basics of liquid chromatography fileintroduction to liquid chromatography ... between the mobile and...
TRANSCRIPT
Rachel Beck, MS, DFTCB
Basics of Liquid Chromatography
Overview Introduction to Liquid chromatography
Theory
Practical Applications
Types of column
Advantages/Disadvantages
What is Liquid Chromatography?
Liquid chromatography (LC) is a separation
technique.
LC Classifications
Liquid-Liquid
Mobile phase – liquid
Stationary phase – liquid
Example: Liquid/Liquid extraction
Liquid-solid
Mobile phase – liquid
Stationary phase – solid
Example: High Pressure Liquid Chromatography
What is HPLC? A separation technique in
which a liquid sample is passed over a solid adsorbent material packed into a column (stationary phase) using a flow of liquid solvent (mobile phase) at high pressures.
http://en.wikipedia.org/wiki/High-performance_liquid_chromatography
How does LC work? A sample mixture is injected
onto the column. Each component partitions between the stationary phase (column packing material) and the mobile phase (solvent system). Molecules that have a greater affinity for the stationary phase take longer to reach the detector.
LC Schematic
To Mass Analyzer
Sample Mixture + Mobile Phase
Injection Time 2.0 min Time 3.5 min Time 5 min
t0 t5 t6
Dete
cto
r S
ignal
Time
Theory
Theory In order to understand liquid
chromatography, one must understand migration rates, resolution, and peak shape.
Migration rates – Retention Times
Resolution
Peak Shape – Van Deemter Equation
Migration Rates Migration rates are equilibrium constants
describing the distribution of the solutes between the mobile and stationary phases.
Mathematically, migration rates are related to retention times.
Retention time is the time difference between injection and analyte detection.
t0 t5 t6
Dete
cto
r S
ignal
Time t1 t2 t3 t4
Inje
ctio
n
Resolution Resolution is the column’s ability to differentiate
between two peaks.
An R value of 1.5 indicates baseline resolution
An R value of >1.0 is necessary for quantification purposes.
α is the selectivity factor
κ is the capacity factor
N is the column efficiency
Ra 1
a
k
K 1
N
4
Selectivity Factor A fraction consisting of the partition ratios of two
retained species.
Selectivity is the chemistry of the system
how does the analyte interact with the stationary phase
The selectivity is always equal to or greater than one. (α 1.0)
α is the selectivity factor
κ is the capacity factor
N is the column efficiency
Ra 1
a
k
K 1
N
4
α = KB
KA
k'B
k'A
= = (tR)B-t0
(tR)A-t0 http://www.chromedia.org/chromedia
Capacity Factor A dimensionless term used in conjunction with retention
time to describe chromatographic behavior (time an analyte spends in the stationary phase)
Calculated from the chromatogram
α is the selectivity factor
κ is the capacity factor
N is the column efficiency
Ra 1
a
k
K 1
N
4
k' = tR-t0
t0 http://www.chromedia.org/chromedia
Theoretical Plates A measure of the degree of broadening of a
chromatographic band; often expressed in terms of plate height (H) or the number of theoretical plates (N).
The more theoretical plates and the smaller the plate height the more efficient separation or increased resolution.
α is the separation factor
κ is the retention factor
N = number of theoretical plates
Ra 1
a
k
K 1
N
4
Time
Ideal
Broad
Fronting
Tailing
Doublet
McNair Basic Gas Chromatography 1998
Peak Broadening
a) Original chromatogram with overlapping peaks
b) Increased band separation
c) Decreased band width
Skoog, West, & Holler. Fundamantals of Analytical Chemistry.
Increasing column efficiency (HETP) increases resolution.
Factors Affecting Column Efficiency Column length
Particle size
Packing quality
Linear velocity (flow)
Instrument quality (dead volume)
Capacity factor
The factors influencing efficiency are described mathematically by the Van Deemter Equation.
Peak Shape Peak Shape is the actual Guassian character
of the analyte peak and is mathematically described by the Van Deemter Equation.
H is the plate height in centimeters
U is the linear velocity of the mobile phase in cm/sec
A = multiple flow paths
B = longitudinal diffusion
C = mass transfer between phases
CS mass transfer associated with stationary phase
CM mass transfer associated with mobile phase
H A B u CS C M
Van Deemter Schematic
http://www.chromacademy.com/resolver-september2010_High_Efficiency_HPLC_Separations.asp
Linear Velocity (Flow)
Magnitude of kinetic effects on column efficiency clearly depends upon the length of time the mobile phase is in contact with the stationary phase
This depends on the flow rate of the mobile phase
Efficiency studies typically determine plate height as a function of mobile phase velocity
Optimum flow: a certain velocity at which the plate number is highest (& the plate height is lowest)
Eddy Diffusion
Eddy diffusion or zone broadening arises from the multitude of pathways by which a molecule can find its way through a packed column.
H A B u CS C M
The more paths the
molecule can take the
broader the peak
Flow Direction
Longitudinal Diffusion
Peak broadening process in which solutes diffuse from the concentrated center of a zone to the more dilute regions.
Dependent upon mobile phase flow rates.
Contributes to peak broadening only at very low flow rates below the minimum (optimum) plate height
Because mobile phase velocity is much higher than the diffusion coefficient of the component in solution, the B term is hardly significant in liquid chromatography
H A B u CS C M
Mass Transfer
H A B u CS C M
A chromatographic system is constantly changing with the mobile phase
The time for restoration of the equilibrium (resistance to mass transfer) results in slight broadened concentration profiles.
Practical Applications
Terminology Mobile phase is a general term that refers to
the moving phase in chromatography.
Gas Chromatography (GC) the mobile phase is a gas and is termed CARRIER GAS
Liquid Chromatography (LC) the mobile phase is liquid and is termed MOBILE PHASE
Stationary Phase refers to the column packing material in chromatography.
Mobile phase of GC In gas chromatography, the moving phase
is a gaseous mixture of sample and carrier
gas.
The liquid sample is injected into the inlet where flash vaporization occurs.
The carrier gas then moves the gaseous sample through the column to the mass analyzer.
Carrier gas should be inert so the integrity of the sample is maintained. Common carrier gases include Helium and Hydrogen
In liquid chromatography, the moving phase is a mixture of liquid sample and liquid mobile phase.
The liquid sample is injected into a loop which is flushed with mobile phase.
The mobile phase moves the liquid sample through the column to the mass analyzer.
The mobile phase can be a single solvent; but typically consists of a mixture of aqueous
and organic solvents. Common solvents are methanol,
acetonitrile, and buffers.
Mobile phase of LC
Mobile Phase Composition There are two theories for elution of solutes from a
liquid chromatographic column. Isocratic elution
Gradient elution
Isocratic elution – The composition of the mobile phase (mixture of A and B) is
maintained throughout the acquisition, or
A single solvent is used.
Gradient elution – the composition of the mobile phase is changed throughout the acquisition.
Similar to temperature programming of GC
Mobile Phase Gradients
To aid in separation, decrease elution time, and optimize peak shape the composition of the mobile phase mixture is changed throughout the sample acquisition.
This change in composition disrupts the interactions occurring between analyte and column stationary phase.
Guard Column • Introduced in front of the analytical column to increase
the life of the analytical column
• Removes particulate matter & contaminants from solvents
• Composition of the guard column should be similar to that of the analytical column
• The particle size is usually larger to minimize pressure drop
http://www.phenomenex.com/guard-hplc-column
Types of LC Columns
LC Stationary Phase Categories Normal Phase - based on
polarity
Reverse Phase – based on polarity
Size Exclusion – based on size
Mixed Phase – utilizes a combination of stationary phases
Ion Exchange – based on charge
Affinity – based on affinity
Chiral – based on chirality
LC Column Application
Skoog, West, & Holler. Fundamantals of Analytical Chemistry.
Size Exclusion Most applicable to high molecular weight species
Packing consists of small silica or polymer particles containing a network of uniform pores
Molecules that are larger than the average pore size have no retention
Separation is based on size not chemical
or physical interactions between analyte
and stationary phase
Two types of packing Polymer beads – polystyrene-divinylbenzene
Silica based particles – greater rigidity; easier
packing; greater stability http://www.shimadzu.com/an/hplc/support/lib/lctalk/55/55intro.html
Ion Exchange Separation and ion determination based upon ion-exchange
resins. (sulfonic acid and quaternary amines)
Stationary phases can be porous beads, pellicular beads, or porous microparticles of silica Porous beads are styrene and
divinylbenzene based with
Anion/cation functional groups
bonded.
Pellicular beads are large
nonporous, spherical glass or
polymer beads coated with a
synthetic ion exchange resin.
Porous silica microparticles are coated with a thin film of exchanger http://www.waters.com/waters/en_US/HPLC-Separation-Modes/nav.htm?cid=10049076
Affinity Separation occurs based on ligand recognition Antigen to antibody
Enzyme to substrate
Receptor to ligand
Most selective chromatography
Utilizes specific interactions between one kind of solute molecule and a second molecule immobilized on a stationary phase.
Not frequently used in
Forensic laboratories.
http://www.rpi.edu/dept/chem-eng/Biotech-Environ/CHROMO/be_types.htm
Chiral Chiral compounds contain atleast one asymmetric carbon
and are often observed as enantiomers and diasteromers same molecular weight and formula; different spatial
arrangements
Separation achieved by the number and type spatially selective interactions
Common stationary phases
are polysaccharide, ligand
exchange, protein, helical
polymers, and macrocyclic.
Not common due to
production demands
http://www.phenomenex.com/chiral-hplc-column#
Normal Phase Contains a POLAR stationary phase
Theory – polar compounds will interact with stationary phase while non-polar compounds will elute early
Normal phase stationary phases typically include polar functional groups (silica, amino, and cyano).
LC gradient transitions from non-polar to polar Interrupts interaction between polar analytes and
column stationary phase
This is NOT commonly used in Forensic Labs
Reverse Phase Contains a NON-POLAR stationary phase.
Theory – non-polar analytes will interact with stationary phase while polar analytes will elute early
Reverse phase stationary phases include C18, C4, and C8
LC gradient composition transitions from polar to non-polar. Disrupts non-polar interaction between analyte and column
stationary phase
Most commonly used stationary phase
Mixed Phase Contains a mixture of stationary phases.
Theory – optimizes separation of complex mixtures through multiple column chemistries
ADFS employs a Synergi Fusion column
Utilizes both polar and non-polar stationary phase
Optimal for a broad range of drugs http://www.phenomenex.com/Products/HPLCDetail/Synergi/Fusion-RP?returnURL=/Products/Search/HPLC
Liquid Chromatography Schematic
Column Pump A
Pump B
6-Port Valve
Mass Analyzer
Sample
Waste
Sample Loop
How does it Work?
Effluent from pump A and B are constantly mixed according to the programmed gradient and are initially flowed through the multiport valve to column.
Sample is injected into the multiport valve where it will fill the sample loop.
The valve switches and the column effluent is diverted to the sample loop where it pushes the sample through to the column and eventually to the mass analyzer.
6-Port Valve Mechanism
http://faculty.uml.edu/david_ryan/84.314/Instrumental%20Lecture%2018.pdf
Advantages Shorter Run Time (<10 mins.) no loss of separation
Volatility issues resolved – liquid effluent vs. gas effluent
Venting of mass analyzer not necessary for column change
Use of guard column prevents constant column maintenance
Ease of troubleshooting
Liquid mobile phase
Leak detection sensors
Advantages Cooled autosampler minimizes sample
evaporation
LC is capable of switching between two columns (different stationary phases).
Easily interfaced with a variety of mass analyzers (TOF, QTRAP, MS/MS, UV, etc)
Disadvantages Costs associated with mobile phase solvents
Autosampler limitations can exist
Clogged lines/injection needle more prevalent
Liquid samples often contain undissolved solutes
Partial clog can cause retention time to be unstable
Air bubbles can damage column and cause RT shifting
Summary
LC is an effective tool in the analysis of drug analytes in a forensic laboratory
setting and should be utilized as a complement to the traditional GC/MS
instrumentation.
Questions and/or Comments?