    TG-900 series

    CO 2 Sensor Kitfor mainstream measurement

    CO 2 measurement The best way to observe the patient's respiratory condition

    F a s t r e s p o n s e w i t h n o l a g t i m e

    N o w a r m - u p t i m e

    N o h e a t e r o r m o t o r

    N o s a m p l i n g t u b e

    C o m p a c t a n d l i g h t w e i g h t

  • 7/26/2019 Co2_sensor-Kit_p900_05 Nihon Kohden Bsm 2301 k

    2/4

    Anti-foggingmembrane

    Normal membrane

    *Both signals and reference are affected by the mist of the nebulizer butCO 2 measurement is not affected.

    CO 2 (mmHg)

    Nebulizer OFF Nebulizer ON

    Sensor output (v)

    60.040.0

    20.0

    0

    3.00

    2.00

    1.00

    060sec

    0 Signals output (Vs)Reference output (Vr)

    S m a r t a d a p t e r w i t ha n t i - f o g g i n g m e m b r a n e

    Window is outsidethe airway

    Water collects

    Anti-fogging membrane isinside the airway

    (Nihon Kohden)

    No water collects

    The TG-950P/970P sensor using quantitative method with anti-foggingmembrane adapter assures accurate measurementin difficult measuring conditions.

    BSM-6701K

    CO2 sensor (TG-970P)

    Actual size

    The TG-950P/970P always provides accurate measurement duringlow flow respiration and oxygen administration.

    OR

    Pediatric

    With Ventilator and Nebulizer

    - Nihon Kohden's anti-fogging membrane makes awater layer which produces even light transmission.There is no need to warm up the window to make thewater layer.

    - Typical membranes cause water drops to collect and produce uneven light reflection and refraction.

    - With no heater and no motor, the sensor is much moredurable (withstands a 2 m drop).

  • 7/26/2019 Co2_sensor-Kit_p900_05 Nihon Kohden Bsm 2301 k

    3/4

    CO2 sensor (TG-920P)

    Mainstream measurement does not require a sampling tube.This eliminates any problems of tube occlusion or contamination.Mainstream is suitable for long-time measurement.

    Small and lightweight CO 2 sensor reduces stress tothe patient so more accurate sleep analysis

    can be perfomed.

    Connect the CO 2 sensor to a defibrillator for airway management.No warm-up time and no calibration are required so you can obtain CO 2 quickly.

    EEG-1200J/K with JE-921A input box

    PVM-2703K

    TEC-8300K

    ER

    PSG

    ICUSedation

    This unique designadapter catchesboth nasal/oralexpiration.

    Comparison of mainstream (cap-ONE) and sidestreamSidestream causes false alarms

    CO 2 (mmHg)

    40

    0

    Actual size

    Nasal/oral adapter YG-121T YG-122T

  • 7/26/2019 Co2_sensor-Kit_p900_05 Nihon Kohden Bsm 2301 k

    4/4

    This brochure may be revised or replaced by Nihon Kohden at any time without notice.

    6865 CAT.No.55-076C 11.10. CLB. C.

    CO 2 Sensor Kits and Adapters Line-up Mainstream Measurement

    Printed in Japan

    cap-ONE is a registered trademark of Nihon Kohden Corporation.

    CO 2 sensor kit Adapter (kg = patient weight) Connector and applicable units

    Semi-quantitative method, intubated and non-intubated

    P907 (TG-920P)

    Intubated - Airway adapter

    R804 (YG-111T) > 7 kg

    Non-intubated - Nasal/oral adapter

    V922 (YG-121T) > 10 kg V923 (YG-122T*) > 10 kg *with oxygen supply tube adapter

    V928 (YG-125T**) > 10 kg V929 (YG-135T**) > 7 kg **for PSG (EEG),

    with airflow pressure sensor tube

    MULTI connector

    BSM series bedside monitors

    TEC-8300 defibrillator

    OLG-2800K CO 2 monitor

    P908 (TG-921T3)

    Mini DIN connector

    TEC-5500/7700 series defibrillators

    EEG-1200/9100 electroencephalographs

    Semi-quantitative method, intubated

    P903 (TG-900P)

    Intubated - Airway adapter

    R801 (YG-101T) > 10 kg

    MULTI connector

    BSM-series bedside monitors

    TEC-8300 defibrillator

    OLG-2800K CO 2 monitor

    P906 (TG-901T3)

    Mini DIN connector

    TEC-5500/7700 series defibrillators

    Quantitative method, intubated

    P909 (TG-970P)

    Intubated - Airway adapter

    R805 (YG-211T) > 7 kg R806 (YG-213T) 2 to 7 kg MULTI connector

    BSM-series bedside monitors

    OLG-2800K CO 2 monitor

    (For TG-950P*)*sales discontinued

    Intubated - Airway adapter

    R802 (YG-201T) > 10 kg R803 (YG-202T) 3 to10 kg