Folie 1 MERIS Validation Team Meeting, JRC Ispra, 9.3.2011 New calibration facility at DLR Oberpfaffenhofen P. Gege, DLR, Remote Sensing Technology Institute

Upload: others

Post on 11-Jul-2020

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Folie 1

MERIS Validation Team Meeting, JRC Ispra, 9.3.2011

New calibration facility at DLR Oberpfaffenhofen

P. Gege, DLR, Remote Sensing Technology Institute

Page 2: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 2

DLR – German Aerospace Center

6700 employees across 33 institutes and facilities at

13 sites.

Offices in Brussels, Paris and Washington.

Cologne

Oberpfaffenhofen

Braunschweig

Goettingen

Berlin

Bonn

Neustrelitz

Weilheim

Bremen Trauen

Lampoldshausen

Hamburg

Stuttgart

Page 3: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 3

Funded partly by ESA to establish Calibration Home Base (CHB) for APEXDesigned for hyperspectral sensors similar to APEX

Mass: 170 kg (excl. adapter)

-range: 380–2500 nm

Bandwidth: 5–10 nm

IFOV: 0.48 mrad

FOV: ±14°Operational since 2007.

Introduction

Page 4: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 4

Close to airfield of DLR OberpfaffenhofenSuited for bulky and heavy instruments up to 500 kg (incl. adapter)Sensor in same position as in aircraftSensor stable on vibrationally isolated calibration bench

Spectral calibration

Geometric calibration

Premises

Page 5: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 5

1. Pillar bearing instrument + adapter2. Folding mirror3. Assembly for geometric measurem.4. Assembly for spectral measurem.

Folding mirror concept

12

4

Page 6: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 6

Large integrating sphere

1.65 m

Aperture 55 x 40 cm²

Inhomogeneity < 0.5 % rms

18 lamps

Various radiance levels (57 – 1524 W m-2)

Flat-field measurements

1. Frame2. Small integrating sphere3. Large integrating sphere

123

AISA (Kuhlbach 2008)

3

Page 7: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 7

Small integrating sphere

0.50 m

Aperture 4 x 20 cm²

Traceable to PTB

Uncertainty (k=2) 1 % in VIS

Radiometric calibration

1. Frame2. Small integrating sphere3. Large integrating sphere

123

400 800 1200 1600 2000 2400Wavelength (nm)

00100100200200300300400400500500600600700700

Rad

ianc

e (m

W m

-2 n

m-1

sr-1

) 0055

1015

U (%

)

1 %1 %

2

Page 8: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 8

Absolute radiometric calibration of radiance sources

Relative radiance1. Calibrated halogen lamp2. Calibrated diffuser3. Spectrometer

Absolute radiance4. 5 Filter radiometers

Uncertainty goal

1

= 1.5 % at 0.35-1.7 µm (2011)

1

= 2.5 % at 1.7-2.5 µm (2012)

1 2 3New

4

3x VIS2x NIR

Page 9: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 9

1. Monochromator Oriel MS257

Range: 0.38–14 µm using 7 gratings

Uncertainty: ± 0.1 nm

Spectral bandwidth: > 0.1 nm (depending on grating and slit width)

2. Parabolic mirror

f = 119 mm

Beam divergence ~ 0.8 x 8 mrad²

Beam cross section ~ 3 x 4 cm²

Spectral measurements

Monochromator

Sensor

2 1

Spectral response function of 3 ROSIS channels (Harder 2008)

Page 10: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 10

New

Spectral measurements: Tunable laser

Specifications

Range: 0.4 – 2.5 µm

Resolution: 3 – 7 cm-1

Repetition rate: 10 HzAdvantages

High energy

No sensor alignment

All pixels simultaneousDisadvantages

High safety requirements

Fix bandwidth

Pulsed (not suited for scanners)

Page 11: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 11

Sensor

CollimatorLamp Slit

1 2

3

Geometric measurements

1. Slit wheel

3 horizontal + 3 vertical slits

Widths: 50, 100, 1000 µm

2. Collimator

f = 750 mm

Divergences: 0.067, 0.13, 1.3 mrad

Beam cross section:

12 cm

3. Angle stepping

Across track: 0.0017 mrad

Along track: 0.0076 mrad

LSFs of AISA pixel no. 192(adapted from Suhr 2008)

-0.4 -0.2 0 0.2 0.4Illumination angle (o)

0

0.2

0.4

0.6

0.8

1

Nor

mal

ized

sig

nal

channel 50channel 50channel 100channel 100channel 150channel 150channel 200channel 200channel 250channel 250

LSF | 12.3.2009

Page 12: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 12

Auxiliary measurements

Detector linearity

Small sphere and neutral density filtersSpectral stray light

Monochromator

Small sphere and bandpass filters

New: Tunable laserSpatial stray light

From inside FOV: set-up for geometric measurements (LSF)

From outside FOV: large sphere and reflectance targets

Polarisation

3 linear polarisers 0.47 – 2.5 µm

Page 13: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 13

Computer control

CHB Slave

Controls CHB hardware components

CHB Master

Measurement concept

Commands sensor

Commands CHB slave

Page 14: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 14

Service

CHB infrastructure accessible to customers on request

DLR contributions:Consultancy (calibration concept, campaign planing)Support for system adaptation (mechanics, software control)CHB mobilisation, wrap-upOperation of CHB subsystems during campaignTbd: scientific support (data evaluation, interpretation)

Timeline:Preparation phase (standard): 2-3 monthsCampaign duration (standard): 1 week +

Cost: 10 k€+

Page 15: Calibration facility at DLR OberpfaffenhofenPG).pdfCalibration Facility DLR > P. Gege > 8 March 2011 Folie 11 Sensor Lamp Slit Collimator 1 2 3 Geometric measurements 1. Slit wheel

Calibration Facility DLR > P. Gege > 8 March 2011

Folie 15

Summary

Facility for characterisation of airborne imaging spectrometers and field spectrometers

Bulky and heavy instruments up to 500 kg

Spectral range: 380 – 2500 nm

Radiometry

Spectroscopy

GeometryContinuously upgraded

Tunable laser

Transfer radiometer

P. Gege et al. (2009): Calibration facility for airborne imaging spectrometers. ISPRS Journal of Photogrammetry & Remote Sensing 64, 387-397. doi:10.1016/j.isprsjprs.2009.01.006

CHB online: http://www.opairs.aero/chb_en.html