cleanliness validation white paper medical device

Upload: shah241

Post on 03-Jun-2018

222 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    1/17

    v1 Dec 2013

    MEDICAL DEVICES CLEANLINESS

    VALIDATION

    Dr Chris Pickles

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    2/17

    2

    INTRODUCTION

    Residues on the surface of medical devices can

    cause implant failure and poor device

    performance. The main source of these residues

    is from materials used in the manufacture of thedevice, although contamination during the

    storage, cleaning and handling of the device is

    also known to occur. Small amounts of these

    surface residues can cause deleterious effects in

    patients, because the residues are in direct

    contact with body tissues and patients often

    have compromised immune systems (Beal,

    2009). In addition, residues may often alter the

    surface chemistry and geometry of the device, so

    even inert residues can be a problem. For

    example, small amounts of non-toxic cutting fluid

    on an implant limit the ability of surrounding

    tissues to attach to the implant (Jackson and

    Ahmed, 2007).

    In order to minimise contamination, the Federal

    Drug Administration (FDA) stipulates that

    medical device manufacturers follow specific

    cleanliness validation procedures (FDA, 2003).

    Firstly, they must identify all possible residues

    present on the device and set an acceptable

    residue limit (Luginbuehl et al, 2006). Then, they

    must use a cleaning regime that reduces residue

    levels below this limit, without leaving significant

    levels of cleaning agent behind. Finallydocumentation to verify that residue limits are

    not exceeded must be submitted to the FDA

    before the device can go on the market

    (Luginbuehl et al, 2006).

    Despite these procedures being in place, some

    medical devices are failing to meet FDA

    requirements for cleanliness verification and

    validation. Since 2001, 173 medical devices have

    been recalled, some due to contamination issues

    (Medical Device Recalls, 2009). In just one year of

    sterility inspections, more than 483 FDA

    observations related to validation deficiencies -more than any other deficiency (Booth, 1999).

    Part of the problem for medical device

    manufacturers and device cleaners is that there

    are no official residue limits or specified analytical

    techniques to measure residue levels. Therefore,

    the manufacturers have to base their judgments

    on existing medical devices on the market, which

    may not be possible if the device has a novel use

    or uses different materials or manufacturing

    processes to devices currently on the market.

    Another issue is that the surfaces of medical

    devices are becoming more complex in terms of

    their geometry and chemistry, making cleaning

    more difficult. Recently there has been a rise in

    the use of combination devices (devices with

    both a drug or biologic component and a device

    component) which creates even more challenges

    for effective device cleaning. These combo

    devices need to simultaneously meet quality

    regulations for both the device and drugscomponent, but the FDA has yet to issue

    guidance for cleanliness validation for these

    devices (Kanegsberg et al, 2008; Staff Report,

    2007).

    Furthermore, highly aggressive cleaning of the

    device may produce undesirable surface

    modifications and inadequate cleaning may leave

    residues that interact with therapeutic chemicals

    (Kanegsberg et al, 2008).

    In light of the challenges facing the medicaldevice industry, this paper will cover the

    following:

    1. Guidance from regulatory bodies on

    cleanliness validation

    2. Residues and sources of contamination

    3. Residue limits

    - Risk categories for medical devices

    - Combination devices

    4. Analytical methods to validate cleanliness:

    description, and pros and cons of each

    method

    - Direct surface analysis

    - Residue analysis

    - Gravimetric

    5.

    Benefits of cleanliness validation

    The advice in this document is intended for

    manufacturers of implantables, combo products

    and of any device that comes into contact with

    the inside of the body, e.g., theatre instruments,

    and users of reusable instruments or companies

    that clean them. Biological contaminants and

    bioburden analysis are not covered in this white

    paper.

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    3/17

    3

    GUIDANCE FROM REGULATORY

    BODIES ON CLEANLINESS

    VALIDATION

    None of the regulatory bodies for medical

    devices stipulate residue limits for medicaldevices, as absolute cleanliness is unobtainable,

    nor do they specify the cleanliness validation

    methods to be used. All of the regulatory bodies

    stipulate a requirement to identify possible

    residues, establish a residue limit and to not

    exceed this limit, and to document and validate

    cleanliness as part of an ongoing process (Daniel

    et al, 2008). For example, ISO 14969 requires

    documented and validated cleaning methods for

    production facilities, manufacturing equipment,

    and for the medical devices themselves (Daniel et

    al, 2008). According to EN ISO 17664, medicaldevice manufacturers are obliged to provide

    validated and documented methods of

    reprocessing (cleaning and disinfection) for

    reusable medical devices (Daniel et al, 2008).

    The UKs BS EN46002 standard, a specification

    for application of ENISO9002, stipulates the

    following requirements for cleanliness of medical

    devices: cleanliness of product and validated

    sterilisation process (Daniel et al, 2008).

    According to the FDA Guide to Inspections of

    Validation of Cleaning Processes and The PDA

    Technical Report No. 29, Validated cleaning

    requires a procedure whose effectiveness has

    been proven by a documented program

    providing a high degree of assurance that a

    specific cleaning procedure, formed

    appropriately, will consistently clean a particular

    piece of equipment, device, or area, to a

    predetermined level of cleanliness - a level

    objectively substantiated by specific chemical

    and microbiological tests (Brunklow et al, 1996).

    RESIDUES AND SOURCES OFCONTAMINATION

    A problem for manufacturers is that there are a

    wide range of possible contaminants/residues for

    a medical device (table A). These residues can

    react with drug components and other residues

    to create new, potentially toxic compounds, or

    can alter the surface geometry of the device by

    corrosion or by creating layers on the surface of

    the device (figure A) (Kanegsberg and

    Kanegsberg, 2006).

    Contaminants usually fall into one of three

    categories: water soluble residue, non water-

    soluble residue and non soluble debris. Water-

    soluble residues are usually ionic compounds

    such as detergents and salts. Non water-soluble

    residues, such as oils, greases, and other

    hydrocarbons, are soluble in solvents other than

    water. Non-soluble debris includes residues such

    as metals, organic and inorganic solids, andceramics.

    Table A. Types of Residues Found on the Surface

    of Medical Devices with Examples

    Residues and Source

    of Residue

    Examples

    Cleaning Agents

    Detergents, IonicCompounds, Acids,Alkalis, Solvents

    Chlorinated cleaners,ethyl and isopropylalcohol, methylchloroform (1,1,1-trichloroethane), andtricholoroethyleneAnionic and non-ionicsurfactants

    Sterilising Agents Especially EthyleneOxide (Estrin, 1990)

    Plastic medicalequipment retainformaldehyderesidues after low-temperature steamand formaldehyde

    (LTSF) sterilization(NEWS, 2005)

    Packaging

    Debris, Extractablecompounds(plasticizers)

    Handling equipment

    Gloves, Lotions,Skin/Oil

    Dithiocarbamatevulcanizationaccelerators on latexproducts (FDA,Technical Guide)

    Oils/lubricants

    hydrocarbon-based)

    Alkanes, olefins,

    additives

    Lubricants aqueous-

    based)

    Emulsifiers

    Inorganics

    Grit blast compoundsLoose metal grainsSalt/ionic speciesHeavy metals

    Silicon, Aluminium,Iron

    Processing Aids

    Polishing compounds(lipid-based)

    Dye penetrantsMoulding aids

    Mould releasingagents can often befound on plastic

    devices.Silicones,fluorocarbons

    (Luginbuehl et al, 2006; Speigelberg, 2003)

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    4/17

    4

    Figure A. Composition of an Initially Contaminated and then Cleaned Surface (Riechl, 2003)

    Residues present as irregular-shaped particles

    have few attachment points with the surface of

    the medical device and can be easily removed

    during cleaning (figure B) (Hazell et al, 2007).

    Residues that form a single layer on the surface

    of a device have many more attachment points

    with the surface of the device and are not so

    easily removed (figure B) (Hazell et al, 2007).

    However, in the saline environment of the body

    these residues can be liberated through corrosion

    reactions and become a potential problem for

    medical device manufacturers (Hazell et al,

    2007).

    Figure B. Comparison of Bonding Sites for a Particle and the Same Volume of Residue Adsorbed onthe Surface of a Metallic Medical Device

    Arrow representing bonding sites -

    not to scale and not representative of the number

    of bonding sites (Hazell et al, 2007)

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    5/17

    5

    The chemical properties of residues present on

    the surface of a device help determine the type

    of analytical method used to validate cleanliness.

    Residues are sub-divided into different categories

    (Table B):

    Table B. Chemical Classifications of Different

    Residues

    Residue

    Classification

    Examples

    Inorganic Salts

    Ceramics

    Metals

    Organic (includingpolymers)

    Polar (soluble in water)

    Apolar (insoluble inwater)

    Biological Natural macromolecules

    Bacterial or viral

    (Luginbuehl et al, 2006)

    RESIDUE LIMITS

    After identifying the potential residues present, it

    is the role of the manufacturer to set residue

    limits and validate that these limits are not

    exceeded in the manufacturing process. The roleof cleaners of reusable devices is to ensure

    residue limits are not exceeded. Although there is

    no official residue limit, there are guidelines for

    setting residue limits based on the current

    product qualities: the risk classification of the

    device*, size of the device and the possible

    residues present (LeBlanc, 2006). For cleanliness

    validation of a process, the FDA wants to see

    evidence that the residue limits are logical,

    practical, achievable and verifiable (Booth, 1999).

    If the device or similar devices are already on the

    market, it is recommended that manufacturers

    look at the devices history of acceptable

    performance then use a mean level of residue

    plus 3 standard deviations for particulates and

    other types of residue (Broad and Kanegsberg,2007).

    For new devices, a series of residue spiking

    biocompatibility studies need to be performed at

    different levels to determine the failure point of

    the device (Broad and Kanegsberg, 2007). At

    half the failure point, analysis can be performed

    to demonstrate that device performance was not

    affected by the residues present and toxicity

    levels were not exceeded (Broad and

    Kanegsberg, 2007). If the expected level of

    residue is known, the device can be spiked at ahigher residue level and then evaluated for

    biocompatibility and functionality. This higher

    residue level is the maximum allowable residue

    limit.

    It is also recommended that manufacturers

    estimate the acceptable daily intake (ADI) for a

    cleaning fluid or residue, if systemic toxicity

    based limits are not known. The ADI can be

    calculated using the LD50 (lethal dose for 50% of

    the population by compatible route of exposure

    depending on device) and a conversion factor

    (usually a value from 100 to 100,000):

    ADI = LD50 (mg/kg) x body weight (kg) /

    conversion factor (LeBlanc, 2006)

    * The Medical Devices Directive (MDD) includes a

    classification system whereby the level of

    regulatory control applied to devices is

    proportionate to the degree of risk associated

    with the device (www.MHRA.gov.uk)

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    6/17

    6

    Table C. Risk Classifications of Medical Devices Used by the Medical Device Directives (the system

    used in the EU)

    Classification Risk

    Nature of

    Contact and

    Tissue Type

    Duration of

    Contact

    Types of Device

    Conformity

    Assessment Process

    including cleanliness

    validation)

    Conducted By

    Class I Low Surface -skin,mucous,membranes

    Limited Examinationgloves, tape,blood-pressurecuff, dentaldams,endoscopes

    Manufacturer/NotifiedBody**

    Class IIa Medium Externalcommunication- Blood,indirect

    Prolonged Dialysis,cardiopulmonarybypass

    Notified Body**

    Class IIb Medium

    Class III High Implant direct,blood andtissue contact

    Permanent Shunts or grafts,orthopaedicimplants

    Notified Body**

    (Booth, 1999; Riechl, 2003; Albert, 2004, www.MHRA.gov.uk)

    ** The Notified Body is an independently verified body and is required for medical device approval in

    the EU.

    N.B. The FDA system differs from the EU system in that the devices are classified as I, II, III and IV

    devices rather than I, IIa, IIb and III.

    For combination devices, setting residue limits is challenging and a good understanding of the

    chemical properties (especially stability) of the biological component of the device is needed; see

    table D below for the issues concerning cleanliness validation for combo devices.

    Table D. Issues Concerning Cleanliness Validation of Combination Devices

    Cleaning or Assembly Issue Comments

    Areas proximal to the drug Cleaning of the device may compromise drug activity

    Consider materials compatibility

    Technical support from cleaning agent supplier is optimal

    Corrosion Avoid chlorine-containing agents with iron-containing alloys

    Thin films Films can interfere with drug release or action

    Particulates Could interfere with drug delivery

    Out gassing Porous materials can absorb and retain cleaning chemicals withunintended slow release

    Cleaning Process Design Balance cleaning action and product modification from wash rinseand dry

    Product design Design for manufacturability including cleaning

    Be aware of complexities that can trap water or other

    contaminants

    Sterilisation Heat or radiation may compromise drug portion of the device

    (Kanegsberg, et al., 2008)

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    7/17

    7

    The therapeutic surface chemicals often used to

    minimise the risk of rejection of implants, for

    example hydroxyapatite, present an opportunity

    for interactions between therapeutic surface

    chemicals and residues (Jackson and Ahmed,

    2007).

    ANALYTICAL METHODS TO

    VALIDATE CLEANLINESS:

    DESCRIPTION AND PROS AND CONS

    OF EACH METHOD

    The challenge for the medical device industry inverifying the cleanliness of their devices andvalidating the performance of their cleaningprocesses is to identify one or moremeasurement techniques that detect the

    contaminants present on a sample at the requiredsensitivity (Booth, 1999). The FDA prefers specificmethods, although considers any validationmethod acceptable if it can be justified(McLaughlin and Zisman, 2002).

    Once possible contaminants are known andresidue limits are set, analysis method (s) must bechosen that measure the analyte (s) at and belowthe residue acceptance limits. For this, the Limitof Detection (LOD) (lowest amount of acompound that can be detected) and the Limit ofQuantitation (LOQ) (lowest amount ofcompound that can be quantified) of the

    analytical tool need to be established. Theresidue acceptance limit should be well above theLOQ, so that the residue can be accuratelyquantified (Kaiser and Minowitz).

    In order to satisfy the FDAs cleanliness validationcriteria, the FDA needs to be assured that thevalidation process is repeatable, the process doesnot adversely affect the product or thepackaging, and the process meets the sterilityassurance limit (SAL) in the worst case scenario.(The SAL is the probability that an implant willremain non-sterile following sterilisation - one in amillion) (Booth, 1999; Arscott et al., 1996). TheFDA recommends that the following four criteria

    are considered before implementing an analyticalprocedure for a cleaning validation application(Booth, 1999):

    Sensitivity - the method is appropriate for theresidue limits in terms of sensitivity and LOD ofthe device (mentioned previously).

    Practicality- the method is practical and rapidand, if possible, uses established pre-existingtechniques and equipment.

    Validation Scheme- the method is readilyvalidated in accordance with regulatoryrequirements for instrumentation.

    Successful Recovery Study- the method should

    include compound recovery studies thatchallenge the sampling and testing methods

    (Booth, 1999).

    The analytical methods can be subdivided intotwo categories: direct and indirect. Directmethods detect the residue directly on thesurface of the device; indirect methods requireresidues to be extracted prior to analysis. Forindirect methods, residues are extracted bywashing the device with water, aqueous solutionor an organic solvent and then collecting therinse water, or by direct surface sampling with aswab. Exceptions to this are volatile organics or

    absorbed gases, e.g. ethylene oxide, which areextracted via thermal evaporation using aheadspace sampling technique.

    At the very minimum, the method requires twodifferent analysts, instruments, columns (ifchromatography is being used), days forexperimentation, and the use of preparedsamples and standards (Kaiser and Minowitz).

    The different cleanliness validation methods aredivided into three main categories: direct surfaceanalysis, residue analysis and gravimetric analysis.Both gravimetric analysis and residue analysis areindirect methods; of course, direct surface

    analysis is a direct method. The choice of methoddepends not only on the anticipated residuespresent and residue acceptance limits, but alsoincludes a consideration of the pros and cons ofthe methods, along with cost considerations(Table E). Usually more than one method is usedto validate the cleanliness of the device to ensurethat the entire range of possible contaminants isdetected.

    Of all of the methods, surface analysis techniques

    are the only methods that can identify the

    location of the contaminant on the device and

    detect extremely low levels of residue. So, the

    surface analysis techniques can potentially beused to identify the manufacturing stage or

    process in which contamination has occurred.

    Two surface analysis techniques that are often

    used to validate cleanliness are ToF-SIMS and

    XPS because they analyse the outermost surface

    layers and therefore directly measure residual

    contamination (Hazell et al., 2007).

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    8/17

    8

    Table E. Comparison of the Three Major Cleanliness Validation Method Categories - Including theResidues Detected, Cost Considerations and Pros and Cons of the Technique

    Analysis

    Technique

    Residues

    Detected or

    Ideal Device

    Material

    Costs

    per

    Sample

    Initial

    Investment

    Pros Cons

    DirectSurfaceAnalysis

    Metals,ceramics,

    plastics

    High High - Location ofcontaminants on thedevice is known

    - Used at each step inthe manufacture ofthe device

    - Used for insolubledried-out residues

    - Identifycontaminantspresent

    - Can detect verysmall levels ofresidue

    - Can assess sub-surface contamination

    - Lab setting is needed

    Gravimetric Non-volatileresiduesincludingnon-watersolublecontaminants

    Low Low - Sensitive

    - Simple

    - Robust

    - Little preparation isneeded

    - Broad range ofcontaminants can betested

    - Can use extractionsolvents other thanpurified water

    - Human error is likely

    - Easy to contaminatethe sample

    - Easy to obtainmisleading data dueto differences inparticle size anddistribution betweenthe residues in thesample and referenceresidues

    - Excludes residuesmore volatile than theextraction solvent

    - Does not identify theresidues present

    ResidueAnalysis(requiresextractionmedia)

    Porousmaterials,materialswith acomplexsurfacegeometry

    Low Low - Can capture residuestrapped in pores inthe device

    - Measure high levelsof contamination

    - Done in a hospitalsetting

    - Some techniques can

    identifycontaminantspresent, some cannot

    - Location ofcontaminants is notknown

    - Does not captureadsorbed contaminant

    - Some residues may beleft on the medicaldevice

    - Residues may not behomogenouslydistributed in theeluant therefore theeluant sampleanalysed may nottruly represent thelevels of contaminantpresent on the device

    - Difficult to assess sub-surface contamination

    (Beal, 2006)

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    9/17

    9

    All of the direct surface analysis techniquesexamine the top few layers of the devicessurface by irradiating the sample with x-rays orion beams and analysing the emitted electrons orsecondary ions. In contrast, the indirect residueanalysis methods use a variety of techniques to

    detect residues present on the surface of thedevice. Descriptions of how the differentmethods work are detailed in tables F and Gbelow.

    Gravimetric analysis is an indirect analysis

    technique. The device is placed in an extraction

    media and the resulting eluant is passed through

    a pre-weighted membrane filter. The filter is then

    dried and re-weighed. It is assumed that any

    difference in mass is due to particles present onthe surface of the medical device and these

    particles are of the same density and size as the

    original tests used to set acceptable mass levels

    for the residues.

    Table F1. Descriptions of the Most Commonly Used Direct Surface Analysis Methods

    MethodDescription

    Method Abbreviation

    X-Ray PhotoelectronSpectroscopy

    (Also known as ElectronSpectroscopy for ChemicalAnalysis (ESCA)

    XPS The medical device is irradiated with x-rays causingthe emission of particles called photoelectrons. Theenergy of the emitted photoelectrons is specific toelements on the surface of the medical device.

    Time-of-Flight SecondaryIon Mass Spectrometry

    ToF-SIMSA pulsed beam of primary ions is focused onto thesurface of the medical device producing particlescalled ions. These ions are analysed providinginformation about the molecules and elementspresent on the surface of the sample.

    3D Non-contact profiling 3D-NCP This technique involves irradiating the sample withtwo or more white-light lasers and analysing thepattern of interference of the light waves. A detailedmap of the outer nanometers of the medical devicessurface can be obtained showing height variation inthe sample

    Transmission electronmicroscopy

    TEM This technique involves passing a high energy (200-300 keV) electron beam through a thinned section ofthe specimen to produce very high-resolution imagesof the material.

    Scanning ElectronMicroscopy

    SEM Rasters a focused electron beam across a samplesurface, providing high-resolution and long-depth-of-field images of the sample surface.

    Laser Ionisation MassAnalysis

    LIMA A high performance reflectron ToF massspectrometer which uses an Nd:YAG laser as its

    primary ionising beam. Uses a finely focussed probe -in this case a laser - for analysis.

    Fourier Transform Infra-RedSpectroscopy

    FTIR Chemical bonds vibrate at characteristic frequencies,and when exposed to infrared radiation, they absorbthe radiation at frequencies that match their vibrationmodes. Measuring the radiation absorption as afunction of frequency produces a spectrum that canbe used to identify functional groups andcompounds.

    Raman Spectroscopy Raman Raman Spectroscopy (Raman) enables you todetermine the chemical structure of a sample andidentify the compounds present by measuringmolecular vibrations, similar to Fourier TransformInfrared Spectroscopy (FTIR). However, the method

    used with Raman yields better spatial resolution andenables the analysis of smaller samples.

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    10/17

    10

    Table F2. Descriptions of the Most Commonly Used Residue Analysis Methods

    Method Abbreviation Description

    Total Organic

    Carbon

    TOC Residuals are extracted from the devices in a known amount of

    purified water and the extract is analyzed on a TOC instrument.The TOC is determined by the oxidation of an organic compoundinto carbon dioxide. This oxidation can occur through a numberof mechanisms depending on the instrument being used. Thecarbon dioxide that is produced from these oxidations is eithermeasured using conductivity or infrared techniques (Kaiser andMinowitz).

    Capillary ZoneElectrophoresis

    CZE A high voltage source is used to apply a potential across twosolutions. One of the solutions contains the analyte and thepotential applied to the solutions causes the analyte to migratethrough the capillary, through the detector, and into the othersolution (Kaiser and Mirowitz).

    GasChromatography-MassSpectroscopy

    GC/MS Identifies and separates volatile and semi-volatile compoundsinto individual components using a temperature-controlled gaschromatograph. During the process, a sample is injected into thechromatograph (or it may come from another sampling device)and passes through the chromatography column, whichseparates mixtures into individual components as they passthrough at different rates. The result is a quantitative analysis ofthe components, along with a mass spectrum of eachcomponent.

    (Zurbruegge, 2006; Speigelberg, 2003; Kaiser and Aiche, 2005)

    Table F3. Gravimetric: Comparison of Specific, Common Cleanliness Validation Techniques - ResiduesIdentified, Pros and Cons

    Analysis

    Method

    Residues Pros Cons

    Detection

    Limits

    Other

    Information

    Gravimetric Same Non-volatileresiduesand abroadrange ofresidues

    - Sensitive

    - Simple

    - Robust

    - Littlepreparation isneeded

    - Broad rangeofcontaminantscan be tested

    - Can useextractionsolventsother thanpurifiedwater

    - Human error islikely

    - Easy tocontaminate thesample

    - Easy to obtainmisleading datadue to differencesin particle sizeand distribution

    between theresidues in thesample andreference residues

    - Excludes residuesmore volatile thanthe extractionsolvent

    - Does not identifythe residuespresent

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    11/17

    11

    Table G: Indirect Residue Analysis (Requires extraction media): Comparison of Specific, CommonCleanliness Validation Techniques - Residues Identified, Pros and Cons

    Analysis

    Method

    Residues Pros Cons

    Detection

    Limits

    Other

    Information

    IonChromatography

    Ionizableorganic acids(cleaningfluidscontainingsodium andpotassiumions)

    - Very low levels ofcleaning agent canbe detected

    - Assumes the rinse

    - Water contains nopotassium

    Specific

    CapillaryElectrophoresis

    Analyseorganic acids,inorganic andtrace drug

    residues

    - Detection limits arehigher than withHPLC

    - All commondetection can be

    used in capillaryelectrophoresisdetection.

    Specific

    GasChromatography-MassSpectroscopy

    (GC-MS)

    Solid, liquid,gas

    Organicsolvents(volatileorganiccompounds)and semi-volatileorganiccompounds(plasticizers,

    phenols, etc.)volatile andsemi-volatilecompounds

    - QuantificationIdentifies very small(trace-level)quantities of aresidue

    - Identifies residues

    - Good for residualsolvents and residueson plastics

    - Elements notdetected, samplemust be volatile

    ppb-ppm

    ng-ug/device

    TOC Ideal for polarorganiccompounds(soluble in thelow ppmrange)

    - Detects residues atlower levels thangravimetry.

    - Faster processingtime than HPLC

    - Does not take long todevelop a method

    - Many possiblesources ofcontamination

    - Organic solventscannot be used

    - No identification ofthe residues

    - The extraction ratioof eluant to residuemust be carefully

    controlled foraccurate analyticalresults.

    0.2mg/device

    Non-specific

    - Quick rapid results

    - A level has beenestablished forpurified water whichrepresents a goodtarget level forresidual analysis.

    - Residue mustcontain significantamounts of organiccarbon, it must bepossible to oxidizethe carbon and theresidues must bewater soluble. Notgood for lubricantsand coolants mineral oil-basedprocessing aids

    - Does not detectinorganiccontamination

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    12/17

    12

    Analysis

    Method

    Residues Pros Cons

    Detection

    Limits

    Other

    Information

    HPLC Can detectany type of

    compound

    - Easy detection ofimpurities or

    potentialcontaminants withinpeaks.

    - Ultraviolet detectionusually requires noadditional reagentsor post column orpre-column reactions.

    - UV detectors are notharmful to thesample, areinexpensive andreadily available

    - Simple

    technique

    - No baseline drift dueto mobile phase

    - Long set up andanalysis time

    - Often requiring oneor two days ofdowntime beforeprocessingequipment can becertified forcleanliness

    - Is not inherentlyspecific

    UV/VisSpectroscopy

    Compoundsthat absorbUV and visiblelight

    Detergents

    - Quantitative - The extraction ratiomust be controlledfor accurateanalytical results

    - Not qualitative

    LCMS Anionic

    - carboxylates

    - sulphates

    Cationic

    - amines

    - quaternaryammoniumcompounds

    Non-ionic

    - glucosides

    - alkylethoxylates

    - Small sample

    size

    - Sample must beliquid or solid

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    13/17

    13

    Table H. Direct Surface Analysis: Comparison of Specific, Common Cleanliness Validation Techniques Residues Identified, Pros and Cons

    Direct

    Surface

    Analysis

    Method

    Residues Pros Cons

    Detection

    Limits

    Other

    Information

    ToF-SIMS Inorganic andorganicmolecularspots >0.01um

    Solidcontaminants

    Localisedresidues

    Dry staining

    residues

    - Does not destroy themedical device sample

    - Produces surface areamap - identifying thelocation of thecontaminants at aresolution

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    14/17

    14

    Direct

    Surface

    Analysis

    Method

    Residues Pros Cons

    Detection

    Limits

    Other

    Information

    FTIR Organic andinorganics,particles >10um

    Solids/liquidsor bothdepending onthe technique

    Best for non-polar organicresidues (dueto required

    solvents) -lubricants,coolants,greases, andeven forresidues fromfingerprints.

    - Detect residues at muchlower levels thangravimetric analysis

    - Sensitivity of 1000ppm

    - Obtain informationabout molecular groups

    - Semi-quantitative

    - Detects all elementspresent

    - Good recovery ofmineral oil-based

    processing aids on PETand metal.

    - Capable of identifyingorganic functionalgroups and oftenspecific organiccompounds

    - Extensive spectrallibraries for compoundidentification

    - Ambient conditions (notvacuum; good forvolatile compounds)

    - Typically non-destructive Minimumanalysis area: ~15 micron

    - Uses chlorinated orchlorofluorinatedozone depletingsolvents, whichrequires special safetyprecautions

    - Limited surfacesensitivity (typicalsampling volumes are~0.8 m)

    - Minimum analysisarea: ~15 micron

    - Limited inorganic

    information

    - Typically notquantitative (needsstandards)

    >1%

    1000ppm

    It can beused as ascreen toidentifypotentialcleaningagents

    SEM/EDX

    SEM:

    spots >5nm

    Solids

    EDX: particles,spots >1um

    Localizedresidues

    - Depth of focus

    - Produce high-resolutionimages giving structuraland morphologicalinformation

    - Rapid results- Identifies the elementspresent above Carbon inthe periodic table

    - Cant quantify amountof contaminantpresent

    - No depth profiling

    - Elements of lowatomic number aredifficult to detect byEDX.

    - SEM may spoil samplefor subsequentanalyses

    - Vacuum compatibilitytypically required

    - Ultimate resolution is astrong function of thesample andpreparation

    - May need to etch forcontrast

    EDA -

    EDX: 0.1 -0.5%

    Can becombinedwithEnergydispersiveanalysis toquantifyamount ofresiduepresent

    and giveelementalinformation

    Light

    Microscopy

    Solids - Create Images - No quantification and

    no elements aredetected

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    15/17

    15

    Direct

    Surface

    Analysis

    Method

    Residues Pros Cons

    Detection

    Limits

    Other

    Information

    Visualexamination

    Visibleorganic andorganicparticles,

    spots > 20um

    - Cheap

    - Quick

    - Inexpensive

    - Not specific

    - Not quantitative

    - Can only be used formedical devices with alarge surface area withlarge quantities ofcontaminants

    1-4mcg/cm2(Booth,1999)

    TEM Solid

    Atomic layer

    - Produces high-resolutionimages giving structuraland morphologicalinformation

    - Does not producequantified andelemental information.

    - Requires a very thinsample

    - Expensive

    Canproducequantifiedandelementalinformationifcombinedwith EEL

    3D NonContactSurfaceProfiling

    Solid - Produce an image withtopographicalinformation such asroughness, peak height,valley depth plus surfaceimages

    - Quantification

    xyresolution

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    16/17

    16

    Class I and II medical devices are exempt frompremarket review or are subject to the 510(K)premarket notification process (FDA, 2003).Under the 510 (K) process, a device can becleared for marketing based on demonstration bythe manufacturer that the device is substantially

    equivalent to one or more legally marketedpredicate devices that dont require a PMA(Hogan and Hartson, 2009).

    Class III devices require a premarket application(PMA) providing evidence demonstrating thesafety and effectiveness of the product prior tomarketing (FDA, 2003).

    Depending on the use of a combo device, it mayrequire either PMA or 510K approval prior toentering the market (FDA, 2003).

    510(K) fees are $3693 as a standard fee and

    $1847 for small businesses (FDA, 2003).

    PMA fees range from $200,725 to $599,366, withlower fees for small businesses (FDA, 2009). Thefees are even higher for combination devicesbecause an efficacy supplement for the biologiccomponent needs to be submitted to the FDA,raising the application fee by an additional$200,725 (FDA, 2009).

    Once the device is on the market, periodicreporting, including cleanliness validationreporting, for the device is needed costing $7.025per annum (FDA, 2009)

    The regulatory system in the EU for medicaldevices is similar to the FDA process. However,independent bodies are often hired to carry outvalidation processes and there is no requirementfor prospective, randomized controlled clinicaltrials for PDA applications. Therefore, theapproval process for drugs in the EU is oftenshorter than the approval process in the US.

    For manufacturers and cleaners of medicaldevices, Ceram offers a range of independently-verified cleanliness analysis techniquesparticularly focusing on surface analysismeasurements. With more than twenty five yearsexperience, Cerams materials and surfaceanalysis group is Europes leading surfacecharacterisation company. Ceram has developedthe unique Validata cleanliness index (CI) whichexpresses surface cleanliness as a single figureranging from 0 to 100% derived from a complexcombinatorial algorithm (Pickles, 2008). Thisindex enables a simple comparison betweensamples and permits the monitoring of processtrends; the index also enables the easycommunication of surface cleanliness validationfindings.

    CASE STUDIES

    Ceram analysed metered dose inhaler (MDI)barrels using XPS and ADXPS. Using thesetechniques, they found significant silicone

    contamination of some of the barrel surfaces,thought to have originated from a lubricantapplied to the barrel assembly duringconstruction or filling of the device. Theinvestigation also identified thinning of thefluorinated surface coating on the device.

    Two direct surface analysis methods were alsoused by Ceram to analyse a stain on the outersurface of Aluminium Metered Dose Inhaler (MDI)cans. Using both XPS and ToF-SIMS, Ceram wasable to characterise the stain and identify it asorganic in nature containing fatty acids oftenfound in lubricating oils.

    The validation methods offered by Ceram havebeen used to verify that changes to cleaningprocesses and revised manufacturing methodsfor a medical device did not detrimentally affectthe cleanliness of the device. For example, Ceramused SEM/EDX to verify for Joint ReplacementIndustries (JRI) that residues limits of embeddedparticulates from the polishing and blastingprocesses in the manufacture of orthopaedicimplants had not been exceeded. Ceram alsoused XPS to compare the elemental andoxidation-state composition of the devicessurface before and after the implementation ofthe new cleaning and manufacturing processesfor the device. Using this information fromCeram, the medical device manufacturer was ableto demonstrate that its new productionprocesses produced cleaner medical devices thanbefore and ensured that a break in productiondid not occur.

    CONCLUSION

    Of the various analytical methods available tomeasure device cleanliness, and thereby trulyvalidate the cleaning process, only surfaceanalysis can inform the level of adsorbedcontaminants.

    Ceram offers a unique form of surface analysistailored for cleanliness process validation for anytype of cleaning method.

  • 8/12/2019 Cleanliness Validation White Paper Medical Device

    17/17

    ABOUT LUCIDEON

    Lucideon is a leading international provider of

    materials development, testing and assurance.

    Through its offices and laboratories in the UK, US

    and the Far East, Lucideon provides materials

    and assurance expertise to clients in a wide range

    of sectors, including healthcare, construction,

    ceramics and power engineering.

    The company aims to improve the competitive

    advantage and profitability of its clients by

    providing them with the expertise, accurateresults and objective, innovative thinking that

    they need to optimise their materials, products,

    processes, systems and businesses.

    ABOUT THE AUTHOR

    DR CHRIS PICKLES - CONSULTANT

    TO LUCIDEON

    EXPERTISE IN: AUTOMOTIVE; POLYMERS;SURFACES & COATINGS

    Chris holds a Degree in Chemistry, a PhD in

    Polymer Science, and a Postdoctoral Fellowship.

    AEROSPACE

    Chris has been supplying surface analysis

    capabilities to the aerospace industry for over

    three years with particular emphasis on carbon

    reduction programmes involving composite

    developments, coating analysis and lubricant

    developments in relation to the introduction of

    biofuels.

    AUTOMOTIVE

    Chris has worked in both the aftercare sector as a

    Company Technical Manager and in tier one

    supply chain manufacturing as Managing

    Director.

    Chris has been responsible for the plasticinjection moulding and blow moulding

    manufacture of automotive component systems

    including highly technical mouldings such as fuel

    tanks and 3D spoilers. In addition Chris has alsomanaged an integral supply chain utilising Toyota

    production system protocols.

    POLYMERS

    During his career, Chris has spent four years

    researching copolymer design for bulk property

    manipulation and the statistical mechanics of

    PVC to determine conformational sequencing.

    Chris's knowledge also encompasses plastics

    manufacturing, including injection moulding of

    glass-filled nylon and co-extrusion blow moulding

    of complex 3D components.

    SURFACES AND COATINGS

    In the field of surface science, Chris has

    conducted research projects on alternative

    material sources for surfactants and detergent

    product re-formulation. These include the re-

    launch of a branded fabric washing product in

    Brazil and the design of a surfactant system

    utilising renewable resources. As Technical

    Manager in the automotive aftercare industry he

    has managed the development and qualitycontrol of spray paints for high speed aerosol

    filling.