Calibration Methods Introduction
1.) Graphs are critical to understanding quantitative relationships One parameter or observable varies in a predictable manner in
relationship to changes in a second parameter 2.) Calibration curve: graph showing the analytical response as a
function of the known quantity of analyte Necessary to interpret response for unknown quantities
Time-dependent measurements of drugs and metabolites in urine samples
Generally desirable to graph data to generate a straight line
Calibration Methods Finding the “Best” Straight Line
1.) Many analytical methods generate calibration curves that are linear or near linear in nature
(i) Equation of Line:
where: x = independent variable
y = dependent variable
m = slope b = y-intercept
bmxy
mx
yslope
Calibration Methods Finding the “Best” Straight Line
2.) Determining the Best fit to the Experimental Data(i) Method of Linear Least Squares is used to determine the best values for
“m” (slope) and “b” (y-intercept) given a set of x and y values Minimize vertical deviation between points and line
Use square of the deviations deviation irrespective of sign
)b)x(my()yy(d iiii
2ii
2i
2i )b)x(my()yy(d
Calibration Methods Finding the “Best” Straight Line
4.) Goodness of the Fit (i) R2: compares the sums of the variations for the y-values to the best-fit line
relative to the variations to a horizontal line. R2 x 100: percent of the variation of the y-variable that is explained by
the variation of the x-variable. A perfect fit has an R2 = 1; no relationship for R2 ≈ 0
R2=0.9952
99.5% of the y-variation is due to the x-variation
R2=0.5298
53.0% of the y-variation is due to the x-variationWhat is the other 47% caused by?
Very weak to no relationship
Strong direct relationship
R2 based on these relative differencesSummed for each point
Calibration Methods Calibration Curve
1.) Calibration curve: shows a response of an analytical method to known quantities of analyte
Procedure:a) Prepare known samples of analyte covering
convenient range of concentrations.
b) Measure the response of the analytical procedure.
c) Subtract average response of blank (no analyte).
d) Make graph of corrected response versus concentration.
e) Determine best straight line.
Calibration Methods Calibration Curve
2.) Using a Calibration Curve
Prefer calibration with a linear response- analytical signal proportional to the quantity of analyte
Linear range- analyte concentration range over which the response is proportional to concentration
Dynamic range- concentration range over which there is a measurable response to analyte
Additional analyte does not result in an increase in response
Calibration Methods Calibration Curve
3.) Impact of “Bad” Data Points Identification of erroneous data point.
- compare points to the best-fit line- compare value to duplicate measures
Omit “bad” points if much larger than average ranges and not reproducible.- “bad” data points can skew the best-fit line and distort the accurate interpretation of data.
Remove “bad” point
Improve fit and accuracy of m and b
y=0.16x + 0.12 R2=0.53261 y=0.091x + 0.11 R2=0.99518
Calibration Methods Calibration Curve
4.) Determining Unknown Values from Calibration Curves(i) Knowing the values of “m” and “b” allow the value of x to be determined once the experimentally y value is known.(ii) Know the standard deviation of m & b, the uncertainty of the determined x-
value can also be calculated
Calibration Methods Calibration Curve
4.) Determining Unknown Values from Calibration Curves(iii) Example:
The amount of protein in a sample is measured by the samples absorbance of light at a given wavelength. Using standards, a best fit line of absorbance vs. g protein gave the following parameters:
m = 0.01630 sm = 0.00022
b = 0.1040 sb = 0.0026
An unknown sample has an absorbance of 0.246 ± 0.0059. What is the amount of protein in the sample?
Calibration Methods Calibration Curve
5.) Limitations in a Calibration Curve(iv) Limited application of calibration curve to determine an unknown.
- Limited to linear range of curve- Limited to range of experimentally determined response for known analyte concentrations
Unreliable determination of analyte concentration
Uncertainty increases further from experimental points
Calibration Methods
syy blankdl 3Signal detection limit:
blanksamplecs yyy Corrected signal:
Detection limit:
m
sc
3
Calibration Methods Calibration Curve
6.) Limitations in a Calibration Curve(vi) Example:
Low concentrations of Ni-EDTA near the detection limit gave the following counts in a mass spectral measurement: 175, 104, 164, 193, 131, 189, 155, 133, 151, 176. Ten measurements of a blank had a mean of 45 counts. A sample containing 1.00 M Ni-EDTA gave 1,797 counts. Estimate the detection limit for Ni-EDTA
Calibration Methods Standard Addition
1.) Protocol to Determine the Quantity of an Unknown(i) Known quantities of an analyte are added to the unknown
- known and unknown are the same analyte- increase in analytical signal is related to the total quantity of the analyte - requires a linear response to analyte
(ii) Very useful for complex mixtures- compensates for matrix effect change in analytical signal caused by
anything else than the analyte of interest.
(iii) Procedure:
(a) place known volume of unknown sample in multiple flasks
Calibration Methods Standard Addition
1.) Protocol to Determine the Quantity of an Unknown(iii) Procedure:
(b) add different (increasing) volume of known standard to each unknown sample
(c) fill each flask to a constant, known volume
XS
X
ff
i
I
I
XS
X
Calibration Methods Standard Addition
1.) Protocol to Determine the Quantity of an Unknown(iii) Procedure:
(d) Measure an analytical response for each sample - signal is directly proportional to analyte
concentration
solutionfinalfromsignal
solutioninitialfromsignal
solutionfinalindardtansplusanalyteofionConcentrat
solutioninitialinanalyteofionConcentrat
Standard addition equation:
V
VSS
V
VXX S
ifo
if
dardtansofvolumeaddedV,volumeinitialunknownV,VVV SoSo Total volume (V):
X-intercept (y=0) yields which is used to calculate from:
Calibration Methods Standard Addition
1.) Protocol to Determine the Quantity of an Unknown(iii) Procedure:
(f) Plot signals as a function of the added known analyte concentration and determine the best-fit line.
ofi V
VXX
fX
iX
Calibration Methods Standard Addition
1.) Protocol to Determine the Quantity of an Unknown(iii) Example:
Tooth enamel consists mainly of the mineral calcium hydroxyapatite, Ca10(PO4)6(OH)2. Trace elements in teeth of archaeological specimens provide anthropologists with clues about diet and disease of ancient people. Students at Hamline University measured strontium in enamel from extracted wisdom teeth by atomic absorption spectroscopy. Solutions with a constant total volume of 10.0 mL contained 0.750 mg of dissolved tooth enamel plus variable concentrations of added Sr. Find the concentration of Sr.
Added Sr (ng/mL = ppb) Signal (arbitrary units)
0 28.0
2.50 34.3
5.00 42.8
7.50 51.5
10.00 58.6
Calibration Methods Internal Standards
1.) Known amount of a compound, different from analyte, added to the unknown.(i) Signal from unknown analyte is compared against signal from internal standard
Relative signal intensity is proportional to concentration of unknown- Valuable for samples/instruments where response varies between runs- Calibration curves only accurate under conditions curve obtained - relative response between unknown and standard are constant
Widely used in chromatography Useful if sample is lost prior to analysis
Area under curve proportional to concentration of unknown (x) and standard (s)
dardtansofionConcentrat
signaldardtansofareaF
analyteofionConcentrat
signalanalyteofArea
S
AF
X
A Sx
Calibration Methods Internal Standards
1.) Example:
A solution containing 3.47 mM X (analyte) and 1.72 mM S (standard) gave peak areas of 3,473 and 10,222, respectively, in a chromatographic analysis. Then 1.00 mL of 8.47 mM S was added to 5.00 mL of unknown X, and the mixture was diluted to 10.0 mL. The solution gave peak areas of 5,428 and 4,431 for X and S, respectively
(a)Calculate the response factor for the analyte(b)Find the concentration of S (mM) in the 10.0 mL of mixed solution.(c)Find the concentration of X (mM) in the 10.0 mL of mixed solution.(d)Find the concnetration of X in the original unknown.