Fast and cost-efficient nitrogen / protein analysis for starch Case study: Dumas nitrogen determination By Tatjana Adamek and Dr. Werner Küppers, Königswinter, Germany

Upload: buitu

Post on 13-Aug-2019

215 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Case study: Dumas nitrogen determination Fast and cost ... · Fast and cost-efficient nitrogen / protein analysis for starch 2 1. Introduction and presentation of pro-blem Due to

Fast and cost-efficient nitrogen / protein analysis for starch

Case study: Dumas nitrogen determination

By Tatjana Adamek and Dr. Werner Küppers, Königswinter, Germany

Page 2: Case study: Dumas nitrogen determination Fast and cost ... · Fast and cost-efficient nitrogen / protein analysis for starch 2 1. Introduction and presentation of pro-blem Due to

Fast and cost-efficient nitrogen / protein analysis for starch

2

1. Introduction and presentation of pro-blemDue to their manufacture, starch products only contain very low concentrations of proteins. Therefore, satisfactory analysis has only been possible to date using high sample weights and long analysis times. With its innovative detection method and fast and complete combustion, DU-MATHERM® from C. Gerhardt, with its innovative detection method and fast and complete com-bustion for the detection of Nitrogen by Dumas, provides an interesting alternative to the cur-rently still common Kjeldahl method.

1.1 Use of starch and starch productsStarches and their derivatives are widely used in the food industry in the manufacture of con-fectionery, baked goods, milk products and par-ticularly beverages in the form of starch-based sugars (mainly glucose syrup, dextrose and isoglucose). Due to the properties of the starch as modifiable polymer and its composition of fermentable sugar units, starch is also widely used as renewable raw material in the chemi-cals industry.1 The many applications explain the high demand for starch-based products and the simultaneous requirement for fast and reliable analysis.

1.2 Starch productionIn our part of the world, starch is usually ob- tained from potatoes or cereals. However, starch is also obtained from numerous other plants, of which rice (broken rice from the rice husking factories) and maize are particularly important in addition to wheat and potatoes.

Internationally, manioc (tapioca) is still an im-portant starch resource. The extraction is per-formed by washing out the starch from the plant parts using a saline solution.2

2. Protein analysis in starchDue to the low concentration, proteins in starch can only be quantitatively verified with great difficulty. For the DIN method according to Kjeldahl3, big sample sizes (approx. 10 g) are weighed, digested with sulphuric acid and then subjected to further analysis steps. Strong foa-ming of the samples at the beginning of Kjeldahl is problematic for heating up the digestion mix-ture quickly to 400 °C. Large sample tubes and long digestion times are absolutely necessary. The consequence is low sample throughput per working day.

Sample type Typical sample weights [g]

Theoretical content [%] Protein

Theoretical content [%] Nitrogen

Potato starch 5–10 0.08 – 0.90 0.0128-0.1441

Maize starch 5–10 0.24 – 0.63 0.0384-0.1008

Rice starch 5–10 0.76 0.1216

Tapioca starch 5–10 0.05 – 0.87 0.0080-0.1392

Wheat starch 5–10 0.21 – 0.50 0.0368-0.0877

Table 1: Protein contents in starch and sample weights in Kjeldahl digestion3

Table 1 shows the expected protein contents in various starches. It can be seen that Nitrogen / Protein analysis in Starch is almost a trace ana-lysis using multiplication of N by the protein fac-tor (e.g. 6.25 for maize and potato products or 5.7 in wheat). With a high sample weight in the

Kjeldahl method of approx. 10 g, there is only 1 mg N (for 0.01% N content) in absolute terms. With Dumas combustion using 100-300 mg, the-re is only 0.01 mg nitrogen in absolute terms for the detection during analyses.

Page 3: Case study: Dumas nitrogen determination Fast and cost ... · Fast and cost-efficient nitrogen / protein analysis for starch 2 1. Introduction and presentation of pro-blem Due to

3

The following official methods have been used for comparison with the results obtained in this article:

– EN ISO 3188:1994 Starches and derived products - Determination of nitrogen con-tent by the Kjeldahl method - Titrimetric method (ISO 3188:1978); Issue date (Ger-man version): 10/1994

– AOAC 992.23, Crude Protein in Cereal Grains and Oilseeds

– ICC Standard method No. 167, 2000.

However, expected precision values are not de-fined in the above so the Dumas method DIN EU 16634, „Determination of the total nitrogen content by combustion according to the Dumas principle and calculation of the crude protein content“ has been used.

2.1 Combustion analysis of starch according to the Dumas principle Only the DUMATHERM® analyser from C. Ger-hardt shown in Figure 1 was used for the analy-ses presented here.

2.2 Theory: combustion method for nitrogen / protein analysis

2.2.1 Sample feeding and combustionWhile the acid digestion according to the Kjeldahl method destroys the sample in a rather rustic way, combustion provides a somewhat more elegant and time saving alternative. The sample is oxidised in a controlled way and the resulting gaseous products are subjected to further ana-lysis. The objective is the complete conversion of the sample into the primary products carbon dioxide, water and nitrogen dioxide (equation 1) without the secondary products carbon monoxi-de (CO) and nitrogen monoxide (NO) being pro-duced by incomplete combustion.

(CHNO) (s) + O2 (g) = CO2 (g) + H2O (g) + NO2 (g) (1)

As air has a nitrogen content far in excess of 70 %, the key to success here is effective sealing against the ambient air in combination with an efficient purging function.An efficient autosampler is needed so that inter-action of the sample and air is ruled out. Figure 2 shows two such models.

Figure 1: DUMATHERM® from C. Gerhardt with pneumatic autosampler and PC controller

Figure 2: Autosampler with purge function - a purge chamber is continuously purged with helium before the sample is mechanically or manually fed to the combustion by the slider.

Page 4: Case study: Dumas nitrogen determination Fast and cost ... · Fast and cost-efficient nitrogen / protein analysis for starch 2 1. Introduction and presentation of pro-blem Due to

Fast and cost-efficient nitrogen / protein analysis for starch

4

Modern combustion apparatus use helium tech-nology to expel atmospheric nitrogen whilst achieving the greatest possible precision of ana-lysis.

Figure 3 shows in detail an example of such an analysis on the DUMATHERM® from C. Ger-hardt. Based on the classic principle of combus-tion analysis according to Jean Dumas3 , almost maintenance-free operation is possible with modern control software.

The pneumatic autosampler (AS) feeds the so-lid or liquid samples (shown as red dots) packed in tin foil to the combustion furnace (CF). At the same time, the helium gas flow is changed over to oxygen to initiate the combustion. The samp-le incinerates and the ash produced is collected in the ash insert (AI). The gaseous combustion products react completely in the presence of catalysts to the required oxides and are trans-ported through the machine by a helium flow after combustion. The nitrogen oxides are first reduced to elementary nitrogen in the reduction reactor (RF) while the secondary products, wa-ter and carbon dioxide, are separated in special traps (F1 to F3). A gas flow consisting of helium and nitrogen remains, from which the nitrogen

can be measured using a thermal conductivity detector. Computer-control enables simulta-neous evaluation of the analysis data.

2.2.2 Detection and evaluationFor ultra low level nitrogen analysis – particu-larly in starch application – helium is recom-mended as carrier gas as it is the only gas which shows very different thermal conductivity in comparison with nitrogen, Table 2.

Gas name λ 300 K, W/m · K

Helium, He 156.7

Carbon dioxide, CO2 16.8

Nitrogen, N2 26.0

Carbon monoxide, CO 16.8

Table 2: Thermal conductivities of various gases at 300 K6. The greater the difference in the thermal conductivity values the better the detection capabi-lity and the precision of the machine.

Figure 3

Page 5: Case study: Dumas nitrogen determination Fast and cost ... · Fast and cost-efficient nitrogen / protein analysis for starch 2 1. Introduction and presentation of pro-blem Due to

5

Helium as carrier gas and nitrogen as analyte are the ideal combination as a reference gas flow in the detector (as used in classic detectors based on a Wheatstone bridge) is unnecessary.

As only minimal amounts of nitrogen are pre-sent in starch samples, the use of helium as carrier gas is therefore crucial for the success of the analysis.

2.3 Practice: Performing a combustion analysisIn addition to the actual analysis by the DUMA-THERM®, other points are still essential for good and reproducible performance:

– Sample preparation- Sample weight- Calibration of the detector- Quality assurance by the user- Selecting measurement conditions correctly- Assessment and evaluation of the results

Tin foil is used in sample weighing and sample balls are made from this which are then kept in the sample tray until analysis.

The foil wrapped samples are placed in the sam-ple tray and the autosampler is then closed with the cover plate before analysis (Figure 4).

2.4 New type of calibration for starch analysis using DUMATHERM®

A new starch application has been developed for the DUMATHERM®. A key component for suc-cess here is the calibration for determining the very small nitrogen peak areas of starch sam-ples.

Figure 4: Sample weighing in tin foil (picture on the left) and placing samples in the sample tray until sample feeding in the autosampler (picture on the right).

Why do we use tin foil?

Tin combusts strongly exothermic to tin(IV)oxide. The energy released during that process supports the full combustion of the sample.

As an alternative, nitrogen free weighing paper could be used to pack the samples allthough additional amounts of ash from the paper must be considered then.

Page 6: Case study: Dumas nitrogen determination Fast and cost ... · Fast and cost-efficient nitrogen / protein analysis for starch 2 1. Introduction and presentation of pro-blem Due to

Fast and cost-efficient nitrogen / protein analysis for starch

6

Figure 5: THAM (Tris(Hydroxymethyl) amino-methane) and EDTA (ethylene diamine tetra-acetic acid), with 9.57 % N) calibration for the DUMATHERM® with a total of20 measuring points

538 Mühle + Mischfutter · 149. Jahrgang · Heft 17 · 6. September 2012

Warum benutzt man Zinnfolie? Zinn verbrennt stark exotherm zu Zinn(IV)oxid. Dabei frei wer­dende Energie unterstützt die vollständige Verbrennung der Probe, sodass die Probe hierbei quasi wie ein Schwamm aufge­saugt wird von Energie. Als Alternative könnte auch stickstoff­freies Wägepapier verwendet werden, allerdings fallen hier große Mengen zusätzlicher Asche an, welche abgetrennt wer­den muss.

2.4 Neuartige Kalibration für die Analytik von Stärke mittels „Dumatherm™“Für die Analytik mittels „Dumatherm™“ wurde eine neue Stär­keapplikation entwickelt. Wesentlicher Baustein zum Erfolg ist hierbei die Kalibration, um die nur sehr kleinen Peakflächen des Stickstoffes aus Stärkeproben quantitativ zu bestimmen.Zur Berechnung der Analysenergebnisse muss eine Kalibration verwendet werden, welche die Signalgrößen der unbekannten Proben vollständig abdeckt. Bei den verwendeten Einwaagen von ca. 150 mg wurden Signalgrößen von ca.150–500 mVs er­reicht. Da die Signalgrößen am absoluten Detektionslimit lie­gen, empfiehlt sich eine mehrstufige Kalibration mit einem Feststoffstandard als auch einem Flüssigstandard. Abb. 5 zeigt eine derartige Kalibrationskurve, erweitert durch Messpunkte nach oben hin mit dem Feststoffstandard EDTA.

Für die Analysen der unbekannten Proben wurden zunächst Standardbedingungen verwendet, d. h. ca. 100–200 mg Einwaa­ge, Sauerstoff­Dosierung mit 400 ml/min, 990 °C Verbrennungs­temperatur. Die folgenden Abb. 6 und 7 zeigen Analysenergebnisse am Bei­spiel der Kartoffelstärke und Weizenstärke.

Als Besonderheit wurde die Kartoffelstärke noch mit stark un­terschiedlichen Einwaagen gemessen, wobei auch hier noch aus­gezeichnete relative Standardabweichungen von <0,6% erhal­ten wurden. Dies zeigt, dass für die Analysen nur 100 mg Ein­ waage notwendig ist. Die erhaltenen Peakflächen sind mit ca. 200 mVs sehr nahe am Detektionslimit von 0,01 mg N absolut.

2.5 Langzeitstabilität der KalibrationDie entwickelte Kalibration hatte eine sehr gute Stabilität. Mit einer Kartoffelstärke­ und einer Weizenstärkeprobe, die über fast 2 Monate an 24 Arbeitstagen kontinuierlich in der täglichen Routine mit analysiert wurden, konnte gezeigt werden, dass die neu ausgearbeitete Kalibration eine hochpräzise Analytik im Bereich von unter 0,25% Protein mit dem „Dumatherm™“ sta­bil möglich macht. Die Ergebnisse sind in Abb. 8 gezeigt.

2.6 FazitDie Stickstoff­/Proteinanalyse nach Dumas wird mit den heute verfügbaren Verbrennungsgeräten eine ernstzunehmende Al­ternative zur Säureaufschlussmethode nach Kjeldahl. Trotz nur geringen Einwaagen ist eine ausgezeichnete Reproduzierbar­keit und Wiederholbarkeit ohne großen Aufwand möglich. Die vorgestellte Kalibrationsstrategie erlaubt die verlässliche Ana­lyse von Stickstoffgehalten am Detektionslimit von 0,01 mg N absolut, bis hin zu hochproteinhaltigen Glutenproben (60–80% Protein). Da die untersuchte Stärke eine ausgezeichnete Homo­genität aufweist, waren sogar beste Analysenergebnisse mit nur 100 mg Probeneinwaage möglich. Hiermit sind die üblichen Pro­bleme im Kjeldahl­Aufschluss mit starkem Schäumen und hohen Chemikalienverbräuchen eliminiert und man kann schnell und sauber eine quantitative Analyse mit einem Verbrennungsgerät erhalten. Nicht ohne Grund etabliert sich deshalb die Verbren­nungsmethode immer mehr als zweite Referenzmethode neben der Kjeldahl­Methode.

3. Literatur1. www.de.wikipedia.org/wiki/St%C3%A4rke

2. Fachverband der Stärke­Industrie e. V.; www.staerkeverband.de/html/zahlen.html

3. DIN EN ISO 3188:1994, Stärken und Stärkederivate­Bestimmung des Stickstoff­gehaltes nach dem Kjeldahl­Verfahren – Volumetrisches Verfahren (ISO 3188:1978); Deutsche Fassung Ausgabedatum: 1994 ­10

4. Souci, S. W., W. Fachmann und H. Kraut: Die Zusammensetzung der Lebens­mittel. Nährwert­Tabellen. – 5. überarbeitete und ergänzte Auflage,; medpharm Scientific Publishers, Stuttgart (1994), CRC PRESS, Tokyo, S. 596 ff

5. Traité Élémentaire de Chimie“, Vol. II (1789), Kapitel VII, S. 493–501

6. www.wikipedia.de (2011)

Stickstoff-/Proteinanalytik

Abb. 5: THAM­ (Tris(hydroxymethyl)­aminomethan) und EDTA­ (Ethylendi­amintetraessigsäure, mit 9,57% N) Kalibration für den „Dumatherm™“ mit insgesamt 20 Messpunkten

oC soC soC

sccm

sccm

sccm

He I 195,3

Flow Rates:

O2 400,1

He II 199,9

AutoRun Time

Sample Table :

649Reduction Reactor

5

Method_1

19.03.2012 11:26:23 3288

Degassing Oven

Analysis Conditions for Method :

989

Temperatures:

Calibration # :

Combustion Reactor

299

9

Analytische Studien

Times:

Sample Delay

Sample Stop

23.03.2012 12:46:49 A 1,2

0,55

0,110

0,000 0,00

0,568

0,018

Protein

[%]

Category

0,0181,256E+02 0,109

Peak Area

[mV*s]

Nitrogen

[%]

N Weight

[mg]

46 (Cubic, With Zero )

3288

RSD [%]

99,487

Average

Standard Deviation

0,018 0,111

3288 149,531

Dumatherm Nitrogen / Protein Analyser

A 1,2 6,25 1,743E-02

Date Time Protein

factor

Sample name Standard nameWeight

[mg]

19.03.2012 11:15:17

Serial Number :

Software Version:

159

DUMATHERM MANAGER V2.06E Adamek

Submitter:

Operator:

151,112 A 1,2 6,25 1,622E+02 2,679E-02

23.04.2012 09:52:04 3288 151,654 A 1,2

2,635E-02

6,25 1,618E+02 2,651E-02 0,018 0,109

1,959E+02 3,527E-02 0,018 0,111

0,018 0,110

03.04.2012 10:16:56 3288 199,620 A 1,2 6,25

6,25 1,896E+02

0

0,05

0,1

0,15

0,2

0,25

0,3

0,35

0,4

0,45

0,5

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

Pro

tein

[%

]

Arbeitstage

Protein [%] inKartoffelstärke

Protein [%] inWeizenstärke

oC soC soC

sccm

sccm

sccm

0,037 0,211

2,690E+02 5,402E-02 0,037 0,208

0,037 0,210

15.05.2012 11:06:20 3289 148,009 A 1,2 5,71

5,71 2,731E+02

151,690 A 1,2 5,71 2,756E+02 5,570E-02

22.05.2012 10:30:41 3289 151,604 A 1,2

5,506E-02

5,71 2,764E+02 5,592E-02

Dumatherm Nitrogen / Protein Analyser

A 1,2 5,71 5,569E-02

Date Time Protein

factor

Sample name Standard nameWeight

[mg]

24.04.2012 14:20:55

Serial Number :

Software Version:

159

DUMATHERM MANAGER V2.06E Adamek

Submitter:

Operator:

0,65

0,210

0,00 0,00

0,62

0,037

Protein

[%]

Category

0,0372,755E+02 0,212

Peak Area

[mV*s]

Nitrogen

[%]

N Weight

[mg]

46 (Cubic, With Zero )

3289

RSD [%]

150,005

Average

Standard Deviation

0,037 0,210

3289 150,058

Sample Table :

649Reduction Reactor

5

Method_1

27.04.2012 09:31:05 3289

Degassing Oven

Analysis Conditions for Method :

989

Temperatures:

Calibration # :

Combustion Reactor

299

9

EDTA 3385

Times:

Sample Delay

Sample Stop

15.05.2012 11:02:32 A 1,2

He I 195,0

Flow Rates:

O2 399,9

He II 199,9

AutoRun Time

Abb. 7: Weizenstärke mit gleichbleibender Einwaage von ca. 150 mg

Abb. 8: Langzeitstabilität der Kalibration über einen Zeitraum von insgesamt zwei Monaten, gemessen an 24 unterschiedlichen Arbeitstagen.

Abb. 6: Analysenergebnisse zur Kartoffelstärke mit variierenden Einwaagen von 100–200 mg

Figure 6: Analysis results for potato starch with varying sample weights from 100 – 200 mg.

538 Mühle + Mischfutter · 149. Jahrgang · Heft 17 · 6. September 2012

Warum benutzt man Zinnfolie? Zinn verbrennt stark exotherm zu Zinn(IV)oxid. Dabei frei wer­dende Energie unterstützt die vollständige Verbrennung der Probe, sodass die Probe hierbei quasi wie ein Schwamm aufge­saugt wird von Energie. Als Alternative könnte auch stickstoff­freies Wägepapier verwendet werden, allerdings fallen hier große Mengen zusätzlicher Asche an, welche abgetrennt wer­den muss.

2.4 Neuartige Kalibration für die Analytik von Stärke mittels „Dumatherm™“Für die Analytik mittels „Dumatherm™“ wurde eine neue Stär­keapplikation entwickelt. Wesentlicher Baustein zum Erfolg ist hierbei die Kalibration, um die nur sehr kleinen Peakflächen des Stickstoffes aus Stärkeproben quantitativ zu bestimmen.Zur Berechnung der Analysenergebnisse muss eine Kalibration verwendet werden, welche die Signalgrößen der unbekannten Proben vollständig abdeckt. Bei den verwendeten Einwaagen von ca. 150 mg wurden Signalgrößen von ca.150–500 mVs er­reicht. Da die Signalgrößen am absoluten Detektionslimit lie­gen, empfiehlt sich eine mehrstufige Kalibration mit einem Feststoffstandard als auch einem Flüssigstandard. Abb. 5 zeigt eine derartige Kalibrationskurve, erweitert durch Messpunkte nach oben hin mit dem Feststoffstandard EDTA.

Für die Analysen der unbekannten Proben wurden zunächst Standardbedingungen verwendet, d. h. ca. 100–200 mg Einwaa­ge, Sauerstoff­Dosierung mit 400 ml/min, 990 °C Verbrennungs­temperatur. Die folgenden Abb. 6 und 7 zeigen Analysenergebnisse am Bei­spiel der Kartoffelstärke und Weizenstärke.

Als Besonderheit wurde die Kartoffelstärke noch mit stark un­terschiedlichen Einwaagen gemessen, wobei auch hier noch aus­gezeichnete relative Standardabweichungen von <0,6% erhal­ten wurden. Dies zeigt, dass für die Analysen nur 100 mg Ein­ waage notwendig ist. Die erhaltenen Peakflächen sind mit ca. 200 mVs sehr nahe am Detektionslimit von 0,01 mg N absolut.

2.5 Langzeitstabilität der KalibrationDie entwickelte Kalibration hatte eine sehr gute Stabilität. Mit einer Kartoffelstärke­ und einer Weizenstärkeprobe, die über fast 2 Monate an 24 Arbeitstagen kontinuierlich in der täglichen Routine mit analysiert wurden, konnte gezeigt werden, dass die neu ausgearbeitete Kalibration eine hochpräzise Analytik im Bereich von unter 0,25% Protein mit dem „Dumatherm™“ sta­bil möglich macht. Die Ergebnisse sind in Abb. 8 gezeigt.

2.6 FazitDie Stickstoff­/Proteinanalyse nach Dumas wird mit den heute verfügbaren Verbrennungsgeräten eine ernstzunehmende Al­ternative zur Säureaufschlussmethode nach Kjeldahl. Trotz nur geringen Einwaagen ist eine ausgezeichnete Reproduzierbar­keit und Wiederholbarkeit ohne großen Aufwand möglich. Die vorgestellte Kalibrationsstrategie erlaubt die verlässliche Ana­lyse von Stickstoffgehalten am Detektionslimit von 0,01 mg N absolut, bis hin zu hochproteinhaltigen Glutenproben (60–80% Protein). Da die untersuchte Stärke eine ausgezeichnete Homo­genität aufweist, waren sogar beste Analysenergebnisse mit nur 100 mg Probeneinwaage möglich. Hiermit sind die üblichen Pro­bleme im Kjeldahl­Aufschluss mit starkem Schäumen und hohen Chemikalienverbräuchen eliminiert und man kann schnell und sauber eine quantitative Analyse mit einem Verbrennungsgerät erhalten. Nicht ohne Grund etabliert sich deshalb die Verbren­nungsmethode immer mehr als zweite Referenzmethode neben der Kjeldahl­Methode.

3. Literatur1. www.de.wikipedia.org/wiki/St%C3%A4rke

2. Fachverband der Stärke­Industrie e. V.; www.staerkeverband.de/html/zahlen.html

3. DIN EN ISO 3188:1994, Stärken und Stärkederivate­Bestimmung des Stickstoff­gehaltes nach dem Kjeldahl­Verfahren – Volumetrisches Verfahren (ISO 3188:1978); Deutsche Fassung Ausgabedatum: 1994 ­10

4. Souci, S. W., W. Fachmann und H. Kraut: Die Zusammensetzung der Lebens­mittel. Nährwert­Tabellen. – 5. überarbeitete und ergänzte Auflage,; medpharm Scientific Publishers, Stuttgart (1994), CRC PRESS, Tokyo, S. 596 ff

5. Traité Élémentaire de Chimie“, Vol. II (1789), Kapitel VII, S. 493–501

6. www.wikipedia.de (2011)

Stickstoff-/Proteinanalytik

Abb. 5: THAM­ (Tris(hydroxymethyl)­aminomethan) und EDTA­ (Ethylendi­amintetraessigsäure, mit 9,57% N) Kalibration für den „Dumatherm™“ mit insgesamt 20 Messpunkten

oC soC soC

sccm

sccm

sccm

He I 195,3

Flow Rates:

O2 400,1

He II 199,9

AutoRun Time

Sample Table :

649Reduction Reactor

5

Method_1

19.03.2012 11:26:23 3288

Degassing Oven

Analysis Conditions for Method :

989

Temperatures:

Calibration # :

Combustion Reactor

299

9

Analytische Studien

Times:

Sample Delay

Sample Stop

23.03.2012 12:46:49 A 1,2

0,55

0,110

0,000 0,00

0,568

0,018

Protein

[%]

Category

0,0181,256E+02 0,109

Peak Area

[mV*s]

Nitrogen

[%]

N Weight

[mg]

46 (Cubic, With Zero )

3288

RSD [%]

99,487

Average

Standard Deviation

0,018 0,111

3288 149,531

Dumatherm Nitrogen / Protein Analyser

A 1,2 6,25 1,743E-02

Date Time Protein

factor

Sample name Standard nameWeight

[mg]

19.03.2012 11:15:17

Serial Number :

Software Version:

159

DUMATHERM MANAGER V2.06E Adamek

Submitter:

Operator:

151,112 A 1,2 6,25 1,622E+02 2,679E-02

23.04.2012 09:52:04 3288 151,654 A 1,2

2,635E-02

6,25 1,618E+02 2,651E-02 0,018 0,109

1,959E+02 3,527E-02 0,018 0,111

0,018 0,110

03.04.2012 10:16:56 3288 199,620 A 1,2 6,25

6,25 1,896E+02

0

0,05

0,1

0,15

0,2

0,25

0,3

0,35

0,4

0,45

0,5

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

Pro

tein

[%

]

Arbeitstage

Protein [%] inKartoffelstärke

Protein [%] inWeizenstärke

oC soC soC

sccm

sccm

sccm

0,037 0,211

2,690E+02 5,402E-02 0,037 0,208

0,037 0,210

15.05.2012 11:06:20 3289 148,009 A 1,2 5,71

5,71 2,731E+02

151,690 A 1,2 5,71 2,756E+02 5,570E-02

22.05.2012 10:30:41 3289 151,604 A 1,2

5,506E-02

5,71 2,764E+02 5,592E-02

Dumatherm Nitrogen / Protein Analyser

A 1,2 5,71 5,569E-02

Date Time Protein

factor

Sample name Standard nameWeight

[mg]

24.04.2012 14:20:55

Serial Number :

Software Version:

159

DUMATHERM MANAGER V2.06E Adamek

Submitter:

Operator:

0,65

0,210

0,00 0,00

0,62

0,037

Protein

[%]

Category

0,0372,755E+02 0,212

Peak Area

[mV*s]

Nitrogen

[%]

N Weight

[mg]

46 (Cubic, With Zero )

3289

RSD [%]

150,005

Average

Standard Deviation

0,037 0,210

3289 150,058

Sample Table :

649Reduction Reactor

5

Method_1

27.04.2012 09:31:05 3289

Degassing Oven

Analysis Conditions for Method :

989

Temperatures:

Calibration # :

Combustion Reactor

299

9

EDTA 3385

Times:

Sample Delay

Sample Stop

15.05.2012 11:02:32 A 1,2

He I 195,0

Flow Rates:

O2 399,9

He II 199,9

AutoRun Time

Abb. 7: Weizenstärke mit gleichbleibender Einwaage von ca. 150 mg

Abb. 8: Langzeitstabilität der Kalibration über einen Zeitraum von insgesamt zwei Monaten, gemessen an 24 unterschiedlichen Arbeitstagen.

Abb. 6: Analysenergebnisse zur Kartoffelstärke mit variierenden Einwaagen von 100–200 mg

Figure 7: Wheat starch with constant sample weight of approx. 150 mg.

Calibration which completely covers the signal levels of the unknown samples must be used for the calculation of the analysis results. Signal le-vels of approx. 150 - 500 mVs have been achieved for the sample weights of approx. 150 mg. As the signal levels are at the absolute detection limit, multistage calibration with a solid standard as well as with a liquid standard is recommended. Figure 5 shows such a calibration curve, exten-ded by measuring points at the top with the solid standard EDTA.

For the analyses of the unknown samples, stan-dard conditions were initially used, i.e. approx. 100 - 200 mg sample weight, oxygen dosing at 400 ml/min, 990 °C combustion temperature.

The following Figures 6 and 7 show analysis re-sults using the examples of potato starch and wheat starch.

Page 7: Case study: Dumas nitrogen determination Fast and cost ... · Fast and cost-efficient nitrogen / protein analysis for starch 2 1. Introduction and presentation of pro-blem Due to

7

As a special feature, the potato starch was also measured with very different sample weights where excellent relative standard deviations of less than 0.6 % were obtained. This shows that sample weights of only 100 mg are necessary for the analyses. The peak areas obtained of approx. 200 mVs are very close to the absolute detection limit of 0.01 mg N.

2.5 Long-term stability of the calibration

The calibration created had very good stability. Using a potato starch and a wheat starch sam-ple which were continuously analysed in a daily routine over almost 2 months for a total of 24 working days, it could be demonstrated that the newly developed calibration makes stable high-precision analysis with the DUMATHERM® pos-sible in the range of less than 0.25 % protein. The results are shown in Figure 8.

2.6 ConclusionNitrogen / protein analysis according to the Du-mas principle with the combustion apparatus available today is a credible alternative to the acid digestion method according to Kjeldahl. Despite only low sample weights, excellent re-producibility and repeatability are possible. The calibration strategy presented here enables reli-

able analysis of nitrogen content at the absolute detection limit of 0.01 mg N up to gluten samples with high protein content (60-80 % protein). As the examined starch had excellent homogeneity, the best analysis results were even possible with only 100 mg sample weight. This eliminates the usual problems of strong foaming and high che-mical consumption in Kjeldahl digestion and a quantitative analysis can be obtained quickly and cleanly using combustion apparatus.

Therefore, it is not without reason that the com-bustion method is increasingly establishing its-elf as second reference method in addition to Kjeldahl.

Figure 8: Long-term stability of the calibration over a total period of 2 months, measured on 24 different working days.

Literature1. http://de.wikipedia.org/wiki/St%C3%A4rke2. Fachverband der Stärke-Industrie e.V.; http://www.staerkeverband.de/html/zahlen.html3. DIN EN ISO 3188:1994 Starches and derived products - Determination of nitrogen content by the Kjeldahl method - Titrimetric method (ISO 3188:1978); German version issue date: 1994 -10.4. Food Composition and Nutrition Tables; 5th revised and expanded edition, S. W. Souci, W. Fachmann, H. Kraut; medpharm Scientific Publishers, Stuttgart 1994, CRC PRESS, Tokyo pages 596ff5. Traité Élémentaire de Chimie, 1789, Vol. II, Chapter VII, pages 493-501.6. Source: www.wikipedia.de, 2011.

Page 8: Case study: Dumas nitrogen determination Fast and cost ... · Fast and cost-efficient nitrogen / protein analysis for starch 2 1. Introduction and presentation of pro-blem Due to

An excerpt from our product portfolio

COMPLETELY AUTOMATIC HYDROLYSIS HYDROTHERM – automatic acid hydrolysis system for fat de-termination according to Weibull-Stoldt. When combined with SOXTHERM®, HYDROTHERM is an ideal system solution for total fat determination.

COMPLETELY AUTOMATIC FAT EXTRACTION SOXTHERM® – automatic fast extraction system for fat determi-nation

COMPLETELY AUTOMATIC WATER STEAM DISTILLATION VAPODEST® – fast distillation system for Kjeldahl nitrogen/protein determination and steam distillation as sample preparation for further analysis.

COMPLETELY AUTOMATIC NITROGEN ANALYSIS DUMATHERM® – nitrogen/protein determination of solid and liquid samples according to the Dumas combustion method. A fast and convenient alternative to the classic Kjeldahl method for almost all sample matrices.

COMPLETELY AUTOMATIC CRUDE FIBRE EXTRACTION FIBRETHERM® – completely automated processing of the boil ing and filt ration processes for determining crude fibre, ADF and NDF.

Completely automated laboratory analysis systems from C. Gerhardt are highly developed special equipment. They automate recurring analysis processes in accordance with national and international standards and norms. They continuously provide precise and reproducible analysis re-sults quickly, at low cost, economically and highly efficiently.

AUTOMATING STANDARDANALYSES

C. Gerhardt – Quality made in Germany