chapter 17. supraglottic airway devices

Upload: adham08

Post on 03-Jun-2018

259 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    1/260

    Chapter 17

    Supraglottic Airway DevicesP.462

    Supraglott ic ai rway devices have become a standard f ixture in airway management,

    f i l l ing a niche between the face mask and tracheal tube in terms of both anatomical

    position and degree of invasiveness. These devices sit outside the trachea but

    provide a handsfree means of achieving a gas-tight airway.

    The f irst s uccessful supraglottic airway d evice, the Laryngeal Mask Airway (LMA)-

    Classic, became available in 1989. As t ime went on, a ddit ional devices were added

    to the LMA family to sat isfy specif ic needs, and a number of other devices were

    developed (1,2). There are a large number of supraglott ic airway devices, some of

    which appear similar to the LMA family and others that work under a different

    concept. I t is not possible to discuss all of these devices, because they are being

    introduced at a rapid rate while others are disappearing. Those that seem to be

    gaining acceptance and longevity at the t ime this writ ing wil l be discussed.

    Laryngeal Mask Airway Family

    T y p e s

    LMA-Classic

    De s c r i p t i o n

    The LMA-Classic (standard LMA, Classic LMA, LMA-C, cLMA) consists of a curved

    tube (shaft) connected to an ell ipt ical spoon-shaped mask (cup) at a 30 angle (Fig.

    17.1). There are two flexible vertical bars where the tube enters the mask to

    prevent the tube from being obstructed by the epiglott is. An inf latable cuff

    surrounds the inner r im of the mask. An inf lat ion tube and self-sealing pilot balloonare attached to the proximal wider end of the mask. A black l ine runs longitudinally

    along the posterior aspect of the tube. At the machine end of the tube is a 15-mm

    connector. The LMA is made from sil icone and contains no latex.

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    2/260

    View Figure

    Figure 17.1LMA-Classic. Note the bars at the junction ofthe tube and mask. (Courtesy of LMA North America.)

    P.463

    TABLE 17.1 Available LMA-Classics

    LMA Size Patient Size

    1 Neonates/infants up to 5 kg

    1.5 Infants between 5 and 10 kg

    2 Infants/children between 10 and 20 kg*

    2.5 Children between 20 and 30 kg

    3 Children 30 to 50 kg

    4 Adults 50 to 70 kg

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    3/260

    5 Adults 70 to 100 kg

    6 Adults over 100 kg

    LMA, Laryngeal Mask Airway.*Size 2.5 may be more suitable for children of this size (16).

    The classic laryngeal mask is available in eight sizes, as shown in Table 17.1. More

    than one size should always be available, because the correct size cannot always

    be predicted. When there is doubt, a larger rather than a smaller size should be

    chosen for the f irst attempt (3).

    Some studies have indicated that the appropriate size for most adult female

    patients is number 4 while the most appropriate size f or adult males is 5

    (4,5 ,6,7,8,9,10,11,12 ,13). Other studies found that both the size four and f ive may

    be suitable for females (11 ,14 ). I t may be more appropriate to use a size 5 for l arge

    adults and a size 4 for normal adults, regardless of gender (12,15 ).

    Al te rnati ve formulas ba sed on we ight have been pro posed (17). A method to choose

    the correct size laryngeal mask for children is to match the widest part of the mask

    to the width of the second to fourth f ingers (18).

    Too small an LMA wil l predispose to gas leaks during posit ive pressure venti lat ion.

    Too large an LMA may tend to come up within the mouth, may interfere with

    procedures in the mouth (5), and may also be associated with a higher incidence of

    sore throat and other symptoms and a greater possibil i ty of pressure on t he l ingual

    nerve.

    I n s e r t i o n

    Standard Technique

    The standard insert ion technique uses a midline or sl i ght ly diagonal approach with

    the cuff ful ly deflated (19 ,20 ,21 ,22). The head should be extended and the neck

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    4/260

    f lexed (snif f ing posit ion) (23). This posit ion is best maintained during insert ion by

    using the noninsert ing hand to stabil ize the occiput (Fig. 17.2). The LMA can be

    inserted without placing the head in this posit ion (24 ). The neutral posit ion may

    cause a small decrease in successful placement compared with the snif f ing posit ion

    (25,26,27). The jaw may be pulled down by an assistant to more fully open the

    mouth.

    The tube port ion is grasped as if i t were a pen, with the index f inger pressing on

    the point where the tube joins the mask (Fig. 17.2). With the aperture facing

    forward (and the black l ine facing the patient 's upper l ip), the t ip of the cuff is

    placed against the inner surface

    P.464

    of the upper incisors or gums. At this point, the tube should be parallel to the f loor

    (23). I f the mouth is being held open, the jaw should be released during further

    insert ion. In the patient with a restricted mouth opening an a lternative method is to

    pass the LMA behind the molar teeth into the pharynx. The tubular part is then

    maneuvered toward the midline (28 ).

    View Figure

    Figure 17.2Initial insertion of the laryngeal mask. Under

    direct vision, the mask tip is pressed upward against thehard palate. The middle finger may be used to push thelower jaw downward. The mask is pressed upward as it is

    advanced into the pharynx to ensure that the tip remainsflattened and avoids the tongue. The jaw should not be held

    open once the mask is inside the mouth. The nonintubatinghand can be used to stabilize the occiput. (Courtesy of

    Gensia Pharmaceuticals, Inc.)

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    5/260

    View Figure

    Figure 17.3By withdrawing the other fingers and with aslight pronation of the forearm, it is usually possible to pushthe mask fully into position in one fluid movement. Notethat the neck is kept flexed and the head extended.(Courtesy of Gensia Pharmaceuticals, Inc.)

    As th e LMA is ad vanced , the mask po rt io n is pres sed agains t th e hard pa la te by

    using the index f inger. This means that the direct ion of applied pressure is dif ferent

    from the direct ion in which the mask moves (29). I f resistance is felt , the t ip may

    have folded on itself or impacted on an irregularity or the posterior pharynx. In this

    case, a diagonal shif t in direct ion is often helpful, or a gloved f inger may be

    inserted behind the mask to l i f t i t over the obstruct ion (30 ). I f at any t ime during

    insert ion the mask fails to stay f lattened or starts to fold back, it should be

    withdrawn and reinserted.

    A chang e of direc t io n can be sensed as the mas k t ip en coun te rs th e posterio r

    pharyngeal wall and follows it downward. By withdrawing the other f ingers as the

    index f inger is advanced and slight ly pronating the forearm, it is often possible to

    insert the mask fully i nto posit ion with a single movement (Fig. 17.3). I f this

    maneuver is not successful, hand posit ion must be c hanged for the next movement.

    The tube is grasped with the other hand, straightened slight ly, and then p ressed

    down with a single quick but gentle movement unti l a definite resistance is felt (Fig.

    17.4). This may co incide with anterior laryngeal displacement (30 ). The longitudinal

    black l ine on the shaft should l ie in the midline facing the upper l ip. Any deviat ion

    may indicate that the cuff is misplaced.

    If the patient has a high, arched palate, a sl ight ly lateral approach may be needed.

    The operator should check that the cuff t ip is correct ly f lattened against the palate

    before proceeding (30 ). Dif f iculty encountered in negotiat ing the angle at the back

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    6/260

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    7/260

    View Figure

    Figure 17.4The laryngeal mask is grasped with the otherhand and the index finger withdrawn. The hand that isholding the tube presses gently downward until resistance isencountered. (Courtesy of Gensia Pharmaceuticals, Inc.)

    180-degree Technique

    Another te chn iqu e is to ins ert the LMA wi th th e laryn gea l ape rtu re po int in g

    cephalad and rotate it 180 degrees as i t enters the hypopharynx (45,81,82 ,83 ). A

    dist inct pop may be felt by the introducing hand (81 ). This method may be as

    satisfactory as the standard technique, especially in pediatric patients. I t has been

    postulated that rotat ion of the bulky LMA cuff in the close proximity of the

    hypopharynx could dislocate the arytenoid cart i lages (84 ).

    Partial Inflation Technique

    Yet another technique is to part ial ly or ful ly inf late the cuff before insert ion

    (85,86,87,88 ,89 ,90). Although this technique may offer some advantages for an

    inexperienced user, the device may frequently be malposit ioned (73 ,91). The

    incidence of sore throat may be reduced with the partial inflation method ( 89 ).

    Thumb Insertion Technique

    The thumb insert ion technique is more suitable for patients where access to the

    head from behind is dif f icult o r impossible. Insert ion is similar to the standard

    technique except that the LMA

    P.466

    is held with the thumb in the posit ion occupied by the index f inger in that technique.

    As th e th umb nea rs the mouth, the f in ge rs are stre tc hed forward ov er th e pati en t' s

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    8/260

    face. The thumb is advanced to i ts ful lest extent. Before removing the thumb, the

    tube is pushed into its f inal posit ion by using the other hand.

    View Figure

    Figure 17.5The laryngeal mask in place. (Courtesy ofGensia Pharmaceuticals, Inc.)

    View Figure

    Figure 17.6Laryngeal mask airway in place. The tip of themask rests against the upper esophageal sphincter while thesides face the puriform fossae.

    T r ac h e al In t u b a t i o n w i t h t h e L MA -C l as s i c

    The LMA can serve as a conduit through which a tracheal tube, stylet, or f iberscope

    is passed (92,93,94,95,96,97 ,98,99 ,100). I t acts to posit ion the device over the

    laryngeal aperture. The 30 angle between the tube and cuff was chosen because it

    was found to be optimal for intubation through the LMA (10 1).

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    9/260

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    10/260

    3 6.0 cuffed 5.0

    4 6.0 cuffed 5.0

    5 7.0 cuffed 5.0

    6 7.0 cuffed 5.0

    LMA, Laryngeal Mask Airway; ID, internal diameter; OD, outer diameter.aIf the connector is removed, a larger tracheal tube can be inserted (104).

    bA larger fiberscope may be accommodated if the aperture bars are removed (105).

    Af ter the LMA is ins erted and f ixed into pos ition , the tra chea l tube is we ll lubricated

    and inserted into the tube. Auscultat ing the end of the tube may be useful during

    spontaneous breathing. The tracheal tube should be rotated 15 to 90

    counterclockwise as i t is advanced to prevent the bevel f rom catching on the bars

    at the junction of the tube and mask. A tracheal tube with a midline bevel point may

    make this maneuver unnecessary. Once through the bars, the tracheal tube is

    rotated clockwise, and the neck is extended to enable the t ip to pass anterior to the

    arytenoids. The tracheal tube is then advanced unti l resistance is felt . The head is

    then f lexed, permitt ing the tube to advance into the trachea.

    An al ternate me thod of t rac he al intubatio n inc lu des load ing the lu bri cated tracheal

    tube with the t ip placed at the level of the bars in the LMA so that when pushed

    down, it will pass smoothly through the middle aperture ( 114).

    I f the trachea is not entered init ial ly, i t is p ossible that the LMA is not well situatedover the laryngeal aperture or the epiglott is is blocking the aperture. Varying

    degrees of neck f lexion and extension at the at lanto-occipital joint may be helpful

    (111 ). I f the tube st i l l does not enter the trachea, it should be withdrawn unti l the

    bevel is just behind the aperture bars. The LMA cuff should be

    P.467

    deflated and the LMA pushed a litt le farther into the hypopharynx. This maneuver

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    11/260

    may cause elevation of the downfolded epiglott is. The tracheal tube is then pushed

    downward.

    Smaller tracheal tubes are easier to place than larger ones ( 115 ). I f the tube is not

    large enough, it may be replaced by a larger tube by using a tube exchanger

    (Chapter 19).

    The LMA can be used to guide a stylet, bougie, or exchange catheter into the

    trachea (42,95 ,99 ,116,11 7,118 ,119 ,120 ,121,122 ,123 ,124,125 ). It may be easier to

    pass a stylet i f i ts angulated end is made to point anteriorly unti l i t passes through

    the grill of the mask and then rotated 180 ( 126 ). Monitoring carbon dioxide through

    a channel in the exchange catheter is a useful way of determining that the t ip is in

    the trachea (127). After catheter placement is confirmed, the LMA can be withdrawn

    and the tracheal tube inserted over the catheter.

    F ib e r s c o p e G u i d e d

    The LMA-Classic can be used to aid f iberoptic-guided intubation. The LMA is

    inserted in the usual manner. The fiberscope, with a well-lubricated tracheal tube

    and a fully deflated cuff threaded over i ts shaft, is advanced through the LMA. The

    fiberscope is advanced into the trachea, and the tracheal tube is advanced over the

    fiberscope into the trachea. The tube should be rotated as it is advanced. I t may be

    useful to posit ion the tracheal tube in the LMA, just proximal to the aperture bars,

    before inserting the fiberscope. A right angle bronchoscopic tracheal tube

    connector with a seal is attached to the tracheal tube. This allows the f iberoptic

    scope to be i nserted through the tracheal tube while the patient is being venti lated.

    With this, posit ive end-expiratory pressure (PEEP) can be applied, which often

    dramatically improves the view by stenting the collapsed upper airway (128)

    Fiberscope size is important, especially in small pat ients. The maximum diameters

    of scopes for various size tracheal tubes are given in Table 17.2.

    The laryngeal mask has proved equal or superior to other devices designed to aid

    with intubation using a f iberscope (129 ,130 ). This i ntubation method has been used

    in infants as small as 3.6 kg (13 1). This method has a higher success rate than the

    blind method and is associated with less risk of trauma or esophageal intubation

    (96,132 ,133,134). An inspection of the tracheobronchial tree can be made. I t is

    useful for patients in whom neck movement needs to be avoided (135).

    A f ibers cop e ma y be us ed to insert a bo ugie or other gu ide into the tra che a th ro ugh

    an LMA

    (97,119 ,125,126,127 ,128 ,129,130,131 ,132,133 ,134,135 ,136,13 7,138,13 9,14 0,14 1,1

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    12/260

    42,143,144,145,14 6,147 ,148). The LMA can then b e removed and a tracheal tube

    inserted over the guide. This technique allows a l arger tracheal tube to be placed

    and avoids the danger of accidental extubation when the LMA is removed (97 ,143).

    Re t r o g r a d e

    The LMA can be used to facil i tate tracheal intubation by using a retrograde wire

    technique (149,150 ,15 1) (Chapter 21). The guide wire is inserted through the

    cricothyroid membrane and passed cephalad. A guide catheter is then threaded

    antegrade over the wire. The LMA is then removed and a tracheal tube passed over

    the catheter.

    L i g h t e d S t y l e t G u i d e d

    Tracheal intubation by using a l ighted stylet placed through a tracheal tube that is

    inserted through an LMA-Classic has been described (152,15 3,15 4,155). The stylet

    t ip is posit ioned at the pat ient end of the tracheal tube and is advanced through the

    LMA. I f a central point of l ight in the anterior neck is observed, the LMA cuff is

    suitably placed around the laryngeal inlet. I f transil lumination is not seen, the LMA

    is reposit ioned in the pharynx according to the location of the l ight. Tracheal

    intubation is accomplished by advancing the l ighted stylet unti l the suprasternal

    notch is transil luminated. The success of intubation with this method is comparable

    to blind intubation (15 5,156).

    Na s o t r a c h e a l In t u b a t i o n

    The LMA has been used to facil i tate nasotracheal intubation (15 7). The LMA is

    inserted and a catheter placed in the trachea. Another catheter is inserted into the

    nose and brought out through the mouth. The LMA is removed with the catheters in

    place. The catheters are s utured together and tract ion applied to the nasal catheter

    so that the cu rve in the mouth is removed. The tracheal tube is then inserted over

    the catheters into the larynx.

    Another metho d to pe rf orm na sotracheal in tu bation is to cu t a window in the

    posterior aspect of the LMA's tube near the mask and remove the aperture bars

    (158 ). A f iberscope mounted with a tracheal tube is inserted via the nose through

    the window in the LMA and into the trachea.

    Another metho d is to in sert the LMA and the n part ial ly wi thdra w the cuff into the

    oropharynx, where it can s upply fresh gas to the spontaneously breathing patient

    (159 ). Nasotracheal intubation is then accomplished by using a f iberscope.

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    13/260

    Problems Associated with Tracheal Intubation through the

    LMA-Classic

    Some standard tracheal tubes may not be long enough to insert through the LMA-

    Classic (160). Deeper placement can be achieved by using a longer tracheal tube

    (134 ,161,162,163 ,164,16 5,166 ,167 ,168,169 ,170 ,171 ,172,173 ,174 ), shortening the

    LMA tube (139 ,175,176), removing the connector from the LMA, deflating the LMA

    cuff (160 ) using a split LMA (92 ,102), or using a device such as a stylet to advance

    the tracheal tube farther (108). Consideration should be given to using an airway

    exchange catheter (17 7).

    As th e trac hea l tube is pas sed th rough th e LMA -C las sic , the p ilot tube ma y be come

    kinked (17 8).

    P.468

    When intubating a pediatric patient with an uncuffed tracheal tube through an LMA-

    Classic, the largest tube that wil l pass through the LMA may be too small to allow a

    good seal during posit ive pressure venti lat ion. Leaving the LMA in place with the

    cuff inf lated wil l reduce the leak (179 ). I f a cuffed tube is used, it may b e

    impossible to remove the LMA without damaging the inflating device ( 18 0).

    Cricoid pressure may make it more dif f icult to pass a tracheal tube

    (38,181 ,182,183). I f intubation is ini t ial ly unsuccessful and intubation is deemed

    vital, a second attempt should be made with transient release of cric oid pressure.

    Changing from one-handed to bimanual cricoid pressure application may be helpful.

    LMA-Classic Removal after Tracheal Intubation

    The decision to remove the LMA-Classic after tracheal intubation or to leave it in

    place depends on the circumstances. Reasons to remove the LMA include concern

    about pressure on the soft t issues, the need to keep it away from the surgical f ield,

    a possible increase in gastroesophageal ref lux, and dif f iculty in placing a gastric

    tube (184).

    Many users prefer to leave the LMA-Classic in place after intubation to provide an

    alternative airway at the conclusion of the anesthetic (185 ,186,187 ). I f this is done,

    the protruding end of the tracheal tube should be f irmly secured.

    There is no simple way to remove the LMA without disturbing the trachea tube. An

    LMA may be modif ied by split t ing the tube and cuff so that it can be more easily

    removed over the tracheal tube (10 2,18 8). A small tracheal tube or other device

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    14/260

    may be placed in the end of the tracheal tube to act as a pusher (extender) to

    prevent extubation as the LMA is withdrawn (94,18 9,190,191,19 2). A f iberscope,

    bougie, jet stylet, or tube changer may be passed through the tracheal tube to

    facil i tate reintubation if extubation occurs as the LMA is being removed (19 3).

    LMA-Unique

    The single-use LMA-Unique (disposable l aryngeal mask ai rway, DLMA) is shown in

    Figure 17.7. Sizes are given in Table 17.3. I t is made of polyvinylchloride and costs

    less than a reusable LMA. While the dimensions are identical to the standard LMA,

    the tube is st if fer and the cuff less compliant. I t may be helpful to warm it prior to

    insert ion to make it sof ter and more compliant.

    Indicat ions are the same as for the LMA-Classic. I t may be a better choice for out-of-hospital or ward use, where it would be dif f icult to clean and s teri l ize a reusable

    LMA after use. Insertion and placement of the LMA-Unique is similar to the LMA-

    Classic. I f i t is used for f iberoptic intubation, shortening the tube may be helpful

    (176 ).

    Comparisons of the LMA-Unique with the LMA-Classic show litt le difference in ease

    of insertion or performance (194 ,195 ,196 ), although the LMA-Unique may be

    somewhat more dif f icult to insert (197 ). The intracuff pressure increases

    signif icantly less in the LMA-Unique when nitrous oxide is used (198,199 ).

    View Figure

    Figure 17.7LMA-Unique.

    One study found that a 3 -mm or 4-mm internal diameter (ID) tracheal tube could not

    be inserted through the LMA-Unique (200). A structural problem result ing in a leak

    has been reported (201 ).

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    15/260

    LMA-Flexible

    De s c r i p t i o n

    The LMA-Flexible (wire-reinforced, reinforced LMA, RLMA, FLMA, f lexible LMA)

    (Fig. 17.8) dif fers from the LMA-Classic in that it has a f lexible, wire-reinforced

    tube (202,203,20 4). This tube is longer and narrower than the tube on the LMA-

    Classic. I t is available in the s izes shown in Table 17.4. The cuff sizes are the

    same as for the LMA-Classic. A single-use v ersion is also available. The sizes for

    the single-use version are the same as for the mult iuse one.

    TABLE 17.3 Sizes of the LMA-Unique

    MaskSize

    Patient Size(kg)

    Maximum CuffVolume (mL of air )

    Largest TrachealTube (ID in mm)a

    Largest F lexibleEndoscope (ID in mm)b

    1 Up to5

    4 3.5 2.7

    1.5 5 to 10 7 4.0 3.0

    2 10 to

    20

    10 4.5 3.5

    2.5 20 to

    30

    14 5.0 4.0

    3 30 to

    50

    20 6.0 5.0

    4 50 to70

    30 6.0 5.0

    5 70 to100

    40 7.0 5.5

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    16/260

    ID, internal diameter.aIf the connector is removed, a larger tracheal tube can be inserted (104).bA larger fiberscope may be accommodated if the aperture bars are removed (105).

    P.469

    View Figure

    Figure 17.8LMA-Flexible. The wire-reinforced tube islonger and has a smaller diameter than the standard LMA.

    The f lexible tube can be bent to any angle without kinking. This allows it to be

    posit ioned away from the surgical f ield without occluding the lumen or los ing the

    seal against the larynx. I t is less l ikely to be displaced during head rotat ion or tube

    reposit ioning than the LMA-Classic.

    I n s e r t i o n

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    17/260

    The LMA-Flexible is more dif f icult to i nsert than the LMA-Classic. A stylet, small

    tracheal tube, or other device may be i nserted into the tube to st if fen i t

    (205 ,206,207,208 ,209,21 0,211 ,212 ,213,214 ,215 ). The manufacturer recommends

    that it be held between the thumb and index finger at the junction of the tube and

    cuff and posit ioned by insert ing the index f inger to i ts ful lest extent into the oral

    cavity until resistance is encountered. It may be necessary to use the other hand to

    achieve full insert ion. I t may be easier to insert b y using the thumb, index, and

    middle f ingers at the junction of the tube and bowl, then using the index and middle

    f inger to advance it into the hypopharynx (21 6). A modif ied Magil l forceps or o ther

    device may be useful (59 ,21 2,217 ). Other methods for insert ion have been

    described (21 8). Some techniques such as the 180-degree technique may not work

    with this LMA (202 ). After insertion, the tube may be brought out through the nose

    (219 ).

    Use

    The LMA-Flexible is designed for use with surgery on the head, neck, and upper

    torso where the LMA-Classic would be in the way. A throat pack should be used if

    there is a r isk of dental fragments becoming wedged behind the cuff (202). I f

    malocclusion test ing is needed, the tubing can be coiled inside the mouth (220).

    Comparisons between the LMA-Classic and LMA-Flexible reveal that both are

    similar in terms of mask posit ion, cl inical performance, and pharyngeal mucosal

    pressures (22 1,222 ).

    P r o b l em s

    The wire reinforcement makes the LMA-Flexible more resistant to kinking and

    compression than the LMA-Classic but does not prevent obstruct ion from bit ing.

    Ai rway obs tructi on an d los s of sea l hav e been re po rte d wh en a Boyle Davis ga g

    was used (223 ,224 ,225 ). This can usually be corrected by reposit ioning the gag.

    The spiral reinforcing wire in the LMA-Flexible may become disrupted. Sometimes,

    the disruption is internal and can only be discovered by looking carefully down the

    shaft (22 6). Defects in the wire may c ause obstruct ion if the tube is bent, or pieces

    of wire could break off and migrate into the tracheobronchial tree (22 6,227).

    The small diameter of the tube l imits the size endoscope or tracheal tube that can

    be passed through the LMA-Flexible (10 5). It has been recommended that

    prolonged spontaneous venti lat ion be avoided because the smaller tube causes

    increased resistance (202,228 ).

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    18/260

    The LMA-Flexible is unsuitable for magnetic resonance imaging (MRI) scanning if

    image quality in the region of the LMA is important (229). The metall ic r ings wil l

    cause image distort ion (22 9).

    Malposit ion is less easily diagnosed with the LMA-Flexible than with the LMA-

    Classic because the tube gives no c lear indicat ion of cuff orientat ion (23 0).

    TABLE 17.4 Size Comparison between Standard and Flexible Laryngeal Mask Airways

    LMA

    Size

    Patient Size

    (kg)

    LMA-Flexible

    (ID in mm)

    LMA-Classic

    (I D in mm)

    LMA-Flexible

    Tube Length

    (cm)

    Maximum Cuff

    I nfl ation Volume

    (mL)

    2 10 to

    20

    5.1 7.0 21.5 Up to 10

    2.5 20 to30

    6.1 8.4 23.0 Up to 14

    3 30 to50

    7.6 10.0 25.5 Up to 20

    4 50 to70

    7.6 10.0 25.5 Up to 30

    5 70 to100

    8.7 11.5 28.5 Up to 40

    6 >100 8.7 28.5 23.5 Up to 50

    LMA, Laryngeal Mask Airway; ID, internal diameter.

    P.470

    LMA-Fastrach

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    19/260

    The LMA-Fastrach (intubating LMA, ILMA, ILM, intubating laryngeal mask airway)

    was designed to overcome some of the l imitat ions of the LMA-Classic during

    tracheal intubation (231,232,233,234 ,235 ,23 6,237,23 8). The LMA-Classic was too

    floppy to optimize alignment with the glott is, and the long narrow tube could not

    accommodate a standard tracheal tube. Another objective was to eliminate the

    need to distort the anterior pharyngeal anatomy in order to v isualize the laryngeal

    inlet, making the device applicable to patients with a history of dif f icult intubation

    and a high or anterior larynx (234 ).

    De s c r i p t i o n

    The LMA-Fastrach (Fig. 17.9) has a short, curved stainless steel shaft with a

    standard 15-mm connector. The tube is of sufficient diameter that a cuffed 9-mmtracheal tube can be inserted and short enough to allow a standard tracheal tube

    cuff to pass beyond the vocal cords. The metal handle is securely bonded to the

    shaft near the connector end to facil i tate one-handed insert ion, posit ion

    adjustment, and maintain the device in a steady posit ion during tracheal tube

    insert ion and removal. There is a si ngle, movable epiglott ic elevator bar in place of

    the two vert ical bars (Fig. 17.10A). A V-shaped guiding ramp is built into the f loor

    of the mask aperture to di rect the tracheal tube toward the glott is. The t ip is sl ight ly

    curved to permit atraumatic insert ion. Figure 17.10Bshows tracheal tube protruding

    through the LMA tube and into the bowl.

    The LMA-Fastrach does not contain latex. I t is available in sizes 3, 4, and 5. These

    fit the same s ize patients as the LMA-Classic. Both reusable and disposable

    versions are available.

    I n s e r t i o n

    The LMA-Fastrach was designed for use with the patient in the neutral posit ion

    (231 ). This i ncludes using a head support, such as a pi l low, but no head extension.

    The insert ion technique consists of one-hand movements in the sagit tal plane. I t

    does not require placing f ingers into the patient 's mouth, thus minimizing the risk of

    injury or infect ion transmission as well as allowing insert ion from almost any

    posit ion (234,239 ).

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    20/260

    View Figure

    Figure 17.9LMA-Fastrach with tracheal tube. The tube onthe LMA is shorter and wider than on the LMA-Classic andhas a metal handle. Note that the tracheal tube connector has

    been removed. (Courtesy of LMA North America.)

    The LMA-Fastrach should be deflated and lubricated in a manner similar to the

    LMA-Classic. I t is held by the handle, which should be approximately parallel to the

    patient 's chest. The mask t ip is posit ioned f lat against the hard palate immediately

    posterior to the upper

    P.471

    incisors, then slid back a nd forth over the palate to distr ibute the lubricant. After

    the mask is f lattened against the hard palate, i t is inserted with a rotat ional

    movement along the hard palate and the posterior pharyngeal wall. The mouth

    opening may need to be increased momentari ly to permit the widest part of the

    mask to enter the oral cavity. The handle should not be used as a lever to force the

    mouth open. As the mask moves toward the pharynx, it should be firmly pressed to

    the soft palate and posterior pharyngeal wall to keep the tip from folding. The

    curved part of the metal tube should be advanced without rotat ion unti l i t contacts

    the patient 's chin, then kept in contact with the chin as the device is rotated inward.

    The handle should not be used to lever upward during insert ion, because this wil l

    cause the mask to press into the tongue.

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    21/260

    View Figure

    Figure 17.10LMA-Fastrach. A:Note the single, movableepiglottic elevator bar and the V-shaped guiding ramp builtinto the floor of the mask aperture to direct the tracheal tubetoward the glottis. B:Tracheal tube emerging from theLMA.

    When properly inserted, the tube should emerge from the mouth directed somewhat

    caudally. Aligning the internal LMA-Fastrach aperture and the glott ic opening by

    f inding the posit ion that produces optimal venti lat ion and then applying a slight

    anterior l i f t with the LMA-Fastrach handle facil i tates correct posit ioning and blind

    intubation (24 0).

    The LMA-Fastrach can be inserted with a 180 rotation technique (24 1).

    Use

    Al though the LMA-Fas tra ch ha s been designed to fac i l i tate tra che al intu ba tio n, i t

    can also be used as a primary airway device. I t is especially useful for the

    anticipated or unexpected dif f icult airway

    (231 ,240,242,243 ,244,24 5,246 ,247 ,248,249 ,250 ,251 ,252,253 ,254 ,255,256,25 7,25 8,

    259,260 ,261 ,26 2,263,26 4,26 5,26 6,267,26 8,269).

    Studies indicate that most insert ion attempts with the LMA-Fastrach are successful,

    and a patent airway is secured in nearly all pat ients

    (231 ,237,240,245 ,247,25 0,253 ,258 ,265,270 ,271 ,272 ,273,274 ,275 ). It has been

    used successfully in children (276), morbidly obese patients (277,278 ,279), and

    acromegalic patients (280).

    The LMA-Fastrach can be inserted with the same or better success than the LMA-

    Classic (28 1,282,28 3,284 ). It is easier to place than the LMA-Classic when manual

    in-l ine stabil izat ion is used (285 ). However, in patients with limited neck movement,

    intubation may be less l ikely to be successful and take longer than if a l ighted

    intubation stylet is used (286 ).

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    22/260

    The LMA-Fastrach has been used successfully in the emergency department and

    prehospital care (265,282,287 ,288,289 ,290 ). It can be used with the patient in the

    lateral posit ion (291 ,292,293 ,29 4).

    T r a c h e a l In t u b a t i o n

    Muscle relaxants are not necessary for intubation through the LMA-Fastrach but

    may increase the success rate (295 ,296 ,297,29 8,299 ). Cricoid pressure wil l

    decrease the likelihood of success and may need to be released to allow intubation

    (182 ).

    The tracheal tube recommended by the manufacturer for use with the LMA-Fastrach

    is a si l icone, wire-reinforced, cuffed tube with a tapered patient end and a blunt t ip

    (236 ,300,301) (Chapter 19). This tube is f lexible, which allows negotiat ion aroundthe anatomical curves of the airway. I t has a high-pressure, low-volume cuff that

    reduces resistance during intubation and makes cuff perforation as the tube passes

    through the LMA less l ikely. There is a stabil i izer that allows the LMA to be

    removed without extubating the patient.

    When using the LMA-Fastrach, standard curved plast ic tracheal tubes are

    associated with a greater likelihood of laryngeal trauma (302 ,303). Warming a

    plast ic tube wil l result in success and complicat ion rates similar to that of the tube

    from the LMA-Fastrach manufacturer (301 ). A s piral-embedded tube should not be

    used. I f a curved plast ic tracheal tube is used, it may be helpful to orient the curve

    opposite the LMA curve (265,272,303).

    Whatever tracheal tube is used, i t is essential that it is possible to remove the

    connector (304). I t is important to lubricate the tracheal tube well and pass it

    through the LMA several t imes before use (305).

    Blind Intubation

    The patient's head is maintained in the neutral position. The tracheal tube

    connector should be loosely f i t ted for easy removal. The tracheal tube should be

    lubricated with a water-soluble lubricant and passed into the metal shaft of the

    LMA-Fastrach unti l the tube t ip is about to e nter the mask aperture. With the

    sil icone tracheal tube specially designed for the LMA-Fastrach, the longitudinal l ine

    should face the handle of the LMA, and the tracheal tube should not be passed

    beyond the point where the transverse l ine on the tube is level with the outer r im of

    the LMA-Fastrach airway tube.

    The LMA-Fastrach handle is grasped with one hand to steady it while the tracheal

    tube is being inserted, then l i f ted l ike a laryngoscope (not levered) to draw the

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    23/260

    larynx forward a few millimeters. This increases the seal pressure and helps to

    align the axes of the trachea and the tracheal tube. It also corrects the tendency for

    the mask to f lex.

    As i t is adv an ced into the LMA-Fas trach, the tube sho uld be rota ted an d mov ed up

    and down to distr ibute the lubricant. Venti lat ion and carbon dioxide monitoring can

    be performed during tracheal tube insert ion by connecting the tracheal tube to the

    anesthesia breathing system.

    The tracheal tube should be advanced gently. The LMA-Fastrach handle should not

    be pressed downward. I f no resistance if felt , i t is l ikely that the epiglott ic elevating

    bar is l i f t ing the epiglott is upward, allowing the tracheal tube to pass into the

    trachea. When the tracheal tube is thought to be in the trachea, the cuff should be

    inf lated and its posit ion in the trachea confirmed (Chapter 19).

    I f the tracheal tube fails to enter the trachea, a number of problems may have

    contributed to the lack of

    P.472

    success (234,270 ). The epiglottis may have folded downward, or the tube may have

    impacted on the periglott ic structures. The LMA-Fastrach may be too small or too

    large for the patient. The l arynx may have been pushed downward during insert ion.

    There may have been inadequate anesthesia or muscle relaxation so that the vocal

    cords were closed.

    During blind tracheal intubation, the operator relies on tact i le perceptions,

    especially a feeling of resistance, while advancing the tracheal tube (23 4). I f the

    mask is not aligned with the glott ic opening or the size of the LMA-Fastrach is

    inappropriate, resistance will be encountered as the tracheal tube tip pushes

    against glott ic or periglott ic structures, such as the downfolded epiglott is,

    valleculae, arytenoids, or aryepiglott ic folds.

    If resistance is felt after the tracheal tube leaves the LMA-Fastrach, there are a

    number of maneuvers that can be taken to relieve the situat ion. I f resistance is felt

    at 2 cm beyond the 15-cm mark on the tracheal tube, it is l ikely that the tube has

    impacted on the vestibular wall. Rotating the tracheal bevel may overcome the

    impaction. Another problem at this level may be a downfolded epiglott is. Without

    deflat ing the cuff, the device should be s wung outward for 6 cm and reinserted

    (234 ). I f resistance is e ncountered 3 cm beyond 15 cm, the epiglott is may be out of

    the reach of the elevating bar, and a larger LMA should be used. If resistance

    occurs immediately after the tracheal tube leaves the LMA-Fastrach, the LMA may

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    24/260

    be too large and should be replaced with a smaller one (270). I f resistance is felt at

    15 plus 4 cm, the LMA may be too large, and a smaller size should be used (239 ).

    Other maneuvers that can be tr ied include slight ly rotat ing the LMA-Fastrach in the

    sagit tal plane by using the metal handle unti l the least resistance to manual

    venti lat ion is achieved, removing the head support, pull ing the metal handle toward

    the user (extension), or pushing it away from the user (flexion) ( 237 ,240 ,258 ,265).

    The LMA-Fastrach can be used to place an airway exchange catheter, which can

    then be used to di rect a tracheal tube into the trachea (30 6).

    Blind intubation using the LMA-Fastrach is faster than f iberoptic-guided intubation

    or intubation using direct laryngoscopy (27 2,307 ). It can be performed awake (248 ).

    When compared with awake f iberoptic intubation for patients with k nown dif f icult

    airways, patient sat isfact ion was greater with the LMA-Fastrach (250 ).

    The success rate of b lind intubation varies from 40% to 100%

    (231 ,235,237,240 ,243,24 5,248 ,256 ,258,265 ,270 ,271 ,272,275 ,276 ,280,300,30 7,30 8,

    309,310 ,311 ,31 2,313,31 4,31 5,31 6,317,31 8), depending on the number of attempts

    and the experience of the operator.

    The blind technique can be time consuming and may result in trauma or esophageal

    intubation (23 7,245,27 1,30 9,31 0,31 2,31 9,320,321,32 2,32 3). I f dif f iculty is

    encountered, the use of an LMA-CTrach should be considered.

    Blind Nasal Intubation

    Blind nasal in tubation can be accomplished (324 ). A f lexible tracheal tube is

    inserted into the trachea through the LMA, which is then removed. A Foley catheter

    is introduced into the nose and withdrawn from the mouth. The Foley catheter is

    inserted into the end of the tracheal tube and inflated with saline to grip the inner

    walls. The Foley catheter is then withdrawn unti l the machine end of the tracheal

    tube exits through the nose.

    Fiberscopic-guided Intubation

    The LMA-Fastrach is useful for f iberoptic intubation in the dif f icult-to-intubate

    patient (26 3). The f iberscope is used to observe correct tracheal tube passage

    through the LMA. The epiglott ic elevating bar is too s t if f to be elev ated by a

    f iberscope without r isk of damaging the t ip or direct ing it downward (239 ,325,32 6).

    I t should be l i f ted by the distal end of the tracheal tube (32 7).

    The tracheal tube is advanced approximately 1.5 cm past the mask aperture while

    the intubating metal handle of the LMA-Fastrach is held to stabil ize it . The t ip of the

    tracheal tube should now have lifted the fiberscope away from the bowl of the

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    25/260

    mask, exposing the glott ic structures. The f iberoptic scope is inserted and

    advanced to, but not beyond, the distal end of the tracheal tube. The tracheal tube

    is advanced unti l the glott is is brought into view. The tracheal tube is then

    advanced into the trachea.

    If dif f icult ies are encountered, the patient 's head and neck may be maneuvered or

    the LMA-Fastrach posit ion adjusted by using the metal handle (22 8).

    Fiberoptic intubation has a high success rate (134,24 0,27 2,31 1,32 8). It can be

    performed awake (257 ,261,32 9,330,33 1). I t has been used in patients with unstable

    necks (261,32 8,33 0). I t is easier than intubation with a r igid laryngoscope or using

    only a f iberscope in p atients with manual in-l ine stabil izat ion (259 ,328 ). It allows an

    examination of the lower airway. I t can be performed easier and in less t ime if the

    recommended tracheal tube is used (33 2).

    During f iberoptic intubation with a size 3 or 4 LMA-Fastrach, venti lat ion may be

    inadequate (333). With a s ize 5 LMA, venti lat ion is generally acceptable.

    The LMA-Fastrach can be used with an optical stylet (334).

    Light-guided Intubation

    An i l lumina ted f lex ib le f iber or a ligh ted intub ati on style t in serte d through the

    tracheal tube extending just beyond the tracheal tube tip can be used to guide a

    tracheal tube into the glott ic opening

    (247 ,270,291,292 ,308,31 0,312 ,322 ,335,336 ,337 ,338 ,339,340 ,341 ,342,343,34 4). A

    dist inct central point of l ight without a halo in the midline indicates correct

    placement. Once correct posit ion is achieved, the tracheal tube is advanced.

    If resistance is felt , correct transil lumination is not observed or the l ight point is

    seen moving laterally, the tracheal tube should be withdrawn 1 cm beyond

    P.473

    the epiglott ic elevator bar and one or more of the following adjust ing maneuvers

    applied, depending on the cause of resistance and observation of the l ight before

    each addit ional intubation attempt: f lexion or extension of the handle, withdrawing

    the LMA-Fastrach by 5 cm followed by reinsert ion, performing manual venti lat ion

    and adjust ing the LMA-Fastrach posit ion unti l the optimal seal is obtained, f lexing

    the neck and extending the head, using a smaller tracheal tube, using a larger or

    smaller LMA-Fastrach, or twist ing the L MA handle (27 0,32 2,33 5,33 6,338,340,34 5).

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    26/260

    Light-guided intubation can decrease the time to successful intubation and the

    number of attempts and failures when insert ing a tracheal tube through the L MA-

    Fastrach compared with blind intubation (308 ,310,312,336,339,340).

    Removing the LMA-Fastrach after Intubation

    Af ter the tra che a ha s been intuba ted , the de c ision need s to be made whether to

    remove the LMA-Fastrach or to leave it in place. It is usually recommended that the

    LMA-Fastrach be removed (346,34 7,348). Alternately, the cuff can be deflated to 20

    to 30 cm H2O and the LMA-Fastrach left in s itu (34 9). The LMA-Fastrach may exert

    mucosal pressures in excess of capil lary perfusion pressure (34 8). Patients in

    whom the LMA-Fastrach is retained have a higher incidence of hoarseness, sore

    throat, and dysphagia (246,346).Removing the LMA is associated with a hemodynamic response (350). Delaying

    removal for a few minutes may s light ly decrease the associated p ressor response

    (273 ,351).

    The tracheal tube needs to be stabilized to prevent extubation during LMA-Fastrach

    removal. The tube connector needs to be removed. A stabil izer rod (extender) that

    is placed in the end of the tracheal tube is available from the manufacturer (Fig.

    17.11). Insert ing the t ip of a smaller tracheal tube into the end of the inserted

    tracheal tube will allow ventilation while the LMA-Fastrach is being removed (35 2).

    View Figure

    Figure 17.11To stabilize the tracheal tube and to prevent

    extubation during LMA-Fastrach removal, a stabilizer rod(extender) is placed in the end of the tracheal tube.

    The LMA-Fastrach cuff is deflated, and the LMA-Fastrach is swung out of the

    pharynx into the oral cavity while applying counterpressure to the tracheal tube

    (239 ). The stabil izing rod is removed when the LMA-Fastrach cuff is clear of the

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    27/260

    mouth. The tracheal tube is then f irmly grasped while unthreading the inf lat ion tube

    and pilot balloon from the LMA-Fastrach. Finally, the tracheal tube connector is

    replaced.

    Problems with Intubation

    Pharyngeal pathology may make intubation through the LMA-Fastrach impossible

    (311 ,331,353).

    Currently, the smallest size available of the LMA-Fastrach is the number 3. This

    has been found to work well for intubation of patients over 30 kg, but for patients

    under this weight, successful intubation through this device is less certain (313 ).

    The LMA-Fastrach tracheal tube is expensive and should not remain in place for

    long periods of t ime because it has a high-pressure cuff.The LMA-Fastrach requires more t ime for i ntubation and results in more esophageal

    intubations and mucosal trauma than rigid laryngoscopy (309 ). Blind intubation

    through the LMA-Fastrach generates cardiovascular responses s imilar to tracheal

    intubation using direct laryngoscopy (309,35 0,354).

    When the LMA-Fastrach is removed, the tracheal tube may be d isplaced downward

    or dislodged (276 ).

    P r o b l em s

    The rigid LMA-Fastrach shaft cannot easily adapt to a change in the posit ion of thepatient 's neck. I t is more l ikely to be dislodged than the LMA-Classic if head or

    neck manipulat ion is required. I t should not be used in cases where the patient wil l

    be in the prone posit ion.

    The LMA-Fastrach is unsuitable for use in the MRI unit .

    A cas e of obs truct io n after the LMA -Fastra ch was ins erte d has bee n rep ort ed (355 ).

    Fiberoscopy revealed that the epiglott ic elevating bar was in the laryngeal aperture,

    and though it l i f ted the epiglott is, the arytenoid cart i lage was pressed anteriorly by

    the LMA-Fastrach cuff, part ial ly obstruct ing the laryngeal aperture. Despite the

    obstruct ion, the trachea was intubated successfully.

    The large diameter of the LMA-Fastrach airway tube can cause dif f iculty during

    insert ion in the patient with a l imited mouth opening and may put dentit ion at r isk

    (356 ,357). Compared with the LMA-Classic, the L MA-Fastrach causes an increased

    incidence of sore throat, sore mouth, and dif f iculty swallowing (358).

    While LMA-Fastrach is easier to place than the LMA-Classic and placement is more

    likely to be successful in

    P.474

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    28/260

    pat ients with immobil ized cervical spines (24 0,28 5,329), the LMA-Fastrach may

    exert pressure on the cervical spine (348,35 9). Intubation through the LMA-

    Fastrach may cause signif icant motion of the cervical spine. I t may be dif f icult to

    insert in the patient with a cervical collar, especially if cricoid pressure is used

    (256 ).

    Anesthes ia prov ide rs who hav e l imi ted us e of the lef t arm wi l l fi nd th e LMA -

    Fastrach dif f icult to use (360 ).

    LMA-CTrach

    De s c r i p t i o n

    The LMA-CTrach (Fig. 17.12) is si milar in construct ion to the LMA-Fastrach

    (361 ,362). I t has two built- in f iberoptic channels, one to convey l ight from and the

    other to convey the image to the viewer. These emerge at the distal end of the

    airway tube under the epiglott ic elevating bar, which l i f ts the epiglott is as the

    tracheal tube passes through the LMA-CTrach into the larynx. The fiberoptic system

    is sealed and robust, so the LMA-CTrach can be autoclaved.

    The monitor (viewer) is attached to the LMA-CTrach via a magnetic latch connector.

    I t has controls for focusing and image adjustment. The viewer is battery operated.

    The battery provides up to 30 minutes of continuous use and can be recharged.The LMA-CTrach is available in sizes 3, 4, and 5 and is reusable up to 20 t imes

    (361 ).

    View Figure

    Figure 17.12The LMA-CTrach. It has two built-in

    fiberoptic channels, one to convey light from and the otherto convey the image to the viewer. The monitor (viewer) is

    attached to the LMA-CTrach via a magnetic latchconnector.

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    29/260

    Use

    The LMA-CTrach is lubricated and inserted similar to the LMA-Fastrach without the

    viewer attached (361 ). An antifogging solut ion should be applied to the optical l ens.

    Af ter the LMA has bee n inserted , the airway sec ured, an d the pa tie nt ventilated, the

    viewer is s witched ON and attached. A real-t ime image of the larynx is then

    displayed. I f posit ioning is not sat isfactory, various maneuvers can be performed to

    improve the view (361). After a s at isfactory glott ic aperture image is achieved, the

    tracheal tube is advanced and viewed as it enters the trachea. Once the patient is

    intubated, the viewer is detached and the laryngeal mask removed, leaving the

    tracheal tube in place.

    E v a l u a t i o n

    The LMA-CTrach has poorer image quality than a f lexible f iberoptic endoscope. In

    contrast to the view through a laryngoscope, the viewer provides visualizat ion from

    the underside of the tracheal tube. I ts advantages include the abil i ty to align the

    LMA outlet with the larynx and a high first intubation attempt success rate with

    minimal neck movement. As with the LMA-Fastrach, the LMA-CTrach cannot be

    used easily in the patient with a l imited mouth opening.

    Tracheal intubation was successful at the f irst or s econd attempt in more than 96%

    of patients, higher than blind intubation via the LMA-Fastrach (36 1,362 ). The viewmay be obstructed by secret ions, lubricant, or blood. I t has proved useful during

    awake intubation in the presence of an unstable cervical spine (363).

    LMA-ProSeal

    De s c r i p t i o n

    The LMA-ProSeal (LMA-PROSEAL, PLM) (Fig. 17.13) has four main parts: the cuff,

    inf lat ion l ine with pilot balloon, airway tube, and drain (gastric access) tube

    (364 ,365,366,367 ). All components are made from sil icone and are l atex-free. I t is

    available in six sizes (Table 17.5). Studies indicate that the size 4 is preferable for

    P.475

    most adult women and the size 5 f or most adult men (368 ,369).

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    30/260

    View Figure

    Figure 17.13LMA-ProSeal. Note theintegral bite block.

    TABLE 17.5 LMA-ProSeal

    LMA

    Size

    Patient

    Size (kg)

    Maximu

    m Cuff

    Inflation

    Volume

    (mL)

    Maximu

    m Gastr ic

    Tube Size

    (French)

    Maximu

    m

    Fiberopti

    c Scope

    Size (mm)

    Length

    of Drain

    Tube

    (cm)

    Largest

    Tracheal

    Tube (ID in

    mm)

    1.5

    5to10

    7 10 18.2

    4.0uncuffe

    d

    2 10to

    20

    10 10 19.0

    4.0uncuffe

    d

    2.5

    20to30

    14 14 23.0

    4.5uncuffe

    d

    3 30to50

    20 16 26.5

    5.0uncuffe

    d

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    31/260

    4 50to

    70

    30 16 4 27.5

    5.0uncuffe

    d

    5 70to

    100

    40 18 5 28.5

    6.0cuffed

    ID, internal diameter.

    The airway (breathing, venti lat ion) tube of the LMA-ProSeal is s horter and smaller

    in diameter than that of the LMA-Classic and is wire reinforced, which makes it

    more f lexible. There is a locating strap on the anterior distal tube to prevent the

    finger sl ipping off the tube and to provide an insert ion slot for the introducer tool.

    An ac cessory ven t unde r the dra inage tube in th e bo wl prev en ts secre ti ons f ro m

    pooling and acts as an accessory venti lat ion port. The LMA-ProSeal has a deeper

    bowl than the LMA-Classic and does not have aperture bars. There is a bite block

    between the tubings at the level where the teeth would contact the device.

    The drain (drainage, esophageal drain) tube is parallel and lateral to the ai rway

    tube unti l i t enters the cuff bowl, where it continues to an opening in the t ip that is

    sloped anteriorly (Fig. 17.14). When the LMA-ProSeal is correct ly posit ioned, the

    cuff t ip l ies behind the cricoid cart i lage at the origin of the esophagus. I t al lows

    liquids and gases to escape from the stomach, reduces the risks of gastric

    insuff lat ion and pulmonary aspirat ion, allows devices to pass into the esophagus,

    and provides information about the LMA-ProSeal posit ion. The drain tube is

    designed to prevent the epiglott is from occluding the airway tube, eliminating the

    need for airway bars (368). A gastric tube, Doppler probe, thermometer,

    stethoscope, or medication can be passed into the esophagus through the drainage

    port (37 0,37 1,372,37 3,37 4). A plast ic support ing ring around the distal drain tube

    prevents the tube from collapsing when the cuff is inf lated.

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    32/260

    View Figure

    Figure 17.14Patient end of LMA-ProSeal. The drain tubecontinues to an opening in the tip.

    The LMA-ProSeal has a second dorsal cuff (Fig. 17.15). This pushes the mask

    anteriorly to provide a better seal around the glott ic aperture and helps to anchor

    the device in place (367 ). The dorsal cuff is not present on sizes 11/2 to 2

    1/2. The

    cuff is softer than that on an LMA-Classic. The ventral cuff is larger proximally to

    improve the seal.

    A silicone-c oated ma l le ab le me ta l in tr od uc er to fac il itate pl ac ement of the LMA -

    ProSeal is available (Fig. 17.16). I t has a curved, malleable si l icone-coated blade

    P.476

    with a guiding handle. The distal end f its into the locating strap, and the proximal

    end f its i nto the airway tube.

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    33/260

    View Figure

    Figure 17.15Posterior of the LMA-ProSeal, showing thedorsal cuff.

    View Figure

    Figure 17.16Metal introducer used to facilitate placementof the LMA-ProSeal.

    I n s e r t i o nInsertion Methods

    I t is recommended that the LMA-ProSeal cuff be deflated into a wedge shape, as

    with the LMA-Classic. The patient should be in the snif f ing posit ion (lo wer neck

    flexion and head extension). I t may be necessary to brief ly release cricoid pressure

    to allow the L MA-ProSeal to pass (375 ).

    In t r o d u c e r T ec h n i q u e

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    34/260

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    35/260

    With this technique, a lubricated stylet, bougie, f iberoptic endoscope, suction

    catheter, lightwand, or gastric tube is first placed through the drain tube

    (366 ,378,379,380 ,381,38 2,383 ,384 ,385,386 ,387 ,388 ,389,390 ,391 ). The patient end

    of the device is then inserted into the esophagus under laryngoscopic or f iberscopic

    guidance. The bougie should be point ing posteriorly, opposite to when it is used for

    intubation. The LMA-ProSeal is then advanced into place over the device. This

    method avoids folding the t ip backward. I t is more successful and less traumatic

    than using the introducer tool or digital me thods (381 ,385,388,392 ). This method

    has been used for patients with known dif f icult airways (39 3,39 4), after failed

    posit ioning of an LMA-Flexible (395 ), and to exchange an LMA-ProSeal (39 6).

    Cuff InflationAf ter the LMA -ProSea l has be en inserted, th e cuff shou ld be inf la ted wi th enough

    air to achieve an intracuff pressure of up to 60 cm H 2O. During insert ion and cuff

    inf lat ion, the front of the neck s hould be observed to see if the c ricoid cart i lage

    moves forward, indicat ing that the mask has correct ly passed behind it . The cuff

    volume required for the LMA-ProSeal to f orm an effect ive seal with the respiratory

    tract is lower than for the LMA-Classic (397 ). In fact, an adequate seal can be

    obtained i n most patients with no air in the cuff; however, the cuff should be

    inflated with at least 25% of the maximum recommended volume to ensure an

    effect ive seal with the gastrointest inal tract (398).

    Tests after Insertion

    A smal l amou nt (1 to 2 mL ) of wa te r-b as ed gel or a soap bubble shou ld be plac ed

    on the end of the drainage tube that protrudes from the mouth and posit ive

    pressure applied to the airway tube (399 ,400 ,401,40 2,403 ). If the LMA-ProSeal is

    properly placed, there should

    P.477

    be a slight up/down movement of the lubricant/soap. The soap bubble may move

    when the lubricant gel does not ( 40 4). I f there is no movement or the bolus is

    ejected, the mask may not be correct ly placed.

    The drainage tube should be tested for patency. This can be done by passing a

    gastric tube, a flexible endoscope, or a lighted stylet through the drainage tube

    (405 ,406). Easy passage indicates c orrect posit ioning; dif f iculty suggests that the

    mask should be reposit ioned, even if venti lat ion is sat isfactory.

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    36/260

    The suprasternal notch tap test involves tapping the suprasternal notch or cricoid

    cart i lage and observing simultaneous movement of the soap bubble at the proximal

    end of the d rainage tube (36 6,40 0). However, this can produce both false posit ive

    and false negative results (40 7).

    Proper LMA-Proseal posit ioning can also be detected by insert ing a l ighted stylet

    (406 ). I f the t ip is folded over, the stylet wil l meet resistance 1 to 2 cm from the t ip

    of the drain tube.

    Insertion Problems

    Several malpositions for the LMA-ProSeal have been described, including

    insuff icient depth, the t ip inserted into the glott is, the t ip folded backward, and

    severe epiglott ic downfolding (408,409 ).I f the LMA-ProSeal is inserted to an insuff icient depth, there wil l be a poor seal. I f

    a bubble of air is se en or the bolus of gel is ejected when the lungs are inf lated

    with less than 20 cm H 2O, the respiratory and gastrointest inal tracts are not

    isolated from one another. I f advancing the LMA-ProSeal does not correct the

    problem, it should be removed and reinserted.

    If the LMA-ProSeal has entered the v estibule of the larynx, venti lat ion may be

    obstructed and gas may leak up the drainage tube or cause the gel in the drain

    tube to move up and down with the cardiac rhythm (401 ,403 ). Pressure on the chest

    leading to bubble formation confirms that the laryngeal vest ibule has been entered

    (409 ). Advancing the mask deeper wil l exacerbate the obstruct ion. The mask should

    be removed and reinserted (377).

    I f the LMA-ProSeal t ip folds backward, the drain tube wil l be pinched off , leaving

    the upper esophageal sphincter open (41 0,411 ,412 ,413). Aspirat ion of gastric

    contents secondary to this malposit ion has been reported (414). The lubricant/soap

    bubble at the end of the drain tube wil l not move, even with gentle tapping on the

    suprasternal notch (39 9,40 0). This malposition may be associated with resistance

    to insert ion, unexpectedly high inf lat ion pressures, and the bi te block port ion

    protruding from the mouth. Inabil i ty to pass a gastric tube freely to the t ip of the

    drain tube confirms the malposit ion. Several techniques may be used to correct the

    situation, including reinsert ion with a lateral approach; reinsert ion with the drainage

    tube st if fened by a gastric tube, bougie, or s tylet; bougie-guided reinsert ion; and

    sweeping a finger behind the cuff (366 ).

    Severe epiglott ic downfolding occurs when the epiglott is is dragged inferiorly and

    completely covers the glott ic inlet. To correct this problem, the LMA-ProSeal should

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    37/260

    be reinserted with the head and neck in a more extreme snif f ing posit ion, with a jaw

    thrust applied or with the epiglott is elevated by using a laryngoscope (366 ).

    In s e r t i n g a Dev i c e t h r o u g h t h e D r a i n T u b e

    The device to be inserted through the drain tube should be well lubricated and

    passed slowly and c arefully, not forced. Warming the device may fac il i tate passage

    (415 ). Some resistance may be felt where the drain tube angulates (41 6). It may be

    helpful to use back-and-forth motions while insert ing a gastric tube to advance

    lubricant into the curved segment and to ease passage through this port ion of the

    drain tube (41 7). Inabil i ty to pass a gastric tube indicates mask misplacement

    (407 ).

    Suction should not be applied to the end of the drain tube as the LMA is beinginserted, because this may cause the drain tube to c ollapse and injure the upper

    esophageal sphincter. Suction should not be applied to a gastric tube unti l i t has

    reached the stomach. The gastric tube may be used to reinsert the LMA-ProSeal if

    i t becomes displaced (38 0). The drain tube should not be c lamped.

    A gas tric tu be ins erted th rough the LMA -ProSeal can be conv erte d to a nasoga stric

    tube (418).

    T r ac h e al In t u b a t i o n t h r o u g h t h e LMA - P r o S ea l

    Tracheal intubation through the LMA-ProSeal requires a long narrow tracheal tubeor an a irway exchange catheter (41 9). After the LMA-ProSeal is removed, a larger

    tube can be substituted, if necessary.

    Use

    The LMA-ProSeal can be used for both spontaneous and controlled venti lat ion, but

    is more suited to controlled venti lat ion (409,420 ). The sealing pressure is higher

    than with the LMA-Classic in adult and pediatric patients, making it a better choice

    for situations where higher airway pressures are required, where better airway

    protect ion is desirable, and for surgical p rocedures in which intraoperative gastricdrainage or decompression is needed

    (364 ,365,397,398 ,411,41 3,421 ,422 ,423,424 ,425 ,426 ,427,428 ,429 ,430,431,43 2,43 3,

    434,435 ,436 ,43 7,438,43 9,44 0,44 1,442). Case reports show no aspirat ion of gastric

    contents despite regurgitat ion or v omit ing unless the LMA-ProSeal is malposit ioned

    (414 ,436,443,444 ,445,44 6,447 ,448 ,449,450 ,451 ,452 ,453,454 ). However, aspiration

    has been reported with malposit ioning (45 5).

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    38/260

    I t may be easier to place the LMA-ProSeal than the LMA-Classic during manual in-

    l ine neck stabil izat ion (424 ). I t has been used in cases of known dif f icult airway

    (428 ) and has been successfully used after fai lure with an LMA-Classic (375,427 ).

    The LMA-ProSeal may be useful in cases where it is important to avoid airway

    trauma, as it exerts lower pressures against the pharyngeal mucosa than the LMA-

    Classic (39 7). However, airway trauma as evidenced by

    P.478

    blood on the device after removal is higher for the LMA-ProSeal than for the LMA-

    Classic (40 9).

    The LMA-ProSeal has been found to be s afe for use in an MRI unit , but imaging

    quality may be compromised, depending on the pulse sequence that is used and

    whether the area of interest is near the LMA (45 6).

    P r o b l em s w i t h t h e LMA -P r o S e a l

    The LMA-ProSeal is less s uitable as an intubation device than the LMA-Fastrach

    because of the narrower airway tube. The fiberscope and tracheal tube sizes that

    can be accommodated by the LMA-ProSeal are given in Table 17.5. The high

    resistance associated with the smaller lumen may make it l ess suitable for use with

    spontaneously breathing patients than other devices ( 45 7).

    The LMA-ProSeal may be somewhat more dif f icult and take s light ly longer to insert

    than the LMA-Classic in adults, although overall success is equivalent

    (364 ,365,397,408 ,411,41 3,421 ,422 ,432,433 ,458 ). The incidence of intraoperative

    complicat ions and postoperative sore throat are similar. In chi ldren, the ease of

    insert ion is similar to the LMA-Classic (43 4,435 ,437,459 ). The LMA-ProSeal

    requires a greater depth of anesthesia for insert ion than does the LMA-Classic

    (460 ).

    The LMA-ProSeal can cause airway obstruction after insertion, either by

    compressing the supraglottic and glott ic s tructures or by cuff infolding (462 ,463).

    Removing air from the cuff or p lacing the patient in the snif f ing posit ion may relieve

    the obstruction.

    Part ial upper airway obstruct ion during spontaneous v enti lat ion may result in

    aspirat ing air through the drain tube into the esophagus (40 4,46 3). Esophageal

    insuff lat ion can occur s imultaneously with venting from the drainage tube during

    posit ive pressure venti lat ion with malposit ion (46 4,46 5). This may result in

    inadequate venti lat ion.

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    39/260

    I t may not be possible to insert a gastric tube in some patients (466 ,467 ). This may

    be due to select ion of too large a tube, inadequate lubricat ion, using a cooled

    gastric tube, cuff overinf lat ion, or malposit ion (415).

    The LMA-ProSeal is relat ively contraindicated for intraoral surgery because it

    cannot be moved easily around the mouth, the drainage tube is vulnerable to

    occlusion, and the larger proximal cuff could interfere with the surgical f ield (36 5).

    The LMA-ProSeal has a shorter l i fe span than the LMA-Classic (468 )

    Us i n g t h e LMA Fam i ly

    An LMA of th e cho sen size plus on e s ize smal le r and larg er sho uld alwa ys be

    immediately available. The sy ringe used to inf late the LMA should contain only air.

    Inject ing organic substances such as propofol from a previously used syringe may

    damage the LMA (469 ).

    Preuse Inspection

    V i s u a l In s p e c t i o n

    The f irst s tep is to examine the tube. The airway tube should not be discolored, as

    this would prevent seeing f luids that may enter the tube. There should be no cuts or

    tears in the tube, and the spiral wires should not be kinked. The rest of the LMA's

    external surface should be examined for damage such as cuts, tears, scratches, or

    foreign part icles.

    The interior should be free f rom obstruct ion or foreign part icles. The LMA-Flexible

    should be examined to make certain that the reinforcing wire is wholly contained

    within the wall of the tube.

    The tube should be f lexed up to, but not beyond, 180. Kinking should not occur.

    Bending the tube beyond 180 could cause permanent damage.

    The next test is to examine the mask aperture. The bars should be gently probed to

    make certain that they are not damaged and the space between them is free from

    particulate matter. If the drain tube in the LMA-ProSeal bowl is torn or perforated,

    the LMA should not be used.

    The connector should f i t t ight ly into the outer end of the airway tube. I t should not

    be possible to remove it easily. The connector should not be twisted. If the

    connector has cracks or surface irregularit ies, i t should not be used.

    De f l a t i o n / I n f l a t i o n

    The next step is to withdraw air from the cuff so that the walls a re f lattened against

    each other. Excessive force should be avoided. The cuff should not reinf late. The

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    40/260

    syringe should be removed from the inf lat ion valve and the cuff checked to make

    certain that it remains deflated. I f i t reinf lates, there is a faulty valve or leaking

    cuff. An LMA cuff with a hole may not reinflate after the air had been removed

    (470 ).

    The next step is to inf late the cuff with 50% more ai r than the recommended

    maximum inf lat ion volume (Table 17.6).

    The cuff should hold the pressure for at least 2 minutes. Any herniat ion, wall

    thinning, or asymmetry is an indicat ion to discard the LMA. The balloon should be

    ell ipt ical, not spherical or irregularly shaped. Excessive pilot balloon width

    indicates weakness and imminent rupture. Failure to perform this test may miss

    problems with the cuff (467 ,471 ).

    Mask Preparation

    The cuff should be fully deflated with a dry syringe to form a f lat oval disc (Fig.

    17.17) This can be done by pressing the hollow side down against a c lean, hard,

    f lat surface. The d eflated cuff should be wrinkle-free.

    A cuf f-d ef lati ng too l is av ailabl e from the ma nu fac tu re r (Fig. 17.18). This device wil l

    provide a superior and more consistent shape than either hand manipulat ion or free

    deflat ion but does not offer any benefits in

    P.479

    terms of residual volume (472 ). The use of this device wil l lengthen the cuff l i fe

    (473 ).

    TABLE 17.6 Maximum Test Cuff Inflation Volumes

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    41/260

    Size LMA-Classic or LMA-

    Unique (mL)

    LMA-Flexible

    (mL)

    LMA-Fastrach

    (mL)

    LMA-ProSeal

    (mL)

    1 6

    1.5 10

    2 15 15

    2.5 21 21

    3 30 30 30

    4 45 45 45 45

    5 60 60 60 60

    6 75 75

    Lubricat ion should be a pplied to the posterior cuff surface cuff just b efore insert ion,

    taking care to avoid gett ing lubricant on the anterior (bowl) su rface. The

    manufacturer recommends water-soluble jelly and does not recommend the use of

    analgesic-containing gels or sprays, because this may delay the return of protect ive

    ref lexes and may provoke an allergic reaction. While some studies show that

    lubricat ion with l idocaine gel or spray wil l result in a l ower incidence of retchingand coughing on emergence (87,47 4), another study showed increased intra- and

    postoperative problems (475). Lubricants or sprays that contain si l icone may cause

    the mask to soften and s well.

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    42/260

    View Figure

    Figure 17.17The laryngeal mask ready for insertion. Thecuff should be deflated as tightly as possible, with the rimfacing away from the mask aperture. There should be nofolds near the tip. (Courtesy of Gensia Pharmaceuticals,Inc.)

    Anesthetic Induction

    Insert ion of the LMA requires suff icient general or topical anesthesia to obtund the

    airway ref lexes. A depth similar to that necessary for insert ing an oropharyngeal

    airway but not as deep as is needed for tracheal in tubation is required (476).Abs ence of a mo to r respo ns e to a ja w thrust is a re liab le me thod for assessing th e

    adequacy of anesthesia for LMA insertion (477). Greater depth is needed for

    insert ing the LMA-ProSeal than for the LMA-Classic (46 0).

    Awake Placement

    The laryngeal mask can be inserted in an awake patient fol lowing topical

    anesthesia of the upper airway and/or nerve blocks

    (99,100 ,135,248,478 ,479 ,480,481,482 ,483). Mask insertion should be coordinated

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    43/260

    with swallowing. I t may be helpful to part ial ly inf late the cuff to simulate a bolus of

    food (484).

    Cuff Inflation and Assessing Position and FunctionThe cuff should be inf lated to a pressure of approximately 60 cm H 2O (485,486 ). A

    cuff pressure gauge is recommended for proper inflation pressure. Cuff pressure

    can be est imated by feeling the tension in the pilot balloon. A spherical pi lot

    balloon is an indicat ion that there is too much gas in the cuff.

    The cuff should be inf lated over 3 to 5 seconds without holding the tube unless the

    posit ion is obviously unstable (e.g., in edentulous patients with slack t issues). This

    usually causes slight upward movement of the airway tube, and a s light bulging at

    the front of the neck is commonly seen. There should be a smooth oval swell ing inthe neck and no cuff visible in the oral cavity.

    The recommended maximum inflation volumes are given in Table 17.7. In pract ice,

    it is rarely necessary to

    P.480

    use the full volume (30,485 ,487 ,488). Using greater-than-recommended volumes

    wil l not improve the seal against the larynx and may worsen it (19 4). A rational

    approach is to inf late the mask with half the maximum inf lat ion volume and to

    determine the oropharyngeal leak pressure, adding more air i f necessary. Cuff size

    is probably more important than inf lat ing volume in determining the seal, so

    upsizing the LMA may provide a better seal than adding more air to the cuff of a

    smaller LMA (10,16 ,489,49 0).

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    44/260

    View Figure

    Figure 17.18Cuff-deflating device for the Laryngeal MaskAirway. A:The laryngeal mask is inserted into the device.The cuff is deflated by using a syringe. At the same time,the device is compressed. B:After cuff deflation, the LMAis ready for insertion.

    I f posit ive-pressure venti lat ion is to be used, the leak pressure should be greater

    than 20 cm H2O (30 cm H 2O with the LMA-ProSeal). If spontaneous respiration is to

    be used, the leak pressure should be greater than 10 cm H 2O. This is the

    approximate pressure of f luid at the posterior pharyngeal wall i f the oral cavity is

    flooded (49 1). Unti l spontaneous respirat ion has resumed, it may be helpful to

    occlude the nose and seal the mouth around the tube to allow posit ive-pressure

    venti lat ion (492 ).

    The airway sealing pressure is determined by observing the pressure gauge in the

    breathing system as the bag is squeezed and the pressure increases. Several

    methods can be used to determine the leak pressure (493,494). A s tethoscope can

    be placed just lateral to the thyroid cartilage. Another method is to listen over the

    mouth for a noise when the bag is squeezed. Carbon dioxide may be detected by

    placing the sample l ine in the oral cavity. Another method is determining a steady

    airway pressure after closing the adjustable pressure l imit ing (APL) valve in the

    circle s ystem.

    It may be possible to improve the seal by adding more air to the cuff ( i f the

    maximum recommended volume has not been injected) or by flexing or rotating the

    head and neck slight ly (22 2,495,496). The leak pressure wil l be higher if the head

    and neck are f lexed or rotated (222 ,433,497,498 ). Higher pressures may be

    achieved by applying pressure on the front and/or side of the neck, by applying

    continuous forward pressure on the LMA, or by lift ing the handle of the LMA-

    Fastrach (499,50 0,501 ,502,503 ,50 4).

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    45/260

    Indications that the LMA is properly positioned include normal breath sounds, chest

    movements, pressure-volume loops and volume monitoring not showing a leak , and

    carbon dioxide waveforms with posit ive-pressure venti lat ion. I f the patient is

    breathing spontaneously, normal reservoir bag excursions and absence of signs of

    obstruct ion are indicat ions of proper

    P.481

    placement (506 ). A f iberscope or r igid endoscope can be inserted through the LMA

    to confirm its posi t ion and rule out airway obstruct ion (73 ,79 ,10 9,506 ,507). X-ray or

    MRI can also be used to confirm the posit ion (75,508 ). An esophageal detector

    device can be used (50 9,510), although its ut i l i ty has been questioned (511).

    TABLE 17.7 Maximum Cuff Dimensions

    Mask

    Size

    Air Volum

    (mL)

    Maximum Bulge of

    Cuff Ti p (mm)

    Maximum Bulge of

    Wide End of Cuf f (mm)

    Maximum Transverse

    Diameter of Cuff (mm)

    1 6 7.8 8.6 26.3

    1.5 10 9.5 10.2 32.6

    2 15 11.5 13.0 39.0

    2.5 21 13.0 14.5 45.0

    3 30 14.8 16.6 51.2

    4 45 17.0 19.0 58.5

    5 60 21.1 22.4 68.3

    I f the airway is obstructed, the cause may be incorrect mask posit ion, a downfolded

    epiglott is, a closed glott ic sphincter, or an overinf lated cuff. In most cases,

    removing and reinsert ing the mask wil l el iminate the obstruct ion. Another technique

    is to l i f t the anterior neck structures by using a gloved hand inserted into the

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    46/260

    mouth, deflate the cuff, and rotate the mask 360 (512). In some cases, the

    epiglott is may be straightened digital ly (513). Jaw manipulat ion or reposit ioning the

    head usually does not relieve airway obstruct ion. Removing air from the cuff may

    be helpful (51 4). I f despite these efforts sat isfactory v enti lat ion cannot be achieved,

    the device should be withdrawn and reinserted or a dif ferent size LMA or tracheal

    tube should be used.

    Fixation

    A bi te bloc k or ro l l of gauze should be inserted into th e mouth bes ide th e tube to

    prevent the patient from bit ing the tube and to i mprove stabil i ty. This is not

    necessary with the LMA-ProSeal. Various other devices have been used

    (515 ,516,517,518 ). An oropharyngeal airway should not be used, because both itand the LMA are designed to be placed in the midline, and the airway t ip might

    compromise the LMA cuff or cause tube compression (519 ). Also, an oropharyngeal

    airway may not prevent the tube from being bi t ten (121 ,520).

    The tube should be secured with tape, taking care that it does not become twisted.

    This can be accomplished by aff ixing the tape f irst to the maxil la, winding over the

    cephalad side of the tube, and down around the caudal side to f ix the tube and bite

    block f irmly to each other and to the opposite maxil la (521). Further security can be

    provided by taping from zygoma to zygoma under the mandible (52 2) or around the

    neck (52 3,52 4). A tracheal tube holder may be used (525 ). Other fixation methods

    have been described (526 ,527 ). The f ixat ion method should not obstruct the

    surgery. A suture around a tooth may be used if tape wil l be in the way.

    Bending the tube against its natural curvature may cause it to become dislodged or

    kink, unless the LMA-Flexible is used. Tract ion from the breathing system should

    be avoided, and several methods to achieve this have been suggested (527,529).

    Intraoperative Management

    During surgery, airway patency and correct LMA orientat ion should be verif ied at

    regular intervals. The patient 's upper abdomen should be periodically observed for

    signs of distention and epigastric auscultat ion performed. A l ighter level of

    anesthesia than would be required if a tracheal tube were used is usually possible.

    I f laryngospasm, wheezing, swallowing, coughing, straining, or breath holding

    occurs, anesthesia should be deepened or muscle relaxants administered. An

    aerosol can be administered by using an LMA (530 ).

    Nitrous oxide and carbon dioxide can dif fuse into the cuff, increasing intracuff

    pressure and volume

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    47/260

    (198 ,199,531,532 ,533,53 4,535 ,536 ,537,538 ,539 ,540 ,541,542 ,543 ,544). Cuff volume

    increases less with the LMA-Unique than with the LMA-Classic ( 198 ). The increase

    in volume may cause airway obstruct ion (514 ). Inf lat ing the cuff with n itrous oxide

    wil l avoid this increase (54 5,546).

    The manufacturer recommends that cuff pressure be checked periodically with a

    pressure gauge, transducer, or other device ( Fig. 19.29) and adjusted to keep it at

    approximately 60 cm H2O. The pilot balloon should feel compliant. I f the balloon

    feels st if f or ol ive shaped, the pressure may be excessive. Others have suggested

    that the logical method of controll ing cuff pressures during nitrous oxide anesthesia

    may be to take the just seal p ressure as a control value and withdrawn volume to

    maintain values close to this pressure ( 546).

    The LMA can be used with controlled (including mechanical) or spontaneous

    venti lat ion. Patient outcome has been found to be s imilar in nonparalyzed patients

    with posit ive-pressure venti lat ion or spontaneous breathing (547). I f controlled

    venti lat ion is used, the peak i nspiratory pressure should be kept below 20 cm H 2O

    (30 cm H2O with the ProSeal). Higher pressures may result in a leak around the

    mask, gastric distention, and operating room pollut ion (15 ,54 8,54 9,550 ,551,55 2).

    Changes in the venti latory pattern to reduce t idal volume and using muscle

    relaxants may result in a lower peak pressure. I f higher pressures are required,

    considerat ion should be given to exchanging the LMA for a tracheal tube. I f cricoid

    pressure is applied, the airway pressures at which the patient is venti lated can

    often be increased to over 30 cm H2O without gastric insuff lat ion occurring (553 ).

    Pressure control venti lat ion (Chapter 12), with or without PEEP, which is available

    on newer anesthesia venti lators, may be the mode of choice for controlled

    venti lat ion with the laryngeal mask because it al lows a lower peak pressure for the

    same t idal volume with less leak around the LMA (439,55 4,555 ). For patients

    breathing spontaneously, pressure-support venti lat ion i mproves gas exchange and

    reduces the work of breathing (556,55 7). The work of breathing can also be

    reduced by using CPAP (558 ).

    A sudde n increas e in lea kag e, sno ri ng , o r ot her sounds often sign als the need for

    more muscle relaxation, although other causes such as LMA displacement, l ight

    anesthesia causing glott ic closure, airway obstruct ion, a leaking cuff, and a

    decrease in lung c ompliance related to the surgical procedure are other possible

    causes

    P.482

  • 8/11/2019 Chapter 17. Supraglottic Airway Devices

    48/260

    (559 ). Adding air to the cuff wil l not al ways co rrect a leak and may make it worse

    by increasing tension in the cuff and pushing it away from the larynx (560).

    Sometimes, removing some air from the cuff wil l help.

    I f regurgitat ion occurs, the f irst s ign may be the appearance of f luid traveling up the

    LMA tube. Breath holding or coughing may occur. The patient should be placed in

    the head-down position, the breathing circuit disconnected, and the airway tube

    suctioned. I t may not be necessary to remove the LMA, although preparations for

    tracheal intubation should be made and the patient intubated, if indicated.

    Insert ing a nasogastric tube behind a non-ProSeal LMA can be aided by using a

    nasal airway or a f lexible endoscope to displace the LMA forward (561).

    Emergence from Anesthesia

    I t is important that the bite block or roll of gauze be left in place unti l the LMA is

    removed to maintain patency and prevent damage to the LMA (56 2,563 ). Cuff

    deflat ion should be performed only when the LMA is removed (564 ,565). I f the cuff

    remains inf lated as the LMA is removed, a greater mass of secret ions wil l also be

    removed (566), but this technique increases the incidence of blood staining (but not

    sore throat) (56 7). Taking off a glove tha t was worn when the LMA was removed

    and invert ing it over the device wil l minimize the spread of contamination (56 8).

    Keeping the LMA in place during transfer to the postanesthesia care unit (PACU)

    wil l maintain a patent airway, while l eaving the anesthesia provider's hands free for

    other tasks. During recovery, supplementary oxygen can be delivered with the L MA

    in place by using a T-piece or other device

    (569 ,570,571,572 ,573,57 4,575 ,576 ,577,578 ,579 ,580 ). With the T-piece, respirat ion

    may be assisted manually by in termittently occluding the expiratory l imb (19).

    There is controversy regarding the optimal t ime for LMA removal. I t should either

    be removed with the patient in a deep level of anesthesia or when full recovery of

    airway ref lexes has occurred. Leaving the LMA in posit ion unti l airway ref lexes

    have recovered and the patient can phonate