Applications of CCD Array Spectrometers in Undergraduate

Education at EIU

Jonathan Blitz and Daniel Sheeran

Department of Chemistry

Page 2: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

The traditional educational spectrophotometer

Single beam

Single channel

Page 3: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Multichannel Spectrometer

Page 4: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Spectrophotometry – the mundaneMalachite Green Absorbance Spectrum from Multichannel Spectrometer

-0.05

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

0.45

400 450 500 550 600 650 700 750 800

Wavelength (nm)

Ab

sorb

ance Output from Spec-20:

Absorbance = 0.390

• Entire spectrum shows λmax and choice of analysis wavelength

• 2 spectrometers for lab of 24 students is sufficient

Page 5: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

The multichannel spectrometer is easily configured for emission spectrometry using a fiber optic probe for light input.

This has been used as a portable fluorescence emission spectrometer.

Page 6: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Fluorescence

A typical luminescence spectrometer Our luminescence spectrometer

Handheld UV Source

Multichannel Spectrometer

Emission Monochromator Mulitchannel Detector

Page 7: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Fluorescence Emission Spectrum

Page 8: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

0

500

1000

1500

2000

2500

3000

3500

4000

200 250 300 350 400 450 500 550 600 650

Wavelength (nm)

Em

iss

ion

Inte

ns

ity

0

0.5

1

1.5

2

2.5

3

Ab

so

rba

nc

e

B. Fluorescence emission spectrum of quinine

C. Absorbance spectrum of quinine

A. Hand held UV source

Absorbance and Fluorescence Emission Spectra of Quinine in

Water

Page 9: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Fluorescence Emission Spectra

0

500

1000

1500

2000

2500

3000

3500

4000

350 400 450 500 550 600 650 700 750

Wavelength (nm)

Em

issi

on

Inte

nsi

ty (

Co

un

ts)

Quinine

Rhodamine BFluorescein

Eosin Yellow

Page 10: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

The multichannel spectrometer has also been used as a detector for qualitative analysis in a traditional flame test …

Page 11: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Fiber Optic Collection of Na Flame Emission Spectrum

Page 12: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Flame Emission Spectra of the Alkalis

0

1000

2000

3000

4000

400 500 600 700 800 900 1000Wavelength (nm)

Em

iss

ion

Inte

ns

ity

Li

Na

K

Cs

Rb

Page 13: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Qualitative Analysis

“Group IV” Cations: Na+, K+, Ca2+, Mg2+, NH4+

Visual Flame Test:

• Na: persistent yellow

• K: transient red or lavender

• Ca: unreliable, transient red

Na masks K when both are present (Co blue filter)

False + and false – for Ca common with visual flame test

Page 14: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Qualitative Analysis

Spectrometer Flame Test:

•Na and K easily detected and distinguished

•Ca not detected, chemical test only

0

1000

2000

3000

4000

400 450 500 550 600 650 700 750 800 850

Wavelength, nm

Em

iss

ion

Inte

ns

ity

KNa

background Ca

Page 15: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Qualitative Analysis

Spectrometer Flame Test Student Results:

•Na > 95% correct on 1st try

•K > 95% correct on 1st try

•Ca > 98% correct on 1st try

Fewer false +, fewer false –

No blue filter for visual observation of K

Page 16: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Quantitative Analysis

• Emission intensity is transient and variable

• Use another alkali as an internal standard

• Sample introduction with a nichrome wire does not provide adequate quantitative results

• Aspirate a sample using a continuous sprayer into the air intake of a Bunsen burner for improved sample introduction

Page 17: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Compressed Air

Spectrometer

Fiber optic

Page 18: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry
Page 19: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Na Calibration (Li Internal Standard)

R2 = 0.996

0

0.5

1

1.5

2

2.5

0 100 200 300 400 500 600

ppm Na

Na

/Li

Em

iss

ion

In

ten

sit

y R

ati

oK Calibration (Li Internal Standard)

R2 = 0.995

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

0 100 200 300 400 500 600

ppm K

K/L

i E

mis

sio

n I

nte

ns

ity

Ra

tio

Cs Calibration (Li Internal Standard)

R2 = 0.999

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

0 100 200 300 400 500 600

ppm Cs

Cs

/Li

Em

iss

ion

In

ten

sit

y R

ati

o

Rb Calibration (Li Internal Standard)

R2 = 0.995

0

0.5

1

1.5

2

2.5

3

0 100 200 300 400 500 600

ppm Rb

Rb

/Li

Em

iss

ion

In

ten

sit

y R

ati

o

Li Calibration (K Internal Standard)

R2 = 0.994

0

0.1

0.2

0.3

0.4

0.5

0.6

0 50 100 150 200 250 300 350 400 450

ppm Li

Li/

K E

mis

sio

n I

nte

ns

ity

Ra

tio

Page 20: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

K unknown = 200.0 ppm Measured = 202.5 ppm

Student results not that good, but calibrations are OK

Detection limits using spray: 30-40 ppm for all alkalis

Detection limits much lower using nichrome wire

Page 21: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry

Conclusions

• A multichannel spectrometer is useful for teaching spectrophotometry

• Using a fiber optic probe the spectrometer was configured for the following tasks:

As a fluorescence emission spectrometer

For qualitative analysis of alkalis

For quantitative analysis of alkalis using an internal standard calibration