cambustion scs smoking machine brochure 2018...scs smoking / vaping machine introduction offering...

6
Tobacco, cannabis, heat-not-burn, e-cigarettes, water pipes ISO 3308, 20768, Health Canada, custom regimes Generate aerosol for a range of analytical techniques Capture aerosol & volatiles on filter pad & cold trap Reproduce real world topographies Unlimited puff volume and duration Actuator option for e-cigarettes Lung simulator option Applications: Environmental effects Device repeatability & reproducibility E-cigarette battery effects Volunteer-recorded regimes Mainstream aerosol Sidestream aerosol Ambient aerosol SCS Smoking / Vaping Machine

Upload: others

Post on 14-Jul-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Cambustion SCS Smoking Machine brochure 2018...SCS Smoking / Vaping Machine Introduction Offering unrivalled flexibility and compatibility with a wide range of products and analytical

Tobacco, cannabis, heat-not-burn, e-cigarettes, water pipes

ISO 3308, 20768, Health Canada, custom regimes

Generate aerosol for a range of analytical techniques

Capture aerosol & volatiles on filter pad & cold trap

Reproduce real world topographies

Unlimited puff volume and duration

Actuator option for e-cigarettes

Lung simulator option

Applications:

Environmental effects

Device repeatability & reproducibility

E-cigarette battery effects

Volunteer-recorded regimes

Mainstream aerosol

Sidestream aerosol

Ambient aerosol

SCSSmoking / Vaping Machine

Page 2: Cambustion SCS Smoking Machine brochure 2018...SCS Smoking / Vaping Machine Introduction Offering unrivalled flexibility and compatibility with a wide range of products and analytical

SCS Smoking / Vaping MachineIntroductionOffering unrivalled flexibility and compatibility with a wide range of products and analytical techniques,

Cambustion's SCS has been adopted worldwide for research and development of combustible, heat-not-burn

and e-cigarettes since 2006.

Conventional MachinesThe linear piston drives used in many smoking and vaping machines place limits on total puff volume, repeat

frequency and mean that the aerosol is only available after a non-negligible aging period in the syringe chamber.

Reproducible profiles are somewhat limited by the capabilities of the drive system.

The SCS AdvantageThe SCS drives flow through the cigarette using

pressure control, with real-time feedback and

control of the cigarette flow. Standard regimes such

as ISO 3308, 20768 and Health Canada can be

reproduced, while users can specify their own regimes

including using real-world topography.

Constant Volume Sampling mode

For combustible tobacco smoke measurements where the aerosol is very high concentration and non-volatile,

the SCS may be operated in a Constant Volume Sampling (CVS) mode, which applies dilution in the Puff

Head. This reduces coalescence effects, enabling accurate measurements of particle size distributions, while

avoiding saturating downstream instruments. CVS allows real-time calculation of mass per second with online

measurement techniques, and real-time integration for total mass, even with real-world smoking profiles when

the cigarette flow rate changes rapidly.

There are no limits to puff duration or total volume and thousands of puff repeats can be made without the need

for cleaning.

Time (s)

Cig

aret

te F

low

Real Human

Health Canada

ISO 3308

ISO 20768

Variable valve

Δp

Blower

Vac Pump

Main Control UnitHEPA

Puff Head

ThermocoupleOrif ice f low meterPinch valve

CVS Configuration

Page 3: Cambustion SCS Smoking Machine brochure 2018...SCS Smoking / Vaping Machine Introduction Offering unrivalled flexibility and compatibility with a wide range of products and analytical

Puff Inhalation Exhalation mode

For volatile aerosol such as cannabis or electronic cigarette aerosol, dilution of the aerosol leads to evaporation

on a sub-second timescale. In the Puff Inhalation Exhalation (PIE) configuration, the SCS may be operated with

a lung bag. Undiluted aerosol generated according to the regime is inhaled through the Puff Head from the

device, and fed into a flexible lung bag. Aerosol may be exhaled from the lung bag via the Breath Head into

the measurement device.

Alternatively, the PIE SCS can dilute sampled aerosol by an inhalation of air through the Breath Head, allowing

dilution in the lung bag, before exhaling the diluted aerosol through the Breath Head for analysis, more

accurately replicating human smoking of combustibles.

The lung bag may be bypassed to allow unlimited puff volume if required. Options to heat and humidify the

lung bag are available.

E-cigarette Button Pusher

An option is available for repeatable and programmable actuation of manually triggered e-cigarettes.

Analytical Techniques

SCS generated aerosol may be sent to a range of online analytical devices, including fast particle sizers such

as Cambustion's DMS500, Gas Chromatographs (GC) from a range of manufacturers, Malvern Spraytec and

more.

Sample CollectionSamples may be extracted using gas syringes for injection into a wide range of measurement instruments.

Δp

Variable valve

Δp

P

Δp

Blower

Vac Pump

Main Control UnitHEPA

Puff Head

Breath HeadThermocoupleOrifice flowmeter

Lung bag Pinch valve

SCS Smoking / Vaping Machine

Puff Inhalation Exhalation Configuration

Page 4: Cambustion SCS Smoking Machine brochure 2018...SCS Smoking / Vaping Machine Introduction Offering unrivalled flexibility and compatibility with a wide range of products and analytical

An optional stainless steel filter holder allows material

to be collected onto a filter pad, for mass or chemical

analysis. A cold trap may be connected inline, to allow

sampling of both volatile and aerosol components.

Comprehensive Data CollectionThe SCS records operating parameters, allowing

instant access to information such as flow rate,

resistance to draw and actual puff volumes. Every

event is recorded in full detail by default, for easy

traceability even months afterwards.

Accurate Flow and VolumeThe SCS uses orifice ∆P measurement flow meters

to measure the true flow delivered, and an internal

reference flowmeter is used to calibrate the orifice flow

meters. Users can easily connect external reference

flow meters for validation and traceability.

Low MaintenanceSyringe free design reduces maintenance requirements

compared with conventional machines. All parts

exposed to aerosol are user cleanable, with long

intervals between maintenance.

Easy to moveThe SCS components easily fit on a small trolley,

allowing them to be moved between different areas

of the laboratory.

Low Temperature TestingThe SCS sampling head supports sub 0°C testing for

e-cigarette development.

SCS Specifications:

Parameter Specification

Maximum puff flow rate 200 cc/s

Maximum puff duration Unlimited

Maximum puff volume Unlimited

Puff repeats Unlimited

Puff frequency Infinitely variable

Vaping angle/orientation Unrestricted

Error in puff volume <5 %

Power supply(market specific)

100, 110 or 220 V AC, 50/60 Hz

Power consumption(including PC)

300 W

System total weight (excluding PC)

Typically 17 Kg

Smoking / vaping profiles ISO 3308, ISO 20768Health Canada, Custom

Profile following Maximum 25 Hz

Sampling headtemperature

Optionally controllable, maximum 50 ° C

Data logging Routine data logging of flow rate, resistance to draw and operational parameters

Compatible products Combustible cigarettes, electronic cigarettes, water pipes

Number of channels 1 (single cigarette or product)

Product size standard 5 – 10 mm diameter(1/8th BSP thread allows connection of other devices)

Combustible product ignition

Manual, by user. (Countdown displayed on screen)

e-cigarette trigger Optional button pusher accessory

SCS Smoking / Vaping MachineSoftware

Intuitive Windows software allows even inexperienced

users to specify profiles, run smoking/vaping routines

and check SCS calibration. Filter Holder

Cold Trap

Page 5: Cambustion SCS Smoking Machine brochure 2018...SCS Smoking / Vaping Machine Introduction Offering unrivalled flexibility and compatibility with a wide range of products and analytical

CAMBUSTION DMS50 Fast Particulate Size Spectrometer

Nucleation mode burst produced by transient lift-off in real world driving: measured by DMS50 in the plume of the vehicle.

Particle Size and Mass MeasurementWith DMS500

Cambustion’s DMS500 particle sizer offers 10 Hz

particle size/number distributions between 5 and

2,500 nm. The T 10-90% response time of 200 ms

allows measurements within a single puff, as well as

identifying puff to puff variability.

The DMS500 can output real-time particle mass rate

(mg/s), together with integrated mass (mg) across a

series of puffs.

This makes the combination of SCS and DMS500

ideally suited to studying a wide range of products.

The graph above shows changes in the average

spectra from successive 2 second puffs during the

ISO profile. Increases in concentration with each puff

are notable, as is the evolution of the aerosol size.

Operating the SCS in Constant Volume Sampling

mode allows calculation of real-time mass emissions

rates (mg/s) from the size distributions, and integration

to give total mass, allowing unprecedented insights

into product behaviour.

The DMS500's rapid time response allows

identification of sub-second transients, for example

lighting transients in the case of a water pipe, as

illustrated in the graph below.

Electronic cigarettes can produce a more complex

aerosol, due to the highly volatile nature and range

of boiling points involved. The graph below shows

size distributions from a single two second puff on an

e-cigarette.

At the start of the puff very small particles (<50 nm)

dominate. As the puff continues, a larger mode

of ~300 nm appears, and dominates both mass

and number. The different size distributions may

be understood in terms of different volatility of

carrier (PG/VG) and flavourings. When volatile

components recondense, in the absence of other

particles homogeneous nucleation occurs, leading

to the very small particles. Once the device has

warmed up further less volatile components (PG/VG)

also condense, and the more volatile compounds

undergo heterogeneous nucleation.

DMS Fast Particle Sizer for SCS

0.00E+00

5.00E+07

1.00E+08

1.50E+08

2.00E+08

2.50E+08

1.00 10.0 100 1000

dN

/dlo

Dp/c

c

Particle Diameter (nm)

Puff by Puff Particle Size Spectra from a Tobacco Cigarette

Puff 1Puff 2Puff 3Puff 4Puff 5Puff 6Puff 7Puff 8

ISO Test

2222.6

23.226.3

26.9

0.0E+00

4.0E+07

8.0E+07

1.2E+08

1.6E+08

5 9 15 27 49 87 154 274 487 866Time (s)

dN/d

logD

p /c

cParticle Diameter (nm)

Size distributions during e-cig 2 second puff

0

20

40

60

80

100

120

140

160

0

500

1,000

1,500

2,000

2,500

3,000

0.0 50.0 100.0 150.0 200.0 250.0

Instantaneous and integrated mass from user triggered e-cigarette

Time (s)

Integrated Mass (µg) Instantaneous Mass (µg/s)

Inst

anta

neou

s M

ass

(µg/

s)

Inte

grat

ed M

ass

(µg)

Page 6: Cambustion SCS Smoking Machine brochure 2018...SCS Smoking / Vaping Machine Introduction Offering unrivalled flexibility and compatibility with a wide range of products and analytical

DMS Fast Particle Sizer for SCS

All specifications subject to change without notice.

Cambustion SCS and DMS500 products are available for both rental and purchase.

For more information or to discuss a specific application, please contact:

[email protected] tel: +44 1223 210250www.cambustion.com/scs fax: +44 1223 210190 N. America tel: +1 (800) 416 9304

ISO 9001:2015 cert.# FS 73100

Cambustion Ltd,J6 The Paddocks,347 Cherry Hinton Road,Cambridge CB1 8DHUnited Kingdom

NOV 2018

DMS500 Specifications:

Size range 5 nm - 1,000 nm(option) 5 nm - 2,500 nm

Data rate Max 10 Hz

Time Response T10-90% 200 ms

Spectral elements 16/decade

Power supply 100-115 or 220-240 VAC1500W

Main unit weight 80 kg

External Pump Weight 44 kg

Calibration NIST traceable for size & number

Within a single puff of a tobacco cigarette, the

temperature of the combustion zone and of the

tobacco rod is increasingly rapidly during the puff.

This leads to increasing size of particles as the puff

progresses, a process which conventional smoking

machines and particle size instruments can not

resolve. This process is easily visualised using a

contour plot.

dN/dlogDp/cc1.E+6 4.E+6 1.6E+7 6.3E+7 2.5E+8 1.E+9

1000

500

200

100

50

20

10

5

0 20 40 60 80 100 120 140 160

6.00E+01

5.00E+01

4.00E+01

3.00E+01

2.00E+01

1.00E+01

0.00E+00Parti

cle

Dia

met

er (n

m)

1000

500

200

100

50

20

10

515 16 17 18 19 20

5.00E+01

4.00E+01

3.00E+01

2.00E+01

1.00E+01

0.00E+00

Time (s)

Tobacco cigarette Health Canada puff by puff particle size dataPa

rticl

e D

iam

eter

(nm

)

Orif

ice

FLow

(cc/

s)O

rific

e FL

ow (c

c/s)

Time (s)

Tobacco cigarette Health Canada single puff particle size data