Arterial Line Analysis
SAMIR EL ANSARY
ICU PROFESSOR
AIN SHAMS
CAIRO
![Page 2: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/2.jpg)
https://www.facebook.com/groups/1451610115129555/#!/groups/1451610115129555/
Wellcome in our new group ..... Dr.SAMIR EL ANSARY
No click
![Page 3: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/3.jpg)
How it works
PRESSURE TUBING
PRESSURE BAG 300 MMHG
![Page 4: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/4.jpg)
How it works
![Page 5: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/5.jpg)
• An arterial vessel is cannulated under aseptic conditions with a 20 guagecannula using seldinger (guideline) technique.
• The arterial catheter is connected to a 1000ml flush bag of NaCl and pressurised to 300mmHg (also required for KVO running @ 3ml/hr).
• At the mid point in the pressurised connecting tube there is a transducer point and this is connected directly to the patients monitoring device.
![Page 6: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/6.jpg)
• The pressure transducer converts the patients arterial blood pressure oscillations into an electrical waveform that is readable on the monitoring device.
• The resulting arterial pressure wave differs depending on the site of vascular cannulation i.e. radial, femoral, etc. this is due to several factors including – Fluid status– Vessel pathology– Cannulation quality (including thrombus, phlebitis
and/ or vasospasm)– Reflection waves throughout the arterial tree
(more evident in the more distal catheters).
![Page 7: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/7.jpg)
• The following slide shows the effects of wave reflection/deflection on the systolic and diastolic arterial blood pressure.
• Note that the mean arterial blood pressure is fairly constant.
• Wave reflection/deflection occurs as the blood passes through the arterial tree under pressure.
• If all of the vessels were straight, and had no branches, then the flow of blood would be direct and the pressure at each end would be the same.
![Page 8: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/8.jpg)
• However, as arteries are under constant pressure adjustment (musclular wall adjustment) and as they have bends and many branches, the flow of blood becomes less laminar – it becomes turbulent
This turbulance that causes
• the systolic pressure to rise and the diastolic pressure to fall slightly in the more
distal arteries.
![Page 9: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/9.jpg)
Arterial trace at different sites
BLOOD turbulance that causes
the systolic pressure to rise and the diastolic
pressure to fall slightly in the more distal arteries.
![Page 10: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/10.jpg)
Before we analyse the arterial waveform,
Always make sure that the gain on the monitor is correctly set!!!!
![Page 11: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/11.jpg)
Is it accurate?
• Failure to notice this may lead to unnecessary, or missed treatments
for our patients.
• There are 2 main abnormal tracing problems that can occur once the
monitor gain is set correctly.
![Page 12: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/12.jpg)
•Dampened trace
wide, slurred, flattened tracing
UNDERESTIMATES blood pressure.
![Page 13: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/13.jpg)
Dampened: wide, slurred, flattened tracing
• Dampening occurs due to:
– air bubles
– overly compliant, distensible tubing
– catheter kinks
– clots
– injection ports
– low flush bag pressure or no fluid in the flush bag
• This type of trace UNDERESTIMATES blood pressure.
![Page 14: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/14.jpg)
•Resonant trace ‘spiked’ tracing
OVERESTIMATES blood pressure
![Page 15: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/15.jpg)
Resonant: ‘spiked’ tracing
• Resonance occurs due to:
– long tubing
– overly stiff, non-compliant tubing
– increased vascular resistance
– reverberations in tubing causing harmonics that distort the trace (i.e. high systolic and low diastolic)
– non-fully opened stopcock valve
• This type of trace OVERESTIMATES blood pressure.
![Page 16: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/16.jpg)
Waveform Anatomy
![Page 17: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/17.jpg)
End diastole - systole • Anacrotic limb reaches from point (a) to point
(b)
point (y) is
known as the
anacrotic
notch
y
a
b
![Page 18: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/18.jpg)
Anacrotic limb
• The anacrotic limb represents the first phase of the arterial pulse cycle• It occurs as the ventricles eject the blood into the arterial tree and gives
a visual record of the arterial pressure rising to that of the end
systole.
![Page 19: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/19.jpg)
Anacrotic limb
• The steepness of the ascending phase can be affected by heart rate,
increased systemic vascular resistance, and through the use of
vasopressors such as noradrenaline(more steep incline)
• And vasodilators such as GTN (less steep incline).
![Page 20: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/20.jpg)
Anacrotic limb
• Myocardial contractility also effects the steepness of the anacrotic limb – during
impaired contractility (post MI for example) the up-sweep, or the rate of pressure increase can be prolonged.
• As the pressure reaches maximum, and the wave makes sharp turn to level off,
this is called the anacrotic notch
![Page 21: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/21.jpg)
• The black line represents the rate of pressure increase
![Page 22: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/22.jpg)
Anacrotic notch abnormality
• In some patients, a second systolic notch can be seen (as point (a) in the next slide).
• It is detected when aortic insufficiency exists in association with aortic stenosis, and may be found in hypertrophic obstructive cardiomyopathy.
• It may OVERESTIMATE systolic blood pressure slightly.
![Page 23: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/23.jpg)
Bisferiens tracing
a
![Page 24: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/24.jpg)
End systole – diastole
• Dicrotic limb reaches from point (b) to point (c)
b
c
![Page 25: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/25.jpg)
Dicrotic limb
• Descending limb of the arterial pressure trace as the pressure falls to that of the end diastolic pressure
• Dicrotic means ‘twice beating’ – meaning that this phase of the arterial pressure pulse should have a second, smaller wave, known as the dicrotic notch
![Page 26: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/26.jpg)
• Dicrotic notch is point (x)
x
![Page 27: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/27.jpg)
Dicrotic notch
• Can occur at any point that there is a fluctuation in pressure during the descending arterial limb.
• The most common time for this to occur is when the aortic and pulmonary valves snap shut causing pressure reverberations through the arterial
![Page 28: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/28.jpg)
Dicrotic pressure changes
![Page 29: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/29.jpg)
Dicrotic pressure changes
this part here is the dicrotic notch
![Page 30: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/30.jpg)
Dicrotic notch – continued..
• Flat or non-existent notch can mean that the patient is dehydrated (line trace will also ‘swing’)
• Low notch can also mean high pulse pressure (due to the low diastole in septic shock for example)
• Flattened notch can be present in cardiopulmonary valve insufficiency.
![Page 31: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/31.jpg)
Dicrotic limb: 2
• The rate of dicrotic ‘fall-off’, or the rate at which the arterial line trace falls from end-systole to early-diastole changes in relation to systemic vascular resistance.
• In patients with a severely reduced arteriolar resistance, fall-off time is rapid. This occurs as soon as end-systole finishes due to the greatly reduced pressure in the arterial tree (representing reduced afterload). The arterial waveform in this clinical state looks thin and pointed (don’t confuse this with resonance).
![Page 32: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/32.jpg)
Dicrotic limb: 2
• In patients with increased vascular resistance, such as main vessel stenosisfor example, the dicrotic fall-off time is greatly increased.
• This occurs due to the length of time it takes to return to end-diastolic pressure. The arterial waveform in this clinical state may be normal, or quite fat!
![Page 33: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/33.jpg)
Stroke volume variance
![Page 34: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/34.jpg)
Stroke volume variance
• The systolic blood pressure reading can vary from time to time – this is known as ‘arterial line swing’ and occurs more in dehydrated patients.
• This phenomenon occurs during respiration –both spontaneous and mechanically ventilated, although is more pronounced during volume controlled mechanical ventilation (SIMV).
![Page 35: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/35.jpg)
Stroke volume variance
• During spontaneous respiration, at the beginning of inspiration, the intrarthoracicpressure briefly drops before building up and becoming much higher than before due to the expansion of the lungs.
• This increased pressure causes a reduction in transmural blood flow back to the heart by compressing the intrarthoracic veins (reduced pre-load) causing stroke volume to drop.
![Page 36: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/36.jpg)
Stroke volume variance
• As expiration begins, after a brief period of further increased pressure as the muscles contract, the pressure drops significantly as the air leaves the lungs greatly increasing transmural blood flow (increased pre-load) causing stroke volume to rise.
• This can also be noted in the systolic blood pressure figure as it fluctuates with respiration.
![Page 37: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/37.jpg)
Stroke volume variance
• During positive pressure ventilation there is always a reduced transmural blood flow – this is why blood pressure is always lower during mechanical ventilation.
• Stroke volume variance is much more pronounced in relation to higher VTi and peak airways pressures.
• Of the next 2 slides, slide 1 represents changes during spontaneous respiration and slide 2 during mechanical ventilation.
![Page 38: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/38.jpg)
Arterial ‘swing’
• sv = stroke volume
![Page 39: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/39.jpg)
Ventilated arterial ‘swing’
• (A) = the expiratory phase – or the period where the only pressure in the chest is due to PEEP.
• (B) = the inspiratory phase, ASB + PEEP
![Page 40: Arterial line analysis](https://reader030.vdocuments.site/reader030/viewer/2022032616/55a74e621a28aba5638b46b8/html5/thumbnails/40.jpg)
Pulse contour analysis: swing
• The use of pulse contour analysis is available in the ICU and this can help in the selection of vasoactive drugs and/ or fluids.
• LiDCO also analyses arterial trace stroke volume variance and translates this into a visual percentage to determine if a patient will be pre-load responsive or not.
• LiDCO IS USELESS IF THE ARTERIAL LINE TRACE IS INACCURATE – please remember this.