1 A moment arm is best defined as:
A A tendency to produce rotation about a pivot when a
force is applied
B The distance a force has to act from a pivot to create
equilibrium
C The generator of torque in a rotating device
D The net effect of all forces acting on a pivot multiplied by
the distance of each from the pivot
E The perpendicular distance of the line of action of a force
from a pivot
2 If an individual’s body weight creates a downward forceof 500N, approximately what joint reaction force isgenerated in the hip on the ipsilateral side if theindividual stands on one leg?
A 50N
B 250N
C 500N
D 1500N
E 5000N
3 Which of the following is the least likely consequence ofincreasing the offset in a new design of femoralcomponent for total hip replacement?
A A tendency to microseparation and edge loading
B Decreased bending moment on the implant
C Decreased generation of force by the abductor muscles
D Increased joint reaction force
E Increased wear rate
4 What is the linear wear threshold for UHMWPE, belowwhich it is thought that osteolysis will not occur as aconsequence of wear debris generation?
A 10 mm
B 50 mm
C 100 mm
D 0.5 mm
E 1 mm
5 Which of the following pairs of steps can be employedto create cross-linked polyethylenes for use in jointreplacement?
A Ethylene oxide treatment and heating
B Heating and Vitamin E incorporation
C High dose electron beam radiation and heating
D Sterilization dose gamma irradiation and heating
E Vitamin E incorporation and high dose gamma radiation
THOPAEDICS AND TRAUMA 27:2 126
6 Approximately what volume of wear occurs in thesecond year after implantation of a second generationmetal-on-metal hip prosthesis?
A 1 mm3
B 5 mm3
C 25 mm3
D 50 mm3
E 150 mm3
7 Which of the following statements concerningceramic femoral head in total hip replacement is nottrue?
A Niobium and Zirconium alloy can be coated onto
ceramic heads to improve the mechanical properties
B Phase change on heating Alumina Oxide heads does not
occur, unlike Zirconia Oxide heads
C Reducing the grain size in Alumina Oxide heads re-
duces the risk of implant fracture by an order of
magnitude
D Wear stripes are thought to result from third body wear
E Zirconia oxide femoral heads should never be
autoclaved
8 Which of the following statements concerningcemented, polished taper stems in hip replacement isnot true when they are compared to roughened stems?
A Distal migration wedges the implant, offloading the
cement mantle
B Radial compressive stresses in the cement mantle are
increased
C Shear stress at the implant cement interface is reduced
D The incidence of initial migration is high
E There is a poorer initial interlock with cement
9 Which of the following is not a perceived advantage ofpartial proximal coating in uncemented femoralstems?
A Better initial fixation as it is easier to accurately ream the
relevant region
B Circumferential coatings form a seal preventing wear
debris from entering the femoral canal beyond the
coating
C Less stress shielding
D Less endosteal lysis
E Load is dissipated rather than transferred to the tip of the
prosthesis
� 2013 Published by Elsevier Ltd.
CME SECTION
10 To which of the following material characteristics of animplant does stress shielding most directly relate?
A Stiffness
B Plasticity
C Toughness
D Ultimate tensile strength
E Yield strength
11 It has been suggested that surface nanograins mayrotate during the sliding of hip prostheses to bringwhich constituent towards the surface?
A Carbon
B Chromium
C Cobalt
D Molybdenum
E Niobium
ORTHOPAEDICS AND TRAUMA 27:2 127
12 Which of the following techniques can be used tocharacterize the chemical nature of materials makingup the biofilm on a retrieved prosthetic femoralhead?
A Atom force microscope
B Small angle X-ray scattering
C Transmission electron microscopy
D Tribometry
E X-ray photoelectron spectroscopy