bsm-1733/1753/1763 bedside monitor -...

15
This technical data may be revised or replaced by Nihon Kohden at any time without notice. Some products may not be available in your country. TD. BSM-1700 BSM-1733/1753/1763 Bedside Monitor Technical Data

Upload: vuonghanh

Post on 18-Mar-2018

395 views

Category:

Documents


8 download

TRANSCRIPT

Page 1: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

This technical data may be revised or replaced by Nihon Kohden at any time without notice.

Some products may not be available in your country.

TD. BSM-1700

BSM-1733/1753/1763

Bedside Monitor

Technical Data

Page 2: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

2

FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters for single adult, pediatric and neonatal patients. The monitors are to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians.

Smart cable concept (MULTI connectors) Smart cable technology gives you high flexibility in the selection of parameters without traditional modules. When you plug a sensor cable into a MULTI connector, the monitor automatically detects the type of parameter and starts measuring. Either IBP, CO, CO2, BIS or SpO2-2 can be measured per MULTI connector. For the basic parameters, ECG, Respiration, SpO2, NIBP and Temperature can be measured with dedicated connectors.

High resolution, TFT color display The monitor has a high resolution, 5,7 inch TFT color LCD. Max. 9 waveforms* can be displayed. *12 traces for 12-lead ECG.

Flexible monitor control A touch panel screen provides easy and quick operation. Active parameters appear automatically on the screen with a variety of screen configurations.

Alarm indicator An alarm indicator at the top of the monitor can be seen clearly from a distance from all angles. It lights in red or yellow depending on the severity of the alarm. A green lamp blinks in synchronization with the QRS when configured.

Upgraded arrhythmia detection Upgraded arrhythmia detection allows you to analyze multi-channel ECG. And the special software ec1 with the multi-template matching method reduces false alarms and assures high quality detection. Up to 23 itemes of arrhythmia can be detected. Up to 32,768 (8,192)* arrhythmia episodes can be stored.

*When transport is OFF (When transport is ON)

12 lead ECG simultaneous display using 10 electrodes

10 electrodes provide the same quality simultaneous 12 lead ECG as a multi-channel electrocardiograph on a screen. This makes it easier to recognize multi-focus VPC, ST change and ventricular hypertrophy.

12 lead ECG interpretation 12 lead ECG interpretation is available. The 12 lead ECG with 10 electrodes provides 12 lead ECG interpretation results with identical quality to Nihon Kohden electrocardiographs with

ECAPS 12C program.

ST recall Up to 72 hours, 4,320 files (24 hours, 1,440 files) *ST events can be saved. All Lead ST Segment analysis is available. You can edit an ST measurement point after reviewing it.

*When transport is OFF (When transport is ON)

Alarm history You can review up to 32,768 (8,192)* files for the numeric data of alarm occurrences from the alarm history table. You can also review the wave by jumping to the full disclosure screen from the alarm history table.

*When transport is OFF (When transport is ON)

72 hours, 5 traces (24 hours, 5 traces)* full disclosure

Up to 4 parameters full disclosure waveform is displayed. Full disclosure makes it easy to search for waveforms.

*When transport is OFF (When transport is ON)

Trend graph/trend table Up to 72 hours (24 hours)*of trend graphs of all parameters, and trend table and NIBP table, 2,048 (512)* files can be stored and reviewed.

*When transport is OFF (When transport is ON)

Short trend The latest 25 minutes parameter data can be displayed as a trendgraph on the home screen. This short trandgraph can be dragged left or right by touch panel operation. OCRG short trend is available.

Mainst ream EtCO 2 fo r both in tubated & non-intubated patient (cap-ONE, TG-920P)

With a unique and very small (about 10 g) mainstream cap-ONE CO2 sensor (TG-920P), smooth, fast (within 5 seconds) and reliable readings of EtCO2 & respiration-rate is available.

PWTT trigger NIBP measurement PWTT is non-invasively measured by continuous monitoring of ECG and SpO2. If a sudden critical circulation change occurs between periodic NIBP measurements, PWTT detects it and trigger NIBP measurement to confirm it.

esCCO With optional QP-172P, esCCO measurement is available. The esCCO is a new technology to calculate cardiac output continuously. It uses PWTT (Pulse Wave Transit Time) which is obtained from the pulse oximetry and ECG-signals from each cycle of the ECG and peripheral pulse wave. It can continuously display CCO, CCI, SV, SVI.

Page 3: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

3

iNIBP iNIBP is Nihon Kohden’s original algorithm to measure NIBP during inflation. It realizes fast and painless NIBP measurement. YAWARA CUFF 2, Nihon Kohden’s original cuffs, are designed to subcutaneous bleeding for patient comfort and to reduce noise for more accurate measurement. Life Scope PT uses Linear inflation method NIBP.

CVP-ET End tidal mean CVP measurement is available. The CVP-ET uses the CO2 signal to calculate the end tidal mean CVP. In this mode, the mean CVP value is stable and less affected by respiration. To measure CVP-ET, mainstream

CO2 measurement (TG-920P, TG-950P or TG-970P) is necessary.

LS-NET (Life Scope Networking) You can connect with Nihon Kohden central monitors and bedside monitors by LS-NET (Life Scope Networking with Ethernet LAN) with AC cradle, SC-170R

Wireless networking By attaching optional ZS-900PK transmitter to the monitor, telemetry networking is available.

Interbed When Life Scope series bedside monitors are connected to each other by LS-NET, you can exchange and view data of other bedside monitors without a need of the central monitor. Both individual display and up to 9 beds display are available.

SYSTEM COMPOSITION

Battery SB-170P

Bedside Monitor

BSM-1733/1753/1763

Transmitter ZS-900P*1

AC Cradle SC-170P

Data Acquisition unit JA-690P/JA-694P

WEP-5200

Telemetry system

LS-NET Multi-patient Receiver

ORG-9100

LS-NET

CNS-6200/9600 series

CNS-6200/9600 series

Central monitor

Central monitor

BSM-6000 series

*1: DI-170P Holder required

Interface QI-600P

ECG/BP Output

Network Printer

Page 4: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

4

SPECIFICATIONS Measuring Parameters

ECG (3/6/10 lead), respiration (impedance method), SpO2, NIBP, IBP*1(up to 3ch), temperature (up to 2ch), etCO2, FiCO2, cardiac output, SpO2-2, BIS *1: MULTI connector parameters. 3 MULTI

connectors are available.

Display Display size: 5.7-inch color TFT Waveform display mode: Non-fade fixed Viewing area: 115.2x 86.4 mm Resolution: 640 x 480 dots Maximum number of traces: 9 traces (12 traces for 12-lead ECG measurement) Sweep speed: 50, 25, 12.5 or 6.25 mm/s Respiration/CO 2 sweep speed: 25, 12.5, 6.25, 1.56 mm/s Waveform display colors: 9 traces (12 traces for 12-lead ECG measurement) Numeric display colors: 12, selectable Displayed waveforms: ECG (maximum 12 traces), respiration, IBP (maximum 3 traces), SpO2 pulse wave, CO2, CO thermodilution curve, BIS Numerical data display: HR, VPC, ST, RR, NIBP (systolic, diastolic, MAP), IBP (systolic, diastolic, mean), SpO2, PR, TEMP, CO, CI, Ti, Tb, CO2 (I), CO2 (E), BIS, SEF95, SR, EMG, SQI, esCCO, esCCI, esSV, esSVI Synchronization mark: Heart rate sync mark, pulse rate sync mark, respiratory sync mark Alarm configurable start-up values : Master Alarm Neonate, Master Alarm Child, Master Alarm Adult

Language Available language: Czech, Danish, English, Finnish, French, German, Greek, Hungarian, Italian, Dutch, Norwegian, Polish, Portuguese, Polish, Rumanian, Russian, Slovak, Spanish, Swedish and Turkish

Sound Type of sound: Alarm, synchronization (HR, IBP, SpO2), click, NIBP completion sound Alarm sound: 4 types (pip, ping, bing bong, ding ding) or IEC standard Synchronization sound: Variable pitch for IBP and SpO2

Alarm Alarm items: Upper/lower limits, apnea, arrhythmia, connector disconnection, noise, electrode off, waveform detecting, probe off, cuff/hose check, sensor check, low battery, operating environment Vital sign alarms: HR, PR, RR, APNEA, CO2, SpO2, NIBP, BIS, IBP, VPC, ST, IBP, TEMP

Arrhythmia alarms: ASYSTOLE, VF, VT, V BRADY, EXT TACHY, EXT BRADY, SV TACHY, VPC RUN, TACHYCARDIA, BRADYCARDIA, COUPLET, EARLY VPC, MULTIFORM, V RHYTHM, PAUSE, BIGEMINY, TRIGEMINY, VPC, IRREGULAR RR, PACER NON-CAPTURE, PROLONGED RR, NO PACER PULSER Alarm types: Crisis (red, blinking), Warning (yellow, blinking), Advisory (yellow or blue, lit), Message Alarm indication: Alarm indicator, highlighted numeric data/ message, alarm sound Alarm suspend: Provided (for 1, 2, 3 min, Off) When connected to a central, this alarm can be adjusted from the central.

Trend graph Trend parameters: HR, PR, VPC, RR, NIBP, CO2(E), CO2(I), SpO2, SpO2-2, PI, Tb, ST-I to ST-V6, IBP, Temperature, BIS, SQI(BIS), None Storage capacity: 72 hours (24 hours)*

*When transport is OFF (When transport is ON) Up to 2 parameters can be user selected for 2separate graphical windows.

Vital Signs Table Parameters: HR, PR, VPC, RR, NIBP, CO2(E), CO2(I), SpO2, SpO2-2, PI, Tb, ST-I to ST-V6, IBP, Temperature, BIS, SQI(BIS), NONE

Storage capacity; Vital sign data: 72 hours (24 hours)* NIBP: 2,048 files (512 files)*

*When transport is OFF (When transport is ON) Interval for periodic list: 1, 5, 10, 15, 30 or 60 min. Interval for NIBP list: at NIBP measurement

HEMO Trend Number of files: 2,048 files (512 files)*

*When transport is OFF (When transport is ON) Parameters: HR/RR, CO, CI, CCO, CCI, ART-MEAN, PAP-MEAN, PCWP, PCWP time, CVP, SV, SVI, LVSW, LVSWI, SVR, SVRI, PVR, PVRI, PVSW, PVSWI,

Recall Item; Arrhythmia: 32, 768 files (8,192 files)*

*When transport is OFF (When transport is ON) Arrhythmia recalled: ASYSTOLE, VF, VT, V BRADY, EXT TACHY, EXT BRADY, SV TACHY, VPC RUN, TA-CHYCARDIA, BRADYCARDIA, COUPLET, EARLY VPC, MULTIFORM, V RHYTHM, PAUSE, BIGEMINY, TRI-GEMINY, VPC, IRREGULAR RR, PACER NON-CAPTURE, PROLONGED RR, NO PACER PULSE, ALL

Alarm History Item: Vital sign alarms, Arrhythmia alarms, Technical alarms and Operational alarms Number of files: 32, 768 files (8,192 files)*

*When transport is OFF (When transport is ON)

Page 5: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

5

Full Disclosure Storage capacity: 72 hours (24 hours)*

*When transport is OFF (When transport is ON) Number of storing parameters: 5 (only 4 waveforms can be displayed) Wave duration: 60 s, up to 4 parameter waveform Actual size ECG waveform display: Provided

ST Recall Item: ST level waveforms of all monitoring ECG Storage capacity: 72 hours, 4,320 files (24 hours, 1,440 files)*

*When transport is OFF (When transport is ON)

12 LEAD Item: 12 lead analysis result data measurement, interpretation, diagnostic Number of files: 24 files (6 files)*

*When transport is OFF (When transport is ON)

Interbed Waveforms and data of other Life Scope series bedside monitor connected by Ethernet can be reviewed without the need of a central station. Alarm silencing from interbed window is available. Display items for personal display: Waveform: 2 traces Numeric data: ECG, SpO2, RR, CO2, NIBP IBP (3 channels) and TEMP ( 2 channels) Display items for 9 other bed display: Items: Bed ID, Patient name, Alarm (if any), Heart rate and SpO2 Waveform: none Numeric data: HR, SpO2

ECG Complies with IEC 60601-2-27: 2005, ANSI/AAMI EC13: 2002, ANSI/AAMI EC57: 1998. Leads: 3-electrode cable: I, II, III 6-electrode cable: I, II, III, aVR, aVL, aVF, 2 from V1 to V6 10-electrode cable: I, II, III, aVR, aVL, aVF, V1 to V6 Defibrillation-proof: ECG input protected against 400 Ws/DC 5 kV IEC 60601-2-27 17.101 compatible Electrode offset potential tolerance: ≥ ±500 mV Input dynamic range: ≥ ±5 mV Internal noise: ≤ 30µVp-p (Referred to input) Noise suppression: RL driving gain: maximum 40 dB Maximum voltage: 1.23 Vrms Common mode rejection ratio: ≥ 95 dB Input bias current: ≤ 100 nA Frequency response: DIAG mode: 0.05 to 150 Hz (–3 dB) MONITOR mode: 0.3 to 40 Hz (–3 dB) MAXIMUM mode: 1 to 18 Hz (–3 dB)

AC hum filter: ≤ –40 dB (at 50 or 60 Hz) Input impedance: ≥ 5 MΩ (at 10 Hz), ≥ 2.5 MΩ (at 0.67 to 40 Hz), IEC 60601-2-27*2005 ESU protection: Provided (IEC 60601-2-27: 2005 compatible) Recovery time after defibrillation: 10 s Leads-off sensing: Each leads has own sensing with user configurable automatic lead change Active electrode: < 100 nA Reference electrode: < 900 nA Waveform display: Display sensitivity: 10 mm/mV ±5% (DIAG mode at ×1 sensitivity) Number of channels: 3 (maximum, with 6 or 10 electrodes on home screen), 12 (maximum, with 10 electrodes at 12 LEAD window) Sensitivity control: ×1/4, ×1/2, ×1, ×2, ×4, or AUTO Pacing mark display: Available Heart rate count: Calculation method: Moving average/Instantaneous beat to beat QRS detection (at × 1 sensitivity): Adult: Width: 70 to 120 ms, Amplitude: 0.5 to 5 mV, rate: 30 to 200 beats/min Child and neonate: Width: 40 to 120 ms, Amplitude: 0.5 to 5 mV, rate: 30 to 250 beats/min Counting range: 0, 15 to 300 beats/min (±2 beats/min) Counting accuracy: ±2 beats/min (0, 15 to 300 beats/min) (Essential performance in EMC standard) Heart rate display update cycle: Every 3 s or when alarm is generated. Heart rate sync mark delay time: within 100 to 200 ms (when QRS is detected) Tall T-wave rejection capability: Complies with the heights of T-waves from 0 mV to 1.2 mV specified in ANSI/ AAMI EC13 Sect. 4.1.2.1(c) Heart rate averaging: Calculated by using the most recent 4 or 12 beats. Pacemaker pulse detector rejection of fast ECG signals: Slew rate at which the pacemaker pulse detector responds: 6 to 8 V/s Tested as specified in ANSI/AAMI EC13 Sect. 4.1.4.3 Pacemaker pulse rejection capability, without overshoot: Complies with the amplitudes of pacemaker pulses ±2 to ±700 mV and widths 0.1 to 2 ms specified in ANSI/AAMI EC13 Sect. 4.1.4.1Pacemaker pulse rejection capability, with overshoot: Overshoot amplitudes and time constants of ±0.12 mV/100 ms to ±2 mV/4 ms (As defined by method B of ANSI/AAMI EC13 Sect. 4.1.4.2, this corresponds to the pacemaker pulses amplitudes and widths of ±4 mV/2 ms to amplitudes ±80 mV/0.1 ms.) Heart rate alarm: Upper limit range: 16 to 300 beats/min, OFF in 1 beat/min steps

Page 6: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

6

Lower limit range: OFF, 15 to 299 beats/min in 1 beat/min steps Alarm items: TACHYCARDIA, BRADYCARDIA, Arrhythmia analysis: Analysis method: Multi-template matching method Number of channels: 2 QRS detection type: Adult, childe, neonate VPC counting rate: 0 to 99 VPCs/min Arrhythmia message: ASYSTOLE, VF, VT, V BRADY, EXT TACHY, EXT BRADY, SV TACHY, VPC RUN, TACHYCARDIA, BRADYCARDIA, COUPLET, EARLY VPC, MULTIFORM, V RHYTHM, PAUSE, BIGEMINY, TRIGEMINY, FREQ VPC, VPC, IRREGULAR RR, PACER NON-CAPTURE, PROLONGED RR, NO PACER PULSE Other messages: NOISE, CHECK ELECTRODES, LEARNING Arrhythmia alarm: Upper limit range: OFF, 1 to 99 VPC/min Number of arrhythmia recall files: 32,768 (72 hours) Storage time per file: 8 s ST level measurement: Number of measurement channels: 3-electrode: 1 ch 6-electrode: 8 ch 10-electrode: 12 ch ST level measuring range: ±2.5 mV Measurement point: automatic with possibility to adjust manually ST level alarm: Upper limit range: –1.99 to 2.00 mV OFF Lower limit range: OFF, –2.00 to 1.99 mV Storage capacity: up to 4,320 files Cascade ECG Waveform: available

12-lead Interpretation 12 lead ECG interpretation: ECAPS 12C (BSM), Measurements and diagnostic, Available when monitoring 12 leads Analysis patient age: 3 years to adult Analysis time: about 30 sec Finding items: 200 Judging items: 5 Interpretation items: Normal sinus rhythm, TACHYCARDIA, BRADYCARDIA, VPC Display and output: Screen, network printer, printer connected to the central monitor File storage: 18 files

Respiration (Transthoracic Impedance Pneumography)

Measuring method: Transthoracic impedance pneumography Number of channels: Selectable from R-F and R-L Measuring impedance available range: 220 Ω to 4 kΩ Excitor current: 40 ±10 µArms at 40 kHz (sine wave) Internal noise: ≤ 0.2 Ω (Referred to input) Respiration rate counting range: 0 to 150 counts/min

Respiration rate counting accuracy*: ±2 counts/min (0 to 150 counts/min) * Essential performance in EMC standard Frequency response (high frequency cut-off): 3 Hz ±1 Hz (–3 dB) Defibrillation proof: Respiration input protected against 400 Ws/DC 5 kV Recovery time after defibrillation: 10 s Impedance respiration: Measurement On/Off available Heart beat rejection: Available Waveform display: Display sensitivity: 10 mm/1 Ω ±25% (at ×1 sensitivity) Sensitivity control: ×1/4, ×1/2, ×1, ×2, ×4 Respiration rate display update cycle: Every 3 s or when alarm is generated Alarm: Upper limit range: 2 to 150 counts/min, OFF Lower limit range: OFF, 0 to 148 counts/min Apnea alarm: OFF, 5 to 40 s in 5 s steps Displayed message: APNEA

SpO2 Complies with ISO 9919: 2005. Display: Display update cycle: Every 3 s or when alarm is generated Sync tone modulation: Changes tone depending on SpO2 value Sweep speed: 6.25, 12.5, 25, 50 mm/s(±10%) Waveform sensitivity: ×1/8, ×1/4, ×1/2, ×1, ×2, ×4, ×8 or AUTO SpO2: Measuring method: Two wavelength light absorption method Wavelength range: BSM-1733: 660/905 nm (LNOP tip clip and LNCS tip clip), 663/880 nm (Other clips) BSM-1753: 660/900 nm BSM-1763: 660/940 nm Averaging time: BSM-1753: 6 to 7 s (approximately 3 seconds in FAST mode), If the dynamic averaging time exceeds 20 seconds for SpO2, the SpO2 and pulse rate will continue to be updated every second Display range: BSM-1733/BSM-1753:1 to 100%SpO2 BSM-1763: 0 to 100%SpO2 Declared range: BSM-1733/BSM-1753/BSM-1763: 70 to 100%SpO2 Measuring accuracy*: BSM-1733: Adult: ±2%SpO2 (No motion), ±3%SpO2 (Motion), Neonate: ±3%SpO2 BSM-1753: Adult: ±2%SpO2, Neonate: ±3%SpO2 BSM-1763: 70%SpO2 ≤ %SpO2 < 80%SpO2 ±3%SpO2, 80%SpO2 ≤ %SpO2 ≤ 100%SpO2 ±2%SpO2 SpO2 accuracy is guaranteed at surrounding temperature of 18 to 40°C (64.4 to 104°F). * Essential performance in EMC standard SpO2 alarm: Upper limit range: 51 to 100%SpO2 in 1%SpO2

Page 7: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

7

steps , OFF Lower limit range: OFF, 50 to 99%SpO2 in 1%SpO2 steps Pulse rate: Display range: BSM-1733: 25 to 240 beats/min BSM-1753: 20 to 300 beats/min BSM-1763: 30 to 300 beats/min Declared range: BSM-1733: 25 to 240 beats/min BSM-1753: 20 to 250 beats/min BSM-1763: 30 to 300 beats/min Counting accuracy (rms)*: BSM-1733: ±3 beats/min (No motion), ±5 beats/min (Motion) BSM-1753: ±3 beats/min BSM-1763: ±3% ±1 beat/min * Essential performance in EMC standard Pulse rate alarm: Upper limit range: 16 to 300 beats/min in 1 beat/min steps, OFF (When SYNC SOURCE is set to ECG) 31 to 300 beats/min in 1 beat/min steps, OFF (When SYNC SOURCE is set to PRESS or SpO2) Lower limit range: OFF, 15 to 299 beats/min in 1 beat/min steps (When SYNC SOURCE is set to ECG) OFF, 30 to 299 beats/min in 1 beat/min steps (When SYNC SOURCE is set to PRESS or SpO2) Escalating Alarm: SpO2 drop, SpO2 sub value time Response time (BSM-1763 only): Selectable from “SLOW”, “NORMAL” and “FAST” The following graphs show the response time example when SpO2 changes 0.6%/s. The following graph shows the response time example when pulse rate changes 10 bpm/s.

Recovering time after defibrillation: 10 s PI: available for BSM-1733, BSM-1763

Non Invasive Blood Pressure, NIBP Complies with IEC 60601-2-30: 1999. Measuring method: Oscillometric Measuring range: 0 to 300 mmHg Cuff pressure display range: 0 to 300 mmHg Accuracy: ±3 mmHg (0 mmHg ≤ NIBP ≤ 300 mmHg) Cuff inflation time: Adult/Child: duty 100%, ≤ 20 s (700 cc, 0 to 200 mmHg) Neonate: duty 50%, ≤ 15 s (70 cc, 0 to 200 mmHg) Measurement mode: Adult, child or neonate is recognized by connected air hose Typical Measurement time: neonatal 20s, pediatric 25s, adult 30 sec. Maximum measurement time: Adult/Child: ≤ 160 s Neonate: ≤ 80 s Operation mode: Manual, STAT (≤ 5 or 10 min), Periodic, lumbar Auto remeasurement: 1 time Air leakage: ≤ 3 mmHg/min

Page 8: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

8

Measurement accuracy with a simulator*: ±10 mmHg * Essential performance in EMC standard Systolic range: Adult: 0 - 300 mmHg Child: 0 - 300 mmHg Neonate: 0 - 150 mmHg Diastolic range: Adult: 0 - 300 mmHg Child: 0 - 300 mmHg Neonate: 0 - 150 mmHg Mean range: Adult: 0 - 300 mmHg Child: 0 - 300 mmHg Neonate: 0 - 150 mmHg Initial pressurization value: Adult: 180 mmHg Child: 140 mmHg Neonate: 100 mmHg Maximum pressurization value: Adult/Child: 300 mmHg Neonate: 150 mmHg Display items: Systolic (SYS), diastolic (DIA), mean (MAP), cuff pressure during NIBP, measurement, delta PWTT NIBP data display update cycle: Updated every measurement Measurement completion sound: Generated at measurement completion (depends on the setting) Alarm: Upper limit range: 15 to 260 mmHg in 5 mmHg steps, OFF Lower limit range: OFF, 10 to 255 mmHg in 5 mmHg steps Safety: Maximum pressurization value cuff inflation limiter: Adult/Child: 300 to 330 mmHg Neonate: 150 to 165 mmHg Cuff inflation time limiter: Adult/Child: 161 to 165 s Neonate: 81 to 84 s Interval time limiter: 25 to 29 s Power discontinuity: Deflate immediately after power down Recovering time after defibrillation: 10 s

Invasive Blood Pressure, IBP Complies with IEC 60601-2-34: 2000 except for clauses 44.6, 45.101 a) and 45.101 b). Labels: P1, P2, P3, ART, ART2, RAD, DORS, AO, FEM, UA, UV, PAP, CVP, RAP, RVP, LAP, LVP, ICP, ICP2, ICP3, (ICP4, P4 to P7) Calculation: CVP-ET Complied transducer: P23XL-1 and P10EZ-1 Argon Medical devices reusable transducers Argon Medical devices disposable transducers DX series 5 µV/V/mmHg, bridge resistor: 200 Ω to 20 kΩ, defibrillation-proof or the equivalents For the warm-up time, refer to the transducer manual. Volume displacement: 0.04 mm3/100 mmHg Auto zero balancing range: ±200 mmHg

Auto zero balancing accuracy: ±1 mmHg Calculation Method: Peak, Standard (Average) Measuring range: –50 to 300 mmHg Measuring accuracy: ±1 mmHg ±1 digit (–50 mmHg ≤ IBP < 100 mmHg) ±1% ±1 digit (100 mmHg ≤ IBP ≤ 300 mmHg) Total measuring accuracy*: ±4% or ±4 mmHg (whichever is greater)** * Essential performance in EMC standard ** When used with ANSI/AAMI BP-22-1994 complied equipments Internal noise: within ±1 mmHg Temperature zero drift: ±0.1 mmHg/1°C Frequency response: DC to 12 Hz or 20 Hz (selectable) Recovering time after defibrillation: 10 s Display items: Systolic (SYS), diastolic (DIA), mean (MEAN) Display update cycle: Every 3 s or when alarm is generated BP sync sound: Systolic value 20 to 120 mmHg, changes in 20 steps every 5 mmHg Alarm: Upper limit range: −48 to 300 mmHg in 2 mmHg steps, OFF Lower limit range: OFF, −50 to 298 mmHg steps in 2 mmHg steps Alarm inactivation: Alarm is inactivated in certain period when zero balancing is performed. Pulse rate: Counting range: 0, 30 to 300 beats/min Display range: 0 to 300 beats/ min Counting accuracy (rms): ±2 beats/min (30 beats/min ≤ PR ≤ 300 beats/min) Alarm: Upper limit range: 16 to 300 beats/min in 1 beat/min steps, OFF (When SYNC SOURCE is set to ECG) 31 to 300 beats/min in 1 beat/min steps, OFF (When SYNC SOURCE is set to PRESS or SpO2) Lower limit range: OFF, 15 to 299 beats/min in 1 beat/min steps (When SYNC SOURCE is set to ECG) OFF, 30 to 299 beats/min in 1 beat/min steps (When SYNC SOURCE is set to PRESS or SpO2)

Temperature Complies with EN 12470-4: 2000 only for clauses 6.2, 6.3 a), 6.5, 6.6, 6.7, 6.8, 6.9, 6.10 and 8. Labels: T1, T2, Tskin, Tskin2, Trect, Tcore, Tnaso, Teso, Ttymp, Tblad, Taxil (Tskin3, T3 and T4) Thermistor probe: 400 series (YSI) Number of channels: 2 channels with Delta Temp Measuring range: 0 to 45°C, 32 to 113°F Measuring accuracy*: ±0.1°C (25°C ≤ TEMP ≤ 45°C) ±0.2°C ( 0°C ≤ TEMP < 25°C) * Essential performance in EMC standard Internal noise: ≤ 0.014°C (at 37°C)

Page 9: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

9

Temperature drift: within ±0.005°C /°C Display range: 0 to 45°C (32 to 113°F) Display update cycle: Every 3 s or when alarm is generated Time response delay from probe to monitor display: ≤ 6 seconds (sensor time constant is not included) Recovering time after defibrillation: 10 s Alarm: Upper limit range: 0.1 to 45.0°C (33 to 113°F) in 0.1°C (1°F) steps, OFF Lower limit range: OFF, 0.0 to 44.9°C (32 to 112°F) in 0.1°C (1°F) steps

Cardiac Output, CO Measuring method: Thermodilution method Measuring parameters: Cardiac output (CO), injectate temperature (Ti), blood temperature (Tb), delta Tb Number of channel: 1 Measuring range: Injectate temperature (Ti): 0°C to 27°C (32 to 81°F) Blood temperature (Tb): 15°C to 45°C (59 to 113°F) Thermodilution curve (delta Tb): 0°C to 2.5°C (32 to 37°F) Cardiac output (CO): 0.5 to 20 L/min Measuring accuracy: Ti: ±0.2°C (0°C to 27°C) Tb: ±0.1°C (25°C ≤ TEMP ≤ 45°C) ±0.2°C (15°C ≤ TEMP < 25°C) CO: ±5% Internal noise: Ti: ≤ 0.025°C RMS Tb: ≤ 0.016°C RMS (correspond to 37°C) Delta Tb: ≤ 0.005°C RMS Temperature drift: Ti: ±0.005°C /°C Tb: ±0.005°C /°C Frequency response (delta Tb): DC to 12 Hz (–3 dB) Injectate volume range: 3, 5, 10 mL Display update cycle: Updated every measurement Tb alarm: Upper limit range: 15.1 to 45.0°C (60 to 113°F) in 0.1°C (1°F) steps, OFF Lower limit range: OFF, 15.0 to 44.9°C (59 to 112°F) in 0.1°C (1°F) steps

Carbon Dioxide, CO 2 (Mainstream Method)

For the TG-900P/TG-920P/TG-950P/TG-970P CO2 sensor kit specifications, refer to the kit manual. Displayed CO 2*: Values detected by the TG-900P/TG-920P/TG-950P/TG-970P CO2 sensor kit * Essential performance in EMC standard Calculation method: TG-900P/TG-920P: semi-quantitative TG-950P/TG-970P: quantitative CO2 measuring range:

TG-900P/TG-920P/TG-950P: 0 to 100 mmHg TG-970P: 0 to 150 mmHg CO2 measuring accuracy: TG-900P/TG-920P: ±0.4 kPa (0 ≤ CO2 ≤ 1.33 kPa) (±3 mmHg (0 ≤ CO2 ≤ 10 mmHg)) ±0.53 kPa (1.33 < CO2 ≤ 5.33 kPa) (±4 mmHg (10 < CO2 ≤ 40 mmHg)) ±10% reading (5.33 < CO2 ≤ 13.3 kPa (40 < CO2 ≤ 100 mmHg)) (At 1 atmospheric pressure, air inspiration, no condensation) TG-950P/TG-970P: ±0.27 kPa (0 ≤ CO2 ≤ 5.33 kPa) (±2 mmHg (0 ≤ CO2 ≤ 40 mmHg)) ±5% reading (5.33 < CO2 ≤ 9.33 kPa (40 < CO2 ≤ 70 mmHg)) ±7% reading (9.33 < CO2 ≤ 13.3 kPa (70 < CO2 ≤ 100 mmHg)) ±10% reading (13.3 < CO2 ≤ 20 kPa (100 < CO2 ≤ 150 mmHg)) *TG-970P only (When no condensation) Warm-up time: TG-900P/TG-920P: 5 s TG-950P: 15 s TG-970P: 10 s Response time: TG-900P: 160 ms (typical) for steps from 10 to 90% TG-920P/TG-950P/TG-970P: 120 ms (typical) for steps from 10 to 90% Respiration rate counting range: TG-900P/TG-920P: 3 to 150 counts/min TG-950P/TG-970P: 0 to 150 counts/min Respiration rate counting accuracy: TG-900P/TG-920P: ±10% (3 to 150 counts/min) TG-950P/TG-970P: ±1 count/min CO2 value display update cycle: Every 3 s or when alarm is generated CO2 alarm: Upper limit: CO2 (I): 1 to 99 mmHg in 1 mmHg steps, OFF 0.1 to 13.0 kPa in 0.1 kPa steps, OFF ETCO2: 2 to 99 mmHg in 1 mmHg steps, OFF 0.2 to 13.0 kPa in 0.1 kPa steps, OFF Lower limit: ETCO2: OFF, 1 to 98 mmHg in 1 mmHg steps OFF, 0.1 to 12.9 kPa in 0.1 kPa steps Respiration rate alarm: Upper limit range: 2 to 150 counts/min in 2 counts/min steps, OFF Lower limit range: OFF, 0 to 148 counts/min in 2 counts/min steps Apnea time: OFF, 5 to 40 s in 5 s steps Displayed message: APNEA Total system response time: ≤ 1.0 second Recovering time after defibrillation: 10 s

Bispectral Index, BIS For the BISx/BIS processor specifications, refer to the BISx/BIS processor manual.

Page 10: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

10

BIS can be also monitored with Covidien’s BIS monitor. Recovering time after defibrillation: 10 s BIS alarm: Upper limit range: 1 to 100 in 1 steps, OFF Lower limit range: OFF, 0 to 99 in 1 steps

ECG/BP Output Outputs 100 mmHg/V IBP waveform and the first trace of 1 mV/V ECG waveform. When more than one IBP waveforms are acquired, the IBP waveform of the top MULTI socket on the monitor is output (when “FIXED POSITION” is set for IBP Analog Output) or the IBP waveform is output following the highest priority label (when “HIGHEST PRIORITY LABEL” is set for IBP Analog Output). Complied medical electrical equipments: Connecting medical electrical equipment must comply to the following standards: IEC 60601-1: 1988 IEC 60601-1 Amendment 1: 1991 IEC 60601-1 Amendment 2: 1995 Medical electrical equipment must be connected by specified method in following standards: IEC 60601-1-1: 2000 Output impedance: ECG: ≤ 100 Ω±20% BP: ≤ 100 Ω±20% Output-waveform: ECG: ±5.0 V (at 1 mV/V ± 5% sensitivity) BP: –0.5 to +3.0 V (at 100 mmHg/V ± 1% sensitivity) HT: 5.0 to 15.0 V (Open collector output: 0.5 to 50 mA) Frequency response: ECG: ≥ 0.5 to 100 Hz (≥ –3 dB) (No reproducibility of pace maker pulse) BP: DC to 20 Hz ±3 Hz (–3 dB) HT pulse width: 15 ms Offset: ECG: ≤ ±50 mV BP: ≤ ±10 mV Delay: ECG: 20ms max BP: 40ms max Sensitivity accuracy: ECG: ±5% BP: ±1%

When ZS-900PK Transmitter is Connected

Frequency capacity deviation: ≤ ±3 ppm (15 to 35°C) Transmission power: 1.0 mW +5%, –40% (15 to 35°C) Spurious emission strength: ≤ 2.5 µW (5 MHz to 1.5 GHz) Occupied bandwidth: 5.0 to 8.5 kHz Adjacent channel leaking power: ≥ 40 dBR Transmission frequency range: 420.0500 to 449.6625 MHz Modulation method: Frequency shift keying

Parameters sendable to central monitor: ECG (1ch), respiration, SpO2, NIBP, IBP (2 ch), temp (2ch), CO2

esCCO (Optional QP-172P is required) Method: estimated continuous noninvasive cardiac output, pulse wave transit time method Measured parameter: esCCO, esCCI, esSV, esSVI, (To calculate esCCI, esSVI, patient information (BSA) is required. Calibration CO source: CO from patient information, CO from Hemo Trend, CCO, Manually enter Calibration pulse pressure source: ART, ART2, RAD, FEM, NIBP Measuring range: esCCO: 0.5 to 20.00 L/min, 0.01 L/min steps esCCI: 0.5 to 20.00 L/min/m2, 0.01 L/min/m2 steps esSV: 0 to 300 mL, 1 mL steps esSVI: 0 to 200 mL/m2, 1 mL/m2 steps Alarm range: esCCO upper limit:0.60 to 20.00 L/min in 0.10 L/min steps, OFF esCCO lower limit: OFF, 0.50 to 19.90 L/min in 0.10 L/min steps esCCI upper limit:0.60 to 20.00 L/min in 0.10 L/min steps, OFF esCCI lower limit: OFF, 0.50 to 19.90 L/min in 0.10 L/min steps Signal Quality Index: available (4 levels)

Battery (SB-170P Battery Pack) Type of battery: Lithium ion Number of battery: 1 Battery operation time: approximately 5-hour (new battery, fully charged and no options are used in normal temperature) Line voltage: Rate Charging current: 800mA Charging time: approximately 4.5h Battery status indication: Operating on battery/ remaining battery power (on screen), charging, charging complete, battery error, battery weak, battery not inserted Operating environment: Charging temperature: 10 to 55°C (50 to 131°F) Discharging temperature: 5 to 50°C (41 to 122°F) Humidity: 30 to 85% RH (noncondensing) Atmospheric pressure: 700 to 1060 hPa Transport and storage environment: When the battery pack is stored more than 6 months, charge and discharge or charge the battery once every 6 months. Temperature: –20 to +60°C (–4 to +140°F) Humidity: 20 to 85% RH (noncondensing) Atmospheric pressure: 700 to 1060 hPa

External Output ZS-900PK: Available LS-NET: Available

Power Requirement Power source: DC only

Page 11: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

11

Line voltage: Rate+12V, operation range +9.0V to 12.6V

DC source: SB-170p battery pack, SC-170R AC cradle or host monitor Power input: 40W Battery charging current: 400mA or 800mA

Clock Accuracy At operating temperature 25°C: approx. ±2 min 40 s/month maximum At storage temperature –20 to +60°C: approx. ±6 min/month maximum Environment

Operating environment Temperature: 10 to 40°C (50 to 104°F) SpO2 accuracy is guaranteed at surrounding temperature of 18 to 40°C (64 to 104°F). Humidity: 30 to 85% RH (10 to 40°C, noncondensing) Atmospheric pressure: 700 to 1060 hPa Transport and storage environment: Temperature: –20 to +65°C (–4 to +149°F) Humidity: 10 to 95% RH Atmospheric pressure: 700 to 1060 hPa

Mechanical Strength MIL-STD-810F

514.5 Vibration environments Category 4 & 9 516.5 Mechanical shock, Procedure IV- Transit

drop

Electromagnetic Compatibility IEC 60601-1-2: 2001 IEC 60601-1-2 Amendment 1: 2004

Safety Standard Safety standard:

EN 12470-4: 2000*1 IEC 60601-1: 1988 IEC 60601-1 Amendment 1: 1991 IEC 60601-1 Amendment 2: 1995 IEC 60601-1-1: 2000 IEC 60601-1-2: 2001 IEC 60601-1-2 Amendment 1: 2004 IEC 60601-1-4: 1996 IEC 60601-1-4 Amendment 1: 1999 IEC 60601-1-6: 2010 IEC 60601-1-8: 2006*2*3 IEC 60601-1-9: 2007 IEC 60601-2-27: 2005 - Particular requirements for the safety, including essential performance, of electrocardiographic monitoring equipment IEC 60601-2-30: 1999 - Particular requirements for the safety, including essential performance of automatic cycling in non-invasive blood pressure monitoring equipment IEC 60601-2-34: 2000 - Particular requirements for the safety, including essential performance of automatic cycling of invasive blood pressure monitoring equipment*4 IEC 60601-2-49: 2001 - Particular

requirements for the safety of multifunction patient monitoring equipment IEC 62304: 2006 ISO 9919: 2005 ISO 21647: 2004 *1 This monitor complies with EN 12470-4: 2000 only for clauses 6.2, 6.3 a), 6.5,6.6, 6.7, 6.8, 6.9, 6.10 and 8. *2 Only the “IEC standard” alarm sound complies with clause 6.3.3.2. *3 This monitor complies with IEC 60601-1-8: 2006 except for interbed alarm. *4 This monitor complies with IEC 60601-2-34: 2000 except for clauses 44.6,45.101 a) and 45.101 b). Type of protection against electrical shock: CLASS I EQUIPMENT (AC Powered by the SC-170R AC cradle or host monitor) Internally Powered EQUIPMENT (BATTERY Powered) Degree of protection against electrical shock: Defibrillator-proof type CF applied part: ECG, SpO2, NIBP, IBP, temperature, multi connector, external output connector. When the multi connector is used for CO measurement, it is not defibrillator proof. Degree of protection against harmful ingress of water: BSM-1733/BSM-1753/BSM-1763: IPX2 (protected against vertically falling water drops when enclosure tilted up to 15゜ , when the TEMP connector cover, ZS cover, USB cover and battery cover are firmly attached to each connector) SC-170R AC cradle: IPX1 (protected against vertically falling water drops) Degree of protection against solid foreign opjects: IP3X (Protected against solid foreign objects of 2.5mm diameter and greater) Degree of safety of application in the presence of FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE: Equipment not suitable for use in the presence of FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE Mode of operation: CONTINUOUS OPERATION

Dimensions and Weight Not including the battery and options: Dimensions: 147W × 194 H × 94 D mm (excluding protruding parts) Weight: 1.4 kg With battery pack: Dimensions: 147W × 194 H × 94 D mm (excluding protruding parts) Weight: 1.57 kg

Electromagnetic Emissions For Electromagnetic Emissions, refer to the Operator's manual.

Electromagnetic Immunity For Electromagnetic Immunity, refer to the

Page 12: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

12

Operator's manual.

STANDARD ACCESSORIES Bedside monitor BSM-1733/1753/1763 Side Cover 1 Operator’s Manual 1 BSM-1700 series Manuals CD-ROM 1

ACCESSORIES Accessory set for BSM-1763 (NK SpO 2) (An accessory set must be ordered with main unit.)

Order code Y212A Y212B Y213A Y213B

Model YO-60IY1 YO-60IY2 YO-60AY1 YO-60AY2

Electrode lead,

3 electrode, cl ip, IEC,

K911, BR-903P

√ √

Electrode lead, 3 electrode, cl ip, AHA, K911A, BR-903PA

√ √

ECG connection cord

K922, JC-906P √ √

ECG connection cord

K922A, JC-906PA √ √

SpO2 connection cord

K931, JL-900P √ √ √ √

Air hose for NIBP

S902, YN-901P √ √ √ √

NIBP cuff, adult, 13 cm cuff width,

S951D, YP-713T √ √

NIBP cuff, adult large, 16 cm cuff width, S951E, YP-714T

√ √

Order power cord separately or purchase locally Power cord L936 European version 200-240V

L935 US version 100-110V Accessory set for BSM-1753(Nellcor SpO2)/BSM-1733 (Masimo SpO 2) (An accessory set must be ordered with main unit).SpO2 connection cord, JL-650P is available for BSM-1753

Order code Y212C Y212D

Model YO-60IY3 YO-60IY4

Electrode lead, 3 electrode, cl ip, IEC,

K911, BR-903P √ √

Electrode lead, 3 electrode, cl ip, AHA, K911A, BR-903PA

ECG connection cord

K922, JC-906P √ √

ECG connection cord

K922A, JC-906PA

Air hose for NIBP

S902, YN-901P √ √

NIBP cuff, adult, 13 cm cuff width,

S951D, YP-713T √

NIBP cuff, adult large, 15 cm cuff width, S951E, YP-714T

Order power cord separately or purchase locally Power cord L936 European version 200-240V

L935 US version 100-110V

OPTIONS esCCO Program, QP-172P AC cradle, SC-170R Ground lead, 544582 Battery, X161 (SB-170P) Transmitter, ZS-900PK Bed rail hook, DZ-170P Holder for transmitter, DI-170P Cart, KC-170P Adapter, DH-170P Mount Adapter, DH-171P Pole clamp, DH-172P Holders, DH-692P Wall mount kit, product of GCX (local purchase) BIS processor, QE-910P Connection cable to BIS processor (QE-910P), YJ-671P Upgrade software, QS-065P Network printer, Local purchase

CONSUMABLES ECG and Respiration (Impedance)

Cable/cord K911 Electrode lead, BR-903P, for 3 electrodes, clip, IEC, 0.8 m K911A Electrode lead, BR-903PA, for 3 electrodes, clip, AHA, 0.8 m K916 Electrode lead, BR-963P, for 3 electrodes on one lead, clip, IEC, 0.8 m K912 Electrode lead, BR-906P, for 6 electrodes, clip, IEC, 0.8 m K912A Electrode lead, BR-906PA, for 6 electrodes, clip, AHA, 0.8 m K922 ECG connection cord, JC-906P, 3/6 electrodes, IEC, 3 m K922A ECG connection cord, JC-906PA, 3/6 electrodes, AHA, 3 m K925 ECG connection cord, JC-916P, 3/6 electrodes, IEC, 1.5 m K921 ECG connection cord, JC-900P, 10 electrodes, IEC, 3 m, use with disposable electrode with lead K921A ECG connection cord, JC-900PA, 10 electrodes, AHA, 3 m, use with disposable electrode with lead K901 ECG patient cable, BJ-900P, 10 electrodes, IEC, 3.8 m K901A ECG patient cable, BJ-900PA, 10 electrodes, AHA, 3.8 m K974 ECG/BP output cable, YJ-910P, 5 m

K975 ECG/BP output cable, YJ-920P, 0.3 m Disposable electrodes (Vitrode series) G203 L-150, for general use, 30 electrodes x 5/set, 35 mm dia, Vitrode L G207 L-150X, for general use, X-ray (radiolucent), 30 electrodes x 5/set, 35 mm dia, Vitrode L

Page 13: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

13

G210D F-150M, for infant/neonate to adult, 3 electrodes x 50/set, 25 x 45 mm, Vitrode F G210C F-150S, for NICU, 3 electrodes x 50/set, 18 x 36 mm, Vitrode F G221 G-600, for ICU/OR, 30 electrodes x 20/set, 47 x 47 mm, Vitrode G G272A V-090M3, for general use, 3 electrodes x 30/set, DIN, 25 x 45 mm, Vitrode V G272B V-040M4, for general use, 4 electrodes x 10/set, DIN, Vitrode V G278A V-09IO3, for general use, 3 electrodes x 30/set, DIN, 25 x 45 mm, Vitrode V G273A V-04IO4, for general use, 4 electrodes x 10/set, DIN, 25 x 45 mm, Vitrode V G272C V-060M6, for general use, 6 electrodes x 10/set, DIN, Vitrode V G274A V-06IO6, for general use, 6 electrodes x 10/set, DIN, 25 x 45 mm, Vitrode V G271A V-120S3, for NICU, 3 electrodes x 40/set, DIN, 20 x 20 mm, Vitrode V G300A N-03IS3, for NICU, 3 electrodes x 10/set, DIN, Vitrode N

SpO2

Cord K931 SpO2 connection cord, JL-900P, 2.5 m K935 Nellcor SpO2 connection cord, JL-650P, 3 m K936 Masimo SpO2 connection cord, for LNOP series, JL-630P, 3.6 m K937 Masimo SpO2 connection cord, for LNCS series, JL-631P, 3 m Probes (BluPRO series) P225F Finger probe, TL-201T, attached to finger or toe by clip P311C Finger probe, TL-631T3, regular size, attached to finger or toe by tape P310C Finger probe, TL-630T3, large size, attached to finger or toe by tape P225G Multi-site probe, TL-220T, attached to finger or toe by tape P259 Cottony tape, 340703, secures probe and cord for stable monitoring, stretchable, 20 pcs/set P267 Probe fastener, 30 pcs/set

Reusable Probe TL-651T3 Finger-tip, 40 kg or more Disposable probes (BluPRO series) P203A TL-271T, for adult finger or toe, 24 pcs/set, 0.8 m P204A TL-271T, for adult finger or toe, 5 pcs/set, 0.8 m P203E TL-271T3, for adult finger or toe, 24 pcs/set, 1.6 m P204E TL-271T3, for adult finger or toe, 5 pcs/set, 1.6 m P203B TL-272T, for child finger or toe, 24 pcs/set, 0.8 m P204B TL-272T, for child finger or toe, 5 pcs/set, 0.8 m

P203F TL-272T3, for child finger or toe, 24 pcs/set, 1.6 m P204F TL-272T3, for child finger or toe, 5 pcs/set, 1.6 m P203C TL-273T, for neonate instep, 24 pcs/set, 0.8 m P204C TL-273T, for neonate instep, 5 pcs/set, 0.8 m P203G TL-273T3, for neonate instep, 24 pcs/set, 1.6 m P204G TL-273T3, for neonate instep, 5 pcs/set, 1.6 m P203D TL-274T, for infant finger or toe, 24 pcs/set, 0.8 m P204D TL-274T, for infant finger or toe, 5 pcs/set, 0.8 m P203H TL-274T3, for infant finger or toe, 24 pcs/set, 1.6 m P204H TL-274T3, for infant finger or toe, 5 pcs/set, 1.6 m Multi-site Y probe P205A Multi-site Y probe, TL-260T (for adult, child, neonate), 5 pcs/set P260A Sponge attachment tape S for TL-260T multi-site Y probe, 24 pcs/set P260B Sponge attachment tape L for TL-260T multi-site Y probe, 24 pcs/set P256 Ear clip adapter for TL-260T, reusable Disposable probes P228A TL-051S, for adult/neonate, 5 pcs/set, 0.8 m cable P228B TL-052S, for adult/neonate, 5 pcs/set, 1.6 m cable P229A TL-061S, for child/infant, 5 pcs/set, 0.8 m cable P229B TL-062S, for child/infant, 5 pcs/set, 1.6 m cable Nellcor SpO2 probes are available from NELLCOR (www.covidient.com) or their suppliers Masimo SpO2 probes are available from MASIMO (www.masimo.com) or their suppliers NIBP Reusable Cuffs (Latex free) S943A YP-960T: for infant, 5 cm cuff width

S943B YP-961T: for child, 7 cm cuff width

S943C YP-962T: for child, 10 cm cuff width

S944B YP-963T: for adult, 13 cm cuff width

S944C YP-964T: for adult, 15 cm cuff width

S944D YP-965T: for thigh, 19 cm cuff width Reusable Cuffs (YAWARA-CUFF series, Latex free) S949A YP-702T: for child, 10 cm cuff width

S949B YP-703T: for adult, 13 cm cuff width Reusable Cuffs (YAWARA-CUFF 2series, Latex free) S951A YP-710T: for infant, 5 cm cuff width

Page 14: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

14

S951B YP-711T: for child, 7cm cuff width

S951C YP-712T: for adult, 10 cm cuff width

S951D YP-713T: for adult, 13 cm cuff width

S951E YP-714T: for adult, 16 cm cuff width

S951F YP-715T: for thigh, 19 cm cuff width Disposable Cuffs S945C YP-810P: for infant, 6 cm cuff width, 20 pcs S945D YP-811P: for child, 8 cm cuff width, 20 pcs S946E YP-812P: for adult, 10 cm cuff width, 20 pcs S946F YP-813P: for adult, 14 cm cuff width, 20 pcs S946G YP-814P: for adult, 15 cm cuff width, 20 pcs S946H YP-815P: for adult, 17 cm cuff width, 20 pcs S946I YP-816P: for adult, 18 cm cuff width, 20 pcs S946J YP-817P: for thigh, 20 cm cuff width, 20 pcs S948A YP-820P: for neonate, 2 cm cuff width, 10 pcs S948B YP-821P: for neonate, 3 cm cuff width, 10 pcs S948C YP-822P: for neonate, 4 cm cuff width, 10 pcs S948D YP-823P: for neonate, 4.5 cm cuff width, 10 pcs S948E YP-824P: for neonate, 5 cm cuff width, 10 pcs Air hoses: S903 YN-990P: extension hose, 1.5 m

S901 YN-900P: 1.5 m

S902 YN-901P: 3.5 m

S904 YN-920P: for neonate, 1.5 m

S905 YN-921P: for neonate, 3.5 m

IBP

Argon Medical: K951 IBP connection cord, JP-900P, Argon Medical transducer, 3.5 m Argon Medical blood pressure transducers are available from Argon Medical Company (http://www.argonmedical.com) or their supplier. Edwards Lifesciences: L901 IBP connection cord, JP-920P, for Edwards Lifesciences transducer, 3.5 m Edwards Lifesciences blood pressure transducers are available from Edwards Lifesciences Corporation (www.edwards.com) or their suppliers. Biosensors International: K957 IBP connection cord, JP-960P, for Biosensor transducer, 3.5 m Biosensors blood pressure transducers are available from BIOSENSORS INTERNATIONAL (www.biosensorsintl.com) or

their suppliers. K952 IBP connection cord, JP-910P, 0.3 m

JP-911P IBP interface isolation cable

Temperature

Thermistor Probe P240B For adult rectum/esophagus, 401J

P241B For child rectum/esophagus, 402J

P242D Disc type, 409J

P249 Probe cover, for 401J, 10 pcs/set

P252 Temperature insulation pad, for 409J, 60 pcs/set

CO2 (Mainstream measurement)

Semi-quantitative method P903 CO2 sensor kit, TG-900P (TG-101T + JG-900P), 3 m P922A CO2 sensor, TG-101T, 1 m

K981 CO2 adapter, JG-900P, 2 m

R801 Airway adapter, YG-101T, 50 pcs/set

P907 CO2 sensor kit, TG-920P (TG-121T + JG-920P), 3.5 m P923 CO2 sensor, TG-121T, 2 m

K984 CO2 adapter, JG-920P, 1.5 m

R804 Airway adapter, YG-111T, 30 pcs/set

V922 Nasal adapter, YG-121T, for nasal/oral, 30 pcs/set V923 Nasal adapter, YG-122T, for nasal/oral, 30 pcs/set, with O2 supply tube adapter V927 Oxygen cannula, 25 pcs/set Quantitative method P905 CO2 sensor kit, TG-950P (TG-201T + JG-950P), 4 m

P921 CO2 sensor, TG-201T, 2 m

K982 CO2 adapter, JG-950P, 2 m

R802 Airway adapter, adult, YG-201T, 30 pcs/set R803 Airway adapter, pediatric, YG-202T, 30 pcs/set P909 CO2 sensor kit, TG-970P (TG-221T + JG-970P), 3.5 m

P924 CO2 sensor, TG-221T, 3 m

K987 CO2 adapter, JG-970P, 0.5 m

R805 Airway adapter, adult, YG-211T, 30 pcs/set R806 Airway adapter, pediatric, YG-213T, 30 pcs/set V910V Connector, YG-270P, 30 pcs/set, for YG-213T

CO

K962 CO connection cord, JT-950P, 2.0 m Argon Medical: Argon Medical injectate supply systems and

Page 15: BSM-1733/1753/1763 Bedside Monitor - Medidynemedidyne.dk/wp-content/uploads/ENG-BSM-1700-Specs.pdf · 2 FEATURES The BSM-1700 bedside monitor is intended for use for transport monitoring,

15

thermodilution cathetes are available from Argon Medical Company (www. argonmedical.com). Edwards Lifesciences: Edwards Lifesciences injectate supply systems and thermodilution cathetes are available from Edwards Lifesciences (www.edwards.com)

ORDERING INFORMATION When ordering, please specify the following information: 1) Power line voltage and its frequency 2) Requirements for consumables 3) Requirements for optional accessories 4) Transmission frequency of any existing

telemetry system—Nihon Kohden’s or any other manufacturers’

5) Specify part order code in