clinical applications of remote implantable cardioverter

12
DEVICE THERAPY REVIEW ARTICLE Clinical Applications of Remote Implantable Cardioverter-Defibrillator Monitoring: Current Status and Future Directions 1 JACOB C. JENTZER, MD and 2 JOHN H. JENTZER, MD 1 Duke University Medical Center, Durham, NC 2 Dixie Regional Medical Center, St. George, UT ABSTRACT. Implantable cardioverter-defibrillator and cardiac resynchronization devices are integral for reducing mortality in many heart failure patients. Monitoring of these devices can be facilitated using wireless remote transmission. Remote monitoring can lead to early detection of important events such as device or lead malfunction, pacing abnormalities, arrhythmias, and device-delivered antiarrhythmia therapy. Studies of remote monitoring show a reduction in the response time to clinical events. Remote monitoring may safely reduce the need for frequent office device checks and facilitate triage of device-detected events. Monitoring of several device-detected parameters can predict heart failure decompensation, and interventions based on this information may prevent heart failure hospitalization. Ongoing clinical studies will further elucidate the benefits of an expanding array of remote monitoring possibilities. KEYWORDS. ambulatory monitoring, artificial cardiac pacing, cardiac arrhythmias, implantable defibrillators, heart failure. ISSN 2156-3977 (print) ISSN 2156-3993 (online) 2011 Innovations in Cardiac Rhythm Management Introduction Implantable cardioverter-defibrillators (ICDs) reduce all- cause mortality by preventing sudden death in patients with a history of sustained ventricular arrhythmias 1 and in patients with left ventricular (LV) systolic dysfunction at high risk of ventricular arrhythmias. 2 Addition of cardiac resynchronization therapy (CRT) to ICDs (CRT-D) via biventricular pacing provides a further reduction in mortality in populations with LV systolic dysfunction, QRS prolongation and moderate-to-severe heart failure (HF) symptoms. 3,4 In patients with LV systolic dysfunc- tion, prolonged QRS and mild-to-moderate HF symp- toms, CRT-D therapy reduces the incidence of HF hospitalizations, improves indices of LV function and remodeling, and may reduce mortality beyond ICD alone. 5 In 2007, over 234,00 ICDs and over 148,000 CRT- Ds were implanted in North America. 6 Monitoring of these devices has placed an increasing burden on the electrophysiology device community due to regularly scheduled in-office ICD follow-ups which usually do not result in clinical intervention. 7–9 Expert consensus recom- mends ICD follow-up every 3–6 months, 6 but clinically important events between clinic visits may be silent until the next scheduled device interrogation. Remote monitor- ing (RM) via periodic wireless downloading of device data can provide more frequent surveillance of devices, allowing for early problem identification and clinical intervention. 10 RM may safely allow reduced frequency of in-office device checks, 9 which could translate into reductions in healthcare costs. 11–14 Remote monitoring basics RM of ICD and CRT-D devices has become an accepted 6 and evolving strategy for device surveillance since its clinical introduction by Biotronik in 2000. 15–17 Each major device manufacturer has a proprietary wireless RM The authors report no conflicts of interest for the published content. Manuscript submitted May 23, 2011, final version accepted June 6, 2011. Address correspondence to: John Jentzer, MD, Southwest Cardiology Associates, Dixie Regional Medical Center, 1380 East Medical Center Drive, St. George, UT 84790. E-mail: john.jentzer@ imail.org The Journal of Innovations in Cardiac Rhythm Management, 2 (2011), 430–441 The Journal of Innovations in Cardiac Rhythm Management, August 2011 430

Upload: others

Post on 03-Feb-2022

1 views

Category:

Documents


0 download

TRANSCRIPT

DEVICE THERAPY

REVIEW ARTICLE

Clinical Applications of Remote ImplantableCardioverter-Defibrillator Monitoring: CurrentStatus and Future Directions1JACOB C. JENTZER, MD and 2JOHN H. JENTZER, MD

1Duke University Medical Center, Durham, NC2Dixie Regional Medical Center, St. George, UT

ABSTRACT. Implantable cardioverter-defibrillator and cardiac resynchronization devices areintegral for reducing mortality in many heart failure patients. Monitoring of these devices can befacilitated using wireless remote transmission. Remote monitoring can lead to early detection ofimportant events such as device or lead malfunction, pacing abnormalities, arrhythmias, anddevice-delivered antiarrhythmia therapy. Studies of remote monitoring show a reduction in theresponse time to clinical events. Remote monitoring may safely reduce the need for frequent officedevice checks and facilitate triage of device-detected events. Monitoring of several device-detectedparameters can predict heart failure decompensation, and interventions based on this informationmay prevent heart failure hospitalization. Ongoing clinical studies will further elucidate thebenefits of an expanding array of remote monitoring possibilities.

KEYWORDS. ambulatory monitoring, artificial cardiac pacing, cardiac arrhythmias,implantable defibrillators, heart failure.

ISSN 2156-3977 (print)ISSN 2156-3993 (online)

’ 2011 Innovations in Cardiac

Rhythm Management

Introduction

Implantable cardioverter-defibrillators (ICDs) reduce all-cause mortality by preventing sudden death in patientswith a history of sustained ventricular arrhythmias1 andin patients with left ventricular (LV) systolic dysfunctionat high risk of ventricular arrhythmias.2 Addition ofcardiac resynchronization therapy (CRT) to ICDs (CRT-D)via biventricular pacing provides a further reduction inmortality in populations with LV systolic dysfunction,QRS prolongation and moderate-to-severe heart failure(HF) symptoms.3,4 In patients with LV systolic dysfunc-tion, prolonged QRS and mild-to-moderate HF symp-toms, CRT-D therapy reduces the incidence of HFhospitalizations, improves indices of LV function andremodeling, and may reduce mortality beyond ICD

alone.5 In 2007, over 234,00 ICDs and over 148,000 CRT-Ds were implanted in North America.6 Monitoring ofthese devices has placed an increasing burden on theelectrophysiology device community due to regularlyscheduled in-office ICD follow-ups which usually do notresult in clinical intervention.7–9 Expert consensus recom-mends ICD follow-up every 3–6 months,6 but clinicallyimportant events between clinic visits may be silent untilthe next scheduled device interrogation. Remote monitor-ing (RM) via periodic wireless downloading of devicedata can provide more frequent surveillance of devices,allowing for early problem identification and clinicalintervention.10 RM may safely allow reduced frequency ofin-office device checks,9 which could translate intoreductions in healthcare costs.11–14

Remote monitoring basics

RM of ICD and CRT-D devices has become an accepted6

and evolving strategy for device surveillance since itsclinical introduction by Biotronik in 2000.15–17 Each majordevice manufacturer has a proprietary wireless RM

The authors report no conflicts of interest for the published content.Manuscript submitted May 23, 2011, final version accepted June 6,2011.Address correspondence to: John Jentzer, MD, SouthwestCardiology Associates, Dixie Regional Medical Center, 1380 EastMedical Center Drive, St. George, UT 84790. E-mail: [email protected]

The Journal of Innovations in Cardiac Rhythm Management, 2 (2011), 430–441

The Journal of Innovations in Cardiac Rhythm Management, August 2011 430

system: Biotronik (Lake Oswego,OR) Home Monitoring,Boston Scientific (St. Paul, MN) Latitude, Medtronic(Mounds View, MN) CareLink, and St. Jude (St. Paul,MN)Merlin.net (previously HouseCall Plus). These sys-tems share the capability to transmit device diagnosticinformation automatically to a local receiver which sendsthis information transtelephonically to a central server,allowing access by providers who can receive clinicalalerts.15–17 Depending on the system, device data can bedownloaded automatically each day or downloads mayonly occur when triggered by the patient or a devicealert.16 Remote device reprogramming is not currentlyavailable on any of these systems. Further details regardingthe technical specifications of the different RM systems arereviewed elsewhere.15–17 Patterned after remote pacemakermonitoring of battery voltage, magnet rate, and currentheart rhythm, modern ICD and CRT-D devices monitor amuch broader array of parameters. Pacing and shock leadimpedance, device-triggered alerts, sensing function, pro-grammed device parameters, percent pacing and sensingin each chamber, and review of detected arrhythmias andtherapies with stored or real-time electrograms can bemonitored (Table 1).15 Device-based parameters and algo-rithms for HF monitoring are being developed forincorporation into clinical decision making.18 We hopethat this evolving RM armamentarium may translate intobetter patient outcomes and health care savings.

Remote monitoring of ICD and CRT-Dpacemaker function

Analogous to standard pacemaker evaluation, RM canconfirm proper pacing function of ICD and CRT-D devicesand can identify the need for in-office reprogramming

to optimize device function. Abnormal pacing function cancontribute to worsening HF in ICD and CRT-D patients,and remote monitoring will disclose the amount of rightventricular (RV) pacing or biventricular pacing. ExcessiveRV pacing (above 40%) is associated with increasedmortality and HF events in patients with ICDs19 andmay predispose to ventricular arrhythmias20 and attenu-ated ICD benefit.21 Excessive RV pacing can often beminimized by lengthening the programmed AV delay oruse of pacing-minimization device algorithms, but mayrequire device upgrade to include CRT, particularly inpatients who become pacing dependent.22

Inadequate percentage of biventricular pacing (below93%) can eliminate the clinical benefits of CRT.23,24 RMcan detect drops in percent biventricular pacing, warrant-ing a full evaluation to identify the etiology.25 Loss of LVpacing capture in CRT-D devices due to LV leadmalfunction is suggested by a dramatic change in theLV pacing resistance, which must be confirmed withformal pacing threshold testing in a device clinic. Anexcessively long programmed AV delay may allow in-trinsic conduction through the native conduction systemto pre-empt biventricular pacing, especially during exer-tion. Evaluation of electrograms with their markerchannels will disclose ventricular sensing rather thanbiventricular pacing. Atrial tachyarrhythmias (ATs),supraventricular tachyarrhythmias (SVTs) or sinus tachy-cardia with ventricular rates exceeding the upper ratelimit of the pacemaker will inhibit biventricular pacing,25

as can frequent ventricular premature beats. In CRT-Dpatients with atrial fibrillation, AV nodal ablation may berequired to yield the full benefit of biventricular pacing.26

Remote monitoring of ICD lead integrity

Oversensing of non-physiological signals is an importantnon-arrhythmic cause of inappropriate ICD shocks,generally due to lead fracture, insulation breaks, orlead–header connection problems.27 The majority of ICDsystem complications are related to lead integrity failure,which continues to be a significant clinical problem.28

Rates of modern ICD lead failure are relatively low (0.6%/year), but can be higher with certain high-risk ICD leadssuch as the Medtronic Sprint Fidelis lead (3.8%/year,hazard ratio 6.4);29,30 yearly lead failure rates appear to

accelerate over time.31 Changes in lead impedance andshort sensing intervals from lead fracture can be identifiedearlier with RM, allowing timely clinical intervention.32,33

Figure 1a illustrates an abrupt increase in ICD lead pacingimpedance due to a lead fracture. Electrograms and theirmarker channels downloaded remotely from Latitude(Figure 1b) reveal oversensing of noise and confirm thediagnosis. Changes in lead impedance alone are notadequately sensitive for lead fracture,33–35 because over-sensing is often the first manifestation of lead fracture.35–37

Device reprogramming with a downloadable lead integ-rity algorithm may improve lead fracture detection byidentifying oversensing and increases in impedance.33

This algorithm can prevent many inappropriate ICDshocks in patients with Sprint Fidelis leads,33 but remains

Table 1: Remotely monitored device parameters

System statusP and R wave sensing and sensing integrityPacing thresholdsLead impedanceBattery status (voltage and charge time)Rhythm statusPercentage paced and sensed events

Right ventricular pacing percentageBiventricular pacing percentage

Average heart rate (day and night)Intrinsic heart rate and AV interval

Tachyarrhythmia burdenMode switch episodesAtrial/ventricular high rate episodesDuration/percentage of time in ATVentricular rate during AT

Device-delivered therapyAntitachycardia pacingShocks (cardioversion, defibrillation)

Device electrogramsHeart failure status (selected devices)Heart rate variabilityPatient activity levelIntrathoracic impedance

Adapted from reference 12. AT, atrial tachyarrhythmia.

J C Jentzer and J H Jentzer

431 The Journal of Innovations in Cardiac Rhythm Management, August 2011

unable to completely prevent inappropriate ICD shocks inpatients with fractured leads.36 Further analysis of down-loaded device data can differentiate lead fracture fromconnection problems and normal lead function in patientswith high-impedance lead integrity algorithm alerts.37

Lead malfunction may also cause ventricular under-sensing, which can prevent delivery of appropriatetherapy for a ventricular arrhythmia. RM may detectmore than 90% of lead-related complications and reducesymptomatic lead failure episodes by half,32 whilereducing the time to recognition of lead complications.38

Continuous follow-up in a device clinic plus RM isrecommended for patients implanted with a high-risklead.39 Oversensing of physiologic signals, such as T-waveoversensing (Figure 2) or sensing of myopotentials40 may

be corrected with a change in device sensitivity or mayrequire lead repositioning or implantation of a new lead.

Remote monitoring of arrhythmias

RM provides an effective means to evaluate device-detected arrhythmias and device-delivered therapy suchas antitachycardia pacing (ATP), low-energy cardiover-sion, and defibrillation. RM systems allow accurateevaluation of stored electrograms taken during sympto-matic or asymptomatic episodes of arrhythmias ordevice therapy.41 Electrogram analysis helps determinethe appropriateness and effectiveness of device-deliveredtherapy and can identify device or lead malfunctions and

Right Ventricular

A

B

Pace Impedance

Shock

Shock

(AS)

VT RVP-VF

VF

VF LVS LVS LVSRVS

VTVF

V-DurV-Detect

RVS488

193

160643

645

LVS VF VT VFVF128 215

308 300

320

320

320 320 320 320

3 0 0 0 0 0

300 198293 208163

1380 155 445

PVP

PVC VT VT LVS

PVP

RVP

LVP LVP LVP LVP LVP

RVP RVP RVP RVP

PVP

605 245LVP-Tr VF

350

PVP

(AS) (AS) AS AS AS ASβ48β58β23β25β25β30β20

A

RV

1,6001,3401,080

820560300

0

0 Nov 2010

Nov 2010

Dec 2010

Dce 2010756555453525

Shock Impedance

Figure 1: (a) Ventricular pacing impedance histogram downloaded from Boston Scientific Latitude showing an abrupt rise inright ventricular (RV) pacing impedance (top) in November 2010 at the time of lead fracture. (b) Electrograms downloadedfrom Boston Scientific Latitude showing right atrium (top), right ventricle (middle) and far-field right ventricle (bottom). Noiseon the RV sensing electrogram (middle) is indicative of RV lead fracture, leading to oversensing interpreted by the device asventricular tachyarrhythmias and treated by antitachycardia pacing (V-Detect, end of strip).

Clinical Applications of Remote Implantable Cardioverter-Defibrillator Monitoring

The Journal of Innovations in Cardiac Rhythm Management, August 2011 432

adverse effects related to device therapy. If device therapywas not effective, a change in device programming,medications, or hardware configuration may be neces-sary. Changes in ICD programming or antiarrhythmicmedication may require formal in-hospital evaluation ofcardioversion and defibrillation thresholds and ATPefficacy, such as with device-based non-invasive pro-grammed stimulation.

Ventricular arrhythmias and ICD shocks

Although ICDs reduce mortality by preventing death dueto ventricular arrhythmias,42 patients receiving ICD shockhave a higher risk of both arrhythmic and pump failuredeath than patients without ICD shocks.43 Patients withmore frequent ICD shocks have a higher mortality thanpatents with fewer ICD shocks.44 The mortality risk is

greater for appropriate ICD shocks (treated ventriculararrhythmias) than for inappropriate ICD shocks,43,45

which are most often due to AT or other SVTs and lessoften due to ventricular oversensing (typically from leadfailure).35 Inappropriate ICD shocks due to lead complica-tions may not pose an excess mortality risk.7 Earlydetection is paramount because ICD shocks correlate withHF instability, often requiring adjustment of medications,clinical assessment and/or device reprogramming.46

Strategies to avoid inappropriate ICD therapy due toSVT include antiarrhythmic agents, catheter ablation, ordevice programming of atrial fibrillation and sinustachycardia discriminators.35,47 ATP can be effective inmore than 70% of ventricular tachycardia (VT) episodes,safely preventing unnecessary ICD shocks.48,49 UnlikeICD shocks, VT episodes terminated by ATP may notconfer an adverse prognosis.44 ATP can occasionally beassociated with proarrhythmia such as acceleration of VT

Figure 2: Electrograms from Biotronik Home Monitoring showing far-field ventricular (top), right atrium (middle) and rightventricle (bottom). VF markers overlie the T wave, indicating T wave oversensing. VS indicates ventricular sensing; VF indicatesventricular sensing in the ventricular fibrillation zone; Det VF indicates ventricular fibrillation detection.

J C Jentzer and J H Jentzer

433 The Journal of Innovations in Cardiac Rhythm Management, August 2011

rate,49 in which case ATP should be eliminated. Frequentepisodes of VT may require pharmacologic or ablativeintervention, particularly when device therapy is ineffec-tive.50,51 Figure 3 shows electrograms from an episode ofunsuccessful ATP therapy for sustained VT, whichprompted referral for VT ablation that successfullyeliminated most of the patient’s ventricular arrhythmiaburden.

Atrial arrhythmias

Symptomatic and asymptomatic AT are common in HFpatients, and may be identified during the first post-implant year in up to one-fourth of CRT-D recipients whoare in sinus rhythm at the time of implantation.52,53

Knowledge regarding the incidence, duration, frequency,and ventricular rate response of AT can be gleaned easilyusing RM (Figure 4).52 AT episodes are one of the mostcommon clinical alerts identified by RM,10,54 and device-detected AT episodes predict worse outcomes in ICDand CRT-D patients.55 AT can interfere with properdevice functioning by preventing optimal biventricularpacing25,26 or causing inappropriate ICD shocks.35 RMmay allow earlier intervention for new or worseningdevice-detected AT,56 but it remains unclear if earlydetection of asymptomatic AT improves clinical out-comes. Device-detected atrial high-rate episodes corre-spond to AT and predict increased stroke risk.57

Mathematical modeling suggests that early initiation ofanticoagulation based on asymptomatic device-detectedAT could prevent embolism and stroke,58 and an ongoingrandomized trial will directly test this hypothesis.59

Clinical trials of remote device monitoring

Whereas transtelephonic pacemaker monitoring appearsuseful primarily for battery status determination,60

recent clinical trials using RM of ICD and CRT-D devicesdemonstrate earlier identification of clinical events10 andsafe reduction in clinic visits.9 The CONNECT (ClinicalEvaluation of Remote Notification to Reduce Time toClinical Decision) trial randomized 1,997 Medtronic ICDand CRT-D patients to standard in-office assessmentwith or without RM, and showed decreased time toclinical decision making by 17 days (more than 75%)and reduced healthcare utilization days with RM.10

The majority of remotely detected clinical events inCONNECT were prolonged AT episodes, consistent withprior studies.54 The TRUST (Lumos-T Safely ReducesRoutine Office Device Follow-up) trial randomized 1,339Biotronik ICD patients to standard office-only follow-upevery 3 months or less frequent office visits plus RM anddemonstrated a 45% reduction of in-office visits and areduction in time to respond to clinical alerts using RM,without increasing adverse events.9 TRUST confirmedthe safety of using RM in lieu of more frequent officevisits for device checks, showing that 85% of follow-upsbetween 3 and 12 months after implantation can beaccomplished remotely and that more than 90% ofscheduled clinic device checks did not require interven-tion. These studies confirm the results of smaller studiessuggesting the ability of RM to reduce response time toclinical events54,61 and reduce in-hospital follow-upvisits.14,62 Earlier identification of clinical events by RMdid not translate into an improvement in clinicaloutcomes in either CONNECT10 or TRUST.9 Smallstudies suggest that a reduction in clinic visits by RM

Figure 3: Electrograms (right atrium, top and right ventricle, bottom) downloaded from Medtronic CareLink demonstratingventricular tachyarrhythmia that is not terminated with ATP (VT Rx 1 Burst), indicated by persistent tachycardia sensing afterantitachycardia pacing. TS indicates tachycardia sensing; TP indicates tachycardia pacing.

Clinical Applications of Remote Implantable Cardioverter-Defibrillator Monitoring

The Journal of Innovations in Cardiac Rhythm Management, August 2011 434

may reduce healthcare costs,12–14 particularly for patientswith longer driving distances,11 but we await formaleconomic analysis from larger trials for confirmation.Further cost savings are possible if early detection of clinicalevents by RM leads to interventions preventing hospitaliza-tion, as suggested in CONNECT by a reduced number ofhospital days.10 The observational ALTITUDE study com-pared 69,556 Boston Scientific ICD and CRT-D patients whounderwent RM with 116,222 similar patients who receivedin-office device follow-up only, and identified a 50%mortality reduction for those patients undergoing RM, withbetter survival at 1 and 5 years.7 The non-randomizeddesign of ALTITUDE limits conclusions about causality, butbaseline demographics did not differ between the twogroups, so a dramatic imbalance in risk factor burden wouldbe required to explain the survival difference. ALTITUDE isthe only published study showing improved clinical out-comes associated with RM, although smaller studies suggestimprovements in clinical care56,63 or patient satisfaction andquality of life64 without clear outcome benefits.

Remote monitoring of heart failure status

The majority of ICD and CRT-D devices are implantedin patients with HF, a group at high risk of hospitaliza-tion.65 Weight gain and worsening of HF symptomsmay occur late66 and have poor sensitivity for predi-cting HF hospitalization.60,67 Several device-measuredparameters (Table 2) can predict HF decompensation,allowing RM to facilitate early clinical intervention andpotentially avert hospitalization.14,15,18,68 Intrathoracic

impedance (measured between the RV lead and thedevice generator) varies inversely with lung fluidcontent, such that a decline in intrathoracic impedance

suggests worsening pulmonary congestion.69 Intra-thoracic impedance correlates inversely with pulmonary

capillary wedge pressure,70 RV pressures,71 and na-triuretic peptide levels.72 Declines in intrathoracicimpedance can also occur due to device pocket com-plications, pneumonia, or pleural effusion.69 A dec-line in intrathoracic impedance may occur nearly 2weeks before worsening symptoms and weight gainpreceding a HF exacerbation.70 Declines in intrathoracicimpedance may be associated with worsening atrial73

and ventricular arrhythmias.74 Intrathoracic impedanceand AT demonstrate a bidirectional relationship, inwhich reduced intrathoracic impedance predicts moreAT and increasing AT burden predicts a reductionin intrathoracic impedance.73 The Medtronic OptiVolfeature calculates a positive numeric OptiVol indexto facilitate recognition of a sustained reduction inintrathoracic impedance over time, measured in ohm-days (Figure 5a); a remote or audible alert can besent when a preset threshold is crossed.69 Heartrate variability and patient activity level also declineas much as 2 weeks before HF hospitalization(Figure 5b).68,75,76 The Medtronic Cardiac Compass alertfeature integrates the OptiVol index with patientactivity, nocturnal heart rate, and heart rate variabilityas well as clinical events warranting provider notifica-tion such as increasing AT, reduction in biventricular

pacing, or ICD shocks.77

Shock

A

RV

AS

VS

VP-MT VP-MT

VP-MT VP-MT

VP-MT

VP-MT

[AS] [AS] [AS] [AS] [AS] [AS] [AS]

AS428 223

213430

398

428

428

428

VP-MT428

VP-MT428

VP-MT428

438435

428 428

428415

AS AF

AFAS

VS

AS430 430420 423 433

AS AS ASAS

Figure 4: Electrograms downloaded from Boston Scientific Latitude in right atrium (top), right ventricle (middle) andventricular far-field (bottom), showing 2:1 conducted atrial flutter. AS indicates atrial sensing; AF indicates atrial sensing inatrial fibrillation zone; VP-MT indicates ventricular pacing at maximum tracking rate.

J C Jentzer and J H Jentzer

435 The Journal of Innovations in Cardiac Rhythm Management, August 2011

Clinical studies of remote heartfailure monitoring

A standard OptiVol threshold of 60 ohm-days has up to60–76% sensitivity for detecting future HF hospitaliza-tions in CRT-D patients,67,70,78 and many missed HFevents are associated with a subthreshold decline inintrathoracic impedance. Each OptiVol threshold cross-ing per year predicts a 36% increased risk of HFhospitalization.76 The FAST (Fluid Accumulation StatusTrial) study demonstrated that an OptiVol threshold of60 ohm-days had superior sensitivity (76% versus 23%)and fewer false positives than daily weight monitoringfor prediction of HF worsening in 156 CRT-D patients.67

The SENSE-HF study of 501 ICD and CRT-D patientsshowed low sensitivity (42%) and positive predictivevalue (38%) of the standard 60 ohm-day OptiVolthreshold for predicting HF hospitalizations, especially

within the first 6 months after device implantation.79

False-positive OptiVol threshold crossings occur at a rateof 0.25–1.9 events per year, including episodes in whichmedication titration avoided hospitalization and lesserdegrees of decompensation not leading to hospitaliza-tion.67,68,78 The majority of OptiVol threshold crossingsat 60 ohm-days may not be associated with a clinical HF

event.67,68,76,79 Higher OptiVol thresholds of 100–120 ohm-days have been proposed to improve specificity,77,80 andrefinements in the OptiVol algorithm may increasespecificity, reduce false-positive threshold crossing events,

and improve positive predictive value of OptiVol.81

Given these limitations in sensitivity and specificity,intrathoracic impedance and OptiVol cannot be used as asole means of patient evaluation, but may supplementdaily weight monitoring and standard clinical assessmentto facilitate telemonitoring programs.82,83 The observa-tional DECODE (Decompensation Detection)study of 699Boston Scientific CRT-D patients found that adding heartrate variability and lead (not intrathoracic) impedancemonitoring to standard telemonitoring yielded poorsensitivity (,50%) for predicting HF hospitalizations, withfrequent false-positive alerts (two per patient-year).84 Anobservational study of 532 Medtronic CRT patients showedthat intrathoracic impedance monitoring using OptiVolwas associated with a reduced risk of HF hospitalization.68

The observational PARTNERS-HF (Program to Access and

Review Trending Information and Evaluate Correlation toSymptoms in Patients With Heart Failure) study of 694Medtronic CRT-D patients showed that the remote CardiacCompass alert predicted a fivefold increased 1-month riskof HF hospitalization with improved sensitivity over theOptiVol index alone.77 Further research is needed todemonstrate that RM of intrathoracic impedance and otherparameters can prevent HF events, and ongoing rando-mized studies will help to clarify this question.85

Remote hemodynamic monitoring by devices

Dedicated implantable devices can remotely monitorintracardiac hemodynamics in HF patients. Parallel to

declines in intrathoracic impedance,70 intracardiac pres-sures rise before weight gain or symptom worseningpreceding HF decompensation,86 allowing RM of hemo-dynamic parameters to facilitate early intervention.18

Investigational wireless implantable devices can measurepulmonary artery pressure (CardioMEMS Champion)87

or left atrial pressure (St. Jude HeartPOD),86 or estimatepulmonary capillary wedge pressure from RV hemody-namics (Medtronic Chronicle).88 Physician-directed med-ication titration based on device hemodynamic readingsmay prevent HF hospitalizations.86–88 The randomizedCOMPASS-HF trial of 274 advanced HF patients foundthat Chronicle-guided patient management produced a

non-significant 21% reduction in HF events.88 HeartPOD-guided medication titration increased utilization ofevidence-based HF therapies and improved functionalclass.86 The randomized CHAMPION trial of 550 ad-vanced HF patients showed that Champion-guidedpatient management reduced total HF hospitalizationsby 36%.87 Although not available on current ICD andCRT-D devices, attempts are being made to integrateintracardiac pressure monitoring into the next generationof devices.89

Integrating remote monitoring intoclinical practice

A number of potential logistical and technical pitfalls limitthe clinical application of RM.16 Currently available RMsystems generally transmit via telephone landline only, butmobile phone compatibility is available for some models.Transmission failure can occur, but technical success ratesgenerally exceed 90%.9 Patients must be actively involvedin RM for best results, and device manufacturer represen-tatives can often facilitate patient training and enrollment.Remote device alerts may occur in the majority of patientswithin 18 months after implantation,90 and frequent alertsfor non-critical events may add to physician workload.Remote alert management may require individualizedpatient alert settings and/or a nurse-run triage system, anda specific plan for device alert management is recom-

mended,16,62 This approach may reduce the number ofalerts reaching the responsible physician by 90%, leading toa minimal increase in physician workload estimated atapproximately 15 min per week per 100 patients.62,90,91

Table 2: Device parameters predicting heart failuredecompensation

Decreased intrathoracic impedanceElevated OptiVol index

Increased nocturnal heart rateDecreased heart rate variabilityDecreased patient activity levelIncreased atrial tachyarrhythmia burdenDecreased biventricular pacing percentageIncreased right ventricular pacing percentageIncreased intracardiac pressures

Right ventricular pressuresPulmonary artery pressureLeft atrial pressure

Adapted from reference 12.

Clinical Applications of Remote Implantable Cardioverter-Defibrillator Monitoring

The Journal of Innovations in Cardiac Rhythm Management, August 2011 436

Physicians could potentially be liable if critical RM data arenot acted on in a timely manner, so direct-to-physicianalerts for critical events may be preferred. False-positive

OptiVol alerts can occur at a rate approaching two perpatient per year,67 but OptiVol monitoring can be success-fully integrated into a multidisciplinary heart failure

Figure 5: (a) Remote monitoring of OptiVol using Medtronic CareLink. Intrathoracic impedance (bottom) declines in mid-June,leading to a OptiVol (top) threshold crossing event in late June, successfully treated by uptitration of diuretics. A similarepisode occurs at the beginning of August, leading to hospitalization. (b) Remote monitoring of patient activity level (top) andheart rate variability (third from top) downloaded from Medtronic CareLink. Patient activity and heart rate variability declinecoincident with decline in intrathoracic impedance and OptiVol threshold crossing.

J C Jentzer and J H Jentzer

437 The Journal of Innovations in Cardiac Rhythm Management, August 2011

disease management program without excessive providerburden.83 Integrating RM data into electronic medicalrecords facilitates documentation and coordination ofclinician responses to remote alerts, and this capability isavailable on some systems.12,13,16 Reimbursement can beavailable for RM encounters as often as every 3 months inplace of reimbursement for an office visit; more frequentRM encounters are generally not reimbursed unless achange in therapy occurs. RM may allow a reducedfrequency of scheduled in-office device checks, but itcannot currently replace all in-office device checks orobviate the need for patient–clinician encounters. Whilethe majority of routine device checks do not yield relevantfindings requiring intervention,7,9 unscheduled visits aremuch more likely to require intervention or hospitaliza-tion.8,62 RM provides an opportunity for early triage ofclinically relevant events, allowing prompt in-officefollow-up for significant events and allowing fewercritical alerts to be managed without face-to-face contactunless necessary.

Conclusion

With the proliferation of ICD and CRT-D devices and theexpanding HF population, RM of these devices maybecome increasingly important for patient care. RM canallow early warning of lead malfunction, pacing abnorm-alities, arrhythmias, and HF deterioration to allow timelyintervention. RM may safely allow less frequent in-officedevice checks, facilitating care of patients living far awayfrom their providers and potentially reducing healthcarecosts. The evolving HF monitoring features of ICD andCRT-D devices are an exciting advance that may simplifyclinical management of a complicated and challengingpatient group. We hope that ongoing clinical studies ofRM will clearly document clinical benefits and costsavings to justify the widespread proliferation of thistechnology.

References

1. Connolly SJ, Hallstrom AP, Cappato R, et al. Meta-analysis ofthe implantable cardioverter defibrillator secondary preven-tion trials. AVID, CASH and CIDS studies. Antiarrhythmicsvs Implantable Defibrillator study. Cardiac Arrest StudyHamburg. Canadian Implantable Defibrillator Study. EurHeart J 2000; 21:2071–2078.

2. Theuns DA, Smith T, Hunink MG, Bardy GH, Jordaens L.Effectiveness of prophylactic implantation of cardioverter-defibrillators without cardiac resynchronization therapy inpatients with ischaemic or non-ischaemic heart disease: asystematic review and meta-analysis. Europace 2010; 12:1564–1570.

3. Bristow MR, Saxon LA, Boehmer J, et al. Cardiac-resyn-chronization therapy with or without an implantabledefibrillator in advanced chronic heart failure. N Engl J Med2004; 350:2140–2150.

4. McAlister FA, Ezekowitz J, Hooton N, et al. Cardiac resynchro-nization therapy for patients with left ventricular systolicdysfunction: a systematic review. JAMA. 2007; 297:2502–2514.

5. Al-Majed NS, McAlister FA, Bakal JA, Ezekowitz JA. Meta-analysis: cardiac resynchronization therapy for patients

with less symptomatic heart failure. Ann Intern Med 2011;154:401–412.

6. Wilkoff BL, Auricchio A, Brugada J, et al. HRS/EHRAexpert consensus on the monitoring of cardiovascularimplantable electronic devices (CIEDs): description oftechniques, indications, personnel, frequency and ethicalconsiderations. Heart Rhythm 2008; 5:907–925.

7. Saxon LA, Hayes DL, Gilliam FR, et al. Long-term outcomeafter ICD and CRT implantation and influence of remotedevice follow-up: the ALTITUDE survival study. Circulation2010; 122:2359–2367.

8. Heidbuchel H, Lioen P, Foulon S, et al. Potential role ofremote monitoring for scheduled and unscheduled evalua-tions of patients with an implantable defibrillator. Europace2008; 10:351–357.

9. Varma N, Epstein AE, Irimpen A, Schweikert R, Love C.Efficacy and safety of automatic remote monitoring forimplantable cardioverter-defibrillator follow-up: the Lumos-T Safely Reduces Routine Office Device Follow-up (TRUST)trial. Circulation 2010; 122:325–332.

10. Crossley GH, Boyle A, Vitense H, Chang Y, Mead RH. TheCONNECT (Clinical Evaluation of Remote Notification toReduce Time to Clinical Decision) trial: the value of wirelessremote monitoring with automatic clinician alerts. J Am CollCardiol 2011; 57:1181–1189.

11. Fauchier L, Sadoul N, Kouakam C, et al. Potential costsavings by telemedicine-assisted long-term care of implan-table cardioverter defibrillator recipients. Pacing ClinElectrophysiol 2005; 28(Suppl 1):S255–259.

12. Raatikainen MJ, Uusimaa P, van Ginneken MM, Janssen JP,Linnaluoto M. Remote monitoring of implantable cardio-verter defibrillator patients: a safe, time-saving, and cost-effective means for follow-up. Europace 2008; 10:1145–1151.

13. Bikou O, Licka M, Kathoefer S, Katus HA, Bauer A. Costsavings and safety of ICD remote control by telephone: aprospective, observational study. J Telemed Telecare 2010; 16:403–408.

14. Elsner CH, Sommer P, Piorkowski C, et al. A prospectivemulticenter comparison trial of home monitoring againstregular follow-up in MADIT II patients: additional visitsand cost impact. Computers Cardiol 2006; 33:241–244.

15. Andrikopoulos G, Tzeis S, Theodorakis G, Vardas P.Monitoring capabilities of cardiac rhythm managementdevices. Europace 2010; 12:17–23.

16. Heidbuchel H. Telemonitoring of implantable cardiacdevices: hurdles towards personalised medicine. Heart2011; 97:931–939.

17. Orlov MV, Szombathy T, Chaudhry GM, Haffajee CI.Remote surveillance of implantable cardiac devices. PacingClin Electrophysiol 2009; 32:928–939.

18. Samara MA, Wilson Tang WH. Device monitoring strate-gies in acute heart failure syndromes. Heart Fail Rev 2011;(in press).

19. Sharma AD, Rizo-Patron C, Hallstrom AP, et al. Percentright ventricular pacing predicts outcomes in the DAVIDtrial. Heart Rhythm 2005; 2:830–834.

20. Gardiwal A, Yu H, Oswald H, et al. Right ventricularpacing is an independent predictor for ventricular tachy-cardia/ventricular fibrillation occurrence and heart failureevents in patients with an implantable cardioverter-defibrillator. Europace 2008; 10:358–363.

21. Barsheshet A, Moss AJ, McNitt S, et al. Long-termimplications of cumulative right ventricular pacing amongpatients with an implantable cardioverter-defibrillator.Heart Rhythm 2011; 8:212–218.

22. Martinelli Filho M, de Siqueira SF, Costa R, et al.Conventional versus biventricular pacing in heart failure

Clinical Applications of Remote Implantable Cardioverter-Defibrillator Monitoring

The Journal of Innovations in Cardiac Rhythm Management, August 2011 438

and bradyarrhythmia: the COMBAT study. J Card Fail 2010;16:293–300.

23. Koplan BA, Kaplan AJ, Weiner S, Jones PW, Seth M,Christman SA. Heart failure decompensation and all-causemortality in relation to percent biventricular pacing inpatients with heart failure: is a goal of 100% biventricularpacing necessary? J Am Coll Cardiol 2009; 53:355–360.

24. Hayes DL, Boehmer JP, Day J, et al. Cardiac resynchroniza-tion therapy and the relationship of percent biventricularpacing to symptoms and survival. Heart Rhythm 2011; (inpress).

25. Leclercq C. Problems and troubleshooting in regular follow-up of patients with cardiac resynchronization therapy.Europace 2009; 11(Suppl 5):v66–71.

26. Wilton SB, Leung AA, Ghali WA, Faris P, Exner DV.Outcomes of cardiac resynchronization therapy in patientswith versus those without atrial fibrillation: A systematicreview and meta-analysis. Heart Rhythm 19 2011; (in press).

27. Swerdlow CD, Friedman PA. Advanced ICD troubleshoot-ing: Part I. Pacing Clin Electrophysiol 2005; 28:1322–1346.

28. Maisel WH, Hauser RG. Proceedings of the ICD LeadPerformance Conference. Heart Rhythm 2008; 5:1331–1338.

29. Hauser RG, Hayes DL. Increasing hazard of Sprint Fidelisimplantable cardioverter-defibrillator lead failure. HeartRhythm 2009; 6:605–610.

30. Borleffs CJ, van Erven L, van Bommel RJ, et al. Risk offailure of transvenous implantable cardioverter-defibrilla-tor leads. Circ Arrhythm Electrophysiol 2009; 2:411–416.

31. Kleemann T, Becker T, Doenges K, et al. Annual rate oftransvenous defibrillation lead defects in implantablecardioverter-defibrillators over a period of .10 years.Circulation. 2007; 115:2474–2480.

32. Spencker S, Coban N, Koch L, Schirdewan A, Muller D.Potential role of home monitoring to reduce inappropriateshocks in implantable cardioverter-defibrillator patientsdue to lead failure. Europace 2009; 11:483–488.

33. Swerdlow CD, Gunderson BD, Ousdigian KT, et al.Downloadable algorithm to reduce inappropriate shockscaused by fractures of implantable cardioverter-defibrillatorleads. Circulation 2008; 118:2122–2129.

34. Kallinen LM, Hauser RG, Lee KW, et al. Failure ofimpedance monitoring to prevent adverse clinical eventscaused by fracture of a recalled high-voltage implantablecardioverter-defibrillator lead. Heart Rhythm 2008; 5:775–779.

35. Tzeis S, Andrikopoulos G, Kolb C, Vardas PE. Tools andstrategies for the reduction of inappropriate implantablecardioverter defibrillator shocks. Europace 2008; 10:1256–1265.

36. Kallinen LM, Hauser RG, Tang C, et al. Lead integrity alertalgorithm decreases inappropriate shocks in patients whohave Sprint Fidelis pace-sense conductor fractures. HeartRhythm 2010; 7:1048–1055.

37. Swerdlow CD, Sachanandani H, Gunderson BD, OusdigianKT, Hjelle M, Ellenbogen KA. Preventing overdiagnosis ofimplantable cardioverter-defibrillator lead fractures usingdevice diagnostics. J Am Coll Cardiol 2011; 57:2330–2339.

38. Varma N, Michalski J, Epstein AE, Schweikert R. Automaticremote monitoring of implantable cardioverter-defibrillatorlead and generator performance: the Lumos-T SafelyRedUceS RouTine Office Device Follow-Up (TRUST) trial.Circ Arrhythm Electrophysiol 2010; 3:428–436.

39. Maisel WH, Hauser RG, Hammill SC, et al. Recommendationsfrom the Heart Rhythm Society Task Force on LeadPerformance Policies and Guidelines: developed in collabora-tion with the American College of Cardiology (ACC) and the

American Heart Association (AHA). Heart Rhythm 2009; 6:869–885.

40. Kowalski M, Ellenbogen KA, Wood MA, Friedman PL.Implantable cardiac defibrillator lead failure or myopoten-tial oversensing? An approach to the diagnosis of noise onlead electrograms. Europace 2008; 10:914–917.

41. Perings C, Bauer WR, Bondke HJ, et al. Remote monitoringof implantable-cardioverter defibrillators: results from theReliability of IEGM Online Interpretation (RIONI) study.Europace 2011; 13:221–229.

42. Packer DL, Prutkin JM, Hellkamp AS, et al. Impact ofimplantable cardioverter-defibrillator, amiodarone, andplacebo on the mode of death in stable patients with heartfailure: analysis from the sudden cardiac death in heartfailure trial. Circulation 2009; 120:2170–2176.

43. Poole JE, Johnson GW, Hellkamp AS, et al. Prognosticimportance of defibrillator shocks in patients with heartfailure. N Engl J Med 2008; 359:1009–1017.

44. Sweeney MO, Sherfesee L, DeGroot PJ, Wathen MS, WilkoffBL. Differences in effects of electrical therapy type forventricular arrhythmias on mortality in implantable cardi-overter-defibrillator patients. Heart Rhythm 2010; 7:353–360.

45. Daubert JP, Zareba W, Cannom DS, et al. Inappropriateimplantable cardioverter-defibrillator shocks in MADIT II:frequency, mechanisms, predictors, and survival impact.J Am Coll Cardiol 2008; 51:1357–1365.

46. Braunschweig F, Boriani G, Bauer A, et al. Management ofpatients receiving implantable cardiac defibrillator shocks:recommendations for acute and long-term patient manage-ment. Europace 2010; 12:1673–1690.

47. Theuns DA, Rivero-Ayerza M, Boersma E, Jordaens L.Prevention of inappropriate therapy in implantable defi-brillators: a meta-analysis of clinical trials comparingsingle-chamber and dual-chamber arrhythmia discrimina-tion algorithms. Int J Cardiol 2008; 125:352–357.

48. Wathen MS, DeGroot PJ, Sweeney MO, et al. Prospectiverandomized multicenter trial of empirical antitachycardiapacing versus shocks for spontaneous rapid ventriculartachycardia in patients with implantable cardioverter-defibrillators: Pacing Fast Ventricular Tachycardia Re-duces Shock Therapies (PainFREE Rx II) trial results.Circulation 2004; 110:2591–2596.

49. Wathen MS, Sweeney MO, DeGroot PJ, et al. Shockreduction using antitachycardia pacing for spontaneousrapid ventricular tachycardia in patients with coronaryartery disease. Circulation 2001; 104:796–801.

50. Ferreira-Gonzalez I, Dos-Subira L, Guyatt GH. Adjunctiveantiarrhythmic drug therapy in patients with implantablecardioverter defibrillators: a systematic review. Eur Heart J2007; 28:469–477.

51. Aliot EM, Stevenson WG, Almendral-Garrote JM, et al.EHRA/HRS Expert Consensus on Catheter Ablation ofVentricular Arrhythmias: developed in a partnership withthe European Heart Rhythm Association (EHRA), a Re-gistered Branch of the European Society of Cardiology(ESC), and the Heart Rhythm Society (HRS); in collabora-tion with the American College of Cardiology (ACC) andthe American Heart Association (AHA). Heart Rhythm 2009;6:886–933.

52. Leclercq C, Padeletti L, Cihak R, et al. Incidence ofparoxysmal atrial tachycardias in patients treated withcardiac resynchronization therapy and continuously mon-itored by device diagnostics. Europace 2010; 12:71–77.

53. Marijon E, Jacob S, Mouton E, et al. Frequency of atrialtachyarrhythmias in patients treated by cardiac resynchro-nization (from the Prospective, Multicenter Mona LisaStudy). Am J Cardiol 2010; 106:688–693.

J C Jentzer and J H Jentzer

439 The Journal of Innovations in Cardiac Rhythm Management, August 2011

54. Lazarus A. Remote, wireless, ambulatory monitoring ofimplantable pacemakers, cardioverter defibrillators, andcardiac resynchronization therapy systems: analysis of aworldwide database. Pacing Clin Electrophysiol 2007;30(Suppl 1):S2–S12.

55. Santini M, Gasparini M, Landolina M, et al. Device-detectedatrial tachyarrhythmias predict adverse outcome in real-world patients with implantable biventricular defibrillators.J Am Coll Cardiol 2011; 57:167–172.

56. Ricci RP, Morichelli L, Santini M. Remote control ofimplanted devices through Home Monitoring technologyimproves detection and clinical management of atrialfibrillation. Europace 2009; 11:54–61.

57. Glotzer TV, Daoud EG, Wyse DG, et al. The relationshipbetween daily atrial tachyarrhythmia burden from implan-table device diagnostics and stroke risk: the TRENDS study.Circ Arrhythm Electrophysiol 2009; 2:474–480.

58. Ricci RP, Morichelli L, Gargaro A, Laudadio MT, Santini M.Home monitoring in patients with implantable cardiacdevices: is there a potential reduction of stroke risk? Resultsfrom a computer model tested through monte carlosimulations. J Cardiovasc Electrophysiol 2009; 20:1244–1251.

59. Ip J, Waldo AL, Lip GY, et al. Multicenter randomizedstudy of anticoagulation guided by remote rhythm mon-itoring in patients with implantable cardioverter-defibrilla-tor and CRT-D devices: Rationale, design, and clinicalcharacteristics of the initially enrolled cohort The IMPACTstudy. Am Heart J 2009; 158:364–370 e361.

60. Crossley GH, Chen J, Choucair W, et al. Clinical benefits ofremote versus transtelephonic monitoring of implantedpacemakers. J Am Coll Cardiol 2009; 54:2012–2019.

61. De Ruvo E, Gargaro A, Sciarra L, et al. Early detection ofadverse events with daily remote monitoring versusquarterly standard follow-up program in patients withCRT-D. Pacing Clin Electrophysiol 2011; 34:208–216.

62. Ricci RP, Morichelli L, Santini M. Home monitoring remotecontrol of pacemaker and implantable cardioverter defi-brillator patients in clinical practice: impact on medicalmanagement and health-care resource utilization. Europace2008; 10:164–170.

63. Santini M, Ricci RP, Lunati M, et al. Remote monitoring ofpatients with biventricular defibrillators through theCareLink system improves clinical management of arrhyth-mias and heart failure episodes. J Interv Card Electrophysiol2009; 24:53–61.

64. Al-Khatib SM, Piccini JP, Knight D, Stewart M, Clapp-Channing N, Sanders GD. Remote monitoring of implan-table cardioverter defibrillators versus quarterly deviceinterrogations in clinic: results from a randomized pilotclinical trial. J Cardiovasc Electrophysiol 2010; 21:545–550.

65. Hunt SA, Abraham WT, Chin MH, et al. 2009 focusedupdate incorporated into the ACC/AHA 2005 Guidelinesfor the Diagnosis and Management of Heart Failure inAdults: a report of the American College of CardiologyFoundation/American Heart Association Task Force onPractice Guidelines: developed in collaboration with theInternational Society for Heart and Lung Transplantation.Circulation 2009; 119:e391–479.

66. Wolfel EE. Can we predict and prevent the onset of acutedecompensated heart failure? Circulation 2007; 116:1526–1529.

67. Abraham WT, Compton S, Haas G, et al. IntrathoracicImpedance vs Daily Weight Monitoring for PredictingWorsening Heart Failure Events: Results of the FluidAccumulation Status Trial (FAST). Congest Heart Fail 2011;17:51–55.

68. Catanzariti D, Lunati M, Landolina M, et al. Monitoringintrathoracic impedance with an implantable defibrillatorreduces hospitalizations in patients with heart failure.Pacing Clin Electrophysiol 2009; 32:363–370.

69. Wang L. Fundamentals of intrathoracic impedance mon-itoring in heart failure. Am J Cardiol 2007; 99(10A):3G–10G.

70. Yu CM, Wang L, Chau E, et al. Intrathoracic impedancemonitoring in patients with heart failure: correlation withfluid status and feasibility of early warning precedinghospitalization. Circulation 2005; 112:841–848.

71. Vanderheyden M, Houben R, Verstreken S, et al.Continuous monitoring of intrathoracic impedance andright ventricular pressures in patients with heart failure.Circ Heart Fail 2010; 3:370–377.

72. Tomasi L, Zanotto G, Zanolla L, et al. Physiopathologiccorrelates of intrathoracic impedance in chronic heartfailure patients. Pacing Clin Electrophysiol 2011; 34:407–413.

73. Jhanjee R, Templeton GA, Sattiraju S, et al. Relationship ofparoxysmal atrial tachyarrhythmias to volume overload:assessment by implanted transpulmonary impedance mon-itoring. Circ Arrhythm Electrophysiol 2009; 2:488–494.

74. Ip JE, Cheung JW, Park D, et al. Temporal associationsbetween thoracic volume overload and malignant ventri-cular arrhythmias: a study of intrathoracic impedance.J Cardiovasc Electrophysiol 2011; 22:293–299.

75. Adamson PB, Smith AL, Abraham WT, et al. Continuousautonomic assessment in patients with symptomatic heartfailure: prognostic value of heart rate variability measuredby an implanted cardiac resynchronization device.Circulation 2004; 110:2389–2394.

76. Perego GB, Landolina M, Vergara G, et al. Implantable CRTdevice diagnostics identify patients with increased risk forheart failure hospitalization. J Interv Card Electrophysiol 2008;23:235–242.

77. Whellan DJ, Ousdigian KT, Al-Khatib SM, et al. Combinedheart failure device diagnostics identify patients at higherrisk of subsequent heart failure hospitalizations: resultsfrom PARTNERS HF (Program to Access and ReviewTrending Information and Evaluate Correlation toSymptoms in Patients With Heart Failure) study. J AmColl Cardiol 2010; 55:1803–1810.

78. Vollmann D, Nagele H, Schauerte P, et al. Clinical utility ofintrathoracic impedance monitoring to alert patients withan implanted device of deteriorating chronic heart failure.Eur Heart J 2007; 28:1835–1840.

79. Conraads VM, Tavazzi L, Santini M, et al. Sensitivity andpositive predictive value of implantable intrathoracicimpedance monitoring as a predictor of heart failurehospitalizations: the SENSE-HF trial. Eur Heart J 2011; (inpress).

80. Ypenburg C, Bax JJ, van der Wall EE, Schalij MJ, van ErvenL. Intrathoracic impedance monitoring to predict decom-pensated heart failure. Am J Cardiol 2007; 99:554–557.

81. Sarkar S, Hettrick DA, Koehler J, et al. Improved Algorithmto Detect Fluid Accumulation via Intrathoracic ImpedanceMonitoring in Heart Failure Patients with ImplanableDevices. Journal of Cardiac Failure 2011; (in press).

82. Hasan A, Paul V. Telemonitoring in chronic heart failure.Eur Heart J 2011; article in press.

83. Mullens W, Oliveira LP, Verga T, Wilkoff BL, Wilson TangWH. Insights from internet-based remote intrathoracicimpedance monitoring as part of a heart failure diseasemanagement program. Congest Heart Fail 2010; 16:159–163.

84. Ewald GA, Gilliam FR, Sweeney RJ. Automated HF De-compensation Detection: Results from the DecompensationDetection Study (DECODE) [abstract]. Journal of Cardiac Failure2009; 15(Suppl):S122.

Clinical Applications of Remote Implantable Cardioverter-Defibrillator Monitoring

The Journal of Innovations in Cardiac Rhythm Management, August 2011 440

85. Braunschweig F, Ford I, Conraads V, et al. Can monitoringof intrathoracic impedance reduce morbidity and mortalityin patients with chronic heart failure? Rationale and designof the Diagnostic Outcome Trial in Heart Failure (DOT-HF).Eur J Heart Fail 2008; 10:907–916.

86. Ritzema J, Troughton R, Melton I, et al. Physician-directedpatient self-management of left atrial pressure in advancedchronic heart failure. Circulation 2010; 121:1086–1095.

87. Abraham WT, Adamson PB, Bourge RC, et al. Wirelesspulmonary artery haemodynamic monitoring in chronicheart failure: a randomised controlled trial. Lancet 2011;377(9766):658–666.

88. Bourge RC, Abraham WT, Adamson PB, et al. Randomizedcontrolled trial of an implantable continuous hemodynamic

monitor in patients with advanced heart failure: theCOMPASS-HF study. J Am Coll Cardiol 2008; 51:1073–1079.

89. Adamson PB, Conti JB, Smith AL, et al. Reducing eventsin patients with chronic heart failure (REDUCEhf) studydesign: continuous hemodynamic monitoring with animplantable defibrillator. Clin Cardiol 2007; 30:567–575.

90. Nielsen JC, Kottkamp H, Zabel M, et al. Automatic homemonitoring of implantable cardioverter defibrillators.Europace 2008; 10:729–735.

91. Theuns DA, Rivero-Ayerza M, Knops P, Res JC, Jordaens L.Analysis of 57,148 transmissions by remote monitoring ofimplantable cardioverter defibrillators. Pacing Clin Electrophysiol2009; 32(Suppl 1):S63–65.

J C Jentzer and J H Jentzer

441 The Journal of Innovations in Cardiac Rhythm Management, August 2011