chapter 20: atomic spectroscopy high temp...

39
high temp (2000-6000K) Chapter 20: Atomic Spectroscopy

Upload: others

Post on 27-Mar-2020

6 views

Category:

Documents


4 download

TRANSCRIPT

Page 1: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

high temp(2000-6000K)

Chapter 20:

Atomic Spectroscopy

Page 2: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

20-1. An Overview

sample

(AES)

(AAS)

(AFS)

Most compounds Atoms in gas phasehigh temp (2000-6000K)

Mass-to-charge(ICP-MS)

Page 3: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Flame in Atomic Spec.Cuvette in Mol. Spec.

(Path length in Flame: 10 cm)

Atomic Absorption experiment

Page 4: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Atomic Fluorescence experiment

Atomic fluorescence from Pb at 405.8 nmAtomization: 6 ns pulse of 1064 nm laser radiation

Page 5: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Bandwidth of radiation that is absorbed or emitted:

Molecular spectroscopy: ~100 nmAtomic spectroscopy: ~0.001 nm (very sharp line spectra little overlap between the spectra of different elements

can measure multi-elements simultaneously; ~70 elements)

Fig.21-3. A portion of emission spectrum

Atomic Emission experiment

Page 6: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Atomizer :converts a sample to an atomic vapor

Atomization: Flames (AAS, AES; liquid sample)Furnaces (AAS, AES; liquid & solid sample)Plasma (AES, ICP-MS; liquid sample)

The precision and accuracy of AS are critically dependent upon atomization step & sample introduction method

20-2. Atomization: Flames, Furnaces, and Plasmas

Page 7: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Premix burner: fuel, oxidant, and sample are premixed.

Nebulization: formation of a small droplets

Aerosol: a fine suspension of liquid (solid)particles in a gas

Nebulizer:create an aerosol from the liquid sample

“aerosol reaching the flame contains only about 5% of initial sample”

Flames

Pneumatic nebulizer

Page 8: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Flames

most common

For refractory elements

Refractory elements: high-boiling elements

Page 9: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Furnaces An electrically heated graphite furnace- offers greater sensitivity than that provided by flames- requires less sample (1-100 L for furnace, 1-2 mL for flames)

Page 10: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Furnaces

Page 11: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Inductively Coupled Plasma

-Exclusively for emission (not absorption) or mass spec. -6000-10000K in ICP (higher and more stable temp than flame)

-Telsa coil ionize Ar gas Free electron is accelerated and oscillated in rf induction coil-Accelerated free electrons collide with Ar atom and transfer energy -> Joule Heating (iR2)

Page 12: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Temp is a critical factor in determining- the degree to which a given sample breaks down to atoms (atomization efficiency)- the extent to which a given atom is found in its ground state, excited state or ionized states

-describes the relative populations of different states at thermal equilibrium-the relative population of any two states is

Boltzman Distribution

g: degeneracy (the number of states available at each energy level)T; temperature, k; Boltzman constant(1.31 X 10-23J/K)

Eo

E*

ΔE

g* = 3,excited states

go = 2,ground states

N*/No = (g*/go) exp(-ΔE/kT)

20-3. How Temperature Affects Atomic Spectroscopy

Page 13: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

N*/No = (g*/go) exp(- ΔE/kT)

Sodium atom at 2600K (in acetylene-air flame)

ΔE = 3.371 x 10-19J/atomgo = 1, g* = 2

Thus, N*/No = (2/1)exp[(-3.371 x 10-19)/(1.381 x 10-23J/K)(2600K)] = 1.67 x 10-4

0.0167% of the atoms are in the excited state (99.9833% in ground state)

If the temperature is 2610K (from 2600K)

N*/No = (2/1)exp[(-3.371 x 10-19)/(1.381 x 10-23J/K)(2610K)] = 1.74 x 10-4

0.0174% of the atoms are in the excited state (99.9826% in ground state)

[(1.74-1.67) x 100%]/1.67 = 4% increase in the population of the excited state[99.9833-99.9826) x 100%]/99.9833 = 0.07% decrease in the population of the ground state

Effect of Temp on Excited-State Population

Page 14: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Sodium atom at 2600K

0.0167% of the atoms are in the excited state (99.9833% in ground state)

If the temperature is 2610K (from 2600K)

0.0174% of the atoms are in the excited state (99.9826% in ground state)

10-K rise in temperature:

Almost no change (0.07% decrease) in the ground-state population:No noticeable change in an atomic absorption

4% increase in the excited-state population: Emission intensity rise by 4%

In atomic emission spectroscopy, the flame temperature should be kept very stable

(ICP: high temp high sensitivity, stable plasma temp reproducible data)

Effect of Temp on Absorption and Emission

Page 15: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

In atomic spec, absorption and emission lines are very sharp

Three sources of line broadening:

(1) Uncertainty effect: ~ 10-4 nm(2) Doppler effect: ~ 10-2 nm – 10-3 nm(3) Pressure effect (collision broadening): ~ 10-2 – 10-3 nm

Uncertainty broadening (natural line width)

Heisenberg uncertainty principle:

the shorter the lifetime of the excited state, the more uncertain in its energy relative to the ground state

δE • δt ≥ h/2π

δE; the uncertainty in the energy difference between the ground and excited statesδt; life time of the excited state before it decays to the ground sate (~ ns in AS)h; Plank’s constant

20-4. Instrumentation

Atomic Linewidths

Page 16: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Doppler broadening

Atoms in hot flame: atomic motions in every direction causes Doppler broadening

Doppler linewidth ; δ = (7 x10-7) (T/M) 1/2

T; temperature, M; atomic mass

A rapidly moving atom in flames

If the motion is toward a photon detector, the wavelength of radiation emitted or absorbed:

decreases (high frequency)

If the atom is receding from the photon detector, the wavelength of radiation emitted or absorbed:

increase (low frequency)

Atomic Linewidths

Page 17: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Doppler broadening

For an emission near = 300 nm from Fe (M = 56) at 2500K

Doppler linewidth ; δ = (7 x10-7) (T/M)1/2 = (300 nm) (7 x10-7) (2500K/56)1/2 = 0.0014 nm

An order of magnitude greater than the natural linewidth

Pressure broadening

Collisions of the absorbing or emitting atoms with other atoms or ions in heating medium

shorten the lifetime of the excited state or

causes the change in the energy level of the ground and excited states.

line broadening

High pressure higher collisional frequency greater line broadening

Atomic Linewidths

Page 18: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Monochromators cannot isolate lines narrower than 10-3 – 10-2 nm. To get narrow lines of the correct frequency, Use of hollow cathode lamp containing the same element that being analyzed

Filled with Ne or Ar at a pressure of 130 ~ 700 Pa

High voltage (~300V) is applied between the anode and cathode Filler gas is ionized and positive ions are accelerated toward the cathodeAccelerated positive ions strike the cathode with enough energy to sputter

metal atoms from the cathode into the gas phase Free atoms are excited by collisions with high-energy electrons: photon emission

Atomic radiation has the same frequency as that absorbed by the analyte atoms

Hollow-Cathode Lamp in AAS

Page 19: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

- The linewidth in hollow cathode lamp is sufficiently narrow

- A different lamp is required for each element- Some lamps are made with more than one element in the cathode

- The purpose of a monochromator in atomic spectroscopy:1) to select one line from hollow cathode lamp (usually filter)2) to reject as much emission from the flame or furnace as possible

Hollow-Cathode Lamp in AAS

Atoms in lamp are cooler than atoms in a flame lamp emission is narrower than the absorption bandwidth in the flame

Page 20: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

1

23 1,2,3:

removed by filter or monochromator

Page 21: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

~ 70 elements can be detected simultaneously

Multielement Detection in ICP-AES

Page 22: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Used in atomic emission spectroscopy

(low dispersion)or grating

Two dispersion elements arranged in series: higher dispersion and resolution

than an echellette of the samesize

Echelle Monochromator

Page 23: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Color Art Insert – Color Plate 23

Polychromator for ICP-Emission with one detector for each element

Page 24: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Color Art Insert – Color Plate 24

Polychromator for ICP-Emission with one detector for all element

Page 25: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the
Page 26: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Background signals arises from absorption, emission, or scattering from matrixas well as flame, plasma or furnace

backgroundsignal(0.3)

analyte singnal(1.0)

Background correction is necessary

Background Correction

Graphite furnace absorption spectrum of bronze dissolved in HNO3

Page 27: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Beam chopping method (or source modulation): eliminate flame emission

Only flame emission (signal b)

Atomic absorption+

Flame emission(signal a)

Signal a – signal b = analyte signal

Source modulation: the power supply of the source isdesigned for ac operation so that the source is switchedon and off at the desired constant frequency

Background Correction

Page 28: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

- The detection limits for furnaces is two orders of magnitude lower (thus, better) than thatobserved with a flame.

- The detection limits for ICP are intermediate between flame and furnace.

Page 29: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Interference is any effect that changes the signal while analyte concentration remains unchanged.

Types of interference:

(1) Spectral interference: overlap of analytical signal with unwanted signals due to other elements in the sample/ or due to flames

20-5. Interference in AS

-Cd line causes spectral interference with the As line in most spectrometers.

-Two peaks are separated with 1-m Czerny-Turner monochromator with a resolution of 0.005 nm from 160 -320 nm and 0.01 nm from 320 -800 nm

Page 30: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

(1) Spectral interference:

Problem solving:Overlap between lines of different elements in the sample: choose another wavelength Or Use of high-resolution spectrometers

Elements that form very stable diatomic oxides are incompletely atomized at the temp of the flame and furnace:much broader and more complexspectrum than that of an atom

Page 31: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Types of interference:

(2) Chemical interference: chemical reactions decreasing the extent of atomization of analytes

Example:

SO42- and PO4

3- hinders the atomization of Ca2+ by forming nonvolatile salts.

Solution:

Add releasing agent (protective agent)

- EDTA, 8-hydroxyquinoline: forms stable but volatile species with Ca2+

- La3+ : preferentially reacts with sulfate or phosphate and frees Ca2+

Page 32: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

(3) Ionization interference:

- ionization of analyte atoms decreases the concentration of neutral atoms- energy levels of ionized atoms are different from those of neutral atoms

<gas phase ionization reaction>M(g) M+(g) + e-(g)

- a problem in the analysis of alkali metals even at relatively low flame temperature(alkali metal: low ionization potential)

at 2450K and 0.1 Pa, Na: 5% ionized, K: 33% ionized

Solution: add an ionization suppressor to decrease the extent of ionization of analyteExample: for the analysis of Sr, add K (more easily ionized)

K K+ + e- --- I.P (K) = 4.3 evSr Sr+ + e- --- I.P (Sr) = 5.7 eV

providing high concentration of electrons to the flame (from K+ + e- ) suppress the ionization of the analyte (Sr)

Page 33: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

1. ICP is twice as hot as a conventional flame 2. The residence time of analyte in the plasma is twice as long

Atomization is more complete Signal is enhanced Formation of oxides and hydroxides are negligible

3. ICP is free of background radiation 4. The temperature in ICP is more uniform almost no self-absorption

Virtues of Inductively Coupled Plasma

Page 34: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the
Page 35: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

20-6. ICP-Mass Spectrometry

Isobaric interference: Interference by ions of similar mass e.g. 40Ar16O+ = 56Fe+, 40Ar2+ = 80Se+

Use of collision cell: dissociate ArO+, Ar2+, ArCl+, -- Or high resolution mass

(negative: attract positive ions)

(water cooling)

Vacuum to avoid collisions between ions and gas molecules that divert the ions from their trajectory in a field

Page 36: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

- Most important techniques for elemental analysis- Mass range: 3-300 (recently up to 4000 m/z)- Ability to resolve ions differing in m/z by 1- 90% elements in periodic table can be determined- Detection limits; 0.1 – 10 ppb- RSD: 2 ~ 4%- Dynamic range: 6 orders- Fast analysis: 10 s for 1 elements

ICP-MS

Page 37: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the
Page 38: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Laser Ablation-ICP-MS

Page 39: Chapter 20: Atomic Spectroscopy high temp (2000-6000K)echem.yonsei.ac.kr/wp-content/uploads/2014/10/강의7... · 2014-10-31 · (AES) (AAS) (AFS) Most compounds ... -describes the

Laser Ablation-ICP-MS