chapter 19 part vi nuclear magnetic resonance dr. nizam m. el-ashgar chemistry department

37
Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department Islamic University of Gaza 01/25/22 1 Chapter 19

Upload: norah

Post on 13-Feb-2016

29 views

Category:

Documents


2 download

DESCRIPTION

Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department Islamic University of Gaza. Instrumentation. The most important parts of an FTNMR instrument are: The magnet, The RF generator, The sample chamber or probe, - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Chapter 19 Part VINuclear Magnetic Resonance

Dr. Nizam M. El-AshgarChemistry Department

Islamic University of Gaza

04/22/231 Chapter 19

Page 2: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Instrumentation

The most important parts of an FTNMR instrument are:•The magnet,•The RF generator, •The sample chamber or probe, (which not only houses the sample but also the RF transmission and detection coils).• In addition, the instrument requires:• A pulse generator•An RF receiver,•Lots of electronics, and a computer for data processing.

Page 3: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department
Page 4: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department
Page 5: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Continuous-Wave NMR

Page 6: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

6

Nuclear Magnetic Resonance Spectroscopy

Page 7: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

A Simplified 60 MHzA Simplified 60 MHzNMR Spectrometer (continuous wave NMR Spectrometer (continuous wave

Instrument)Instrument)

TransmitterReceiver

Probe

h

SN

RF Detector Recorder

RF (60 MHz)Oscillator

~ 1.41 Tesla (+/-) a few ppm(+/-) a few ppm

absorption signal

MAGNETMAGNET

Page 8: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Magnet Legs

Prob

e

Sample Bore

Cryogens

Magnet Coil

NMR Magnet Cross-Section

Page 9: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Fortunately, different types of protons precess at different rates in the same magnetic field.

N

S

CH2 CO

CH3

59.999995 MHz

59.999820 MHz

59.999700 MHz

h60 MHz

To cause absorption of the incoming 60 MHzthe magnetic field strength,Bo , must be increased toa different value for eachtype of proton.

Bo = 1.41 Tesla

Differences are very small, in the parts per million range.

EXAMPLE:

Page 10: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

scan

increasing Bo

HIGHFIELD

LOWFIELD

UPFIELDDOWNFIELD

IN THE CLASSICAL NMR EXPERIMENT THE INSTRUMENTIN THE CLASSICAL NMR EXPERIMENT THE INSTRUMENT SCANS FROM “LOW FIELD” TO “HIGH FIELD”SCANS FROM “LOW FIELD” TO “HIGH FIELD”

NMR CHART

Page 11: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Continuous wave (CW) or field sweep instruments.• Older low field (e.g., 60 MHz) NMR instruments used a fixed RF

and a permanent magnet or electromagnet with a set of Helmholtz coils in the pole faces of the magnet.

• These coils could be adjusted to vary the applied magnet field slightly by passing a current through them, causing each chemically different nucleus to come into resonance sequentially. The sample holder was placed in the magnetic field.

• Two RF coils surrounded the sample so as to be orthogonal to each other and to the applied magnetic field.

• One coil applied a constant RF frequency; • the second coil detected the RF emission from the excited nuclei

as they relaxed. • These systems were simple and rugged, but limited in resolution

and capability. • There are no longer any manufacturers of CW NMR

instrumentation.

Page 12: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

MODERN INSTRUMENTATIONMODERN INSTRUMENTATION

PULSED FOURIER TRANSFORM TECHNOLOGY

requires a computer

FT-NMR

Page 13: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

FT-NMR

Page 14: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

FT-NMR Instrumentation• Main features of a basic NMR include:

– A radio transmitter coil that produces a short powerful pulse of radio waves

– A powerful magnet that produces strong magnetic fields– The sample is placed in a glass tube that spins so the

test material is subject to uniform magnetic field. • Solid samples are dissolved in a solvent that will not give a signal

– A radio receiver coil that detects radio frequencies emitted as nuclei relax to a lower energy level

– A computer that analyses and record the data

Page 15: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

A Modern NMR Instrument

Radio WaveTransceiver

Page 16: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Sample Holder• The sample holder in NMR is normally tube-shaped and is

therefore called the sample tube.• The tube must be transparent to RF radiation, durable, and

chemically inert.• Glass or Pyrex tubes are commonly used. • These are sturdy, practical, and cheap. • They are usually about 6–7 in. long and ~1/8 in. in diameter,

with a plastic cap to contain thes ample.• This type of tube is used for obtaining spectra of bulk samples

and solutions.• Sample tubes range in size from this “standard” size down to

tubes designed to hold 40 L of sample, such as the Nano Probe version from Varian Associates (www.varianinc. com).

• Flow-through cells are used for hyphenated techniques such as HPLC-NMR and on-line analysis.

Page 17: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Sample Probe• The sample chamber into which the sample holder is placed is called

the probe in an NMR spectrometer. • The probe holds the sample fixed in the magnetic field, contains an air

turbine to spin the sample holder while the spectrum is collected and houses the coil(s) for transmitting and detecting NMR signals.

• The probe is the heart of the NMR system. • The most essential component is the RF transmitting and receiving

coil, which is arranged to surround the sample holder and is tuned to the precession frequency of the nucleus to be measured.

• Modern NMR probes use a single wire coil to both excite the sample and detect the signal.

• The coil transmits a strong RF pulse to the sample; the pulse is stopped and the same coil picks up the FID signal from the relaxing nuclei

• For maximum sensitivity, a fixed frequency probe is needed.• This means that a separate probe is required for each nucleus to be

studied: 1H, 13C, 19F, and so on.

Page 18: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

An NMR Probe

Page 19: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

• Probes usually have variable temperature control to run experiments at temperatures selected by the analyst.

• New probe designs with flow-through sample holders are commercially available, for use in coupled HPLC-NMR instruments and HPLC-NMR-MS instruments.

• The probe is installed in the spectrometer magnet so that the coils are centered in the magnet.

• The sample tube is inserted into the top of the probe and is moved by an air column through the magnet bore and centered among the probe coils.

• The tube exits the spectrometer at the top of the probe, again moved by an air column from the bore.

Page 20: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

NMR Sample & Probe Coil

Page 21: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Magnet• The magnet in an NMR spectrometer must be strong, stable,

and produce a homogeneous field. • Homogeneous in this context means that the field does not

vary in strength or direction from point to point over the space occupied by the sample.

• The field must be homogeneous to a few ppb within the sample area.

• It is common to express the magnetic field strength in terms of the equivalent proton frequency from the Larmor equation.

• Commercial magnets range from the now obsolete 60 MHz (1.4 T) to 700 MHz (16.4 T) and higher.

• The magnet is so large that this instrument comes with its own staircase درجso that the analyst can insert and remove the sample tube from the top of the probe.

Page 22: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

• Modern NMR spectrometers use superconducting solenoid لولبي ,magnets ملف

• The magnet consists of a main field coil made of superconducting Nb/Sn or Nb/Ti wire with a dozen or more superconducting shim coils رقيقة wound around the main coil to improve field homogeneity.

• The superconducting coils must be submerged in liquid helium.

• The magnet and liquid helium reservoir are encased in a liquid nitrogen reservoir to decrease the evaporative loss of the more expensive liquid helium.

• The sample probe is mounted in the bore along with a set of room temperature shim coils.

• These coils are adjusted with every new sample placed in the probe to compensate for sample composition, volume, and temperature.

Page 23: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

• Adjusting the field homogeneity, called “shimming”, used to be a time-consuming task.

• Now computers with multivariate optimization NMR Spectroscopy procedures perform this task automatically.

• The cryogenic fluids must be replenished on a regular basis, usually weekly for the liquid nitrogen and every 1–6 months for the liquid helium.

• The superconducting coils must be kept cold; if permitted to warm up they stop functioning.

• The magnet is said to have “quenched”.

Page 24: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

RF Generation and Detection• The RF radiation is generated by using an RF crystal oscillator.• The output of the oscillator is amplified, mixed, and filtered to

produce essentially monochromatic RF radiation for an NMR instrument.

• The RF radiation delivered to the sample is pulsed. • A simple pulse might be a rectangular pulse of 500 MHz

frequency for a 10 s duration. • The process of pulsing actually widens the RF band, providing a

range of frequencies that is able to excite all nuclei whose resonances occur within the band of frequencies.

• All resonances within the band are excited simultaneously. • A 10 s, 500 MHz rectangular pulse leads to a power distribution

around the 500 MHz frequency as shown.• All signals are collected simultaneously. The RF pulse delivered is

generally on the order of watts while the NMR signal collected is on the order of microwatts.

• The FID signal in the time domain must be converted to a frequency domain spectrum by application of a Fourier transformation or other mathematical transformation

Page 25: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Signal Integrator and Computer• All NMR spectrometers are equipped with a signal

integrator to measure the area of peaks.

• The area often appears on the NMR spectrum as a step function superimposed on the spectrum, and also printed out in a data report as well.

• All NMR spectrometers require a computer to process the data and much of the instrument is under computer control.

• Computer control of the room temperature shim currents that control the field homogeneity, of sample spinning rates, autosamplers, pulse sequences, and many other aspects of the instrument operation is a necessity.

Page 26: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

ANALYTICAL APPLICATIONS OF NMR

• NMR spectroscopy is used for both qualitative and quantitative analyses.

Page 27: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Samples and Sample Preparation for NMRLiquid Samples:Simplest samples to analyze by NMR.• Neat nonviscous liquids are run “as is” by placing about 0.5

mL of the liquid in a glass NMR tube.• Liquids can be mixed in a suitable solvent and run as

solutions; the analyte concentration is generally about 2–10%.

• Sensitivity is sufficient to determine concentrations down to about 0.1%

• NMR is not considered a “trace” analytical technique, but that is changing as instrumentation continues to improve Microtubes with as little as 1 mg of sample in 100 mL of solvent are now in use.

• Samples in solution and neat liquids are degassed to remove oxygen and filtered to remove iron particles; both O2 and iron are paramagnetic and cause undesired line broadening.

Page 28: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Solid Samples:• Soluble solid samples are dissolved in a suitable solvent

for analyses. • A typical sample size is 2–3 mg dissolved in 0.5 mL of

solvent.• Some solid polymer samples may be run under “liquid”

conditions, that is, without MAS, by soaking the solid in solvent and allowing it to “swell”.

• This gives enough fluidity to the sample that it behaves as a liquid with respect to the NMR experiment.

• Other solid samples must be run in an instrument equipped with MAS, as has been discussed.

Page 29: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

• Gaz Samples:

• NMR does not have sufficient sensitivity to analyze gas phase samples directly.

• Gases must be concentrated by being absorbed in a suitable solvent, condensed to the liquid phase, or adsorbed onto an appropriate solid phase.

Page 30: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

NMR Solvent

• A suitable solvent for NMR should meet the following requirements:

(1)Chemically inert toward the sample and the sample holder, (2) Have no NMR absorption spectrum itself or a very simple

spectrum, and (3) Easily recovered, by distillation, for example, if the original

sample is required for other testing. • The best solvents for proton NMR contain no protons and

therefore give no proton NMR signals. • Carbon tetrachloride and carbon disulfide fall into this

category.

Page 31: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Deuterated solvents• Replacing protons, 1H, with deuterium, 2H or D, will remove

most of the proton signal for the solvent from the spectrum. • Deuterated chloroform, CDCl3, deuterated water, D2O, and

many other deuterated solvents are commercially available for use.

• They have two disadvantages: They are expensive and they generally contain a small amount of

1H, so some small signal from the solvent may be seen. • A spectrum of the solvent (the blank spectrum) should be run

regularly and whenever a new lot of solvent is used.• • Deuterated chloroform is the solvent used for most of the 13C

spectra shown in the chapter and the small signal from the naturally occurring 13C in the solvent is visible in these spectra.

• An example will be pointed out when discussing these spectra.

Page 32: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

Quantitative AnalysesBoth high and low resolution NMR are used for:• Quantitative analyses of mixtures. • Quality control of both incoming raw materials and finished

products.• Determination of percent purity of pharmaceuticals and chemicals.• Quantitative determination of fat and water in vivo in animals, and

many other applications.

Advantage of NMR • Data can be obtained under experimental conditions where the area

under each resonance is directly proportional to the number of nuclei contributing to the signal.

• One type of organic compound can be determined quantitatively in the presence of a different type, such as

The percentage of alcohols in alkanes, amines in alcohols, aromatics and aliphatics in petroleum, olefins in hydrocarbon mixtures, organic halides and organometallic compounds in other organic compounds, or the number of side chains in a hydrocarbon.

Page 33: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

• NMR can be used to provide determination of chemical purity and quantitative measurements of impurities in materials.

• The accuracy and precision of quantitative NMR measurements are comparable to other instrumental analyses methods.

• Major components can be accurately determined with precisions better than 1% RSD while impurities in materials may be quantified at 0.1% or lower.

• The most common nuclei used for quantitative analyses are 1H, 13C, and 31P.

• Quantitative 31P NMR has been used to determine phospholipids, inorganic phosphorus, and organophosphorus compounds, such as phosphorus-based insecticides.

Page 34: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

HYPHENATED NMR TECHNIQUE

• We have seen that NMR can provide detailed molecular structure information.

• It is possible to join together or “couple” an HPLC instrument with an NMR spectrometer.

• The HPLC performs the separation of a complex mixture and the NMR spectrometer takes a spectrum of each separated component to identify its structure.

• We now have a “new” instrument, an HPLC-NMR instrument.

• We call a coupled instrument like this a “hyphenated” instrument.

Page 35: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

The coupling of two instruments to make a new techniquewith more capabilities than either instrument alone provides results in a hyphenated technique or hybrid technique.

HPLC-NMR is made possible with a specially designed flow probe instead of the standard static probe.

For example, Bruker Instruments (www.brukerbiospin. com) has a flow probe for proton and 13C NMR with a cell volume of 120 mL.

Complex mixtures of unknown alkaloids extracted from plants have been separated and their structures completely characterized by HPLC-NMR using a variety of 2D NMR

Page 36: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

• HPLC-NMR and another hyphenated, more powerful instrument, HPLC-NMR-MS are used in pharmaceutical research and development.

• These hyphenated techniques identify not only the structures of unknowns, but with the addition of MS, the molecular weight of unknown compounds.

• The HPLCNMR-MS instrument separates the sample on the HPLC column, takes the NMR spectra as the separated components flow through the probe and then acquires the mass spectrum of each separated component to determine the molecular weight and additional structural information from the mass spectral fragmentation pattern.

• • The MS must be placed after the NMR, since MS is a

destructive technique.

Page 37: Chapter 19 Part VI Nuclear Magnetic Resonance Dr. Nizam M. El-Ashgar Chemistry Department

LIMITATIONS OF NMR• There are two major limitations to NMR:

(1) it is limited to the measurement of nuclei with magnetic moments and (2) it may be less sensitive than other spectroscopic and chromatographicmethods of analyses.

• As we have seen, although most elements have at least one nucleus that responds in NMR, that nucleus is often of low natural abundance and may

• have a small magnetogyric ratio, reducing sensitivity.

• The proton, 1H, and fluorine, 19F, are the two most sensitive elements.

• Elements in the ionic state do not respond in NMR, but the presence of ions in a sample contributes to unacceptable line broadening.

• Paramagnetic contaminants such as iron and dissolved oxygen also broaden NMR lines.

• Nuclei with quadrupole moments, such as 81Br, broaden the NMR signal. Line broadening in general reduces the NMR signal and hence the ensitivity.