chapter-vii a novel method to detect and estimate traces...

26
CHAPTER-VII A novel method to detect and estimate traces level contamination of crystalline polymorphic form in Amorphous form by XRD

Upload: others

Post on 03-Oct-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHAPTER-VII

A novel method to detect and estimate traces level contamination of crystalline polymorphic

form in Amorphous form by XRD

Page 2: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

chapter Vll

Introduction:

Wilhelm Conrad RSntgen discovered 1895 the X-rays. 1901 he was honoured

by the Noble prize for physics.

1912: von Laue, Friedrich, and Knipping: "Interference Effects with Roentgen Rays"

passed x-rays through crystal of sphalerite (zinc sulfide); distinct diffraction pattern

observed.

(a) Crystals cause distinct x-ray diffraction patterns due to atoms.

(b) Crystals are composed of periodic arrays of atoms.

1914: English physicists Sir William Henry Bragg and his son Sir William Lawrence.

Bragg showed that the scattering of x-rays could be represented as a "reflection" by

successive planes of atoms within a crystal diffraction pattern can be used to

determine relative positions of atoms within a single crystal (i.e., molecular structure)

now called Bragg's law, d= nX/2 sin9. First single crystal structure: NaCl. 1915:

Braggs awarded Nobel Prize.

There are several X-ray diffraction techniques. Two of the most common are:

Single crystal X-ray diffraction: used to solve structure of crystalline materials

ranging from inorganic compounds to complex macromolecules such as proteins or

polymers. You can learn everything about a crystal structure, but requires a single

crystal. Although obtaining single crystals is difficult, single crystal X-ray

crystallography is a primary method for determining the molecular conformations of

biological interest such as DNA, RNA and proteins.

Powder X-ray diffraction: used to characterize crystallographic structure, grain size,

and preferred orientation in polycrystalline or powder solid samples. This is a

preferred method of analysis for characterization of unknown crystalline materials.

Compounds are identified by comparing diffraction data against a database of known

materials. It can be used to follow phase changes as a function of variable such as

temperature, pressure.

Uses of X-Ray Powder Diffi-action:

The most widespread use of x-ray powder diffraction, and the one we focus on

here, is for the identification of crystalline compounds in amorphous compound by

their diffraction pattern. Listed below are some specific uses:

194

Page 3: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHapterVlI

Identification of single-phase materials - minerals, chemical compounds,

ceramics or other engineered materials.

Identification of multiple phases in microcrystalline mixtures (i.e., rocks).

Determination of the crystal structure of identified materials.

Identification and structural analysis of clay minerals

Recognition of amorphous materials in partially crystalline mixtures

Crystallographic structural analysis and unit-cell calculations for crystalline

materials.

Quantitative determination of amounts of different phases in multi-phase

mixtures by peak-ratio calculations.

Quantitative determination of phases by whole-pattern refinement.

Determination of crystallite size fi-om analysis of peak broadening.

Determine of crystallite shape from study of peak symmetry.

Study of thermal expansion in crystal structures using in-situ heating stage

equipment.

Basic overview of solid forms

Organic molecular solid Molecular weight

I I Amorphous Crystalline

^ — ^ X Polymorphs Hydrates

.—^-. r^—. Monotropic Enantiotrqpic Reversible Non>reverstt>le .4 p.

Ail can have numercnis habits

Figure 1: Basic overview of solid forms.

From the above figure 1 we can say that a solid can exist in two forms viz.

either amorphous or crystalline. In crystalline form a solid can exist as polymorph,

hydrate, solvate or co-crystal.

195

Page 4: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHapterVlI

Polymorphs arise when molecules of a compound attain different

conformations with differing energies of stabilization. Although their chemical

properties may be same but they differ in their structural orientation. Generally a less

stable polymorph converts into its stable polymorph. But the conversion might take

years or few seconds to occur as it depends on the activation energy supplied to the

polymorph and free energy of that solid. Lowest energy form is more stable. At

transition temperature when one form (polymorph) is more stable than the other forms

then they are called as enantiotropic polymorphs and when the all forms are stable at

transition temperature, then they are called monotropic polymorphs. When the

crystal's repeating arrangement unit has a molecule of solvent attached to it then it is

called as a solvate. When the solvate molecule is replaced by water molecule then it is

called as hydrate. A co crystal is a distinct solid-state material with unique,

unpredictable structure and physical property profile '̂̂ . Different polymorphs give a

drastic effect on the dissolution rate which is in turn dependent on the solubility '^l

Amorphous state means lack of form. It has no long range order of molecular packing.

They generally possess higher internal energy, have greater thermodynamic properties

and have increased mobility due to greater intermolecular distances. We can say that

an amorphous system is actually fluid but appears to be a solid in the time scale

observation. This statement can be explained by the following figure 2:

Til T „ T. ,

T«nip*r«tui«

RGURg .Sch«matteftpf««»fttttioftof4KWh»lpy(ofwhim<^v«t»u»t«iiy«wtuwdH»^^ for a Nqijld caiMbI* of oryttatl l^ and forming ̂ fforant glassM at dHforant cooling ratat (Tic Kauzmann tamparatura. T^ glass transition tamparatura. T«: fosien tamparatura).

Figure 2 [3]

196

Page 5: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHapterVlI

Cooling from melt, there is a chance that it may convert to a crystalline form

at temperature Tm. This is due to an exothermic crystallization which occurs due to

decrease in volume and enthalpy. Secondly, if we don't give enough time for the

molecule to crystallize then we don't see discontinuity at Tm and the system enters

into supercooled state. Now on further cooling the viscosity of the compound

decreases and it appears to be in a rubbery state. When we further cool it we observe

that at some point we don't see the decrease in viscosity as rubbery state is converted

to glassy state. The temperature at which this phenomenon occurs is called as glass

transition temperature (Tg). We can say that Tg is a thermodynamic requirement

because otherwise the material may get lower enthalpy than crystalline material or

even get negative enthalpy Tk. This is a significant difference between amorphous

and crystalline materials. Amorphous materials are easy to formulate than crystalline

materials but they degrade faster than the latter (generally).

Paracetamol:

NHCOCH,

o Paracetamol

H

Figure 3: The Structure of paracetamol [4]

Paracetamol is used as an antipyretic and analgesic. Commercially it is

available as tablets, capsules, solutions, suspensions etc. Paracetamol has three known

polymorphic form viz. Monoclinic (Form I), Orthorhombic (Form II) and a third

instable form III. Efforts were made in characterising form III with use of rapid

heating DSC and HPMC (crystal growth modifier) ^̂ l Thermal analysis of the

polymorphs was carried out using FT-IR spectroscopies "̂l Form II is the

commercially available form of paracetamol. However form I has a stififer

arrangements in the crystal lattice compared to the orthorhombic metastable form II.

Hence form I shows poor compression ability *̂̂ . Although form II shows better

compression ability compared to form I but still form I is preferred as API. This is due

to the stability issues related with form II. The stability of the form II varies with

alteration in temperature and humidity. Hence, the stability issue prompts the

197

Page 6: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHapterVlI

manufacturer to liken form I. Efforts should be made to make form II more stable

under ambient temperature and humidity.

Itraconazole;

A I H v_y \ /

Ct

Figure 4: The Structure of Itraconazole '̂ ^

Itraconazole (Figure 4) is an antimicrobial agent. It is commercially used for

treatment of onychomychosis. The main disadvantage for this drug is its solubility. It

is a poorly soluble drug. Hence attempts were made to form a co crystal of the drug

which will increase the solubility of the drug. Amorphous state of the drug has a

greater solubility than the crystalline form but the amorphous form is less stable and

degrades quickly in presence of humidity and temperature. Hence in one such attempt

a CO crystal of itraconazole was prepared using 1,4-dicarboxylic acids ^^\ A trimeric

CO crystal was prepared with itraconazole atoms at the two ends of the dicarboxylic

acid (Succinic acid).

anWtmgal agent

«»i r"^

Co-crystallization The dissolution proflie of co<ry$tal$ with L-malic acid matches that of amorphous cis-itraconazol*

1. Sporjrwx* 2. >-malic JCi<) co-crystar 3. MMtMtc aetd c<xrystai 4. Succtnsc aetd co-crystals 9. c^-itraconazol«

mmtwMH JAmCmm&m 2603, J2S,*««-««?

Figure 5: Dissolution profile of itraconazole

From the figure 5 we can conclude that the co crystal with malic acid matches

the solubility that of the amorphous itraconazole ^^\ Thus this study suggests that it is

198

Page 7: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHapterVlI

possible to have a form which matches the bioavailability of the amorphous form but

has the physical and chemical stability of the crystalline form.

Basic overview of Amorphous Forms:

Celecoxib:

Celecoxib is a nonsteroidal anti-inflammatory drug indicated for Rheumatic

arthritis. It acts as a COX2 inhibitor. It is commercially marketed as Celebrex and is

supplied in the form of capsules.

Figure 6: The structure of Celecoxib *̂̂

The crystalline form of the drug is marketed. However it was found that the

higher bioavailability was shown by the amorphous state. The downfall of the

amorphous state was its stability. This was due to the structural relaxation which

causes devitrification of the celecoxib if stored at room temperature ^'l This was

enhanced by mixing it with polymers like PVP, PVP-Meglumine which helped in

stabilising the amorphous system 1'°""]. A new solid state form was developed by

Pharmacia Corporation and patented ^̂ l The polymer used form a composite with the

amorphous form and prevents its conversion to the crystalline form. Crystalline form

of the Celecoxib is less pharmaceutically active. The amorphous lactose was prepared

from the crystalline one by spray drying techniques with isopropanol. Studies were

conducted to check the stability and solubility of the drug-polymer amorphous system

'^l The measurement of the crystalline content, glass transition temperatures and

enthalpy relaxation were characterised using DSC. Dissolution and solubility studies

were carried out to check the release mechanisms. Hence by these techniques the new

form of Celecoxib was discovered and an improved formulation was developed.

199

Page 8: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

Simvastatin:

CHaperVlI

H.C

Figure 7: The structure of Simvastatin ''^'

Simvastatin is a cholesterol reducing agent belonging to the statin family. It

belongs to class II (low solubility but high permeability). It has two known

polymorphs, however the second polymorph requires extreme conditions to form and

form I is the most known stable form available. Simvastatin is suitable candidate for

amorphous form. In one such study the physicochemical properties of two differently

prepared amorphous forms were prepared and characterized '̂̂ l The two methods

were cryo-milling and quench cooling. In quench cooling the compound was melted

in oven and then quench cooled by putting liquid nitrogen on the plate carrying the

compound. In cryo milling the compound was ball-milled and the mill was

surrounded liquid nitrogen. The amorphous material obtained by two different

methods were further characterized by using FT-IR, DSC, High Speed DSC, TGA,

XRPD, PLS and SEM.

Figure 8: SEM images of simvastatin (a) crystalline, (b) cryo-milled, (c) 4°C milled, (d) RT milled, (e) recrystallized cryo-milled, and (f) quench cooled''^^.

200

Page 9: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

chapter VII

Figure 8 depict the difference in the crystal structure of the amorphous

material prepared by different techniques. Accelerated stability studies were carried

out on the amorphous material prepared by two techniques. The cryo-milled

amorphous product showed less stability compared to quench cooled product. The

logic behind making amorphous form is that it shows more solubility than the

crystalline form. Amorphous form has more bioavailability compared to the

crystalline form. The main disadvantage of this form is that it has excess of

thermodynamic properties hence it tends to recrystallize into its crystalline form. If

we are able to keep the amorphous form stable enough till the expiry of the medicine

then we can achieve a novel and alternative formulation.

Summary of literature survey:

The crystalline and amorphous form of the drug have same molecular

configuration but different physical and chemical properties. The crystalline form of

the drug is not easy to formulate but the formulator is sure that the formulation will be

stable enough till its expiry date. But it is essential to identify the most stable

polymorph early in the development. This is of utter importance as the goal is to avoid

the scenario of ritonavir where the stable polymorph occurred after manufacturing

240 batches which had impaired bioavailability '̂̂ l Compounds used as active

pharmaceutical ingredients (APIs) must be screened for polymorphism according to

ICH guidelines. Whether to use a hydrate form or a solvate form will depend on the

ability of the compound to be stable over varied temperature and humidity. Co

crystals offer a new dimension to the formulator as it has the solubility as that of the

amorphous material but also retains the physical and chemical properties of the

crystalline form. However to choose an ideal co crystal former is of the utmost

importance. The amorphous form of the drug has increased solubility than its

crystalline counterpart and hence has increased bioavailability. The amorphous

material is easy to formulate than the crystalline material due to its high internal

energy and increased mobility. But amorphous state is not a state where the molecules

love to be and hence the molecules try to give out energy and recrystallize to a more

stable crystalline form. This is the major issue with amorphous materials. Little

change in temperature or humidity may largely affect the stability of the dosage form.

Various pharmaceutical processes affect the final form of the drug. Processes like

spray- drying, lyophilisation may cause formation of amorphous form. Amorphous

201

Page 10: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

material is formed during mechanical shearing processes (Ball milling, wet

granulation, compaction). To conclude one must do extended preformulation studies

and then formulate the best possible stable drug available.

Polymorphism specifying the diversity of nature is widely observed in

pharmaceutical compounds. Differences in their physico-chemical and mechanical

properties led to the emergence of characterization based stringent quality control

measures of these altered solid-state forms in active pharmaceutical ingredients

(APIs) and drug products '̂'*l This complexity generally observed due to the

differences inflow properties, compatibility, hygroscopisity, crystal morphology and

hence process ability of different forms ''^l The sudden appearance or disappearance

of a crystalline form can threaten process development, and can lead to serious

pharmaceutical consequences if the transformation occurs during the manufacturing

or storage of the dosage form. Therefore, qualitative and quantitative analyses of

polymorphs must be incorporated early on in the drug development stage, both in the

API manufacturing and the formulation stage ̂ '*l

Pharmaceutical companies engaged in generic products aim at developing a

formulation that is qualitatively and quantitatively closest to the innovator product in

terms of excipients and the API polymorphic form. However, USFDA allows the use

of alternate polymorphic forms as long as the criteria of pharmaceutical equivalence

and bioequivalence are met ^"l

Amorphous materials can be naturally occurring and offer several advantages,

such as insulin (where as Insulin Zn complex is highly crystalline), large molecule

excipients (mostly polymers), and some small drug molecules that are difficult to

crystallize. The pharmaceutical industry often engages in costly high energy

processes, such as lypholization, high energy milling, melt extrusion and

coprecipitation with polymers to obtain an amorphous material. The desired benefits

of amorphous drugs include higher rate of solution and higher kinetic solubility.

Improved bioavailability of poorly soluble drugs is an especially desired consequence

of higher kinetic solubility. Several additional excellent references exist which discuss

the use and characteristics of amorphous materials in pharmaceuticals t'*'̂ °J.

The use of amorphous materials in any field is associated with some challenges.

202

Page 11: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHapterVlI

The transformation from amorphous to a crystalline phase is dependent on diffusion

properties of the amorphous material, as well as the kinetics of nucleation and growth

of the crystalline phase ^ '̂"^^1. Additionally, the identity of the resultant crystalline

phase(s) cannot be absolutely predicted. First, though the identity of the most stable

crystalline form of a drug is sought, it is possible that the most stable form is not yet

identified. Second, there is no guarantee that the amorphous phase will transform to

the most stable crystalline phase '̂*l

Pure amorphous materials are often produced in the pharmaceutical industry.

Materials in the amorphous phase are thermodynamically less stable than any

crystalline form, leading to a tendency for the amorphous materials to transform to a

known or potentially unknown crystalline phase '̂*l The time scale for any

transformation is also unknown and can, in part, be evaluated via stability studies. The

detection of crystalline phases in mixed systems is often performed by powder X-ray

diffraction.

In this work, a limit of detection method was established to detect low levels

of crystalline material in an amorphous material ['*"''].

Materials and Methods:

Instrumentation:

XRD patterns of pure amorphous, pure crystalline and amorphous/crystalline

mixes were collected at room temperature on Bruker D8 Advance diffractometer

(Kralschue, West Germany) in which used the Bragg-Brentano configuration, a

copper X-ray tube (Cu Ka=1.5406 A) and Scintillation detector.

Experimental conditions/Method:

The Bruker D8 Advance configuration included power of 40 kV x 40 mA, Cu

Ka radiation passing through nickel filter, 0.3 mm divergence slit, 0.3 mm anti-

scattering slit, 0.1 mm detector slit and 0.6 mm receiving slit. The scan 20 ranges 3.0

to 60°, step size 0.03° and time per step O.Ssec. The sample holder was rotated in

plane parallel to their surface at 30 rpm during the measurement. Obtained

diffractograms were analysed with DIFFRAC'"* diffraction software.

203

Page 12: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHapter'UlI

Samples:

This report will mainly focus on two material Atorvastatin Calcium

Amorphous (figure 9), Atorvastatin Calcium crystalline known form (figure 10) and

Fluvastatin Sodium Amorphous, Fluvastatin Sodium crystalline known form. To

determine the limit of detection in each case, blends using the amorphous material and

its related most stable known crystalline API phase, were prepared directly on the

sample holder using the manual blend method. All standard mixes are expressed as

%crystalline phase/total API. For example, a 2.0% crystalline standard contains 2.0%

crystalline per total API. The Si Zero background sample holders (Supplied by

Bruker). All sample holders were prepared with flat 511 Si inserts to assure sample

depth of 0.5 mm.

iUi '̂ M'̂ ^*.l»uw^<..l,A,̂ >w. spsa

•KSr.rarr*

Figure 9 Figure 10

METHOD DEVELOPMENT, EXPERIMENTAL DESIGN AND STUDY:

Several data analysis methods were evaluated to obtain a reproducible analysis

technique to observe the limit of detection (LOD) of a crystalline Form 1 in

amorphous material A. With the noise generated by the diffuse disorder scattering

fi-om the amorphous material, reproducible interpretation of a peak is difficult, but

necessary for a validated method. The FDA (food and drug administration) published

guidelines for LOD methods, which suggested LOD can be determined by visual

evaluation, signal-to-noise evaluation and Standard Deviation of the Response and

Slope. This last method is not possible for this case because the signal is not linear

with the degree of crystallinity. Also, as pointed out previously, the method must be

general enough to detect any peak, not just the peaks fi"om the most stable known

form. This means that peak detection at known 2 D values was not appropriate, though

204

Page 13: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

chapter'V'lI

this is the most common method of analysis and was certainly used for method

development and identifying limit of detection standards. Often, a sufficiently trained

scientist can recognize a peak, which would make the visual evaluation possible.

However, validation of that visual recognition is close to impossible. Therefore, a

numeric approach to analysis must be taken. Jade, 10a program which reads several

PXRD file formats, was used for the analyses. Smoothing algorithms, background

subtraction and peak finding routines were evaluated. The Fast Fourier Transform

(FFT) smoothing algorithms work well with crystalline materials but tend to introduce

low intensity peaks for an amorphous material, which lead to problems distinguishing

real peaks from FFT-generated peaks. The traditional smoothing methods, including

the Sovitsky-Golay least squares smoothing algorithm, tended to reduce the

detectability of low intensity peaks. Background subtraction techniques often did not

adequately describe the background of a halo. This version of Jade allows the scientist

to select the background more accurately, but again, that lead to validation and

repeatability concerns. The analysis that appeared most reproducible was a simple

peak finds routine, with a parabolic filter, variable filter length, summit peak location,

0.1% intensity cutoff, 1.0 20 range to find background, and 7 points for the

background average. Normally, a threshold of 3a is used for peak detection, but 5a

was required due to the occasional interference of noise from the amorphous halo.

The limit of detection of the known most stable crystalline form for each

system was determined by making mixes of crystalline compound in the amorphous

compound.

Method development:

The calculated penetration depth information was used as a guide, but was not

expected to be accurate since the sample materials were imagined has a not pure

crystalline materials, nor isotropic but homogeneous. Tests were performed to

determine the effect of depth of penetration so that the appropriate sample depth,

especially for standard preparation, could be designed. Instead of thoroughly mixing

the 5.0% crystalline form with the amorphous blend, 5.0% crystalline form was

layered in the sample holder at 0.5 mm depth and at the top surface. Figure 11, which

demonstrates the effect of depth of penetration, shows that crystalline form can be

detected more precisely when layered on top surface, but not at 0.5 mm depth. The

powder pattern for 5.0% crystalline form on the top surface shown similar pattern

205

Page 14: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHapterVlI

with that of crystalline form. Figure 11 is the diffractogram of crystalline form layered

on upper surface and lower surface of the sample holder, overlay. But in case of

pharmaceutical industry if there is a contamination or conversion of amorphous to

more stable crystalline form then it will be for entire bulk. So uniform mixing of

crystalline form with that of the amorphous method is adapted for the study rather

than layering on the top surface. In-order to suffice the intended purpose.

*»»k*M*rtnkH*W

•Uli'l^lllVHtl i

Figure 11 Figure 12

It begins with defining the requirements and the information desired from the

technique. The purpose of the method discussed here is to demonstrate that the test

samples contain amorphous material without evidence of crystalline materials, within

the limits of detection. A "low" limit of detection is desired, with the eventual desire

to relate degree of crystallmity to product performance, such as dissolution rate. The

development of a technique to understand the structure (truly amorphous and/or

crystallites too small to be detected by X-ray) of the amorphous material is not a goal

of this method, and would require more advanced instrumentation. The method,

including interpretation of the powder patterns, must be able to be run and analyzed

reproducibly by several scientists, and possibly may need to be transferred to a

production site.

A full understanding of the material to be tested is also important. The test

samples are defined as containing API that is amorphous by X-ray. There is a

possibility that some of the samples will also contain API that has transformed to a

crystalline phase. Thus, the expected PXRD pattern will be an amorphous halo,

sometimes with peaks due to crystalline excipients. The limit of detection was

determined using the most stable crystalline phase. One assumption is that the

206

Page 15: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHapterVlI

amorphous-to-crystalline transition can be simulated with the physical mixture added

to the amorphous material. This assumes that any preferred orientation in the physical

mixture is similar to the naturally occurring crystallization in amorphous material.

Since there is no guarantee that the most stable known phase will be the result

of the transition from the amorphous phase, the method must be general enough to

pick up any new crystalline peaks. The kinetics of the amorphous-to-crystalline

transition is unknown and likely to depend on temperature, humidity and physical

treatment such as grinding. The standard samples are defined as amorphous by X-ray

and/or the related crystalline form. Preparing standard samples to determine the limit

of detection introduced a sample uniformity challenge that needed to be handled. To

reduce errors from sample loss, standards were prepared by adding amorphous

material directly on the sample holder. The instrument and sample holder

configuration can also be selected to reduce sample uniformity issues. The first

selection is for an instrument with a wide beam so that the largest sample surface can

be analyzed ''*l

The scan range, scan step size, time per step and the analysis method were

evaluated simultaneously. The scan ranges were selected based on strong peak

positions and limiting interference, which lead to scan ranges set to 3° to 60° 29. This

allowed the appearance of multiple peaks from the most likely crystal form and a

large enough range to allow detection of a form other than the most stable known

form. The scan step size and time per step were set to optimize signal/noise, with the

optimum appearing to be 0.03° 26 step size and 0.5 seconds per step. Several data

analysis methods were evaluated to obtain a reproducible analysis technique to

observe the limit of detection (LOD) of a crystalline form in amorphous material.

With the noise generated by the diffuse disorder scattering from the amorphous

material, reproducible interpretation of a peak is difficult, but necessary for a

validated method. The FDA published guidelines for LOD methods, which suggested

LOD can be determined by visual evaluation, signal-to-noise evaluation and Standard

Deviation of the Response and Slope.

This last method is not possible for this case because the signal is not linear

with the degree of crystallinity. Also, as pointed out previously, the method must be

general enough to detect any peak, not just the peaks fi-om the most stable known

207

Page 16: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CftapterVlI

form. This means that peak detection at known 2Q values was not appropriate, though

this is the most common method of analysis and was certainly used for method

development and identifying limit of detection standards. Often, a sufficiently trained

scientist can recognize a peak, which would make the visual evaluation possible.

However, validation of that visual recognition is close to impossible. Therefore, a

numeric approach to analysis must be taken, a program Eva which reads PXRD file

formats, was used for the analyses. Smoothing algorithms, background subtraction

and peak finding routines were evaluated. The Fast Fourier Transform (FFT)

smoothing algorithms work well with crystalline materials but tend to introduce low

intensity peaks for an amorphous material, which lead to problems distinguishing real

peaks from FFT-generated peaks. The traditional smoothing methods, including the

Sovitsky-Golay least squares smoothing algorithm, tended to reduce the detectability

of low intensity peaks. Background subtraction techniques often did not adequately

describe the background of a halo. This version of Eva allows the scientist to select

the background more accurately, but again, that lead to validation and repeatability

concerns. The analysis that appeared most reproducible was a simple peak find

routine, normally, a threshold of 1 is used for peak detection I'*"'*!.

The limit of detection of the known most stable crystalline form for system

was determined by making mixes of crystalline in the amorphous material. The limit

of detection was determined to be 3% (refer figure 17) by observing that 3.0%

crystalline form was repeatedly detected, but that lower amounts were not

reproducibly detected. To assure that 3% crystalline form could always be detected,

the source intensity was varied by decreasing the source energy to 35 mA, 30 mA and

25 mA. With 25 mA source energy crystalline peaks were found to be undetected. It

gives an inference that source energy to be monitored using external standard

corundum (AI2O3), intensity counts were recorded for every by day (Diffractogram

were not presented).

Experimental design and Study:

Talcum (figure 12) and NaCl (figure 13) are selected as an external standard

and diffractogram of both are recorded. NaCl is selected as an amorphous blank as it

does not show interference in 20 region were major peaks of Atorvastatin calcium

crystalline form appear and shows straight pattern, mean no halo pattern where as

208

Page 17: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHapUrVlI

amorphous pattern shows. Refer figure 14 for overlay of Talcum and Atorvastatin

calcium crystalline form and for overlay of NaCl and Atorvastatin calcium crystalline

form refer figure 15.

It has been observed major peaks of Atorvastatin calciimi crystalline form, are at 20

9.13 ±0.2° and 21.56±0.2°. Limit of detection is considered based on the detection

of these two peaks of Atorvastatin calcium crystalline form.

Jii.iMiyM.ww " A M T &«K-.jiCACr.<m.liM •

iUMMlfM^Mik

Figure 13 Figure 14

Figure 15 Figure 16

Study details for finding limit of detection by considering NaCl as an External

amorphous standard:

Detection of Atorvastatin calcium crystalline form with NaCl.

a. 1.0% w/w of Atorvastatin calcium crystalline form in NaCl - diifi-actogram

shows absence of peak at 28 9.l3±0.2°and 21.56±0.2°.

b. 2.0% w/w of Atorvastatin calcium crystalline form in NaCl - diifractogram

shows absence of peak at 28 9.13±0.2° and 21.56±0.2°.

209

Page 18: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CfiapterVlI = ^

c. 3.0% w/w of Atorvastatin calcium crystalline form in NaCI - diifractogram

shows presence of peak at 29 9.13±0.2° and 21.56±0.2°; refer figure 16.

In similar manner Atorvastatin calcium crystalline form with Atorvastatin

Calcium amorphous

a. 1.0% w/w of Atorvastatin calcium crystalline form in Atorvastatin

Amorphous - diifi-actogram shows absence of peak at 20 9.13±0.2°and

21.56±0.2°.

b. 2.0% w/w of Atorvastatin calcium crystalline form in Atorvastatin

Amorphous - diifractogram shows absence of peak at 29 9.13±0.2° and

21.56±0.2°.

c. 3.0% w/w of Atorvastatin calcium crystalline form in Atorvastatin

Amorphous - diifractogram shows presence of peak at 29 9.13±0.2° and

21.56±0.2°, refer figure 17.

m —,

ao -ao — m -__ an — 290 -

** ̂ 230 -

HD -

m '

UiJjCt^ 1T0

MO —

ISO —

MO -

I M -

130 -

110 —

10 -

«e —

70 ^

M —

90

« 10 —

at — M —

,

nts)

CMS7B12

1

A

k l/\ ' I uJ V AM/T^TM*u^^JM iJ^ • Yyyjvfr Ti|^v«nrT"

w

^ A T V C a - F M aWctyMillMtnwnorphMO

* > 4 1 W 1

fcAPi .

Mwi* ft ("7 jy ' " H N L

ir^Hf

zo

i^lMm. • M I F M V W I I h A J ^ 1 * 1 u i> II . . u

^^^"^'>''*'wM^wHfV*A<^^

M M

2-Theta - Scale

rf# jj-y^. iij.a ii 1 . . • ' ^ * ^ ' > * * M A A / f k / W v v - « f « i y H r ^ ^

90 00

Figure 17

Demonstrating that the method will work reproducibly is an important step to

completing method validation. In the pharmaceutical industry, the ICH has prepared

guidelines for method validation ^^^\ Most of the guidelines were initially based on

chromatography method development, but many of the concepts, such as specificity,

linearity, precision (refer figure 18), accuracy, sensitivity and robustness, can be used

for other methods, including PXRD methods.

Specificity is the ability to measure a specific analyte free fi-om interference of

other components. In this case, the method was developed to test the 20 range where

210

Page 19: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CfiapterVlI s = s

the amorphous halo or excipients provide the least interference while crystalline form

presence is obvious.

The analysis method was not designed to be specific, in that it is designed to

pick up any peak associated with a known or unknown crystal form. Linearity is

defined as showing direct and proportional response to changes in concentration. The

interferences by diffuse disorder scattering changes with the relative amounts of

amorphous/crystalline content, which interferes with linearity. In practice, we found

that the response is not linear near the limit of detection, which is an expected result.

Therefore, linearity is not a parameter that is tested to validate this limit of detection

PXRD method. Accuracy is also a concept that is not applicable to this method. Since

there is no linearity, we are not using this method to predict the amount of crystalline

Form present; we are simply determining that the amount of crystalline phases present

are either greater than or less than the limit of detection. Precision, which is the

closeness of agreement among a series of measurements, is a parameter that should be

evaluated for this technique. In this case, standards at the limit of detection (LOD) are

run repeatedly and are demonstrated to consistently detect peaks of crystalline form.

Precision:

Atorvastatin calcium crystalline form 9.0% w/w in Atorvastatin calcium

amorphous - diffractogram shows presence of peak at 29 9.13±0.2° and 21.56±0.2°

the six replicate sample preparation of same sample shows almost same intensity at

26 9.13±0.2° and 21.56±0.2°, refer figure 18.

211

Page 20: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

Cfiapterl'n

[A

fJ>»*̂ «rvj*Wv-̂ w.»w»̂ '*

.•«V^w...W^>V*MA>"*v^„,^, / ^ - ~ , wVt.-.v-v--'»«™^yA~''*->-<^(—\-- ••'-^^^,-wU>.~

•"Atorvastatin Ca 9%crystaUine preparation 1 '^Atorvastatin Ca9%crystalline preparation 2 "'Atorvastatin Ca9%crvstalline preparation 3 ^Atorvastatin Ca9%crvstaliine preparation 4

2-Theta - Scale ^AtOTvastatinCa 9%crystalline preparations ^Atorvastatin Ca9%crystaUine preparation 6

Figure 18

In a similar manner, study has been conducted for Fluvastatin Na based on the

study it has been observed, form of NaCl does not have any peak which corresponds

to the major peaks of Fluvastatin crystallhne form, refer figure 19.

_J liL k m fkt\ fi ft l-'-ra -SyK

•]pC£,i t»ttSM*. T\)aS;ee V i ^ *«|iakiiM*NM«Mjblfl rti>ii a | l I i i^^i i • » JL

%l3<-sa> K i s^a^lM'

Figure 19 Figure 20

It has been observed major peaks of Fluvastatin crystallhne form, are at 28

12.21±0.2° and 12.97±0.2°. Limit of detection is considered based on the detection of

these two peaks of Fluvastatin crystallline form.

Detection of.Eluvastatin Crystallline form in NaCl,

a) 1.0% w/w of Fluvastatin crystallline form in NaCl - diifractogram shows

absence of peak at 29 12.21±0.2° and 12.97±0.2°.

212

Page 21: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHapterl^I --—

b) 2.0% w/w of Fluvastatin crystallline form in NaCl- diifractogram shows

presence of peak at 29 12.21±0.2° (12.130°) and 12.97±0.2°. (12.910°), refer

figure 20.

Detection of FluvastatinCrystallline form in Fluvastatin amorphous.

a) Diffractogram of Fluvastatin amorphous refer figure 21.

b) 2.0% w/w of Fluvastatin crystallline form in Fluvastatin amorphous -

diifractogram shows presence of peak at 20 12.21±0.2° (12.279°) and 12.97±0.2°

(12.951°) refer figure 22.

c) Precision: Fluvastatin crystallline form 6.0% w/w in Fluvastatin amorphous -

diifractogram shows presence of peak at 29 12.21±0.2° and 12.97±0.2°, all the six

replicate sample preparation shows same intensity at 29 12.21±0.2° and

12.97±0.2° refer figure 23.

Figure 21 Figure 22

213

Page 22: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CfuifterVn = =

/v̂ fr̂ %

'/ww^v»«Jl^^•^u^

-v.V*,JV,t. ^1^,, . , • 'V •H-''t",-^fi,M-niy!y^i_^ ^.^

•"Fluvastatin Na 6% crystalline preparation I "•Fluvastatin Na 6% crystalline preparation 2 ™Fluvastatin Na 6% crystalline preparation 3 'Fluvastatin Na 6% crystalline preparation 4

2-Theta - Scale

'^Fluvastatin Na 6% crystalline preparation 5 ^'Fluvastatin Na 6% crystalline preparation 6

Figure 23

Demonstrating that the method will work reproducibly is an important step to

completing method validation. In the pharmaceutical industry, the FDA has prepared

guidelines for method validation. Most of the guidelines were initially based on

chromatography method development, but many of the concepts, such as specificity,

linearity, precision, accuracy, sensitivity and robustness, can be used for other

methods, including PXRD methods.

Specificity is the ability to measure a specific analyte free from interference of

other components. In this case, the method was developed to test the 29 range where

the amorphous halo or excipients provide the least interference while crystalline form

presence is obvious. The analysis method was not designed to be specific, in that it is

designed to pick up any peak associated with a known or unknown crystal form

Linearity is defined as showing direct and proportional response to changes in

concentration. The interferences by diffuse disorder scattering changes with the

relative amounts of amorphous/crystalline content, which interferes with linearity. In

214

Page 23: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHapterVlI

practice, we found that the response is not linear near the limit of detection, which is

an expected result. Therefore, linearity is not a parameter that is tested to validate this

limit of detection PXRD method. Accuracy is also a concept that is not applicable to

this method. Since there is no linearity, we are not using this method to predict the

amount of crystalline present; we are simply determining that the amount of

crystalline phases present are either greater than or less than the limit of detection.

Precision, which is the closeness of agreement among a series of

measurements, is a parameter that should be evaluated for this technique. In this case,

standards at the limit of detection (LOD) are run repeatedly and are demonstrated to

consistently detect peaks of crystalline. The related phenomenon, robustness, is often

defined as the capacity of a method to remain unaffected by small, deliberate

variations in the method. In regards to robustness, we demonstrated that the

amorphous material A was not induced to form crystalline API by grinding or holding

the sample for several days beyond the normal analysis time. We also demonstrate

that additional grinding of a standard LOD mix results in a sample from which

crystalline content can still be detected.

Limitations of the Procedure,

(i) The demarcation of the crystalline (Ic) and amorphous (la) intensities from the

experimentally obtained X-ray diffraction pattern is done arbitrarily and is subject

to individual judgment. The procedure may thus lead to high variability in the

results, especially when the degree of crystallinity is low.

(ii) Samples may show preferred orientation effects, which may cause significant

errors in peak intensity measurement.

An effective way to minimize preferred orientation is to decrease the particle size

of the sample by grinding. However, the grinding process itself can cause

undesired changes in the solid state of the material.

(iii) It is necessary to generate a standard curve using samples with different degrees

of crystallinity. A common procedure is to mix crystalline and amorphous

standard phases in appropriate proportions. Non-homogenous mixing becomes a

potentially important issue, especially when preparing mixtures of exfreme

compositions.

215

Page 24: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CHapter'Un

Our investigation had two objectives: (i) to develop an X-ray diffractometric

technique, based on synchrotron X-rays and a 2-D area detector, having increased

sensitivity and rapid data acquisition capability compared to conventional instruments

and (ii) to develop an objective method for calculating the degree of crystallinity

based on an algorithm. The key analytical challenge, from a pharmaceutical

perspective, has been to develop a technique capable of discerning subtle changes in

lattice order in substantially amorphous materials. Our focus was therefore on the

detection and quantification of the first evidence of crystallization in an amorphous

material. An in situ crystallization approach was used to overcome the problem of in

homogeneity in mixing, which is a particularly serious issue at extreme mixture

compositions.

Conclusions:

The method described in this report was developed to determine if amorphous

material has crystalline character that is detectable by PXRD. Limit of detection of

Atorvastatin Ca crystalline from I in Atorvastatin amorphous is 3.0% and method is

found to be precise at concentration 3 times of its limit of detection.(i.e9.0%) and

Limit of detection of Fluvastatin crystalline from B in Fluvastatin amorphous is 2.0%

and method is found to be precise at concentration 3 times of its limit of detection.(i.e

6,0%).The most stable crystalline known form, was used to prepare standards to

determine its limit of detection, though the method was also designed to detect peaks

from other crystal forms. This methods, and similar methods for other amorphous

materials, have been used for formulation feasibility and stability studies. In the case

of amorphous materials, stability studies are especially important because, besides the

possibility of crystallization, the amorphous materials are also more hygroscopic and

often less chemically stable than related crystalline forms.

This study deals with quantification of crystalline content in amorphous form

using crystallographic (XRD) method. A simple novel XRD method has been

developed and validated for the quantification of crystalline form in amorphous form.

The two drug substances namely Atorvastatin Calcium and Fluvastatin Sodium have

been considered for the study. XRD pattern of pure amorphous, crystalline and

mixture of amorphous and crystalline of different concentration (w/w) has been

collected by using Bruker DS-Advance diffractometer, on in which the Braggs

bretano configuration, a copper X-ray tube (Cu Ka= 1.5406), power of 40kv x 40mA,

scan range 3 to 60°, step size 0.03° and time per step 0.5 sec. Limit of detection for

Atorvastatin Calcium and Fluvastatin Sodium is found to be 3% and 2% respectively.

216

Page 25: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

copter'UII : =

References:

1. Christer B. Aakero'y et al. practical guidelines for co-crystal Synthesis.

Pharmaceutical co-crystals meeting Amsterdam, The Netherlands, 2008.

2. Tros de Ilarduya et al. Drug. Dev.Ind Pharm, 1997, 23, 1095.

3. Aditya Mohan Kaushal et.al.. amorphous drug deliver systems: molecular aspects,

design and performance. Critical reviews in therapeutic drug carrier system, 2004,

21(3), 133-193.

4. Zimmermann B et.al., thermal analysis of paracetamol polymorphs by FT-IR

spectroscopies. J. pharm Biomed Anal, 2011, 54(2), 295-302.

5. S. Gaisford, et al.. Characterisation of paracetamol form III with rapid-heating

DSC, J. Pharm. Biomed. Anal, 2010.

6. Etienne Joiris et.al. Compression behaviour of orthorhombic paracetamol.

Pharmaceutical Research, 1998, 15(7), 1122-1130.

7. Julius F. Remenar et.al.. Crystal engineering of novel cocrystals of a Triazole

drug with 1,4-Dicarboxylic acids. J. Am. Chem. Soc, 2003, 125 (28), 8456-8457.

8. Patent No. U.S. 6.964,978. Solid-State Form of Celecoxib Having Enhanced

Bioavailability. 2005.

9. K. Grzybowska et al. Molecular dynamics and physical stability of amorphous

anti-inflammatory drug: Celecoxib. J. Phys. Chem. B, 2010, 114, 12792-12801.

10. Piyush Gupta et.al., stability and solubility of Celecoxib-PVP amorphous

aispersions: A molecular perspective. Pharm Research 2004,21(10), 1762-1769.

11. Piyush Gupta et al. Molecular interactions in celecoxib-PVP-Meglumine

amorphous system. JPP 2005, 57,1-8.

12. Kirsten A. Graeser et al. physicochemical properties and stability of two

differently prepared amorphous forms of Simvastatin. Crystal growth and design

(2008), 8(1), 128-135.

13. Scott L. Childs etal.A metastable polymorph of Metformin hydrochloride:

isolation and characterization using capillary crystallization and thermal

microscopy techniques. Crystal growth and design, 2004, 4 (3), 441-449.

14. S. Yamamura and Y. Momose, Quantitative analysis of crystalline

pharmaceuticals in powders and tablets by a pattern-fitting procedure using X-ray

diffraction data, Int. J. Pharm, 2001, 212,203-212.

217

Page 26: CHAPTER-VII A novel method to detect and estimate traces ...shodhganga.inflibnet.ac.in/bitstream/10603/88339/15/15...Figure 1: Basic overview of solid forms. From the above figure

CfiapterVlI

15. M. J. Kontny and G. Zografi. Sorption of water by solids. In H. G. Brittain (ed.),

physical characterization of pharmaceutical solids. Marcel Dekker, New York,

1995, 387-418.

16. Vishal koradia, Garimachawla, Arvind k. Bansal, qualitative and quantitative

analysis of clopidogrel bisulphate polymorphs, acta pharm., 2004, 54, 193-204

17. The United States Pharmacopoeia 32, United States Pharmacopoeial Convection,

Rockville, USA.

18. B. A. Sarsfield, M. Davidovich, S. Desikan, M. Fakes, S. Futemik, J. L. Hilden,J.

S.Tan, S. Yin, G. Young, B. Vakkalagadda, and, K. Volk, Bristol Myers Squibb

Co., New Brunswick, NJPurdue University, West Lafayette, powder x-ray

diffraction detection of crystalline phases in amorphous pharmaceuticals.

International Centre for Diffraction Data, 2006, ISSN 1097-0002 , 322-327.

19. Cletus Nunes, ArumugamMahendrasingam, and Raj Suryanarayanan,

Quantification of crystallinity in substantially amorphous materials by synchrotron

x-ray powder diffractometry. Pharmaceutical Research, 2005, 22 (11).

20. R. Surana and R. Suryanarayanan. Quantitation of crystallinity in substantially

amorphous pharmaceuticals and study of crystallization kinetics by X-ray powder

diffractometry. Powder Diffr, 2000, 15, 2-6.

21. C. J. Kedward, W. MacNaughtan, and J. R. Mitchell. Isothermal and non-

isothermal crystallization in amorphous sucrose and lactose at low moisture

contents. Carbohydr. Res, 2000, 329, 423-430.

22. S. L. Shamblin, E. Y. Huang, and G. Zografi. The effects of co-lyophilized

polymeric additives on the glass transition temperature and crystallization of

amorphous sucrose. J. Therm. Anal, 1996,47, 1567-1579.

23. ICH Topic Q 2 (Rl) Validation of analytical procedures text and methodology

note for guidance on validation of analytical procedures, text and methodology

(CPMP/ICH/381/95), June 1995.

218