    BSBS--200 Parameter Setting200 Parameter Setting

    (Kinetics)(Kinetics)

    WendyWendyApr, 2008Apr, 2008

  • 8/7/2019 BS-200 Kinetics

    2/52

  • 8/7/2019 BS-200 Kinetics

    3/52

    PrefacePreface

    Here we set ALT as the example forall Kinetics items on

    Generally, Kinetics item includes: ALT, AST, ALP,ACP, CK, CK-MB, AMY, LIP, GGT, LDH,

  • 8/7/2019 BS-200 Kinetics

    4/52

  • 8/7/2019 BS-200 Kinetics

    5/52

    Input the item name forTest.

    ALT

  • 8/7/2019 BS-200 Kinetics

    6/52

    No. will be generated automatically by the system, eg 1 for the first one.

    1

    ALT

  • 8/7/2019 BS-200 Kinetics

    7/52

    Full Name of the item, can be blank if the user dont want to input anything.

    1

    ALT

  • 8/7/2019 BS-200 Kinetics

    8/52

    StandardNo. of the item, can be blank or involved with the bar code.

    1

    ALT

  • 8/7/2019 BS-200 Kinetics

    9/52

    Reac.Type of the item, can be referred from the reagent insert as ...

    1

    ALT

  • 8/7/2019 BS-200 Kinetics

    10/52

    So we can set it as Kinetics type.

    Indicates the absorbance should be monitored in exact minutes.

    Basic parameter for ALTBasic parameter for ALT

  • 8/7/2019 BS-200 Kinetics

    11/52

    1

    ALT

    Reac.Type is set as Kinetics as the reagent insert indicates.

    1

    Kinetics

  • 8/7/2019 BS-200 Kinetics

    12/52

    1

    ALT

    1

    Kinetics

    Pri.Wave & Sec.Wave should be selected as reagent insert indicates

  • 8/7/2019 BS-200 Kinetics

    13/52

    So we can set it as 340 nm for Pri. Wave; since nosecondary wavelength is mentioned, Sec. Wave could be

    blank.

    Basic parameter for ALTBasic parameter for ALT

  • 8/7/2019 BS-200 Kinetics

    14/52

    1

    ALT

    1

    Kinetics

    Pri.Wave is selected as 340 nm, Sec.Wave is blank.

    340

  • 8/7/2019 BS-200 Kinetics

    15/52

    1

    ALT

    1

    Kinetics

    340

    Direction of the absorbance changing during the reaction, refer to ...

  • 8/7/2019 BS-200 Kinetics

    16/52

    From Reaction

    Principle, we can

    see NADH(routineindicator under

    340nm) is a to-be-consumed

    reactant, which

    will be less and

    less during testing,

    thus, indicatingthe absorbance is

    lower and lower in

    the measurement.

    Thus, we can set

    it as Descending

    Basic parameter for ALTBasic parameter for ALT

  • 8/7/2019 BS-200 Kinetics

    17/52

    1

    ALT

    1

    Kinetics

    340

    Direction is selected as Descending.

    Descending

  • 8/7/2019 BS-200 Kinetics

    18/52

  • 8/7/2019 BS-200 Kinetics

    19/52

    Mono reagent items' reaction time begins when Sample is dispensed.

    Dual reagent items reaction time begins when R2 is dispensed.

    The time between S and R2 dispensing is called Incubation time,

    only dual reagent measurements have incubation time.

    R1 Sample

    Preheating time Reaction time

    To be set

    from the

    insert

    Fixed 14

    cycles

    Basic parameter for ALTBasic parameter for ALT

    Incubation

    time

    0 cycle of

    reaction time

    R2

  • 8/7/2019 BS-200 Kinetics

    20/52

    Time is all calculated into cycles on BS-200, and 1 cycleequals to 16 seconds.

    Roughly speaking, we change minutes into cycles forReac. Time and Incuba. Time as below.

    minute cycle

    0.5 2

    1 4

    2 8

    3 12

    4 16

    5 20

    Basic parameter for ALTBasic parameter for ALT

  • 8/7/2019 BS-200 Kinetics

    21/52

  • 8/7/2019 BS-200 Kinetics

    22/52

    1

    ALT

    1

    Kinetics

    340

    Reac.Time is set as 4 and 16, and Incuba.Time as 20.

    4 16

    20

    Descending

  • 8/7/2019 BS-200 Kinetics

    23/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    Unit of the item, should be set according to the reagent insert

    4 16

    20

  • 8/7/2019 BS-200 Kinetics

    24/52

    Basic parameter for ALTBasic parameter for ALT

    Kinetics method is always testing enzymes, and usuallythe Unit is U/L.

    According to the part of Reference Values, we can alsofind its proper to set the Unit as U/L.

  • 8/7/2019 BS-200 Kinetics

    25/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    4 16

    20

    We can set the Unit as U/L.

    U/L

  • 8/7/2019 BS-200 Kinetics

    26/52

  • 8/7/2019 BS-200 Kinetics

    27/52

    Basic parameter for ALTBasic parameter for ALT

    Its normal to report ALT results without decimals, so wecan set the Precision as 0.

  • 8/7/2019 BS-200 Kinetics

    28/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    4 16

    20

    U/L

    Precisionof the item is set as 0.

    0

  • 8/7/2019 BS-200 Kinetics

    29/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    4 16

    20

    U/L

    R1, R2and Sample Volumeof the item, should be referred to

    0

  • 8/7/2019 BS-200 Kinetics

    30/52

    To process with the sample testing at 37, we need 100ul sample,

    1000ul R1 and 250ul R2 (namely, the ratio of S/R1/R2 is 2/20/5);

    According to the volume range of BS-200 (sample: 3-45ul, R1: 180-

    450ul; R2: 30-450ul), and the indicated S/R1/R2 ratio;

    We can set Samplevolume as 18ul, R1 as 180ul, and R2as 45ul.

    Basic parameter for ALTBasic parameter for ALT

  • 8/7/2019 BS-200 Kinetics

    31/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    4 16

    20

    U/L0

    R1, R2 and Sample Volumeare set as 180, 45 and 18.

    18

    180

    45

  • 8/7/2019 BS-200 Kinetics

    32/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    4 16

    20

    U/L0

    Absorbance ofR1 Blank(mono rgt.) and MixedRgt. Blank(dual rgt.),

    set according to experience or reagent insert; if no idea, can be blank.

    18

    180

    45

  • 8/7/2019 BS-200 Kinetics

    33/52

  • 8/7/2019 BS-200 Kinetics

    34/52

    Basic parameter for ALTBasic parameter for ALT

    From the part of Linearity, we can get aware of that, themaximum absorbance change is 0.160 A/min, and theupper limit of Linearity Range is 350 U/L.

  • 8/7/2019 BS-200 Kinetics

    35/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    4 16

    20

    U/L

    0

    18

    180

    45

    We can set the upper limit ofLinearityRangeas 350 U/L.

    350

  • 8/7/2019 BS-200 Kinetics

    36/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    4 16

    20

    U/L

    0

    18

    180

    45

    LinearityLimit should be set for Kinetics tests, refer to the instruction

    350

  • 8/7/2019 BS-200 Kinetics

    37/52

    Background ofLinearity Limit check

    Its always testing enzymes with the method of Kinetics;

    In this kind of measurement, its supposed that theabsorbance changing speed remains constant during thereaction;

    So it should be a straight line as the reaction curve; Linearity is to check if the reaction curve is straight

    enough (thus, if the reaction is correct) to measure theconcentration of Kinetics items.

    Value ofLinearity Limit setting

    The linearity ranges from 0 to 1, and the default value is0.20000.

    If set, BS-200 will automatically calculate thelinearity during every measurement (as the next slide )and compare the result to the linearity limit and flag the

    result exceeding the defined limit.

    Basic parameter for ALTBasic parameter for ALT

  • 8/7/2019 BS-200 Kinetics

    38/52

    Basic parameter for ALTBasic parameter for ALT

    A1 A6. . . . At-5 . . . .At

    Linearity = 100% (A6 A1) (At At-5)

    At A1

  • 8/7/2019 BS-200 Kinetics

    39/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    4 16

    20

    U/L

    0

    18

    180

    45

    According to the instruction, LinearityLimit can be set as 0.200000.

    350

    0.200000

  • 8/7/2019 BS-200 Kinetics

    40/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    4 16

    20

    U/L

    0

    18

    180

    45

    SubstrateLimitshould be set for Kinetics tests, refer to the instruction

    350

    0.200000

  • 8/7/2019 BS-200 Kinetics

    41/52

    Background ofSubstrate Limit check

    Its always testing enzymes with the method of Kinetics;

    In this kind of measurement, its supposed that there issurplus substrate in the reagent (enough for the testing);

    When the samples concentration is higher than its linearrange, the substrate will be exhausted, and then thetesting result will not be correct;

    Substrate Limit is to check if the reaction is processedcorrectly.

    Value ofSubstrate Limit setting

    The Substrate Limit can be set from 0 to 50000(absorbance);

    In ascending reactions, it could be 20000~25000; and itcould be 0~5000 for descending reactions (eg nextslide ).

    If set, BS-200 will automatically compare absorbancevalue with the Linearity Limit and flag the result exceedingthe defined limit.

    Basic parameter for ALTBasic parameter for ALT

  • 8/7/2019 BS-200 Kinetics

    42/52

    Basic parameter for ALTBasic parameter for ALT

    Since its a descending reaction, we can set 5000 for ALT; itmeans when the absorbance becomes lower than 5000, BS-200

    will flag the result reminding you to rerun this sample.

    5000

  • 8/7/2019 BS-200 Kinetics

    43/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    4 16

    20

    U/L

    0

    18

    180

    45

    350

    0.200000

    According to the instruction, SubstrateLimit can be set as 5000.

    5000

  • 8/7/2019 BS-200 Kinetics

    44/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    4 16

    20

    U/L

    0

    18

    180

    45

    Once Factoris set, no calibration is required; it should be referred from

    350

    0.200000

    5000

  • 8/7/2019 BS-200 Kinetics

    45/52

    According to Calculation in the reagent insert and thereaction temperature of 37, we can set 2143 for theFactor.

    Basic parameter for ALTBasic parameter for ALT

  • 8/7/2019 BS-200 Kinetics

    46/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    4 16

    20

    U/L

    0

    18

    180

    45

    Factoris set as 2143. However, we recommend to calibrate ourselves.

    350

    2143

    0.200000

    5000

  • 8/7/2019 BS-200 Kinetics

    47/52

    1

    ALT

    1

    Kinetics

    340

    Descending

    4 16

    20

    U/L

    0

    18

    180

    45

    350

    2143

    Prozonecheck is set for the turbidimetric tests, which can also be blank.

    0.200000

    5000

  • 8/7/2019 BS-200 Kinetics

    48/52

    The basic parameters for ALT havebeen finished

    Next, we should set the reagentinformation as following

    Reagent parameter for ALTReagent parameter for ALT

  • 8/7/2019 BS-200 Kinetics

    49/52

  • 8/7/2019 BS-200 Kinetics

    50/52

  • 8/7/2019 BS-200 Kinetics

    51/52

    Basic reagent information (Position andBottle Type) has been set up

    We should set relevant calibration and QC

    information for the item

    (Please refer to the video of Calibration & QC setting)

    Calibration and QC for ALTCalibration and QC for ALT

  • 8/7/2019 BS-200 Kinetics

    52/52

    Thank you!Thank you!NOTE:Kinetics items should be set in the same way as ALT.