benefits we’ve got them....catheters are suitable for use with power injectors. powermidline™...
TRANSCRIPT
Bard’s goal is to provide you with exceptional vascular products. We leave it up to you, the skilled clinician, to choose the best device for your patients. We provide the options, you provide the expertise. With our PowerMidline™ Catheter configurations, you have excellent MST midline options for your vascular access needs.
Unique hub design further identifies the catheter as a midline which
helps eliminate catheter confusion and promotes patient safety.
20 cm Length - distally trimmable allows for customization to
individual patient’s anatomy.
*Please consult PowerMidline™ Catheter labels and inserts for power injection warnings, precautions, and directions for use.
Reverse taper gently plugs the insertion site and provides
excellent kink-resistance.
Clearly labeled as a midline catheter to assist with catheter identification.
POWER-INJECTABLE MIDLINE*
BENEFITS• Available in Full and Max Barrier kit nursing
configurations with Modified Seldinger
Technique (MST) components
• Helps to improve patient comfort due
to the potential for decreased needle sticks
• Body-softening polyurethane helps
minimize vessel wall trauma
• No x-ray required during the placement
procedure
• StatLock® Stabilization Device compatible
5 Fr.Dual-Lumen
4 Fr. Single-Lumen
OPTIONS. WE’VE GOT THEM.
Please consult product labels and inserts for any indications, contraindications, hazards, warnings, cautions and instructions for use.
BARD ACCESS SYSTEMS, INC. • 605 North 5600 West • Salt Lake City, Utah 84116 USA • Main: 801-522-5000 • Customer Service: 1-800-545-0890 • Clinical Information: 1-800-446-3385 www.bardaccess.com
Bard, GuardIVa, PowerMidline, and StatLock are trademarks and/or registered trademarks of C. R. Bard, Inc. All other trademarks are the property of their respective owners. © 2017 C. R. Bard, Inc. All rights reserved. BAS/PWRM/1116/0012
Indications for Use: The PowerMidline™ Catheters are indicated for short term access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The PowerMidline™ Catheters are suitable for use with power injectors.
PRODUCT CODESPOWERMIDLINE™CATHETER
PRODUCT CODE CATHETER SIZE LUMENS LUMEN SIZE TRAY AVERAGE PRIMING
VOLUME (ML)MAX ACHIEVABLE PUMP FLOW RATE1 (ML/HR)
POWER INJECTION RATE (ML/SEC)
AVERAGE GRAVITY FLOW RATE2 (ML/HR) QUANTITY
P4153108 3F SL 20G
Full Tray
0.31 Saline (1.0 cP) > 999 3 mL/sec 729
5/case
P4154108 4F SL 18G 0.37Saline (1.0 cP) > 999Blood (3.5 cP) = 959
7 mL/sec 2116
P4254108 4F DL 21G, 19G 0.30, 0.38Saline (1.0 cP) > 999
Blood (3.5 cP) = 904**19g lumen only
4 mL/sec 220, 1523
P4255108 5F DL 18G, 18G 0.37Saline (1.0 cP) > 999 Blood (3.5 cP) = 887
7 mL/sec 1439
P4254108G 4F DL 21G, 19GFull Tray with
GuardIVa® Dressing
0.30, 0.38Saline (1.0 cP) > 999
Blood (3.5 cP) = 904**19g lumen only
4 mL/sec 220, 1523
P4255108G 5F DL 18G, 18G 0.37Saline (1.0 cP) > 999Blood (3.5 cP) = 887
7 mL/sec 1439
P4153108D 3F SL 20G
Max BarrierTray
0.31 Saline (1.0 cP) > 999 3 mL/sec 729
3/case
P4154108D 4F SL 18G 0.37Saline (1.0 cP) > 999Blood (3.5 cP) = 959
7 mL/sec 2116
P4254108D 4F DL 21G, 19G 0.30, 0.38Saline (1.0 cP) > 999
Blood (3.5 cP) = 904* *19g lumen only
4 mL/sec 220, 1523
P4255108D 5F DL 18G, 18G 0.37Saline (1.0 cP) > 999Blood (3.5 cP) = 887
7 mL/sec 1439
P4254108DG 4F DL 21G, 19GMax
Barrier Tray with
GuardIVa® Dressing
0.30, 0.38Saline (1.0 cP) > 999
Blood (3.5 cP) = 904**19g lumen only
4 mL/sec 220, 1523
P4255108DG 5F DL 18G, 18G 0.37Saline (1.0 cP) > 999Blood (3.5 cP) = 887
7 mL/sec 1439
PowerMidline™ Catheter | Product Ordering Information & Specifications
1. Estimates based on a pump alarm setting of 500 mmHg and full length catheter at 20 cm. 2 Testing performed using water.