Atomic Absorption. Where g = degeneracy. Lines are broadened by two effects:. Doppler Collisional Operating conditions for lamp are chosen so that the Doppler broadening in the lamp (low P, few collisions) is less than the Doppler and collisional broadening in the flame or furnace. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Atomic Absorption
Page 2: Atomic Absorption
Page 3: Atomic Absorption
Page 4: Atomic Absorption
Page 5: Atomic Absorption

kT

E

eg

g

N

N

0

*

0

*

Where g = degeneracy

Page 6: Atomic Absorption
Page 7: Atomic Absorption
Page 8: Atomic Absorption
Page 9: Atomic Absorption
Page 10: Atomic Absorption

Lines are broadened by two effects:

• Doppler

• Collisional

• Operating conditions for lamp are chosen so that the Doppler broadening in the lamp (low P, few collisions) is less than the Doppler and collisional broadening in the flame or furnace.

Page 11: Atomic Absorption

Atomic Absorption

Page 12: Atomic Absorption

Ideally, we want the selected

emission line from the lamp to be

narrower than the absorption spectrum

Monochromator is used to select one of the emitted lines

Page 13: Atomic Absorption
Page 14: Atomic Absorption

Absorption lines are narrow

• The selected bandwidth of light from source must be narrower than chosen absorption line

• If a monochromator was able to select a narrow enough bandwidth from the output of a deuterium lamp, the power of the light would be negligible

• Therefore lamps that emit spectral lines are used

Page 15: Atomic Absorption

Hollow Cathode Lamp

Page 16: Atomic Absorption
Page 17: Atomic Absorption

HCL

• Inert gas is ionized by discharge

• Is accelerated to cathode

• Causes some element to dislodge and form atomic cloud (sputtering)

• Some are excited (in collisions with ions) and emit line spectra.

• Usually lamps are for one element – but can be for as many as six.

Page 18: Atomic Absorption

Electrodeless Discharge Lamp

Page 19: Atomic Absorption

Electrodeless Discharge Lamp• Microwave excited discharge tubes• Intensities 10-100 x greater than from HCL• Small amount of element or halide of an element

in a small sealed tube containing a few torr of inert gas

• Placed in microwave cavity (2450 MHz)• Argon is ionized, the ions are accelerated and

excite the metal atoms• Less stable than HCL, but more intense.• Not available for all elements

Page 20: Atomic Absorption

High-resolution continuum source AAS:the better way to perform atomic absorption

spectrometry

Page 21: Atomic Absorption

• Single xenon arc lamp

• Today, multiple hollow cathode lamps are no longer used.

• With the use of a single xenon arc lamp, all the elements can be measured from 185-900 nm.

• This takes AAS into a true multi-element technique with the analysis of 10 elements per minute.

Page 22: Atomic Absorption

• CCD technology - For the first time in an AAS CCD chips are now available with 200 pixels which act as independent detectors.

• Simultaneous background correction - Background is now measured simultaneously compared to sequential background on conventional AAS.

• Better detection limits - Due to the high intensity of the Xenon Lamp there is better signal/noise ratio thus giving better detection limits. In some cases it is up to 10 times better than conventional AAS.

Page 23: Atomic Absorption
Page 24: Atomic Absorption

Sample

• There are a variety of different sampling methods:

• Flame• Furnace (electrothermal atomizer)• Arc, spark• ICP• Cold vapour atomization• Hydride generation

Page 25: Atomic Absorption
Page 26: Atomic Absorption

Flame

• Stable

• Safe

• Cheap to maintain

• High temperature

• Reducing Atmosphere - many metals form stable oxides, not easily atomized just by flame temperatures

Page 27: Atomic Absorption

Flame

• Typical system – spray chamber and burner

• Sample is aspirated into spray chamber using nebulizer (sucked in by Venturi effect)

• Produces aerosol.

• Aerosol strikes obstruction – spoiler – to break it into smaller drops

• Only smallest drops proceed to flame

• Larger drops go down drain

Page 28: Atomic Absorption
Page 29: Atomic Absorption

Sequence of Events in Flame• Evaporation of Solvent (leaving fine salt

particles suspended in flame)

• Loss of water of hydration

• Vaporization of solid particles to free atoms (due to heat and chemical reaction)

• Excitation

• Ionization (not always desirable)

Page 30: Atomic Absorption

Nebulization• Controls fraction of sample to reach flame• Drop size is governed by viscosity, surface

tension, gas flow, density, design of nebulizer• Organic solvents have lower viscosity and lower

surface tension than water (0.25 - 0.3 x) They also allow preconcentration

• But change flame conditions – not always so beneficial

• Salt increases viscosity, decreasing efficiency• The smaller the drop, the more easily it is

desolvated and vaporized

Page 31: Atomic Absorption

Ultrasonic breakup of drops

• High frequency vibrations

• Uniform and controllable drop size

• BUT

• Drops are larger and equipment more expensive

Page 32: Atomic Absorption

Desolvation

• Critical to number of free atoms

• Usually occurs at base of flame

• Solvent then water of crystallization

• Depends on droplet size and solvent

Page 33: Atomic Absorption

Vaporization• Atomization to free atoms• Ideally – want high temperature and long

residence time (slow burn rate of the gases) - lots of time for atomization

• Depends on nature of molecules and atoms• Al2O3 atomizes more slowly than NaCl

particle of the same size• Important if analyzing mixtures – different

conditions are needed for different atoms

Page 34: Atomic Absorption

Ionization

• Ions undergo different transitions than atoms

• Want one or the other

• Ions not desirable in flame method

• In ICP, ions are the desired species

• For alkali and alkaline earths, ions form above 2000K

Page 35: Atomic Absorption

Ionization• Increases• at low sample concentration• With increasing flame temperature• With decreasing ionization potential• Prevent by:• Low flame temperature• Excess of easily ionizable metal eg Li• Called a SUPPRESSOR• Eg: add lots of Li to solution to be

analyzed for K

Page 36: Atomic Absorption

Premix (Laminar flow) Burner

• Most common

• Gases premixed before entering burner

• Stable flame

• Use long narrow flames – long path length for light absorption

• Use at right angles for emission

• Small (narrow) flame keeps atom concentration high

Page 37: Atomic Absorption

Width of slot depends on gases

• Narrow slots prevent flame backing into mixing chamber and causing explosion

• But must allow enough gas through to support rate of burning

• Too narrow:

• Cooling by adjacent air

• Salt deposition clogs burner

• Use different burners for different gases

Page 38: Atomic Absorption

Gases

• Gas mixtures with high-burning velocities are less safe

• Also want long residence times

• C2H2-N2O (220cm/s) is better than C2H2-O2 (1130 cm/s). They have similar flame temperatures.

Page 39: Atomic Absorption
Page 40: Atomic Absorption

Air-C2H2

Yellow

Colorless

Blue

C2H2-N2O

Red (CN., NH.)

Whitish-blue

Page 41: Atomic Absorption
Page 42: Atomic Absorption

Flame AtomizerAdvantages

• Convenient

• Rapid

• Suitable for all AA-determinable elements

Limitations

• Limited Sensitivity

• Large Sample Volume

• Cannot handle some sample types

Page 43: Atomic Absorption

Sensitivity Limitation

Page 44: Atomic Absorption

Interferences

• Spectral

• Vaporization

• Chemical

Page 45: Atomic Absorption

Spectral

• Mg 285.21 nm

• Na 285.28 nm

• Not usually much of a problem – can change to another wavelength

• Problem worse in emission because more lines – High T – lots of excitation

• Choice of line dictates concentration range

able to be analyzed

Page 46: Atomic Absorption
Page 47: Atomic Absorption

Vaporization Interferences

• When one component of a sample influences the rate of vaporization of the species of interest

• Physical – changes matrix it vaporizes from

• Chemical – changes the species to be vaporized

Page 48: Atomic Absorption

Chemical Vaporization Interferences

• Metal oxides form

• Metal ions form thermally stable complexes with anions

• The effects usually occur during formation of the solid particle

• CaPO4 formation – a well known example.

• CaPO4 is harder to vaporize than Ca2+

Page 49: Atomic Absorption

CaPO4 - Interference Prevention

• Put light path higher in flame to allow a longer residence time

• Add releasing agent – La2+ or Sr2+ (added in excess) will preferentially combine with PO4

3- and leave Ca2+ free to be analyzed

• Protective agent – add EDTA. Ca-EDTA complex is easily destroyed in flame

• Glucose – burns easily and helps droplets shatter apart

• Hotter flame – then need ionization suppressor

Page 50: Atomic Absorption