BRITISH STANDARD BS EN ISO 10993-14:2001 Incorporating Corrigendum No. 1 Biological evaluation of medical devices — Part 14: Identification and quantification of degradation products from ceramics The European Standard EN ISO 10993-14:2001 has the status of a British Standard ICS 11.100 NO COPYING WITHOUT BSI PERMISSION EXCEPT AS PERMITTED BY COPYRIGHT LAW

Upload: lybao

Post on 17-Mar-2018

216 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

BRITISH STANDARD BS EN ISO 10993-14:2001Incorporating Corrigendum No. 1

Biological evaluation of medical devices —

Part 14: Identification and quantification of degradation products from ceramics

The European Standard EN ISO 10993-14:2001 has the status of a British Standard

ICS 11.100

NO COPYING WITHOUT BSI PERMISSION EXCEPT AS PERMITTED BY COPYRIGHT LAW

Page 2: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

BS EN ISO 10993-14:2001

This British Standard, having been prepared under the direction of the Health and Environment Sector Policy and Strategy Committee, was published under the authority of the Standards Policy and Strategy Committee on 14 December 2001

© BSI 15 March 2002

ISBN 0 580 38774 7

National foreword

This British Standard is the official English language version of EN ISO 10993-14:2001. It is identical with ISO 10993-14:2001.

The UK participation in its preparation was entrusted to Technical Committee CH/26, Biological Testing of Medical and Dental Materials and Devices, which has the responsibility to:

A list of organizations represented on this committee can be obtained on request to its secretary.

Cross-referencesThe British Standards which implement international or European publications referred to in this document may be found in the BSI Standards Catalogue under the section entitled “International Standards Correspondence Index”, or by using the “Find” facility of the BSI Standards Electronic Catalogue.A British Standard does not purport to include all the necessary provisions of a contract. Users of British Standards are responsible for their correct application.

Compliance with a British Standard does not of itself confer immunity from legal obligations.

— aid enquirers to understand the text;

— present to the responsible international/European committee any enquiries on the interpretation, or proposals for change, and keep the UK interests informed;

— monitor related international and European developments and promulgate them in the UK.

Summary of pagesThis document comprises a front cover, an inside front cover, the EN ISO title page, the EN ISO foreword page, the ISO title page, pages ii to vi, pages 1 to 11, a blank page, two Annex ZA pages, an Annex ZB page and a back cover.

The BSI copyright date displayed in this document indicates when the document was last issued.

Amendments issued since publication

Amd. No. Date Comments

13748 Corr. No. 1

15 March 2002

Incorporation of Annex ZA and Annex ZB

Page 3: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

EUROPEAN STANDARD

NORME EUROPÉENNE

EUROPÄISCHE NORM

EN ISO 10993-14

November 2001

ICS 11.100

English version

Biological evaluation of medical devices - Part 14: Identificationand quantification of degradation products from ceramics (ISO

10993-14:2001)

Evaluation biologique des dispositifs médicaux - Partie 14:Identification et quantification des produits de dégradation

des céramiques (ISO 10993-14:2001)

Biologische Beurteilung von Medizinprodukten - Teil 14:Qualitativer und quantitativer Nachweis von keramischen

Abbauprodukten (ISO 10993-14:2001)

This European Standard was approved by CEN on 14 October 2001.

CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this EuropeanStandard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such nationalstandards may be obtained on application to the Management Centre or to any CEN member.

This European Standard exists in three official versions (English, French, German). A version in any other language made by translationunder the responsibility of a CEN member into its own language and notified to the Management Centre has the same status as the officialversions.

CEN members are the national standards bodies of Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece,Iceland, Ireland, Italy, Luxembourg, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland and United Kingdom.

EUROPEAN COMMITTEE FOR STANDARDIZATIONC OM ITÉ EUR OP ÉEN DE NOR M ALIS AT IONEUROPÄISCHES KOMITEE FÜR NORMUNG

Management Centre: rue de Stassart, 36 B-1050 Brussels

© 2001 CEN All rights of exploitation in any form and by any means reservedworldwide for CEN national Members.

Ref. No. EN ISO 10993-14:2001 E

Page 4: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

I NESO 01-39914:20( 10E)

2

CORRECTED 2002-02-06

Foreword

This document (ISO 10993-14:2001) has been prepared by Technical Committee ISO/TC 194"Biological evaluation of medical devices" in collaboration with Technical Committee CEN/TC206 "Biocompatibility of medical and dental materials and devices", the secretariat of which isheld by NEN.

This European Standard shall be given the status of a national standard, either by publication ofan identical text or by endorsement, at the latest by May 2002, and conflicting national standardsshall be withdrawn at the latest by May 2002.

This document has been prepared under a mandate given to CEN by the European Commissionand the European Free Trade Association, and supports essential requirements of EUDirective(s).

For relationship with EU Directive(s), see informative Annex ZA, which is an integral part of thisdocument.

According to the CEN/CENELEC Internal Regulations, the national standards organizations ofthe following countries are bound to implement this European Standard: Austria, Belgium, CzechRepublic, Denmark, Finland, France, Germany, Greece, Iceland, Ireland, Italy, Luxembourg,Netherlands, Norway, Portugal, Spain, Sweden, Switzerland and the United Kingdom.

Endorsement notice

The text of the International Standard ISO 10993-14:2001 has been approved by CEN as aEuropean Standard without any modifications.

NOTE Normative references to International Standards are listed in annex ZB (normative).

Page 2EN ISO 10993−14:2001

Page 5: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

Reference numberISO 10993-14:2001(E)

INTERNATIONALSTANDARD

ISO10993-14

First edition2001-11-15

Biological evaluation of medical devices —

Part 14:Identification and quantification ofdegradation products from ceramics

Évaluation biologique des dispositifs médicaux —

Partie 14: Identification et quantification des produits de dégradation descéramiques

EN ISO 10993−14:2001

Page 6: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

ii

EN ISO 10993−14:2001

Page 7: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

iii

Contents Page

Foreword.....................................................................................................................................................................iv

Introduction ................................................................................................................................................................vi

1 Scope ..............................................................................................................................................................1

2 Normative references ....................................................................................................................................1

3 Terms and definitions ...................................................................................................................................2

4 Test procedures.............................................................................................................................................24.1 Principle..........................................................................................................................................................24.2 Testing of dental devices..............................................................................................................................24.3 General testing techniques...........................................................................................................................34.4 Extreme solution test ....................................................................................................................................44.5 Simulation solution test ................................................................................................................................6

5 Analysis of filtrate..........................................................................................................................................95.1 General............................................................................................................................................................95.2 Choice of chemicals or elements to be analysed.......................................................................................95.3 Sensitivity of the analysis method...............................................................................................................9

6 Test report ......................................................................................................................................................9

Bibliography ..............................................................................................................................................................11

EN ISO 10993−14:2001

Page 8: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

iv

Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISOmember bodies). The work of preparing International Standards is normally carried out through ISO technicalcommittees. Each member body interested in a subject for which a technical committee has been established hasthe right to be represented on that committee. International organizations, governmental and non-governmental, inliaison with ISO, also take part in the work. ISO collaborates closely with the International ElectrotechnicalCommission (IEC) on all matters of electrotechnical standardization.

International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 3.

Draft International Standards adopted by the technical committees are circulated to the member bodies for voting.Publication as an International Standard requires approval by at least 75 % of the member bodies casting a vote.

Attention is drawn to the possibility that some of the elements of this part of ISO 10993 may be the subject ofpatent rights. ISO shall not be held responsible for identifying any or all such patent rights.

International Standard ISO 10993-14 was prepared by Technical Committee ISO/TC 194, Biological evaluation ofmedical devices.

ISO 10993 consists of the following parts, under the general title Biological evaluation of medical devices:

— Part 1: Evaluation and testing

— Part 2: Animal welfare requirements

— Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity

— Part 4: Selection of tests for interactions with blood

— Part 5: Tests for in vitro cytotoxicity

— Part 6: Tests for local effects after implantation

— Part 7: Ethylene oxide sterilization residuals

— Part 8: Selection and qualification of reference materials for biological tests

— Part 9: Framework for identification and quantification of potential degradation products

— Part 10: Tests for irritation and delayed-type hypersensitivity

— Part 11: Tests for systemic toxicity

— Part 12: Sample preparation and reference materials

— Part 13: Identification and quantification of degradation products from polymeric medical devices

— Part 14: Identification and quantification of degradation products from ceramics

— Part 15: Identification and quantification of degradation products from metals and alloys

— Part 16: Toxicokinetic study design for degradation products and leachables

EN ISO 10993−14:2001

Page 9: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

v

— Part 17: Establishment of allowable limits for leachable substances using health-based risk assessment

— Part 18: Chemical characterization of materials

EN ISO 10993−14:2001

Page 10: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

vi

Introduction

This part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solutiontest and a simulation solution test. The extreme solution test is developed as a worst-case environment and thesimulation test is developed as a very common environment.

Degradation products covered by this part of ISO 10993 are formed primarily by dissolution in an aqueousenvironment. It is recognized that additional biological factors such as enzymes and proteins can alter the rate ofdegradation. Degradation by such outside factors is not addressed in this part of ISO 10993.

It should be kept in mind that a ceramic device might have extraneous chemical phases and/or elements inextremely minor amounts. Whilst these components might not be named in the original specification, they can oftenbe suspected by the relationship that the material in question has to other materials and the expected history of thematerial’s processing.

Once identified and quantified, the chemical composition of the degradation products form the basis for riskassessment and, if appropriate, biological safety studies according to the principles of ISO 10993-1.

EN ISO 10993−14:2001

Page 11: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

INTENRATIONAL TSANDADR ISO -3990114:(1002E)

© ISO 1002 – All rights rsedevre 1

Biological evaluation of medical devices —

Part 14:Identification and quantification of degradation products fromceramics

1 Scope

This part of ISO 10993 specifies two methods of obtaining solutions of degradation products from ceramics(including glasses) for the purposes of quantification. It also gives guidance on the analysis of these solutions inorder to identify the degradation products. Because of the generalized nature of this part of ISO 10993, productspecific standards, when available, that address degradation product formation under more relevant conditions ofuse, should be considered first.

This part of ISO 10993 considers only those degradation products generated by a chemical dissociation ofceramics during in vitro testing. No degradation induced by mechanical stress or external energy is covered. It isnoted that while ISO 6872 and ISO 9693 cover chemical degradation tests, they do not address the analysis ofdegradation products.

Because of the range of ceramics used in medical devices and the different requirements for accuracy andprecision of the results, no specific analytical techniques are identified. Further, this part of ISO 10993 provides nospecific requirements for acceptable levels of degradation products.

Although these materials are intended for biomedical applications, the biological activity of these degradationproducts is not addressed in this part of ISO 10993.

2 Normative references

The following normative documents contain provisions which, through reference in this text, constitute provisions ofthis part of ISO 10993. For dated references, subsequent amendments to, or revisions of, any of these publicationsdo not apply. However, parties to agreements based on this part of ISO 10993 are encouraged to investigate thepossibility of applying the most recent editions of the normative documents indicated below. For undatedreferences, the latest edition of the normative document referred to applies. Members of ISO and IEC maintainregisters of currently valid International Standards.

ISO 3310-1, Test sieves — Technical requirements and testing — Part 1: Test sieves of metal wire cloth

ISO 3696, Water for analytical laboratory use — Specification and test methods

ISO 5017, Dense shaped refractory products — Determination of bulk density, apparent porosity and true porosity

ISO 6474, Implants for surgery — Ceramic materials based on high purity alumina

ISO 6872:1995, Dental ceramic

ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing

EN ISO 10993−14:2001

1

Page 12: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

ISO -3990141:(1002E)

2 © ISO 1002 – All rights rsedevre

ISO 10993-9, Biological evaluation of medical devices — Part 9: Framework for identification and quantification ofpotential degradation products

3 Terms and definitions

For the purposes of this part of ISO 10993, the terms and definitions given in ISO 10993-1 and ISO 10993-9 as wellas the following apply.

3.1ceramicstypically crystallized materials that are physically nonmetallic and chemically inorganic

3.2blank discnoncoated circular plate made of the substrate material to be used in the finished device

3.3retentateundissolved solids remaining in the filter paper after filtration

3.4filtratesolution which passes through the filter paper

4 Test procedures

4.1 Principle

This part of ISO 10993 consists of two tests. The first test, an extreme solution test conducted at low pH, serves asa screen for most ceramics for the observation of possible degradation products. The second test simulates a morefrequently encountered in vivo pH. A flowchart of the decision process for using these test methods is given inFigure 1.

The test methods described in this part of ISO 10993 shall be used for ceramics in bulk and granular form as wellas ceramic coatings.

When deviations from the recommended test specimen or solution volumes are used, full justification shall beprovided.

4.2 Testing of dental devices

4.2.1 General

This part of ISO 10993 is intended to simulate worst-case exposure to tissue environments. For dental ceramicsexposed to the oral cavity (e.g. ceramic veneering material), a more appropriate test environment is given inISO 6872. However, for dental devices not exposed to the oral cavity, such as dental implant stems, thespecifications given in 4.4 of this part of ISO 10993 shall apply.

4.2.2 Test methods for dental devices exposed to the oral cavity

For dental devices exposed to the oral cavity, the method given in 8.4 of ISO 6872:1995 shall be used as theextreme solution test.

EN ISO 10993−14:2001

2

Page 13: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

ISO 1-399041:(1002E)

© ISO 1002 – All rights rsedevre 3

Figure 1 — Flowchart of the decision-making process for the extreme solution test and the simulationsolution tests (see text for details)

4.2.3 Specimen characterization

The specimen shall be characterized as described in 4.4.4. If the specimen density is greater than 99 % of thetheoretical maximum density, and the specimen has an average surface roughness (Ra) of less than 5 µm, thesurface area may be calculated by direct geometrical measurement.

Low surface roughness is required for geometrical measurement in order to avoid grossly underestimating thesurface area.

4.2.4 Analysis

The filtrate for analysis shall be separated from the retentate as described in 4.4.7.6 to 4.4.7.11.

4.3 General testing techniques

4.3.1 Mass determination

Mass shall be determined using a balance with an accuracy of no less than 0,000 5 g. All mass determinationsshall have 6 replicates.

EN ISO 10993−14:2001

3

Page 14: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

ISO -3990141:(1002E)

4 © ISO 1002 – All rights rsedevre

4.3.2 Drying techniques

Drying in an oven at a temperature of (100 � 2) °C shall continue until a mass change of � 0,1 % occurs betweenmass determinations. This is normally accomplished by drying overnight and weighing at 2-h intervals the next day.

4.4 Extreme solution test

4.4.1 Principle

The extreme solution test is a test based on a low pH citric acid buffer solution. The pH value of 3 is defined hereas a worst-case low-end service environment. For devices exposed to an environment where the pH is lower than3, an alternative lower pH solution shall be used and justification shall be provided. In the event of a chemicalreaction between the extreme solution and the test specimen, an alternative extreme test at similar pH shall bejustified and performed.

4.4.2 Application range

This test is applicable to all ceramics. It should be noted that the mechanisms of degradation may not be the samefor all materials at low pH as they are at blood pH (approximately pH 7,35 to pH 7,45). Nonetheless, as an extremecondition for the production of possible degradation products, this severe test can serve as a screen for mostmaterials.

It is expected that materials will dissolve up to their solubility limit in the solution. To accelerate the test to thesolubility limit endpoint, the test is carried out on a granulated specimen (see 4.4.3.3).

4.4.3 Extreme test sample preparation

4.4.3.1 Specimen configuration

Specimens shall be granulated from a specimen manufactured according to the method intended for material use.If the specimen is a ceramic coating, it shall be removed from the substrate material and granulated to anappropriate size. Under some circumstances (e.g. thin coatings), insufficient ceramic material is available toperform the extreme test. In these cases a scaled-down test may be used in which a sample may be preparedusing the ratio of 1 g per 20 ml of test solution. When this is done, the precision and accuracy of the massdetermination shall be appropriately scaled and justified in order to accommodate the alternative sample size.

4.4.3.2 Granulation

Granulation shall be accomplished by grinding with a tungsten carbide mortar and pestle.

4.4.3.3 Sizing

The granulated specimen shall pass through a 400 µm screen but be retained on a 315 µm screen using a dryscreen method such as that described in ISO 3310-1.

If it is not possible to produce granules of this size (e.g. due to the grinding of a coating), granules of a size smallerthan that defined in this part of ISO 10993 shall be used, and the size shall be reported.

NOTE The use of a particle size smaller than that specified in this clause is likely to lead to increased dissolution andtherefore is not expected to reduce the yield of dissolution products and not expected to compromise risk analysis for biologicalsafety.

4.4.3.4 Specimen preparation

The mass of starting material is dependent upon the solubility of the material as determined by the solubilitycharacterization in 4.4.4.3:

EN ISO 10993−14:2001

4

Page 15: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

ISO 1-399041:(1002E)

© ISO 1002 – All rights rsedevre 5

� for low solubility granulated material, (5,00 � 0,05) g shall be used;

� for high solubility granulated material, (10,00 � 0,05) g shall be used.

4.4.4 Extreme test sample characterization

4.4.4.1 Surface area characterization

The specimen shall be characterized by gas adsorption in accordance with an appropriate method, e.g. such asthose given in ASTM D4780.

4.4.4.2 Density

The specimen shall be characterized for density in accordance with ISO 5017.

4.4.4.3 Solubility characterization

From information about the material available from the producer or other sources, the materials shall becharacterized as "high" or "low" solubility materials in the following manner.

� Consult Figure 1 for a flowchart of the decision-making process.

� If (5,00 � 0,05) g of the material are expected to totally dissolve in 100 ml during testing as described in 4.4.7.1to 4.4.7.5, the material shall be considered as high solubility.

� If (5,00 � 0,05) g of the material are not expected to totally dissolve in 100 ml, the material shall be consideredas low solubility.

� If the information is unavailable, the material shall be considered as high solubility material.

4.4.4.4 Microstructural and X-ray characterization

X-ray diffraction shall be performed with an X-ray diffractometer. The 2� resolution and reproducibility shall bebetter than 0,02°. Microstructure analysis shall conform with that specified in ISO 6474.

4.4.5 Test equipment

4.4.5.1 Test container

A 250 ml polypropylene or high-density polyethylene container shall be used. A fresh specimen container shall beused for each test. Glass containers shall not be used since they may contaminate test solutions.

4.4.5.2 Büchner funnel

A Büchner or similar type funnel fitted appropriately to retain undissolved particles shall be used.

4.4.6 Citric acid buffer solution

The buffered citric acid solution shall be freshly prepared and have a pH of 3,0 � 0,2 at a temperature of(37 � 1) °C. The solution shall be prepared as follows:

Dissolve 21 g of citric acid monohydrate in 500 ml water (ISO 3696, grade 2) in a 1 000 ml volumetric flask. Add200 ml of 1 mol/l sodium hydroxide solution and dilute to the mark with water (ISO 3696, grade 2). Mix 40,4 ml ofthis solution with 59,6 ml of 0,1 mol/l hydrochloric acid yielding the buffered citric acid solution.

EN ISO 10993−14:2001

5

Page 16: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

ISO -3990141:(1002E)

6 © ISO 1002 – All rights rsedevre

4.4.7 Test procedure

4.4.7.1 Weigh the container without the top.

4.4.7.2 Weigh the container and specimen. Report the difference in mass between the container withspecimen and the container without specimen as the mass of the specimen.

4.4.7.3 Add (100 � 1) ml of buffered citric acid. Care should be taken to ensure that all of the specimen is incontact with the solution.

4.4.7.4 Place the container with specimen in a controlled-temperature environment at (37 � 1) °C for(120 � 1) h. The container shall be agitated at 2 Hz using a longitudinal or circular movement. If the test specimenis totally dissolved before 120 h, terminate the test and note the time in the test report.

4.4.7.5 Remove the container and specimen and allow them to cool to room temperature.

4.4.7.6 Weigh the filtering medium (e.g. filter paper) to determine its mass without retentate.

4.4.7.7 Remove the specimen via filtration and retain the filtrate for analysis. Filtrate should not be stored inglass containers.

4.4.7.8 Rinse the filtering medium and retentate three times with small amounts of water (ISO 3696, grade 2)to remove the citric acid buffer.

4.4.7.9 Dry the specimen and filtering medium with retentate to a constant mass (see 4.3.2).

4.4.7.10 Weigh the filtering medium with retentate. The difference in mass between the filtering medium withand without retentate is the mass of the retentate.

4.4.7.11 The difference between the mass of the specimen and the mass of retentate is the mass of thedissolved material.

4.5 Simulation solution test

4.5.1 Principle

The simulation test is based on a buffer solution of pH 7,4 � 0,1 as defined in 4.5.6. This will simulate the body’snormal pH level.

4.5.2 Application range

This test is applicable to all ceramics.

NOTE The mechanism of degradation in this test may not be the same as in the extreme test.

4.5.3 Simulation test specimen configuration

4.5.3.1 Coated ceramics

4.5.3.1.1 Blank discs

Test specimens shall be prepared as coatings on blank discs.

Blank discs shall be of diameter (36 � 1) mm and thickness of (2 � 0,1) mm using the same substrate material andpreparation techniques as in the finished device.

EN ISO 10993−14:2001

6

Page 17: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

ISO 1-399041:(1002E)

© ISO 1002 – All rights rsedevre 7

4.5.3.1.2 Coated discs

Blank discs shall be coated on all sides using coating techniques that are used in the production of the finisheddevice.

NOTE Because of the reduced surface area to volume ratio, the sensitivity of the test will be reduced using this method.

4.5.3.2 All other ceramics

Test specimens shall be granulated using the methods described in 4.4.3.2 and 4.4.3.3 from a specimenmanufactured by methods used to produce the finished device.

4.5.4 Simulation test sample characterization

4.5.4.1 General

For coated samples, surface area, microstructure and X-ray characterization shall be recorded. For all otherceramics, density, surface area, microstructure and X-ray characterization shall be recorded.

4.5.4.2 Density

The specimen shall be characterized for density in accordance with ISO 5017.

4.5.4.3 Microstructural and X-ray characterization

X-ray diffraction shall be performed with an X-ray diffractometer. The resolution and reproducibility shall be betterthan 0,02°. Microstructure analysis shall conform with that specified in ISO 6474.

4.5.4.4 Surface area characterization

The specimen shall be characterized by gas adsorption in accordance with an appropriate method, e.g. such asthose given in ASTM D4780.

4.5.5 Test equipment

4.5.5.1 Test container

A 250 ml polypropylene or high-density polyethylene container shall be used. A fresh specimen container shall beused for each test. Glass containers shall not be used since they may contaminate test solutions.

4.5.5.2 Büchner funnel

A Büchner or similar type funnel fitted appropriately to retain undissolved particles shall be used.

4.5.6 Buffer solution

The solution shall be freshly prepared TRIS-HCI buffer. It shall be prepared by dissolving 13,25 g oftris(hydroxymethyl)aminomethane in 500 ml of water (ISO 3696, grade 2). Adjust the pH with an appropriateamount of 1 mol/l hydrochloric acid to pH 7,4 � 0,1 at a temperature of (37 � 1) °C. Make up to 1 000 ml with water(ISO 3696, grade 2).

4.5.7 Coated disc test procedure

4.5.7.1 General

Both the coated and uncoated discs are exposed to the simulation test solution in order to determine whetherdegradation products are generated under simulated test conditions.

EN ISO 10993−14:2001

7

Page 18: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

ISO -3990141:(1002E)

8 © ISO 1002 – All rights rsedevre

4.5.7.2 Blank disc test

4.5.7.2.1 Place the blank disc in the test container for the exposure test.

4.5.7.2.2 Add (100 � 1) ml of buffer solution to the container with the blank disc. Care shall be taken to ensurethat the entire blank disc is in contact with the solution.

4.5.7.2.3 Maintain the container with blank disc at (37 � 1) °C in a controlled-temperature chamber for(120 � 1) h. The container shall be agitated at 2 Hz using a longitudinal or circular movement.

4.5.7.2.4 Remove the container with specimen and allow them to reach room temperature.

4.5.7.2.5 Filter the solution and retain the filtrate for analysis (see clause 5).

4.5.7.3 Coated disc test

4.5.7.3.1 Determine the mass of the ceramic coating by subtracting the mass of the coated disc from the massof the blank disc for each test specimen.

Each disc shall be weighed before and after coating to determine the mass of the coating.

4.5.7.3.2 Place the coated disc in a test container for the exposure test.

4.5.7.3.3 Add (100 � 1) ml of buffer solution to the container with the coated disc. Care shall be taken to ensurethat the entire coated disk is in contact with the solution.

4.5.7.3.4 Maintain the container with coated disc at (37 � 1) °C in a controlled-temperature chamber for(120 � 1) h. The container shall be agitated at 2 Hz using a longitudinal or circular movement.

4.5.7.3.5 Remove the container with specimen and allow them to cool to room temperature.

4.5.7.3.6 Weigh the filtering medium (e.g. filter paper).

4.5.7.3.7 Filter the solution and retain the filtrate for analysis (see clause 5).

4.5.7.3.8 Rinse the filtering medium and retentate three times with small amounts of water (ISO 3696, grade 2).

4.5.7.3.9 Dry the coated disc and filtering medium with retentate to a constant mass.

4.5.7.3.10 Weigh the filtering medium with retentate. The difference in mass between the filtering medium withand without retentate is the mass of the retentate.

4.5.7.3.11 The difference between the original mass of the coating and the mass of the retentate is the mass ofthe materials dissolved.

4.5.8 Test procedure (all other ceramics)

4.5.8.1 Weigh the container without the top.

4.5.8.2 Weigh the container and specimen. Report the difference in mass between the container withspecimen and the container without specimen as the mass of the specimen.

4.5.8.3 Add (100 � 1) ml of buffer solution. Care shall be taken to ensure that the entire specimen is in contactwith the solution.

4.5.8.4 Place the container with the specimen in a controlled-temperature environment at (37 � 1) °C for(120 � 1) h. The container shall be agitated at 2 Hz using a longitudinal or circular movement. If the test specimenis totally dissolved before 120 h, terminate the test and note the time in the test report.

EN ISO 10993−14:2001

8

Page 19: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

ISO 1-399041:(1002E)

© ISO 1002 – All rights rsedevre 9

4.5.8.5 Remove the container and specimen and allow them to cool to room temperature.

4.5.8.6 Weigh the filtering medium (e.g. filter paper) to determine its mass without retentate.

4.5.8.7 Remove the specimen via filtration and retain the filtrate for analysis. Filtrate should not be stored inglass containers.

4.5.8.8 Rinse the filtering medium and retentate three times with small amounts of water (ISO 3696, Grade 2).

4.5.8.9 Dry the specimen and filtering medium with retentate to a constant mass (see 4.3.2).

4.5.8.10 Weigh the filtering medium with retentate. The difference in mass between the filtering medium withand without retentate is the mass of retentate.

4.5.8.11 The difference between the mass of the specimen and the mass of retentate is the mass of thedissolved material.

5 Analysis of filtrate

5.1 General

After each experiment a qualitative and quantitative analysis of the solution shall be performed in triplicate.

The number of test methods, practices, accuracy and precision on analytic techniques is large and changeable.Samples should be analysed using Inductively Coupled Plasma Spectroscopy (ICP) if possible. Other tests, suchas Atomic Absorption Spectroscopy (AAS), whilst less useful, may provide information at the desired concentrationlevels.

5.2 Choice of chemicals or elements to be analysed

The chemicals or elements to be analysed in the filtrate solutions should include both chemical constituents knownto exist in the material and possible impurities such as small amounts of elements that are due to commonly knownsubstitutions in the raw material and possible additions to the material during processing. The volume of filtrateshall be made up to a fixed volume of 125 ml or 250 ml, depending on the starting volume. Larger volumes shall bejustified.

5.3 Sensitivity of the analysis method

Applied analysis methods shall be of adequate sensitivity (e.g. at least 10�6 by atomic absorption or massspectroscopy). Record only compositional constituents that have been detected above the limits of quantification.Hazardous materials shall be recorded in accordance with the appropriate International Standards, if available.

6 Test report

The test report shall include all data identified in accordance with this part of ISO 10993 during characterization,testing and analysis:

a) test institution;

b) date of measurement;

c) a statement that this test was conducted in accordance with ISO 10993-14 and describing any deviations fromthe standard protocols, with justifications;

d) description of test material including batch or lot number;

EN ISO 10993−14:2001

9

Page 20: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

ISO -3990141:(1002E)

10 © ISO 1002 – All rights rsedevre

e) type of test:

1) high solubility,

2) low solubility,

3) extreme (intra-oral),

4) extreme (10993-14),

5) simulated;

f) surface area and method;

g) sample density, microstructure and X-ray diffraction pattern;

h) duration of test;

i) test results:

1) specimen mass,

2) volume of solution added,

3) drying time,

4) mass of retentate,

5) mass of material dissolved,

6) volume of filtrate,

7) chemical analysis and method (for coated specimens, the degradation products from the ceramic aredifferentiated from the substrate by comparing the analysis of filtrate of the blank disc with that from thecoated disc),

8) for each element identified in the filtrate, calculation of mass dissolved per total surface area.

EN ISO 10993−14:2001

10

Page 21: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

ISO 1-399041:(1002E)

© ISO 1002 – All rights rsedevre 11

Bibliography

[1] ISO 9693, Metal-ceramic dental restorative systems

[2] ISO 10993-12, Biological evaluation of medical devices — Part 12: Sample preparation and referencematerials

[3] ISO 10993-16, Biological evaluation of medical devices — Part 16: Toxicokinetic study design fordegradation products and leachables

[4] ISO 10993-17, Biological evaluation of medical devices — Part 17: Establishment of allowable limits forleachable substances using health-based risk assessment

[5] ASTM C92, Standard Test Methods for Sieve Analysis and Water Content of Refractory Materials

[6] ASTM D4780, Standard Test Method for Determination of Low Surface Area of Catalysts by MultipointKrypton Adsorption

EN ISO 10993−14:2001

11

Page 22: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation
Page 23: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

Annex ZA (informative)

Clauses of this European Standard addressing essential requirementsor other provisions of EU Directives

This European Standard has been prepared under a mandate given to CEN by the EuropeanCommission and the European Free Trade Association and supports essential requirementsof EU Directive 93/42/EEC.

WARNING Other requirements and other EU Directives may be applicable to theproduct(s) falling within the scope of this standard.

The following clauses of this standard are likely to support requirements of Directive93/42/EEC.

Compliance with these clauses of this standard provides one means of conforming with thespecific essential requirements of the Directive concerned and associated EFTA regulations.

Table ZA.1— Correspondence between this European Standard and EU Directives

Clause/subclause of thisEuropean Standard

Corresponding EssentialRequirement of Directive93/42/EEC

Comments

4 7.5 - Annex 1

5

EN ISO 10993−14:2001 EN ISO 10993−14:2001

2

Page 24: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

EN ISO 10993-14:2001 (E)

3

Annex ZA (informative)

Clauses of this European Standard addressing essential requirements orother provisions of EU Directives

This European Standard has been prepared under a mandate given to CEN by the EuropeanCommission and the European Free Trade Association and supports essential requirements ofEU Directive 93/42/EEC.

WARNING Other requirements and other EU Directives may be applicable to theproduct(s) falling within the scope of this standard.

The following clauses of this standard are likely to support requirements of Directive 93/42/EEC.

Compliance with these clauses of this standard provides one means of conforming with thespecific essential requirements of the Directive concerned and associated EFTA regulations.

Table ZA.1— Correspondence between this European Standard and EU Directives

Clause/subclause of thisEuropean Standard

Corresponding EssentialRequirement of Directive93/42/EEC

Comments

4 7.5 - Annex 1

5

EN ISO 10993−14:2001

2

Page 25: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

I NESO 01-39914:20( 10E)

4

Annex ZB(normative)

Normative references to international publicationswith their relevant European publications

This European Standard incorporates by dated or undated reference, provisions from otherpublications. These normative references are cited at the appropriate places in the text and thepublications are listed hereafter. For dated references, subsequent amendments to or revisions ofany of these publications apply to this European Standard only when incorporated in it byamendment or revision. For undated references the latest edition of the publication referred toapplies (including amendments).

NOTE Where an International Publication has been modified by common modifications, indicatedby (mod.), the relevant EN/HD applies.

Publication Year Title EN Year

ISO 3696 1987 Water for analytical laboratory use –Specification and test methods

EN ISO 3696 1995

ISO 6872 1995 Dental ceramic EN ISO 6872 1998

ISO 10993-1 1997 Biological evaluation of medical devices– Part 1: Evaluation and testing

EN ISO 10993-1 1997

ISO 10993-9 1999 Biological evaluation of medical devices- Part 9: Framework for identificationand quantification of potentialdegradation products

EN ISO 10993-9 1999

EN ISO 10993−14:2001

2

Page 26: Biological evaluation of medical · PDF fileThis part of ISO 10993 consists of two tests for the biological evaluation of medical devices: an extreme solution test and a simulation

BS EN ISO 10993-14:2001

BSI

389 Chiswick High Road

London

W4 4AL

BSI — British Standards InstitutionBSI is the independent national body responsible for preparing British Standards. It presents the UK view on standards in Europe and at the international level. It is incorporated by Royal Charter.

Revisions

British Standards are updated by amendment or revision. Users of British Standards should make sure that they possess the latest amendments or editions.

It is the constant aim of BSI to improve the quality of our products and services. We would be grateful if anyone finding an inaccuracy or ambiguity while using this British Standard would inform the Secretary of the technical committee responsible, the identity of which can be found on the inside front cover. Tel: +44 (0)20 8996 9000. Fax: +44 (0)20 8996 7400.

BSI offers members an individual updating service called PLUS which ensures that subscribers automatically receive the latest editions of standards.

Buying standards

Orders for all BSI, international and foreign standards publications should be addressed to Customer Services. Tel: +44 (0)20 8996 9001. Fax: +44 (0)20 8996 7001. Email: [email protected]. Standards are also available from the BSI website at http://www.bsi-global.com.

In response to orders for international standards, it is BSI policy to supply the BSI implementation of those that have been published as British Standards, unless otherwise requested.

Information on standards

BSI provides a wide range of information on national, European and international standards through its Library and its Technical Help to Exporters Service. Various BSI electronic information services are also available which give details on all its products and services. Contact the Information Centre. Tel: +44 (0)20 8996 7111. Fax: +44 (0)20 8996 7048. Email: [email protected].

Subscribing members of BSI are kept up to date with standards developments and receive substantial discounts on the purchase price of standards. For details of these and other benefits contact Membership Administration. Tel: +44 (0)20 8996 7002. Fax: +44 (0)20 8996 7001. Email: [email protected].

Information regarding online access to British Standards via British Standards Online can be found at http://www.bsi-global.com/bsonline.

Further information about BSI is available on the BSI website at http://www.bsi-global.com.

Copyright

Copyright subsists in all BSI publications. BSI also holds the copyright, in the UK, of the publications of the international standardization bodies. Except as permitted under the Copyright, Designs and Patents Act 1988 no extract may be reproduced, stored in a retrieval system or transmitted in any form or by any means – electronic, photocopying, recording or otherwise – without prior written permission from BSI.

This does not preclude the free use, in the course of implementing the standard, of necessary details such as symbols, and size, type or grade designations. If these details are to be used for any other purpose than implementation then the prior written permission of BSI must be obtained.

Details and advice can be obtained from the Copyright & Licensing Manager. Tel: +44 (0)20 8996 7070. Fax: +44 (0)20 8996 7553. Email: [email protected].