Clearance measurements as a tool for waste minimization during decommissioning Patrik Konnéus, Veronica Andersson and Maria Lindberg Studsvik Nuclear AB, SE-611 82 Nyköping, Sweden Detector equipment Counts-log scale Energy(keV) 250 500 750 1000 1250 1500 1750 0 10 1 10 2 10 3 10 10 Counts log scale Energy(keV) 250 500 750 1000 1250 1500 1750 2000 2250 2500 0 10 1 10 2 10 3 10 4 Counts log scale 2 10 3 10 4 4 Detector Name: B00009 Sample Title: QA Bakground Count Live time: 60000 Peak Energy FWHM Net Peak Net Area Continuum Net Peak No (keV) (keV) Area Uncert Counts cps 3 72,8 0,78 5,27E+02 28,87 4,50E+02 0,008783 4 74,96 0,78 9,15E+02 36 4,97E+02 0,015250 5 84,76 0,91 2,91E+02 37,6 5,18E+02 0,004850 6 238,54 0,93 9,34E+01 31,18 3,88E+02 0,001557 7 351,97 0,83 9,14E+01 27,4 2,78E+02 0,001523 8 510,78 2,71 1,21E+03 47,25 3,40E+02 0,020167 9 609,16 1,29 6,58E+01 22,68 1,69E+02 0,001097 12 1172,97 0,99 9,93E+01 17,64 6,67E+01 0,001655 13 1332,45 1,21 9,38E+01 18,47 6,83E+01 0,001563 14 1460,69 0,79 5,80E+01 12,48 3,90E+01 0,000967 Cell entrance showing the rotating table To the left: scale measuring the object’s weight. The detector distance is 150 cm from table centre, placed for minimizing background. During measurement the table is able to rotate. Background shielding of 160 mm steel and 3 mm lead lining. Standard waste containers used for routine counting • 1200 or 600 L box (Berglöfs box) 200 L or 100 L steel drum. • Maximum dimensions: 1140 x 1450 x 1140 mm. • Maximum weigth: 3500 kg. Complete background spectrum for 60 000 s Spectrum from daily QA background Peak analysis report from background spectrum • HPGe broad energy 50 % relative efficiency. • ISOCS is used for efficiency calculations and APEX for detector operating. • Operating – In order to have full control of the equip- ment, daily qualilty assurance checks are carried out for control of detector stability and background influences. Acquisition time normally 1200 - 3600 seconds to reach acceptable detection limits of main gamma emitting radionuclides. Clearance, exemption or free release of material from regulatory control are three ways of expres- sing the same thing, minimizing the amount of waste that needs disposal as radioactive waste and therefore preserving a repository as a national asset. Studsvik Nuclear has utilized the former whole body measurement cell as a shielded compartment used for gamma spectroscopy measurements of materials that may be subject to clearance. The cell was moved from its original position by heavy lifting and transport, as it was transported in one piece weighing 53 tonnes.

Upload: others

Post on 09-Jul-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Clearance measurements as a tool for waste minimization during … · 2012-05-11 · Clearance measurements as a tool for waste minimization during decommissioning Patrik Konnéus,

Clearance measurements as a tool for waste minimization during decommissioning

Patrik Konnéus, Veronica Andersson and Maria LindbergStudsvik Nuclear AB, SE-611 82 Nyköping, Sweden

Detector equipment

Counts-log

scale

Energy(keV)250 500 750 1000 1250 1500 1750

010

110

210

310

10

Counts

log scale

Energy(keV)250 500 750 1000 1250 1500 1750 2000 2250 2500

010

110

210

310

410

Counts

log scale

210

310

4410

Detector Name: B00009

Sample Title: QA Bakground Count

Live time: 60000

Peak Energy FWHM Net Peak Net Area Continuum Net PeakNo (keV) (keV) Area Uncert Counts cps3 72,8 0,78 5,27E+02 28,87 4,50E+02 0,008783

4 74,96 0,78 9,15E+02 36 4,97E+02 0,015250

5 84,76 0,91 2,91E+02 37,6 5,18E+02 0,004850

6 238,54 0,93 9,34E+01 31,18 3,88E+02 0,001557

7 351,97 0,83 9,14E+01 27,4 2,78E+02 0,001523

8 510,78 2,71 1,21E+03 47,25 3,40E+02 0,020167

9 609,16 1,29 6,58E+01 22,68 1,69E+02 0,001097

12 1172,97 0,99 9,93E+01 17,64 6,67E+01 0,001655

13 1332,45 1,21 9,38E+01 18,47 6,83E+01 0,001563

14 1460,69 0,79 5,80E+01 12,48 3,90E+01 0,000967

Cell entrance showing the rotating table• To the left: scale measuring the object’s weight.• The detector distance is 150 cm from table centre,

placed for minimizing background.• During measurement the table is able to rotate.• Background shielding of 160 mm steel and 3 mm lead

lining.

Standard waste containers used for routine counting• 1200 or 600 L box (Berglöfs box) 200 L or 100 L steel drum.• Maximum dimensions: 1140 x 1450 x 1140 mm.• Maximum weigth: 3500 kg.

Complete background spectrum for 60 000 s

Spectrum from daily QA background

Peak analysis report from background spectrum

• HPGe broad energy 50 % relative effi ciency.• ISOCS is used for effi ciency calculations and APEX for

detector operating.• Operating – In order to have full control of the equip-

ment, daily qualilty assurance checks are carried out for control of detector stability and background infl uences.

Acquisition time normally 1200 - 3600 seconds to reach acceptable detection limits of main gamma emitting radionuclides.

Clearance, exemption or free release of material from regulatory control are three ways of expres-sing the same thing, minimizing the amount of waste that needs disposal as radioactive wasteand therefore preserving a repository as a national asset.

Studsvik Nuclear has utilized the former whole body measurement cell as a shielded compartment used for gamma spectroscopy measurements of materials that may be subject to clearance. The cell was moved from its original position by heavy lifting and transport, as it was transported in one piece weighing 53 tonnes.