biopot. amp 1
TRANSCRIPT
-
8/11/2019 Biopot. Amp 1
1/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
1
Biopotential Amplifiers
These are very important part of modern medicalinstrumentation
We need to amplify biopotentials which are generated in the
body at low levels with a high source impedance
Biopotentials amplifiers are required to increase signal strengthwhile maintaining fidelity
Basic Requirements of Biopotential AmplifiersEssential functions of a bioamplifier are:
To take a weak biopotential and increase its amplitude so that it
can be processed, recorded or displayed
To amplify voltage, but it could be considered as a poweramplifier as well
To amplify current since in some cases a biopotential amplifier
is used to isolate the load from the sourcecurrent gain only
Input Impedance (Zin)
All biopotential amplifiers must have high input impedance
minimize loading (remember the characteristics of biopotential
electrodes resulting into loading and distortion if input
impedance of the amplifier is not high enough) typical values
of Zin over the frequency range of the measurand = 10 M(remember the loading rule)
-
8/11/2019 Biopot. Amp 1
2/17
-
8/11/2019 Biopot. Amp 1
3/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
3
Bandwidth (BW)
Frequency responserequirements
The biopotential amplifier must be sensitive to important
frequency components of the biosignal
Since biopotentials are low level signals, it is important to limitbandwidthoptimize signal-to-noise ratio
Gain (G)
Biopotential amplifiers have a gain of 1000 or greater
Mode of Operation
Very frequently biosignals are obtained from bipolar electrodes
Electrodes symmetrically located with respect to groundneed
differential amplification
High CMRRrequired because:
1.
Common mode signals much greater than the biosignal appear
on bipolar electrodes
2.
Symmetry with respect to ground is not perfect (mismatch
between electrode impedances) more on this later
Calibration Signal
Medical and clinical equipment require quick calibration
The gain of the biopotential amplifier must be calibrated to
provide us with an accurate indication of the signals amplitude
Push button to apply standard signal to the input of the
biopotential amplifier
Adjustable gain switch carefully selects calibrated fixed gains
(in microprocessorbased systems, gain adjustment can be
achieved by software setting)
-
8/11/2019 Biopot. Amp 1
4/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
4
Electrocardiography
A very widely used medical instrument, which is utilized todiagnose and monitor cardiac beat abnormalities is the
electrocardiograph
It measures the electrical activity of the heart (more preciselybiopotential differences arising from the electrical activity ofmyocardium). Weve already talked about the genesis of the
ECG signal.
The ECG machine uses surface electrodes and high input
impedance
Differential amplifiers with good common mode rejection ratio
to record the electrocardiogram
Normal ECG amplitude ranges between 0.5-4 mV. Normalfrequency content of ECG (for diagnostic purposes) is 0.05-100Hz. A typical ECG waveform is shown below:
Obviously all human hearts are not the same and this leads into
variability in different parts of the ECG signalSignificant diagnostic features of the ECG signal are:
Duration of component parts of the signal
Polarities and magnitudes
The details of the ECG signal and the degree of variability in differentparts of the ECG signal is shown below:
-
8/11/2019 Biopot. Amp 1
5/17
-
8/11/2019 Biopot. Amp 1
6/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
6
Some ECG waveform abnormalities that may indicate illness are:
An extended PR interval may be diagnosed as AV node block
A widening of the QRS complex may indicate conduction
problems in the bundle of His
An elevated ST segment may indicate occurrence of myocardial
Infarction (MI)
A negative polarity in the T wave may be due to coronaryinsufficiency
ECG Leads
Normal ECG recordings for the standard lead connections
leads I, II and III(Lead IIprovides the strongest signal)
-
8/11/2019 Biopot. Amp 1
7/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
7
Obviously, all human hearts are not the same and this resultsinto a high degree of variability
Note the degree of variability of different parts of the ECG
signal
Some abnormalities that may indicate illness:
An extended P-R interval may be diagnosed as AV node
block
Widening of the QRS complex conduction problems in the
bundle of His
Elevated ST segment may indicate occurrence of MI
Negative polarity T wave may be due to coronary insufficiency
QRS amplitude, polarity, time domain, PR interval (indicator of
heat beat per min. & T-wave amplitude are some very important
distinctive features.
1.
Loss of P-wave (un-coordinated twitching)
2.
Atria undergoing very rapid regular flapping movement
3.
An irritable focus discharge in the ventricle (Extrasystole)
4.Loss of QRS blood circulation, severely impaired lots of ectopic
foci
-
8/11/2019 Biopot. Amp 1
8/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
8
Origin of the ECG signal
We have already covered this concept extensively in theprevious lectures (The Dipole filedof the heart, the Eindhovens
Triangle , the electrical circuit modelfor the electrocardiographic
problem, etc.)
Standard Limb Leads (I, II, III)
The lead wires are color-coded according to some conventions.One example is: White RA(Right Arm), Black LA (Left Arm),
Green RL (Right Leg), Red LL (Left Leg), and Brown C
(Chest)Note:There is a CM(common mode) amplifier connected to the
right leg. We will discuss this in detail later.
-
8/11/2019 Biopot. Amp 1
9/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
9
Augmented Limb LeadsThese leads offer a free 50% increase over leads VR, VL, and VF
connections (unipolar leads) with respect to Wilson terminal
AVR= -I III/2, AVL= I II/2, aVF= II I/2
Note:Each measurement is made from the reflected limb and the
average of the other two limbs.
-
8/11/2019 Biopot. Amp 1
10/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
10
Chest Leads (Precordial)
Chest Lead Anatomical Positions
V1- 4th
intercostal space Right sternal margin
V2- 4th
intercostal space Left sternal margin
V3 Midway between V2& V4V4 5
thintercostal space on mid-clavicular line
V5 Same as V4, on the anterior axillary line
V6 Same as V5, on the mid-axillary line
-
8/11/2019 Biopot. Amp 1
11/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
11
12 Lead Clinical Electrocardiography
-
8/11/2019 Biopot. Amp 1
12/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
12
The ECG MachineMost representative Specs:
Zin= 10 M
Frequency response = 0.05 100 Hz
Strip Chart Recorder Speed = 25 mm/sec.
Fast Speed = 100 mm/sec.
For detailed Specs. Refer to the Table in your text Summary of
performance requirements for electrocardiographs
Block Diagram of a modern electrocardiograph
SensingElectrodes
Lead-failDetector
Amplifier
ProtectionCircuit
ALead
Selector
DrivenRight-legCircuit
Right legElectrode
IsolationCircuit
Driver
Amplifier
Recorder
Printer
Auto-
Calibration
Baseline
Restoration
Isolated
PowerSu l
ADC
Memory
Microcomputer
Control
Program
ECG AnalysisProgram
Operator
Display
Keyboard
-
8/11/2019 Biopot. Amp 1
13/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
13
Problems frequently encountered in electrocardiography
There are many factors that must be considered in the design &application of the ECG machine
These considerations should address the common problems inelectrocardiography
The more common problems are:
Frequency distortion
Saturation or cut-off distortion
Ground loops
Open lead wires
Artifact from large electric transients
Interference from electric devices
Other sources of electric interferenceLets go through these problems carefully and focus on some of
their causes.
Frequency distortion
If the ECG machine does not meet the frequency response
requirements for amplification of the ECG signal, frequencydistortion occurs
a. True ECG
b. High frequency distortion due to inadequate gain of ECG amplifier at highfrequencies
-
8/11/2019 Biopot. Amp 1
14/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
14
c. Low frequency distortion due to inadequate gain of ECG amplifier at lowfrequencies
Note:
In high frequency distortion, the corners are rounded off and the
amplitude of the QRS is diminished In low frequency distortion, the baseline gets distorted andmonophasic waves appear to be more biphasic
Saturation or cutoff distortion
This type of distortion occurs if there are high offset voltages at
the electrodes or the amplifiers in the ECG machine are improperly
adjusted.
a. Undistorted ECG recordingb. Clipping of peak of the ECG due to positive saturation effects in
amplifier
c. Clipping of lower voltages in the ECG due to negative saturation
or cutoff effects in amplifier
Note:
In saturation distortion, the combination of the ECG signal and
offset voltages drives the amplifier into saturation. As a result theQRS peaks are cutoff because the output of the amplifier can not
exceed the saturation voltage
-
8/11/2019 Biopot. Amp 1
15/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
15
In cutoff distortion, the lower portions of the ECG are cutoff(clipped).
Ground loops
Ground loops can arise when patients having their ECGs takenon either a clinical ECG machine or on a cardiac monitor are alsoconnected to other pieces of apparatus
Each electric device has its own ground connection either
through the power line or in some cases through a heavy groundwire attached to some ground point in the room
Ground loop currents may present a safetyproblem
-
8/11/2019 Biopot. Amp 1
16/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
16
They elevate the patients body potential to some voltageabove the lowest ground to which the instrumentation is attached
(ground A in the figure)
This potential produces common mode voltages that mayincrease the amount of interference seen if the CMRR of thedevice is not high enough
In electrocardiographs the ground leads usually run alongside
the ECG signal leads, magnetic fields caused by the groundingcircuit can induce small voltages in the signal lead wires
interferenceon the tracing
Ground loops represent closed current paths. They subtend some
geometrical areas within the loop. If this area is large and is in a
strong time varying magnetic filed, a current can be induced in the
ground loop, the current flows through the patientand commonmode voltages are generated
Open lead wires
A frequently occurring problem in electrocardiography is one ofthe lead wires becoming disconnected as a result of rough handling
This results in relatively high potentials being induced on the
open lead wire due to electric fields emanating from the power
lines or other electrical devices in the vicinity of the ECG machine
This in turn leads to a big constant-amplitude deflection of the
ECG pen at 50 or (60) Hz. Such an electrode is not making goodcontact with the patient. Therefore, this phenomena translates into
having a circuit for detecting poor electrode contact
-
8/11/2019 Biopot. Amp 1
17/17
Department of Electrical and Computer Engineering UTEPEE 4385 Biomedical Instrumentation 1
BI1 Lecture Series Prepared by Homer Nazeran PhD
Please read carefully and check for typos!
17
Artifacts from large electric transients
These transients occur during the following situations:
Defibrillating the patient with a high-current high-voltageelectric pulse called the defibrillating pulse (more on this in BI)
Motion artifacts
The patient encountering a build-up of electric charges that canbe partially discharged through the body
Switching leads manually in older ECG machines (not aproblem in new machines)
Note: These effects can be alleviated by reducing the source of the
artifact.
Include an electronic protection circuitry to limit max. Input
voltage across ECG amp. To minimize saturation duringdefibrillation
Reduce the build up of charge on personnel by having them use
conductive clothing, shoes, etc and touch the bed frame beforetouching the patient