i-CHROMAT
M
mmunofluorometer Equipment
Reader
Operation Manual
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 2/20
I Quick Manuals E.1
II Introduction and Principle of Operation E.2
III Contents of System E.3
IV Specifications E.4
V Initial Installation Procedure E.5
VI Printer Installation E.6
VII Operation E.7
VIII Warnings, Precautions, and Limitations E.8
IX Service, Maintenance, and Disposal E.9
X Trouble Shooting E.10
XI Contact Information E.11
En gl i s h
TABLE OF CONTENTS
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 3/20
Section I
Q ui c k
M an u al
I
Rev. 10 E.1
For finish a test,
press “In/Out” button to put the holder back
Power Switch OFF
Quick Manual for TEST MODE
I
II
III
IV
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 4/20
This manual contains the introductions for the operation of i -CHROMATM
Reader and for the sample testing procedures. Refer to the test procedure of
specific product for specific test and operating instructions.
i -CHROMATM Reader, manufactured by BodiTech Med Inc., is a portable
instrument for fluorescence detection to quantify concentration of various kind
of analytes in blood or urine. The assays and the instrument are for in-vitro
diagnostic use only. i -CHROMATM Reader can be used in a central
laboratory or in a point-of-care testing facility. i -CHROMATM Reader uses a
laser as the excitation light source. The emitted light from the fluorescence dye
is collected and converted into an electrical signal. The signal is closely related
to the amount of fluorescing dye molecules present on the spot under
examination.
After a buffer-mixed sample is applied to the Test Device, the Test Device is
inserted into i-CHROMA Reader and the concentration of analyte is
calculated by a pre-programmed calibration process. i -CHROMATM Reader
can accept Test Devices that are designed specifically for use with this
instrument.
Section II
Introduction and Principle of OperationI nt rod ct i on
II
Rev. 10 E.2.1
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 5/20
i -CHROMATM Reader is equipped with a built-in Test Device
holder and does not require an external holder to place the Test
Device in. The Test Device holder will appear through the
opening on the front surface of the Reader.
The main power switch, located on the left side of the instrument,
powers the i -CHROMATM
Reader. The optional printer comes
with its own power supply. These two power supplies have
different power ratings, and a critical damage on instrument can
thus be resulted in if power supplies are connected incorrectly.
Power supplies are labeled, indicating that which unit are
supposed to work with.
I n t r o d u c t i on
II
Rev. 10 E.2.2
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 6/20
Section III
Contents of System
Unpacking
After opening the box, compare the contents with the list below.
Contents i -CHROMATM Reader set
i -CHROMATM Reader 1
Operation manual 1
Power cable for Reader 1
Connection cable 1
Printer set (Optional)
Thermal printer 1
Manual for printer 1
Printer paper (roll) 1
Pipette set (Optional)
Pipette (10, 20, 30, 75, 100ul) 1
Capillary tube (5ul) 1
System Check Chip set
Cartridge 1
ID chip 1
Cont ent s
III
Rev. 10 E.3
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 7/20
Physical Description
Dimensions 250 (L) x 185 (W) x 80 (H) mm Weight 2 kg
Power supply 100-240V AC, 50/60Hz, 0.5-1.3A
Data output: On-board screen / Printer
Environmental Set-up
Temperature 15oC ~30oC
Humidity 10 ~ 80%
Location Dry, clean, flat, horizontal surface away from
direct sunlight and mechanical vibration.
Optical Description
Light source Laser diode, 2.5 mW, 637nm
Detector Silicon photo diode
Other
Driver Motor 12V
Interface RS-232 serial (I/O) port
Printer Thermal
Display LCD (16x4 character)
Key pad 5 function keys
*** This device meets the EMI guideline as per EN60601-1-2.
Section IV
Specifications
S p e c i f i c a t i on s
IV
Rev. 10 E.4
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 8/20
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 9/20
4. Turn on the power switch on the left side of i -CHROMATM Reader.
5. i -CHROMATM Reader will execute a series of self-checking routine.
Upon successful completion, the display monitor will show the initial
screen as shown below. (The actual contents of the display could be differ,
depending on the specification and customer requirements.)
Caution: Keep the cables, cords and wirings for the power adapter and
the printer away the front and the top surface of i -CHROMATM
Reader. These could render it hazardous to connect/disconnect the
power adapter (and the printer).
6. i -CHROMATM Reader is now ready for use.
7. At the end of the run, remove the Test Device from i -CHROMATM
Reader. Also, make sure the sample holder is retracted before turning
the power off.
M E N U 22D13:13:00
* Test Mode
Setting Mode
R e a d er
I n s t al l a t i on
V
Rev. 10 E.5.2
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 10/20
ection VI
rinter Installation
i -CHROMATM Reader comes with an optional printer that preconfigured to
work. For proper handling and installation of the printer, refer to the
instruction manual accompanying the printer. With the printer ready for
operation, you can connect the printer to i -CHROMATM Reader following
steps listed below.
1. Connect the other of the cable to i -CHROMATM Reader.
2. Now the printer is ready for operation.
3. Turn on the power switch on the left side of Printer.
P r i n t e r
VI
Rev. 10 E.6
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 11/20
Section VII
Operation
The operation of i -CHROMATM Reader is performed by five primary
functions each of which can be represented by corresponding key pad on thefront panel. The software in i -CHROMATM Reader is designed concisely
and user-friendly to be used in a point-of-care or a central laboratory setting.
Each of these functions is addressed in this section.
Function Keys
The “cursor” in these descriptions stands for the “>>” notation.
This function selects and executes the command at the cursor
position
This function moves the sample carrier in and out of the main
body of i -CHROMATM Reader. This key is also used to finish
the run: the sample holder stays extended at the end of the run
and this key retracts the holder back into the main body.
This function returns the screen to the previous mode.
This function moves the cursor up a line on the display window.
This function moves the cursor down a line on the display
window.
O p e r a t i on
VI
Rev. 10 E.7.1
Down
Up
Reset
Select
In/Out
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 12/20
Procedure for System Check
The following steps describe recommended procedures for a test in i -
CHROMATM Reader with a System Control Device.
On-Screen displays and applicable steps
When you turn power switch on, you will
see ”MENU” on–Screen after “Self Test
Mode”.
Insert
ID Chip
Choose “Test mode” by moving cursor with
“Up” or “Down” key. Press “Select” key to move
next stage after you chose a specific mode.
“Insert ID Chip” is displayed on the screen.
Insert a System Check ID chip.
Ope r at i on
VII
M E N U 22D
13:13:00
* Test Mode
Setting Mode
Rev. 10 E.7.2
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 13/20
ystem Check
Reading
Progressing
i -CHROMATM Reader will scan the System
Control Device automatically.
Prepare a System Control Device
And then, you should do a test
Insert a System Control Device into the holder
after a sample loading .
Press “Select” key to start.
Insert Control
Cartridge
Select to Start
System Check
System OK
The test result will show up on-screen. You can
also get the print-out of test result.
If you press “Reset” button, Screen will return to
“MENU” mode. Now you can run another test.
O p e r a t i on
V
Rev. 10 E.7.3
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 14/20
Procedure for Test Mode
The following steps describe recommended procedures for a test in i -
CHROMATM Reader with a specific Test Device. You can use onlywhole blood for test (No serum or commercial controls)
The actual display may differ from the ones shown below, depending on
the factory configuration of the Reader .
If the scanning is stopped by accident or error message is displayed on the screen, you
will hear an alarm sound. Please check out the cause of alarm sound.
On-Screen displays and applicable steps
When you turn power switch on, you will
see ”MENU” on–Screen after “Self Test
Mode”.
Insert
ID Chip
Choose “Test mode” by moving cursor with
“Up” or “Down” key. Press “Select” key to move
next stage after you chose a specific mode.
“Insert ID Chip” is displayed on the screen.
Insert a proper ID chip which you want to test.
Ope r at i on
VII
M E N U 22D
13:13:00
* Test Mode
Setting Mode
Rev. 10 E.7.4
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 15/20
sCRP Test
Reading
Progressing
OO %
i -CHROMATM Reader will scan the Test Device
automatically after 3 min for immune reaction.
Prepare a Test Device and apply sample mixture
onto the Test Device.
And then, you should do a testInsert a Test Device into the holder after a sample
loading .
Press “Select” key to start.
Insert hsCRP
Cartridge
Select to Start
hsCRP Test
O.O mg/L
The test result will show up on-screen. You can
also get the print-out of test result.
If you press “Reset” button, Screen will return to
“MENU” mode. Now you can run another test.
O p e r a t i on
V
sCRP TestReaction
Progressing
05 : 00
Rev. 10 E.7.5
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 16/20
MENU * SELECT
SET * RESET
year. month. day
hour : min
“Time Set” mode is for date/time setting only
when you activate the instrument for the first time
or reuse after not using the instrument more than a
month. After pressing “Reset” key, you can set year
blinking with “Up” or “Down” key.
Procedure for Time Setting and Recall
Previous Results
Choose “Setting Mode” and “Time Set” or
“Recall Result” by moving cursor with “Up” or
“Down” key. Press “Select” key to move next stage
after you chose a specific mode.MENU 22D
13:13:00
Recall Result
* Time Set
PRE * RESET
NEXT * SELECT
year. month. day
hour : min
Use “Select” key to move year to month to dateto time while setting month/date/time with “Up”
or “Down” key. Press “Select” key to go back to
Menu mode when you finish setting date/time.
Ope r at i on
VII MENU 22D
13:13:00
Test Mode
* Setting Mode
hsCRP Test 01
00.0 mg/L
year. month. day 13:13
“Recall Result” mode is used to store and recall
previous tests. You can store up to 100 test results
and can also print the test result. You can find thetest result by searching date and time of the test
with “Up” or “Down” key. For printing the
previous results, press “Select” key.
Print Data
* OneAll
Choose “One” mode for single result or “All”
mode for total data with “Up” or “Down” key and
then press “Select” key.
Rev. 10 E.7.6
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 17/20
The test sample should be applied to the “B” well, as shown below.
Caution in handling the cartridge
• When inserting the Test Device into the holder, orient the Test
Device in such a way that the “B” well should be toward i -
CHROMATM Reader. This direction is also marked with an arrow
on the Test Device .
• Push the Test Device all the way into the holder until the Test Device
comes to a stop. Use only the push of your thumb or the index finger.
Don’t use of excessive force on the Test Device, or it will result in a
mechanical failure.
• The spent Test Device should be treated as a potential bio-hazard and
should be disposed of according to the standing ordinance. We
recommend high temperature autoclaving, followed by incineration.
A
B
A B
O p e r a t i on
VI
Rev. 10 E.7.7
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 18/20
Do not insert anything other than a Test Device provided by themanufacturer into the sample holder.
Do not immerse i -CHROMATM Reader into water.
Do not spill any liquid on i -CHROMATM Reader. This may disable the
system.
Do not drop i -CHROMATM Reader.
Do not subject i-CHROMA to mechanical shocks heavier than light taps
with hands.
Do not open i -CHROMATM Reader without written authorization from
its representative.
Do not place heavy objects on i -CHROMATM Reader. This may damage
the optical alignment to result in degraded performance or in mechanical
damage.
Use only the power adapter provided with i -CHROMATM Reader.
Use i -CHROMATM Reader only in environment as specified in section
IV .
Use the test result as a pre-screening guide. The result should be
interpreted by trained medical personnel only.
Section VIII
Warnings, Precautions and Limitations
Cautions should be taken to use i -CHROMATM Reader in settings
meeting the requirements specified in Section IV of this manual and in
observance of the warnings in Section VIII . If used otherwise, the
built-in safety features of i -CHROMATM Reader may be compromised
to present severe electrical, mechanical or biological hazards to the
user.
W ar ni n g s
P r e c aut i ons
VIII
Rev. 10 E.8
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 19/20
Section X
Trouble Shooting
Call customer service.Internal Electrical or Mechanical Fault
Message:“Error 21~24”
User’s action is not required. Normal mechanicalsetting
Clicking noisewhen the holder extends
Call customer service.Electric circuit failure
Ground the surface of i -CHROMATM Reader to electricground terminal. Remove anysource of electric charge. Resetthe system.
Electrostatic shock Display does notfunction properly.
Turn off the power and startfresh
Software Abnormality
Call customer service.Mechanical failure(Buzzing noise)Sample holder
does not show.
Turn the power off and on again.ComputationalAbnormality
Wait till the computation isfinished.
Excessivecomputational loadRun finished and
no resultreturned.
Call customer service.Faulty power adapter
Turn on the main switch.Main switch is off.
Remove the power cable (fromSMPS to i -CHROMATM Reader) and reconnect firmly.
Poor connection between the power adapter and the cable.
Check if the wall outlet is alive.Power failure
Nothing happens
Recommended MeasuresProbable CausesSymptoms
Symptoms users can take corrective measures
Tr o u b l e
X
Rev. 10 E.10.1
8/3/2019 Chroma Manual 090120
http://slidepdf.com/reader/full/chroma-manual-090120 20/20
Symptoms Probable Causes Recommended Measures
Message:
“Error 25~28”
Technical problem with the
Test Device
Retest using fresh sample
or device.
Message:“Error S1 or S2”
Optical failure Call customer service.
Message:“Value Error 1-10”
Unusual sample or incorrectstorage of reagents
Retest using fresh sampleor device.
Message:“Caution Error” or
“Error”
Technical problem with theTest Device
Replace the Test Devicewith a new one and repeat
the test.
Message:“Caution BarcodeError”
Test Device alignment is off. Remove the Test Deviceand try again. Repeatwith a new Test Device .
Call customer service.Message:“Lot Number Error”
Message:
“ID Chip Lot Number Mismatch”
Does not match the Test
Device lot # with ID Chip #
Check the Test Device lot
# and ID Chip #.
Message:“ID Chip CartridgeMismatch”
Does not match the TestDevice Item with ID ChipItem
Check the Test DeviceItem and ID Chip Item.
Message:“Insert ID Chip”
No ID Chip Insert ID Chip onto theinstrument.
Message:“Hb Low”
Hb conc. of the sample is below 5.0 g/dL
Retest the sample andmake sure that a correctvolume of detector isapplied.
Message:“Hb High”
Hb conc. of the sample isabove 25.0 g/dL.The volume of detector is toosmall
Retest the sample andmake sure that a correctvolume of detector isapplied.
Tr o u bl e
S h o o t i n g