biopot. amp 1

Upload: etome05

Post on 03-Jun-2018

231 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/11/2019 Biopot. Amp 1

    1/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    1

    Biopotential Amplifiers

    These are very important part of modern medicalinstrumentation

    We need to amplify biopotentials which are generated in the

    body at low levels with a high source impedance

    Biopotentials amplifiers are required to increase signal strengthwhile maintaining fidelity

    Basic Requirements of Biopotential AmplifiersEssential functions of a bioamplifier are:

    To take a weak biopotential and increase its amplitude so that it

    can be processed, recorded or displayed

    To amplify voltage, but it could be considered as a poweramplifier as well

    To amplify current since in some cases a biopotential amplifier

    is used to isolate the load from the sourcecurrent gain only

    Input Impedance (Zin)

    All biopotential amplifiers must have high input impedance

    minimize loading (remember the characteristics of biopotential

    electrodes resulting into loading and distortion if input

    impedance of the amplifier is not high enough) typical values

    of Zin over the frequency range of the measurand = 10 M(remember the loading rule)

  • 8/11/2019 Biopot. Amp 1

    2/17

  • 8/11/2019 Biopot. Amp 1

    3/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    3

    Bandwidth (BW)

    Frequency responserequirements

    The biopotential amplifier must be sensitive to important

    frequency components of the biosignal

    Since biopotentials are low level signals, it is important to limitbandwidthoptimize signal-to-noise ratio

    Gain (G)

    Biopotential amplifiers have a gain of 1000 or greater

    Mode of Operation

    Very frequently biosignals are obtained from bipolar electrodes

    Electrodes symmetrically located with respect to groundneed

    differential amplification

    High CMRRrequired because:

    1.

    Common mode signals much greater than the biosignal appear

    on bipolar electrodes

    2.

    Symmetry with respect to ground is not perfect (mismatch

    between electrode impedances) more on this later

    Calibration Signal

    Medical and clinical equipment require quick calibration

    The gain of the biopotential amplifier must be calibrated to

    provide us with an accurate indication of the signals amplitude

    Push button to apply standard signal to the input of the

    biopotential amplifier

    Adjustable gain switch carefully selects calibrated fixed gains

    (in microprocessorbased systems, gain adjustment can be

    achieved by software setting)

  • 8/11/2019 Biopot. Amp 1

    4/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    4

    Electrocardiography

    A very widely used medical instrument, which is utilized todiagnose and monitor cardiac beat abnormalities is the

    electrocardiograph

    It measures the electrical activity of the heart (more preciselybiopotential differences arising from the electrical activity ofmyocardium). Weve already talked about the genesis of the

    ECG signal.

    The ECG machine uses surface electrodes and high input

    impedance

    Differential amplifiers with good common mode rejection ratio

    to record the electrocardiogram

    Normal ECG amplitude ranges between 0.5-4 mV. Normalfrequency content of ECG (for diagnostic purposes) is 0.05-100Hz. A typical ECG waveform is shown below:

    Obviously all human hearts are not the same and this leads into

    variability in different parts of the ECG signalSignificant diagnostic features of the ECG signal are:

    Duration of component parts of the signal

    Polarities and magnitudes

    The details of the ECG signal and the degree of variability in differentparts of the ECG signal is shown below:

  • 8/11/2019 Biopot. Amp 1

    5/17

  • 8/11/2019 Biopot. Amp 1

    6/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    6

    Some ECG waveform abnormalities that may indicate illness are:

    An extended PR interval may be diagnosed as AV node block

    A widening of the QRS complex may indicate conduction

    problems in the bundle of His

    An elevated ST segment may indicate occurrence of myocardial

    Infarction (MI)

    A negative polarity in the T wave may be due to coronaryinsufficiency

    ECG Leads

    Normal ECG recordings for the standard lead connections

    leads I, II and III(Lead IIprovides the strongest signal)

  • 8/11/2019 Biopot. Amp 1

    7/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    7

    Obviously, all human hearts are not the same and this resultsinto a high degree of variability

    Note the degree of variability of different parts of the ECG

    signal

    Some abnormalities that may indicate illness:

    An extended P-R interval may be diagnosed as AV node

    block

    Widening of the QRS complex conduction problems in the

    bundle of His

    Elevated ST segment may indicate occurrence of MI

    Negative polarity T wave may be due to coronary insufficiency

    QRS amplitude, polarity, time domain, PR interval (indicator of

    heat beat per min. & T-wave amplitude are some very important

    distinctive features.

    1.

    Loss of P-wave (un-coordinated twitching)

    2.

    Atria undergoing very rapid regular flapping movement

    3.

    An irritable focus discharge in the ventricle (Extrasystole)

    4.Loss of QRS blood circulation, severely impaired lots of ectopic

    foci

  • 8/11/2019 Biopot. Amp 1

    8/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    8

    Origin of the ECG signal

    We have already covered this concept extensively in theprevious lectures (The Dipole filedof the heart, the Eindhovens

    Triangle , the electrical circuit modelfor the electrocardiographic

    problem, etc.)

    Standard Limb Leads (I, II, III)

    The lead wires are color-coded according to some conventions.One example is: White RA(Right Arm), Black LA (Left Arm),

    Green RL (Right Leg), Red LL (Left Leg), and Brown C

    (Chest)Note:There is a CM(common mode) amplifier connected to the

    right leg. We will discuss this in detail later.

  • 8/11/2019 Biopot. Amp 1

    9/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    9

    Augmented Limb LeadsThese leads offer a free 50% increase over leads VR, VL, and VF

    connections (unipolar leads) with respect to Wilson terminal

    AVR= -I III/2, AVL= I II/2, aVF= II I/2

    Note:Each measurement is made from the reflected limb and the

    average of the other two limbs.

  • 8/11/2019 Biopot. Amp 1

    10/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    10

    Chest Leads (Precordial)

    Chest Lead Anatomical Positions

    V1- 4th

    intercostal space Right sternal margin

    V2- 4th

    intercostal space Left sternal margin

    V3 Midway between V2& V4V4 5

    thintercostal space on mid-clavicular line

    V5 Same as V4, on the anterior axillary line

    V6 Same as V5, on the mid-axillary line

  • 8/11/2019 Biopot. Amp 1

    11/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    11

    12 Lead Clinical Electrocardiography

  • 8/11/2019 Biopot. Amp 1

    12/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    12

    The ECG MachineMost representative Specs:

    Zin= 10 M

    Frequency response = 0.05 100 Hz

    Strip Chart Recorder Speed = 25 mm/sec.

    Fast Speed = 100 mm/sec.

    For detailed Specs. Refer to the Table in your text Summary of

    performance requirements for electrocardiographs

    Block Diagram of a modern electrocardiograph

    SensingElectrodes

    Lead-failDetector

    Amplifier

    ProtectionCircuit

    ALead

    Selector

    DrivenRight-legCircuit

    Right legElectrode

    IsolationCircuit

    Driver

    Amplifier

    Recorder

    Printer

    Auto-

    Calibration

    Baseline

    Restoration

    Isolated

    PowerSu l

    ADC

    Memory

    Microcomputer

    Control

    Program

    ECG AnalysisProgram

    Operator

    Display

    Keyboard

  • 8/11/2019 Biopot. Amp 1

    13/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    13

    Problems frequently encountered in electrocardiography

    There are many factors that must be considered in the design &application of the ECG machine

    These considerations should address the common problems inelectrocardiography

    The more common problems are:

    Frequency distortion

    Saturation or cut-off distortion

    Ground loops

    Open lead wires

    Artifact from large electric transients

    Interference from electric devices

    Other sources of electric interferenceLets go through these problems carefully and focus on some of

    their causes.

    Frequency distortion

    If the ECG machine does not meet the frequency response

    requirements for amplification of the ECG signal, frequencydistortion occurs

    a. True ECG

    b. High frequency distortion due to inadequate gain of ECG amplifier at highfrequencies

  • 8/11/2019 Biopot. Amp 1

    14/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    14

    c. Low frequency distortion due to inadequate gain of ECG amplifier at lowfrequencies

    Note:

    In high frequency distortion, the corners are rounded off and the

    amplitude of the QRS is diminished In low frequency distortion, the baseline gets distorted andmonophasic waves appear to be more biphasic

    Saturation or cutoff distortion

    This type of distortion occurs if there are high offset voltages at

    the electrodes or the amplifiers in the ECG machine are improperly

    adjusted.

    a. Undistorted ECG recordingb. Clipping of peak of the ECG due to positive saturation effects in

    amplifier

    c. Clipping of lower voltages in the ECG due to negative saturation

    or cutoff effects in amplifier

    Note:

    In saturation distortion, the combination of the ECG signal and

    offset voltages drives the amplifier into saturation. As a result theQRS peaks are cutoff because the output of the amplifier can not

    exceed the saturation voltage

  • 8/11/2019 Biopot. Amp 1

    15/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    15

    In cutoff distortion, the lower portions of the ECG are cutoff(clipped).

    Ground loops

    Ground loops can arise when patients having their ECGs takenon either a clinical ECG machine or on a cardiac monitor are alsoconnected to other pieces of apparatus

    Each electric device has its own ground connection either

    through the power line or in some cases through a heavy groundwire attached to some ground point in the room

    Ground loop currents may present a safetyproblem

  • 8/11/2019 Biopot. Amp 1

    16/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    16

    They elevate the patients body potential to some voltageabove the lowest ground to which the instrumentation is attached

    (ground A in the figure)

    This potential produces common mode voltages that mayincrease the amount of interference seen if the CMRR of thedevice is not high enough

    In electrocardiographs the ground leads usually run alongside

    the ECG signal leads, magnetic fields caused by the groundingcircuit can induce small voltages in the signal lead wires

    interferenceon the tracing

    Ground loops represent closed current paths. They subtend some

    geometrical areas within the loop. If this area is large and is in a

    strong time varying magnetic filed, a current can be induced in the

    ground loop, the current flows through the patientand commonmode voltages are generated

    Open lead wires

    A frequently occurring problem in electrocardiography is one ofthe lead wires becoming disconnected as a result of rough handling

    This results in relatively high potentials being induced on the

    open lead wire due to electric fields emanating from the power

    lines or other electrical devices in the vicinity of the ECG machine

    This in turn leads to a big constant-amplitude deflection of the

    ECG pen at 50 or (60) Hz. Such an electrode is not making goodcontact with the patient. Therefore, this phenomena translates into

    having a circuit for detecting poor electrode contact

  • 8/11/2019 Biopot. Amp 1

    17/17

    Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1

    BI1 Lecture Series Prepared by Homer Nazeran PhD

    Please read carefully and check for typos!

    17

    Artifacts from large electric transients

    These transients occur during the following situations:

    Defibrillating the patient with a high-current high-voltageelectric pulse called the defibrillating pulse (more on this in BI)

    Motion artifacts

    The patient encountering a build-up of electric charges that canbe partially discharged through the body

    Switching leads manually in older ECG machines (not aproblem in new machines)

    Note: These effects can be alleviated by reducing the source of the

    artifact.

    Include an electronic protection circuitry to limit max. Input

    voltage across ECG amp. To minimize saturation duringdefibrillation

    Reduce the build up of charge on personnel by having them use

    conductive clothing, shoes, etc and touch the bed frame beforetouching the patient