Chemical Analysis of metal samplesusing

Optical Emission Spectroscopy (OES)

Glen ThieleSpectrosource Pty Ltd

1Spectrosource Pty Ltd copyright 2019

Page 2: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

What is Optical emission spectroscopy (OES) also known as Atomic emission spectroscopy (AES)?

2Spectrosource Pty Ltd copyright 2019

Page 3: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Aggregate state of matter

3Spectrosource Pty Ltd copyright 2019

Page 4: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Atomic physics – How photons are emitted

4Spectrosource Pty Ltd copyright 2019

Page 5: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Optical emission Spectrometers

ICP OES

Arc/Spark OES

LIBS OES

5Spectrosource Pty Ltd copyright 2019

Page 6: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Why Use Spark OES Analysis?

•Most widely used, reliable and reproducible metal analysis technique.

•Wide choice of instruments.

•Excellent LOD

•CCD based spectrometers allow a high number of elements and bases.

•Detection of N is possible with high end portable and high end laboratory spectrometers.

•Detection of O is possible with high end laboratory spectrometers.

6Spectrosource Pty Ltd copyright 2019

Page 7: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

PMT based Spark OES architecture

7Spectrosource Pty Ltd copyright 2019

Page 8: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

CCD based Spark OES architecture

8Spectrosource Pty Ltd copyright 2019

Page 9: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Spark Stand architecture

9Spectrosource Pty Ltd copyright 2019

Page 10: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Typical spark sequence

10Spectrosource Pty Ltd copyright 2019

Page 11: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Grating propertiesPrism:

• non-linear refraction

• affected by temperature

due to influence of

medium density

Grating:

• linear diffraction

• different orders

11Spectrosource Pty Ltd copyright 2019

Page 12: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

12

Reciprocal dispersion as criteria of line separation

500

750

1000[mm]

diameters of

Rowland circuit

lines/mm

focal length [mm]

3600

500

0,51

2400

750

0,52

1800

1000

0,51reciprocal dispersion

[nm/mm]

center

entrance slit

area for

detectors

grating

3600 grooves/mm grating

Spectrosource Pty Ltd copyright 2019

Page 13: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Spark OES Spectrum 130nm – 800nm

13Spectrosource Pty Ltd copyright 2019

Page 14: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Calculation of results from Calibration Curve

result comparison to previouslymeasured set of calibration samples

14

Emission spectrometry is a relative measurement method

Spectrosource Pty Ltd copyright 2019

Page 15: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Recalibration, Standardization or Drift correction

15Spectrosource Pty Ltd copyright 2019

Page 16: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Type Calibration

16

Illustration of exaggerated concentration error Δ in a calibration curve where Type calibration correction can be used to improve accuracy.

Spectrosource Pty Ltd copyright 2019

Page 17: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Measurement accuracy and reproducibility

17Spectrosource Pty Ltd copyright 2019

Page 18: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

The analytical Process

18

Sample taking Sample preparation Chemical analysisSpectrosource Pty Ltd copyright 2019

Page 19: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Type Calibration and measure program

19Spectrosource Pty Ltd copyright 2019

Page 20: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Material Data Base / Grade Table

double click

• software compares analysis results with

element concentration ranges and

reports grade maches

• virtually unlimited material data entries

• international material standards as well

as customised data base can be loaded

• Specific material data base can be

linked to a measuring program

20Spectrosource Pty Ltd copyright 2019

Page 21: Chemical Analysis of metal samples using Optical Emission ...€¦ · Typical spark sequence Spectrosource Pty Ltd copyright 2019 10. Grating properties Prism: • non-linear refraction

Thank you. For more information or feedback contact Glen Thiele

mobile: 0420323577email: [email protected]

• References

• (1) V. Thomsen, Modern Spectrochemical Analysis of Metals: An Introduction for Users of Arc/Spark Instrumentation (ASM International, Materials Park, Ohio, 1996).

• (2) Power Point Presentations – definitions and images: Belec Spectrometrie Opto-Electronik(2012 – 2016).

• (3) Various definitions and images: Wikipedia.

• (4) Melting Spectro Analysis Version 6: G Henderieckx – Gietech BV

21Spectrosource Pty Ltd copyright 2019