AUTO LENSMETER Mode Model LM-600/600P/600PD LM-600/600P/600PD OPERATOR’S MANUAL OPERATOR’S MANUAL

Upload: vuonglien

Post on 15-May-2019

220 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

AUTO LENSMETER

ModeModel LM-600/600P/600PDLM-600/600P/600PD

OPERATOR’S MANUALOPERATOR’S MANUAL

Page 2: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

NIDEK CO., LTD. : 34-14, Maehama, Hiroishi-cho, Gamagori, Aichi 443-0038, Japan(Manufacturer) Telephone: +81-533-67-6611

Facsimile: +81-533-67-6610NIDEK CO., LTD : 3F Sumitomo Fudosan Hongo Bldg., 3-22-5, Hongo,(Tokyo Office) Bunkyo-Ku, Tokyo 113-0033, Japan

Telephone: +81-3-5844-2641Facsimile: +81-3-5844-2642

NIDEK INCORPORATED : 47651 Westinghouse Drive, Fremont, California 94539, U. S. A.(United States Agent) Telephone: +1-510-226-5700

Facsimile: +1-510-226-5750NIDEK S.A. : Europarc 13, rue Auguste Perret, 94042 Créteil, France(EU Authorized Representative) Telephone: +33-1-49 80 97 97

Facsimile: +33-1-49 80 32 08

June 201031018-P902C

Printed in Japan

Original instructions

Page 3: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

I

Use this device properly and safely.

Safety precautions

BEFORE USE, READ THIS MANUAL.

This operator’s manual includes operating procedures, safety precautions, andspecifications for the NIDEK AUTO LENSMETER, LM-600/LM-600P/LM-600PD.

Cautions for safety and operating procedures must be thoroughly understood beforeusing the device.

Keep this manual handy for reference.

The device complies with ISO 8598 (Optics and optical instruments - Lensmeters).

There are no parts within the device that requires servicing by the user other than printerpaper.

This manual is described mainly using the screen samples of the LM-600P.

If you encounter any problems or have questions about the device, please contact NIDEKor your authorized distributor.

In this manual, a signal word is used to designate the degree or level of safety alerting. The defini-tion is as follows.

CAUTION • Indicates a potentially hazardous situation which, if not avoided, may result in minor ormoderate injury or property damage accident.

Even situations that are labeled CAUTION may result in serious injury under certain conditions.Safety must be followed strictly at all times.

Page 4: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

II

Usage precautions

Before use

CAUTION• Do not use the device for other than the intended purpose.NIDEK will not be responsible for accidents or malfunction caused by such misuse.

• Never modify nor touch the internal structure of the device.This may result in electric shock or malfunction.

• Install the device in an environment that meets the following conditions. The followingconditions must be maintained during use.

Temperature: 10�to 35°C (50�to 95°F)Humidity: 30 to 85% (Non-condensing)Pressure: 800 to 1060 hPaA place with low dustA place with little external lightA place free of vibration and impact

• Do not store the device in an area that is exposed to rain or water, or containspoisonous gas or liquid.

Corrosion or malfunction of the device may occur.

• Avoid installing the device where it is exposed to direct air-conditioning flow.Changes in temperature may result in condensation inside the device or adverselyaffect measurements.

• Avoid using the device in a place exposed to direct sunlight or near incandescentlight. A light-reflecting surface such as a glass showcase or a shiny table is also notappropriate for precise work.

Under such circumstances, the device may work irregularly or issue error messages.

• Be sure to use a wall outlet which meets the power specification requirements.If the line voltage is too high or too low, the device may not give full performance. Mal-function or fire may occur.

• The electrical outlet must have a grounding terminal.Electric shock or fire may occur in the event of malfunction or power leakage.

• Insert the mains plug into an outlet as far as the prongs of the plug will go.Fire may occur if the device is used with a loose connection.

• For supplying the device with the power, never use a table tap and extension cable.The electrical safety may be lowered.

Page 5: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

III

During use

CAUTION• Install the device in area where the outlet that the mains plug is inserted into is easilyaccessible during use. In addition, ensure that the power cord can be disconnectedwithout the use of a tool.

Otherwise, it may interfere with disconnecting of the power from the input power sourcein case of abnormality.

• Do not use the power cord other than equipped. Do not use the equipped power cordfor purpose other than intended.

Malfunction or fire may result.

• Do not place heavy objects on the power cord.The damaged power cord may cause fire or electric shock.

• Before connecting a cable, turn off the power switch and disconnect the power cordfrom an outlet.

Malfunction of the device may result.

• To transport the device, use the special packing materials to protect the device fromimpact of dropping.

Excessive vibration or impact to the device may cause malfunction.

CAUTION• Immediately replace the power cord if the internal wires are exposed, the table turnson or off when the power cord is moved, or the cord and/or plug are too hot to be heldwith hands.

This may result in electric shock or fire.In the event of malfunction, disconnect the power cord from the wall outlet. Never touchthe inside of the device and contact your authorized distributor.

• There may be a few constantly-lit, missing, or dead pixels in your LCD which are acharacteristic of the LCD. This does not represent failure of the LCD; continuouslyuse the display.

• Class 1 LED Product output beam is emittedfrom the LED aperture. (UV model only)

LED aperture

[UV model only]

Page 6: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

IV

CAUTION• The International Electromagnetic Commission sets the essential requirements forelectrical and electronic equipment that may disturb, or be disturbed by, otherequipment. The LM-600/LM-600P/LM-600PD complies with these requirements astabled on pages 113 to 116. Follow the guidance in the tables for use of the device inan electromagnetic environment.

• This device has been tested and found to comply with the limits for medical devicesto the IEC 60601-1-2: 2001 + A1: 2004.These limits are designed to provide reasonable protection against harmfulinterference in a standard medical installation.This device generates, uses and can radiate radio frequency energy and, if notinstalled and used in accordance with the instructions, may cause harmfulinterference to other devices in the vicinity.However, there is no guarantee that interference will not occur in a particularinstallation. If this device does cause harmful interference to other devices, which canbe determined by turning the device off and on, the user is encouraged to try tocorrect the interference by one or more of the following measures:

Reorient or relocate the receiving device.Increase the separation between the devices.Connect the device to an outlet on a circuit different from that to which the otherdevice(s) are connected.Consult the manufacturer or field service technician for help.

• Never use the device with cable or accessories other than the designated ones.Malfunction caused by deteriorated electromagnetic compatibility (EMC) characteristicsmay result.

• Never use portable and mobile radio frequency (RF) devices in the vicinity of thisdevice.

These devices may adversely affect medical electrical equipment and malfunction mayresult.

Page 7: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

V

After use

Maintenance and check

Disposal

CAUTION• While the device is not in use, turn it off.

• The LM-600P/LM-600PD uses a heat-sensitive printer paper. To keep the printed datafor a long period of time, make copies of the printouts.

The paper degrades over time and the printed data may become illegible.

• Do not yank the power cord to disconnect it from an outlet but hold the plug.This can damage the metal core of the cord and may result in electric shock, short cir-cuit, or fire.

• Occasionally clean the prongs of the mains plug with a dry cloth.If dust settles between the prongs, the dust will collect moisture, and short circuit or firemay occur.

• If the device will not be used for a long time, disconnect the power cord from the walloutlet.

Electric leak may occur.

CAUTION• Only service technicians properly trained by NIDEK can repair the device.NIDEK is not responsible for any accidents resulted from improper servicing.

• When performing maintenance work, secure a sufficient maintenance space.Maintenance work in an insufficient space may result in injury.

• Take special care not to scratch the protective glass under the nosepiece.Flaws on the glass substantially lower the reliability of the measurement.

• Clean the protective glass under the nosepiece every once in a while with a blower.If dust settles on the protective glass, it may affect the measurement accuracy.

CAUTION• Follow the local governing ordinances and recycling plans regarding disposal orrecycling of device components. It is recommended to commission the disposal to adesignated industrial waste disposal contractor.

• When disposing of packing materials, sort them by material and follow localgoverning ordinances and recycling plans.

Page 8: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

VI

Page 9: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

Table of Contents

VII

1. BEFORE USE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1 Outline of Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1

1.2 Intended Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1

1.3 Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2

1.4 Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61.4.1 Measurement screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61.4.2 Prism layout screens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .131.4.3 UV measurement screen (UV model only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .171.4.4 Parameter screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19

1.5 Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20

1.6 Checking Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21

1.7 Before First Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22

2. OPERATING PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . 252.1 Operation Flow. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25

2.2 Preparation for Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26

2.3 Setting Lenses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .292.3.1 Setting uncut lenses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .292.3.2 Setting mounted lenses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .312.3.3 Setting for PD measurement (LM-600PD only) . . . . . . . . . . . . . . . . . . . . . . . . . . .32

2.4 Measuring Single Vision Lenses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35

2.5 Measuring Bifocal Lenses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38

2.6 Measuring Progressive Power Lenses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .432.6.1 Measuring uncut lenses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .432.6.2 Measuring mounted lenses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43

2.7 Measuring Prism Lenses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52

2.8 Measuring Contact Lenses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54

2.9 Measuring UV Transmittance (UV model only) . . . . . . . . . . . . . . . . . . . . . . . . . .57

2.10 Marking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .602.10.1 Marking at the optical center. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .602.10.2 Marking for prism prescription (Prism Layout) . . . . . . . . . . . . . . . . . . . . . . . . . . . .622.10.3 Marking for prism prescription (Measurement screen) . . . . . . . . . . . . . . . . . . . . . .67

2.11 Printing (LM-600P/LM-600PD only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .692.11.1 Connecting to the AR, ARK or RKT and setting parameters . . . . . . . . . . . . . . . . .70

2.12 Data save to the Eye Care card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .722.12.1 Writing of fixed data to the Eye Care card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .722.12.2 Erasing Data on the Eye Care Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74

Page 10: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

:

VIII

2.13 After Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

2.14 Setting Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 762.14.1 Parameter tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 782.14.2 Entering comments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88

3. OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED. .913.1 Connecting to the NIDEK Motorized Refractor (RT) or Computer . . . . . . . . . 91

3.1.1 Outline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 913.1.2 Connecting procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 923.1.3 Operating procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

3.2 Connecting the Foot Switch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94

4. MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 954.1 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95

4.2 Error Messages and Countermeasures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96

4.3 Replacing Printer Paper (LM-600P/LM-600PD only) . . . . . . . . . . . . . . . . . . . . . 98

4.4 Ink Refilling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1004.4.1 Ink cartridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1004.4.2 Ink pad type (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101

4.5 Cleaning the Protective Glass. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102

4.6 Adjusting the Lens Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103

4.7 Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1044.7.1 Cleaning the device exterior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1044.7.2 Cleaning the printer (LM-600P/LM-600PD only) . . . . . . . . . . . . . . . . . . . . . . . . . 1044.7.3 Cleaning the Eye Care card slot. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105

4.8 List of Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106

5. SPECIFICATIONS AND ACCESSORIES . . . . . . . . . . . 1075.1 Classifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107

5.2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108

5.3 Standard Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1115.3.1 Standard accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1115.3.2 Optional accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111

6. EMC (ELECTROMAGNETIC COMPATIBILITY) . . . . . . 113

7. INDEX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117

Page 11: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

1

1

1. BEFORE USE

1.1 Outline of Device

The NIDEK AUTO LENSMETER LM-600/LM-600P/LM-600PD measures the optical performance ofspectacle lenses such as single vision, bifocal (trifocal), and progressive power lenses or contactlenses. This device has a measuring unit and a display unit in front, and a printer unit on the right side(Only the LM-600P/LM-600PD is equipped with the printer). The measuring unit is comprised of anosepiece, a lens table, a lens holder, and a lens marker. Below the nosepiece, there is a read buttonfor reading measured data.

The display unit employs a full-graphic color LCD, displaying measured data, alignment condition of alens, and settings of the device.The device also offers the following features as well as the above features:

� Performs area measurement, which measures the refractive power at multiple positions ofa lens simultaneously.

Makes measurement easier and more correct by detecting changes in the refractive power of aprogressive power lens.

Changes measurement mode automatically after detecting a progressive power lens.

� Performs easier and quicker measurement with automatic functions such as auto read,auto R/L selection, and auto ADD measurement.

� Measures the Pupillary Distance (PD). (LM-600PD only)

� Connected to a personal computer with the USB interface in addition to the RS-232C inter-face of the standard communication protocol among NIDEK optometry devices.

� Adopts the marker for repellent or super repellent lenses.

� Equipped with the Eye Care card writer and enables to transfer the measured data to otherdevices wirelessly.

� Contains a speedy, advanced, and quiet line printer. (LM-600P/LM-600PD only)

� Measures ultraviolet rays (UV-A, center wavelength 365 nm) transmittance. (UV modelonly)

1.2 Intended Use

This device is designed to measure vertex powers and prismatic effects of spectacle and contactlenses, to orientate and mark uncut lenses, and to verify the correct mounting of lenses in spectacleframes.

LM-600 basic model (without the printer)

LM-600P equipped with the printer.

LM-600PD equipped with the printer and PD measurement function.

Page 12: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

2

BEFORE USE Configuration

1.3 Configuration

1. Display

2. Pilot lamp

3. Function buttons

4. Eye Care card slot

5. Lens holder lever

6. Nosepiece

7. Read button

8. Contrast knob

9. Marking lever

10. Lens table

11. Lens table lever

12. Printer cover

[LM-600P]

The LM-600 has no printer.

[LM-600]

13. Nose slider

[LM-600PD]

Page 13: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

3

BEFORE USE Configuration

1

1. Display320 × 240 dots color LCD.

2. Pilot lampIndicates ON or OFF of the power to the device.

• Display auto off modeWhen no button is pressed for the time set with the Auto off parameter, the display auto off function works and thenthe display turns off automatically. The measurement light source also turns off.To recover from auto off state, press any button. During lens measurement, the device will not be placed into displayauto off mode.The LM-600PD cancels the display auto off mode after the nose slider is moved.

3. Function buttonsFunctions assigned to the function buttons are displayed with icons on the screen.

4. Eye Care card slotUsed to save the measured data.

Never remove the Eye Care card while it is being accessed.When the measured data is saved in the LM-600/LM-600P/LM-600PD memory and the Eye Care card is inserted,the data is automatically written.

5. Lens holder leverUsed to move the lens holder.

6. NosepieceUsed to hold a lens; the base point for measurements.

When measuring contact lenses, replace the standard nosepiece with the supplied one for contact lenses.Before measurement, make sure that the contact part of the lens has no foreign matters or scratches.

7. Read buttonUsed to read measured data.

This button fixes the measured data on the screen so that they are saved in the memory.The read button is inoperative during occurrence of a measurement error.UV model: Holding this button for about 1 second starts UV measurement.Models without the UV function: Holding this button for about 1 second on the normal measurement screen, PPLmeasurement screen, or auto measurement screen changes to the prism entry screen.

8. Contrast knobUsed to adjust the screen contrast.

9. Marking leverUsed to mark a lens for its basic line. To do so, push the lever down.

ON Power turned on

OFF Power turned off

Blinking Display auto off mode

To fix a lens Lift it once and lower it gently.

To remove a lens Lift it until it clicks.

Page 14: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

4

BEFORE USE Configuration

10. Lens tableTouched to the bottom of lens frames.

The PD can be measured with the scale on the lens table. (LM-600/LM-600P)Mount the lens frames so that their nosepads come into contactwith the nosepiece while aligning the frame center. Bring thelens table close to the lens to read the single-eye PD.* The nosepads may not touch the nosepiece depending on

the lens frame size and nosepad shape.

11. Lens table leverUsed to move the lens table back and forth.

12. Printer cover (LM-600P/LM-600PD only) Opened by sliding the lever at the rear of the device when replacing the printer paper.

13. Nose slider (LM-600PD only)Used to measure the PD of mounted lenses.

Place the frames so that their nosepads straddle the nose slider. The nose slider position determines the right or leftlens. The PD is measured from the nose slider position when the measured data is read.

When the state is changed from single to R/L by pressing the clear button , moving the nose slider returnsthe state to R/L and displays the PD.The nose slider functions when the lens table is positioned within 50 mm from the nosepiece center. When thenosepiece center is separated from the lens table by more than 50 mm, the slider automatically turns off.When the slider is not in use, move it to the far left and lift it upright so that it does not obstruct measurement.

Page 15: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

5

BEFORE USE Configuration

1

[Rear side]

14. Power switchUsed to turn on or off the power of the device.

15. Printer cover lever (LM-600P/LM-600PD only)Used to unlock the printer cover when replacing the printer paper.

16. Power inletUsed to connect a removal power cord.

17. USB port*1

Used to connect a computer.In conformity with USB 1.1. Connect a computer using the optional USB cable.The communication software needs to be installed in the computer.

18. Interface connectorUsed for communication.

RS-232C interface connector. Connect the AR, ARK, RT-2100/RT-5100, or a computer.When the device is connected to the AR-600/AR-330A series, ARK-700/ARK-530A series, RKT-7700/TONOREF II,it is possible to print the data measured by the LM-600P/LM-600PD with the printer of the connected device.To connect the optional foot switch, also use this connector.To connect a computer, the communication software needs to be installed in a computer.

Connectable devicesAR-600/AR-330A series, ARK-700/ARK-530A series, RKT-7700/TONOREFII, RT-2100 series, RT-5100 series

17. USB port

15. Printer cover lever

16. Power inlet

18. Interface connector

* Remove the cap be-fore connection.

14. Power switch

*1. Accessory equipment connected to the analog and digital interfaces must be certified according to the representative ap-propriate national standards (for example, UL 1950 for Data Processing Equipment, UL 60601-1 for Medical Equipment,and CSA C22.2 No. 601-1, EN 60601-1, and IEC 60601-1.) Furthermore, all configurations shall comply with the system stan-dard IEC 60601-1-1. Anyone who connects additional equipment to the signal input part or signal output part configures amedical system, and is therefore responsible that the system complies with the requirements of the system standard IEC60601-1-1. If in doubt, consult the technical service department or your local representative.

Page 16: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

6

BEFORE USE Displays

1.4 Displays

1.4.1 Measurement screenThere are four measurement screens, Auto measurement screen, Normal measurement screen, Pro-gressive power lens (PPL) measurement screen, and Contact lens (CL) measurement screen.

The screen below shows the normal measurement screen to measure single vision lensesor bifocal (trifocal) lenses.

1. R/L indicationShows measurement in the R/L state. Data below the indication shows measured data of the lens side.

To change the lens side, press the L-selection button or R-selection button . The R or L indication

of the side that is being specified at the moment will blink.

To change the state from R/L to single , press the clear button .

1. L indication

3. Target

4. Measurement mode/lens indication

2. Setting indications

7. Measurement selection button8. Cylinder +/- button

9. R/L selection button10. Clear button

1. R indication

5. AXIS bar

6. Alignment circle

11. Print button

The lens side is not selected (single state).

The left-eye lens is selected.

The right-eye lens is selected.

The measured data of the lens in single state is saved in the memory.

The measured data of the left-eye lens is saved in the memory.

The measured data of the right-eye lens is saved in the memory.

Page 17: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

7

BEFORE USE Displays

1

2. Setting indicationsShow each setting of the abbe number, wavelength, indication step, CYL mode, auto read, and state of theEye Care card.

Each setting is changed on the parameter screen.

•Auto readWhen the target is aligned, as well as when the read button is pressed, measured data is automatically saved in thememory and fixed on the screen.

3. TargetShows the optical center of a lens to be measured.

The shape of the target changes according to the setting of the Target parameter and alignment level.

For the normal target, when auto read is set and the target changes to (cross) or (large cross), the currentaxis is displayed to the upper right of the target.

Abbe number Set a value in the range of 20 to 60.

Wavelength :e-line or :d-line

Indication step 0.01, 0.06, 0.12, or 0.25 (D)

CYL mode +/- (MIX), +, or -

Auto read

Setting state of auto read In the single state, settings of the Auto RD S parameter are displayed. In the R/L state,settings of the Auto RD R/L parameter are displayed. Normal target

Off: , : , : Prism target

Off: , : , :

Eye Care card

State of the Eye Care card

(White): The Eye Care card is not inserted.

(Yellow): The Eye Care card is inserted.

(Green): The Eye Care card is being accessed. Never touch the card.

(Red): An error occurs while the Eye Care card is being accessed.

Eye Care card

Auto read

CYL mode

Abbe number

Wavelength

Indication step

Normal target Prism target Details of target

Misaligned

(cross)

Placed within approximately 0.5�. Measured data can beread.

When auto read ( ) is set, the measured data is fixed.

(large cross)

Aligned (marking point)

When auto read ( ) is set, measured data is fixed.

When auto read ( ) is set, measured data is automaticallyread again and the updated values are fixed.

Page 18: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

8

BEFORE USE Displays

4. Measurement mode/Lens indicationShows a measurement mode and whether the lens side is specified or not.

The R/L indication on the top of the screen also shows the lens side.

See “� Screen change in auto measurement” (page 48) for details of the screen in auto measurementand change of the measurement mode indication.

Indication samples by combination of the measurement mode and lens side

5. AXIS barThe thick line displayed on the protractor in the alignment circle shows the axis.

6. Alignment circleThe target is displayed on this circle. The center shows the optical center position.

7. Measurement selection buttonUsed to change the measurement screen.Each press changes the measurement screen in the order of Auto measurement ��Normal measurement �PPL measurement ��CL measurement ��Auto measurement...

The icon on the button shows which screen will be displayed next when the button is pressed.

Normal measurementMeasurement of single vision lenses or bifocal (trifocal) lenses

PPL measurementWhen it is detected that a lens is placed in an inverted position, the icon changes to

to call attention to it.

CL measurement

Auto measurementWhen the lens on the nosepiece is detected as a progressive power lens, the normalmeasurement screen changes to the PPL measurement screen automatically.

When the distance portion of a progressive power lens is placed on the nosepiece, itmay not be detected as a progressive power lens.In such a case, put the intermediate portion (progressive zone) or near portion area onthe nosepiece.

When it is detected that a lens is inverted, the icon changes to to call

attention to it.

Right-eye lens Left-eye lens Single state

Normal measurement

PPL measurement

CL measurement

Auto measurement

Changes the screen to the normal measurement screen.

Changes the screen to the PPL measurement screen.

Changes the screen to the CL measurement screen.Only when the Contact parameter is set to “On”, this icon is displayed.

Changes the screen to the normal measurement screen in auto measurement.

Page 19: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

9

BEFORE USE Displays

1

When the Icon parameter is set to “Fixed”, the icon of the measurement selection button is displayed without toggleas follows:

In addition, extended pressing of the measurement selection button functions as the parameter button .

When the icon changes, release the button to display the parameter screen.

8. Cylinder +/- buttonUsed to simply change the CYL reading between + and -.

Pressing this button again returns to the CYL mode set with the parameter.

Pressing any button other than the read button or the parameter button also returns to the mode set with

the parameter.

When the Prism parameter is set to “P-B” or “BU/D BI/O”, the button can also switch the prism coordinate display.Extended pressing of the button changes the icon display as shown below. Releasing the button correspondingly switches the prism coordinate display.

9. R/L selection buttonUsed to specify the right-eye or left-eye lens.

When the Icon parameter is set to “Fixed”, the icon of the R/L selection button is displayed without toggle as follows:

When R or L is detected by the nose slider with the LM-600PD, following icon (blank) is displayed.

10. Clear button

Used to clear data in the memory.

The fixed data is cleared and the R/L indication returns to .

When the Single parameter is set to “Off”, the R/L indication returns to .

In the setting for measuring contact lenses only, the parameter button is displayed instead.

Only when the Contact parameter is set to “Only”, this icon is displayed.

Changes the screen in the order of Auto measurement ��Normal measurement ��PPL measurement ��CL measurement ��Auto measurement...

(White) Shows that data is displayed in the CYL mode set with the parameter.

(Yellow)Shows that data (SPH, CYL, and AXIS) is displayed after reversing the CYL modeset with the parameter.

(White) Shows that data is displayed in the coordinates set with the parameter.

(Yellow)Shows that data is displayed in the coordinates other than those set with theparameter.

Changes the screen to the right-eye side.

Changes the screen to the left-eye side.

Toggles the screen between the right-eye and left-eye lenses.

Shows that the lens side is being detected by the nose slider. The button is disabled. While this icon is displayed, the lens side cannot be selected by the button.

Page 20: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

10

BEFORE USE Displays

Extended pressing of the clear button functions as the clear button of the Eye Care card.After the button is pressed and held, a beep sounds and the pilot lamp blinks. Releasing the button erases all thedata in the Eye Care card.

11. Print button

Used to print the fixed data out, output it by communication, and write it to the Eye Care card. Functions andindications of the button change depending on the parameter settings.

When this button is pressed before the measured data is fixed, the data is fixed and printed out.

LM-600P/LM-600PD

LM-600

Icons FunctionsParameters

Printer Com modePrints measured data out. Pressing and holding this button for one second or more feeds the paper.

When the device is connected to an external device, the measured data is transmitted at the same time of printing (When the Com mode parameter is set to other than “Off”).When the Eye Care card is inserted, the data is written.

When this button is pressed and held with the clear button

together, the printer paper is fed a little and cut.

On –

Prints measured data using the printer of a connected AR, ARK or RKT.

For details of “Communication print (AR print)”, see “2.11.1 Connecting to the AR, ARK or RKT and setting parameters” (page 70).When the Eye Care card is inserted, the data is written.

AR print NIDEK

Measured data is transmitted to a connected external device.Printing is not performed.When the Eye Care card is inserted, the data is written.

Off

NIDEKPC

NCP10NCP20

Data is written to the Eye Care card.Printing and data transmission are not performed.

Off Off

Icons FunctionsParameters

Printer Com modePrints measured data using the printer of a connected AR, ARK or RKT.

When the device is connected to an external device, the measured data is transmitted at the same time of printing.When the Eye Care card is inserted, the data is written.

AR print NIDEK

Measured data is transmitted to a connected external device.Communication print (AR print) is not performed.When the Eye Care card is inserted, the data is written.

Off

NIDEKPC

NCP10NCP20

Data is written to the Eye Care card.Communication print (AR print) and data transmission are not performed.

Off Off

Page 21: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

11

BEFORE USE Displays

1

� PD indications of the measurement screen (LM-600PD only)

This is the measurement screen to measure the PD using the lens table and nose slider.The PD can be measured on the screen other than CL measurement screen.

The PD of mounted lenses (binocular PD, RPD, LPD) is indicated in 0.5 mm increments.

The PD is measured at the point where the nose slider is located at the press of the readbutton.

This value is displayed only when PD parameter is set to “On”. However, it is not displayedwhen the nose slider is placed where the PD cannot be measured.

� Progressive Power Lens (PPL) measurement screen

This is the screen to measure progressive power lenses.

PD

Crossline

Page 22: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

12

BEFORE USE Displays

� Contact Lens (CL) measurement screen

This is the screen to measure contact lenses.

The auto read setting is not displayed on the setting indication.

In CL measurement, auto read, auto R/L selection, and auto print do not work.A SE value is displayed in addition to lens measured data.

� Explanations about icons of function buttons

Curling up of the lower-left corner of the icon like indicates that the button has multiple func-tions.Extended pressing of the button sounds a short beep. Then, the icon indication changes and the pilotlamp blinks. Releasing the button executes the function displayed by the icon.

For the functions of the extended pressing of each button, refer to the explanations of eachicon.

SE value

Page 23: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

13

BEFORE USE Displays

1

1.4.2 Prism layout screensTwo screens are provided for blocking lenses for heterophoria prescriptions using the prism layoutfunction: Prism entry screen used for entering prism prescriptions and measurement mode screen formeasuring lenses. With the prism layout function, the on-screen position of the target is shifted based on the enteredprism prescriptions. Easy alignment of lenses is allowed only by aligning the target to the center of thealignment circle.In addition, the target changes depending on the alignment conditions in the same manner as themeasurement screen: � (cross) � (large cross). Therefore, an alignment error can bereduced compared to the marking of lenses for prism prescriptions using the measurement screen.

� Prism entry screen

This screen is used for entering prism prescriptions. To display this screen, press the read button andmeasurement selection button simultaneously on the normal measurement screen, PPL measure-ment screen, or auto measurement screen. For models without the UV function, pressing and holdingthe read button for about one second on the normal measurement screen, PPL measurement screen,or auto measurement screen also changes the screen to this screen.This screen allows the operator to mark lenses and to measure lenses with limited functions.

When lenses are measured on the prism entry screen, the CYL mode switching, measureddata clear, and printing are unavailable.

1. Prism layout indication /This icon represents the prism layout screen.

7. UP/DOWN button4. Next screen button

1. Prism layout indication

2. Setting indications

3. Preset prism prescription

5. Prism data button 6. R/L selection button

Page 24: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

14

BEFORE USE Displays

2. Setting indicationsThe indication representing the Auto read setting is not displayed.

On the prism layout screens, Auto read is disabled regardless of the setting.The indications of the abbe number, wavelength, indication step, CYL mode, and state of the Eye Care card aredisplayed in the same manner as those on the measurement screen.

3. Preset prism prescriptionEntered prism prescriptions are displayed.

As soon as prism prescriptions are entered, the target moves opposite to the direction of the entered prism data.

4. Next screen button

Used to switch the screen to the measurement mode screen.

Extended pressing of the next screen button functions as the exit button .Extended pressing of the button changes the icon indication. Releasing the button returns to the measurementscreen.

5. Prism data buttonUsed to switch the type of prism data to be entered or modified. In addition, a current selection is shown.

The indication of the button differs according to selection of the rectangular or polar coordinate. When the type ofprism data is selected, the indication of the UP and DOWN buttons changes:

Extended pressing of the prism data button functions as the coordinate system button / .Used to switch the prism representation method between the rectangular and polar coordinates.Extended pressing of the button changes the icon indication as shown below. Releasing the button correspondinglyswitches the prism coordinate.

6. R/L selection buttonUsed to select whether the entered prism prescriptions are for the right-eye or left-eye lens.

When the Icon parameter is set to “Fixed”, the icon of the R/L selection button is displayed without toggle as follows:

Rectangular coordinates

(DU)Enter the prism power in the Base Down/Up direction with and

.

(IO)Enter the prism power in the Base In/Out direction with and

.

Polar coordinates

(�) Enter an absolute prism value (�: PRISM) with and .

(�) Enter the prism base direction (°: BASE) with and .

(XY) Selects the rectangular coordinate.

(���) Selects the polar coordinate.

The entered prescription is for the left-eye lens.

The entered prescription is for the right-eye lens.

Changes the screen from the currently displayed side to the opposite side.

Page 25: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

15

BEFORE USE Displays

1

7. UP/DOWN buttonUsed to enter prism prescriptions. The indication of the buttons differs according to selection of therectangular or polar coordinate.

Each time the button is pressed, the preset prism data changes in a specified increment. The prism value changescontinuously when the button is pressed and held.Allowable range: Up to 20��of prism data expressed in polar coordinates When the rectangular coordinate is selected, and the data entry of even less than 20� may be rejected, or the prismdata of the other prism base side* may be modified to confine the prism data to 20� expressed in polar coordinates.* For example, if the prism data is modified in the Base IN/OUT direction, the prism data in the Base UP/DOWNdirection may be modified.

Page 26: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

16

BEFORE USE Displays

� Measurement mode screen

This screen is used for measuring and marking lenses in accordance with preset prism prescriptionsentered on the prism entry screen. Enter the measurement mode screen through the prism entryscreen.

The cylinder +/- button and the print button function in the same manner ason the measurement screen.

8. Preset prism prescriptionPrism prescriptions entered on the prism entry screen are displayed.

During lens measurement, the preset prism data remains unchanged. The measured prism data is printed out.

9. Previous screen button

Used to return to the prism entry screen.

Extended pressing of the previous screen button functions as the exit button .Extended pressing of the button changes the icon indication. Releasing the button returns to the measurementscreen.

10. Clear buttonUsed to erase the measured data that has been saved.

The preset prism prescriptions are not cleared. To clear the preset prism data, return to the prism entry screen.Extended pressing of the clear button functions as the clear button of the Eye Care card.

After the button is pressed and held, a beep sounds and the pilot lamp blinks. Releasing the button erases all thedata in the Eye Care card.

9. Previous screen button

8. Preset prism prescription

10. Clear button

Page 27: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

17

BEFORE USE Displays

1

1.4.3 UV measurement screen (UV model only)This is the screen to display the UV (ultraviolet rays) transmittance. This screen is displayed onlywhen the UV display parameter is set to “Compare”.The UV measured result is displayed visually and emphatically. The UV protective effect can be com-pared after additional measurement.

1. UV transmittance indicationDisplays the UV transmittance in percentages.

The UV transmittance is displayed in the range of 0 to 100% by increments of 5% (or 1% dependingon the parameter setting).

2. UV 100% correction buttonUsed to correct the UV transmittance to 100%. This button is used when 100% is not displayed without thelens on the nosepiece.

Extended pressing of the button for about one second starts correction. The correction is not renewedwhen the lens is set.

3. Remeasurement buttonUsed to measure the UV transmittance again.

4. Target indicationThe target is displayed for lens alignment.

The range inside the circle is within 2�. When the target is in circle, the target changes from circle tocross.

1. UV transmittance indication

6. Exit button

4. Target indication

2. UV 100% correction button5. Compare measurement

button3. Remeasurement button

Page 28: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

18

BEFORE USE Displays

5. Compare measurement button

Used to compare with the UV transmittance of other lenses.The UV compare screen is displayed when the comparemeasurement button is pressed after the lens on the nosepiece ischanged.The right side 2) is a result of the additional measurement.Additional measurement can be performed repeatedly.

6. Exit buttonUsed to return to each measurement screen.

� When the UV display parameter is set to “Simple”

The UV transmittance (%) is displayed on thelower part of the screen.

The screen cannot be switched to the UV com-pare screen.

UV compare screen

UV transmittance indication (%)

Page 29: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

19

BEFORE USE Displays

1

1.4.4 Parameter screenThis is the screen to set parameters related to each setting of the device.

1. ParametersParameters are displayed.

2. Parameter settingsParameter settings are displayed.

The contents within the frame are those currently selected for setting.

3. Exit buttonUsed to save current settings and change the parameter screen to the measurement screen.

4. Back page buttonUsed to display the previous page of the parameter screen. There are six pages (seven pages for the UVmodel).

5. Forward page button

Used to display the next page of the parameter screen.

6. Up button /Down button

Used to select a parameter.The read button toggles the icon.

7. Selection button

Used to change parameter settings.Move the frame to change the parameter settings by pressing this button.The read button toggles the icon. The icon indication changes depending on the parameter settings.

1. Parameters 2. Parameter settings

3. Exit button

5. Forward page button 6. Up/down button

7. Selection button

Page No.

4. Back page button

Page 30: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

20

BEFORE USE Labels

1.5 Labels

To call attention to operators, some labels and indications are provided on the device.If labels are curling up or characters fade and become barely legible, contact NIDEK or your autho-rized distributor.

Indicates that caution must be taken. Refer to the operator’s manual before use.

Indicates the state of the power switch. If the symbol side of the switch is presseddown, power is not supplied to the device.Indicates the state of the power switch. If the symbol side of the switch is presseddown, power is supplied to the device.

Indicates the contrast knob that is used for adjusting the contrast on the display.

Indicates that the device must be supplied only with alternating current.

Indicates the date of manufacture.

Indicates the manufacturer.

Indicates that this product shall be disposed of in a separate collection of electrical andelectronic equipment in EU.

Page 31: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

21

BEFORE USE Checking Contents

1

1.6 Checking Contents

Unpack the contents from the shipping carton and check them.

The following is included in the standard configuration.

AUTOMATISCHER

FRONTOFOCÓMETRO AUTOMÁTICO

FRONTO-FOCOMETRE AUTOMATIQUEFRONTIFOCOMETRO AUTOMATICO

SCHEITELBRECHWERTMESSER

AUTOMATISCHER

FRONTOFOCÓMETRO AUTOMÁTICO

FRONTO-FOCOMETRE AUTOMATIQUEFRONTIFOCOMETRO AUTOMATICO

SCHEITELBRECHWERTMESSER AUTOMATISCHER

FRONTOFOCÓMETRO AUTOMÁTICO

FRONTO-FOCOMETRE AUTOMATIQUEFRONTIFOCOMETRO AUTOMATICO

SCHEITELBRECHWERTMESSER

LM-600P/LM-600PD

LM-600

UV model only

[Rear side]

• Main body • Nosepiece for contact lenses • Printer paper (3 rolls): LM-600P/LM-600PD only • Power cord • Dust cover • Operator’s manual (this book) • Measuring Progressive Power Lenses explanation sheet

Page 32: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

22

BEFORE USE Before First Use

1.7 Before First Use

Put the device on a stable table and connect the power cord.

1 Put the device on a stable table.

2 Lay the device down gently.

3 Connect the power cord to the power inlet.

4 Stand up the device.Draw the power cord from the indentation atthe rear of the device so as not to be caught.

5 Connect peripheral devices if necessary.For details of connecting peripheral devices, see “3 OPERATION WHEN PERIPHERALDEVICES ARE CONNECTED” (page 91).

6 Make sure that the power switch is turned off ( ) and plug the power cord in the walloutlet.

Power inlet

CAUTION • The electrical outlet must have a grounding terminal.Electric shock or fire may occur in the event of device malfunction or power leakage.

Indentation

Power switch

Page 33: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

23

BEFORE USE Before First Use

1

7 Turn the power switch on ( ).The initial screen is displayed on the display and the device is initialized. Wait for a while until the screen is changed to the measurement screen.

8 Make sure that the measurement screen is dis-played.

The measurement screen displayed at power-oncan be set with the Intl. screen parameter.

• If an error appears, it may be due to disturbance light.Avoid installing the device on a place where it is exposedto direct sunlight or near the illumination. In particular,ensure that the device is not illuminated by a spot lightfrom the upper front.

In particular, measured data of lenses of 10� or more islikely to be affected adversely by disturbance light.

In such cases, relocate or reorient the device.

• Before putting the device on a light-reflecting surface such as a glass showcase or a shinytable, put a cloth on it.

Reflected light such as illumination may adversely affect measurements.

Initial screen Initial screen of UV model

Auto measurement screen

Spot light

Page 34: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

24

BEFORE USE Before First Use

9 When the screen is too bright or too dark, adjustthe brightness with the contrast knob.

10 Set the supplied printer paper. (LM-600P/LM-600PD only)For details of the setting procedure, see “4.3 Replacing Printer Paper (LM-600P/LM-600PDonly)” (page 98)."

This is all you have to do before use.

• Set the parameters to suit operator’s needs or preferences.For details of the parameters and setting procedure, see “2.14 Setting Parameters” (page 76).

• For details of connecting peripheral devices, see “3 OPERATION WHEN PERIPHERALDEVICES ARE CONNECTED” (page 91).

Contrast knob

Page 35: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

25

2

2. OPERATING PROCEDURE

2.1 Operation Flow

Turning ON the device

2.2 Preparation for Measurement (page 26)Turn the power switch on and set measurement conditions.

Measurement

2.3 Setting Lenses (page 29)

2.4 Measuring Single Vision Lenses (page 35)

2.5 Measuring Bifocal Lenses (page 38)

2.6 Measuring Progressive Power Lenses (page 43)

2.7 Measuring Prism Lenses (page 52)2.8 Measuring Contact Lenses (page 54)

Marking

2.10 Marking (page 60)2.10.1 Marking at the optical center (page 60)2.10.2 Marking for prism prescription (Prism Layout) (page 62)2.10.3 Marking for prism prescription (Measurement screen) (page 67)

Printing/Data recording

2.11 Printing (LM-600P/LM-600PD only) (page 69)2.12 Data save to the Eye Care card (page 72)Transcribe the measured data onto a prescription or such.For transmitting data to connected devices:

3 OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED (page 91)

Turning OFF the device

2.13 After Use (page 75)

• The operating procedure for printing the measured data with the LM-600P/LM-600PD isdescribed. Writing of fixed data to Eye Care card, communication print (AR print), and datatransmission are the same procedure.

Page 36: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

26

OPERATING PROCEDURE Preparation for Measurement

2.2 Preparation for Measurement

1 Make sure that the power switch is turned off ( ) and plug the power cord in the walloutlet.

2 Turn the power switch on ( ).The initial screen appears.

Wait for a while until the screen changes to themeasurement screen.

3 After initialization, the screen changes to themeasurement screen.

CAUTION• The electrical outlet must be equipped with a grounding terminal.Electric shock may occur in the event of trouble or power leak.

Initial screen

Auto measurement screen

• The measurement screen displayed at power-on can be set with the Intl. screen parameter.

• Do not turn on the power with a lens placed on the nosepiece.If a lens is on the nosepiece before the measurement screen appears, the 0D initialization errormessage appears. Remove the lens and turn on the power again.

Page 37: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

27

OPERATING PROCEDURE Preparation for Measurement

2

� Measurement screen at power-on

According to the “Intl. screen” and “Contact” parameters, and the type of the attached nosepiece, themeasurement screen displayed after initialization differs.

• Settings of the Intl. screen parameter

* The factory setting is “Auto”.

* According to the setting of the Contact parameter, the different measurement screen may appear.

• Settings of the Contact parameter• When the parameter is set to “On” or “Off”:

Except for the following cases, the measurement screen set with the Intl. screen parameterappears.

• When the Contact parameter is set to “Off” and the Intl. screen parameter is set to “Contact“, theauto measurement screen appears.

• When the Intl. screen parameter is set to otherthan “Contact”, and the power is turned on withthe nosepiece for contact lenses on, the “Usecontact nosepiece?” message appears on theinitial screen.

Parameter settings Measurement screen displayed at start-up

Auto Normal measurement screen in auto measurement

Normal Normal measurement screen

Progressive PPL measurement screen

Contact CL measurement screen

Restart button Execute button

To measure lenses of glasses

Change the nosepiece to the standard one, then press the restartbutton .

The measurement screen set with the Intl. screen parameter appears.

To measure contact lenses only

Press the execute button to change the screen to the CLmeasurement screen.

In such a case, the CL measurement screen cannot be changed to theauto measurement screen, normal measurement screen, or PPLmeasurement screen unless the nosepiece is replaced and the power isturned off and on again.

Page 38: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

28

OPERATING PROCEDURE Preparation for Measurement

• When the parameter is set to “Only”:

Regardless of the setting of the Intl. screen param-eter, the CL measurement screen appears.

Make sure that the nosepiece is for contact lenses.

� Details of target

The target form displayed on the measurement screen can be selected with the Target parameter.

Settings of the Target parameter

CL measurement screen

Parameter setting Details of target to be displayed

(Normal target)

The target shows the distance and direction from the center of thenosepiece.The direction and movement distance are constant regardless of therefractive power of lenses.

The target changes � (within 0.5�) � (within ø0.8

mm) as it gets close to the center.

(Prism target)

The target moves based on prisms as with the eyepiece-type orprojection-type lensmeter.The target indicates the axis while rotating.

The target changes � (within 0.5�) � (within ø0.8

mm) as it gets close to the center.In measurement of progressive power lenses, the target is always displayed

with .

• This operator’s manual describes each screen with the Target parameter at “ “.

When “ ” is selected When “ ” is selected

Page 39: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

29

OPERATING PROCEDURE Setting Lenses

2

2.3 Setting Lenses

2.3.1 Setting uncut lenses

1 Set a lens on the nosepiece.Place the lens center on the nosepiece with theconvex side up.

2 Fix the lens with the lens holder.Lift the lens holder lever to unlock, then lower itslowly to fix the lens.

• Make sure that there are no dusts on the tips of the nosepiece and the lens holder beforesetting lenses.

It may damage the lenses.

• When a uncut lens is set, the lens table does not need to be used.

• Set the lens with the upper side of the lens towards the operator.For the LM-600/LM-600P/LM-600PD, the zero degree direction of the prism base is on the opera-tor’s left as viewed from the front.

• When there are still scratches on the lenses after cleaning of the nosepiece and lensholder, request the maintenance such as part replacement.

Page 40: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

30

OPERATING PROCEDURE Setting Lenses

• Check that the nose slider is lifted uprighton the far left.

Detection by the nose slider position has ahigher priority to select R or L. In such acase, the R/L selection button ( or

) is disabled.

Nose slider

Page 41: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

31

OPERATING PROCEDURE Setting Lenses

2

2.3.2 Setting mounted lenses

1 Set mounted lenses.Place the frames on the nosepiece with the frontsurface up.

2 Set the lens table.Pull the lens table lever towards the operator until it touches the bottom of the frames.

3 Fix the lens with the lens holder.Lift the lens holder lever to unlock, then lower itslowly to fix the lens.

• When measuring glasses in use or processed lenses, remove dusts and processing wasteon the lens before measurement.

Otherwise, foreign matters may be attached to the nosepiece or lens holder and damage thelenses.

• Touch the bottom of both frames onto the lens table.If the frames are off the lens table, an error may occur in the axis value.

• Check that the nose slider is lifted uprighton the far left.

Detection by the nose slider position has ahigher priority to select R or L. In such acase, the R/L selection button ( or

) is disabled.

Nose slider

Page 42: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

32

OPERATING PROCEDURE Setting Lenses

2.3.3 Setting for PD measurement (LM-600PD only)For mounted lenses, the optical center distance betweenthe right and left lenses can be measured.

1 Set the nose slider.Move the lens table forward, and pull down thenose slider.

2 Place the frames on the nose slider.Place the nosepads of the frames on the noseslider.

3 Set the lens on the nosepiece.The left-eye or right-eye lens is automaticallydetected by the nose slider position. Blinking

or indicates which lens is beingmeasured.

• Precise PD measurement may be difficult with glasses with cylinder-only lenses, highpower lenses, or sharply warped frames.

LPD RPD

PD

Nose slider

Page 43: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

33

OPERATING PROCEDURE Setting Lenses

2

4 Set the lens table.Pull the lens table lever towards the operator until it touches the bottom of the frames.

5 Fix the lens with the lens holder.Lift the lens holder lever to unlock, then lower itslowly to fix the lens.

6 When the read button is pressed after lens align-ment or measurement is finished with the autoread function, the PD data is fixed.

7 Changing between R and L displays the binocu-lar PD.

8 When the read button is pressed after lens alignment or measurement is finished withthe auto read function, the PD data is fixed.

• Touch the bottom of both frames onto the lens table.If the frames are off the lens table, an error may occur in the axis value.

• In PD measurement, when the Auto RD R/L parameter is set to (cross), the parametersetting changes to (large cross) automatically.

• See “2.11 Printing (LM-600P/LM-600PD only)” (page 69) for the printout sample after PDmeasurement.

Page 44: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

34

OPERATING PROCEDURE Setting Lenses

� Message during measurement (LM-600PD)

When either of the following operations is performed during R or L detection by the nose slider, a cau-tion message appears.

During R or L detection: The R/L selection button is displayed blank .

• When the R/L selection button is pressed.• When the lens side can be switched automatically after the lens is measured and removed with

the Auto R/L parameter set “On”.

R or L detection by the nose slider has higher priority than the other operations. When themessage appears, adjust the nose slider position or move it to the far left so that the slider isoff the detectable range.

Page 45: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

35

OPERATING PROCEDURE Measuring Single Vision Lenses

2

2.4 Measuring Single Vision Lenses

Single vision lenses are measured on the auto measurement screen or normal measurement screen.

1 Specify the lens side if necessary.Press the L-selection button or R-selec-tion button to specify the left-eye lens orright-eye lens.

The R or L indication blinks indicating the mea-surement lens side.

2 Perform alignment of a lens.Move the lens to bring the target ( ) close tothe center of the alignment circle.

When the target comes within a range of 0.5Dfrom the center, the target changes from to

(cross).

For mounted lenses, move the lens table along the frames. When alignment is complete,make sure that the bottom of the frames is touched to the lens table.

3 Press the read button.The measured data is fixed.

• When the lens side is specified after measurement in the single state (R/L not specified),the measured data is cleared. Specify the lens side before measurement.

• When the target changes from to (cross), correct measurement can be performedin normal measurement.

When marking a lens, perform more correct alignment until (large cross) appears.

• The auto read function can be selected with the “Auto RD S” or “Auto RD R/L” parameter.

Page 46: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

36

OPERATING PROCEDURE Measuring Single Vision Lenses

• Fixing measured data

After the measured data on the screen is fixed and the data is saved in the memory, the R/Lindication is inverted.

R/L indication when data is fixed

The target is not fixed. In order to retry measurement, align the target and press the readbutton again.

The CYL mode can be changed with the cylinder +/- button even after data is fixed.

• Fixing measured data with the auto read function

When the “Auto RD S” or “Auto RD R/L” parameter is set to “ ” or “ ”, the data is auto-matically fixed after the target is aligned as well as when the read button is pressed.

To retry measurement, align the target and press the read button again.

When the “Auto RD S” and/or “Auto RD R/L” parameter is set to “ ”, the measured data isautomatically fixed after the target is aligned. When alignment is continued until the targetchanges to “ ”, auto read is performed again.

4 Measure the other lens if necessary (in the R/L state).Press the L-selection button or R-selection button to specify the left-eye lensor right-eye lens.

Follow the same steps as the first lens.

5 Press the print button to print out the measured results (or transmit the data orwrite the fixed data to the Eye Care card).

For details of printing, see “2.11 Printing (LM-600P/LM-600PD only)” (page 69).

Shows that measured data of the lens in the single state is saved in the memory.

Shows that measured data of the left-eye lens is saved in the memory.

Shows that measured data of the right-eye lens is saved in the memory.

Auto RD S Auto read function in the single state

Auto RD R/L Auto read function in the R/L state

• In the R/L state, when the Auto R/L parameter is set to “On”, R and L are switchedautomatically after the measured data is fixed and the lens is removed.

• When the Auto prt. S parameter is set to “On”, the measured data is automatically printedout after the data is fixed in the single state and the lens is removed.

• When the Auto Prt. R/L parameter is set to “On”, the measured data is automaticallyprinted out after the data of both eyes is fixed and the lens is removed.

Page 47: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

37

OPERATING PROCEDURE Measuring Single Vision Lenses

2

� Changing CYL mode

Pressing the cylinder +/- button switches the CYL mode between + and -. Pressing the buttonagain returns to the preset mode.

� Changing prism indication

To display the prism value, set the Prism parameter to “P-B” or “BU/D BI/O”.When this parameter is set to “Off“, the measured prism value is not displayed.

When marking with the prism prescription, follow the prescription.

� Guide mark

To align the target precisely, a guide mark ( ) is displayedin the alignment circle.It is displayed by setting the Guide parameter to “On”.

Align the target by moving the lens in the direction in whichthe guide line is displayed.Proper alignment makes the guide mark disappear.

(White) Shows that data is displayed in CYL the mode set with the parameter.

(Yellow)Shows that data (SPH, CYL, and AXIS) is displayed after reversing the CYL modeset with the parameter.

P-BThe measured prism value is displayed by polar coordinates.

Absolute value of prism (�)Direction of base (°)

BU/D BI/O

The measured prism value is displayed by rectangular coordinates.Base InBase OutBase UpBase Down

Off The measured prism value is not displayed.

• When the Prism parameter is set to “Off”, measured values are displayed larger than whenprism measured values are displayed.

For a display sample, see “2.5 Measuring Bifocal Lenses” (page 38).

Page 48: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

38

OPERATING PROCEDURE Measuring Bifocal Lenses

2.5 Measuring Bifocal Lenses

Bifocal lenses (or trifocal lenses) are measured sequentially in the order of the distance portion �near portion (for trifocal lenses, distance portion � intermediate portion � near portion).

1 Specify the lens side if necessary.Press the L-selection button or R-selection to specify the left-eye lens or right-eye lens.

2 Bring the distance portion onto the nosepiece.

3 Measure the distance power.After the target changes from to (cross),press the read button.

The distance power is fixed.

Distance portion

Intermediate portion

Near portion

Bifocal lens Trifocal lens

• In measurement of the distance portion, auto read works according to the setting of the“Auto RD S” or “Auto RD R/L” parameter.

Page 49: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

39

OPERATING PROCEDURE Measuring Bifocal Lenses

2

4 Display the near ADD power (Add: 1st ADDpower).

Pull the lens towards the operator to bring thenear portion onto the nosepiece.

When the ADD power is detected, the auto ADDmeasurement function places the device intoADD power measurement.

For trifocals, bring the intermediate portion ontothe nosepiece.

5 Press the read button.The measured data of the ADD power (Add) isfixed.

Auto read is disabled so press the read button tofix the measured data.When the measured prism value is notdisplayed, “Ad” is displayed.

For bifocal lenses, measurement is complete.

For trifocals, proceed to Steps 6 and 7.

• It is not necessary to align the target.

• Measure the near ADD power without removing the lens from the nosepiece.If the lens has been removed once, restarting distance measurement is necessary.

• For trifocal lenses, when the vertical length of the intermediate portion is narrow, themeasuring beam may interrupt the boundary and correct measurement may not beperformed.

Intermediateportion

Page 50: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

40

OPERATING PROCEDURE Measuring Bifocal Lenses

6 Bring the near portion onto the nosepiece.The near ADD power (Ad2: 2nd ADD power)appears on the screen.

When the measured prism value is not dis-played, “A2” is displayed.

7 Press the read button.The near ADD power (Ad2) is fixed.

Auto read is disabled so press the read button to fix the measured data.

8 Measure the other lens if necessary (in the R/L state).Press the L-selection button or R-selection button to specify the left-eye lensor right-eye lens.

Follow the same steps as the first lens.

9 Press the print button to print out the measured results (or transmit the data orwrite the fixed data to the Eye Care card.).

For details of printing, see “2.11 Printing (LM-600P/LM-600PD only)” (page 69).

� Sphere indication for near portion

By setting the Near parameter on the parameter screen to“Near Sph”, the measured data of the near portion (ADD)is displayed by sphere instead of ADD.

N: 1st near power (distance power + 1st ADDpower)

2: 2nd near power (distance power + 2nd ADDpower)

• It is not necessary to align the target.

• Auto R/L selection works as in the same case with single vision lenses.

Page 51: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

41

OPERATING PROCEDURE Measuring Bifocal Lenses

2

� Measuring ADD power more accurately

These is the procedure for measuring the ADD power without including an error in the measured datamade by the distance between the segment and nosepiece.It is especially effective when the base lens is thick.

1 Set the distance portion with the concave side upas shown to the right.

Bring the opposite side of the segment positionfrom the midpoint of the lens center onto thenosepiece.

2 Press the read button.The distance power is fixed.

3 Set the near portion.Pull the lens towards the operator to bring thenear portion onto the nosepiece.

The ADD power (Add) appears.

• It is not necessary to align the target.

• This measured data is only a reference for obtaining the ADD power and is not the actualdistance power.

Record the distance power in normal measurement with the convex side up.

• It is not necessary to align the target.

Page 52: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

42

OPERATING PROCEDURE Measuring Bifocal Lenses

4 Press the read button.The measured data of the near ADD power(Add) is fixed.

Auto read is disabled so press the read button tofix the measured data.

Page 53: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

43

OPERATING PROCEDURE Measuring Progressive Power Lenses

2

2.6 Measuring Progressive Power Lenses

2.6.1 Measuring uncut lensesMeasure a lens at the marks of the distance portion and near portion printed on the lens in the samemanner as “2.5 Measuring Bifocal Lenses” (page 38).

2.6.2 Measuring mounted lensesMeasure a lens on the PPL measurement screen or auto measurement screen.

On the auto measurement screen, when the lens on the nosepiece is detected as a pro-gressive power lens, the screen changes to the PPL measurement screen automatically.

After the distance and near portions are detected according to power changes in the progressivepower lens, the guides (target and arrow) to align the lens appears on the screen.

1) Move the lens in the direction of the arrow toroughly align the lens position.

2) Precisely align the target with the crossline.

• A lens must be placed with its horizontal reference line parallel to the lens table.

• If the marks printed on the lens interrupt the measuring beam, the target may wiggle.In such a case, measure the lens when the target stabilizes by moving the lens slightly.

Distance power measuring position

Near power measuring positionHorizontal reference line

Distance portion

Intermediate portion (progressive zone) Power changes progressively.

Near portion

Arrow

Crossline

Target

Page 54: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

44

OPERATING PROCEDURE Measuring Progressive Power Lenses

1 Change the screen to the PPL measurementscreen.

Press the measurement selection button on the normal measurement screen to displaythe PPL measurement screen.

To switch from the auto measurement screen tothe PPL measurement screen, see “� Screenchange in auto measurement” (page 48).

2 Specify the lens side if necessary.Press the L-selection button or R-selection button to specify the left-eye orright-eye lens.

3 Set mounted lenses.See “2.3.2 Setting mounted lenses” (page 31).

The following is the measurement procedure for the right-eye lens.

4 Measure the distance power.1) Bring the intermediate portion (progressive

zone) onto the nosepiece.

Place a portion that is slightly lower than the centerof the lens on the nosepiece.

The target ( ) indicating the distance portion isdisplayed.

The arrow shows the direction in which the lensshould be moved.

• Tips on moving the lens: To measure progressive power lenses accurately, pay attention to the following:

• Move the lens vertically after horizontal alignment of the lens.

• Move the lens while the frames are constantly in contact with the lens table.

• Ensure that the lens rear surface is constantly in contact with the nosepiece.

Page 55: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

45

OPERATING PROCEDURE Measuring Progressive Power Lenses

2

2) Move the lens in the horizontal direction toalign the target with the vertical line of thecrossline.

3) Slowly move the lens inward (upward on thescreen) referring to the target to align it withthe crossline.

When the target is off the vertical line of thecrossline in the horizontal direction while thelens is moved, move the lens in the horizontaldirection.

When the vertical position is roughly aligned,the arrow disappears.

4) Precisely adjust the lens vertically and hori-zontally to align the target with the crosslineuntil the target changes from to (cross).

5) When the target changes to (cross) andmeasured data stabilizes, the measured datais automatically fixed.

The “Distance measurement complete” messageappears. The screen changes to ADD power mea-surement for the near portion automatically.

• When the Dist. auto RD parameter is set to “Off”, press the read button when the targetchanges to (cross) in Step 5) and the measured data stabilizes.

• When the distance portion cannot be detected in normal measurement, bring the distanceportion onto the center of the nosepiece, and then press the read button.

• For progressive power lenses corrected with the horizontal prism, the position of 3 mmupper from the pupil for the distance vision with glasses is marked as distance portion. Insuch a case, press the read button manually.

The near portion can be measured in the normal procedure.

Page 56: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

46

OPERATING PROCEDURE Measuring Progressive Power Lenses

5 Measure the near ADD power.1) The target indicating the near portion

appears.

The arrow shows the direction in which the lensshould be moved.

2) Move the lens slowly in the direction of thearrow (towards the operator).

When the intermediate portion (progressive zone)is detected, the “Conducting near measurement...”message appears and the current ADD power(Add) appears.

The indicator for near portion (�) appears above,below, on the right of, or on the left of the target,which shows the direction and movement distanceto move the lens.

Move the lens in the direction of (�).

The number of indicators (up to three) shows how much the lens should be moved.

3) Align the lens horizontally.

Move the lens in the horizontal direction untilthe indicator for near portion (�) on the leftor right of the target disappears.

4) Align the lens vertically.

Move the lens in the vertical direction untilthe indicator for near portion (�) above orbelow the target disappears.

• When the indicator for near portion appears on the left or right of the target, performhorizontal alignment again.

Perform vertical alignment of the lens ensuring that the indicator for near portion is not displayedon the left or right of the target.

• When the indicator for near portion above the target does not disappear:For small vertical frames, the indicator for near portion may not disappear due to steep variationsin lens powers even though the lens is moved towards the operator.

In such cases, press the read button to measure the lens with the lens table at the front.

Move the lens in the vertical direction afterhorizontal alignment of the lens.

Page 57: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

47

OPERATING PROCEDURE Measuring Progressive Power Lenses

2

5) Precisely adjust the lens vertically and hori-zontally to align the target to the crosslineuntil the target changes from to (cross).

6) When the target changes to (cross) andthe measured data stabilizes, the data is fixedautomatically.

The “Near measurement complete” messageappears.

6 Measure the other lens if necessary (in the R/L state).Press the L-selection button or R-selection button to specify the left-eye orright-eye lens.

Follow the same steps as the first lens.

7 Press the print button to print out the measured results (, transmit the data, orwrite the fixed data to Eye Care card.).

For details of printing, see “2.11 Printing (LM-600P/LM-600PD only)” (page 69).

• Whether to display the indicator for nearportion (�) can be set with the Indicatorparameter.

When the Indicator parameter is set to“Off”, align the lens referring to the target.When the vertical position is aligned, thearrow disappears.

After horizontal and vertical alignment ofthe lens, the target changes from to

(cross).

• When the Near auto RD parameter is set to “Off”, press the read button when the targetchanges to (cross) in Step 6) and the measured data stabilizes.

• When the near portion cannot be detected in normal measurement, bring the near portiononto the center of the nosepiece, and then press the read button.

Page 58: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

48

OPERATING PROCEDURE Measuring Progressive Power Lenses

� Measurement screen in the single state

When the lens side is not specified (single state), a round lens is displayed in form on the screen.The measurement procedure is the same as when thelens side is specified.

� Screen change in auto measurement

When the lens on the nosepiece is detected as a progressive power lens on the auto measurementscreen (normal measurement), the screen changes to the PPL measurement screen (auto measure-ment mode) automatically.

1) Set a progressive power lens onto the nose-piece on the auto measurement screen.

Place the intermediate portion (progressive zone)that is slightly lower than the center of the lens ornear portion area onto the nosepiece.

• Auto R/L selection works after ADD power measurement.

• When the distance portion is placed on the nosepiece, the lens may not be detected as aprogressive power lens.

• The progressive power lens detection is continuously performed while the lens is set on thenosepiece in auto measurement mode.

Page 59: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

49

OPERATING PROCEDURE Measuring Progressive Power Lenses

2

2) The screen changes to the PPL measure-ment screen automatically.

After lens measurement, pressing the print button or clear button returns the screen tothe auto measurement screen (normal measurement).

� Progressive power lenses for small vertical frame

When the height of the processed lens is too small for the near portion to be included, the displaychanges to alert that the proper near power has not been measured.When the edge of the lens is found before the near portionis detected, a small lens image is displayed on the screen.At this time, the lens image and measured ADD value aredisplayed in red.

The near ADD power is fixed by the auto readfunction, and measurement completes.

The ADD value is printed out with an exclamation point forattention.

• If the lack of near portion is detected when the Near auto RD parameter is set to “Off”, thenear ADD power is automatically read.

An exclamation point is added.

Page 60: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

50

OPERATING PROCEDURE Measuring Progressive Power Lenses

When the height of the processed lens is small and near portion is lacked, the ADD power cannot bemeasured with the usual auto read function.

For such lenses, measure the point which is nearest to thenear portion with the auto read function.

The measurement point is out of the near por-tion, so the measured data is not near power ofthe original uncut lens.

The measuring procedure is same as normal lenses, but note the following things.

• Be sure to move the lenses in the vertical direction when the indicator for near portion ( )is not displayed on the right and left sides.Otherwise, the ADD power may be underestimated.

• Align the nosepiece so that more than half of it is covered with a lens.When the nosepiece is widely out of the lens, a measurement error occurs.

• The lack amount of the ADD power cannot be judged. Consider the additional poweracquired with this function as lower than actual.

When the height of the processed lens is sufficient

When the height of the processed lens is small

The near portion is lacked from a lens shape.

Measurement point (Nosepiece position)

Page 61: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

51

OPERATING PROCEDURE Measuring Progressive Power Lenses

2

� PPL measurement screen in PD measurement (LM-600PD only)

When PD measurement is performed on the PPL measurement screen, messages such as “Conduct-ing distance measurement...”, “Distance measurement complete”, and “Near measurement complete”and the PD value is alternately displayed.

While “Conducting near measurement...” is displayed, the PD value is not displayed.

� PD indication in measurement of progressive power lenses (LM-600PD only)

The PD displayed in red indicates that PD measurement is not performed correctly due to a lens withsimple astigmatism at the distance portion or equivalent.

When the PD is displayed in red, the position of the noseslider is displayed as the PD data. Refer to the value.

The PD data is printed out with “*” for attention.

These displays are al-ternately switched

Message display PD value display

“*” indication

Page 62: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

52

OPERATING PROCEDURE Measuring Prism Lenses

2.7 Measuring Prism Lenses

There is the procedure to measure prism power of mounted prism lenses.Set the Prism parameter in advance.

1 Mark the pupil center (eye point).Instruct the customer to wear glasses and lookstraight. Mark the pupil center on each lens.

2 Specify the lens side.Press the L-selection button or R-selection button to specify the left-eye orright-eye lens.

3 Set mounted lenses.Place the frames on the nosepiece with the lensconvex side up.

4 Hold the lens with the lens holder.

P-B The prism measured value is displayed by polar coordinates.

BU/D BI/O The prism measured value is displayed by rectangular coordinates.

Off The prism measured value is not be displayed.

• The mark should be 1 mm or smaller in diameter.If the mark is too big, it may interfere with the measurement.

Page 63: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

53

OPERATING PROCEDURE Measuring Prism Lenses

2

5 Bring the eye point marked in Step 1 to the cen-ter of the nosepiece.

6 Press the read button.The measured data is fixed.

7 Measure the other lens if necessary.Press the L-selection button or R-selection button to specify the left-eye orright-eye lens.

Follow the same steps as the first lens.

8 Press the print button to print out the measured results (, transmit the data, orwrite the fixed data to Eye Care card.).

For details of printing, see “2.11 Printing (LM-600P/LM-600PD only)” (page 69).

• It is not necessary to align the target.

• When the “Measurement error” message appears, the mark may interrupt the measuringbeam. Move the lens slightly to achieve the measured data.

Page 64: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

54

OPERATING PROCEDURE Measuring Contact Lenses

2.8 Measuring Contact Lenses

1 Replace the nosepiece with the one for contactlenses.

The top of the nosepiece for contact lenses issmaller than the standard one.

2 Change the screen to the CL measurementscreen.

Press the measurement selection button on the PPL measurement screen to change thescreen to the CL measurement screen.

3 Specify the lens side if necessary.Press the L-selection button or R-selec-tion button to specify the left-eye or right-eye lens.

• When the Contact parameter is set to “Only”, the measurement selection button isdisabled.

The mode is restricted to the CL measurement.

• When the Contact parameter is set to “Off”, the screen cannot be changed to the CLmeasurement screen.

Select “On” or “Only”.

Page 65: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

55

OPERATING PROCEDURE Measuring Contact Lenses

2

4 Set a contact lens.Put the lens on the nosepiece with the convexside up.

For soft contact lenses, set the lens after remov-ing moisture from the surface with a soft cloth.

5 Align the contact lens.Align the contact lens, pushing its end lightly withtweezers tips.

6 Press the read button.In CL measurement, a SE value is displayedalong with the S, C and A measured data.

• SE (Spherical Equivalent) value

This is the half of the cylinder value added to the sphere value. When a non-cylindrical con-tact lens is measured and still a cylinder value is detected, the SE value will be more reliablethan the SPH value to know the total sphere value.

It reduces the error in the measured data made by the unintended cylinder value.

CAUTION• Hold a contact lens with tweezers or fingers. Be careful not to damage the lens.

• Only use tweezers with round tips.

• Do not press the contact lens down with the lens holder.It damages the contact lens.

• In CL measurement, auto read is disabled regardless of the parameter setting.Press the read button for measurement.

Page 66: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

56

OPERATING PROCEDURE Measuring Contact Lenses

7 Remove the contact lens from the nosepiece.

8 Measure the other lens if necessary.Press the L-selection button or R-selection button to specify the left-eye orright-eye lens.

Follow the same steps as the first lens.

9 Press the print button to print out the measured results (, transmit the data, orwrite the fixed data to Eye Care card.).

For details of printing, see “2.11 Printing (LM-600P/LM-600PD only)” (page 69).

• In CL measurement, auto print is disabled.

• Measure a soft contact lens as quickly as possible before the lens surface becomes dry.Since the lens contains water and is made of soft material, the lens cannot stay spherical for a longtime, altering the measured data.

Page 67: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

57

OPERATING PROCEDURE Measuring UV Transmittance (UV model only)

2

2.9 Measuring UV Transmittance (UV model only)

A UV (ultraviolet rays) transmittance is measured with the central wavelength of 365 nm (UV-A) bypercentage. The UV transmittance can be measured on all measurement screens.To measure the UV transmittance, set the UV, UV step, UV display, Auto correct parameters.

To measure the UV transmittance, set the UV parameter to “On” (initial setting) or “With“.

For details of parameter setting, see “2.14 Setting Parameters” (page 76).

1 Set a lens on the nosepiece.

2 Align the distance portion of the lens.Move the lens to bring the target ( ) close to the center of the alignment circle.

To maintain the measurement accuracy of the UV transmittance, the measurement range isrecommended to be within 2�.

3 Press the read button for about one second.UV measurement is performed and the measured result is displayed.

The indication of the measurement result differs according to the setting of the UV displayparameter.

• When the UV display parameter is set to “Simple”:

The UV transmittance (%) is displayed on thelower part of the screen.

• Perform UV measurement just after measurement for the distance portion.UV measurement is not performed after measurement for the near portion (intermediate portion)even if the read button is pressed.

• When the UV parameter is set to “With“, perform alignment correctly because powermeasurement is performed simultaneously with UV measurement.

UV transmittance indication (%)

Page 68: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

58

OPERATING PROCEDURE Measuring UV Transmittance (UV model only)

When the UV display parameter is set to “Compare”:

The screen changes to the UV measurementscreen and the measured result is displayed.

Buttons on UV measurement screen

Compare measurement of the UV transmittance

It is possible to measure the other lenses and compare the UV transmittance.

1) Change the lens and set its optical center on the nosepiece. Then, press the comparemeasurement button .

When the target becomes cross (within 2�), measurement can be performed more precisely.

UV 100% correction button

Used to correct the UV transmittance to 100%.This button is used when 100% is not displayed without the lens on thenosepiece.Pressing the button for about one second starts the correction. When the UV100% correction is correctly performed, a beep sounds.

Remeasurement button

Used to measure the UV transmittance again.The last result of the UV transmittance 1) is saved as lens measurementdata.

When the target becomes cross (within 2�), measurement can be performedmore precisely.

Compare measurement button

Used to compare with the UV transmittance of other lenses.

Exit button Used to return to each measurement screen.

• When the UV transmittance is unstable, it is recommended to perform UV measurementafter UV 100% correction.

Target indicationThe range of inside of the circle is within 2�.

Page 69: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

59

OPERATING PROCEDURE Measuring UV Transmittance (UV model only)

2

2) The measured results of the two lenses aredisplayed on the UV compare screen simulta-neously.

The left side 1) is original measured result andright side 2) is that of additional one.

3) Press the exit button to return to each measurement screen.

� Screen display at the time of auto correction

When the Auto correct parameter is set to “On” or “Silent”, the UV transmittance is automat-ically corrected to 100%. During the auto correction, the message “UV100% during mea-surement” is displayed.

When the message disappears, 100% correction is renewed. However, it is not renewed ifthe lens is put when the message is displayed.

• Pressing of the remeasurement button (or compare measurement button )renews the result of 1) (or2)).

• UV 100% correction button functions as on UV measurement screen.

• Compared lens data 2) is not displayed on the measurement screen.

The UV transmittance displayed in the lower part of the measurement screen is result of 1).

• Lens measurement value is not influenced even if the message “UV100% duringmeasurement” is displayed during lens measurement.

• 100% correction is performed at 5 to 60 minute intervals according to change amount ofthe camera sensitivity of the UV measurement.

Page 70: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

60

OPERATING PROCEDURE Marking

2.10 Marking

2.10.1 Marking at the optical centerMarking is performed to indicate the optical center position and axis direction.

1 Set a lens on the nosepiece.

2 Align the lens.Move the lens until the target changes from to (cross) and then from (cross) to (large cross).

(large cross) shows that the lens is within 0.8mm in diameter from the optical center wheremarking is allowed.

3 When the lens contains cylindrical power, set the cylinder axis according to the prescrip-tion.

While watching the axis value, rotate the lens until the prescribed axis value is displayed.

When the “Auto RD S” or “Auto RD R/L” parameteris set to “ ” or “ ”, the measured data will beautomatically fixed after the target is aligned.

Watch the axis value displayed in the upper-rightof the target.

When (large cross) changes back to (cross), align the target again.

For the axis of 90° or 180°, the shape of the tar-get changes as shown to the right.

• To measure the PD of mounted lenses, mark a lens at the optical center of both the right-eye and left-eye lenses. Measure the distance between the marked center points.

• For simple astigmatism lenses such as S 0.00 D and C -1.00 D or equivalent, the targetcannot be moved in the bus direction.

Mark a lens at the geometric center in the correct direction.

AXIS 90° AXIS 180°

Page 71: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

61

OPERATING PROCEDURE Marking

2

4 Mark the lens.Press down the marking lever to mark the lens.Three points in a line parallel to the lens tableare marked.

5 Remove the lens from the nosepiece.To remove the lens, lift the lens holder lever untilit clicks.

• When marking a cylinder lens in the bus direction, adjust the axis to 180°.

• Press down the marking lever slowly and lightly.Pressing down firmly may damage the tip of the ink cartridge.

• Do not touch the marked points.Faded ink makes the cylinder axis hard to see.

• It is recommended to make its mark on the uncut lens with a marking pen to identify theright or left side of the lens or nasal side after marking.

• After marking the repellent or super repellent lenses, pull up the marking lever slowly sothat the tip of the ink cartridge gradually becomes off from the lens.

To do so causes fewer ink spread and the ink markings become easy to see.

Page 72: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

62

OPERATING PROCEDURE Marking

2.10.2 Marking for prism prescription (Prism Layout)This is the procedure to mark lenses for heterophoria prescription.The prism alignment function allows the operator to easily determine the marking point on the lens.

Prism layout function

Prior entry of the prism prescription causes the on-screen target to move opposite to thedirection by the amount of the prism data.

Align the lens so that the target is aligned to the center of the alignment circle and then markthe lens.

� Only when marking lenses

1 Press the read button and measurement selectionbutton simultaneously on the normal measurementscreen, PPL measurement screen, or auto measure-ment screen. The screen changes to the prism entryscreen.

For models without the UV function, pressing theread button for about one second on the normalmeasurement screen, PPL measurementscreen, or auto measurement screen alsochanges the screen to the prism entry screen.

2 As necessary, designate the left-eye or right-eye lens.Press the L selection button or R selection button for the lens to be marked.

• The prism entry screen can be switchedfrom the parameter screen

a. Change the screen to the parameterscreen.

b. Align the cursor ( ) to “Prism lay-out”.

c. Press the selection button toswitch the screen to the prism entryscreen.

Prism entry screen

• It is possible to enter prism prescriptions for the both left-eye and right-eye lenses in a rowbefore marking them.

Page 73: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

63

OPERATING PROCEDURE Marking

2

3 Set the prism representation method with the coordinate system button in accor-dance with the prism representation method of the prism data on the prescription.

Each the coordinate system button is pressed, the coordinate system toggles between therectangular coordinate and polar coordinate systems.

4 Select the type of prism data to be entered withthe prism data button and enter or modify theprim prescription with the UP or DOWN button.

The indication of the UP and DOWN buttonchanges according to the selected coordinatesystem. As soon as the coordinate system isselected, the indication of the UP and DOWNbutton changes accordingly.

Each the UP or DOWN button is pressed, thepreset prism data is changed in the specifiedincrements. When the button is pressed, theprism data changes continuously.

The target moves opposite to the direction by the amount of entered prism data.

In the case of the rectangular coordinate representation (XY):

In the case of the polar coordinate representation (� ):

5 As necessary, enter the prism prescription of the other lens.Press the L selection button or R selection button to specify the left-eye orright-eye lens.Enter the prism prescription in the same manner.

(XY) Used to select the rectangular coordinate system.

��� Used to select the polar coordinate system.

Prism data button

UP/DOWN button

(DU) Used to enter a prism power in the Base Down/Up direction with and .

(IO) Used to enter a prism power in the Base In/Out direction with and .

(�� Used to enter an absolute prism value (�: PRISM) with and .

Used to enter the prism base direction (°: BASE) with and .

• Up to 20�� of prism data expressed in polar coordinates can be entered. When therectangular coordinate system is selected, the entry of data of even less than 20� may berejected, or the prism data of the other prism base side* may be modified to confine theprism data to 20� expressed in polar coordinates. (* For example, when the prism data ismodified in the Base IN/OUT direction, the prism data in the Base UP/DOWN direction maybe modified.)

Page 74: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

64

OPERATING PROCEDURE Marking

6 Press the L selection button or R selection button to specify the left-eye orright-eye lens.

7 Set the lens on the nosepiece with the upper side of the lens towards the operator.

8 Align the lens.Move the lens to align the target to the center ofthe alignment circle until the target changes from

to (cross) and then changes from (cross) to (large cross).

(large cross) represents that the lens iswithin an area of 0.8 mm in diameter from theoptical center of the lens and the lens can bemarked properly.

9 If the lens contains any cylindrical power, set thecylinder axis according to the prescription.

While watching the axis value (A), turn the lensuntil the axis reading shows the cylinder axis onprescription.

The current axis is also displayed at the upperright of the target.

10 Mark the lens.Press down the marking lever to mark the lens.

11 As necessary, mark the other lens.

Change the selected lens with the L selection button or R selection button .Mark the other lens in the same manner.

CAUTION • When marking lenses for prism prescriptions, be sure to set the lens with the upperside of the lens towards the operator.

If the lens is set with its bottom side towards the operator, the marked lens does not fol-low the prescription.With the LM-600/LM-600P/LM-600PD, the zero-degree direction of the prism base ison the operator’s left as viewed from the front. Be aware that the orientation of the lensset on the LM-600/LM-600P/LM-600PD is opposite in orientation of the lens set on amanual lensmeter or blocker.

• When the read button is pressed at that time, the measured data is fixed.The saved measured data can be printed from the measurement mode screen.

• To mark both lenses in a row, enter prism prescriptions for both lenses before making.

Page 75: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

65

OPERATING PROCEDURE Marking

2

12 Hold the next screen button to return to the measurement screen.Extended pressing of the button changes the icon display to the exit button . Releas-ing the button returns to the measurement screen.

� When measuring and marking lenses in a row

1 Enter prism prescriptions according to Steps 1 to 7 of “� Only when marking lenses”(page 62).

2 Press the next screen button to switchthe screen to the measurement mode screen.

3 As necessary, designate the left-eye lens or right-eye lens.Press the L selection button or R selection button for the lens to be marked.

4 Set and align the lens.Set and align the lens in the same manner asSteps 9 to 11 of “� Only when marking lenses”(page 62).

5 Press the read button.The measured data is fixed.

6 Mark the lens.Press down the marking lever to mark the lens.

Measurement mode screen

• Skip this step only when marking lenses on the measurement mode screen.

Page 76: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

66

OPERATING PROCEDURE Marking

7 As necessary, measure and mark the other lens.Change the selected lens with the L selection button or R selection button .

Measure and mark the lens in the same manner.

8 Press the print button to print out the mea-sured result (, transmit the data, or write the fixeddata to Eye Care card.).

The prism prescription is printed out togetherwith the other measured data.

Printing the data clears the measured data butmaintains preset prism prescriptions.

9 Press and hold the previous screen button to return to the measurement screen.Extended pressing of the button changes the icon display to the exit button . Releas-ing the button returns to the measurement screen.

• To mark lenses in a row, press the previous screen button to return to the prism entryscreen once, and then enter prism prescriptions for the other lens.

If measured data has been saved, press the clear button to clear the data in advance.

Prism prescriptions

Page 77: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

67

OPERATING PROCEDURE Marking

2

2.10.3 Marking for prism prescription (Measurement screen)This is the procedure to mark lenses for heterophoria prescription using the measurement screen.

� For rectangular coordinates

1 Set the Prism parameter on the parameterscreen to “BU/D BI/O”.

For details of the setting procedure, see “2.14Setting Parameters” (page 76).

2 Press the L-selection button or R-selection button to specify the left-eye orright-eye lens.

3 Set the lens on the nosepiece with the upper side of the lens towards the operator.

4 While watching the prism values on the screen,move the lens until the values show the prescrip-tion.

Prism indication

5 If the lens contains any cylindrical power, while watching the axis value on the screen,rotate the lens until the value shows the prescription.

6 Mark a lens.Press down the marking lever to mark the lens.

• Specify the left-eye lens or right-eye lens for rectangular coordinates.

CAUTION• When marking for prism prescription, be sure to set the upper side of the lenstowards the operator.

If the lens is set with the wrong side towards the operator, the marked lens does not fol-low prescription. For the LM-600/LM-600P/LM-600PD, the zero degree direction is onthe operator’s left. Be aware that the lens should be set with the upper side of the lenstowards the operator, which is opposite in orientation of the lens with a manual lensme-ter or a blocker.

Prism values

I Base In

O Base Out

U Base Up

D Base Down

Page 78: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

68

OPERATING PROCEDURE Marking

� For polar coordinates

1 Set the Prism parameter on the parameterscreen to “P-B”.

For details of the setting procedure, see “2.14Setting Parameters” (page 76).

2 Press the L-selection button or R-selection button to specify the left-eye orright-eye lens.

3 Set the lens on the nosepiece with the upper side of the lens towards the operator.

4 While watching the prism values on the screen,move the lens until the values show the prescrip-tion.

The prism power is displayed with the absolutevalue (�: PRISM) and base direction (°: BASE).

5 If the lens contains any cylindrical power, while watching the axis value on the screen,rotate the lens until the value shows the prescription.

6 Mark a lens.Press down the marking lever to mark the lens.

CAUTION• When marking for prism prescription, be sure to set the upper side of the lenstowards the operator.

If the lens is set with the wrong side towards the operator, the marked lens does not fol-low prescription. For the LM-600/LM-600P/LM-600PD, the zero degree direction is onthe operator’s left. Be aware that the lens should be set with the upper side of the lenstowards the operator, which is opposite in orientation of the lens with a manual lensme-ter or a blocker.

Prism values

Page 79: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

69

OPERATING PROCEDURE Printing (LM-600P/LM-600PD only)

2

2.11 Printing (LM-600P/LM-600PD only)

Press the print button to print out the measured data.

[Sample printout]

• Print the data out after making sure that the measured data is fixed.When the print button is pressed before the measured data is fixed, the measured data isfixed and printed out.

• When an external device is connected, printing and data transmission are performed.When the Printer parameter is set to “Off”, only data transmission is performed by pressing theprint button .

• The device uses the thermal printer paper. When saving the printings, make a copy of it.The thermal paper may be difficult to read due to the aged deterioration.

Sample printout in the singlestate

Sample printout in the R/Lstate (LM-600P)

Sample printout in CL measurement

Measurement modeNot printed in normal measurement

When printing is performed after onlyone side is measured in the R/L state,“ ” is printed on the numeric field.

Depending on the setting of the Eco print parame-ter, printing is performed with line space reduced.

Sample printout in the R/Lstate (LM-600PD)

Page 80: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

70

OPERATING PROCEDURE Printing (LM-600P/LM-600PD only)

2.11.1 Connecting to the AR, ARK or RKT and setting parameters It is possible to print measured data using the printer of the connected AR, ARK or RKT.

Connectable devices:AR-600/AR-330A series, ARK-700/ARK-530A seriesRKT-7700, TONOREF II

� Connecting procedure

1 Connect the AR, ARK or RKT to the interface connector of the LM-600/LM-600P/LM-600PD with the interface cable (option: OPIF-6).

CAUTION• Be sure to turn off each device before connecting the interface cable, etc.Connecting the cable with power on may cause malfunction.

Interface connector

To AR, ARK or RKT

Interface cable (optional)

Page 81: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

71

OPERATING PROCEDURE Printing (LM-600P/LM-600PD only)

2

� Setting procedure

Set the following parameters on the LM-600/LM-600P/LM-600PD and AR, ARK or RKT.

1 Set the following parameters related to communications on the LM-600/LM-600P/LM-600PD:

For details of the parameter setting procedure, see “2.14 Setting Parameters” (page 76).

Set Parameters No. 31 to 37 to the factory settings. Setting the Printer parameter to “AR print”automatically sets those parameters to their factory settings.

2 Set the following parameters on the AR, ARK or RKT.Parameter settings on the AR or ARK series

Parameter settings on the RKT-7700 or TONOREF II

For details of the parameter setting procedure, refer to the corresponding Operator’s Manualto the connected AR, ARK or RKT.

No. Parameter Settings

16 Printer AR print

31 Com Mode NIDEK

32 Baud Rate 9600 (bps)

33 Parity Odd

34 Data bits 8bit

37 Stop bits 1bit

Parameter Settings

IN port (LM) NIDEK

LM Data Prt. YES

Baud-Rate 9600

Bit Length 8

Parameter Settings

BAUD-RATE 9600

BIT LENGTH 8

LM DATA PRINT YES

• The printout produced by the printer of the AR, ARK or RKT is different in font etc. from theprintout produced by the printer of the LM-600P/LM-600PD.

• “Eco print” is not supported for communication print.Printing is performed with normal line space.

Page 82: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

72

OPERATING PROCEDURE Data save to the Eye Care card

2.12 Data save to the Eye Care card

The measured data is saved to the Eye Care card and LM data is transferred to a device such as theRT-5100.SPH, CYL, AXIS, ADD, and PRISM data are each saved.

The 2nd ADD power is saved for the data with it.The prism value is saved in the form set with the Prism parameter.

2.12.1 Writing of fixed data to the Eye Care card

� When the Eye Care card is inserted before measurement

1 Insert the Eye Care card when the LM-600/LM-600P/LM-600PD has no measured data in theinternal memory.

The Eye Care card icon is changed from white toyellow.

2 Start measurement.

3 Press the print button .

Regardless of print button icon, the data is written to the Eye Care card. The Eye Care card icon becomes green, and the data is written to the Eye Care card.

After the data is written successfully, the icon changes to yellow.

4 When the Eye Care card icon changes to yellow,remove the Eye Care card.

Remove the card straight.

• When the lens side is not specified (single state), the data is not written to the Eye Carecard.

Even if data is attempted to be written, the data is not written.

Eye Care card

Eye Care card slot

Page 83: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

73

OPERATING PROCEDURE Data save to the Eye Care card

2

� When the Eye Care card is inserted after measurement

When the measured data is saved in the memory and the Eye Care card is inserted, the data is auto-matically written.

1 After measurement, insert the Eye Care card tothe Eye Care card slot.

The Eye Care card icon is displayed in green,and the data is written to the Eye Care card.

After the data is written successfully, the settingindication changes to yellow.

2 When the Eye Care card icon changes to yellow,remove the Eye Care card.

Remove the card straight.

• Never remove an Eye Care card while it is being accessed.While the card is being accessed, the icon is displayed in green.

When trying to remove the card while it is being accessed, data is not written successfully and theEye Care card may be irreparably damaged.

• The data is written for Auto Prt. S and Auto Prt. R/L settings.

Eye Care card

Eye Care card slot

• Never remove an Eye Care card while it is being accessed.While the card is being accessed, the icon is displayed in green.

The red icon indicates an error occurrence. In such a case, the operator is only allowed to removethe card.

When trying to remove the card while it is being accessed, the data is not written successfully andthe Eye Care card may be irreparably damaged

Page 84: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

74

OPERATING PROCEDURE Data save to the Eye Care card

Indications of the Eye Care card icon

2.12.2 Erasing Data on the Eye Care Card

This is the procedure for erasing all the data on the Eye Care card.

1 Insert the Eye Care card to the Eye Care card slot.The Eye Care card icon is changed from white to yellow.

2 Press and hold the clear button.After the button is pressed and held, a beep sounds and the pilot lamp blinks.Releasing the button erases all the data in the Eye Care card.While the data is being erased, the Eye Care card icon is displayed in green.

3 When the Eye Care card icon is changed to yellow, remove the Eye Care card.Remove the card straight.

(White): The Eye Care card is not inserted.

(Yellow): The Eye Care card is inserted.

(Green):The Eye Care card is being accessed.

Never touch the card.

(Red):An error occurs while the Eye Care card is being accessed.

The card can be removed.

• Only one LM data (1 frame) can be saved to the Eye Care card.

Page 85: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

75

OPERATING PROCEDURE After Use

2

2.13 After Use

1 Turn the device off.When the measurement screen is displayed, turn the power switch off ( ).

2 Put the dust cover over the device.

• Turning off the power when the parameter screen is displayed may not save the setparameter settings.

Parameter settings are saved when the exit button , back page button , or forwardpage button is pressed on the parameter screen.

CAUTION• If the device will not be used for a long period of time, disconnect the power cord fromthe grounded power outlet.

If dust on the device attracts moisture, it may cause short circuit or fire.

• When the device is not in use, turn it off and put the dust cover over the device.Dust may affect the measurement accuracy.

Page 86: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

76

OPERATING PROCEDURE Setting Parameters

2.14 Setting Parameters

The LM-600/LM-600P/LM-600PD contains the function to change each parameter setting of thedevice according to your needs and preference.There are parameters related to the indication, measurement, printing, and communication.The procedure for changing and checking each parameter setting is described as follows.

1 Display the parameter screen.Press and hold the measurement selection but-ton until the icon changes to and releasethe button.

2 Display the page on which the parameter settingneeds to be changed.

Select the page on which the parameter settingneeds to be changed with the forward page but-ton or back page button .

There are six pages (seven pages for the UVmodel).

Each the read button is pressed, the iconschange. Select the icon combination accordingto the operation.

3 Move the cursor to the parameter to be changed.With the up button or down button

, move the cursor ( ) to the param-eter to be changed.

Pressing the read buttonchanges the icons.

Page 87: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

77

OPERATING PROCEDURE Setting Parameters

2

4 Change the parameter setting with the selectionbutton / .

The contents within the frame are those currentlyselected for setting.

For details of the parameter settings, see “2.14.1Parameter tables” (page 78).

In the parameters necessary for entering numerical values, the selection buttons are dis-played as follows.

5 Change the other parameter settings to be changed in the manner as Steps 2 to 4.

6 Press the exit button to return to the measurement screen.The changed parameter settings are saved in memory.

Increase a numerical value.

Decrease a numerical value.

Page 88: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

78

OPERATING PROCEDURE Setting Parameters

2.14.1 Parameter tables<Page 1> * Underlined options indicate factory settings.

1 : StepThis is for selecting the indication step of measured data (D).The indication step of axis and prism � is 1° at all times.

2 : CylinderThis is for selecting the CYL mode from +, +/- (MIX), or -.

3 : PrismThis is for selecting the indication form of the measured prism value. When “Off” is selected, the prismvalue is not displayed.

4 : Abbe selectThis is for compensating an error in measured data produced due to measurement of high-power lensesusing the abbe number. Select the abbe number for compensation from A:, B:, C:, or Com.

Set A: (factory setting: 41)Set B: (factory setting: 58)Set C: (factory setting: 32)

Abbe numbers can be entered in each of Set A, Set B, and Set C in the range of 20 to 60 according tothe lens material.

“Com” can be selected only when the abbe number is transmitted from a connected PC. When “Com”is selected, the abbe number through communication is set and displayed on the setting indication ofthe measurement screen. (In this case, it is displayed in black with yellow background.)

Move the cursor to A, B, or C and press the selection button to increase the value.

A is automatically selected at power-on. Set the abbe number that is usually used at A.

5 : WavelengthThis is for selecting the measured data from the one to be obtained with the e-line wavelength (546.07nm) or that with the d-line wavelength (587.56 nm).

No. Parameter Settings Category

1 Step 0.25, 0.12, 0.06, 0.01 (D)Measurementindication

2 Cylinder +, +/-, -

3 Prism Off, P-B, BU/D BI/O

4 Abbe select A:, B:, C:, Com Standard of measurement5 Wavelength e, d

6 Prism layout – Blocking function

CYL+ Displays the cylindrical power by positive reading.

CYL±(MIX)

Displays the cylindrical power by positive reading when the refractive power ispositive for any axis angle.Indicates the cylindrical power by negative reading in other cases.

CYL- Displays the cylindrical power by negative reading.

Off

The measured prism value is not displayed.Measured values are displayed larger than when prism measured values aredisplayed.See “2.5 Measuring Bifocal Lenses” (page 38).

P-B The measured prism value is displayed by polar coordinates.

BU/D BI/O The measured prism value is displayed by rectangular coordinates.

Page 89: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

79

OPERATING PROCEDURE Setting Parameters

2

6 : Prism layoutThis parameter is for switching the screen to the prism layout screen.

<Page 2> * Underlined options indicate factory settings.

7 : Auto R/LThis is for changing the R/L indication automatically or manually when the measuring lens side ischanged.

On the CL measurement screen, auto R/L is disabled.

8 : Auto RD SThis is for enabling or disabling the auto read function in the single state.

* The target inside the parentheses is the setting when the Target parameter is set to .

On the CL measurement screen, auto read S is disabled.

9 : Auto RD R/LThis is for enabling or disabling the auto read function in the R/L state.

No. Parameter Settings Category

7 Auto R/L Off, On

Auto function

8 Auto RD S Off, ,

9 Auto RD R/L Off, ,

10 Dist. auto RD Off, On

11 Near auto RD Off, On

12 Indicator Off, On

On-screenindication

13 Single Off, On

14 Near Near Sph, Add

15 Contact Off, On, Only

Off The lens side is specified by pressing the button.

On

After the measured data is fixed, the R/L indication changes automatically byremoving the lens without pressing the button.

However, when both R and L data are fixed, the sign does not change by removingthe lens.

Off The measured data is fixed by pressing the read button.

( )

When the lens is aligned until the target changes to (cross) (or ), themeasured data is automatically fixed without pressing the read button.

After auto read, if alignment is performed again until the target changes to “ ” (or), auto read is performed again.

( )When the lens is aligned until the target changes to (large cross) (or

), the measured data is automatically fixed without pressing the readbutton.

Off The measured data is fixed by pressing the read button.

Page 90: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

80

OPERATING PROCEDURE Setting Parameters

* The target inside the parentheses is the setting when the Target parameter is set to “ ”.

On the CL measurement screen, auto read R/L is disabled.

10 : Dist. auto RDThis is for enabling or disabling the auto read function for the distance portion in measurement of pro-gressive lenses.

When the lens is aligned until the target changes to (cross), the measured data is fixed withoutpressing the read button.

11 : Near auto RDThis is for enabling or disabling the auto read function for the near portion in measurement of progres-sive lenses.

When the lens is aligned until the target changes to (cross), the measured data is fixed withoutpressing the read button.

12 : IndicatorThis is for selecting whether to display the indicator for near portion (�) for alignment of the near por-tion in PPL measurement.When this parameter is set to “On”, the indicator for near portion (�) appears above, below, on theright of, or on the left of the target, which shows the direction and movement distance to move the lens.

13 : SingleThis is for selecting whether to display the single state.

When “On” is selected, the single state is set by pressing the clear button . When “Off” isselected, R is set.

14 : NearThis is for selecting the indication form of the measured data for the near portion.

15 : ContactThis is for enabling or disabling the CL measurement mode.

( )

When the lens is aligned until the target changes to (cross) (or ), themeasured data is automatically fixed without pressing the read button.

After auto read, when alignment is performed again until the target changes to (or ), auto read is performed again.When PD measurement is performed with the LM-600PD, the operation is same aswhen the setting is (large cross) even though the parameter is set to (cross).

( )When the lens is aligned until the target changes to (large cross) (or

), the measured data is automatically fixed without pressing the readbutton.

Near Sph The near portion power is displayed by sphere value (SPH).

Add The near portion power is displayed by ADD value (ADD).

Off

When CL measurement is not performed:The screen changes by pressing the measurement selection button in the order ofAuto measurement screen ��Normal measurement screen � PPL measurementscreen ��Auto measurement screen...

On

When CL measurement is performed:The screen changes by pressing the measurement selection button in the order ofAuto measurement screen ��Normal measurement screen � PPL measurementscreen ��CL measurement screen ��Auto measurement screen...

OnlyWhen CL measurement only is performed.

The measurement selection button is disabled. Only the CL measurement screen isavailable.

Page 91: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

81

OPERATING PROCEDURE Setting Parameters

2

<Page 3> * Underlined options indicate factory settings.

16 : PrinterThis is for setting the condition when the print button is pressed.

17 : Print densityThis is for setting the line space of printing.

18 : Paper cutThis is for selecting the cut method of the printer paper.

19 : Auto cutterThis is for selecting whether to cut the paper automatically after printing.

20 : Eco printThis is for setting the line space of printing.

When the parameter is set to “On”, the printer paper can be saved by reducing the line space.

For printout samples, see “2.11 Printing (LM-600P/LM-600PD only)” (page 69).

No. Parameter Settings Category

16 Printer Off, On, AR print

Printing

17 Print density 60, 80, 100, 120, 140 (%)

18 Paper cut Partial, Full

19 Auto cutter Off, On

20 Eco print Off, On

21 Auto prt. S Off, On

22 Auto prt. R/L Off, On

23 Print & Clear Off, On

24 Comments Sets the comments.

OffThe print button is displayed by the icon.Printing is not performed by pressing the print button.

Only data transmission is performed.

On Printing is performed by pressing the print button .Data transmission is also performed.

AR print

The print button is displayed by the icon.Printing by the printer of the connected AR, ARK or RKT is performed bypressing the print button.

Data transmission is also performed.

• “On” is disabled with the LM-600.

• When the Printer parameter is set to “AR print”, the parameters “Print density”, “Paper cut”,“Auto cutter”, and “Eco print” are disabled.

Partial Printer paper is cut with a center part left.

Full Printer paper is cut off.

Off Printed with normal line space.

On Printed with line space reduced to 1/4 of the normal space.

Page 92: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

82

OPERATING PROCEDURE Setting Parameters

21 : Auto prt. SThis is for selecting whether to print out the measured data by removing the lens after the data is fixedin the single state (R/L not specified).On the CL measurement screen, this function is disabled.

22 : Auto prt. R/LThis is for selecting whether to print out the measured data by removing the lens after the data is fixedin the R/L state.

On the CL measurement screen, this function is disabled.

23 : Print & ClearThis is for selecting whether to clear the measured data from memory after printing, data communica-tion, or data writing to the Eye Care card.

24 : CommentsThis is for setting the contents of comments in communication print.Selecting the Comments parameter and then pressing the selection button switches the screen to thecomments setting screen.For details, see “2.14.2 Entering comments” (page 88).

OffMeasured data is not cleared after printing, data communication, or datawriting to the Eye Care card.

Previous measured data is overwritten when next measurement is performed.

OnMeasured data is cleared after printing, data communication, or data writing tothe Eye Care card (except for during occurrence of error).

Page 93: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

83

OPERATING PROCEDURE Setting Parameters

2

<Page 4> * Underlined options indicate factory settings.

25 : Intl. screenThis is for selecting the measurement screen displayed at power-on.

26 : TargetThis is for selecting the indication format of the target displayed on the measurement screen.

For display samples, see “� Details of target” (page 28).

27 : GuideThis is for setting whether or not to display the guide mark ( ) in the center of the alignment circlefor precise alignment of the target.

See “� Guide mark” (page 37) for details of the guide mark.

28 : BeepThis is for selecting the musical quality of beeps, which sound for each press of the button and whenthe data is auto read.

29 : Auto offThis is for enabling or disabling the display auto off function or setting the display auto off time.

During lens measurement, the device will not be placed into display auto off mode.

30 : ColorThis is for screen color setting.

31 : IconThis is for enabling or disabling the switch icon function.

When “Fixed” is selected, the icons of the measurement selection button and the R/L selection buttonare displayed without toggle as follows:

Measurement selection button: , R/L selection button:

No. Parameter Settings Category

25 Intl. screen Auto, Normal, Progressive, ContactOn-screenindication

26 Target , 27 Guide Off, On

28 Beep Off, Low, Mid, HiOther function

29 Auto off Off, 1, 3, 5, 10, 15, 30, 60 (min)

30 Color Blue, Red, Green On-screenindication31 Icon Fixed, Changeable

Auto Normal measurement screen in auto mode

Normal Normal measurement screen

Progressive PPL measurement screen

Contact CL measurement screen

Normal targetThe target indicates the distance and direction from the center of thenosepiece.The direction and movement distance are constant regardless of the refractivepower of lenses.

Prism targetThe target moves based on prisms as with the eyepiece- type or projection-type lensmeter.The target rotates itself, indicating the axis.

Page 94: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

84

OPERATING PROCEDURE Setting Parameters

<Page 5> * Underlined options indicate factory settings.

32 : Com modeThis is for selecting the device to communicate with.

When the Printer parameter is set to “AR print”, “NIDEK” is automatically selected and it cannot

be changed.

33 : Baud rateThis is for selecting the baud-rate (bit transmission speed) for communication.When the Com mode parameter is set to “NIDEK”, “9600” is automatically selected and it cannot bechanged.

34 : ParityThis is for setting whether to perform parity check. Select either odd parity or even parity.

When the Com mode parameter is set to “NIDEK”, “Odd” is automatically selected and it cannot bechanged.

35 : Data bitsThis is for selecting the bit number for a single character used for communication.

When the Com mode parameter is set to “NIDEK”, “8bit” is automatically selected and it cannot bechanged.

36 : Stop bitsThis is for selecting the bit number of stop bit for communication.When the Com mode parameter is set to “NIDEK”, “1bit” is automatically selected and it cannot bechanged.

37 : CR codeThis is for selecting whether to attach CR (carriage return) code at the end of data to be transmitted.

38 : Prism TxThis is for selecting whether to transmit the measured prism value.

No. Parameter Settings Category

32 Com mode Off, NIDEK, PC, NCP10, NCP20

Communicationfunction

33 Baud rate 1200, 2400, 4800, 9600, 19200 (bps)

34 Parity Off, Odd, Even

35 Data bits 7bit, 8bit

36 Stop bits 1bit, 2bit

37 CR code Off, On

38 Prism Tx Off, On, Display

Off When no data is to be transmitted or printed

NIDEK Communication with a NIDEK-brand device

PC Computer

NCP10 NCP10-compliant NIDEK-brand device

NCP20 NCP20-compliant NIDEK-brand device

Off The measured prism value is not transmitted at any time.

On The measured prism value is transmitted at all times.

DisplayWhen the measured prism value is displayed on the screen, it is transmitted. Ifnot, it is not be transmitted.

Page 95: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

85

OPERATING PROCEDURE Setting Parameters

2

PD parameter (LM-600PD only)

The PD parameter is displayed on page 5 of theLM-600PD parameter screen.

Sets whether or not to display the PD.

When R/L change with the nose slider is performed (set to “On” or “R/L Only”), the R/L selection buttonis disabled with (blank).

The nose slider functions when the lens table is positioned within 50 mm from the center of the nose-piece. When the lens table is beyond 50 mm, the nose slider automatically turns off.

• In order to connect the LM-600/LM-600P/LM-600PD with external computers, ask yourauthorized distributor for the interface manual which explains the details of thecommunication protocols.

Off The PD is not displayed. R/L change with the nose slider is not performed.

On The PD is displayed. R/L change with the nose slider is performed.

R/L Only The PD is not displayed. R/L change with the nose slider is performed.

Page 96: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

86

OPERATING PROCEDURE Setting Parameters

<Page 6> * Underlined options indicate factory settings.

Parameters No. 39 to 45 are the items when the Com mode parameter is set to “NCP20”.

39 : SourceThis is for setting the port number of the source (LM-600/LM-600P/LM-600PD) from 0 to 255.

40 : DestinationThis is for setting the port number of the destination (receiving device) from 0 to 255.

41 : RequestThis is for setting the response waiting time (time-out) for command requests from 0.1 to 10.0 seconds.

42 : ReceptionThis is for setting the waiting time (time-out) for receiving commands from 1 to 100 seconds.

43 : ResponseThis is for setting the waiting time (time-out) for responding to commands from 0.1 to 10.0 seconds.

44 : Re-requestThis is for setting the response waiting time (time-out) for another command requests from 0.1 to 10.0seconds.

45 : RetryThis is for setting the number of retry times from 0 to 5 when communication is not established.

No. Parameter Settings Category

39 Source 0 to 255, factory setting: 1

Communicationfunction (NCP20)

40 Destination 0 to 255, factory setting: 0

41 Request 0.1 to 10.0 (seconds), factory setting: 2

42 Reception 1 to 100 (seconds), factory setting: 10

43 Response 0.1 to 10.0 (seconds), factory setting: 2

44 Re-request 0.1 to 10.0 (seconds), factory setting: 2

45 Retry 0 to 5 (times), factory setting: 3

Page 97: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

87

OPERATING PROCEDURE Setting Parameters

2

<Page 7> (UV model only)

46 : UVThis is for selecting whether to perform UV measurement

47 : UV stepThis is for setting the indication step of the UV transmittance (1%/5%).

48 : UV displayThis is for setting the screen in UV measurement.

49 : Auto correctThis is for setting the auto correction of the UV transmittance to 100%.

No. Parameter Settings Category

46 UV Off, On, With

UV measurementfunction

47 UV step 1, 5 (%)

48 UV display Compare, Simple

49 Auto correct Off, On, Silent

Off UV measurement is not performed by extended pressing of the read button.

On UV measurement is performed by extended pressing of the read button.

With

Reading of the lens power and UV measurement are performed by extended pressing of the read button.When ADD powers are measured, only reading of the lens power is performed.

• Even though the indication step is set to 1%, the measurement accuracy is within ±5%.

Compare

Displays the measured results by switching the screen to the UV measurement screen.

The UV transmittance can be compared with that of other lenses by comparemeasurement in the UV compare screen.

Simple Displays the measured results on the measurement screen.

Off The UV transmittance is not automatically corrected to 100%.

OnThe UV transmittance is automatically corrected to 100%. When correction is renewed, a beep sounds.

SilentThe UV transmittance is automatically corrected to 100%. When correction is renewed, a beep does not sound.

• 100% correction is performed at 5 to 60 minute intervals according to change amount ofthe camera sensitivity of the UV measurement.

Page 98: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

88

OPERATING PROCEDURE Setting Parameters

2.14.2 Entering commentsThe comments in communication print can be changed. (Factory setting: NIDEK LM-600/NIDEK LM-600P/NIDEK LM-600PD).

1 Display the parameter screen.Press and hold the measurement selection but-ton until the icon changes to and releasethe button.

2 Press the forward page button to displayparameter page 3.

3 Press the down button to select “Com-ments”.

4 Press the selection button to switch thescreen to the comments settings screen.

The comments setting screen is displayed.

In the upper part of the screen, the currently-entered comments are displayed.

Page 99: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

89

OPERATING PROCEDURE Setting Parameters

2

5 Press the right switch or left switch to select the entry position.

The cursor indicates the position where a char-acter is entered or changed.

Up to 24 characters can be entered in two lines.

6 Change the contents with the selection button and/or .

Move the cursor to the desired character in thelist and change the display.

7 Repeat Steps 5 to 6 to enter the desired comments.

8 When comments change is completed, press the exit button to finish the settings.When the exit button is pressed, the comments are saved in the memory and the screenreturn to parameter page 3.

9 Press the exit button to return to the measurement screen.

• The line with “DEL ” at the top can not be printed out.Entering “DEL” in a line makes the character position blank.

• The display on the screen may differ with the printed character depending on theconnected device.

It occurs for the marks such as “#”, “$”, “@”, “[”, “]”. Check the actual printout.

Move the cursor in the entry position.

Move the cursor and change the contents.

Page 100: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

90

OPERATING PROCEDURE Setting Parameters

Page 101: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

91

3

3. OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED

The LM-600/LM-600P/LM-600PD transmits data to an external device such as the NIDEK motorizedrefractor, computer, and Eye Care card system.The optional foot switch can be used instead of the read button to read measured data.

3.1 Connecting to the NIDEK Motorized Refractor (RT) or Computer

3.1.1 OutlineAny data on printout can be transmitted to the NIDEK motorized refractor (hereafter referred to as theRT) or a computer.The measured data transmitted to the RT is used as former lensmetory data (LM data) in subjectiverefraction.

Connectable device: RT-2100 series, RT-5100 seriesData transmitted to the computer can be managed by various database software.

CAUTION• Be sure to turn off each device before connecting the interface cable, etc.Connecting the cable with power on may cause malfunction.

• LM data can be transmitted through RS-232C compliant communication.

• USB communication can be performed to transmit data to a computer.

Measurement of eyeglasses

Subjective refraction

LM data

LM data

Managed by database softwareData output to RT

NIDEK motorized refractor

Computer

Subjective test guided by the LM data

Page 102: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

92

OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED Connecting to the NIDEK Motorized Refractor (RT) or Computer

3.1.2 Connecting procedure

1 Connect the RT (or a computer) to the interface connector of the LM-600/LM-600P/LM-600PD with the interface cable (optional).

� Connecting a computer with USB

Connect the optional USB cable between the USB port of the LM-600/LM-600P/LM-600PD and theUSB port of a computer.

The USB driver included in the optional USB cable needs to be installed in the computer.

Interface connector

To RT or computer Interface cable (optional)

• The USB interface is compliant with USB 1.1.

• Do not use the commercially available USB cable since the EMC performance (interference produced) may deteriorate.

USB port

To computer

USB cable (optional)

Page 103: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

93

OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED Connecting to the NIDEK Motorized

3

When connecting the cable to the interface connector or USB port, remove the cap on the connector.Bending the removed cap separates into two units. Cover the not-used connector with the cap.

3.1.3 Operating procedure

1 After measurement, press the print button .Perform operation in the same manner as the normal printing.

See “2 OPERATING PROCEDURE” (page 25) for details of the measuring procedure.

See “2.11 Printing (LM-600P/LM-600PD only)” (page 69) for details of the printing proce-dure.

2 The LM-600/LM-600P/LM-600PD automatically transmits data to the RT (or computer).When the LM-600/LM-600P/LM-600PD is connected to the RT, it receives a Data No. (IDNo.).

When the LM-600/LM-600P/LM-600PD is connected to a computer, it does not receive aData No. (ID No.).

3 The measured data is printed.When the LM-600/LM-600P/LM-600PD is connected to the RT, the data No. (ID No.) is alsoprinted.

Connector cap

Bend it to separate.

• Disconnect the interface cable while pressing the button on the connector.

When the cable is connected, the button islocated on the side of the connector.

Button

Page 104: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

94

OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED Connecting the Foot Switch

3.2 Connecting the Foot Switch

The optional foot switch can be used instead of theread button to read measured data.In such cases when a number of lenses are mea-sured in the Lab system, the foot switch allows formeasurement while holding lenses both hands.

� Connecting procedure

1 Connect the cable of the foot switch to the interface connector.

• The read button works with the foot switch connected.

Foot switch

Cable of foot switch

Interface connector

Page 105: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

95

4

4. MAINTENANCE

4.1 Troubleshooting

In the event that the device does not work correctly, correct the problem according to the followingtable before contacting your authorized distributor.

If the symptom cannot be corrected with the above actions, contact NIDEK or your authorized distributor.

Symptom Action

The LCD does not turn on. • The power cord may not be correctly connected.

Reconnect it securely. • The power switch may not have been turn on. Check the power switch.

The LCD does not turn on (not clear) even though the power is on.

• It seems that the display auto off function has been executed. Press abutton to recall the ON state.

• The brightness may be darkly adjusted.Adjust the contrast knob.

The screen disappears suddenly. • It seems that the display auto off function has been executed. Press a

button to recall the ON state.

Data is not printed out.

• Check the printer paper. If the paper has been used up, set new printerpaper.

• The Printer parameter may be set to “Off”.Reset the parameter.

The printer runs, but the paper is blank.

• The printer paper may be set with the wrong side up.Set it with the correct side up.

Data cannot be read even though the Eye Care card is inserted.

• The contact in the Eye Care card slot may be dirty. Clean it (see page105).

• When an error occurs while data is read in the other device such as arefractor, clean its card reader.

Page 106: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

96

MAINTENANCE Error Messages and Countermeasures

4.2 Error Messages and Countermeasures

If a message appears on the lower part of the screen, take measures according to the table below.Should a service visit be necessary, please prepare the details of the message and symptom alongwith the serial number of the device.

Error message Cause and countermeasure

0D initialization error • The measuring beam was interrupted during initialization.

Check the nosepiece.Press the restart button to reboot the device.

Please remove lens from nosepiece. • A lens was on the nosepiece at power-on.

After removing the lens, press the restart button to reboot the device.

Dust detected. Please clean lens.

• There is something which interrupts the measuring beam on the nosepiece atpower-on.

Check the nosepiece. Remove the dust and dirt from the protective glass.Press the restart button to reboot the device.

Use contact nosepiece?

• The nosepiece for contact lenses is placed at power-on.Replace with the standard nosepiece, press the restart button to reboot thedevice.To measure a contact lens, press the execution button.

Measurement error • There is something which interrupts the measuring beam on the nosepiece.

Check the nosepiece. Remove the one which interrupts the measuring beam.

SPH overflow error • The sphere value exceeds ±27 D.

Check the power of the lens.

CYL overflow error • The cylinder value exceeds ±11 D.

Check the power of the lens.

ADD overflow error • The ADD value exceeds 11 D.

Check the power of the lens.

Center overflow error • The lens with the prism value exceeding 21� is on the nosepiece.

Perform alignment of the lens or remove the lens.

Printer error. • The print button was pressed with no printer paper set.

Fill the printer paper.

Printer paper error • Printer paper runs out during printing

Fill the printer paper. After the paper is filled, the printing continues.

[Message display sample]

Execute buttonRestart button

Page 107: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

97

MAINTENANCE Error Messages and Countermeasures

4

� Screen displayed at the time of overflow error

The measured value at the time of overflow error is displayed in red and the red indication remainseven after having been read. Also, auto read is disabled at the time of error.

If printing is performed, erroneous measured values are denoted with an asterisk.

Communication error

• Correct communication is not performed with other devices.Check the interface cable.Check that the connected device is turned on.Check that the parameters related to communication are correctly set.

AR print error

• A request for data transmission was not sent from the AR, ARK, or RKT duringprinting by the connected AR, ARK, or RKT.

Check the interface cable.Check the parameter settings on the AR, ARK, or RKT.

Eye Care card error • Correct communication is not performed with the Eye Care card.

Replace the Eye Care card and check if the communication is possible.

CCD error • CCD signals are not correct.

Failure in the interior of the device. Contact NIDEK or your authorized distributor.

UV initialization error • UV-LED is not lit up.

Failure in the interior of the device. Contact NIDEK or your authorized distributor.

UV measurement error • Significant change such as interference light occurred during measurement.

Make the lens stable and perform measurement again.

Nose slider is activated

• This message appears when following operations are performed in thecondition that the R/L switch button is displayed blank.

1. The R/L selection button (blank) is pressed.2. R or L can be switched automatically after the lens is measured and

removed when the Auto R/L parameter is set to “On”.

Error message Cause and countermeasure

The measured value is displayed in red toindicate that the measured value is errone-ous.

The asterisks are printed to indicate that themeasured values are erroneous.

Page 108: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

98

MAINTENANCE Replacing Printer Paper (LM-600P/LM-600PD only)

4.3 Replacing Printer Paper (LM-600P/LM-600PD only)

When a red line appears on the side of the printer paper, it means that the paper is running short.In such a case, stop using the printer and replace the roll with a new one.

1 Pull the printer cover towards the operator.Slide the printer cover lever at the rear of thedevice to open the cover.

2 Remove any remaining printer paper.

• Do not run the printer while the printer paper is not set.It may ruin the printer head.

• Do not pull the printer paper forcefully.Doing so could cause malfunction of the printer.

CAUTION• When replacing the printer paper, be sure not to touch the printer head on the upperpart inside the printer paper holder.

The printer head can get very hot and cause burns.

Page 109: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

99

MAINTENANCE Replacing Printer Paper (LM-600P/LM-600PD only)

4

3 Set a new printer paper..Set the paper as shown in the figure to the right.

Leave the end of the paper out from the cover.

4 Close the printer cover.Press the right and left sides of the printer coverto close the printer.

• When the roll is set upside down, the data is not printed correctly.

• Make sure that the printer paper is set straight and the core of the roll paper is not misaligned.

The printer paper may not feed properly.

• Make sure that the cover is closed securely.If the cover is closed incompletely, the auto-cutter may not operate properly.

Page 110: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

100

MAINTENANCE Ink Refilling

4.4 Ink Refilling

4.4.1 Ink cartridgeWhen markings become faint, replace the ink cartridge. Use the specified cartridge for refills.Replace the optional red or blue ink cartridge in the same manner.

1 Press the tip of the ink cartridge and remove theC-ring with tweezers.

During work, take enough care not to lose the C-ring.

It is recommended to remove the C-ring withtweezers so as not to lose.

When the C-ring is removed, the spring and car-tridge fall off. Take care not to lose them.

2 Insert the spring and the new cartridge into themarker holder, and set the C-ring in the grooveon the end of the ink cartridge.

Match the shape of the cartridge to the hole ofthe marker holder by rotating until the cartridgecan be inserted smoothly.

Use the C-ring which is removed in the proce-dure 1 or the one attached to the cartridge as areplacement part.

C-ring

Marker holder

ink cartridge

Spring

C-ring

Page 111: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

101

MAINTENANCE Ink Refilling

4

4.4.2 Ink pad type (optional)When markings become faint, add some ink.

1 Remove the ink pad.Push the pad to the right and pull it out of the holder.

2 Add ink.Add some ink to the pad.

• If the refilled ink is too much, the periphery of the lens marker may be soiled. Be careful ofthe amount of the ink.

Page 112: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

102

MAINTENANCE Cleaning the Protective Glass

4.5 Cleaning the Protective Glass

When the “Dust detected. Please clean lens.” message appears at power-on, clean the protectiveglass. If dust settles on the protective glass, it may affect the measurement accuracy.

1 Remove the nosepiece.

2 Clean the protective glass.Remove any dust on the glass with a blowerbrush.

3 If any dust remains, wipe gently with a lens cleaning cloth.

CAUTION• Clean the protective glass under the nosepiece with a blower brush.The dust may include sharp particles. Wiping them without the blower brush damagesto the lens coating. Be sure to remove any dust on the protective glass with the blowerbrush.

• Take special care not to scratch the protective glass.Flaws on the glass substantially lower the reliability of measurement.

Nosepiece

Protectiveglass

Page 113: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

103

MAINTENANCE Adjusting the Lens Table

4

4.6 Adjusting the Lens Table

Adjust the feel of the lens table lever movement to be heavy or light.

1 Remove the lens table lever.Pull it out straight.

2 Loosen the setscrew on the adjustable screwwith a flatblade screwdriver.

3 Rotate the adjustable screw to adjust the feel ofthe lens table lever movement.

Rotate the screw without rotating the center axis.

4 Tighten the setscrew on the adjustable screw.

5 Attach the lens table lever and check the lever movement.Align the lens table lever attachment hole with the center axis and push it as far as it will go.

Repeat the adjustment until the lever movement feels proper.

Rotate the screw clockwise.

The lens table lever feels heavier.

Rotate the screw counterclockwise.

The lens table lever feels lighter.

• The effective rotation range of the adjustable screw is one rotation, either clockwise orcounterclockwise.

Even if the lever is rotated counterclockwise by more than one rotation, the lever does not becomelighter.

Moreover, continuing to rotate the screw counterclockwise (about five rotations or more) removesit. It may cause device malfunction.

Lens table lever

Adjustable screw

Setscrew

Lighter Heavier

Page 114: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

104

MAINTENANCE Cleaning

4.7 Cleaning

4.7.1 Cleaning the device exteriorWhen the cover or panel of the device becomes dirty, wipe with a soft cloth. For severe stains,immerse the cloth in a neutral detergent, wring well, and wipe. Finally wipe with a dry and soft cloth.

4.7.2 Cleaning the printer (LM-600P/LM-600PD only)Following an extended period of use, the paper feed on the printer auto-cutter becomes dirty due tocutting residue. If left as is, failure of the auto-cutter may occur. So perform cleaning of the printer.

1 Open the printer cover and remove the printerpaper roll.

See “4.3 Replacing Printer Paper (LM-600P/LM-600PD only)” (page 98).

2 Put the nozzle of an electric vacuum cleaner on the auto-cutter to vacuum the residue.Do not blow the residue. If may cause failure due to the residue becoming attached to theinternal operation mechanism.

3 Set the printer paper as it was.

CAUTION• Never use an organic solvent such as paint thinner.It may ruin the surface of the device.

• Lightly wipe the exterior of the LCD.Scratches or malfunction of the display may result.

• Never use a sponge or cloth soaked in water.The water may leak into the inside of the device and cause device malfunction.

Auto-cutter

Page 115: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

105

MAINTENANCE Cleaning

4

4.7.3 Cleaning the Eye Care card slotFollowing an extended period of use, grime may become adhered to the contacts of the Eye Carecard slot. If left as is, the data may not be read or written properly.In such a case, clean the contacts with a contact cleanerfor contact-type IC card reader/writer.

Use a commercial contact cleaner or an optionalone (part number 34086-7110).

When the contact part of the Eye Care card becomes dirty,wipe with a soft cloth. For persistent grime, immerse thecloth in a neutral detergent, wring well, and wipe. Finallywipe with a dry and soft cloth.

Eye Carecard slot

Contactcleaner

• Be sure to turn off the power of the device before contact cleaner cleaning.

• For the contact cleaner operating procedure, see a provided manual.

• The Eye Care card reader contacts are on the lower surface. Repeatedly insert the contactcleaner with the wiper part (cloth-attached surface) facing downward several times to cleanthe reader contacts.

Page 116: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

106

MAINTENANCE List of Replacement Parts

4.8 List of Replacement Parts

* Restock the spare consumables after replacing.

Part name Part number Note

Printer paper 31001-E031 Width 58 mm, Length 25 m

Ink cartridge (White) 31001-3373 White ink (3 packs), with C-ring

Ink cartridge (Red) 31001-3371 Red ink (3 packs), with C-ring

Ink cartridge (Blue) 31001-3372 Blue ink (3 packs), with C-ring

Ink for ink pad (White) 82001-IK004 Refill ink

Contact cleaner 34086-7110 For cleaning the Eye Care card slot

• When storing the ink cartridge, keep the marker pen upright with the marking point downward.

Otherwise, filled ink may be separated and hardly come out during marking.

Page 117: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

107

5

5. SPECIFICATIONS AND ACCESSORIES

5.1 Classifications

[Classification of LED product] Class 1The LM-600/LM-600P/LM-600PD (UV model only) is classified as a Class 1 LED Prod-uct.*1

The Class 1 LED Product that is safe during use, including long-term direct intrabeam view-ing, even when exposure occurs while using optical viewing instruments (eye loupes or bin-oculars).

VISIBLE AND INVISIBLE LED RADIATION.CLASS 1 LED PRODUCT. WAVELENGTH 365 nm, 523 nm.

[Classification under the provision of 93/42/EEC (MDD)] Class I

The LM-600/LM-600P/LM-600PD is classified as a Class I device.

[Form of protection against electrical shock] Class I

The LM-600/LM-600P/LM-600PD is classified as a Class I device.

A Class I is a device in which the protection against electric shock does not rely on basicinsulation only, but which includes an additional safety precaution in such a way that meansare provided for the connection of the device to the protective (ground) conductor in the fixedwiring of the installation in such a way that accessible metal parts cannot become live in theevent of a failure in the basic insulation.

[Degree of protection against liquid entry] IP20

The LM-600/LM-600P/LM-600PD is an ordinary device (enclosed system without protectionagainst ingress of liquids).

Avoid exposing water to the device.

[Degree of safety application in the presence of flammable anesthetics and/or flammable cleaningagents]

The LM-600/LM-600P/LM-600PD should be used in an environment where no flammableanesthetics and/or flammable cleaning agents are present.

Do not operate the device near flammable type materials.

[Method (s) of sterilization or disinfection recommended by the manufacturer]

The LM-600/LM-600P/LM-600PD does not have any part to be sterilized or be disinfected.

[Mode of operation]

The LM-600/LM-600P/LM-600PD is a continuous operation device.

[Mode of transport]

Stationary equipment

*1. In accordance with IEC60825-1:1993/A1:1997/A2:2001.

Page 118: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

108

SPECIFICATIONS AND ACCESSORIES Specifications

5.2 Specifications

� Measuring items

• Spherical power (lens for glasses)-25.00 to +25.00 D0.01/ 0.06/ 0.12/ 0.25 D increments[Accuracy]

The accuracy specifications are based on the results of test lens measured in accordance with ISO8598 (lensmeter).

• Spherical power (contact lenses)-25.00 to +25.00 D (BC=6.00 to 9.00)0.01/ 0.06/ 0.12/ 0.25 D increments

• Cylindrical power0.00 to ±10.00 D (-, +/-, +)0.01/ 0.06/ 0.12/ 0.25 D increments

• Axis 0° to 180° 1° increment

• ADD power 0.00 to +10.00 D (Add, Ad2)0.01/ 0.06/ 0.12/ 0.25 D increments

• Prism power 0.00 � to 20.00 �0.01/ 0.06/ 0.12/ 0.25� increments[Accuracy]

The accuracy specifications are based on the results of test lens measured in accordance with ISO8598 (lensmeter).

• Prism measurement modePolar coordinates (�, �)Rectangular coordinates (BASE IN/OUT, BASE UP/DOWN)

Measurable range (D) Accuracy (D)��0�-5

0��+5 ±0.06

� -5�-10

+5��+10 ±0.09

� -10�-15

+10��+15 ±0.12

� -15�-20

+15��+20 ±0.18

� -20 +20 ±0.25

Measurable range (�) Accuracy (�) 0��5 0.1

5��10 0.2

10��15 0.3

15��20 0.4

Page 119: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

109

SPECIFICATIONS AND ACCESSORIES Specifications

5

� Measuring time

� Measurement mode

� Applicable lenses

� Other functions

� External configuration

• PD measurement (LM-600PD only)20 mm to 49.5 mm (monocular PD)40 mm to 99 mm (binocular PD)

• UV transmittance (UV model only)0 to 100%: Transmittance with central wavelength 365 nm (UV-A)1/5% step

• Updating of display 0.09 second (Minimum time)

Continuous measurement of SPH, CYL, AXIS, ADD, and PRISM.The measured data is saved in the memory and fixed on the screen by pressing the read button.The fixed data is automatically released by setting a new lens on the nosepiece.

• Lens diameter Lenses for glasses:20 to 120 mm in diameter (LM-600/LM-600P) 20 to 100 mm in diameter (LM-600PD)

Contact lenses: Larger than the inner diameter of the nosepiece (5 mm in diameter)

• Measurable transmittance (SPH, CYL, AXIS, ADD, and PRISM)10% or more (0.00 to ±15.00 D)20% or more (±15.00 to ±25.00 D)

• Compensation function of high index lensesThe abbe number is changeable in the range of 20 to 60.

• Lens marker Ink cartridge type

• Display 4.7 inch color full graphic LCD320 × 240 dots with a backlight

• Printer (LM-600P/LM-600PD only)Thermal line printer with auto cutterWidth 58 mm

• Interface connectorsRS-232C: 1 portUSB: 1 port

• Dimensions 194 (W) mm × 210 (D) mm × 408 (H) mm7.64 (W) inch × 8.27 (D) inch × 16.07 (H) inch

• Mass LM-600: 3.6 kg (7.94 lbs)LM-600P: 3.9 kg (8.60 lbs)LM-600PD: 4.0kg (8.82 lbs)

Page 120: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

110

SPECIFICATIONS AND ACCESSORIES Specifications

� Environmental conditions (during use)

� Environmental conditions (during transport and storage)

� There are other functions as follows:

*1 Internal fuses are not user-serviceable parts.

• Power source AC 100 (±10%) to 240 (±10%) V, 50/60 Hz* Fuses (240 V, 4 A, two units) are included in the internal electrical power source.*1

• Power consumption 40 VA

• Temperature 10 to 35°C (50 to 95°F)

• Humidity 30 to 85%

• Pressure 800 to 1060 hPa

• Altitude Up to 1000 m at sea level

• Installation location Indoors

• Temperature -10 to 55°C (14 to 131°F)

• Humidity 10 to 85%

• Pressure 700 to 1060 hPa

• Installation categoryII (OVERVOLTAGE CATEGORIES)

• Pollution degree 2 (IEC60664)

• Service life 8 years from the date of initial operation* Proper maintenance is necessary.

• Packing unit 1 unit

Page 121: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

111

SPECIFICATIONS AND ACCESSORIES Standard Configuration

5

5.3 Standard Configuration

5.3.1 Standard accessories

5.3.2 Optional accessories

• Power cord 1 unit

• Dust cover 1 unit

• Nosepiece for contact lenses 1 unit

• Printer paper (LM-600P/LM-600PD only) 3 rolls

• Operator’s manual 1 volume

• Measuring Progressive Power lenses explanation sheet 1 sheet

• Eye Care card

• Interface cable (OPIF-6) *for RT connection (Din-8 to Din-8)

• Interface cable (OPIF-11) *for PC connection (Din-8 to Dsub-9)

• USB cable *with special USB driver

• Foot switch (OPFS-3)

• Ink cartridge (red)

• Ink cartridge (blue)

• Ink pad type marking unit

Page 122: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

112

SPECIFICATIONS AND ACCESSORIES Standard Configuration

Page 123: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

113

6

6. EMC (ELECTROMAGNETIC COMPATIBILITY)

The LM-600/LM-600P/LM-600PD complies with the International Electrotechnical Commission stan-dards (IEC 60601-1-2: 2001 + A1: 2004) for electromagnetic compatibility as listed in the tables below.Follow the guidance in the tables for use of the device in an electromagnetic environment.

EMC (IEC 60601-1-2: 2001 + A1: 2004)

*1 For the regions where the rated voltage is 220 V or greater, this device complies with class A. For the regions wherethe rated voltage is less than 127 V or less, this standard is not applicable.

*2 For the regions where the rated voltage is 220 V or greater, this device complies with this standard. For the regionswhere the rated voltage is less than 127 V or less, this standard is not applicable.

Guidance and manufacturer's declaration - electromagnetic emissions

The LM-600/LM-600P/LM-600PD is intended for use in the electromagnetic environment specified below. The customeror the user of the LM-600/LM-600P/LM-600PD should assure that it is used in such an environment.

Emissions test Compliance Electromagnetic environment - guidance

RF emissionsCISPR 11

Group 1 The LM-600/LM-600P/LM-600PD uses RF energy only for its internalfunction. Therefore, its RF emissions are very low and are not likely tocause any interference in nearby electronic equipment.

RF emissionsCISPR 11

Class B The LM-600/LM-600P/LM-600PD is suitable for use in allestablishments, including domestic establishments and those directlyconnected to the public low-voltage power supply network thatsupplies buildings used for domestic purposes.

Harmonic emissionsIEC 61000-3-2

*1

Voltage fluctuations/Flicker emissionsIEC 61000-3-3

*2

Page 124: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

114

EMC (ELECTROMAGNETIC COMPATIBILITY)

Guidance and manufacturer's declaration - electromagnetic immunity

The LM-600/LM-600P/LM-600PD is intended for use in the electromagnetic environment specified below. The customer or the user of the LM-600/LM-600P/LM-600PD should assure that it is used in such an environment.

Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance

Electrostatic Discharge (ESD)IEC 61000-4-2

±6 kV contact±8 kV air

±6 kV contact±8 kV air

Floor should be wood, concrete or ceramic tile.If floors are covered with synthetic material, therelative humidity should be at least 30%.

Electrical fast transient/burstIEC 61000-4-4

±2 kVfor power supply lines±1 kVfor input/output lines

±2 kVfor power supply lines±1 kVfor input/output lines

Mains power quality should be that of a typicalcommercial or hospital environment.

SurgeIEC 61000-4-5

±1 kVdifferential mode±2 kVcommon mode

±1 kVdifferential mode±2 kVcommon mode

Mains power quality should be that of a typicalcommercial or hospital environment.

Voltage, dips, short interruptions and voltage variations on power supply input linesIEC 61000-4-11

<5% UT

(>95% dip in UT)for 0.5 cycle40% UT

(60% dip in UT)for 5 cycles70% UT

(30% dip in UT)for 25 cycles< 5% UT

(> 95% dip in UT)for 5 sec

<5% UT

(> 95% dip in UT)for 0.5 cycle40% UT

(60% dip in UT)for 5 cycles70% UT

(30% dip in UT)for 25 cycles < 5% UT

(> 95% dip in UT)for 5 sec

Mains power quality should be that of a typicalcommercial or hospital environment. If the userof the LM-600/LM-600P/LM-600PD requirescontinued operation during power mainsinterruptions, it is recommended that the LM-600/LM-600P/LM-600PD be powered from anuninterruptible power supply or a battery.

Power frequency (50/60 Hz) magnetic fieldIEC 61000-4-8

3 A/m 3 A/m Power frequency magnetic fields should be atlevels characteristic of a typical location in atypical commercial or hospital environment.

NOTE UT is the a.c. mains voltage prior to application of the test level.

Page 125: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

115

EMC (ELECTROMAGNETIC COMPATIBILITY)

6

Guidance and manufacturer's declaration - electromagnetic immunity

The LM-600/LM-600P/LM-600PD is intended for use in the electromagnetic environment specified below. The customeror the user of the LM-600/LM-600P/LM-600PD should assure that it is used in such an environment.

Immunity test IEC 60601 test level Comliance level Electromagnetic environment - guidance

Conducted RF IEC 61000-4-6

Radiated RF IEC 61000-4-3

3 Vrms150 kHz to 80 MHz

3 V/m80 MHz to 2.5 GHz

3 Vrms(V1=3)

3 V/m(E1=3)

Portable and mobile RF communications equipment should be used no closer to any part of the LM-600/LM-600P/LM-600PD, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended separation distanced=1.2 P 150 kHz to 80 MHz

d=1.2 P 80 MHz to 800 MHzd=2.3 P 800 MHz to 2.5 GHzwhere P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).Field strengths from fixed RF transmitters, as

determined by an electromagnetic site survey,a should be less than the compliance level in

each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobileradios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should beconsidered. If the measured field strength in the location in which the LM-600/LM-600P/LM-600PD is used exceedsthe applicable RF compliance level above, the LM-600/LM-600P/LM-600PD should be observed to verify normaloperation. If abnormal performance is observed, additional measures may be necessary, such as reorienting orrelocating the LM-600/LM-600P/LM-600PD.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Page 126: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

116

EMC (ELECTROMAGNETIC COMPATIBILITY)

Recommended separation distances between portable and mobile RF communications equipment and the LM-600/LM-600P/LM-600PD

The LM-600/LM-600P/LM-600PD is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the LM-600/LM-600P/LM-600PD can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the LM-600/LM-600P/LM-600PD as recommended below, according to the maximum output power of the communications equipment.

Rated maximum output power of transmitter

W

Separation distance according to frequency of transmitterm

150 kHz to 80 MHzd=1.2 P

80 MHz to 800 MHzd=1.2 P

800 MHz to 2.5 GHzd=2.3 P

0.01 0.12 0.12 0.23

0.1 0.38 0.38 0.73

1 1.2 1.2 2.3

10 3.8 3.8 7.3

100 12 12 23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Page 127: AUTO LENSMETER l LM-600/600P/600PD - revolutionehr.com · V After use Maintenance and check Disposal CAUTION† While the device is not in use, turn it off. † The LM-600P/LM-600PD

7

117

7. INDEX

A

Alignment circle •••••••••••••••••••••••••••••••••••••••••••••••••• 8Auto read •••••••••••••••••••••••••••••••••••••••••••••••••••••••••• 7AXIS bar ••••••••••••••••••••••••••••••••••••••••••••••••••••••••••• 8

B

Back page button •••••••••••••••••••••••••••••••••••••••••••••• 19

C

Clear button ••••••••••••••••••••••••••••••••••••••••••••••••••••••• 9Compare measurement button ••••••••••••••••••••••••••••• 18Contrast knob •••••••••••••••••••••••••••••••••••••••••••••••••••• 3Coordinate system button •••••••••••••••••••••••••••••••••••• 14CYL mode ••••••••••••••••••••••••••••••••••••••••••••••••••••••• 37Cylinder +/- button ••••••••••••••••••••••••••••••••••••••••••••••• 9

D

Display ••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••• 3Display auto off mode ••••••••••••••••••••••••••••••••••••••••••• 3

E

Exit button ••••••••••••••••••••••••••••••••••••••••••••••••••••••• 19

F

Fixing the measured data •••••••••••••••••••••••••••••••••••• 36Forward page button •••••••••••••••••••••••••••••••••••••••••• 19Function buttons ••••••••••••••••••••••••••••••••••••••••••••••••• 3

I

Indicator for near portion ••••••••••••••••••••••••••••••••••••• 46Initial screen •••••••••••••••••••••••••••••••••••••••••••••••••••• 26Interface connector •••••••••••••••••••••••••••••••••••••••••••••• 5Intl. screen •••••••••••••••••••••••••••••••••••••••••••••••••••••• 27

L

Lens holder lever •••••••••••••••••••••••••••••••••••••••••••••••• 3Lens table ••••••••••••••••••••••••••••••••••••••••••••••••••••••••• 4Lens table lever •••••••••••••••••••••••••••••••••••••••••••••••••• 4

M

Marking lever ••••••••••••••••••••••••••••••••••••••••••••••••••••• 3Measurement mode indication •••••••••••••••••••••••••••••••• 8

Measurement mode screen •••••••••••••••••••••••••••••••••• 16Measurement selection button ••••••••••••••••••••••••••••••••8

N

Next screen button ••••••••••••••••••••••••••••••••••••••••••••• 14Nose slider •••••••••••••••••••••••••••••••••••••••••••••••••••••• 32Nosepiece •••••••••••••••••••••••••••••••••••••••••••••••••••••••••3

P

Parameter settings ••••••••••••••••••••••••••••••••••••••••••••• 19Parameters •••••••••••••••••••••••••••••••••••••••••••••••••••••• 19PD ••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••• 11PD measurement •••••••••••••••••••••••••••••••••••••••••••••• 32Pilot lamp ••••••••••••••••••••••••••••••••••••••••••••••••••••••••••3Power inlet ••••••••••••••••••••••••••••••••••••••••••••••••••••••••5Power switch ••••••••••••••••••••••••••••••••••••••••••••••••••••••5Preset prism prescription ••••••••••••••••••••••••••••••••••••• 14Previous screen button ••••••••••••••••••••••••••••••••••••••• 16Print button •••••••••••••••••••••••••••••••••••••••••••••••••••••• 10Prism data button •••••••••••••••••••••••••••••••••••••••••••••• 14Prism entry screen ••••••••••••••••••••••••••••••••••••••••••••• 13Prism layout function •••••••••••••••••••••••••••••••••••••••••• 13Prism layout indication •••••••••••••••••••••••••••••••••••••••• 13

R

R/L indication •••••••••••••••••••••••••••••••••••••••••••••••••••••6R/L selection button •••••••••••••••••••••••••••••••••••••••••••••9Read button •••••••••••••••••••••••••••••••••••••••••••••••••••••••3Remeasurement button ••••••••••••••••••••••••••••••••••••••• 17

S

Selection button •••••••••••••••••••••••••••••••••••••••••••••••• 19Setting indications •••••••••••••••••••••••••••••••••••••••••••••••7

T

Target ••••••••••••••••••••••••••••••••••••••••••••••••••••••• 7 28

U

UP/DOWN button •••••••••••••••••••••••••••••••••••••••••••••• 15Up/down button ••••••••••••••••••••••••••••••••••••••••••••••••• 19USB port •••••••••••••••••••••••••••••••••••••••••••••••••••••••••••5UV 100% correction button •••••••••••••••••••••••••••••••••• 17UV measurement •••••••••••••••••••••••••••••••••••••••••••••• 57UV measurement screen ••••••••••••••••••••••••••••••17 58