ciency.asgaschromatographyisbasedonvolatility,makesuretotakethetemperaturerangeofyourcolumnintoconsideration.
d. Other factors to take into consideration when selecting the stationary phase.
�. Non-polar phases offer several advantages: •inertnessandlowbleed:usefulforMSdetection •user-friendlycolumns •widetemperaturerange •moreresistanttotracesofoxygenandwater •widerangeoffilmthicknesses inthecarriergas
�. Only immobilized phases are solvent resistant. �. Gas chromatography is based on volatility. The maximum operating temperature of the
column should be sufficiently high to elute all components.
General Considerations for Column Selection (continued)
��
General Considerations for Column Selection (continued)
TableTable4:PhasePolarity
ZB-1
ZB-1ms
ZB-5
ZB-XLB
ZB-MR-1
ZB-MR-2
ZB-624
ZB-35
ZB-1701
ZB-1701P
ZB-50
ZB-WAX
ZB-FFAP
ZB-5ms
ZB-5MSi
ZB-1HT Inferno
ZB-5HT Inferno
ZB-WAXPLUS
5
LOW-POLARITY MEDIUM-POLARITY HIGH-POLARITY
8 9 11 1513 52 57 5818 19 24
STARTING AdVICE: Non-Polar Phase ZB-5ms or Polar Phase ZB-WAXpluS
��
ZB-1Alternative to any 100 % Dimethylpolysiloxane Phase:
REPLACES
*Thicker films (≥1.0 µm df) are rated to 340/360 °C (Isothermal / TPGC)
Temperature Limits: -�0 to ��0/��0 °C (Isothermal / TPGC)* • Lowpolaritycolumn• Usedfor“fingerprinting”androutinequalitycontrolanalyses(e.g.,citrusoils)• equivalenttoUSPPhaseG2
DB-1
HP-1
MET
-1HP
-101
HP-P
ONA
Rtx-
1Rt
x-1
F&F
Rtx-
1PON
ASP
B-1
CP-S
il 5
CBBP
100
7-1
OV-1
SE-3
0AT
-1Ul
tra 1
DB-1
EVD
XDB
-288
7
ZB-� Applications:
Amines Oxygenates and GRO Drugs of abuse Pesticides
Ethanol PCBs Essential oils Semi-volatiles
Gases (refinery) Simulated distillation Hydrocarbons Sulfur compounds (light)
MTBE Solvent impurities Natural gas odorants Mercaptans
Polarity
Bleed
Temperature Limits
Stability
Low High
Column Profile
��
ZB-1msAlternative to Any MS-Certified 100 % Dimethylpolysiloxane Phase:
REPLACES
Temperature Limits: -�0 to ��0/��0 °C (Isothermal / TPGC) • Loweredbleed(MSCertified)especiallysuitedtohighsensitivityGC/MS• extremelyinertforactivecompoundssuchasdrugsorpesticides• Improvedsignal-to-noiseratioforbettersensitivityandmassspectralintegrity• IdenticalselectivitytotheZB-1• availablewithGuardianIntegratedGuardColumns• equivalenttoUSPPhaseG2
DB-1
ms
HP-1
ms
Rtx-
1MS
VF-1
ms
SolG
el-1
ms
CP-S
il 5
CB m
s
Equi
ty-1
MDN
-1AT
-1m
sColumn Profile
Low High
Polarity
Bleed
Temperature Limits
Stability
ZB-�ms Applications:
Amines Polychlorinated biphenyls (EPA Method 1668)
Pesticides
Acids Drugs of abuse
Diesel Fuel Flavors & fragrances
�4
ZB-1HT Inferno™
Alternative to Any 100 % Dimethylpolysiloxane High-Temperature Phase:
* 0.53 mm ID columns are rated to 400 °C max operational temperature
Temperature Limits: -�0 to 400/4�0 °C (Isothermal / TPGC)* • Firstnon-metal100%dimethylpolysiloxanephasestableto430°C• Individuallytestedforlowbleed,MScertified• Ruggedhightemperature,polyimidecoated,fusedsilicatubing• Providestrueboilingpointseparationforhydrocarbondistillationmethods• Lowactivity,providesgoodpeakshapeforacidicandbasicsamples• Providesrobustcolumnperformanceforhightemperaturebakeouts
ZB-�HT InfernoApplications:
High boiling petroleum products High molecular weight waxes Motor oils
Simulated distillation methods Polymers/plastics
Long-chained hydrocarbons Diesel fuel
DB-1
HT
MXT
-1 S
imDi
stPe
troco
l 288
7CP
-Sim
Dist
REPLACES
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
�00� R&d �00 Award Recipient
�5
ZB-5Alternative to Any 5 %-Phenyl-95 %-Dimethylpolysiloxane Phase:
*Thicker films (≥1.0 µm df) are rated to 340/360 °C (Isothermal / TPGC)
Temperature Limits: -�0 to ��0/��0 °C (Isothermal / TPGC)* • Versatilelowpolaritycolumn• Lowbleed(MSCertified)especiallysuitedtohighsensitivityworkusingGC/MS• extremelyinertforactivecompoundssuchasdrugsorpesticides• Resilienttodirtysamples-longcolumnlife• Greatcolumnforunknownsamples• equivalenttoUSPPhaseG27
DB-5
HP
-5HP
-PAS
-5HP
-5 Tr
ace
An
alys
isRt
x-5
SPB-
5M
DN-5
CP-S
il 8
CB00
7-5
OV-5
SE-5
4AT
-5Eq
uity
-5EC
-5Ul
tra 2
BP5
BPX5
HP-H
P-10
1
ZB-5 Applications:
Alkaloids Residual solvents Drugs PCBs/aroclors
FAMEs Essential oils/flavors Halo-hydrocarbons Phenols
Semi-volatiles Pesticides/herbicides Dioxins
REPLACES
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
��
ZB-5msAlternative to Any MS-Certified 5 %-Phenyl-Arylene-95 %-Dimethylpolysiloxane Phase:
Temperature Limits: -�0 to ��5/�50 °C (Isothermal / TPGC) • aryleneMatrixTechnology(aMT)• Fullyconditionedwithin35minutes• Highresponseforacidsandbases=Verylowactivity• enhancedresolutionofPolyaromanticHydrocarbons(PaHs)andothermulti-ringaromaticcompounds• TheperfectchoiceforePamethods525,610,625,8100,and8270• equivalenttoUSPPhaseG27
DB-5
ms
DB-5
.625
DB-5
ms
EVDX
VF
-5m
sCP
-Sil
8 CB
MS
ZB-5ms Applications:
Alkaloids Essential oils/flavors Pesticides/herbicides Amines
FAMEs Semi-volatiles PCBs/aroclors Acids
Phenols Drugs Solvent impurities Dioxins
Residual solvents Halo-hydrocarbons EPA methods
REPLACES
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
��
ZB-5mSiAlternative to Any 5 %-Phenyl- 95 %-Dimethylpolysiloxane Phase:
REPLACES
Temperature Limits: -�0 to ��0/��0 °C (Isothermal / TPGC) •Lowbleedandhighlyinert5%phenylcolumn•Improvedpeakshapeforacidic/basiccompounds•MaximumsensitivityfromMScertifiedbleedlevels•Reproduciblecolumn-to-columnperformanceinsuredbyindustryleadingQCspecifications•Highphasestabilityandhightemperaturelimits•TraditionalbondingchemistryprovidesthesameselectivityastheZB-5columns
Rtx-
5ms
MDN
-5S
HP-5
ms
Rtx-
5Am
ine
HP-5
msi
Rxi-5
ms
DB-5
ZB-5mSi Applications:
Drugs of abuse FAMEs Pesticides Nitrosamines Phenols EPA methods
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
��
ZB-5HT Inferno™
Alternative to Any 5 %-Phenyl-95 %-Dimethylpolysiloxane High-Temperature Phase:
REPLACES
* 0.53 mm ID columns are rated to 400 °C max operational temperature
Temperature Limits: -�0 to 400/4�0 °C (Isothermal / TPGC)* • Firstnon-metal5%-phenyl95%-dimethylopolysiloxanephasestableto430°C• Individuallytestedforlowbleed,MScertified• Ruggedhightemperature,polyimidecoated,fusedsilicatubing• Lowactivity,providesgoodpeakshapeforacidicandbasicsamples• Providesrobustcolumnperformanceforhightemperaturebakeouts
DB-5
HTHT
-5VF
-5HT
St
x-5H
TXT
I-5HT
ZB-5HT InfernoApplications:
High boiling petroleum products Polymers/plastics Diesel fuel
Simulated distillation methods High molecular weight waxes Motor oils
Long-chained hydrocarbons Triglycerides Surfactants
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
�00� R&d �00 Award Recipient
��
ZB-35Alternative to Any 35 %-Phenyl-65 %-Dimethylpolysiloxane Phase:
REPLACES
Temperature Limits: 50 to �40/��0 °C (Isothermal / TPGC) • Intermediatepolaritycolumnwithtemperaturelimitsupto360°C-allowshighmolecularweightanalysis• excellentinertnesstominimizeanalyteadsorption,improveefficiency,andreproducibility• Morerugged(longercolumnlife)thanotherpolarphases• excellentfortraceanalysiswithbleed-sensitivedetectors(MS,FID,eCD,nPD)• equivalenttoUSPPhaseG42
DB-3
5ms
HP-3
5Rt
x-35
SPB-
35M
DN-3
5BP
X 35
AT-3
500
7-11
OV-1
1Su
p-He
rbSP
B-60
8BP
X608
Rtx-
35M
SDB
-35
HP-3
5ms
EC-3
5
ZB-�5 Applications:
Aroclors Pesticides Semi-volatiles Steroids
Amines Pharmaceuticals Drugs of abuse EPA methods 508, 608, 8081, 8141, 8151
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
�0
ZB-50Alternative to Any 50 %-Phenyl-50 %-Dimethylpolysiloxane Phase:
REPLACES
Temperature Limits: 40 to ��0/�40 °C (Isothermal / TPGC) • Highpolaritycolumnwithtemperaturelimitsupto340°Callowshigh-temperaturebake-outto removecontaminants•excellentinertnesstominimizeanalyteadsorption,improveefficiency,andreproducibility•Morerugged(longercolumnlife)thanotherpolarphases•excellentfortraceanalysiswithbleed-sensitivedetectors•Greatfordrugscreeningandenvironmentalcompounds •equivalenttoUSPPhaseG3
DB-1
7HP
-50+
Rtx-
50DB
-17h
tSP
B-50
SP-2
250
CP-S
il 24
CB
BPX5
0AT
-50
007-
17DB
-17m
sDB
-17
EVDX
SPB-
17
ZB-50 Applications:
Antidepressants Pesticides/herbicides Cholesterols
Drugs of abuse Steroids Triglycerides
Glycols Aroclors EPA methods 508, 608, 8081, 8141, 8151
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
��
ZB-624Alternative to Any 6 %-Cyanopropylphenyl-94 %-Dimethylpolysiloxane Phase:
REPLACES
Temperature Limits: -�0 to ��0 °C • Formulatedforlowbleed• excellentforUSePaMethods501.3,502.2,503.1,524.2,601,602,8010,8015,8020,8240,8260,8021• Specificallydesignedfortheseparationofvolatileorganiccompounds(VOCs)• Increasedtemperaturelimitspeedsruntimesandre-equilibration• Widelyusedphasetoseparatevolatileorganicflavorandfragranceadditivesand residualsolventsinindustrialorpharmaceuticalproducts(OVIs)• equivalenttoUSPPhaseG43
DB-6
24HP
-VOC
Rtx-
624
BP 6
24AT
-624
007-
624
007-
502
CP-S
elec
t 624
CB
DB-V
RXRt
x-VM
SRt
x-13
01DB
-130
1CP
-130
1SP
B-13
01SP
B-62
4
ZB-��4 Applications:
Volitile organic compounds (VOCs) Residual solvents EPA methods 524, 624, 8260
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
��
ZB-1701Alternative to Any 14 %-Cyanopropylphenyl-86 %-Dimethylpolysiloxane Phase:
REPLACES
Temperature Limits: -�0 to ��0/�00 °C (Isothermal/TPGC)*
• Fastrunandre-equilibrationtimesforenhancedsamplethroughputandproductivity• Providesalternateselectivitytophenylphaseswithsimilarpolarity• equivalenttoUSPPhaseG46
DB-1
701
Rtx-
1701
SPB-
1701
CP-S
il 19
CB
OV-1
701
007-
1701
AT-1
701
BP10
Equi
ty 1
701
Rtx-
VMS
*Thicker films (≥1.0 µm df) are rated to 260/280 °C (Isothermal / TPGC)
ZB-��0� Applications:
Pharmaceutical Esters PAHs Solvents
Alcohols Drugs Steroids
Phenols PCBs TMS sugars
Tranquilizers Aromatic hydrocarbons Amines
intermediates
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
��
ZB-1701PAlternative to Any 14 %-Cyanopropylphenyl-86 %-Dimethylpolysiloxane Phase:
REPLACES
Temperature Limits: -�0 to ��0/�00 °C (Isothermal/TPGC)*
• SpeciallytestedtoensureresponseofDDT,endrin,endrinaldehyde,andendrinKetone• Fastrunandre-equilibrationtimesforenhancedsamplethroughputandproductivity• equivalenttoUSPPhaseG46
DB-1
701P
*Thicker films (≥1.0 µm df) are rated to 260/280 °C (Isothermal / TPGC)
ZB-��0�P Applications:
Organochlorine pesticides
Organophosphorous pesticides
Nitrogen containing pesticides
Aroclors
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
�4
ZB-WAXAlternative to Any Polyethylene Glycol Phase:
REPLACES
Temperature Limits: 40 to �50/��0 °C (Isothermal/TPGC)
• Lowbleed,(MSCertified)• Highlystable,longlifetime• Lowactivityforamines• Bonded,solventrinsible• excellentchromatographyofpolarcomplexmixtures• Widelyusedforprofilingand“fingerprinting”• CloseequivalenttoUSPPhaseG16
HP-IN
NOW
axRt
x-W
AXCP
-Wax
57
CB00
7-CW
EC-W
axSt
abilw
ax-D
BSo
lGel
-WAX
FAM
EWAX
DB-W
AXet
rM
et-W
axOm
egaw
axBP
20
ZB-WAX Applications:
Alcohols OVIs Aldehydes Pharmaceuticals
Aromatics Solvents Essential oils Styrene
Flavors/fragrances Xylenes Glycols Basic compounds
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
�5
ZB-WAXplusAlternative to Any Polyethylene Glycol Phase:
REPLACES
Temperature Limits: �0 to �50/��0 °C (Isothermal/TPGC)*
• 100%aqueousstable• extremelyinertforacidiccompounds• enhancedselectivityforlowboilingsolvents• Highretentionofalcoholsandotherchlorinatedsolvents• Increasedefficiencyat20°C• Bondedandsolventrinsible• equivalenttoUSPPhaseG16
DB-W
AXSu
pelc
owax
-10
Carb
owax
-20M
PEG
20M
AT-W
AXPe
rmab
ond
CW 2
0MHP
-20M
CP-W
AX 5
2 CB
CAM
Stab
ilwax
AT-A
quaW
ax B
P20
BP 2
0
ZB-WAXPluS Applications:
Alcoholic beverages Glycols Alcohols Pharmaceuticals
OVIs Aldehydes Solvents Aromatic
Styrene Essential oils Xylene isomers Flavors/fragrances
Acids (free)
*Thicker films (≥1.0 µm df) are rated to 230/240 °C (Isothermal / TPGC)
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
��
ZB-FFAPAlternative to Any Nitroterephthalic Acid Modified Polyethylene Glycol:
REPLACES
Temperature Limits: 40 to �50/��0 °C (Isothermal/TPGC)
• Highpolaritycolumn• especiallysuitedfororganicacids,freefattyacids,andalcohols• excellentthermalandchemicalstability• BondedFFaPPhase• ReplacesOV-351• CloseequivalenttoUSPPhaseG35
DB-F
FAP
HP-F
FAP
BP21
Nuko
l CP
-WAX
58
CB00
7-FF
APSt
abilw
ax-D
APE
-FFA
PAT
-100
0CP
-FFA
P-CB
SPB-
1000
EC-1
000
ZB-FFAP Applications:
Acrylates Ketones Alcohols Volatile free acids
Aldehydes Organic acids Free fatty acids Phenols
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
��
ZB-XLBAlternative to Any 5 %-Phenyl- 95 %-Dimethylpolysiloxane Phase:
SimiLAR To
Temperature Limits: �0 to �40/��0 °C (Isothermal / TPGC) •Uniquelowpolaritysi-arylenecolumn•engineeredspecificallyforusewithbleedsensitivedetectorssuchasMS•Providesalternateselectivitytostandard5-typephases•Oftenusedforconfirmationofpesticides,PCBs,orotherenvironmentalsamples•Goodtoolforsamplescreeningtoidentifyunknowncontaminants
DB-X
LBRt
x-XL
BRt
x-CL
Pest
icid
esSt
x-CL
Pest
icid
es
ZB-5msi Applications:
Polychlorinated biphenyls (PCBs) Pesticides Herbicides
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
*Thicker films (≥1.0 µm df) are rated to 260/280 °C (Isothermal / TPGC)
��
ZB-mULTiRESiDUE™ (mR)-1
SimiLAR* TO
Temperature Limits: -�0 to ��0/�40 °C (Isothermal/TPGC)
• Proprietaryphasespeciallydesignedfortheseparationofalltypesofpesticides,herbicides,and
insecticides• WhenusedinparallelwiththeZebronMR-2column,providesbaselineresolutionandconfirmationofall
20chlorinatedpesticidesregulatedunderePaMethod8081in<10min• MSCertifiedphaseprovideslowbleedperformanceforpesticideconfirmationbyMS• Lowactivity,decreasedbreakdownofsensitivepesticidessuchasDDT• Providesrobustcolumnperformanceforhightemperaturebakeouts
Rtx-
CLPe
stic
ides
Stx-
CLPe
stic
ides
DB-X
LBRt
x-XL
B
ZB-mULTIRESIdUE (mR)-� Applications:
Organochlorine pesticides Nitrogen containing pesticides
Insecticides Multi-pesticide residue methods
Organophosphorous pesticides
Herbicides Aroclors/PCBs Habacetic acids
*notexactequivalent,selectivitymightbedifferent
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
��
ZB-mULTiRESiDUE™ (mR)-2
SimiLAR* TO
Temperature Limits: -�0 to ��0/�40 °C (Isothermal/TPGC)
• Uniqueproprietarypolymerchemistry-unlikeanyothercolumnonthemarket• Speciallydesignedfortheseparationofalltypesofpesticides,herbicides,andinsecticides• WhenusedinparallelwiththeZebronMR-1column,providesbaselineresolutionandconfirmationofall
20chlorinatedpesticidesregulatedunderePaMethod8081in<10min• MSCertifiedphaseprovideslowbleedperformanceforpesticideconfirmationbyMS• Lowactivity,decreasedbreakdownofsensitivepesticidessuchasDDT•Providesrobustcolumnperformanceforhightemperaturebakeouts
Rtx-
CLPe
stic
ides
2St
x-CL
Pest
icid
es 2
*notexactequivalent,selectivitymightbedifferent
ZB-mULTIRESIdUE (mR)-� Applications:
Organochlorine pesticides Nitrogen containing pesticides
Insecticides Multi-pesticide Residue methods
Organophosphorous pesticides
Herbicides Aroclors/PCBs Habacetic acids
Column Profile
Polarity
Bleed
Temperature Limits
Stability
Low High
�0
Please Note:Remembertowearsafetyglasseswhennearanygassystem,cuttingcapillarytubing,andgenerallyasyouworkaroundthelab.
Column Installation
Pre-InstallationCheckList • Replaceoxygen,moistureandhydrocarbontrapsasnecessary.
• Checkgascylinderpressurestoensurethatanadequatesupplyofcarrier,make-upandfuelgasesis available.Carriergasesshouldbeofthehighestpossible(affordable)purity.Note:Itiscriticalthat oxygenandwater,normallypresentingascylinders,beremovedfromthecarriergasbythe appropriateuseoffiltersandadsorbents.
• ensurethattheinjectionportiscleanandfreeofsampleresidues,septumorcapillarydebris.
• Checkandreplaceasnecessarythecriticalinjectorcomponentssuchasseals,linersandsepta.
• Checkandreplacedetectorsealsasnecessary.
• Carefullyinspectyourcolumnfordamageorbreakage.
InstallationToolsandSupplies • Ceramicwaferorsapphire-tippedpencil
• Magnifier(10-20X)
• Ruler
• Marker(e.g.,correctionfluid)
• Wrenches
• Ferrules
• Vialofinjectionsolvent
• Injectionsyringe
• Supplyofanappropriatenon-retainedcompound(e.g.,methane)
• appropriatecolumntestmixture
• Gasflowmeter
• Supplyofreplacementpartsandaccessories e.g.,septa,liners,ferrules,O-rings
��
Thefollowingisabriefreminderofthegeneralprecautionsrequiredinhandlingandinstallinganyorganic-coatedfusedsilicacapillarycolumn.ConsultyourGCmanualformoredetails.
Fusedsilicacapillarycolumnsbecomebrittleifthepolyimidecoatingappliedduringmanufactureisdamaged.avoidtemperaturesaboverecommendedoperatinglimitsandexcessivebending,twisting,andabrasionofcolumns,whichwilldamagethisprotectivecoating.Remember,evenifthecolumndoesnotbreakimmedi-ately,whentheprotectivecoatingisdamagedthecolumnmaybreakspontaneouslylater.
Thestationaryphase,whichcoatstheinsideofthecolumn,mustalsobeprotected.Theendsofthecolumnwillbesealedorprotectedbyaseptumwhenyoureceivethecolumn.Oncetheendsareopeninprepara-tionforinstallation,thecolumnshouldbeinstalledinachromatographassoonaspracticalandaflowofdry,oxygen-freecarriergasmaintaineduntilthecolumnisremovedandresealed.
allforeignmaterialincludingdebrisfromtheseptaorferrulesmustbekeptoutofthecolumn.
Detailed Column Installation Instructions
scorecapillarywithsmoothedgeofwaferata45°angle
applyforceinadownwarddirection
tubingshouldbreakcleanly
inspectcutwithamagnifyingglass
Figure �:CuttingFusedSilicaTubingFigure �:ProperandImproperCutCapillaryend
ba
Examples of: (a) A Properly Cut Fused Silica Column End
(b) An Improperly Cut End
Correct Incorrect4 X
BAD
GOOD
��
1.Installanutandferrule.Cutacentimeter 1.Installanutandferrule.Cutacentimeter ortwooffanendofthecolumn(Figure2). ortwooffanendofthecolumn(Figure2). ensurecutiscleanandsquare(Figure1).Be ensurecutiscleanandsquare(Figure1).Be suretheferruleistherightsizeandthe suretheferruleistherightsizeandthe taperedendistowardtheend. taperedendisfacingthecorrectdirection.
2.MountthecolumnintheGCovenwithout 2.Inserttheoutletendofthecolumnintothe damagingthecolumncoating.(Figure3) detectorexactlythedistanceprescribedin theinstrumentmanual.3.Insertthecolumnintotheinjectorexactly thecorrectdistancespecifiedinthe 3.Tightentheferrulenutuntilthecolumn instrumentmanual(Figure4).Tightentheferrule resistsmovement.One-quarterturnpast nutuntilthecolumnresistsmovement. finger-tightisaboutright.afterthecolumnis One-quarterturnpastfinger-tight equilibrated,injectadetectableunretained isaboutright. samplesuchasmethanetodeterminedead volumetimeandlineargasvelocity.adjust4.adjusttheheadpressuretoobtaintheflow gaspressuretoobtainpropervaluesforyour ratelistedonthetestchromatogram. analyticalmethod.
5.Checktheinletconnectionsforleaks. 4.Themethanepeakmusthaveidealpeak-shape ortheinstallationisfaulty!
Critical Column Installation Steps
Injector Installation: detector Installation:
Figure 4:MeasureproperinjectionportdistanceFigure �:Hangcolumnwithoutdamagingpolyimidecoating
��
Installing the Column
A. Conditioning and Testing the Capillary Column1.Purgethecolumnwithcarriergasforapproximately15minutes.Furtherconditioningmaybe desirable.2.Inserttheoutletendofthecolumnintothedetectorexactly the distance prescribedinthe instrumentmanual.Useallpurposecorrectionfluidtomarktheexact insertiondistance.3.Setgas-flowratestoinstrumentspecifications.Warning! Some detectors may be damaged by heating without proper gas flow.4.Checkthesystemforleaks.Itispreferabletouseathermal-conductivity-typeleakdetector.Donot usesoapsorliquid-basedleakdetectorswithcapillarycolumns.Never heat the column without checking thoroughly for leaks first. 5.Setinjectoranddetectortemperatures.Turnthedetectoronwhensteadystatetemperaturesare achieved.6.Increasetheoventemperaturetothemaximumcontinuousoperatingtemperatureforthecolumn. Warning! do not exceed the maximum operating temperature of the column.Maintainthat temperatureuntilaflatbaselineisobserved.Ifthistakesmorethananhour,itcouldindicatea problem.7.Injectadetectableunretainedsamplesuchasmethane(seeTables11and12)todeterminedead volumetimeandlineargasvelocity.adjustgaspressuretoobtainpropervaluesforyour analyticalmethod.8.Setoventostartingtemperature.Injectanothersampleofadetectableunretainedsubstance.Re setthecarriergasvelocitytodesiredvalue.9.ChecktheperformanceoftheGCandthecolumnbyinjectingaknownsampleorperformancetest mix.Ifallpeakstail,itcouldindicateloosefittings,impropercolumninstallation,orbrokenliner. SeetheSectiononTroubleshooting Installation Problems on page 34.
10.Calibratetheinstrument.11.Injectasample,ensuringthatthevaporizedsamplevolumedoesnotexceedtheinletsleeve’sbuffervolume(seeTables6and9).12.Forshort-termstandbyoperationoftheGCinstrument,continuecarriergasflowat100-200°C. Long-termstandbyconditionsrequirethatthecolumnberemovedfromtheinstrument,flame- sealedorend-cappedwithsepta,andstoredawayfromlightinitsoriginalbox.
�4
Troubleshooting Installation problems
Protecting the ColumnItisimportanttoprotectthecolumnandinstrumentcomponentsfromexposuretodirtysamples.non-volatileorhighmolecularweightcomponentscancontaminatethestationaryphase,causingpeakresolutionandquantitativeaccuracytosignificantlydegrade.Ifpossible,filteryoursamplespriortoinjectionandusepacking(glasswoolorsilanizedfusedsilica)intheinletsleevetoremovesampleresiduebeforeitcanenterthecolumn.Refertoyourinstrumentmanualforspecificinstructionsonpackingtheinletsleeve.
Poorcolumnlifetimeisusuallycausedbynon-volatileorcausticcontaminantsinthesampledamagingthefirst4-6in.ofthecolumn.Cuttingoffthedamagedportionwillusuallyrestorethecolumnperformance,butovertimeperformancewilldegradetoapointwherethecolumncannolongerbeused.Ifyouareexperiencingrapiddegradationofcolumnperformance,thereareseveralsimplewaystohelpprotectyourcolumnandincreaselifetime.
PleasecontactaPhenomenexTechnicalRepresentativeforstepstoimprovecolumnlifetime.
Hints and Tips
Moreoftenthannot,GCcolumnproblemsaretraceabletosomethingimproperlydoneduringinstallation.Foramorecompletetreatmentofthesubject,seeourFREEguidebook,“GCTroubleshooting”.
�5
LinersThelineristhefirstlineofdefenseforthecolumnandthestylechosencanmakeabigdifferenceinhowmuchcontaminationgetsontothecolumn.Theeasiestthingtodoistoaddasmallamountofsilanizedglasswooltoaliner,whichtrapsthenon-volatilecompoundsandpreventsthemfromenteringthecolumn.Glasswoolwillalsohelpcatchpiecesofseptathatresultafterrepeatedinjections.
Caution: glass wool can also add activity for acids, bases, and pesticides. Activity is the result of free silanol (-OH) groups on the surface of the glass wool that form even after it has been deactivated. Crushing the glass wool can lead to increased activity, so it is recommended to purchase pre-packed liners, rather than try to pack your own.
Herearesomelinersthatareavailablepre-packedwithglasswoolorprovideadditionalcolumnprotection:
Hints and Tips (continued)
Description GC Model No.
Dimensions IDxLxOD (mm)
Material*(deactivated)
Quartz Wool(Y / N) Mfr. No. Part No. Unit
Split/Splitless
5880/5890/6890 4 x 78.5 x 6.3 B (y) Y 092002 AG0-7515 5/pk
092219 AG0-7582 25/pk
Split/Splitless, Recessed Gooseneck Liner
5880/5890/6890 1.5 x 78.5 x 6.3 B (y) N 5183-46915183-4692
AG0-4661AG0-4662
5/pk25/pk
Cup Splitter/Split Liner
5880/5890/6890 4 x 78.5 x 6.3 B (n) N 5183-4699 AG0-4647 5/pk
5183-4700 AG0-4648 25/pk
Cup Splitter/Split Liner
Autosystem 3.5 x 100 x 5 B (n) N 0330-5181 AG0-4663 5/pk
Split/Splitless Single Taper/Gooseneck Liner
5880/5890/6890 4 x 78.5 x 6.3 B (y) Y 5183-46935183-4694
AG0-4657AG0-4658
5/pk25/pk
* B = Borosilicate; Deactivated = Yes (y) or No (n)
Guard Columns – Standard GuardsZ-Guardcolumnsare5or10mpiecesofdeactivatedtubingthatareconnectedtoananalyticalcolumnusingaglasspress-fitconnector(Figure5).Thetubingactslikeatrapfornon-volatileresiduesthatwouldotherwisedamagethestationaryphaseofyouranalyticalcolumn.SincethereisnostationaryphaseintheZ-Guard,cuttingthecolumndoesnotsignificantlyaffectretentiontimesorchromatographicseparation.additionalZ-Guardsmaybeattachedasnecessary,aslongastheanalyticalcolumnisstillprovidinggoodseparation.
Sinceguardcolumnscanalsobeasourceofactivity,eachZ-GuardisindividuallyQCtestedtoensureitiscompletelydeactivated.Thisaddedlevelofqualityensuresthatyouwillgetthebestperformancepossibleoutofyourcolumn.
��
Figure 5:Z-Guardconnectionusingpress-fitunion
Guardian™ Integrated Guard ColumnsGuardiancolumnshavethe5mor10mguardbuiltdirectlyintotheanalyticalcolumninonecontinuouslengthoftubing(Figure6).Unliketraditionalguardcolumns,thereisnomechanicalconnectionbetweentheguardandtheanalyticalcolumn.eachGuardiancolumnundergoesaspecialdeactivationandQC-testingprocedurethatensuresgoodperformanceforacids,bases,andothersensitivecompounds.The result: all the benefits of a guard column without the possibility of leaks or activity resulting from a faulty connection.
TheGuardiansystemistheidealsolutiontoalltheproblemsassociatedwithtraditionalguardcolumns!
Figure6.RepresentationofaGuardiancolumn
Hints and Tips (continued)
Column Bake OutTheeasiestwaytoreducecolumncontaminationistoaddashort,hightemperaturebakeoutattheendofthestandardGCmethod.Thisbakeouthelpsremovehighboilingcontaminantsthatwouldotherwiseremaininthecolumnandcausedamage.Ifaddingabakeoutsignificantlyincreasesthemethodruntime,aseparatehightemperaturecleaningprogramcanberunafterevery10injectionsorso.
Thetemperatureusedforbakeouthasadirectimpactontheamountofcontaminantsremoved.Usingcolumnswithhigheruppertemperaturelimitswillincreasetheabilitytoremoveunwantedcontaminants.Forexample,theZebronZB-530meterx0.25mmx0.25μmcolumncansustaintemperatureprogramsupto370ºC,whichis20ºChigherthantheZB-5mscolumnofsimilardimension.Checktheuppertemperaturelimitofthecolumncurrentlyused.Ifahightemperaturebakeoutcouldimprovecolumnlife,considerusingaphasewithincreasedthermalstability.
��
Hints and Tips (continued)
applythisthoughtprocesstothecolumnphasechosenwhendevelopingamethod.Forexample,manypesticideapplicationsaredoneonpolarphasessuchasaZebronZB-1701(14%-cyanopropylphenyl-86%-dimethylpolysiloxane),whichhasanuppertemperaturelimitof300ºC.Insomecases,thesesamemethodscanbedoneonalternatephasessuchastheZebronZB-35(35%-phenyl–65%-dimethylpolysiloxane),whichhasanuppertemperaturelimitof360ºC.
Considerations when Optimizing a GC method: Phase Ratio (β)Whenchangingcolumndimensions,itisimportanttoconsidertheaffectitwillhaveontheretentioncharacteristicsofthecolumn.TheDistributionConstant(K)describestheconcentrationofcompoundainthestationaryvs.thecarriergasmobilephase(equation1).Sinceacompoundisonlymovingwhenithasenteredthecarriergas,changesinthisratioshifttheequilibriumandcanaffectcolumnretentionandselectivityifconditionsdonotchange.
Equation �:K=[am]
=kβ[as]
am=ConcentrationofthesoluteintheMobilePhaseas=ConcentrationofthesoluteintheStationaryPhase
k=capacityfactorβ=PhaseRatio
When looking to optimize column dimensions, it is important to consider phase ratio (β) to ensure thatselectivitywill remain thesame.Phaseratio foragivencolumn iscalculatedusingequation2;smallerβvalues result ingreater retention.Chromatographically thismeans thatwhenusingcolumnsof thesameID,thecolumnwithathickerfilmwillhavegreaterretentionforagivenanalyte.Table5liststheβvaluesforcommonIDsandfilmthicknesses.
Equation �:β=ID
4xdƒ
ID=InternalDiameter(µm)dƒ=FilmThickness(µm)
��
WhenusingcolumnsoftwodifferentIDs,thesamefilmthicknessdoesnottranslatetothesameretentioncharacteristics.Figure5demonstratesthephaseratioofa0.25µmfilmthicknessona0.53mmanda0.25mmIDcolumn.
Figure �:Phaseratioofa0.25µmfilmthicknessoncolumnsofdifferinginternaldiameters
Hints and Tips (continued)
Film Thicknessdƒ (µm)
Column Diameter (mm)
0.10 0.18 0.20 0.25 0.32 0.53
0.10 250 450 500 625 800 1325
0.18 139 250 278 347 444 736
0.25 100 180 200 250 320 530
0.33 — — 151 — — —
0.50 — 90 100 125 160 265
1.00 — — 50 63 80 133
1.50 — — — 42 53 88
3.00 — — — 21 27 44
5.00 — — — 13 16 27
Table5:Phaseratio(β)valueforcommoncolumns
IncreasingRetention
Incr
easi
ngR
eten
tion
kStationary Phase kMobile Phase kStationary Phase kMobile Phase
β=250 β=530 0.25mmID 0.53mmID 0.25µmdƒ 0.25µmdƒ
Using Phase Ratio to our Advantage:Theoptimumphaseratiodependsonthegoaloftheseparation.Ifanalyteretentionislow,acolumnwithalowβcanbeusedtoincreaseretention.Ifcolumnprovidesgoodretention,βcanbereducedtoincreasecolumnefficiencyanddecreaseruntime.
Let’susetheseparationoflighthydrocarbonimpuritiesfoundinbutaneasanexample.OnacolumnwithahighβsuchastheZebronZB-160meterx0.32mmx0.25µm(β=360)theisomersco-eluteduetothelackofinteractionwiththestationaryphase(Figure8).ByusingacolumnwithalowerBetasuchastheZebronZB-160meterx0.32mmx3.00µm(β=27),weareabletoachieveseparation(Figure9).
��
0 1 2
2
1
3
4
3 4 5 6 7 min
15426
Figure �:ButaneIsomerseparationon60meterx0.32mmx0.25µm(β=360)
ap
pID
154
26
Figure �:ButaneIsomerseparationon60meterx0.32mmx3.00µm(β=27)
ap
pID
148
21
Analytes:
1.ethane2.Propane3.Isobutane4.Butane
Analytes:
1.ethane2.Propane3.Isobutane4.Butane
Shortening Run Times:TheincreaseinefficiencyofferedbynarrowboreGCcolumnsoftenimprovesseparationenoughtoallowthesameseparationtobedoneinmuchlesstime.Figure10showstheseparationof17priorityPolyaromaticHydrocarboncontaminantsusingastandard30meterx0.25mmx0.25µmcolumn(β=250).Bychoosingacolumnwithsimilarphaseratio,butsmallerIDthemethodcanbeshortenedbyover50%whilestillmeetingresolutionrequirementsforkeyanalytes(Figures11&12).
40
Figure �0:SeparationofPaHsonZB-5ms30meterx0.25mmx0.25µm(β=250)
ap
pID
160
17
Analytes: (forfigures8-10)
1.naphthalene 10.Benz[a]anthracene2.2-Methylnaphthalene 11.Chrysene3.acenaphthalene 12.Benzo[b]fluoranthene4.acenaphthene 13.Benzo[k]fluoranthene5.Fluorene 14.Benzo[a]pyrene 6.Phenanthrene 15.Indeno[1,2,3-cd]pyrene7.anthracene 16.Dibenz[a,h]anthracene8.Fluoranthene 17.Benzo[g,h,i]perylene9.Pyrene
PhaseratioisacriticalstepinoptimizingGCseparation.Ifyouwouldlikemoreinformationonhowitcanbeusedtoimproveyourchromatography,pleasecontactyourPhenomenexTechnicalConsultant.
Figure ��:SeparationofPaHsonZB-5ms10meterx0.10mmx0.10µm(β=250)
ap
pID
158
06Hints and Tips (continued)
Figure ��:SeparationofPaHsonZB-5ms20meterx0.18mmx0.18µm(β=250)
ap
pID
155
60
4�
Hints and Tips (continued)
Carrier Gas Selection and Flow OptimizationItisadvisabletousethehighestpuritygaspossible.Ultrahighpurity(99.99%),ultrapurecarrier(99.995%),orevenresearchgrade(99.9999)ispreferredtominimizecriticalimpurities,instrumentdowntimeandtroubleshooting.air,moistureandorganictrapsshouldbeused,butitisbettertostartwiththehighestpuritygasandreducetheloadongaspurifiersasmuchaspossible.
HeliumshouldbeusedforcapillaryGCwheneverpossible;nitrogenshowsinferiorperformanceduetoslowoptimumlinearvelocityandsteepvanDeemterprofile.
Threetypesofgasarecommonlyusedasacarriergas: 1.Hydrogen(H2):Hydrogenwillyieldmaximalnumberoftheoreticalplatesforthinfilmcolumnsand thehighefficiencyislargelyretainedatvelocitieshigherthanuopt.Hydrogenisnotgenerallyrecom- mendedduetoitshazardousnature. 2. Helium(He):Whenhydrogenisnotused,heliumisthebestalternativeforspeedandsensitivity. 3. nitrogen(n2):nitrogenisthelastchoiceforthinfilmcolumns.Forthickfilmcolumns,nitrogen yieldsthehighestnumberoftheoreticalplates.However,theoptimalvelocityisfairlylow(long analysistimes),andthelossinefficiencyathighervelocitiesishigh.Ifresolutionissufficient, hydrogenorheliumaregoodalternatives.
Figure ��:CarrierGasSelectionandVelocityOptimizationPlots
H (u) curves for H(u) curves for H(u) curves for H(u) curves for different gases different column different film different k values lengths thicknesses
Optimal velocities uopt Optimal velocities uopt for low df values: (cm/sec) for high df values: (cm/sec) H2: 40 H2: 25 He: 25 He: 15 n2: 10 n2: 7
CarrierGas Length FilmThickness RetentionFactorAnalytes: (forfigures8-10)
1.naphthalene 10.Benz[a]anthracene2.2-Methylnaphthalene 11.Chrysene3.acenaphthalene 12.Benzo[b]fluoranthene4.acenaphthene 13.Benzo[k]fluoranthene5.Fluorene 14.Benzo[a]pyrene 6.Phenanthrene 15.Indeno[1,2,3-cd]pyrene7.anthracene 16.Dibenz[a,h]anthracene8.Fluoranthene 17.Benzo[g,h,i]perylene9.Pyrene
Figure ��:SeparationofPaHsonZB-5ms20meterx0.18mmx0.18µm(β=250)
4�
Hints and Tips (continued)
Figure �4:InletPressurevs.Velocity
Thevelocitydiagramsbelowshowtherelationbetweeninletpressureandvelocityfordifferentinternaldiametersandlengths,measuredforhydrogenat120°Cisothermalandsplitlessinjection(approximatevalues).
Temperature ProgrammingTherearenostrictrulestodeterminetheoptimaloventemperatureofyouranalysis.Suggestionscanbefoundinliteraturedataandinthisguidebook.experienceandtrialanderrorareusuallythemostvaluabletools.
General considerations: 1.IsothermalTemperature:forseparationofcomponentswithslightlydifferingboilingpoints. 2.Temperatureprogramming(singleormultistep):forseparationofsampleswithwideboiling pointrange.
Checking for Leaks Useathermoconductivitydetectortocheckforleaks.ItishighlysensitivetoH2,Heanditwon’tcontaminatetheinstrumentorcolumn.Liquidleakindicatorsarenotrecommendedforcapillarycolumnsbecausethereisalwaystheriskofdrawingtheliquidintothecolumnorcolumnfittingsandcontaminatingthesystem.
Note:IfgraphitizedVespelferrulesarebeingused,leakagecanoccuraftertheinitialheatingphaseduetoferruleshrinkageand/ordeformation.Besurethatthefittingisre-tightenedafterthisinitialheatingphasethencarefullycheckedforleaks.Betteryet,alsopre-conditionvespel-containingferrulesinanovenat250°Cforatleast4hourspriortouse.
4�
Hints and Tips (continued)
Injection Techniques
Table6:InjectionModesandSelectedSpecification
Table7:GasesusedwithCommonDetectors
detector Carrier Gas Fuel Gas make-up GaseCD n2,ar/5%CH4 none n2,ar/5%CH4
eCD H2 none ar/5%CH4
eLCD,Hall He,H2 H2 noneFID He,H2,n2 air+H2 n2,He,H2
FPD n2,He air+H2 SameascarrierHID He none HenPD He,n2,H2, air+H2 HePID He,H2,n2 none n2,HeTCD He,H2 none Sameascarrier
Parameter Split Splitless direct On-Column
ColumnID(mm) 0.10-0.53 0.18-0.53 0.53 0.18-0.53*
InjectionTemp(°C) High250-300
Moderate200-250
Moderate200-250
Low(15°Cbelowbpofsolvent)
VentorPurge Continuous afterinitialtime none none
TypicalSampleSize(µL) 1-2 2-4 2-4 1-4
ConcentrationRange(ppm)
10-1000 0.01-100 0.1-100 0.01-100
* Special needles required
44
Hints and Tips (continued)
Table9:expansionVolumesofCommonSolvents
Injection Volume (µL) H�O (µl) CS2 (µL) CH�Cl� (µL) Hexane (µL) 0.1 142 42 40 20 0.5 710 212 200 98 1.0 1420 423 401 195 2.0 2840 846 802 390 3.0 4260 1270 1200 585 4.0 5680 1690 1600 780 5.0 7100 2120 2000 975
Calculating Split Ratio and Column Flow Rate:
Split Ratio=splitventflow(mL/min)+columnflow(mL/min)/columnflow(mL/min)
Flow(mL/min)=(π)columnradius(cm)2columnlength(cm)/deadvolumetime(min)
Average Linear Velocity u(cm/sec)=columnlength(cm)/retentiontimeofnon-retainedpeak(sec)
Column Length (m) Id (mm) Column Head Pressure (psig)
15 0.25 6
30 0.25 12
60 0.25 24
Gas Flow Settings Table8:TypicalHeadpressures(forHelium)
45
Sample and solvent expansion volume = nRT / P
Where: n = numberofmolesofsolventandsample,calculatedas: volume(mL)xdensity(g/mL)/MW(g/mole) T= absolutetemperatureoftheinjector(°K) P= columnheadpressure(atm)+1atm R= Universalgasconstant
Volume of inlet sleeve = πr�L
Where: r = radius(cm) L= length(cm)
Hints and Tips (continued)
Table10:SplitlessHoldTimes
Column Hold Time (min) Column Flow Sample Transfer Id (mm) Rate (mL/min) Time (min) He H� He H�
0.25 1.0 0.71.4 1.20.6 0.32 0.75 1.22.4 0.70.4 0.53 0.50 2.65.2 0.30.2
1. From a disposable lighter2. Place 1-2 drops in an autosampler vial and tightly cap. Shake and inject 1-2 μL from the headspace of the vial. Do not inject any liquid.
3. Use a column temperature above 55 ºC.4. Use a column temperature above 95 ºC.
Column dead Times and markers
Table11:RecommendedDeadVolumeTimeMarkers
detector Type marker CompoundeCD Methylenechloride2,3,DichlorodifluoromethaneFID Methane,Butane1
nPD acetonitrile2,4
PIDeLCD VinylchlorideTCD,MS Methane,Butane1,air
4�
Table12:RecommendedMethaneRetentionTimes
Length (m) H� (sec) He (sec) N� (sec)15 38 75 15030 75 150 300
60 150 300 600
examinetheshapeofthedead Volume Peak or Solvent Peak.acorrectcolumninstallationwillyieldasharpnarrowsymmetricalpeak.anytailingordistortioninpeakshapeindicatesafaultyinstallation.
Figure �5:DeadVolumePeakShapeTest
Ifthepeakisbroadand/ortailing,checkthefollowingpotentialproblemareas:
• Impropercolumnpositioning/insertionintoinletordetector • Grosscontaminationofthesplittersleeve • Chippedorcrackedsplittersleever • Impropersweepingofcolumnendbymakeupgas • Damagedorcrushedcolumnend • Obstructedcolumn
Sample Capacity
Table13:TypicalSampleCapacity(maxforsinglecomponent)
Id (mm) df (µm) Capacity (ng) 0.18 0.18 20-75 0.25 0.25 50-125 0.32 0.50 100-250 0.53 1.0 500-1,000
Note: Capacityrepresentsmaximumloadingpercomponent.Samplecapacityincreaseswithfilmthickness.
Hints and Tips(continued)
Tailing peak indicatesimproper installation
Symmetrical peak indicatesproper installation
methane with FId / TCd. Calculateaveragelinearvelocitybyinjecting25-100µLof1%methaneinn2gasblend.Measuretheretentiontimeofthemethanepeakandcalculatethefollowing:
Average Linear Velocity (u) = L/tO
4�
Hints and Tips (continued)
Solvent Rinsing of Zebron Capillary ColumnsSelecttherinsesolventfromtheTablebelow.
Table14:PhaseCompatibilitywithRinsingSolvents
Table15:RinsingConditions
Column Id (mm) Rinse Solvent Volume (mL) Pressure (psig) 0.25 5 40 0.32 5 40 0.53 10 20
Important:Rinsefromthebacktothefrontofthecolumntoavoidpushinginletcontaminantsfurtherintothecolumn.
Phase Water methanol CH�Cl� CHCl� Acetone Hexane
ZB-1
ZB-1ms
ZB-1HTInferno
ZB-5
ZB-5ms
ZB-5HTInferno
ZB-624
ZB-35
ZB-1701
ZB-1701P
ZB-50
ZB-WaXplus aVOID
ZB-WaX aVOID
ZB-FFaP aVOID
MR-1
MR-2
OK
4�
Chemical CompatibilityImportant! WaterandorganicsolventssuchasthoselistedinthetableabovewillnotdamageZebroncolumnstationaryphases.However,inorganicacidsandbasesshouldbecompletelyavoidedorrapiddegradationandpermanentdamagetothestationaryphasewillresult.Intheeventchemicaldamageisincurred,theremovalof0.5-2mofcapillaryoffthefrontendwilloftenrestorecolumnperformance.
“Baking Out” the ColumnColumncontaminationanddegradationinanalyticalperformancecanoccurifthehighestboilersarenotelutedwitheveryrun.Thefinaloventemperatureneedstobesethighenoughtoensureelutionofthesecompounds,butnotsohighastocausethermaldamage.Thiscanbedoneeitherisothermally,ormorecommonly,viaagradientorballisticincreaseuntilthelastcomponentselutefromthecolumn.
NEVER exceed the upper temperature limits of the column.Severedegradationandlossofsta-tionaryphase,aswellaspermanentdamagetothetubingsurface,mayresult.Chromatographicallythismaymanifestitselfinexcessivecolumnbleed,peaktailing,decreasedresolution,shortenedruntimes,reducedcolumnlifetimesorevencolumnfailure.Topreventaccidentaloverheatingandthermaldamagetothecolumn,settheoven’smaximumtemperatureatorslightlybelowthecolumn’suppertemperaturelimit.
extractedsamplesoftencauseabuildupofcontaminantsthatrequireamoreconcerted“bakingout”,althoughthistechniqueshouldbeusedcarefullyandsparingly.DO NOT exceed more than 15 min-utes at the upper isothermal temperature limit specified for the column.Bakingoutacontami-natedcolumnmaycausesomesampleresiduestobeconvertedtoinsolublematerialsthatcannolongerberemoved,evenbysolventrinsing.Thecolumnmaybeirreversiblydamagedasaresult.Thebestwaytoguardagainstcolumncontaminationanddegradedanalyticalperformanceistointroduceonlysamplesthathavefirstbeenthoroughlyandcarefullyextractedandfiltered.Beforebakingouttoremovecontaminants,trysolventrinsingthecolumnfirst.
Column StorageImportant! Thecolumnmaybeleftintheinstrumentforshort-termstorage.ensureaflowofcarriergasthroughthecolumnat100-200°C.Forlong-termstorage,disconnectfromtheGCandcaporsealthecapillaryends.Oxygenandmoisturecandegradeorirreversiblydamagethecolumn,especiallycyanopropyl-basedphases.Wax(polyethyleneglycol)andcyanopropyl-basedphasesarealsosus-ceptibletoUV-induceddegradationandshouldbeshieldedfromlight(fluorescentorsunlight).Storethecolumnintheoriginalbox.Uponreinstallation,cutcolumnendstoensurethatseptumfragmentsorotherdebrishavenotbeenleftinthecolumn.
Hints and Tips (continued)
4�
References
Baugh,P.,Gas Chromatography: A Practical Approach,Oxford,1994. Blau,K.,andHalket,J.M.eds.,High Resolution Gas Chromatography,3rdedition,Hewlett-Packard,
1989. Braithwaite,a.,andSmith,F.J.,Chromatographic methods,5thedition,Chapman&Hall,1996. Fowlis,l.,Gas Chromatography,2ndedition,JohnWileyandSons,1995. Grant,D.W.,Capillary GC,JohnWileyandSons,1996. Grob,K.,Split and Splitless Injection in GC,3rdedition,Huthig,1993. Grob,K.,On-Column Injection In Capillary Gas Chromatography,2ndedition,Huthig,1991. Grob,R.L.,Chromatographic Analysis of the Environment,MarcelDekker,1983. Grob,R.L.,modern Practice of Gas Chromatography,3rdedition,JohnWileyandSons,1995. ThisbookisavailablefromPhenomenex,Part No. AA0-4455. Hill,H.H.,ed.,detectors for Capillary Chromatography,JohnWileyandSons,1992. Hyver,K.J.,andSandra,P.eds.,High Resolution Gas Chromatography,3rdedition,Hewlett-Packard,
1989. Ioffe,B.V.andVitenberg,a.G.,Head-Space Analysis and Related methods In Gas Chromatography,
JohnWileyandSons,1984. Jennings,W.,etal.,Analytical Gas Chromatography,2ndedition,academicPress,1997. Jennings,W.G.,andRapp,a.,Sample Preparation for Gas Chromatographic Analysis,Huthig,1983. Katz,e.,Quantitative Analysis Using Chromatographic Techniques,JohnWileyandSons,1987. Kitson,G.G.,Larsen,B.S.,andMcewen,C.n.,Gas Chromatography and mass Spectrometry. A
Practical Guide,academicPress,1996. Mcnair,H.M.andMiller,J.M.,Basic Gas Chromatography,1997. ThisbookisavailablefromPhenomenex,Part No. AA0-4454. Poole,C.F.,andPoole,S.K.,Chromatography Today,elsevierSciencePublishing,1991. Robards,K.,etal.,Principles and Practice of modern Chromatographic methods,academicPress,
1995. Rood,D.,A Practical Guide to the Care, maintenance, and Troubleshooting of Capillary Gas Chromatographic Systems,2ndedition,Huthig,1995. Schomburg,G., Gas Chromatography,VCH,1990. Scott,R.P.W.,Introduction to Analytical Gas Chromatography,2ndedition,MarcelDekker,1997. Scott,R.P.W.,Techniques and Practices of Chromatography,MarcelDekker,1995. Sonchik,S.M.,J. Chrom. Sci.,17,277,1979. Unger,K.K.,Packings and Stationary Phases In Chromatographic Techniques,Chrom.Sci.Series,
Vol.47,MarcelDekker,1990.
Top Recommendations
50
Adsorption mode:Chromatographyinwhichthestationaryphaseisasolid.Separationoccursthroughmechanismsofadsorptionandresorption.Band broadening:extracolumnbandbroadeningisduetoalossinefficiencythroughthechromatographicsystem.allpartsofachromatographicsystemcancausebandbroadening.Baseline:Detectionofthemobilephase(i.e.carriergas).Baseline:anysignalnotresultingformanalytes.Baseline noise:Lowlevel,highfrequencysignalssuperimposedonthebaselinesignal,duetocolumnbleed,impuritiesininjectoranddetector,etc.Bleed:Decompositionandvaporizationofthestationaryphase.Mostapparentwhenusingthecolumnnearitsmaximalallowableoperatingtemperature.Capacity:SeeSamplecapacity.Capacity factor:anumberwhichindicatestherelativetimeasolutespendsintheliquidphasewithregardtothetimeitspendsinthemobilephase.Capillary columns:narrow(from0.050to0.530mminternaldiameter),long(from10to100m)chromatographiccolumns,offeringhighefficiencies.CGC:CapillarygaschromatographyChemical bonded:Termwhichindicatesachemicalreactionofthestationaryphasewiththecapillarywall,renderingthecolumnsolvent-resistant.See“Immobilized”.Christmas tree effect:aChristmastree-likeshapeofasolutepeakduetouneventemperaturedistributionintheGCsystem.Coating efficiency (CE):expressesthepercentageofthemeasuredefficiencyovertheefficiencytheoreticallypossible.Coelution:Thesimultaneouselutionoftwocomponents.Crosslinked:Interlinkingofthestationaryphasetoobtainastableandsolvent-resistantfilm.See“Immobilized”.dead volume:Thevolumeofacapillarycolumn.Theretentiontimeofanunretainedsolute(tM)isusedtoconvertcarriergasflowintocarriergasvelocity.distribution constant(Kd): KDdescribestheequilibriumbetweentheconcentrationofasoluteinthestationaryphaseandinthemobilephase.
Glossary
ECd:electronCaptureDetection.Effective plate number:Thenumberofeffectiveplates.Efficiency:atermindicatingtheabilityofthecolumntoeluteacomponentinanarrowchromatographicpeak.Columnefficiencyisdescribedby‘n’and‘H’;seeequations.Elution:Thetransferofacomponentthroughthecolumn.FId:FlameIonizationDetection.Film thickness:Thethicknessofthestationaryphasecoatedontothecolumnwall.FSOT:FusedSilicaOpenTubular.FTIR:FourierTransformInfra-Redspectroscopicdetection.Fused silica:Syntheticpolymerofsilicondioxide.Duetoitshighpurity,fusedsilicahasreplacedglassasmaterialforcapillarycolumns.Golay equation:Chromatographicequation,describingthecolumnefficiencyincapillarygaschromatography.Height equivalent to a theoretical plate (H):Thelengthofthecolumnoccupiedbyonetheoreticalplate.ThesmallerH,thehighertheplatenumberforagivenlength.Seeequation.Immobilized:Theimmobilizationofthephaseduetocross-linkingand/orchemicalbonding.Inertness:Theabsenceofactivegroupsinthecolumnwhichwouldreactwithpolarsolutes,resultingintailingpeaks.Liquid phase:Thestationeryphase,coatedontoacolumnwhichisinaliquidstateatoperatingtemperatures.mAOT:MaximalallowableOperatingTemperatureofthecolumn.mcReynolds constants:aclassifyingsystemforthepolarityandselectivityofstationaryphases.mobile phase:Themediumusedtocarrythesolutesthroughthecolumn.IncapillaryGCthreegasesareappliedasmobilephase:hydrogen,heliumandnitrogen.Non-polar:Theabsenceofpolargroupsinthestationaryphase.On column injection:Injectiontechniqueincapillarygaschromatography.Thesampleisinjecteddirectlyintothecolumn.
5�
Overloading:exceedingthesamplecapacityforasoluteonagivencolumn,resultinginasymmetricalpeakshapesandvariableretentiontimes.Phase ratio:numberwhichindicatestheratiobetweenthemobilephase(columnradius)andthestationaryphase(filmthickness).PLOT:PorousLayerOpenTubular:anadsorbentsuchasaluminiumoxideisdepositedontheinnercolumnwall,increasingthecontactsurface.Polar:Thepresenceofpolargroupsinthestationaryphase.Polarity:Stationaryphasesareoftenclassifiedaccordingtotheirdegreeofpolarity,indicatingthewaysofinteractioninachromatographicseparation.Polyimide:Theouterpolymercoatingoffusedsilicatubinggivingitstrengthandflexibility.Resolution:ameasureoftheabilityofacolumntoseparatetwocomponents.Seeequation.Retention:Interactionbetweenstationaryphaseandsolutewillretainthesoluteinthecolumn.Thedegreeoftheinteractionswilldeterminethestrengthoftheretention.Retention time:Thetimeneededforasolutetoelutefromthecolumn,startingfromtheinjection.Totalretentiontimeisthesumofthetimethesolutespendsinthemobilephaseandthetimeitspendsintheliquidphase (to=adjustedretentiontime).Sample capacity:Maximumsamplesizeofacompoundwhichcanbeintroducedontoacapillarycolumncoatedwithaparticularstationaryphasewithoutdisturbingthechromatographicperformanceofthecolumn.Seeequation.SCOT:SupportCoatedOpenTubularcolumns:Capillarycolumnswheretheliquidstationaryphaseiscoatedontoafinesolidsupport.Selectivity:Thedegreebywhichthestationaryphasedifferentiatessolutes.SeparationinCGCisbasedondifferentselectivitiesforcomponentspresentinasample.Selectivity factor:anumberdescribingtheselectivityofthestationaryphasefortwochemicals,relativetoeachother.aselectivityfactorgreaterthanoneisrequiredto
achieveseparation.Seeequation.Solid State Injection:InjectiontechniqueinCGCwherethesolventisvaporizedpriortoinjectionintothecolumn.Solutes:Thecomponentspresentinasample.Split-splitless injection:InjectiontechniqueinCGCwherethesampleisinjectedintoaheatedchamber,vaporizedandthencarriedbythemobilephaseintothecolumn.aftervaporization,thesamplecanbepartiallyvented(splitmode)ornot(splitlessmode).Stabilized:anuncrosslinkablephasecanbedissolvedfromthecolumnwhenrinsedwithasolvent.Toavoidbreakdownofthefilm,astabilizingagentisadded.Stationary phase:Polymerwhichiscoatedontotheinnercolumnwalltoprovidetheselectivityforseparation.Tailing:Thetail-likeshapeofachromatographicpeak,causedbyactivesitesinthecolumn.TCd:ThermalConductivityDetection.Theoretical plate:Therepresentationofaunitofefficiencyofacolumn.TPGC:TemperatureProgrammedGasChromatography.Trennzahl:Theresolutionbetweentwoconsecutivemembersofthee-paraffinhomologousseries.Seeequation.van deemter equation:Thebasicequationdescribingtherateofbandbroadeningofacolumningaschromatography.ThisequationwasadaptedtocapillaryGCbyGolay(see“Golayequation”).WCOT:WallCoatedOpenTubularcolumns:Capillarycolumnswherethestationaryphaseiscoatedontothecolumnwallasasmoothuniformfilm.
Phenomenexfor useful information
on capillary gas chromatography
Glossary
5�
�. Resolution (R)
• Graphically:
R=
• Chromatographically:
R=
• Trennzahl (TZ):
TZ=-1
�. Efficiency (N and H)
• N = number of theoretical plates
n=5.54 =16 n=
• H = height equivalent to a theoretical plate
Golayequation:
H=B/u+CGu+DLu
u=L/tO
H=+u+u
////
//
//
////
tm (to)
H(u) diagram (Golay plot)
LH
(1+6k+11K)2r2
24(1+k)2DG
2kdf2
3(1+k)2DL
////
//
//
////
tm (to)
u(cm/s)
Equations
√n4
α − 1α
kk+1( )()
2∆tR
Wb1+Wb2
∆tR
Wh1+Wh2
()tR
Wh
2()tR
Wb
2
2DG
u
5�
Equations (continued)
�. Effective plate number 4. Coating efficiency
neff=16 Ce%=()100
5. Selectivity factor (α) �. Capacity factor (Retention Factor)
α= k==
�. distribution constant �. Phase ratio
�. Sample capacity �0. minimal analysis time for a two component mixture
��. Split Ratio=splitventflow+columnflow/columnflow(mL/min)
��. Flow(mL/min)=(π)(columnradius(cm)2(columnlength(cm)/deadvolumetime(min)
��. Average Linear Velocity u(cm/sec)=columnlength(cm)/retentiontimeofnon-retained peak(sec)
�4. Sample and solvent expansion volume = nRT/P Where:n =numberofmolesofsolventandsample,calculatedas:
volume(mL)xdensity(g/mL)/MW(g/mole)T=absolutetemperatureoftheinjector(°K)
P=columnheadpressure(atm)+1atm R =Universalgasconstant
�5. Volume of inlet sleeve = πr�L Where: r=radius(cm) L=length(cm) ��. Average Linear Velocity (u)= L/tO
tR’2
tR’1
tR’=tR-tO
tR2>tR1
()tR’Wb
2
KD=k•β
2dfKd
r
SC~r2(1+k) √LH
~r2(1+)√LH
tR-tO
tO
tR’tO
r2df
β =
nexp
ntheor
α(α-1)
(1+k)3
k2tne=16()()(R2)Hu
2
54
List of symbols and abbreviations
α = Selectivityfactor
Β = Longitudinaldiffusionterm
β = Phaseratio
Ce = Coatingefficiency
CG = Resistancetomasstransferinthegasphase
CL = Resistancetomasstransferintheliquidphase
df = Filmthickness
DG = Diffusioncoefficientinthegasphase
DL = Diffusioncoefficientintheliquidphase
H = Heightequivalenttoatheoreticalplate
k = Capacityfactor
KD = Distributionconstant
L = Columnlength
Equations (continued)
u = averagelineargasvelocity
u = Lineargasvelocity
n = numberoftheoreticalplates
n = effectiveplatenumber
R = Resolution
r = Columnradius
SC = Samplecapacity
tO = Retentiontimeofunretainedsolute
tR = Retentiontimeofthesolute
tR’ = adjustedretentiontimeofthesolute
wb = Peakwidthatbase
wh = Peakwidthathalfheight
/
55
NOTE:
Whileeveryattempthasbeenmadetoensuretheaccuracyoftheinformation
containedinthisguide,Phenomenexassumesnoresponsibilityforitsuse.
Wewelcomeanyadditionsorcorrectionsforincorporationintofutureeditions.
Zebron,Inferno,MultiResidue,andengineeredSelfCross-linking(eSC)aretrademarksofPhenomenex,Inc.©2007PhenomenxInc.allrightsreserved.
5�
mail:
tel.:fax:
email:
mail:
tel.:fax:
email:
www.phenomenex.comPhenomenex products are available worldwide. For the distributor in your country, contact Phenomenex USA, International Department by telephone, fax or email: [email protected].
New ZealandP O Box 31-601Milford 0741North Shore CityNew Zealand
IrelandQueens Avenue,Hurdsfield Ind. Est.,Macclesfield, Cheshire SK10 2BN, UK
01 247 5405+44 [email protected]
DenmarkGydevang 39-413450 AllerødDenmark
4824 80484810 [email protected]
United KingdomQueens Avenue,Hurdsfield Ind. Est.,Macclesfield, Cheshire SK10 2BN, UK
FranceParc des Grillons, Bat.360 route de Sartrouville78232 Le Pecq CedexFrance
01 30 09 21 1001 30 09 21 [email protected]
USA411 Madrid Ave.Torrance, CA90501-1430USA
(310) 212-0555(310) [email protected]
AustriaZeppelinstr. 563741 AschaffenburgGermany
ItalyVia Serenari, 51/D40013 Castel Maggiore (BO)Italy
051 736176051 [email protected]
Canada411 Madrid Ave.Torrance, CA90501-1430USA
(800) 543-3681(310) [email protected]
Puerto Rico273 Sierra Morena,Suite #104San Juan,Puerto Rico 00926
(800) 541-HPLC(310) [email protected]
GermanyZeppelinstr. 563741 AschaffenburgGermany
AustraliaPO Box 4084Lane Cove, NSW 2066Australia
5549
_L